K Number
K200141
Device Name
Oxi-W System
Manufacturer
Date Cleared
2020-04-27

(97 days)

Product Code
Regulation Number
870.2300
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Oxi-W system is intended for use:

To spot check or monitor Oxygen saturation of arterial hemoglobin (SpO2%) and pulse rate.

To analyze the pulse waveform (Photoplethysmography or PTG) provided by the oximeter. It only provides mathematical analysis of the input of the PTG using the first and second derivatives of the PTG values related to the microvascular condition.

To analyze the basic rhythms of the NN or RR intert rate from the PTG, both in the time domain and in the frequency domain (short time 5 minutes). It only provides mathematical analysis of the heart rate to the autonomic nervous system function.

The system only provides mathematical analysis of input PTG values. It is practitioner to make proper judgement based on these values. The software provides a visual alarm for the heart rate and/or SpO2 percent out of the normal range and for the bad quality signal transmission.

The data are stored in PC in the Backup system of the device is intended use only for adult subjects (> 20 years old) This Oximeter is intended to be used in spot-checking (5 minutes).

The device is intended for use in licensed practitioner's office

This device is no intended to be used at home, in hospital or out-of-hospital transport

The device is not intended for use in support life and not for continuously monitoring

The system will be used by practitioner.

Device Description

The Oxi-W System is a programmable electro medical system (PEMS) including:

  • Pulse oximeter with data transmission via Bluetooth. -
  • -Software installed on a computer

Hardware :

  • collects SpO2%, pulse rate value and vertical bar graph pulse amplitude -
    (photoplethysmography) and sends the encrypted numbers to the Software.

Software:

  • Display SP02%, and photoplethysmography. -
  • Mathematical analysis of the pulse waveform (Photoplethysmography) using the first and second derivative of the wave to provide values related to the microvascular condition.
  • Mathematical analysis of the pulse waveform (Photoplethysmography) to detect the heart rate and provides values of the heart rate variability related to the autonomic nervous system function.
AI/ML Overview

The Oxi-W System, a programmable electromedical system consisting of a pulse oximeter with Bluetooth data transmission and computer software, was evaluated for performance and effectiveness.

Here's a breakdown of the acceptance criteria and the study that proves the device meets them:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (from predicate device K160956)Reported Device Performance (Oxi-W System)
Intended UseTo spot check or monitor Oxygen saturation of arterial hemoglobin (SpO2%) and pulse rate. To analyze the pulse waveform (Photoplethysmography or PPG) provided by the oximeter. It only provides mathematical analysis of the input of the PTG using the first and second derivatives of the PTG values related to the microvascular condition. To analyze the basic rhythms of the NN or RR intervals in heart rate from the PTG, both in the time domain and in the frequency domain (short time 5 minutes). It only provides mathematical analysis of the heart rate variability values related to the autonomic nervous system function. The system only provides mathematical analysis of input PTG values. It is the practitioner's responsibility to make proper judgement based on these values. The software provides a visual alarm for the values of the heart rate and/or SpO2 percent out of the normal range and for the bad quality signal transmission. The data are stored in PC in the Backup system of the Oxi-W software. The device is intended for use only for adult subjects (> 20 years old). This Oximeter is intended to be used in spot-checking (5 minutes). The device is intended for use in licensed practitioner's office. This device is not intended to be used at home, in hospital or out-of-hospital transport. The device is not intended for use in support life and not for continuously monitoring. The system will be used by practitioner.
SpO2 Accuracy70%100%: ±2%, 069%: unspecified
Pulse Rate (PR) Range30~235 bpm
PR Accuracy±3
IR light Wavelength890 nm ± 10 (Predicate: 880nm ± 10)
RED light Wavelength663 nm ± 10 (Predicate: 660nm ± 10)
Data Transmission Speed19200 Bauds
Software Measurements and ResultsSpO2%, Pulse rate, Photoplethysmography, Heart rate variability analysis at rest and using Ewing tests, Photoplethysmography Second derivative analysis.
Data Transmitted by HardwareSpO2%, Pulse rate, Photoplethysmography,
Standards Met60601-1 3rd Ed, 60601-1-2 3rd Ed, ISO 80601-2-61
Mode of Data ReceptionHardware Bluetooth microcontroller to PC
Power SupplyLithium battery 3.7 V rechargeable by USB port
Data TransmissionBluetooth
Module Oximeter Circuit Board and ComponentsOEM (K120502), Bluetooth microcontroller (Soc), Circuit board integrated to the probe

Study Proving Acceptance Criteria Met:

The Oxi-W system addresses differences from its predicate device (LD-Oxi system K160956) through a series of performance tests and analyses, rather than a clinical trial. The justification for not requiring clinical tests is that the Oxi-W algorithms and hardware share the same intended use and results as the predicate device, which itself did not require clinical tests.

Specifically, the following tests were conducted:

  • Laboratory tests: These included compliance with international standards such as IEC 60601-1, IEC 60601-1-2, and ISO 80601-2-61, addressing general safety, electromagnetic compatibility, and particular requirements for pulse oximeter equipment.
  • CRC (Cyclic redundancy check) Coding: This was performed to demonstrate the software's ability to accurately capture, store, and analyze data measured by the hardware.
  • Calibration tests: These were conducted using a simulator oximeter.
  • Software verification and validation (SRS/SDS/STD/STR): This comprehensive process confirmed the software's functionality and reliability, particularly for the new protocol communication required by the Bluetooth data transmission.
  • Bluetooth testing and cybersecurity risk analysis: These were conducted in accordance with FDA Wireless Guidance (2014), Cybersecurity Guidance (2013), and relevant Bluetooth white papers and FDA safety communications. This addressed the increased risk associated with Bluetooth data transmission and a different microcontroller.

Absence of Specific Information:

The provided text does not include the following information:

  • Sample size used for the test set and data provenance: No specific sample sizes for testing are mentioned beyond references to "laboratory tests" and "calibration tests." The origin of any test data (e.g., country of origin, retrospective/prospective) is also not provided.
  • Number of experts used to establish the ground truth for the test set and their qualifications: No information is given about experts or their qualifications for establishing ground truth.
  • Adjudication method for the test set: No adjudication method is described.
  • Multi-reader multi-case (MRMC) comparative effectiveness study: There is no indication that an MRMC study was performed. The document focuses on demonstrating substantial equivalence to a predicate device, not on comparing human reader performance with and without AI assistance.
  • Standalone (i.e., algorithm only without human-in-the-loop performance) study: While the device provides mathematical analysis, the evaluation is not explicitly described as a standalone study of the algorithm's performance without human interaction. The device is intended to provide values for practitioner judgment.
  • Type of ground truth used: The type of ground truth (e.g., expert consensus, pathology, outcomes data) used for the tests is not specified. Given the nature of the device (pulse oximetry and physiological signal analysis), it can be inferred that the ground truth for calibration and accuracy would likely come from established measurement standards and potentially comparison against clinical reference devices, but this is not explicitly stated.
  • Sample size for the training set: There is no mention of a training set or its sample size, as the document focuses on regulatory submission and performance testing rather than algorithm development specifics.
  • How the ground truth for the training set was established: As no training set is mentioned, this information is not provided.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

LD Technology LLC Albert MAAREK Quality Manager 100 N. Biscayne Blvd Suite 502 Miami, Florida 33132

Re: K200141

Trade/Device Name: Oxi-W System Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (Including Cardiotachometer And Rate Alarm) Regulatory Class: Class II Product Code: MWI, DQA Dated: March 31, 2020 Received: April 1, 2020

Dear Albert MAAREK:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Todd Courtney Assistant Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number

Device Name

Oxi-W system

Indications for Use (Describe)

Oxi-W system is intended for use:

To spot check or monitor Oxygen saturation of arterial hemoglobin (SpO2%) and pulse rate.

To analyze the pulse waveform (Photoplethysmography or PTG) provided by the oximeter. It only provides mathematical analysis of the input of the PTG using the first and second derivatives of the PTG values related to the microvascular condition.

To analyze the basic rhythms of the NN or RR intert rate from the PTG, both in the time domain and in the frequency domain (short time 5 minutes). It only provides mathematical analysis of the heart rate to the autonomic nervous system function.

The system only provides mathematical analysis of input PTG values. It is practitioner to make proper judgement based on these values. The software provides a visual alarm for the heart rate and/or SpO2 percent out of the normal range and for the bad quality signal transmission.

The data are stored in PC in the Backup system of the device is intended use only for adult subjects (> 20 years old) This Oximeter is intended to be used in spot-checking (5 minutes).

The device is intended for use in licensed practitioner's office

This device is no intended to be used at home, in hospital or out-of-hospital transport

The device is not intended for use in support life and not for continuously monitoring

The system will be used by practitioner.

Type of Use (Select one or both, as applicable)

ال Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW! *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/2 description: The image shows the logo for LD Technology. The logo consists of a green plant inside of a circle on the left, and the words "LD Technology" in a light green color on the right. The text is in a serif font and is slightly italicized.

510(k) Premarket Notification Number: date: April,17,2019

510(k) Summary K200141

Oxi-W system

This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of 21 CFR §807.92

1. Submitter's Identification:

Manufacturer: L.D TECHNOLOGY
Address: 100 N. Biscayne Blvd, Suite 502 Miami, FL, 33132, USA
Tel: 305-379-9900
Email: albert.ldteck@gmail.com

2. Device Name / Classification

Trade name: Oxi_W system
Device Common Name: Cardiac Monitor
Regulation Number: 21 CFR 870.2300
Product Codes: MWI, DQA
Classification: Class II
Classification Panel: Anesthesiology

3. Predicate legally marketed devices

Primary Predicate Device: LD-Oxi system cleared under K160956
Applicant: LD Technology LLC .
Product code s: MWI and DQA
Reference Device:PC-60 NW Fingertip Oximeter cleared under K120502
Applicant: Shenzhen creative industry Co .Ltd
Product code: DQA

4. Device Description

The Oxi-W System is a programmable electro medical system (PEMS) including:

  • Pulse oximeter with data transmission via Bluetooth. -
  • -Software installed on a computer

{4}------------------------------------------------

Description of the features

Hardware :

  • collects SpO2%, pulse rate value and vertical bar graph pulse amplitude -
    (photoplethysmography) and sends the encrypted numbers to the Software.

Software:

  • Display SP02%, and photoplethysmography. -
  • Mathematical analysis of the pulse waveform (Photoplethysmography) using the first and second derivative of the wave to provide values related to the microvascular condition.
  • Mathematical analysis of the pulse waveform (Photoplethysmography) to detect the heart rate and provides values of the heart rate variability related to the autonomic nervous system function.

Software specifications

The system carries out the following operations:

Image /page/4/Figure/12 description: The image shows a block diagram of a system. The diagram includes blocks labeled "BACK UP", "Derivatives", "Photoelectrical plethysmography analysis", "Heart rate", "SpO2 %", "Pulse wave", "Heart rate variability analysis", "PROTOCOL COMMUNICATION", and "HARDWARE". Arrows connect the blocks, indicating the flow of information or processes within the system.

5. Intended use and indications for use

Oxi-W system is intended for use:

To spot check or monitor Oxygen saturation of arterial hemoglobin (SpO2%) and pulse rate. To analyze the pulse waveform (Photoplethysmography or PTG) provided by the oximeter. It only provides mathematical analysis of the input of the PTG using the first and second derivatives of the PTG values related to the microvascular condition.

To analyze the basic rhythms of the NN or RR intervals in heart rate from the PTG, both in the time domain and in the frequency domain (short time 5 minutes). It only provides mathematical analysis of the heart rate variability values related to the autonomic nervous system function. The system only provides mathematical analysis of input PTG values. It is practitioner to make proper judgement based on these values.

{5}------------------------------------------------

The software provides a visual alarm for the values of the heart rate and/or SpO2 percent out of the normal range and for the bad quality signal transmission.

The data are stored in PC in the Backup system of the Oxi-W software.

The device is intended for use only for adult subjects (> 20 years old)

This Oximeter is intended to be used in spot-checking (5 minutes).

The device is intended for use in licensed practitioner's office

This device is no intended to be used at home, in hospital or out-of-hospital transport The device is not intended for use in support life and not for continuously monitoring The system will be used by practitioner.

6. Device specifications and comparison with predicate device.

SpecificationsLD-OXI system K K160956Oxi-W systemComparison
Intended useTo spot check or monitor Oxygensaturation of arterial hemoglobin(SpO2%) and pulse rate.To analyze the pulse waveform(Photoelectrical Plethysmography orPP) provided by the oximeter. It onlyprovides mathematical analyses ofthe input of the SpO2 measurement.To analyze the basic rhythms of theNN or RR intervals in heart rate, bothin the time domain and in thefrequency domain (short time 5minutes). It only providesmathematical analysis of the input ofthe heart rate variability.The mathematical analysis ofPhotoelectrical Plethysmography andHRV ARE NOT intended use fordiagnosis.The software provides a visual alarmfor the values of the heart rate and/orSpO2 percent out of the normal rangeand for the bad quality signaltransmission.The data are stored in PC in theBackup system of the LD-Oxisoftware.To spot check or monitor Oxygensaturation of arterial hemoglobin(SpO2%) and pulse rate.To analyze the pulse waveform(Photoplethysmography or PPG)provided by the oximeter. It onlyprovides mathematical analysis of theinput of the PTG using the first andsecond derivatives of the PTG valuesrelated to the microvascularcondition.To analyze the basic rhythms of theNN or RR intervals in heart rate fromthe PTG, both in the time domain andin the frequency domain (short time 5minutes). It only providesmathematical analysis of the heartrate variability values related to theautonomic nervous system function.The system only providesmathematical analysis of the inputPTG values. It is the practitioner'sresponsibility to make properjudgement based on these values.The software provides a visual alarmfor the values of the heart rate and/orSpO2 percent out of the normal rangeand for the bad quality signaltransmission. The data are stored inPC in the Backup system of the Oxi-Wsoftware.Difference (1)

6.1 Intended use and indication for use comparison

{6}------------------------------------------------

Indication andPrescription foruseThe device is intended for use only foradult subjects (> 20 years old) ThisOximeter is intended to be used inspot-checking (5 minutes) Thedevice is intended for use in licensedpractitioner's officeThis device is no intended to be usedat home, in hospital or out-of-hospital transportThe device is not intended for use insupport life and not for continuouslymonitoringThe system will be use by practitioner.The device is intended for use only foradult subjects (> 20 years old)This Oximeter is intended to be usedin spot-checking (5 minutes).The device is intended for use inlicensed practitioner's officeThis device is no intended to be usedat home, in hospital or out-of-hospitaltransportThe device is not intended for use insupport life and not for continuouslymonitoringThe system will be used bypractitioner.Only grammatical corrections
-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

6.2 Software specification and comparison with predicate device

SpecificationsLD-OXI software K K160956Oxi-W softwareComparison
Softwaremeasurementsand ResultsSpO2%, Pulse rate,Photoplethysmography, Heart ratevariability analysis at rest and usingEwing tests, PhotoplethysmographySecond derivative analysis.SpO2%, Pulse rate,Photoplethysmography, Heart ratevariability analysis at rest and usingEwing tests, PhotoplethysmographySecond derivative analysisSame
Data transmittedby the hardwareSpO2%, Pulse rate,Photoplethysmography,SpO2%, Pulse rate,Photoplethysmography,Same
Mode of datareceptionUSB port of the PCHardware Bluetooth microcontrollerto PCDifference (2).

6.3 Hardware specifications and comparison with reference predicate device.

SpecificationsLD-OXI hardware K160956PC-60NW K120502Comparison
ScientificBackgroundBased on the red and infrared lightabsorption characteristics ofoxygenated and deoxygenatedhemoglobin.Based on the red and infrared lightabsorption characteristics ofoxygenated and deoxygenatedhemoglobin.Same
Material incontact with thepatientReusable SPo2 probe Thermoplasticpolyurethane (TPU) material and latexfreeReusable SPo2 probe Thermoplasticpolyurethane (TPU) material and latexfreeSame
Placement of theprobeIndex fingerIndex fingerSame
ClassificationClass IIClass IISame
Degree ofprotectionagainst electricshocksBFBFSame
Operating modeContinuous useContinuous useSame

{7}------------------------------------------------

FeaturesTo spot check or monitor Oxygensaturation of arterial hemoglobin,pulse rate and photoplethysmographyfor 5 minutes and transmission of thedata to a software for mathematicalanalysis.To spot check or monitor Oxygensaturation of arterial hemoglobin,pulse rate and photoplethysmographyfor 5 minutes and transmission of thedata to a software for mathematicalanalysis.Same
Power supplyLithium battery 3.7 V rechargeable byUSB port2 x LR03 (AAA) alkaline batteries 3VDifference (3)
IR lightWavelength880nm +/- 10890 nm +/- 10Same , the small different does notchange the reading according to thetolerance +/- 10
RED lightWavelength660nm +/- 10663 nm +/- 10Same , the small different does notchange the reading according to thetolerance +/- 10
Pulse rate (PR)Range30~235 bpm30~235 bpmSame
PR Accuracy+3+3Same
SpO2%accuracy70%100%:±2% , 069%, unspecified70%100%:±2% , 069%, unspecifiedSame
DatatransmissionUSBBluetoothDifference(4)
Standard met60601-1 3rd Ed60601-1-2 3rd EdISO 80601-2-6160601-1 3rd Ed60601-1-2 3rd EdISO 80601-2-61Same
Module oximeterCircuit board andcomponentsOEM (K090671)USB microcontrollerCircuit board integrated to the probeOEM (K120502)Bluetooth microcontroller (Soc)Circuit board integrated to the probeDifference (5) for themicrocontroller.
DatatransmissionSpeed19200 Bauds19200 BaudsSame

7. Substantial equivalence

Predicate legally marketed device

LD-Oxi system K K160956 Applicant: LD Technology LLC. Product code MWI, DQA

Similarities:

  • Has the same intended and indication for use ( I.e. details analysis of the intended use > writing),
  • Same technological characteristics as the predicate device

  • Same safety and effectiveness

  • Has the same material in contact with the patient

  • Do not affect the Fundamental Scientific Technology

  • Do not change the prescription use

Differences:

Difference (1) The new writing of the intended use is NOT a change or a new intended use, it is only to clarify the reading of the intended use and the utility of the device.

{8}------------------------------------------------

The sentence "To analyze the pulse waveform (Photoelectrical Plethysmography or PP) provided by the oximeter. It only provides mathematical analyses of the input of the SpO2 measurement".

Was replaced by :

To analyze the pulse waveform (Photoplethysmography or PPG) provided by the oximeter. It only provides mathematical analysis of the input of the PPG using the first and second derivatives of the PTG values related to the vascular microcirculation condition.

Reason of replacement:

  • a) The word Photoplethysmography or PPG is used in peer reviews regarding the technology and therefore, it is commonly understood by the user.
  • b) The sentence "mathematical analyses of the input of the SpO2 measurement" is not clear. In fact, the software performs a mathematical analysis of the pulse waveform (PPG) and not of the SPo2 measurement. The mathematical analysis provides the second derivative PPG and the software results come from the detected points ( a,b,c,d and e) of the second derivative PPG.

Therefore, the mathematical analysis will be better understood by the users.

  • c) Adding "related to the microvascular condition" is NOT a new intended use, it only clarifies the utility of the performed mathematical analysis of the second derivative PPG.
    The sentence "It only provides mathematical analysis of the input of the heart rate variability".

Was replaced by:

It only provides mathematical analysis of the heart rate variability values related to the autonomic nervous system function.

Reason of replacement:

  • Adding "related to the autonomic nervous system function" is NOT a new intended use: it only clarifies the utility of the performed mathematical analysis of the heart rate variability
    The sentence "The mathematical analysis of Photoelectrical Plethysmography and HRV ARE NOT intended use for diagnosis."

Was replaced by:

The system only provides values. It is the practitioner's responsibility to make proper judgements based on these values.

Reason of replacement:

The device provides values related to the autonomic nervous system function and microvascular condition. These values are not intended to provide diagnosis, but only allow the practitioner to judge how those values will be used in the practice.

Difference (2) Mode of Data reception ( Bluetooth vs USB)

Difference (3) Power supply 2 x LR03 (AAA) alkaline batteries 3V vs USB port

Difference (4) and (5) Data transmission Bluetooth versus USB using different microcontrollers.

{9}------------------------------------------------

8. Performances and Effectiveness General safety concerns.

Clinical Test:

Since the Oxi W algorithms and hardware have the same intended use and results as the predicate device, the Oxi-W does not require clinical tests since the predicate device does not provide any clinical Test.

Oxi_W testing:

  1. Laboratory tests

IEC 60601-1:2005 + CORR.1 (2006) + CORR.2 (2007) + AM1 (2012), Medical -Electrical Equipment -- Part 1: General Requirements For Basic Safety And Essential Performance.

IEC 60601-1-2: 2007, Medical Electrical Equipment - Part 1-2: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Compatibility -Requirements And Tests.

ISO 80601-2-61:2011 - Medical electrical equipment - Part 2-61: Particular requirements for basic safety and essential performance of pulse oximeter equipment. 2. CRC (Cyclic redundancy check) Coding was performing to demonstrate the software performance to accurately capture, store, and analyze the data measured by the hardware 3. Calibration tests (simulator oximeter)

  1. Software verification and validation(SRS/SDS/STD/STR)

  2. Bluetooth testing and cybersecurity risk analysis according to the FDA Wireless Guidance (2014), Cybersecurity Guidance (2013) and Design of Bluetooth version 4.0 white paper and FDA released a Safety Communication on March 3,2020.

Difference discussion:

Difference (1) is only to clarify the reading of the intended use and the utility of the device and does not affect the performances and effectiveness and General safety concerns of the proposed device.

Difference (2) requires Software New protocol communication.

The new protocol communication has been verified and validated by the software verification and validation(SRS/SDS/STD/STR) and CRC coding and therefore, does not affect the performances and effectiveness and General safety concerns of the proposed device

Difference (3) regarding the Power supply 2 x LR03 (AAA) alkaline batteries 3V vs USB port Will reduce the risk during the shipping and therefore, does not affect the performances and effectiveness and General safety concerns of the proposed device

Difference (4) and (5) Data transmission Bluetooth versus USB using different microcontrollers. may increase the risk of interference and cybersecurity. The risk of interference and cybersecurity has been tested and evaluated according to according to the FDA Wireless Guidance (2014), Cybersecurity Guidance (2013) and Design of Bluetooth version 4.0 white paper and FDA released a Safety Communication on March 3,2020 and the difference 4 and 5 do not affect the performances and effectiveness and General safety concerns of the proposed device

Therefore, the proposed device using a different cleared hardware than the predicate device does not affect its performance, safety, and effectiveness.

{10}------------------------------------------------

Conclusions

In conclusion, the proposed device, Oxi-W system, has the same intended use, and similar technological characteristics compared to the predicate device. The difference in technological characteristics are addressed by performance testing, and do not raise different questions of safety and effectiveness.

Signature:

No text is visible in this image.

Albert MAAREK

Premarket notification [510K] Number: K200141

§ 870.2300 Cardiac monitor (including cardiotachometer and rate alarm).

(a)
Identification. A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.(b)
Classification. Class II (performance standards).