(300 days)
The Creo Electrosurgical System is intended to provide microwave energy for coagulation (ablation) of soft tissue and is for use only with compatible Creo Medical instruments and accessories.
The System is not intended for use in cardiac procedures.
The NP1 instrument, when used with the Creo Medical Electrosurgical System, is intended for use in coagulation (ablation) of soft tissue.
The NP1 Instrument is not intended for use in cardiac procedures.
The Creo Medical Electrosurgical System with NP1 Accessory includes the following:
- . NP1 Instrument
- . Electrosurgical Generator
- . Footswitch
- Interface Cable with Sterile Sheaths
The Electrosurgical Generator is designed to deliver bipolar radiofrequency (RF) energy and microwave (MW) energy for the purpose of cutting and coagulating tissue. Only the microwave output is used to power the NP1 Instrument. The Electrosurgical Generator output is actuated via a two-pedal Footswitch. One pedal activates the bipolar RF energy output for cutting; the other pedal activates the MW energy output for coagulation. The Electrosurgical Generator incorporates proprietary software developed by Creo Medical for generating and controlling the two energies delivered. The Electrosurgical Generator and Footswitch are non-sterile and reusable.
The single-use Interface Cable connects compatible Creo electrosurgical instruments to the Electrosurgical Generator and is supplied with a sterile sheath that is fitted over its distal end during use to minimize the potential for contamination of the sterile field.
The sterile, single-use NP1 soft tissue ablation accessory (Instrument) consists of a handle, 1.4 m long x 2.7 mm diameter shaft, active nitinol antenna and pointed ceramic tip. Stainless-steel bipolar electrodes separated by dielectric materials couple microwave energy to target tissue when NP1 is in use. NP1 is only compatible for use with the Creo Medical Electrosurgical System, and the System is only intended for use with compatible Creo instruments and accessories.
I am sorry, but the provided text from the FDA 510(k) Pre-Market Notification for the Creo Medical Electrosurgical System with NP1 Accessory does not contain the specific information required to answer your request.
The document is a notification of substantial equivalence for a medical device, and while it mentions performance testing, it does not include:
- A table of acceptance criteria and reported device performance related to a specific study. It broadly states, "The system meets all design specifications, risk mitigation requirements, and applicable medical device standard requirements." and references "Results of mechanical and ablation performance tests, including bench tests and simulated use studies conducted in ex-vivo tissue, demonstrate that the NP1 Electrosurgical System meets design specifications and performs as intended." but no specifics.
- Sample sizes used for a test set or data provenance for specific performance studies. It mentions "simulated use studies conducted in ex-vivo muscle, liver, and kidney tissue" but does not give sample sizes or specific data provenance beyond being ex-vivo tissue.
- Number of experts used or their qualifications for establishing ground truth. This information is entirely absent for any performance claims.
- Adjudication method for a test set. Not mentioned.
- Details about a multi-reader multi-case (MRMC) comparative effectiveness study, including effect size. This type of study is not described or referenced.
- Details about a standalone (algorithm only) performance study. The device is an electrosurgical system, not an AI algorithm, so this concept does not apply in the typical sense of AI/ML software. Its performance is tied to its physical operation and effect on tissue.
- The type of ground truth used (e.g., pathology, outcomes data). For the ablation studies, it implies the "ablation volumes created" as a measure of performance, but it doesn't describe who established this ground truth or how precisely it was measured in terms of a clinical "ground truth."
- The sample size for a training set. As this is a hardware device with embedded software for control, not a machine learning model, the concept of a "training set" for an AI algorithm is not applicable here. The software verification and validation are mentioned but not in the context of a machine learning training set.
- How the ground truth for the training set was established. Not applicable for the reasons above.
The document primarily focuses on demonstrating substantial equivalence to a predicate device through:
- Comparison of technological characteristics.
- Compliance with electrical safety, EMC, and biocompatibility standards.
- Software verification and validation (for control software, not AI/ML).
- Bench testing and simulated use testing to show the system meets design specifications and performs as intended.
Therefore, I cannot extract the requested information from the provided text.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 28, 2020
Creo Medical Ltd. % Phil Triolo President and Senior Consultant Phil Triolo and Associates LC 86 Skycrest Lane Salt Lake City, Utah 84108
Re: K200003
Trade/Device Name: Creo Electrosurgical System with NP1 Accessory Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories Regulatory Class: Class II Product Code: NEY Dated: December 29, 2019 Received: September 21, 2020
Dear Phil Triolo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K200003
Device Name
Creo Electrosurgical System with NP1 Accessory
Indications for Use (Describe)
The Creo Electrosurgical System is intended to provide microwave energy for coagulation (ablation) of soft tissue and is for use only with compatible Creo Medical instruments and accessories.
The System is not intended for use in cardiac procedures.
The NP1 instrument, when used with the Creo Medical Electrosurgical System, is intended for use in coagulation (ablation) of soft tissue.
The NP1 Instrument is not intended for use in cardiac procedures.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| ☑ Prescription Use (Part 21 CFR 801 Subpart D) | |
| ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) Summary
Date Prepared: October 26, 2020
1. SUBMITTER
- 510(k) Sponsor: Creo Medical Ltd Creo House, Unit 2, Beaufort Park Way Chepstow NP16 5UH United Kingdom
- Contact Persons: Phil Triolo, PhD, RAC President and Senior Consultant Phil Triolo and Associates LC Phone: 801-699-9846 / Email: philt@philt.com Shaun Preston Senior Design and Development Engineer Creo Medical Ltd. +44 1291 606005 (office) / +44 7539 732362 (mobile) Email:
- Prepared By: Phil Triolo and Associates, LC 86 Skycrest Lane SLC UT 84108
2. DEVICE
| Trade/Device Name: | Creo Electrosurgical System with NP1 Accessory | |
|---|---|---|
| Common Name: | Electrosurgical System | |
| Classification Name: | Electrosurgical cutting and coagulation device and accessories | |
| Regulation number: | CFR 878.4400 | |
| Product Code: | NEY (Class II) |
3. PREDICATE DEVICES
K163105 Emprint Ablation System
{4}------------------------------------------------
SYSTEM DESCRIPTION 4.
The Creo Medical Electrosurgical System with NP1 Accessory includes the following:
- . NP1 Instrument
- . Electrosurgical Generator
- . Footswitch
- Interface Cable with Sterile Sheaths
The Electrosurgical Generator is designed to deliver bipolar radiofrequency (RF) energy and microwave (MW) energy for the purpose of cutting and coagulating tissue. Only the microwave output is used to power the NP1 Instrument. The Electrosurgical Generator output is actuated via a two-pedal Footswitch. One pedal activates the bipolar RF energy output for cutting; the other pedal activates the MW energy output for coagulation. The Electrosurgical Generator incorporates proprietary software developed by Creo Medical for generating and controlling the two energies delivered. The Electrosurgical Generator and Footswitch are non-sterile and reusable.
The single-use Interface Cable connects compatible Creo electrosurgical instruments to the Electrosurgical Generator and is supplied with a sterile sheath that is fitted over its distal end during use to minimize the potential for contamination of the sterile field.
The sterile, single-use NP1 soft tissue ablation accessory (Instrument) consists of a handle, 1.4 m long x 2.7 mm diameter shaft, active nitinol antenna and pointed ceramic tip. Stainless-steel bipolar electrodes separated by dielectric materials couple microwave energy to target tissue when NP1 is in use. NP1 is only compatible for use with the Creo Medical Electrosurgical System, and the System is only intended for use with compatible Creo instruments and accessories.
5. INDICATIONS FOR USE
The Creo Electrosurqical System is intended to provide microwave energy for coagulation (ablation) of soft tissue and is for use only with compatible Creo Medical instruments and accessories.
The System is not intended for use in cardiac procedures.
The NP1 instrument, when used with the Creo Medical Electrosurgical System, is intended for use in coagulation (ablation) of soft tissue.
The NP1 Instrument is not intended for use in cardiac procedures.
6. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICES
The Electrosurgical Generator (when used to power the NP1 Instrument) is designed to provide microwave energy at 5.8 GHz. The generator output voltage, power, current, duty cycle, and duration are under the control of software and hardware circuits to assure the energy delivered to the NP1 instrument via the Interface Cable is within specifications. When NP1 is connected to the Electrosurgical Generator and selected for use using either the Footswitch or membrane switches on the Generator, only microwave energy is delivered.
The NP1 Instrument is a flexible electrosuraical accessory that employs stainless-steel bipolar electrodes separated by dielectric materials to couple microwave energy delivered by the generator to target tissue as directed by the user. NP1 is only compatible for use with the Creo Medical Electrosurgical System, and the System is only intended for use with compatible Creo instruments and accessories.
{5}------------------------------------------------
Creo Electrosurgical System with NP1 Accessory
K200003
The subject and predicate devices are based on the following same technological elements, as both systems use microwave energy to effect ablation of soft tissue. Both devices utilize the fundamental mechanism for electromagnetic energy coupling to tissue employed by all microwave-powered ablation devices. The interaction between the microwave electromagnetic fields and soft tissue primarily results in rapid oscillation of dipolar water molecules, frictional resistance to such movement, and resultant active localized tissue heating. Thermal conduction, due to the temperature gradient created by the active tissue heating region, results in adjacent tissue heating (thermal spread). The mechanism of microwave energy absorption is known as dielectric heating. For a specified tissue type and microwave frequency, the rate of microwave energy absorption is predictable, which allows for control of the energy absorption and subsequent thermal effects.
The following technological differences between the subject and predicate devices were compared:
| Difference | Evidence supporting substantial equivalence |
|---|---|
| The NP1 Instrument is supplied with pulsed62 W microwave power at 5.8 GHzwhereas the predicate antenna is suppliedwith microwave power at 2.45 GHz | Results of simulated use studies conducted in exvivo muscle, liver, and kidney tissue were used tocharacterize the ablation volumes created by thenew and predicate systems, which have beendetermined to be equivalent. |
| The predicate antenna is actively cooled bywater circulated through the antenna; NP1is not cooled. | Evaluations in ex vivo bovine tissue demonstratethat the temperature of applied parts does notexceed permissible limits and that no non-targettissue is thermally injured as a result of NP1treatments at its maximum power setting. |
| Different patient-contacting materials | Results of biocompatibility and performanceevaluations demonstrate an acceptable biologicalrisk and that acceptance criteria are met. |
| NP1 Interface Cable is single use,predicate cable is reusable | Both cables conduct microwave energy for ablationand neither poses a significant risk for infection. |
| NP1 System is controlled by software andhardware, predicate only by hardware | EMC, electrical safety, and verification test resultsdemonstrate acceptable safety and performance. |
7. PERFORMANCE DATA
The system meets all design specifications, risk mitigation requirements, and applicable medical device standard requirements.
Sterilization/Shelf-life
Per ISO 11135 and AAMI TIR28 standard evaluation, the NP1 Sterility Assurance Level is 10°
The shelf-life for the NP1 Instrument is 6 months. The Interface Sheath is 3 years.
{6}------------------------------------------------
Creo Electrosurgical System with NP1 Accessory
Biocompatibility
The biocompatibility evaluation for the NP1 Instrument was conducted in accordance with ISO 10993-1:2009 and the FDA guidance document, Use of International Standard ISO 10993-1, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process: Guidance for Industry and Food and Drug Administration Staff," issued by the FDA on June 16, 2016.
According to Table A.1: Biocompatibility Evaluation Endpoints in the FDA Biocompatibility Guidance, the antenna assembly of NP1 and the distal portion of its shaft are categorized, from a biological risk perspective, as externally communicating with limited (< 24h) contact with tissue.
The battery of tests whose results support a finding of a low and acceptable biological and toxicological risk included the following:
- Cytotoxicity
- . Sensitization
- . Irritation
- . Systemic toxicity
- . Pyrogen Testing
Electrical Safety, Microwave Safety and Electromagnetic Compatibility
Electrical safety and EMC tests were conducted on the Creo Medical Electrosurgical System with NP1 Accessory. The system complies with FDA-recognized versions of IEC 60601-2-2 and IEC 60601-2-6 standards for safety and the IEC 60601-1-2 standard for EMC.
Software Verification and Validation Testing
Software verification and validation testing were conducted, and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered a "major" level of concern, since a failure or latent flaw in the software could directly result in serious injury or death to the patient or operator.
Mechanical and Ablation Performance Testing
- . Handle and shaft maximum temperature evaluation
- . Post-activation blanch testing
- Tensile strength testing of shaft/antenna components
- Strength testing for handle connection, antenna locking knob and shaftadjustment locking knob
- Standing wave ratio / Reflected power threshold testing
- . Simulated use testing in ex-vivo tissue (ablation zone repeatability and predicate comparison studies)
Results of mechanical and ablation performance tests, including bench tests and simulated use studies conducted in ex-vivo tissue, demonstrate that the NP1 Electrosurgical System meets design specifications and performs as intended, and support Creo's assertion of substantial equivalence to the predicate device.
{7}------------------------------------------------
CONCLUSION 8.
The nonclinical bench testing, simulated use testing, hardware and software verification and validation and standards testing demonstrate there are no new or different questions of safety or effectiveness raised by any differences between the subject device and predicate device. The Creo Medical Electrosurgical System with NP1 Accessory as determined by the indications for use, technological characteristics and the performance testing demonstrate it is as safe and effective as the legally marketed device identified and is substantially equivalent to the identified predicate device.
§ 878.4400 Electrosurgical cutting and coagulation device and accessories.
(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.