(1211 days)
The AUTION MAX AX-4060 Urinalysis System (AUTION MAX) is comprised of the AUTION MAX AX-4060 automated urine analyzer and AUTION Sticks 9EB multi-parameter test strips.
The AUTION MAX AX-4060 urine analyzer, when used with AUTION Sticks 9EB test strips is a fully automated urinalysis system intended for the in vitro qualitative or semi-quantitative measurement of the following analytes: glucose, protein, bilirubin, urobilinogen, pH, blood, ketones, nitrite, leukocytes, turbidity, and color. The test results of these parameters can be used in the evaluation of kidney, urinary, liver and other metabolic disorders. This system is intended to be used by trained operators in clinical laboratories.
AUTION Sticks 9EB test strips are test strips for the in vitro qualitative or semi-quantitative measurement of the following analytes: glucose, protein, bilirubin, urobilinogen, pH, blood, ketones, nitrite, and leukocytes with the AUTION MAX AX-4060 urine analyzer. The test results of these parameters can be used in the evaluation of kidney, urinary, liver and other metabolic disorders.
Special conditions for use statements: Prescription use only. AUTION Sticks 9EB test strips are not to be read visually.
The AUTION MAX AX-4060 Urinalysis System is a fully automated urine analyzer that provides a semi-quantitative or qualitative measurement for glucose, protein, bilirubin, urobilinogen, pH, blood, ketones, nitrites, leukocytes, specific gravity, turbidity, and color tone. The AUTION MAX AX-4060 Urinalysis System consists of AUTION MAX AX-4060 urine analyzer and AUTION Sticks 9EB test strips.
The AUTION Sticks 9EB test strips consist of a plastic strip containing 9 dry chemistry reagent pads impregnated with chemical substances for the determination of the above analytes in urine. These substances if present in urine leads to a chemical reaction that results in a color change, which is measured by the AUTION MAX AX-4060 urine analyzer based on spectrophotometry. In addition, three additional parameters, specific gravity, turbidity, and color tone are directly measured based on reflectance refractometry, transmitted and scattered light measurements.
Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided FDA 510(k) summary:
Device: AUTION MAX AX-4060 Urinalysis System
Predicate Device: AUTION MAX AX-4030 Fully Automated Urinalysis System
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are not explicitly stated as distinct numerical targets in the document. Instead, the study aims to demonstrate substantial equivalence to the predicate device. Therefore, the "acceptance criteria" are implicitly understood as matching or performing comparably to the predicate device's established performance, particularly in terms of agreement percentage. The reported device performance is presented as the agreement between the proposed device (AUTION MAX AX-4060) and the predicate device (AUTION MAX AX-4030).
Key Performance Metric: Percentage of Exact Match and Percentage within +/- 1 Color Block Match compared to the predicate device.
| Analyte | Acceptance Criteria (Implicit: Substantial Equivalence to Predicate) | Reported Device Performance (Exact Match %) | Reported Device Performance (+/- 1 CB Match %) |
|---|---|---|---|
| GLU | High agreement (Exact and +/- 1 CB match) | 98.1% | 100.0% |
| PRO | High agreement (Exact and +/- 1 CB match) | 94.5% | 100.0% |
| BIL | High agreement (Exact and +/- 1 CB match) | 99.5% | 100.0% |
| URO | High agreement (Exact and +/- 1 CB match) | 99.2% | 100.0% |
| pH | High agreement (Exact and +/- 1 CB match) | 92.9% | 100.0% |
| BLD | High agreement (Exact and +/- 1 CB match) | 95.8% | 100.0% |
| KET | High agreement (Exact and +/- 1 CB match) | 99.0% | 100.0% |
| NIT | High agreement (Exact and +/- 1 CB match) | 99.7% | 100.0% |
| LEU | High agreement (Exact and +/- 1 CB match) | 98.6% | 100.0% |
| Turbidity | High agreement (Exact and +/- 1 CB match) | 98.6% | 100.0% |
| Color tone | High agreement (Exact and +/- 1 CB match) | 82.7% | 100.0% |
(Note: The +/- 1 CB Match indicates results falling within one color block difference from the predicate, which is often considered acceptable for semi-quantitative tests.)
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size for Test Set (Method Comparison): A total of 1374 samples were used. This included:
- 1333 natural patient samples
- 41 spiked samples
- Data Provenance: The study was conducted at two (2) clinical laboratories. The document does not specify the country of origin, but given the FDA submission, it's likely conducted in or in accordance with standards for the U.S. and/or Japan (where the manufacturer is located). It is a prospective comparison study.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
This study does not involve human experts establishing a "ground truth" for each sample in the typical sense of diagnostic imaging or pathology. Instead, the "ground truth" for the method comparison study is the result obtained from the legally marketed predicate device (AUTION MAX AX-4030 Automated Urinalysis System). The study demonstrates the correlation and agreement of the new device's readings with the predicate device's readings. No information is provided about experts interpreting results aside from the automated systems.
4. Adjudication Method for the Test Set
Since the "ground truth" is established by the predicate device's results and the comparison is between two automated systems, there was no human adjudication process (e.g., 2+1, 3+1 consensus) described for the test set.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No. A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not performed. This study evaluates the performance of an automated urinalysis system against a predicate automated system, not the improvement of human readers with AI assistance.
6. Standalone Performance (Algorithm Only without Human-in-the-Loop Performance)
Yes. The primary study detailed, especially the Method Comparison (Section 14), assesses the standalone performance of the AUTION MAX AX-4060 Urinalysis System. It compares the results generated by the new automated system directly against those generated by the predicate automated system, without human intervention in the result interpretation or decision-making process during the comparison phase. The device itself is an automated system intended to be used by trained operators in clinical laboratories, but the performance data presented is for the automated analyzer's output.
7. Type of Ground Truth Used
The ground truth for the method comparison study was the results obtained from the legally marketed predicate device (AUTION MAX AX-4030 Automated Urinalysis System). For the initial precision study (Tables 3-5), the "Expected Result" for quality controls served as the reference.
8. Sample Size for the Training Set
The document is a 510(k) summary for a medical device (urinalysis system), not a typically "AI-driven" software device that undergoes explicit training. The device determines analytes based on chemical reactions and spectrophotometry (reflectance refractometry, transmitted and scattered light measurements). Therefore, there is no explicit "training set" in the context of machine learning algorithms usually discussed. The system is likely calibrated and validated during its development, but this is a different process than training an AI model.
9. How the Ground Truth for the Training Set Was Established
As noted above, there is no explicit "training set" in the machine learning sense for this device. The chemical principles and measurement methods are well-established for urinalysis. Any calibration or internal development would have relied on known concentrations and reference methods for each analyte to ensure accurate colorimetric or other measurements. The basis of the technology relates to established chemical reactions on the test pads (e.g., Glucose oxidase reaction for glucose, Protein-error reaction for protein, etc.).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left side of the image is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
April 13, 2023
ARKRAY Inc. Daya Ranamukhaarachchi VP/Scientific and Regulatory Affairs 5198 West 76th Street Minneapolis, MN 55439
Re: K193514
Trade/Device Name: AUTION MAX AX-4060 Urinalysis System Regulation Number: 21 CFR 862.1340 Regulation Name: Urinary glucose (nonquantitative) test system Regulatory Class: Class II Product Code: JIL, JIO, KQO, CDM, JIR, JMT, LJX, JJB, CEN, JIN Dated: October 19, 2022 Received: October 20, 2022
Dear Daya Ranamukhaarachchi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR
{1}------------------------------------------------
- for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Paula V. Caposino -S
Paula Caposino, Ph.D. Acting Deputy Division Director Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K193514
Device Name AUTION MAX AX-4060 Urinalysis System
Indications for Use (Describe)
The AUTION MAX AX-4060 Urinalysis System (AUTION MAX) is comprised of the AUTION MAX AX-4060 automated urine analyzer and AUTION Sticks 9EB multi-parameter test strips.
The AUTION MAX AX-4060 urine analyzer, when used with AUTION Sticks 9EB test strips is a fully automated urinalysis system intended for the in vitro qualitative measurement of the following analytes: glucose, protein, bilirubin, urobilinogen, pH, blood, ketones, nitrite, leukocytes, turbidity, and color. The test results of these parameters can be used in the evaluation of kidney, urinary, liver and other metabolic disorders. This system is intended to be used by trained operators in clinical laboratories.
AUTION Sticks 9EB test strips are test strips for the in vitro qualitative or semi-quantitative measurement of the following analytes: glucose, protein, bilirubin, urobilinogen, pH, blood, ketones, nitrite, and leukocytes with the AUTION MAX AX-4060 urine analyzer. The test results of these parameters can be used in the evaluation of kidney, urinary, liver and other metabolic disorders.
Special conditions for use statements: Prescription use only. AUTION Sticks 9EB test strips are not to be read visually.
| Type of Use (Select one or both, as applicable) | ||
|---|---|---|
| Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) | Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
| Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
5. 510(K) Summary
The 510(k) Summary is included on the following pages.
{4}------------------------------------------------
510(k) Summarv
This 510(k) Summary of the AUTION MAX AX-4060 Urinalysis System is submitted in compliance with 21 CFR807.92 for the purposes of safety and effectiveness.
| Date Prepared: | October 19, 2022 |
|---|---|
| Establishment: | ARKRAY Factory Inc.1480 Koji, Konan-cho,Koka-shiShiga Japan, 520-3306 |
| Submitter: | ARKRAY Inc.Yousuien-nai, 59 Gansuin-cho, Kamigyo-kuKyoto Japan, 602-0008Official US Agent: Daya RanamukhaarachchiARKRAY Factory USA Inc.VP/Scientific and Regulatory AffairsPhone (408) 234-0804Fax: (952) 646-3240Email: ranamukd@arkrayusa.com |
| Primary Contact: | Daya RanamukhaarachchiVP/Scientific and Regulatory AffairsPhone (408) 234-0804Fax: (952) 646-3240Email: ranamukd@arkrayusa.com |
| Secondary Contact: | Dhwani ThakkarRegulatory Affairs Project ManagerEmail: thakkard@arkrayusa.comPhone: 437-247-1628 |
| Device Name: | |
| Trade Name: | 1. AUTION MAX AX-4060 Urinalysis System |
| Common Name: | 1. Automated Urinalysis System2. Urinalysis Test Strip |
{5}------------------------------------------------
Predicate Device:
Regulatory Information:
Regulatory Classification for Urinalysis system and each of the tests are listed in Table 1.
| Test Description | ProductCode | Device Class | Regulation | Device Panel |
|---|---|---|---|---|
| Automated urinalysissystem | KQO | Class I | 21 CFR 862.2900 | 75 Chemistry |
| Glucose, urinary,nonquantitative | JIL | Class II | 21 CFR 862.1340 | 75 Chemistry |
| Blood, occult,colorimetric, in urine | JIO | Class II | 21 CFR 864.6550 | 81 Hematology |
| Urobilinogen, urinary,non-quantitative | CDM | Class I | 21 CFR 862.1785 | 75 Chemistry |
| Protein or albumin,urinary, nonquantitative | JIR | Class I | 21 CFR 862.1645 | 75 Chemistry |
| Nitrite, nonquantitative | JMT | Class I | 21 CFR 862.1510 | 75 Chemistry |
| Leukocyte, peroxidasetest | LJX | Class I | 21 CFR 864.7675 | 81 Hematology |
| Bilirubin and itsconjugates, urinary,non-quantitative | JJB | Class I | 21 CFR 862.1115 | 75 Chemistry |
| pH, urinary,nonquantitative | CEN | Class I | 21 CFR 862.1550 | 75 Chemistry |
| Ketones nonquantitative | JIN | Class I | 21 CFR 862.1435 | 75 Chemistry |
| Table 1. Regulatory Classification for AUTION MAX AX-4060 Urinalysis System | ||
|---|---|---|
| -- | ----------------------------------------------------------------------------- | -- |
Intended Use
The AUTION MAX AX-4060 Urinalysis System (AUTION MAX) is comprised of the AUTION MAX AX-4060 automated urine analyzer and AUTION Sticks 9EB multi-parameter test strips.
The AUTION MAX AX-4060 urine analyzer, when used with AUTION Sticks 9EB test strips is a fully automated urinalysis system intended for the in vitro qualitative or semi-quantitative measurement of the following analytes: glucose, protein, bilirubin, urobilinogen, pH, blood, ketones, nitrite, leukocytes, turbidity, and color. The test results of these parameters can be used in the evaluation of kidney, urinary, liver and other metabolic disorders. This system is intended to be used by trained operators in clinical laboratories.
AUTION Sticks 9EB test strips are test strips for the in vitro qualitative or semi-quantitative
{6}------------------------------------------------
measurement of the following analytes: glucose, protein, bilirubin, urobilinogen, pH, blood, ketones, nitrite, and leukocytes with the AUTION MAX AX-4060 urine analyzer. The test results of these parameters can be used in the evaluation of kidney, urinary, liver and other metabolic disorders.
Special conditions for use statements: Prescription use only. AUTION Sticks 9EB test strips are not to be read visually.
Device Description
The AUTION MAX AX-4060 Urinalysis System is a fully automated urine analyzer that provides a semi-quantitative or qualitative measurement for glucose, protein, bilirubin, urobilinogen, pH, blood, ketones, nitrites, leukocytes, specific gravity, turbidity, and color tone. The AUTION MAX AX-4060 Urinalysis System consists of AUTION MAX AX-4060 urine analyzer and AUTION Sticks 9EB test strips.
The AUTION Sticks 9EB test strips consist of a plastic strip containing 9 dry chemistry reagent pads impregnated with chemical substances for the determination of the above analytes in urine. These substances if present in urine leads to a chemical reaction that results in a color change, which is measured by the AUTION MAX AX-4060 urine analyzer based on spectrophotometry. In addition, three additional parameters, specific gravity, turbidity, and color tone are directly measured based on reflectance refractometry, transmitted and scattered light measurements.
Substantial Equivalence
AUTION MAX AX-4060 Urinalysis System was developed as the successor to the predicate device AUTION MAX AX-4030 Urinalysis System. The device uses the same intendeduse claims as the predicate. The AX-4060 system utilizes the same principles and technology and shares the identical 9EB test strips for measuring the same urine analytes in a semiquantitative or qualitative manner similar to the predicate device. Some important device parts such as LED and Nozzle dispensing units were changed in the AX-4060 System due to a change in manufacturing source and the device was optimized to generate similar performance to the predicate device.
As described in the Performance Characteristics Section below, the verification and validation (bench and clinical) testing demonstrated substantial equivalence for the AX-4060 System with AUTION Sticks 9EB test strips to the predicate device. Bench testing supports that the AX-4060 System performs to meet specifications for precision, linearity, limits of detection (LOD), sensitivity, carryover and interference. Additionally, software testing, electrical safety testing and EMC testing also met performance criteria.
A list of similarities and differences between the proposed device and the predicate device is provided in Table 2.
{7}------------------------------------------------
| Table 2. Substantial Equivalence comparison between AUTION MAX AX-406 |
|---|
| Urinalysis System and AUTION MAX AX-4030 Urinalysis System |
| COMPONENT/CHARACTERISTIC | PROPOSED | PREDICATE |
|---|---|---|
| 510(k) Number | To Be Determined | K093098 |
| Device | AUTION MAX AX-4060 UrinalysisSystem | AUTION MAX AX-4030Urinalysis System |
| Similarities between Proposed and Predicate Devices | ||
| Device Type | Automated Urinalysis System | Same |
| FDA Product Code | KQO, JIL, JIO | Same |
| FDA Authorized Use | Prescription Use | Same |
| FDA Classification | Class II | Same |
| SpecimenType/Measurement Object | Urine | Same |
| Reagent(Chemical Test Strips) | AUTION Sticks 9EB | Same |
| Measurement Items | GLU (Glucose), PRO (Protein), BIL(Bilirubin), URO (Urobilinogen),PH (pH), BLD (Blood), KET(Ketones), NIT (Nitrite), LEU(Leukocytes), S.G. (Specificgravity), turbidity, and color tone. | Same |
| Sample Consumption | Max. 0.90 mL | Same |
| Required Sample Volume | Min. 2.0 mL | Same |
| Test Strip Reaction Time | Approx. 60 seconds | Same |
| Test Strip Storage | Two test strip storage compartments,each of which can contain differenttest strips | Same |
| Test Strip StorageCapacity | Maximum 200 test strips × twostorage compartments | Same |
| Max. Processing Speed | 225 samples/hour | Same |
| Built-in Printer | 58-mm width printer paper (24digits) | Same |
| COMPONENT/CHARACTERISTIC | PROPOSED | PREDICATE |
| Memory Capacity | Normal and STAT measurements:2500 testsControl measurement: 200 testsCheck measurement: 50 testsTrouble list: 100 tests | Same |
| Communication System | RS-232C compliant (Switchablebetween one-way and two-way) | Same |
| Power Requirements | 100 - 240 V AC (Maximum powerline fluctuation of ±10%), 50/60 Hz | Same |
| Power Input | Max. 150 VA | Same |
| Intended UseEnvironment | For indoor use only | Same |
| Measurement environment | Temperature: 10 – 30°C (50 to86°F); Humidity: 30 – 60% RH(No condensation) | Same |
| Altitude | Up to 2000m (6560 feet) | Same |
| Differences between Proposed and Predicate Devices | ||
| Intended Use andIndications for Use | The AUTION MAX AX-4060Urinalysis System (AUTION MAX) iscomprised of the AUTION MAX AX-4060 automated urine analyzer andAUTION Sticks 9EB multi-parametertest strips.The AUTION MAX AX-4060 urineanalyzer, when used with AUTIONSticks 9EB test strips is a fullyautomated urinalysis system intendedfor the in vitro qualitative or semi-quantitative measurement of thefollowing analytes: glucose, protein,bilirubin, urobilinogen, pH, blood,ketones, nitrite, leukocytes, turbidity,and color. The test results of theseparameters can be used in the | The AUTION MAX AX-4030 Urinalysis System(AUTION MAX) is anautomated urine analyzerintended for the in vitromeasurement of the followinganalytes: glucose, protein,bilirubin, urobilinogen, pH,blood, ketones, nitrite,leukocytes, specific gravity,turbidity, and color. TheAUTION MAX is intendedfor use only with AUTIONSticks 9EB multi-parametertest strips. |
| other metabolic disorders. This systemis intended to be used by trainedoperators in clinical laboratories. | ||
| AUTION Sticks 9EB test strips aretest strips for the in vitro qualitative orsemi-quantitative measurement of thefollowing analytes: glucose, protein,bilirubin, urobilinogen, pH, blood,ketones, nitrite, and leukocytes withthe AUTION MAX AX-4060 urineanalyzer. The test results of theseparameters can be used in theevaluation of kidney, urinary, liver andother metabolic disorders. | ||
| Special conditions for use statements:Prescription use only.AUTION Sticks 9EB test strips arenot to be read visually. | ||
| Configuration | Instrument, Sampler, Hand-heldbarcode reader and Accessories | Same, but the Hand-held barcode isoptional. |
{8}------------------------------------------------
{9}------------------------------------------------
{10}------------------------------------------------
| COMPONENT/CHARACTERISTIC | PROPOSED | PREDICATE |
|---|---|---|
| LED Wavelength | 5 LED wavelengths (450, 525,562, 635 and 760 nm) | 5 LED wavelengths (430,500, 565, 635, 760 nm) |
| Warm-up Time | Max. 4 minutes | Max. 2 minutes |
| Display | Large color liquid crystal display (800× 480 dots) with touch panel | Large color liquid crystaldisplay (320 × 240 dots)No touch panel |
| External Output | 1 port (Replaceable with theoptional Ethernet terminal) | 2 ports (One of these ports canbe optionally used as anEthernet port) |
| Transmission Speed | RS-232C: Selectable from 300,600, 1200, 2400, 4800, 9600 and19200 bpsEthernet: 10BASE-T, 100BASE-TX | RS-232C: Selectable from300, 600, 1200, 2400, 4800,9600, and 19200 bpsEthernet: 10BASE-T |
| Dimensions | 530 (W) × 530 (D) × 550 (H) mm(including the sampler) | 530 (W) × 530 (D) ×530(H) mm(including the sampler) |
| Weight | Instrument: Approx. 40 kg (88 lb)Sampler: Approx. 4 kg (9 lb) | Instrument: approx. 37 kgSampler: approx. 4 kg |
| Storage Environment | Temperature: -10 – 50°C (14 to122°F); Humidity: 20 – 80% RH(No condensation) | Temperature: 1 – 30°C (34 to86°F);Humidity: 20 – 80% RH(No condensation) |
| Environment duringTransport | Temperature: -10 – 50°C (14 to122°F); Humidity: 20 – 80% RH(No condensation) | Temperature: -13 – 43°C (9 to109°F); Humidity: 20 – 80%RH (No condensation) |
Test Principle:
The AUTION MAX AX-4060 urine analyzer measures analyte-specific test strips (AUTION Sticks 9EB) by dual-wavelength reflectance measurement (single wavelength for blood [BLD]). The AUTION MAX AX-4060 Urinalysis System detects urine analytes based on chemical reaction on each test pad relative to the concentration of measured analytes present in the urine sample. The chemical reaction leading to the color change that occurs on each chemical test pad is listed below.
{11}------------------------------------------------
| PRINCIPLE |
|---|
| ----------- |
| Glucose: Glucose oxidase reaction. |
|---|
| ------------------------------------ |
| Glucose | GOD Gluconic acid + H2O2 |
|---|---|
| H 2 O 2 + 4-AAP + 1-Naphthol-3,6-disulfonic acid | POD Quinone imine dye (purple color) oxidation |
| Protein: Protein-error reaction. |
|---|
| ---------------------------------- |
| Protein + pH indicator | acid pH indicator changes to a cyan color |
|---|---|
| ------------------------ | ------------------------------------------------------------------------------------------------------------------------------- |
| Bilirubin: Azo-coupling reaction. |
|---|
| ----------------------------------- |
| 2-Methyl-5-nitroaniline + Sodium nitrite | acid Diazonium salt diazo reaction |
|---|---|
| Bilirubin + Diazonium salt | acid Azo dye (reddish brown color) coupling reaction |
| Urobilinogen: Azo-coupling reaction. |
|---|
| -------------------------------------- |
| Urobilinogen + Diazonium salt | acid Azo dye (reddish brown color) coupling reaction |
|---|---|
| ------------------------------- | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ |
| pH: pH indicator. |
|---|
| ------------------- |
| H + + mixed pH indicator | mixed pH indicator shows range of colors covering the urinary pH range (yellow - cyan color) |
|---|---|
| ------------------------------------- | ---------------------------------------------------------------------------------------------- |
| Blood: Activity measurement of pseudoperoxidase in hemoglobin. |
|---|
| ---------------------------------------------------------------- |
| CHP + TMBZ | hemoglobin H2O + Cumene + Oxidation dye (cyan color) pseudoperoxidase action |
|---|---|
| ------------ | ----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------- |
| Ketones: Legal reaction. |
|---|
| -------------------------- |
| Ketones + Sodium nitroprusside | alkaline Ketones complex (purple color) |
|---|---|
| -------------------------------- | ----------------------------------------------------------------------------------------------------------------------------- |
| Nitrite: Griess reaction. |
|---|
| --------------------------- |
| Nitrite + Sulfanilamide | acid Diazo-compound + NEDA-2HCI |
|---|---|
| acid Azo dye (pink color) coupling reaction |
| Leukocytes: Measurement of leukocyte esterase activity. |
|---|
| --------------------------------------------------------- |
| TAI | esterase from leukocytes Indoxyl + MMB Azo dye (purple color) |
|---|---|
| hydrolysis coupling reaction |
{12}------------------------------------------------
Performance Characteristics
Precision
Precision study included repeatability and reproducibility testing in accordance with the CLSI standard EP5-A3. Three (3) levels of quality controls were used to attain results at negative, mid-level positive and high-positive analyte ranges. Two (2) levels of commercial controls were used to obtain the 3 levels by mixing as needed to obtain the highest and midlevel controls for testing of all the analyte ranks. Each analyte was tested using 4 operators on 3 devices combined with 3 lots of reagents for a minimum of 20 days. Results for each level are presented in Table 3, Table 4 and Table 5.
Table 3. Repeatability and Reproducibility of AUTION MAX AX-4060 Urinalysis System Combined results for Level 1 (Negative/Normal) samples
| Results of all combinations | |||||
|---|---|---|---|---|---|
| Repeatability(N=120) | Reproducibility(N=240) | ||||
| Analyte | Expected Result | Exactmatch | Within+/- 1 CB | Exactmatch | Within+/- 1 CB |
| GLU | Normal | 100.0% | 100.0% | 100.0% | 100.0% |
| PRO | Negative | 100.0% | 100.0% | 100.0% | 100.0% |
| BIL | Negative | 100.0% | 100.0% | 100.0% | 100.0% |
| URO | Normal | 100.0% | 100.0% | 100.0% | 100.0% |
| pH | 5.0 | 99.2% | 100.0% | 99.6% | 100.0% |
| BLD | Negative | 100.0% | 100.0% | 100.0% | 100.0% |
| KET | Negative | 100.0% | 100.0% | 100.0% | 100.0% |
| NIT | Negative | 100.0% | 100.0% | 100.0% | 100.0% |
| LEU | Negative | 100.0% | 100.0% | 100.0% | 100.0% |
| TURB | Negative | 100.0% | 100.0% | 100.0% | 100.0% |
| Color | Colorless | 100.0% | 100.0% | 100.0% | 100.0% |
{13}------------------------------------------------
| Results of all combinations | ||||||
|---|---|---|---|---|---|---|
| Expected Result | ReproducibilityRepeatability(N=120)(N=240) | |||||
| Analyte | Qualitative | Semi-quantitative(mg/dL) | Exactmatch | Within+/- 1 BC | Exactmatch | Within+/- 1 BC |
| GLU | +3 | 300-500 | 100.0% | 100.0% | 100.0% | 100.0% |
| PRO | +3 | 300-600 | 90.8% | 100.0% | 94.6% | 100.0% |
| BIL | +3 | 6-10 | 100.0% | 100.0% | 100.0% | 100.0% |
| URO | +4 | >12 | 100.0% | 100.0% | 95.8% | 100.0% |
| pH | 8.0 | 8.0 | 99.2% | 100.0% | 99.6% | 100.0% |
| BLD | +2 | 0.2-0.5 | 100.0% | 100.0% | 100.0% | 100.0% |
| KET | +3 | 80-100 | 98.3% | 100.0% | 99.2% | 100.0% |
| NIT | +2 | 100.0% | 100.0% | 100.0% | 100.0% | |
| LEU | >500 (Leu/μL) | 100.0% | 100.0% | 100.0% | 100.0% | |
| TURB | +2 | 100.0% | 100.0% | 100.0% | 100.0% | |
| Color | dark yellow | 100.0% | 100.0% | 100.0% | 100.0% |
Table 4. Repeatability and Reproducibility of AUTION MAX AX-4060 Urinalysis System Combined results for Level 2 (High) samples
Table 5. Repeatability and Reproducibility of AUTION MAX AX-4060 Urinalysis System Combined results for Level 3 (Mid) samples
| Results of all combinations | |||||||
|---|---|---|---|---|---|---|---|
| Expected Result | Repeatability(N=120) | Reproducibility(N=240) | |||||
| Analyte | Qualitative | Semi-quantitative(mg/dL) | Exactmatch | Within+/- 1 BC | Exactmatch | Within+/- 1 BC | |
| GLU | +1 | 70-100 | 100.0% | 100.0% | 100.0% | 100.0% | |
| PRO | +1 | 30-70 | 100.0% | 100.0% | 100.0% | 100.0% | |
| BIL | +2 | 2-4 | 100.0% | 100.0% | 100.0% | 100.0% | |
| URO | +2 | 4-6 | 99.2% | 100.0% | 99.2% | 99.6% | |
| pH | 6.5 | 6.5 | 100.0% | 100.0% | 100.0% | 100.0% | |
| BLD | +1 | 0.06-0.1 | 100.0% | 100.0% | 100.0% | 100.0% | |
| KET | +1 | 10-20 | 99.2% | 99.2% | 99.2% | 99.6% | |
| NIT | +1 | 100.0% | 100.0% | 100.0% | 100.0% | ||
| LEU | 250 (Leu/μL) | 100.0% | 100.0% | 99.2% | 100.0% | ||
| TURB | +1 | 100.0% | 100.0% | 100.0% | 100.0% | ||
| Color | yellow | 100.0% | 100.0% | 100.0% | 100.0% |
{14}------------------------------------------------
Method Comparison
A method comparison study was conducted to evaluate the agreement of results from AUTION MAX AX-4060 Urinalysis System to the FDA-cleared predicate device AUTION MAX AX-4030 Automated Urinalysis System. Testing was performed at two (2) clinical laboratories by using two (2) instruments and two (2) different lots of test strips. A total of 1374 samples, consisting of 1333 natural patient samples and 41 spiked samples, were tested across the sites.
The data from each site was combined and are presented in Table 6 showing percentages of Exact Match and within +/- 1 Color Block Match for all analytes on test strips. Non-test strip items for turbidity and color tone are in the same templates of Table 7 and Table 8 as the test strip analytes.
| Analyte | AllSamples(n) | ExactMatch(n) | %ExactMatch | 95% CI | % +/- 1CB Match |
|---|---|---|---|---|---|
| GLU | 1044 | 1024 | 98.1% | 97.2% - 98.8% | 100.0% |
| PRO | 1048 | 990 | 94.5% | 93.0% - 95.8% | 100.0% |
| BIL | 1338 | 1331 | 99.5% | 99.0% - 99.8% | 100.0% |
| URO | 1341 | 1330 | 99.2% | 98.6% - 99.6% | 100.0% |
| pH | 1063 | 987 | 92.9% | 91.2% - 94.3% | 100.0% |
| BLD | 1044 | 1000 | 95.8% | 94.5% - 96.9% | 100.0% |
| KET | 1336 | 1322 | 99.0% | 98.3% - 99.4% | 100.0% |
| NIT | 1046 | 1043 | 99.7% | 99.3% - 99.9% | 100.0% |
| LEU | 1044 | 1029 | 98.6% | 97.8% - 99.2% | 100.0% |
Table 6. Overall agreement of results for all test analytes between proposed and predicate device
Table 7. Summary table for turbidity
| Analyte | AllSamples(n) | Exact Match(n) | % ExactMatch | 95% CI | % +/- 1 CBMatch |
|---|---|---|---|---|---|
| Turbidity | 1044 | 1029 | 98.6% | 97.8%-99.2% | 100% |
Table 8. Summary table for color tone
| Analyte | All Samples (n) | Exact Match (n) | % Exact Match | 95% CI | % +/- 1 CB Match |
|---|---|---|---|---|---|
| Color tone | 1044 | 863 | 82.7% | 80.3% - 84.9% | 100.0% |
{15}------------------------------------------------
Interfering Substances
Evaluation of interfering substances influencing the accuracy of results on AUTION MAX AX-4060 Urinalysis System was conducted with AUTION Sticks 9EB.
A list of interfering substances for a given analyte is provided in Table 9 below. The interfering substances and their highest concentrations tested with no interference are listed in Table 10 below.
{16}------------------------------------------------
| Analyte | Substance | Concentration limit with nosignificant interference | Effect when above the concentration limit |
|---|---|---|---|
| Ascorbic Acid | 20 mg/dL | At 50 mg/dL decreased from +2 to +1 | |
| Glucose | Urobilinogen | 4 mg/dL | At 8 mg/dL decreased from +2 to +1 |
| MESNA | 100 mg/dL | At 150 mg/dL increased from NORMAL to +/- | |
| Sodium Bicarbonate | 200 mg/dL | At 250 mg/dL increased from +3 to +4 | |
| Protein | Specific Gravity | 1.02 | At 1.025 decreased from +3 to +2 |
| Hemoglobin | 10 mg/dL | At 12 mg/dL increased from - to +/- | |
| Urobilinogen | Bilirubin | 3.5 mg/dL | At 7 mg/dL increased from NORMAL to +1 |
| Blood | Ascorbic Acid | 50 mg/dL | At 100 mg/dL decreased from +2 to +1 |
| MESNA | 50 mg/dL | At 100 mg/dL decreased from +2 to +1 | |
| Urine pH | pH 8.5 | At pH 9 decreased from +2 to +/- | |
| Methyl-dopa | 50 mg/dL | At 67 mg/dL increased from - to +/- | |
| Ketones | MESNA | 1.5 mg/dL | At 3 mg/dL increased from - to +/- |
| Ascorbic Acid | 50 mg/dL | At 100 mg/dL decreased from +2 to +1 | |
| Nitrite | MESNA | 150 mg/dL | At 200 mg/dL decreased from +2 to +1 |
| Urobilinogen | 5 mg/dL | At 8 mg/dL decreased from +2 to +1 | |
| Leukocytes | Bilirubin | 7 mg/dL | At 10.5 mg/dL increased from negative to 25 |
| Hemoglobin | 3 mg/dL | At 6.25 mg/dL decreased from 250 to 75 | |
| MESNA | 370 mg/dL | At 660 mg/dL decreased from 250 to 75 |
Table 9. Substances interfering with test analytes
Based on the interference testing, following limitations have been included in the labeling:
-
- Elevated hematuria (6.25 mg/dL Hb, approximately 2,156 RBC/uL) has been shown to cause falsely lowered Leukocyte readings and should be considered when interpreting Leukocyte results.
-
- Samples from patients taking methyl dopa may be falsely elevated for blood, negative to trace at 67 mg/dL and negative to +1 at 225 mg/dL. Patients taking the maximum daily dose of methyl dopa may have approximately 210 mg/dL methyl dopa excreted in the urine.
- Patients taking SGLT2 or similar inhibitors may have extremely high urinary glucose values. Due to a 3. hook effect on the glucose pad, if a patient's urine glucose level is greater than 20,000 mg/dL, then falsely decreased glucose results may be obtained with the test.
- When acidic or alkaline compounds, such as ascorbic acid and sodium bicarbonate are taken internally 4. for treatment or as supplements, they may affect urine pH.
{17}------------------------------------------------
| Interfering Substances | Highest Concentration Tested with No Interference** |
|---|---|
| Therapeutic drugs | |
| Acetaminophen | 40 mg/dL |
| Acetoacetic Acid Lithium | 150 mg/dL |
| Ammonium Chloride | 400 mg/dL |
| Amoxicillin | 360 mg/dL |
| Calcium Chloride | 300 mg/dL |
| Citric Acid Monohydrate | 75 mg/dL |
| Oxalic Acid Dihydrate | 70 mg/dL |
| Potassium Chloride | 1000 mg/dL |
| Sodium Chloride | 2000 mg/dL |
| Sodium Dihydrogen Phosphate | 1000 mg/dL |
| Sodium Nitrate | 10 mg/dL |
| Sodium Nitrite | 2 mg/dL |
| Endogenous substances | |
| Albumin | 1000 mg/dL |
| Creatinine | 600 mg/dL |
| Fructose | 100 mg/dL |
| Galactose | 100 mg/dL |
| Glucose | 2000 mg/dL |
| Glycine | 450 mg/dL |
| Lactose | 108 mg/dL |
| Low Urine pH | pH 5 |
| Riboflavin | 10 mg/dL |
| Urea | 3000 mg/dL |
Table 10. Substances with no interference for the highest levels tested
** Not all of the substances tested for interference demonstrated no significant interference for all analytes. For those substances that on initial screening were found to interfere with certain analytes, dose-response testing was conducted to establish the concentration limit below which no significant interference is expected. The results are given in Table 10 above.
{18}------------------------------------------------
Limits of Detection and Sensitivity
Limits of detection and Minimum detection sensitivity were determined based on three instruments, and three reagent lots. Listed below in Table 11-14 are the quantitative values indicating the lower limit of the rank in which 50% or more of the results are falling into the specific rank.
| Analyte | Minimum detection sensitivity |
|---|---|
| GLU | 30 mg/dL (+/-) |
| PRO | 10 mg/dL (+/-) |
| BIL | 0.5 mg/dL (+1) |
| URO | 2.0 mg/dL (+1) |
| BLD | 0.023 mg/dL (+/-) |
| KET | 4.0 mg/dL (+/-) |
| NIT | 0.075 mg/dL (+1) |
| LEU | 14 Leu/µL |
Table 11. Minimum detection sensitivity
Table 12. Detection sensitivity for each Rank
| Blockoutput | Semi-quantitativerank | EvaluationConc.(mg/dL) | %Positive | |
|---|---|---|---|---|
| GLU | +/- | 30,50 | 30 | 100% |
| +1 | 70, 100 | 60 | 93% | |
| +2 | 150,200 | 125 | 97% | |
| +3 | 300,500 | 250 | 100% | |
| +4 | 1000,over | 750 | 100% | |
| PRO | +/- | 10, 20 | 10 | 67% |
| +1 | 30, 70 | 30 | 77% | |
| +2 | 100, 200 | 85 | 53% | |
| +3 | 300, 600 | 250 | 73% | |
| +4 | over | 800 | 93% | |
| BIL | +1 | 0.5, 1.0 | 0.5 | 53% |
| +2 | 2.0, 4.0 | 1.5 | 93% | |
| +3 | 6.0, 10.0 | 5 | 93% | |
| +4 | over | 12 | 93% | |
| URO | +1 | 2.0, 3.0 | 2 | 60% |
{19}------------------------------------------------
AUTION MAX AX-4060 Urinalysis System K193514
| +2 | 4.0, 6.0 | 4 | 80% | |
|---|---|---|---|---|
| +3 | 8.0, 12.0 | 7 | 100% | |
| +4 | over | 14 | 100% | |
| BLD | +/- | 0.03 | 0.023 | 100% |
| +1 | 0.06, 0.10 | 0.045 | 100% | |
| +2 | 0.20, 0.50 | 0.15 | 97% | |
| +3 | 1.00, over | 0.75 | 73% | |
| KET | +/- | 4 | 97% | |
| +1 | 10, 20 | 7.5 | 100% | |
| +2 | 40, 60 | 30 | 100% | |
| +3 | 80, 100 | 70 | 63% | |
| +4 | 150, over | 150 | 100% | |
| NIT | +1 | 0.075 | 57% | |
| +2 | 0.3 | 100% |
Table 13. Detection sensitivity for each Rank of pH
| Blockoutput | Semi-quantitativerank | EvaluationpH | %Positive | |
|---|---|---|---|---|
| pH | 5.5 | 5.5 | 5.3 | 100% |
| 6 | 6 | 5.8 | 100% | |
| 6.5 | 6.5 | 6.2 | 100% | |
| 7 | 7 | 6.8 | 93% | |
| 7.5 | 7.5 | 7.3 | 80% | |
| 8 | 8 | 8 | 100% | |
| 8.5 | 8.5 | 8.5 | 100% | |
| 9 | 9 | 8.8 | 87% |
{20}------------------------------------------------
| Blockoutput | Semi-quantitativerank | EvaluationConc. | %Positive | |
|---|---|---|---|---|
| LEU | 25 | 25 | 14 cells/ μ L | 100% |
| 75 | 75 | 48 cells/ μ L | 70% | |
| 250 | 250 | 172 cells/ μ L | 60% | |
| 500 | 500 | 396 cells/ μ L | 90% |
Linearity
Linearity was evaluated for Non-strip based analytical parameters. Tables 15 and 16 below respectively show the linearity for Color tone and Turbidity.
| Color | Exact match |
|---|---|
| COLORLESS | 100% (54/54) |
| BROWN | 100% (54/54) |
| ORANGE | 100% (54/54) |
| YELLOW | 100% (54/54) |
| GREEN | 100% (54/54) |
| BLUE | 100% (54/54) |
| VIOLET | 100% (54/54) |
| RED | 100% (54/54) |
Table 15. Linearity for Color tone
| Table 16. Linearity for Turbidity |
|---|
| ----------------------------------- |
| Analyte | Rank | Exact match |
|---|---|---|
| Turbidity | - | 100% (60/60) |
| +1 | 100% (60/60) | |
| +2 | 100% (60/60) |
Proposed Labeling
Labeling adequately communicates the device's intended use, safety precautions and directions for use. It satisfies 21 CFR Part 809.10 for in vitro diagnostic devices.
Conclusion
The information provided in this 510(k) premarket notification supports that the AUTION MAX AX-4060 Urinalysis System is substantially equivalent to the AUTION MAX AX-4030 Fully-Automated Urinalysis System.
§ 862.1340 Urinary glucose (nonquantitative) test system.
(a)
Identification. A urinary glucose (nonquantitative) test system is a device intended to measure glucosuria (glucose in urine). Urinary glucose (nonquantitative) measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, hypoglycemia, and hyperglycemia.(b)
Classification. Class II.