(62 days)
Swan-Ganz flow-directed monitoring catheters are indicated for the assessment of a patient's hemodynamic condition through direct intracardiac and pulmonary artery pressure monitoring. Secondary indications are for sampling blood and infusing solutions.
Swan-Ganz thermodilution catheters are indicated for the assessment of a patient's hemodynamic condition through direct intracardiac and pulmonary artery pressure monitoring, cardiac output determination, and for infusing solutions. The distal (pulmonary artery) port also allows sampling of mixed venous blood for the assessment of oxygen transport balance and the calculation of derived parameters such as oxygen consumption, oxygen utilization coefficient, and intrapulmonary shunt fraction.
ControlCath thermodilution catheters and Torque Support Thermodilution catheters are indicated for the assessment of a patient's hemodynamic condition through direct intracardiac and pulmonary artery pressure monitoring, cardiac output determination, and for infusing solutions. The distal (pulmonary artery) port also allows sampling of mixed venous blood for the assessment of oxygen transport balance and the calculation of derived parameters such as oxygen consumption, oxygen utilization coefficient, and intrapulmonary shunt fraction.
Swan-Ganz oximetry TD catheters are indicated for the assessment of a patient's hemodynamic condition through direct intracardiac and pulmonary artery pressure monitoring, cardiac output determination, continuous mixed venous oxygen saturation monitoring, and for infusing solutions. For 631NX55, the distal (pulmonary artery) port also allows sampling of mixed venous blood for the assessment of oxygen transport balance and the calculation of derived parameters such as oxygen consumption, oxygen utilization coefficient, and intrapulmonary shunt fraction.
The Swan-Ganz catheters are cardiovascular diagnostic catheters intended for use on critical care patients. Swan-Ganz catheters are used to monitor the hemodynamic status of critically ill and injured patients. The catheters give clinicians the ability to measure right heart pressures, pulmonary artery occlusion pressure ("wedge"), sample mixed venous blood from the pulmonary artery, as well as measure cardiac output through thermodilution when used with a bedside physiologic monitor.
The subject Edwards Swan-Ganz product line is comprised of Base and Advanced Technology monitoring catheters. The Base Swan-Ganz catheter measures: right heart pressures, thermodilution cardiac output (room temperature and chilled) and provides a mechanism for pulmonary artery blood sampling for laboratory analysis. In addition to providing most of the same functionality as the Base models, the Advanced Swan-Ganz catheters provide the ability to continuously monitor the patient's mixed venous oxygen saturation, the balance between oxygen delivery and consumption, when used with an Edwards Lifesciences oximetry monitor or compatible bedside module system. It also monitors cardiac output with a compatible cardiac output computer.
This is a 510(k) premarket notification for a Class II medical device, specifically Swan-Ganz Flow-Directed Monitoring Catheters. The document is about demonstrating substantial equivalence to a legally marketed predicate device, not about proving the performance of a novel AI/ML device against specific acceptance criteria in a clinical study. As such, most of the requested information (e.g., sample size for test set, number of experts for ground truth, MRMC study, standalone performance) is not applicable or present in this document.
The "acceptance criteria" discussed here are related to demonstrating that the modified device (with SEBS balloons) performs equivalently to the predicate device (with latex balloons) in terms of its functional and safety aspects.
Here's the information that can be extracted or deduced from the provided text:
1. A table of acceptance criteria and the reported device performance:
| Acceptance Criteria (Evaluation Areas) | Reported Device Performance |
|---|---|
| Functional Performance | Passed (performance not affected) |
| Biocompatibility | Passed (in accordance with ISO 10993-1:2018 and FDA guidance) |
| Sterilization | Passed (in accordance with ISO 11135:2014 and FDA Quality System Regulation) |
| Usability and Human Factors | Passed (in accordance with IEC 62366:2015 and FDA guidance) |
| Overall Substantial Equivalence | Demonstrated to be substantially equivalent to the predicate device for intended use. |
2. Sample size used for the test set and the data provenance:
- Sample Size: Not specified for any of the tests. The document refers to "functional testing of the catheter balloon," "biocompatibility testing," "sterilization testing," and "usability and human factors testing" but does not provide quantities for the test articles or subjects involved in these tests.
- Data Provenance: Not explicitly stated. The tests are bench studies and laboratory tests related to device manufacturing and biological effects, not clinical data from patients.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not applicable. This information is typically relevant for studies involving human interpretation or subjective assessment of medical images or conditions where expert consensus is needed to establish ground truth. The tests described (functional, biocompatibility, sterilization, usability) are objective engineering and biological assessments.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:
- Not applicable. Adjudication methods are used in clinical studies, particularly for diagnostic devices, to resolve disagreements among expert readers when establishing ground truth. The tests mentioned are objective evaluations and do not involve human adjudication in this context.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No, an MRMC comparative effectiveness study was not done. This type of study is relevant for AI/ML diagnostic devices. The device described is a medical catheter and the submission focuses on demonstrating substantial equivalence of a modified material.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- Not applicable. This submission is not for an AI/ML algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- Objective Test Standards and Regulations: The "ground truth" for the performance claims comes from objective engineering and biological standards and regulations.
- Functional testing: Likely based on pre-defined engineering specifications for balloon performance.
- Biocompatibility: ISO 10993-1:2018 and FDA guidance.
- Sterilization: ISO 11135:2014 and FDA Quality System Regulation (21 CFR 820.30).
- Usability and Human Factors: IEC 62366:2015 and FDA guidance.
8. The sample size for the training set:
- Not applicable. This submission is not for an AI/ML device that requires a training set.
9. How the ground truth for the training set was established:
- Not applicable. As above, no training set is relevant for this device type.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
February 16, 2020
Edwards Lifesciences, LLC Anne Lo Specialist. Regulatory Affairs One Edwards Way Irvine, California 92614
Re: K193466
Trade/Device Name: Swan-Ganz Flow-Directed Monitoring Catheters- Double and Triple Lumen
Swan-Ganz- Thermodilution, True Size
Swan-Ganz Polymer Blend True Size ControlCath Thermodilution catheters, and Polymer Blend True Size Torque Support Thermodilution catheter
Swan-Ganz Oximetry TD catheter
Regulation Number: 21 CFR 870.1240 Regulation Name: Flow-Directed Catheter Regulatory Class: Class II Product Code: DYG, DQE, DQO Dated: December 12, 2019 Received: December 16, 2019
Dear Anne Lo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
{1}------------------------------------------------
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Jessica Paulsen Division Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
510(k) Number: K193466
Device Name: Swan-Ganz Flow-Directed Monitoring catheters- Double and Triple Lumen
Models: 111NX7P, S111NX7, T111NX7P, 123NX6P, T123NX6, 110NX5P, 114NX7P
Indications For Use:
Swan-Ganz flow-directed monitoring catheters are indicated for the assessment of a patient's hemodynamic condition through direct intracardiac and pulmonary artery pressure monitoring. Secondary indications are for sampling blood and infusing solutions.
Device Name: Swan-Ganz- Thermodilution Catheters, True Size
Models: 131NX7P,131VNX7P,141NX7P,143TNX7,151NX7,096NX6P, TS105NX5, 132NX5
Indications For Use:
Swan-Ganz thermodilution catheters are indicated for the assessment of a patient's hemodynamic condition through direct intracardiac and pulmonary artery pressure monitoring, cardiac output determination, and for infusing solutions.
The distal (pulmonary artery) port also allows sampling of mixed venous blood for the assessment of oxygen transport balance and the calculation of derived parameters such as oxygen consumption, oxygen utilization coefficient, and intrapulmonary shunt fraction.
Device Name: Swan-Ganz
Polymer Blend True Size ControlCath Thermodilution catheters, and Polymer Blend True Size Torque Support Thermodilution catheter
Models: C144NX7, S144NX7, C145NX6, T173NX6
Indications For Use:
ControlCath thermodilution catheters and Torque Support Thermodilution catheters are indicated for the assessment of a patient's hemodynamic condition through direct intracardiac and pulmonary artery pressure monitoring, cardiac output determination, and for infusing solutions.
The distal (pulmonary artery) port also allows sampling of mixed venous blood for the assessment of oxygen transport balance and the calculation of derived parameters such as oxygen consumption, oxygen utilization coefficient, and intrapulmonary shunt fraction.
{3}------------------------------------------------
Device Name: Swan-Ganz Oximetry TD catheter
Model: 631NX55
Indications For Use:
Swan-Ganz oximetry TD catheters are indicated for the assessment of a patient's hemodynamic condition through direct intracardiac and pulmonary artery pressure monitoring, cardiac output determination, continuous mixed venous oxygen saturation monitoring, and for infusing solutions.
For 631NX55, the distal (pulmonary artery) port also allows sampling of mixed venous blood for the assessment of oxygen transport balance and the calculation of derived parameters such as oxygen consumption, oxygen utilization coefficient, and intrapulmonary shunt fraction.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
{4}------------------------------------------------
K193466 - 510(k) SUMMARY
| Swan-Ganz Base and Advanced catheters (5F-7F size) with SEBS balloons | |
|---|---|
| 510(k) Submitter | Edwards Lifesciences, LLC |
| Contact Person | Anne LoSpecialist, Regulatory AffairsEdwards LifesciencesOne Edwards WayIrvine, CA 92614Tel: (949) 250–3386Fax: (949) 809–7779 |
| Date Prepared | December 12, 2019 |
| Trade Name | Swan-Ganz Flow-Directed Monitoring catheters- Double and TripleLumenSwan-Ganz- Thermodilution Catheters, True SizeSwan-GanzPolymer Blend True Size ControlCath Thermodilution catheters, andPolymer Blend True Size Torque Support Thermodilution catheterSwan-Ganz Oximetry TD catheter |
| Common Name | Cardiovascular Diagnostic Catheters |
| ClassificationName | Flow-directed catheter(21 CFR 870.1240)Fiberoptic oximeter catheter(21 CFR 870.1230)Diagnostic intravascular catheter(21 CFR 870.1200) |
| Regulation Class /Product Code | Class IIDYG, DQE, DQO |
| Primary PredicateDevice | K160084- Swan Ganz catheters |
| ReferencePredicate Device | K001063- Swan-Ganz Synthetic ControlCath Thermodilution catheters |
| Device Description | The Swan-Ganz catheters are cardiovascular diagnostic cathetersintended for use on critical care patients. Swan-Ganz catheters are usedto monitor the hemodynamic status of critically ill and injured patients.The catheters give clinicians the ability to measure right heart pressures,pulmonary artery occlusion pressure ("wedge"), sample mixed venous |
| IndicationsforUse/Intended Use | blood from the pulmonary artery, as well as measure cardiac output through thermodilution when used with a bedside physiologic monitor. |
| The subject Edwards Swan-Ganz product line is comprised of Base and Advanced Technology monitoring catheters. The Base Swan-Ganz catheter measures: right heart pressures, thermodilution cardiac output (room temperature and chilled) and provides a mechanism for pulmonary artery blood sampling for laboratory analysis. In addition to providing most of the same functionality as the Base models, the Advanced Swan-Ganz catheters provide the ability to continuously monitor the patient's mixed venous oxygen saturation, the balance between oxygen delivery and consumption, when used with an Edwards Lifesciences oximetry monitor or compatible bedside module system. It also monitors cardiac output with a compatible cardiac output computer. | |
| The intended use for the Base Swan Ganz catheters are identical:The Swan-Ganz Catheters are cardiovascular catheters intended for use on critical care patients. | |
| The Indications for Use are slightly different among the different Base models: | |
| Swan Ganz Flow-Directed Monitoring catheters- Double and Triple LumenSwan-Ganz flow-directed monitoring catheters are indicated for the assessment of a patient's hemodynamic condition through direct intracardiac and pulmonary artery pressure monitoring. Secondary indications are for sampling blood and infusing solutions. | |
| Swan-Ganz Thermodilution Catheters, True SizeSwan-Ganz thermodilution catheters are indicated for the assessment of a patient's hemodynamic condition through direct intracardiac and pulmonary artery pressure monitoring, cardiac output determination, and for infusing solutions. The distal (pulmonary artery) port also allows sampling of mixed venous blood for the assessment of oxygen transport balance and the calculation of derived parameters such as oxygen consumption, oxygen utilization coefficient, and intrapulmonary shunt fraction. | |
| Swan-GanzPolymer Blend True Size ControlCath Thermodilution catheters,and Polymer Blend True Size Torque Support ThermodilutioncatheterControlCath thermodilution catheters and Torque SupportThermodilution catheters are indicated for the assessment of apatient's hemodynamic condition through direct intracardiacand pulmonary artery pressure monitoring, cardiac outputdetermination, and for infusing solutions.The distal (pulmonary artery) port also allows sampling ofmixed venous blood for the assessment of oxygen transportbalance and the calculation of derived parameters such asoxygen consumption, oxygen utilization coefficient, andintrapulmonary shunt fraction. | |
| The intended use for the Advanced Swan Ganz catheters is the sameas the Base Swan Ganz catehters with the addition of hemodynamicmonitoring capabilities.Swan-Ganz catheters are intended to be used with compatible criticalcare patient monitors, for monitoring the hemodynamic condition of acritically ill patient. | |
| Swan-Ganz Oximetry TD catheterSwan-Ganz oximetry TD catheters are indicated for theassessment of a patient's hemodynamic condition through directintracardiac and pulmonary artery pressure monitoring, cardiacoutput determination, continuous mixed venous oxygensaturation monitoring, and for infusing solutions.For 631NX55, the distal (pulmonary artery) port also allowssampling of mixed venous blood for the assessment of oxygentransport balance and the calculation of derived parameters suchas oxygen consumption, oxygen utilization coefficient, andintrapulmonary shunt fraction. | |
| ComparativeAnalysis | Technological Modifications:• Change balloon material from latex to a non-latex StyreneEthylene Butylene Styrene (SEBS) material. |
| Additional device modifications were made that complement the changein balloon material to SEBS as listed below:• Add an adhesive primer to the balloon bonding process to ensureballoon bond strength.• Modify the tip dimension and balloon bonding process for model110NX5P to be the same as other 5F and 6F catheters. | |
| Modify the inflation volume of the 5F and 6F catheters to maintain the same inflation diameter as the predicate devices. No changes were made to the catheter body; the device modifications are limited to the balloon and catheter tip. Labeling Modifications: IFU modifications as a result of the technological modifications made to the subject device. Updates to labeling that are administrative. Updates which provide consistency across subject device labeling and clarity for users. | |
| Functional/ SafetyTesting | The Swan-Ganz Base and Advanced catheters (5F-7F size) with SEBS balloons are identical to the predicate device in terms of intended use and indications for use. To ensure that the Swan-Ganz Base and Advanced catheters (5F-7F size) with SEBS balloons are substantially equivalent to the predicate devices, the following testing was performed: Functional testing of the catheter balloon to show performance was not affected. Biocompatibility testing in accordance to ISO 10993-1:2018 and the FDA guidance document Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process. Sterilization testing in accordance to ISO 11135:2014 Sterilization of health-care products- Ethylene oxide- Requirements for the development, validation and routing control of a sterilization process for medical devices and the FDA Quality System Regulation, 21CFR820.30. Usability and human factors testing in accordance to IEC 62366:2015 and the FDA guidance document, Applying Human Factors and Usability Engineering to Medical Devices. The Swan-Ganz Base and Advanced catheters (5F-7F size) with SEBS balloons have successfully passed functional, biocompatibility testing, usability and human factors, performance testing and bench studies, demonstrating that the subject device is substantially equivalent to the predicate device. |
| Conclusion | The Swan-Ganz Base and Advanced catheters (5F-7F size) with SEBS balloons have been shown to be substantially equivalent to the predicate device for their intended use. |
{5}------------------------------------------------
{6}------------------------------------------------
{7}------------------------------------------------
§ 870.1240 Flow-directed catheter.
(a)
Identification. A flow-directed catheter is a device that incorporates a gas-filled balloon to help direct the catheter to the desired position.(b)
Classification. Class II (performance standards).