K Number
K193514
Manufacturer
Date Cleared
2023-04-13

(1211 days)

Product Code
Regulation Number
862.1340
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The AUTION MAX AX-4060 Urinalysis System (AUTION MAX) is comprised of the AUTION MAX AX-4060 automated urine analyzer and AUTION Sticks 9EB multi-parameter test strips.

The AUTION MAX AX-4060 urine analyzer, when used with AUTION Sticks 9EB test strips is a fully automated urinalysis system intended for the in vitro qualitative or semi-quantitative measurement of the following analytes: glucose, protein, bilirubin, urobilinogen, pH, blood, ketones, nitrite, leukocytes, turbidity, and color. The test results of these parameters can be used in the evaluation of kidney, urinary, liver and other metabolic disorders. This system is intended to be used by trained operators in clinical laboratories.

AUTION Sticks 9EB test strips are test strips for the in vitro qualitative or semi-quantitative measurement of the following analytes: glucose, protein, bilirubin, urobilinogen, pH, blood, ketones, nitrite, and leukocytes with the AUTION MAX AX-4060 urine analyzer. The test results of these parameters can be used in the evaluation of kidney, urinary, liver and other metabolic disorders.

Special conditions for use statements: Prescription use only. AUTION Sticks 9EB test strips are not to be read visually.

Device Description

The AUTION MAX AX-4060 Urinalysis System is a fully automated urine analyzer that provides a semi-quantitative or qualitative measurement for glucose, protein, bilirubin, urobilinogen, pH, blood, ketones, nitrites, leukocytes, specific gravity, turbidity, and color tone. The AUTION MAX AX-4060 Urinalysis System consists of AUTION MAX AX-4060 urine analyzer and AUTION Sticks 9EB test strips.

The AUTION Sticks 9EB test strips consist of a plastic strip containing 9 dry chemistry reagent pads impregnated with chemical substances for the determination of the above analytes in urine. These substances if present in urine leads to a chemical reaction that results in a color change, which is measured by the AUTION MAX AX-4060 urine analyzer based on spectrophotometry. In addition, three additional parameters, specific gravity, turbidity, and color tone are directly measured based on reflectance refractometry, transmitted and scattered light measurements.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided FDA 510(k) summary:

Device: AUTION MAX AX-4060 Urinalysis System
Predicate Device: AUTION MAX AX-4030 Fully Automated Urinalysis System


1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are not explicitly stated as distinct numerical targets in the document. Instead, the study aims to demonstrate substantial equivalence to the predicate device. Therefore, the "acceptance criteria" are implicitly understood as matching or performing comparably to the predicate device's established performance, particularly in terms of agreement percentage. The reported device performance is presented as the agreement between the proposed device (AUTION MAX AX-4060) and the predicate device (AUTION MAX AX-4030).

Key Performance Metric: Percentage of Exact Match and Percentage within +/- 1 Color Block Match compared to the predicate device.

AnalyteAcceptance Criteria (Implicit: Substantial Equivalence to Predicate)Reported Device Performance (Exact Match %)Reported Device Performance (+/- 1 CB Match %)
GLUHigh agreement (Exact and +/- 1 CB match)98.1%100.0%
PROHigh agreement (Exact and +/- 1 CB match)94.5%100.0%
BILHigh agreement (Exact and +/- 1 CB match)99.5%100.0%
UROHigh agreement (Exact and +/- 1 CB match)99.2%100.0%
pHHigh agreement (Exact and +/- 1 CB match)92.9%100.0%
BLDHigh agreement (Exact and +/- 1 CB match)95.8%100.0%
KETHigh agreement (Exact and +/- 1 CB match)99.0%100.0%
NITHigh agreement (Exact and +/- 1 CB match)99.7%100.0%
LEUHigh agreement (Exact and +/- 1 CB match)98.6%100.0%
TurbidityHigh agreement (Exact and +/- 1 CB match)98.6%100.0%
Color toneHigh agreement (Exact and +/- 1 CB match)82.7%100.0%

(Note: The +/- 1 CB Match indicates results falling within one color block difference from the predicate, which is often considered acceptable for semi-quantitative tests.)


2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set (Method Comparison): A total of 1374 samples were used. This included:
    • 1333 natural patient samples
    • 41 spiked samples
  • Data Provenance: The study was conducted at two (2) clinical laboratories. The document does not specify the country of origin, but given the FDA submission, it's likely conducted in or in accordance with standards for the U.S. and/or Japan (where the manufacturer is located). It is a prospective comparison study.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

This study does not involve human experts establishing a "ground truth" for each sample in the typical sense of diagnostic imaging or pathology. Instead, the "ground truth" for the method comparison study is the result obtained from the legally marketed predicate device (AUTION MAX AX-4030 Automated Urinalysis System). The study demonstrates the correlation and agreement of the new device's readings with the predicate device's readings. No information is provided about experts interpreting results aside from the automated systems.


4. Adjudication Method for the Test Set

Since the "ground truth" is established by the predicate device's results and the comparison is between two automated systems, there was no human adjudication process (e.g., 2+1, 3+1 consensus) described for the test set.


5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No. A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not performed. This study evaluates the performance of an automated urinalysis system against a predicate automated system, not the improvement of human readers with AI assistance.


6. Standalone Performance (Algorithm Only without Human-in-the-Loop Performance)

Yes. The primary study detailed, especially the Method Comparison (Section 14), assesses the standalone performance of the AUTION MAX AX-4060 Urinalysis System. It compares the results generated by the new automated system directly against those generated by the predicate automated system, without human intervention in the result interpretation or decision-making process during the comparison phase. The device itself is an automated system intended to be used by trained operators in clinical laboratories, but the performance data presented is for the automated analyzer's output.


7. Type of Ground Truth Used

The ground truth for the method comparison study was the results obtained from the legally marketed predicate device (AUTION MAX AX-4030 Automated Urinalysis System). For the initial precision study (Tables 3-5), the "Expected Result" for quality controls served as the reference.


8. Sample Size for the Training Set

The document is a 510(k) summary for a medical device (urinalysis system), not a typically "AI-driven" software device that undergoes explicit training. The device determines analytes based on chemical reactions and spectrophotometry (reflectance refractometry, transmitted and scattered light measurements). Therefore, there is no explicit "training set" in the context of machine learning algorithms usually discussed. The system is likely calibrated and validated during its development, but this is a different process than training an AI model.


9. How the Ground Truth for the Training Set Was Established

As noted above, there is no explicit "training set" in the machine learning sense for this device. The chemical principles and measurement methods are well-established for urinalysis. Any calibration or internal development would have relied on known concentrations and reference methods for each analyte to ensure accurate colorimetric or other measurements. The basis of the technology relates to established chemical reactions on the test pads (e.g., Glucose oxidase reaction for glucose, Protein-error reaction for protein, etc.).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left side of the image is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

April 13, 2023

ARKRAY Inc. Daya Ranamukhaarachchi VP/Scientific and Regulatory Affairs 5198 West 76th Street Minneapolis, MN 55439

Re: K193514

Trade/Device Name: AUTION MAX AX-4060 Urinalysis System Regulation Number: 21 CFR 862.1340 Regulation Name: Urinary glucose (nonquantitative) test system Regulatory Class: Class II Product Code: JIL, JIO, KQO, CDM, JIR, JMT, LJX, JJB, CEN, JIN Dated: October 19, 2022 Received: October 20, 2022

Dear Daya Ranamukhaarachchi:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR

{1}------------------------------------------------

  1. for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Paula V. Caposino -S

Paula Caposino, Ph.D. Acting Deputy Division Director Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K193514

Device Name AUTION MAX AX-4060 Urinalysis System

Indications for Use (Describe)

The AUTION MAX AX-4060 Urinalysis System (AUTION MAX) is comprised of the AUTION MAX AX-4060 automated urine analyzer and AUTION Sticks 9EB multi-parameter test strips.

The AUTION MAX AX-4060 urine analyzer, when used with AUTION Sticks 9EB test strips is a fully automated urinalysis system intended for the in vitro qualitative measurement of the following analytes: glucose, protein, bilirubin, urobilinogen, pH, blood, ketones, nitrite, leukocytes, turbidity, and color. The test results of these parameters can be used in the evaluation of kidney, urinary, liver and other metabolic disorders. This system is intended to be used by trained operators in clinical laboratories.

AUTION Sticks 9EB test strips are test strips for the in vitro qualitative or semi-quantitative measurement of the following analytes: glucose, protein, bilirubin, urobilinogen, pH, blood, ketones, nitrite, and leukocytes with the AUTION MAX AX-4060 urine analyzer. The test results of these parameters can be used in the evaluation of kidney, urinary, liver and other metabolic disorders.

Special conditions for use statements: Prescription use only. AUTION Sticks 9EB test strips are not to be read visually.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

5. 510(K) Summary

The 510(k) Summary is included on the following pages.

{4}------------------------------------------------

510(k) Summarv

This 510(k) Summary of the AUTION MAX AX-4060 Urinalysis System is submitted in compliance with 21 CFR807.92 for the purposes of safety and effectiveness.

Date Prepared:October 19, 2022
Establishment:ARKRAY Factory Inc.1480 Koji, Konan-cho,Koka-shiShiga Japan, 520-3306
Submitter:ARKRAY Inc.Yousuien-nai, 59 Gansuin-cho, Kamigyo-kuKyoto Japan, 602-0008Official US Agent: Daya RanamukhaarachchiARKRAY Factory USA Inc.VP/Scientific and Regulatory AffairsPhone (408) 234-0804Fax: (952) 646-3240Email: ranamukd@arkrayusa.com
Primary Contact:Daya RanamukhaarachchiVP/Scientific and Regulatory AffairsPhone (408) 234-0804Fax: (952) 646-3240Email: ranamukd@arkrayusa.com
Secondary Contact:Dhwani ThakkarRegulatory Affairs Project ManagerEmail: thakkard@arkrayusa.comPhone: 437-247-1628
Device Name:
Trade Name:1. AUTION MAX AX-4060 Urinalysis System
Common Name:1. Automated Urinalysis System2. Urinalysis Test Strip

{5}------------------------------------------------

Predicate Device:

    1. AUTION MAX AX-4030 Fully Automated Urinalysis System (K093098)
    1. AUTION Sticks 9EB (K013783)

Regulatory Information:

Regulatory Classification for Urinalysis system and each of the tests are listed in Table 1.

Test DescriptionProductCodeDevice ClassRegulationDevice Panel
Automated urinalysissystemKQOClass I21 CFR 862.290075 Chemistry
Glucose, urinary,nonquantitativeJILClass II21 CFR 862.134075 Chemistry
Blood, occult,colorimetric, in urineJIOClass II21 CFR 864.655081 Hematology
Urobilinogen, urinary,non-quantitativeCDMClass I21 CFR 862.178575 Chemistry
Protein or albumin,urinary, nonquantitativeJIRClass I21 CFR 862.164575 Chemistry
Nitrite, nonquantitativeJMTClass I21 CFR 862.151075 Chemistry
Leukocyte, peroxidasetestLJXClass I21 CFR 864.767581 Hematology
Bilirubin and itsconjugates, urinary,non-quantitativeJJBClass I21 CFR 862.111575 Chemistry
pH, urinary,nonquantitativeCENClass I21 CFR 862.155075 Chemistry
Ketones nonquantitativeJINClass I21 CFR 862.143575 Chemistry
Table 1. Regulatory Classification for AUTION MAX AX-4060 Urinalysis System
---------------------------------------------------------------------------------

Intended Use

The AUTION MAX AX-4060 Urinalysis System (AUTION MAX) is comprised of the AUTION MAX AX-4060 automated urine analyzer and AUTION Sticks 9EB multi-parameter test strips.

The AUTION MAX AX-4060 urine analyzer, when used with AUTION Sticks 9EB test strips is a fully automated urinalysis system intended for the in vitro qualitative or semi-quantitative measurement of the following analytes: glucose, protein, bilirubin, urobilinogen, pH, blood, ketones, nitrite, leukocytes, turbidity, and color. The test results of these parameters can be used in the evaluation of kidney, urinary, liver and other metabolic disorders. This system is intended to be used by trained operators in clinical laboratories.

AUTION Sticks 9EB test strips are test strips for the in vitro qualitative or semi-quantitative

{6}------------------------------------------------

measurement of the following analytes: glucose, protein, bilirubin, urobilinogen, pH, blood, ketones, nitrite, and leukocytes with the AUTION MAX AX-4060 urine analyzer. The test results of these parameters can be used in the evaluation of kidney, urinary, liver and other metabolic disorders.

Special conditions for use statements: Prescription use only. AUTION Sticks 9EB test strips are not to be read visually.

Device Description

The AUTION MAX AX-4060 Urinalysis System is a fully automated urine analyzer that provides a semi-quantitative or qualitative measurement for glucose, protein, bilirubin, urobilinogen, pH, blood, ketones, nitrites, leukocytes, specific gravity, turbidity, and color tone. The AUTION MAX AX-4060 Urinalysis System consists of AUTION MAX AX-4060 urine analyzer and AUTION Sticks 9EB test strips.

The AUTION Sticks 9EB test strips consist of a plastic strip containing 9 dry chemistry reagent pads impregnated with chemical substances for the determination of the above analytes in urine. These substances if present in urine leads to a chemical reaction that results in a color change, which is measured by the AUTION MAX AX-4060 urine analyzer based on spectrophotometry. In addition, three additional parameters, specific gravity, turbidity, and color tone are directly measured based on reflectance refractometry, transmitted and scattered light measurements.

Substantial Equivalence

AUTION MAX AX-4060 Urinalysis System was developed as the successor to the predicate device AUTION MAX AX-4030 Urinalysis System. The device uses the same intendeduse claims as the predicate. The AX-4060 system utilizes the same principles and technology and shares the identical 9EB test strips for measuring the same urine analytes in a semiquantitative or qualitative manner similar to the predicate device. Some important device parts such as LED and Nozzle dispensing units were changed in the AX-4060 System due to a change in manufacturing source and the device was optimized to generate similar performance to the predicate device.

As described in the Performance Characteristics Section below, the verification and validation (bench and clinical) testing demonstrated substantial equivalence for the AX-4060 System with AUTION Sticks 9EB test strips to the predicate device. Bench testing supports that the AX-4060 System performs to meet specifications for precision, linearity, limits of detection (LOD), sensitivity, carryover and interference. Additionally, software testing, electrical safety testing and EMC testing also met performance criteria.

A list of similarities and differences between the proposed device and the predicate device is provided in Table 2.

{7}------------------------------------------------

Table 2. Substantial Equivalence comparison between AUTION MAX AX-406
Urinalysis System and AUTION MAX AX-4030 Urinalysis System
COMPONENT/CHARACTERISTICPROPOSEDPREDICATE
510(k) NumberTo Be DeterminedK093098
DeviceAUTION MAX AX-4060 UrinalysisSystemAUTION MAX AX-4030Urinalysis System
Similarities between Proposed and Predicate Devices
Device TypeAutomated Urinalysis SystemSame
FDA Product CodeKQO, JIL, JIOSame
FDA Authorized UsePrescription UseSame
FDA ClassificationClass IISame
SpecimenType/Measurement ObjectUrineSame
Reagent(Chemical Test Strips)AUTION Sticks 9EBSame
Measurement ItemsGLU (Glucose), PRO (Protein), BIL(Bilirubin), URO (Urobilinogen),PH (pH), BLD (Blood), KET(Ketones), NIT (Nitrite), LEU(Leukocytes), S.G. (Specificgravity), turbidity, and color tone.Same
Sample ConsumptionMax. 0.90 mLSame
Required Sample VolumeMin. 2.0 mLSame
Test Strip Reaction TimeApprox. 60 secondsSame
Test Strip StorageTwo test strip storage compartments,each of which can contain differenttest stripsSame
Test Strip StorageCapacityMaximum 200 test strips × twostorage compartmentsSame
Max. Processing Speed225 samples/hourSame
Built-in Printer58-mm width printer paper (24digits)Same
COMPONENT/CHARACTERISTICPROPOSEDPREDICATE
Memory CapacityNormal and STAT measurements:2500 testsControl measurement: 200 testsCheck measurement: 50 testsTrouble list: 100 testsSame
Communication SystemRS-232C compliant (Switchablebetween one-way and two-way)Same
Power Requirements100 - 240 V AC (Maximum powerline fluctuation of ±10%), 50/60 HzSame
Power InputMax. 150 VASame
Intended UseEnvironmentFor indoor use onlySame
Measurement environmentTemperature: 10 – 30°C (50 to86°F); Humidity: 30 – 60% RH(No condensation)Same
AltitudeUp to 2000m (6560 feet)Same
Differences between Proposed and Predicate Devices
Intended Use andIndications for UseThe AUTION MAX AX-4060Urinalysis System (AUTION MAX) iscomprised of the AUTION MAX AX-4060 automated urine analyzer andAUTION Sticks 9EB multi-parametertest strips.The AUTION MAX AX-4060 urineanalyzer, when used with AUTIONSticks 9EB test strips is a fullyautomated urinalysis system intendedfor the in vitro qualitative or semi-quantitative measurement of thefollowing analytes: glucose, protein,bilirubin, urobilinogen, pH, blood,ketones, nitrite, leukocytes, turbidity,and color. The test results of theseparameters can be used in theThe AUTION MAX AX-4030 Urinalysis System(AUTION MAX) is anautomated urine analyzerintended for the in vitromeasurement of the followinganalytes: glucose, protein,bilirubin, urobilinogen, pH,blood, ketones, nitrite,leukocytes, specific gravity,turbidity, and color. TheAUTION MAX is intendedfor use only with AUTIONSticks 9EB multi-parametertest strips.
other metabolic disorders. This systemis intended to be used by trainedoperators in clinical laboratories.
AUTION Sticks 9EB test strips aretest strips for the in vitro qualitative orsemi-quantitative measurement of thefollowing analytes: glucose, protein,bilirubin, urobilinogen, pH, blood,ketones, nitrite, and leukocytes withthe AUTION MAX AX-4060 urineanalyzer. The test results of theseparameters can be used in theevaluation of kidney, urinary, liver andother metabolic disorders.
Special conditions for use statements:Prescription use only.AUTION Sticks 9EB test strips arenot to be read visually.
ConfigurationInstrument, Sampler, Hand-heldbarcode reader and AccessoriesSame, but the Hand-held barcode isoptional.

{8}------------------------------------------------

{9}------------------------------------------------

{10}------------------------------------------------

COMPONENT/CHARACTERISTICPROPOSEDPREDICATE
LED Wavelength5 LED wavelengths (450, 525,562, 635 and 760 nm)5 LED wavelengths (430,500, 565, 635, 760 nm)
Warm-up TimeMax. 4 minutesMax. 2 minutes
DisplayLarge color liquid crystal display (800× 480 dots) with touch panelLarge color liquid crystaldisplay (320 × 240 dots)No touch panel
External Output1 port (Replaceable with theoptional Ethernet terminal)2 ports (One of these ports canbe optionally used as anEthernet port)
Transmission SpeedRS-232C: Selectable from 300,600, 1200, 2400, 4800, 9600 and19200 bpsEthernet: 10BASE-T, 100BASE-TXRS-232C: Selectable from300, 600, 1200, 2400, 4800,9600, and 19200 bpsEthernet: 10BASE-T
Dimensions530 (W) × 530 (D) × 550 (H) mm(including the sampler)530 (W) × 530 (D) ×530(H) mm(including the sampler)
WeightInstrument: Approx. 40 kg (88 lb)Sampler: Approx. 4 kg (9 lb)Instrument: approx. 37 kgSampler: approx. 4 kg
Storage EnvironmentTemperature: -10 – 50°C (14 to122°F); Humidity: 20 – 80% RH(No condensation)Temperature: 1 – 30°C (34 to86°F);Humidity: 20 – 80% RH(No condensation)
Environment duringTransportTemperature: -10 – 50°C (14 to122°F); Humidity: 20 – 80% RH(No condensation)Temperature: -13 – 43°C (9 to109°F); Humidity: 20 – 80%RH (No condensation)

Test Principle:

The AUTION MAX AX-4060 urine analyzer measures analyte-specific test strips (AUTION Sticks 9EB) by dual-wavelength reflectance measurement (single wavelength for blood [BLD]). The AUTION MAX AX-4060 Urinalysis System detects urine analytes based on chemical reaction on each test pad relative to the concentration of measured analytes present in the urine sample. The chemical reaction leading to the color change that occurs on each chemical test pad is listed below.

{11}------------------------------------------------

PRINCIPLE
-----------
Glucose: Glucose oxidase reaction.
------------------------------------
GlucoseGOD Gluconic acid + H2O2
H 2 O 2 + 4-AAP + 1-Naphthol-3,6-disulfonic acidPOD Quinone imine dye (purple color) oxidation
Protein: Protein-error reaction.
----------------------------------
Protein + pH indicatoracid pH indicator changes to a cyan color
-------------------------------------------------------------------------------------------------------------------------------------------------------
Bilirubin: Azo-coupling reaction.
-----------------------------------
2-Methyl-5-nitroaniline + Sodium nitriteacid Diazonium salt diazo reaction
Bilirubin + Diazonium saltacid Azo dye (reddish brown color) coupling reaction
Urobilinogen: Azo-coupling reaction.
--------------------------------------
Urobilinogen + Diazonium saltacid Azo dye (reddish brown color) coupling reaction
-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
pH: pH indicator.
-------------------
H + + mixed pH indicatormixed pH indicator shows range of colors covering the urinary pH range (yellow - cyan color)
-----------------------------------------------------------------------------------------------------------------------------------
Blood: Activity measurement of pseudoperoxidase in hemoglobin.
----------------------------------------------------------------
CHP + TMBZhemoglobin H2O + Cumene + Oxidation dye (cyan color) pseudoperoxidase action
-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Ketones: Legal reaction.
--------------------------
Ketones + Sodium nitroprussidealkaline Ketones complex (purple color)
-------------------------------------------------------------------------------------------------------------------------------------------------------------
Nitrite: Griess reaction.
---------------------------
Nitrite + Sulfanilamideacid Diazo-compound + NEDA-2HCI
acid Azo dye (pink color) coupling reaction
Leukocytes: Measurement of leukocyte esterase activity.
---------------------------------------------------------
TAIesterase from leukocytes Indoxyl + MMB Azo dye (purple color)
hydrolysis coupling reaction

{12}------------------------------------------------

Performance Characteristics

Precision

Precision study included repeatability and reproducibility testing in accordance with the CLSI standard EP5-A3. Three (3) levels of quality controls were used to attain results at negative, mid-level positive and high-positive analyte ranges. Two (2) levels of commercial controls were used to obtain the 3 levels by mixing as needed to obtain the highest and midlevel controls for testing of all the analyte ranks. Each analyte was tested using 4 operators on 3 devices combined with 3 lots of reagents for a minimum of 20 days. Results for each level are presented in Table 3, Table 4 and Table 5.

Table 3. Repeatability and Reproducibility of AUTION MAX AX-4060 Urinalysis System Combined results for Level 1 (Negative/Normal) samples

Results of all combinations
Repeatability(N=120)Reproducibility(N=240)
AnalyteExpected ResultExactmatchWithin+/- 1 CBExactmatchWithin+/- 1 CB
GLUNormal100.0%100.0%100.0%100.0%
PRONegative100.0%100.0%100.0%100.0%
BILNegative100.0%100.0%100.0%100.0%
URONormal100.0%100.0%100.0%100.0%
pH5.099.2%100.0%99.6%100.0%
BLDNegative100.0%100.0%100.0%100.0%
KETNegative100.0%100.0%100.0%100.0%
NITNegative100.0%100.0%100.0%100.0%
LEUNegative100.0%100.0%100.0%100.0%
TURBNegative100.0%100.0%100.0%100.0%
ColorColorless100.0%100.0%100.0%100.0%

{13}------------------------------------------------

Results of all combinations
Expected ResultReproducibilityRepeatability(N=120)(N=240)
AnalyteQualitativeSemi-quantitative(mg/dL)ExactmatchWithin+/- 1 BCExactmatchWithin+/- 1 BC
GLU+3300-500100.0%100.0%100.0%100.0%
PRO+3300-60090.8%100.0%94.6%100.0%
BIL+36-10100.0%100.0%100.0%100.0%
URO+4>12100.0%100.0%95.8%100.0%
pH8.08.099.2%100.0%99.6%100.0%
BLD+20.2-0.5100.0%100.0%100.0%100.0%
KET+380-10098.3%100.0%99.2%100.0%
NIT+2100.0%100.0%100.0%100.0%
LEU>500 (Leu/μL)100.0%100.0%100.0%100.0%
TURB+2100.0%100.0%100.0%100.0%
Colordark yellow100.0%100.0%100.0%100.0%

Table 4. Repeatability and Reproducibility of AUTION MAX AX-4060 Urinalysis System Combined results for Level 2 (High) samples

Table 5. Repeatability and Reproducibility of AUTION MAX AX-4060 Urinalysis System Combined results for Level 3 (Mid) samples

Results of all combinations
Expected ResultRepeatability(N=120)Reproducibility(N=240)
AnalyteQualitativeSemi-quantitative(mg/dL)ExactmatchWithin+/- 1 BCExactmatchWithin+/- 1 BC
GLU+170-100100.0%100.0%100.0%100.0%
PRO+130-70100.0%100.0%100.0%100.0%
BIL+22-4100.0%100.0%100.0%100.0%
URO+24-699.2%100.0%99.2%99.6%
pH6.56.5100.0%100.0%100.0%100.0%
BLD+10.06-0.1100.0%100.0%100.0%100.0%
KET+110-2099.2%99.2%99.2%99.6%
NIT+1100.0%100.0%100.0%100.0%
LEU250 (Leu/μL)100.0%100.0%99.2%100.0%
TURB+1100.0%100.0%100.0%100.0%
Coloryellow100.0%100.0%100.0%100.0%

{14}------------------------------------------------

Method Comparison

A method comparison study was conducted to evaluate the agreement of results from AUTION MAX AX-4060 Urinalysis System to the FDA-cleared predicate device AUTION MAX AX-4030 Automated Urinalysis System. Testing was performed at two (2) clinical laboratories by using two (2) instruments and two (2) different lots of test strips. A total of 1374 samples, consisting of 1333 natural patient samples and 41 spiked samples, were tested across the sites.

The data from each site was combined and are presented in Table 6 showing percentages of Exact Match and within +/- 1 Color Block Match for all analytes on test strips. Non-test strip items for turbidity and color tone are in the same templates of Table 7 and Table 8 as the test strip analytes.

AnalyteAllSamples(n)ExactMatch(n)%ExactMatch95% CI% +/- 1CB Match
GLU1044102498.1%97.2% - 98.8%100.0%
PRO104899094.5%93.0% - 95.8%100.0%
BIL1338133199.5%99.0% - 99.8%100.0%
URO1341133099.2%98.6% - 99.6%100.0%
pH106398792.9%91.2% - 94.3%100.0%
BLD1044100095.8%94.5% - 96.9%100.0%
KET1336132299.0%98.3% - 99.4%100.0%
NIT1046104399.7%99.3% - 99.9%100.0%
LEU1044102998.6%97.8% - 99.2%100.0%

Table 6. Overall agreement of results for all test analytes between proposed and predicate device

Table 7. Summary table for turbidity

AnalyteAllSamples(n)Exact Match(n)% ExactMatch95% CI% +/- 1 CBMatch
Turbidity1044102998.6%97.8%-99.2%100%

Table 8. Summary table for color tone

AnalyteAll Samples (n)Exact Match (n)% Exact Match95% CI% +/- 1 CB Match
Color tone104486382.7%80.3% - 84.9%100.0%

{15}------------------------------------------------

Interfering Substances

Evaluation of interfering substances influencing the accuracy of results on AUTION MAX AX-4060 Urinalysis System was conducted with AUTION Sticks 9EB.

A list of interfering substances for a given analyte is provided in Table 9 below. The interfering substances and their highest concentrations tested with no interference are listed in Table 10 below.

{16}------------------------------------------------

AnalyteSubstanceConcentration limit with nosignificant interferenceEffect when above the concentration limit
Ascorbic Acid20 mg/dLAt 50 mg/dL decreased from +2 to +1
GlucoseUrobilinogen4 mg/dLAt 8 mg/dL decreased from +2 to +1
MESNA100 mg/dLAt 150 mg/dL increased from NORMAL to +/-
Sodium Bicarbonate200 mg/dLAt 250 mg/dL increased from +3 to +4
ProteinSpecific Gravity1.02At 1.025 decreased from +3 to +2
Hemoglobin10 mg/dLAt 12 mg/dL increased from - to +/-
UrobilinogenBilirubin3.5 mg/dLAt 7 mg/dL increased from NORMAL to +1
BloodAscorbic Acid50 mg/dLAt 100 mg/dL decreased from +2 to +1
MESNA50 mg/dLAt 100 mg/dL decreased from +2 to +1
Urine pHpH 8.5At pH 9 decreased from +2 to +/-
Methyl-dopa50 mg/dLAt 67 mg/dL increased from - to +/-
KetonesMESNA1.5 mg/dLAt 3 mg/dL increased from - to +/-
Ascorbic Acid50 mg/dLAt 100 mg/dL decreased from +2 to +1
NitriteMESNA150 mg/dLAt 200 mg/dL decreased from +2 to +1
Urobilinogen5 mg/dLAt 8 mg/dL decreased from +2 to +1
LeukocytesBilirubin7 mg/dLAt 10.5 mg/dL increased from negative to 25
Hemoglobin3 mg/dLAt 6.25 mg/dL decreased from 250 to 75
MESNA370 mg/dLAt 660 mg/dL decreased from 250 to 75

Table 9. Substances interfering with test analytes

Based on the interference testing, following limitations have been included in the labeling:

    1. Elevated hematuria (6.25 mg/dL Hb, approximately 2,156 RBC/uL) has been shown to cause falsely lowered Leukocyte readings and should be considered when interpreting Leukocyte results.
    1. Samples from patients taking methyl dopa may be falsely elevated for blood, negative to trace at 67 mg/dL and negative to +1 at 225 mg/dL. Patients taking the maximum daily dose of methyl dopa may have approximately 210 mg/dL methyl dopa excreted in the urine.
  • Patients taking SGLT2 or similar inhibitors may have extremely high urinary glucose values. Due to a 3. hook effect on the glucose pad, if a patient's urine glucose level is greater than 20,000 mg/dL, then falsely decreased glucose results may be obtained with the test.
  • When acidic or alkaline compounds, such as ascorbic acid and sodium bicarbonate are taken internally 4. for treatment or as supplements, they may affect urine pH.

{17}------------------------------------------------

Interfering SubstancesHighest Concentration Tested with No Interference**
Therapeutic drugs
Acetaminophen40 mg/dL
Acetoacetic Acid Lithium150 mg/dL
Ammonium Chloride400 mg/dL
Amoxicillin360 mg/dL
Calcium Chloride300 mg/dL
Citric Acid Monohydrate75 mg/dL
Oxalic Acid Dihydrate70 mg/dL
Potassium Chloride1000 mg/dL
Sodium Chloride2000 mg/dL
Sodium Dihydrogen Phosphate1000 mg/dL
Sodium Nitrate10 mg/dL
Sodium Nitrite2 mg/dL
Endogenous substances
Albumin1000 mg/dL
Creatinine600 mg/dL
Fructose100 mg/dL
Galactose100 mg/dL
Glucose2000 mg/dL
Glycine450 mg/dL
Lactose108 mg/dL
Low Urine pHpH 5
Riboflavin10 mg/dL
Urea3000 mg/dL

Table 10. Substances with no interference for the highest levels tested

** Not all of the substances tested for interference demonstrated no significant interference for all analytes. For those substances that on initial screening were found to interfere with certain analytes, dose-response testing was conducted to establish the concentration limit below which no significant interference is expected. The results are given in Table 10 above.

{18}------------------------------------------------

Limits of Detection and Sensitivity

Limits of detection and Minimum detection sensitivity were determined based on three instruments, and three reagent lots. Listed below in Table 11-14 are the quantitative values indicating the lower limit of the rank in which 50% or more of the results are falling into the specific rank.

AnalyteMinimum detection sensitivity
GLU30 mg/dL (+/-)
PRO10 mg/dL (+/-)
BIL0.5 mg/dL (+1)
URO2.0 mg/dL (+1)
BLD0.023 mg/dL (+/-)
KET4.0 mg/dL (+/-)
NIT0.075 mg/dL (+1)
LEU14 Leu/µL

Table 11. Minimum detection sensitivity

Table 12. Detection sensitivity for each Rank

BlockoutputSemi-quantitativerankEvaluationConc.(mg/dL)%Positive
GLU+/-30,5030100%
+170, 1006093%
+2150,20012597%
+3300,500250100%
+41000,over750100%
PRO+/-10, 201067%
+130, 703077%
+2100, 2008553%
+3300, 60025073%
+4over80093%
BIL+10.5, 1.00.553%
+22.0, 4.01.593%
+36.0, 10.0593%
+4over1293%
URO+12.0, 3.0260%

{19}------------------------------------------------

AUTION MAX AX-4060 Urinalysis System K193514

+24.0, 6.0480%
+38.0, 12.07100%
+4over14100%
BLD+/-0.030.023100%
+10.06, 0.100.045100%
+20.20, 0.500.1597%
+31.00, over0.7573%
KET+/-497%
+110, 207.5100%
+240, 6030100%
+380, 1007063%
+4150, over150100%
NIT+10.07557%
+20.3100%

Table 13. Detection sensitivity for each Rank of pH

BlockoutputSemi-quantitativerankEvaluationpH%Positive
pH5.55.55.3100%
665.8100%
6.56.56.2100%
776.893%
7.57.57.380%
888100%
8.58.58.5100%
998.887%

{20}------------------------------------------------

BlockoutputSemi-quantitativerankEvaluationConc.%Positive
LEU252514 cells/ μ L100%
757548 cells/ μ L70%
250250172 cells/ μ L60%
500500396 cells/ μ L90%

Linearity

Linearity was evaluated for Non-strip based analytical parameters. Tables 15 and 16 below respectively show the linearity for Color tone and Turbidity.

ColorExact match
COLORLESS100% (54/54)
BROWN100% (54/54)
ORANGE100% (54/54)
YELLOW100% (54/54)
GREEN100% (54/54)
BLUE100% (54/54)
VIOLET100% (54/54)
RED100% (54/54)

Table 15. Linearity for Color tone

Table 16. Linearity for Turbidity
-----------------------------------
AnalyteRankExact match
Turbidity-100% (60/60)
+1100% (60/60)
+2100% (60/60)

Proposed Labeling

Labeling adequately communicates the device's intended use, safety precautions and directions for use. It satisfies 21 CFR Part 809.10 for in vitro diagnostic devices.

Conclusion

The information provided in this 510(k) premarket notification supports that the AUTION MAX AX-4060 Urinalysis System is substantially equivalent to the AUTION MAX AX-4030 Fully-Automated Urinalysis System.

§ 862.1340 Urinary glucose (nonquantitative) test system.

(a)
Identification. A urinary glucose (nonquantitative) test system is a device intended to measure glucosuria (glucose in urine). Urinary glucose (nonquantitative) measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, hypoglycemia, and hyperglycemia.(b)
Classification. Class II.