(217 days)
The synthetic nitrile patient examination gloves, powder free, blue color and tested for use with chemotherapy drugs are disposable devices intended for medical purposes that is worn on the examiner's hands or fingers to prevent contamination between patient and examiner. The glove was tested for use with chemotherapy drugs as per ASTM D6978-05 (Reapproved 2019) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.
Synthetic Nitrile Patient Examination Gloves, Powder Free, Blue Color, and Tested for Use with Chemotherapy Drugs
This document describes the acceptance criteria and the study results for Synthetic Nitrile Patient Examination Gloves, Powder Free, Blue Color, and Tested for Use with Chemotherapy Drugs (K200093).
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria for these gloves are defined by their resistance to permeation by chemotherapy drugs, as outlined in ASTM D6978-05 (Reapproved 2019). The "acceptance criteria" here refer to the performance benchmark set by the test method. For most drugs, the goal is to achieve a breakthrough time greater than 240 minutes.
| Chemotherapy Drug and Concentration | Acceptance Criteria (Breakthrough Detection Time in Minutes) | Reported Device Performance (Breakthrough Detection Time in Minutes) |
|---|---|---|
| Blenoxane (Bleomycin), 15 mg/ml (15,000 ppm) | >240 (Implied by standard and other values) | >240 |
| Carmustine (BCNU), 3.3 mg/ml (3,300 ppm) | >240 (Implied by standard and other values) | 22.6 |
| Cyclophosphamide, 20 mg/ml (20,000 ppm) | >240 (Implied by standard and other values) | >240 |
| Cytarabine, 100 mg/ml (100,000 ppm) | >240 (Implied by standard and other values) | >240 |
| Etoposide (Toposar), 20 mg/ml (20,000 ppm) | >240 (Implied by standard and other values) | >240 |
| Fluorouracil (5 Flu), 50 mg/ml (50,000 ppm) | >240 (Implied by standard and other values) | >240 |
| Idarubicin, 1 mg/ml (1,000 ppm) | >240 (Implied by standard and other values) | >240 |
| Mesna, 100 mg/ml (100,000 ppm) | >240 (Implied by standard and other values) | >240 |
| Mitomycin C, 0.5 mg/ml (500 ppm) | >240 (Implied by standard and other values) | >240 |
| Paclitaxel, 6 mg/ml (6,000 ppm) | >240 (Implied by standard and other values) | >240 |
| Thiotepa, 10 mg/ml (10,000 ppm) | >240 (Implied by standard and other values) | 46.9 |
| Trisenox, 1 mg/ml (1,000 ppm) | >240 (Implied by standard and other values) | >240 |
| Vincristine Sulfate, 1 mg/ml (1,000 ppm) | >240 (Implied by standard and other values) | >240 |
Note on meeting criteria: The device meets the criteria for most drugs, achieving breakthrough times >240 minutes. However, it specifically does not meet the implied >240 minutes criteria for Carmustine (22.6 minutes) and has a significantly shorter breakthrough time for Thiotepa (46.9 minutes), leading to "CAUTION" and "WARNING" statements for these specific drugs.
The document does not provide information for the following points, as they are not relevant to the type of device (medical gloves) and the study conducted (material performance testing, not an AI or diagnostic device study). The study described is a laboratory performance test rather than a clinical study with human or image data.
2. Sample size used for the test set and the data provenance: Not applicable. This is a material permeation test for gloves, not a data-driven AI study. The "test set" would refer to the number of glove samples subjected to each drug test, which is not specified but would follow standard laboratory protocols for ASTM D6978-05. The data provenance would be from laboratory testing.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for material permeation is established by the physical testing procedure of ASTM D6978-05, not by expert consensus on data.
4. Adjudication method for the test set: Not applicable. The test results are objective measurements from the ASTM D6978-05 standard, not subjective interpretations requiring adjudication.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is not an AI or diagnostic device.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is not an algorithm or AI device.
7. The type of ground truth used: The "ground truth" for this study is defined by the objective measurements obtained through the standardized laboratory test method ASTM D6978-05 (Reapproved 2019) for assessing the permeation of medical gloves by chemotherapy drugs.
8. The sample size for the training set: Not applicable. This is not an AI or machine learning device and therefore does not have a "training set."
9. How the ground truth for the training set was established: Not applicable. As there is no training set for this type of device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 20, 2020
Anhui Intco Medical Products Co., Ltd. % Jason Ji Official Correspondent lntco Medical Industries, Inc. 805 Barrington Ave Ontario, California 91764
Re: K200093
Trade/Device Name: Synthetic Nitrile Patient Exam Gloves, Powder Free, Blue, Tested for Use w/Chemotherapy Drug Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA, LZC Dated: May 16, 2020 Received: Jun 02, 2020
Dear Jason Ji:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
CAPT Elizabeth Claverie, M.S. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K200093
Device Name
Synthetic Nitrile Patient Examination Gloves, Powder Free, Blue Color, and Tested for Use with Chemotherapy Drugs
Indications for Use (Describe)
The synthetic nitrile patient examination gloves, powder free, blue color and tested for use with chemotherapy drugs are disposable devices intended for medical purposes that is worn on the examiner's hands or fingers to prevent contamination between patient and examiner. The glove was tested for use with chemotherapy drugs as per ASTM D6978-05 (Reapproved 2019) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.
| Chemotherapy Drug and Concentration | Breakthrough Detection Timein Minutes |
|---|---|
| Blenoxane (Bleomycin), 15 mg/ml (15,000 ppm) | >240 |
| Carmustine (BCNU), 3.3 mg/ml (3,300 ppm) | 22.6 |
| Cyclophosphamide, 20 mg/ml (20,000 ppm) | >240 |
| Cytarabine, 100 mg/ml (100,000 ppm) | >240 |
| Etoposide (Toposar), 20 mg/ml (20,000 ppm) | >240 |
| Fluorouracil (5 Flu), 50 mg/ml (50,000 ppm) | >240 |
| Idarubicin, 1 mg/ml (1,000 ppm) | >240 |
| Mesna, 100 mg/ml (100,000 ppm) | >240 |
| Mitomycin C, 0.5 mg/ml (500 ppm) | >240 |
| Paclitaxel, 6 mg/ml (6,000 ppm) | >240 |
| Thiotepa, 10 mg/ml (10,000 ppm) | 46.9 |
| Trisenox, 1 mg/ml (1,000 ppm) | >240 |
| Vincristine Sulfate, 1 mg/ml (1,000 ppm) | >240 |
- CAUTION: Testing showed an average breakthrough time of 46.9 minutes with Thiotepa
and 22.6 minutes with Carmustine
- WARNING: Do not use with Carmustine.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions. search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
§ 880.6250 Non-powdered patient examination glove.
(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.