(99 days)
BlueStar® Rx is indicated for use by healthcare providers (HCPs) and their patients – aged 18 years and older – who have type 1 or type 2 diabetes. The BlueStar® Rx is intended to provide secure capture, storage, and transmission of blood glucose data as well as information to aid in diabetes self-management. The BlueStar® Rx analyzes and reports blood glucose test results and supports medication adherence. In addition, the BlueStar® Rx provides coaching messages (motivational, behavioral, and educational) based on real-time blood glucose values and trends. It includes software intended for use on mobile phones or personal computers in the home or in professional healthcare settings. The software also allows for entry of other diabetes-related healthcare information and provides educational information.
For bolus insulin users with type 1 and type 2 diabetes, BlueStar® Rx includes an insulin dose calculator to allow patients to use their prescribed regimen to calculate a dose of bolus insulin for a given amount of carbohydrates and/or blood glucose value.
For basal insulin users with type 2 diabetes, BlueStar® Rx includes an Insulin Adjustment Program (IAP) which calculates appropriate long-acting basal insulin doses for titrating insulin levels based on configuration by a healthcare provider knowledgeable in the care and management of diabetes. The healthcare provider must activate the Insulin Adjustment Program and configure it for patient-specific parameters.
The BlueStar® Rx is not intended to replace the care provided by a licensed healthcare professional, including prescriptions, diagnosis, or treatment.
BlueStar Rx is a modified version of the primary predicate (WellDoc BlueStar Rx cleared under K190013). BlueStar Rx maintains all of the features of the primary predicate and adds the titration of long-acting basal insulin doses for qualified type 2 diabetes patients who are not using bolus insulin. This long-acting basal insulin titration feature is included as part of the Insulin Adjustment Program (IAP) in BlueStar Rx. WellDoc BlueStar® as cleared under K190013, is a stand-alone software system intended to be used by healthcare providers (HCPs) and their patients – aged 18 years and older - who have type 1 or type 2 diabetes. The system is intended to assist type 1 and type 2 diabetes patients to self-manage their disease. Patients receive guidance on diabetes selfmanagement and are encouraged to reach out to their healthcare team when needed. There are two versions of the primary predicate WellDoc BlueStar® (cleared under K190013) -BlueStar® (OTC version) and BlueStar® Rx (prescription only) applications. In this submission, Welldoc does not propose to change the OTC version of BlueStar application. Like the bolus insulin dose calculator already included in the primary predicate (BlueStar Rx), the new long-acting basal insulin titration feature is also limited to the prescription use (Rx) version. The bolus insulin calculator feature will be disabled in patients using long-acting basal insulin titration feature in the BlueStar Rx. Health care providers (HCPs) will be required to initiate and manage the basal insulin titration for their qualified type 2 diabetes patients using the following two interfaces:
- Web based HCP interface for use by the providers to prescribe long-acting basal insulin doses for the qualified type 2 diabetes patients.
- Web and mobile patient interface for use by patients to follow provider's basal insulin titration plan.
The IAP feature in BlueStar Rx provides directions to the patients based on prescription by their HCP for titrating long-acting insulin doses only.
BlueStar® Rx will also maintain the following features of the primary predicate (WellDoc BlueStar Rx cleared under K190013): (1) ability to connect to the One Touch Verio Flex Blood Glucose Meter via Bluetooth which allows users to send data from their BG meter to the BlueStar® Rx app. which will provide coaching messages (motivational, and educational) based on the real-time blood glucose values and trends. (2) The BlueStar Server will also have the ability to transmit data to the OneTouch Reveal Server. With this application, 3-hour delayed continuous glucose monitoring device data can be accessed via API and uploaded for data visualization purposes only.
Here's a breakdown of the requested information based on the provided text:
1. Table of acceptance criteria and the reported device performance
Acceptance Criteria (from FDA Guidence) | Reported Device Performance |
---|---|
Software verification and validation per the FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices (May 11, 2005) for a Major Level of Concern and "Guidance for Industry, FDA Reviewers and Compliance on Off-the-Shelf Software Use in Medical Devices". | "Software verification and validation per the FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices (May 11, 2005) for a Major Level of Concern FDA Guidance “Guidance for Industry, FDA Reviewers and Compliance on Off-the-Shelf Software Use in Medical Devices”" (It's stated that this bench testing was performed). |
Cybersecurity evaluation per the FDA Guidance Content of Premarket Submissions for Management of Cybersecurity in Medical Devices Guidance for Industry and Food and Drug Administration Staff, (October 2, 2014). Specifically, addressing: Identify and Protect, Detect, Response and Recover. | "Cybersecurity was evaluated per the FDA Guidance Content of Premarket Submissions for Management of Cybersecurity in Medical Devices Guidance for Industry and Food and Drug Administration Staff, (October 2, 2014). Specifically, addressing the following areas: Identify and Protect, Detect, Response and Recover" (It's stated that this bench testing was performed). |
Human factors validation (user interface adequately validated for use per labeling) with intended user populations. | "Human factors testing was conducted with the intended user populations of patients and healthcare providers. The human factors, design, and labeling information provided in the submission confirm that the user interface has been adequately validated for use per the labeling." (It's stated that this testing was conducted). |
2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)
The document mentions "Performance testing" including "Software verification and validation," "Cybersecurity," and "Human Factors" testing. However, it does not specify sample sizes for any test sets, nor does it provide information about the data provenance (e.g., country of origin or whether it was retrospective or prospective).
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
The document mentions "intended user populations of patients and healthcare providers" for human factors testing, but does not specify the number or qualifications of experts used to establish ground truth for any of the performance tests.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set
The document does not specify any adjudication method for the test set.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
A multi-reader multi-case (MRMC) comparative effectiveness study was not specifically mentioned or described in the provided text. The document refers to the device assisting self-management and providing calculated insulin doses, but not in the context of human reader performance improvement with AI assistance.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
The document describes the "Insulin Adjustment Program (IAP)" which calculates appropriate long-acting basal insulin doses based on configuration by a healthcare provider. While the calculation itself is algorithmic, the system requires a "human-in-the-loop" for activation, configuration, and management by the healthcare provider, and the patient follows directions. Therefore, a purely standalone algorithm-only performance study in isolation from human interaction is not explicitly described in the context of the device's function. The performance testing focuses on software, cybersecurity, and human factors interaction with the system, implying a human-in-the-loop context.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The document does not explicitly state the type of ground truth used for the performance testing. For software verification and validation, ground truth would typically be derived from functional specifications and expected outputs. For human factors, it would likely involve observations and feedback from users regarding usability and clarity.
8. The sample size for the training set
The document does not provide any information regarding the sample size for a training set. This is generally because the device is a rule-based system (an "Insulin Adjustment Program" that "calculates appropriate long-acting basal insulin doses based on configuration by a healthcare provider") rather than a machine learning model that requires explicit training data.
9. How the ground truth for the training set was established
Since the document does not mention a training set, it does not describe how ground truth for a training set was established.
§ 880.5725 Infusion pump.
(a)
Identification. An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.(b)
Classification. Class II (performance standards).