(307 days)
The Eclipse DermaFlex Cannula is intended to inject fluids intradermally.
The Eclipse DermaFlex Cannula is provided as a single-use, sterile device. It is comprised of a needle. hub and cap. This device comes in a variety of needle gauges and lengths. The Eclipse DermaFlex Cannula offers AN type and B Type. The AN type is a hypodermic single lumen needle intended to prepare the site for injection. The tip of this pilot needle is sharpened at one end, while the other end is joined to a hub. The B type has a metal tube with the tip, which is closed and blunt, while the cannula has an opening laterally at a lower point under the tip. On the other end, the device is joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or secondary medication sets to prepare and administer fluids/ medications/ drugs to a patient.
The Eclipse DermaFlex Cannula is a hypodermic single lumen needle intended to inject fluids intradermally. The submission demonstrates substantial equivalence to predicate device MV INTRADERMIC NEEDLES; MAGIC NEEDLE (K110606) by M.V. S.R.L.
Here's an analysis of the provided information regarding acceptance criteria and the study that proves the device meets them:
1. A table of acceptance criteria and the reported device performance
The document does not explicitly state acceptance criteria in a quantifiable manner for each test. Instead, it refers to performance testing conducted in accordance with various ISO and ASTM standards. The reported device performance is a general statement that the device "meets the established characteristics and performance requirements needed to perform its intended function." It also states, "Substantial equivalence is established as these are the same standards established for the predicate device as well. As the Eclipse meets the required criteria of the standards, its performance may be considered to support substantial equivalence."
Therefore, the table below reflects what can be inferred from the document:
| Test Category | Acceptance Criteria (Inferred from standards) | Reported Device Performance |
|---|---|---|
| Sterilization Validation | In accordance with ISO 11135, ISO 11138-1, ISO 11138-2, ISO 11737-1, ISO 10993-7, AAMI TIR28 | Meets established characteristics and performance requirements. |
| Shelf Life Validation | In accordance with ISO 11608-2, ASTM F1929, ISO 11737-2 | Meets established characteristics and performance requirements. |
| Biocompatibility Testing | In accordance with ISO 10993-1, ISO 10993-10, ISO 10993-4 | Meets established characteristics and performance requirements. |
| Performance Tests | In accordance with ISO 9626:2016, ISO 7864:2016, ISO 80369-7, ISO 6009:2016 | Meets established characteristics and performance requirements. |
| Overall Equivalence | Demonstrates substantial equivalence to predicate device (K110606) | Established through meeting required criteria of the referenced standards. |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document does not provide details on the specific sample sizes used for each test or the data provenance (country of origin, retrospective/prospective). It only lists the standards followed for the tests.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This information is not applicable to this type of device (hypodermic needle). The ground truth for performance testing of a medical device like this is established through adherence to recognized international and national standards, not by expert consensus on clinical findings.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This information is not applicable as the evaluation of this device is based on objective performance testing against established standards, not interpretation of data by multiple readers requiring adjudication.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
An MRMC comparative effectiveness study is not applicable to this device. This is a physical medical device (cannula/needle), not an AI-powered diagnostic or assistive tool.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
A standalone algorithm performance study is not applicable as this is a physical medical device, not an algorithm or AI system.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
The ground truth for this device's performance is based on established national and international standards for medical device manufacturing, sterility, biocompatibility, and functional performance (e.g., needle strength, connection integrity, color coding). These standards define objective, measurable parameters.
8. The sample size for the training set
This information is not applicable as this is a physical medical device and does not involve a training set for machine learning or AI.
9. How the ground truth for the training set was established
This information is not applicable as there is no training set for this type of device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in large, bold, blue letters, and the word "ADMINISTRATION" in smaller, blue letters below.
November 5, 2020
Eclipse MedCorp, LLC Julie Summerville Senior Director of Product Management 5916 Stone Creek Drive Suite#120 The Colony, Texas 75056
Re: K200017
Trade/Device Name: Eclipse DermaFlex Cannula Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: Class II Product Code: FMI Dated: October 1, 2020 Received: October 5, 2020
Dear Julie Summerville:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-
combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For CAPT Alan Stevens Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
510(k) Number (if known)
K200017
Device Name Eclipse DermaFlex Cannula
Indications for Use (Describe)
The Eclipse DermaFlex Cannula is intended to inject fluids intradermally.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (7/17)
PSC Publishing Services (301) 443-6740
EF
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
{3}------------------------------------------------
510(K) Summary-K200017 Prepared: June 30, 2020
1. 510k Submitter
Eclipse MedCorp, LLC. 5916 Stone Creek Dr. Suite #120 The Colony, TX 75056 U.S.A. Tel: (972) 380-2911 Fax: (866) 558-0415
2. Contact Person
Julie Summerville Sr. Dir. of Product Management Eclipse MedCorp, LLC 972.380.2911 x2405 jsummerville@eclipsemed.com
3. Device Name:
- · Trade Name : Eclipse DermaFlex Cannula
- · Requlation Description: Hypodermic single lumen needle
- · Device Class: Class II
- Product Code: FMI
- · Requlation Number: 21 CFR Part 880.5570
- · Review Panel: General Hospital
4. Predicate Device:
MV INTRADERMIC NEEDLES; MAGIC NEEDLE (K110606) by M.V. S.R.L.
5. Device Description:
The Eclipse DermaFlex Cannula is provided as a single-use, sterile device. It is comprised of a needle. hub and cap. This device comes in a variety of needle gauges and lengths. The Eclipse DermaFlex Cannula offers AN type and B Type. The AN type is a hypodermic single lumen needle intended to prepare the site for injection. The tip of this pilot needle is sharpened at one end, while the other end is joined to a hub. The B type has a metal tube with the tip, which is closed and blunt, while the cannula has an opening laterally at a lower point under the tip. On the other end, the device is joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or secondary medication sets to prepare and administer fluids/ medications/ drugs to a patient.
Indication: 6.
The Eclipse DermaFlex Cannula is intended to inject fluids intradermally.
7. Technological Characteristics:
The subject device has the same indications for use, principle of operation and structure. The subject device is comprised of the same materials as the predicate, offers the same range of needle gauges and has the same syringe connection type. The needle tube length of the subject device, DermaFlex Cannula is within the size range of the predicate device. The subject device is marketed with a blunt tip needle and pilot needle. The predicate device is marketed with blunt tip needle and with or without a hypodermic needle (pilot needle).; however, both perform the same intended use. Additionally, there are slight differences in the design of the needle and cannula tips between the subject and predicate devices. The inclusion of the hypodermic needle and the slight differences in design are not considered to raise new questions of safety and effectiveness. Further, successful testing of the subject device to appropriate performance testing has been performed and these differences does not raise new questions with regard to safety and effectiveness.
{4}------------------------------------------------
| Table of Technological Characteristics | |||
|---|---|---|---|
| Subject Device | Predicate Device | Comments | |
| Device Name | Eclipse DermaFlex Cannula | MV INTRADERMIC NEEDLES; MAGIC NEEDLE; MAGIC NEEDLE | NA |
| 510k Applicant | Eclipse MedCorp, LLC. | M.V. S.R.L. | NA |
| Product Code | FMI | FMI | Same |
| 510k Number | K200017 | K110606 | NA |
| Indications for Use | The Eclipse DermaFlexCannula is intended to injectfluids intradermally. | The MV intradermic needlesare intended to inject fluidsintradermally. | Same |
| Structure | Hub, Needle, Protective Cap | Hub, Needle, Protective Cap | Same |
| Principle of Operation | This device is used in conjunctionwith a piston syringe to deliver fluids/medications/drugs into the body.The device consists of a metal tube(needle), joined to a femaleconnector (hub). This device comesin a variety of needle gauges andlengths. | This device is used in conjunctionwith a piston syringe to deliverfluids/ medications/drugs into thebody. This device consists of metaltube (needle), joined to a femaleconnector (hub). This device comesin a variety of needle gauges andlengths. | Same |
| Tip Configuration | Sharpened tip (pilot needle) Closedblunt tip, lateral opening(Intradermic needle) | Sharpened tip (pilot needleClosed blunt tip, lateral opening(Intradermic needle) | Same |
| Needle Gauge | 21, 22, 23, 25, 26, 27, 30 | 21, 22, 23, 25, 26, 27, 30 | Same |
| Needle TubeLength | 25, 38, 40, 50, 60, 70mm | 13, 25, 27, 35, 37, 40, 50, 57,70mm | Equivalent: Needlelengths withinpredicate's range |
| Needle TubeMaterial | Stainless Steel 304 | Stainless Steel 304 | Same |
| Connection toSyringe | Luer taper | Luer taper | Same |
| Needle HubMaterial | Polypropylene | Polypropylene | Same |
| Needle CapMaterial | Polypropylene | Polypropylene | NA |
| LubricantComposition | Silicone | Silicone | Same |
| Sterilizationmethod | Ethylene Oxide Gas | Ethylene Oxide Gas | Same |
Performance Testing
Performance testing for the DermaFlex Cannula was conducted in accordance with the following standards.
- Sterlization Validation Test in accordance with ISO 11135, ISO 11138-1, ISO 11138-2, ISO 11737-1, ISO . 10993-7, and AAMI TIR28
- . Shelf Life Validation Test in accordance with ISO 11608-2, ASTM F1929, and ISO 11737-2
- Biocompatibility testing in accordance with ISO 10993-1. ISO 10993-10, ISO 10993-10, ISO ● 10993-4
- Performance Tests in accordance with ISO 9626:2016, ISO 7864: 2016, ISO 80369-7
- ISO 6009: 2016–Hypodermic needles for single use Colour coding for identification ●
The results of the performance testing for the device meets the device meets the established characteristics and performance requirements needed to perform its intended function. Substantial equivalence is established as these are the same standards established for the predicate device as well. As the Eclipse meets the required criteria of the standards, its performance may be considered to support substantial equivalence.
Conclusion
Based on the information provided in this premarket notification, Eclipse MedCorp, LLC concludes that the Eclipse DermaFlex Cannula is substantially equivalent to predicate device.
§ 880.5570 Hypodermic single lumen needle.
(a)
Identification. A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.(b)
Classification. Class II (performance standards).