(190 days)
The FreeStyle Libre 2 Flash Glucose Monitoring System is a continuous glucose monitoring (CGM) device with real time alarms capability indicated for the management of diabetes in persons age 4 and older. It is intended to replace blood glucose testing for diabetes treatment decisions, unless otherwise indicated.
The System also detects trends and tracks patterns and aids in the detection of episodes of hyperglycemia and hypoglycemia, facilitating both acute and long-term therapy adjustments. Interpretation of the System readings should be based on the glucose trends and several sequential readings over time.
The System is also intended to autonomously communicate with digitally connected devices. The System can be used alone or in conjunction with these digitally connected devices where the user manually controls actions for therapy decisions.
The FreeStyle Libre 2 Flash Glucose Monitoring System (herein referred to as the 'FreeStyle Libre 2 System or 'System') is an integrated continuous glucose monitoring system (iCGM) that provides continuous glucose measurements every minute to provide glucose levels, trends and alerts. The FreeStyle Libre 2 System consists of two primary components: a sensor that transmits via Bluetooth Low Energy (BLE), and a BLE enabled display device (Reader). User-initiated RFID scanning provides the user with real-time glucose measurements (glucose values) accompanied by trend information (glucose arrows) and historical glucose information (glucose graph). The user may use the sensor glucose results and information provided by the System in making treatment decisions. The System also provides configurable alerts and alarms designed to warn the user of Low Glucose, High Glucose or Signal Loss.
Here's a breakdown of the acceptance criteria and the studies conducted for the FreeStyle Libre 2 Flash Glucose Monitoring System, based on the provided document:
Acceptance Criteria and Device Performance
| Acceptance Criteria Category | Specific Criteria/Metric | Reported Device Performance (Adult) | Reported Device Performance (Pediatric 6-17) | Reported Device Performance (Pediatric 4-5) |
|---|---|---|---|---|
| Accuracy (Overall) | Percent of iCGM values within 20% of YSI reference | 90.2% (95% LCL: 88.7%) | 90.3% (95% LCL: 88.1%) | 85.4% (95% LCL: 80.3%) (SMBG as comparator) |
| Accuracy (Point/Percent, iCGM Range < 70 mg/dL / <54 mg/dL / 54-69 mg/dL) | % within 15 mg/dL | 89.0% (<70mg/dL), 85.9% (<54mg/dL), 89.5% (54-69mg/dL) | 82.7% (<70mg/dL), 71.9% (<54mg/dL), 86.4% (54-69mg/dL) | N/A (SMBG used for 4-5) |
| Accuracy (Point/Percent, iCGM Range 70-180 mg/dL / 181-250 mg/dL / >250 mg/dL) | % within 15% | 75.9% (70-180mg/dL), 89.1% (181-250mg/dL), 94.0% (>250mg/dL) | 78.0% (70-180mg/dL), 86.0% (181-250mg/dL), 92.2% (>250mg/dL) | N/A (SMBG used for 4-5) |
| Accuracy (Point/Percent, YSI Range < 70 mg/dL / <54 mg/dL / 54-69 mg/dL) | % within 15 mg/dL | 95.3% (<70mg/dL), 91.1% (<54mg/dL), 94.7% (54-69mg/dL) | 94.9% (<70mg/dL), 93.9% (<54mg/dL), 96.5% (54-69mg/dL) | N/A (SMBG used for 4-5) |
| Accuracy (Point/Percent, YSI Range 70-180 mg/dL / 181-250 mg/dL / >250 mg/dL) | % within 15% | 76.5% (70-180mg/dL), 87.9% (181-250mg/dL), 90.9% (>250mg/dL) | 74.3% (70-180mg/dL), 86.6% (181-250mg/dL), 90.2% (>250mg/dL) | N/A (SMBG used for 4-5) |
| Precision | Mean Coefficient of Variation (Mean %CV) | 5.7% | 5.8% | 4.8% |
| Low Glucose Alarm (e.g., 70 mg/dL) | True Alarm Rate | 86.0% | 80.3% | N/A |
| False Alarm Rate | 14.0% | 19.7% | N/A | |
| Correct Detection Rate | 89.3% | 93.5% | N/A | |
| Missed Detection Rate | 10.7% | 6.5% | N/A | |
| High Glucose Alarm (e.g., 200 mg/dL) | True Alarm Rate | 99.2% | 98.0% | N/A |
| False Alarm Rate | 0.8% | 2.0% | N/A | |
| Correct Detection Rate | 97.1% | 98.0% | N/A | |
| Missed Detection Rate | 2.9% | 2.0% | N/A | |
| Sensor Stability (MARD) | Mean Absolute Relative Difference (MARD) throughout wear duration | 9.9% (Beginning), 8.5% (Early Middle), 8.8% (Late Middle), 9.1% (End) | 10.7% (Beginning), 8.0% (Early Middle), 9.7% (Late Middle), 10.2% (End) | N/A |
| Sensor Life | Survival Rate by Day 14 | 71.1% | 78.1% | N/A |
| Glucose Reading Availability | Capture Rate over wear duration | Ranges from 98.1% to 98.6% | Ranges from 94.6% to 96.0% | N/A |
| Reportable Range | 40 to 400 mg/dL | Supported by clinical study data (display 'LO' or 'HI' outside range) | Supported by clinical study data (display 'LO' or 'HI' outside range) | Supported by clinical study data (display 'LO' or 'HI' outside range) |
| Ascorbic Acid Interference | Max average sensor bias after 3x1000mg doses | Less than 20 mg/dL (after second and third 1000mg doses) | Not explicitly stratified by age for this study, assumed adults. | N/A |
Study Information
-
Sample Size and Data Provenance:
- Clinical Accuracy Studies (Study 1 & Study 2):
- Study 1 (Adults): 146 subjects (91.1% Type 1, 8.9% Type 2 diabetes, all requiring insulin).
- Study 2 (Pediatric): 139 subjects (98.6% Type 1, 1.4% Type 2 diabetes, all requiring insulin).
- Age Groups in Study 2 (for YSI comparison): 77 pediatric subjects for <70 mg/dL range, 124 for 70-180 mg/dL and >180 mg/dL range. Specific breakdown for 4-5 and 6-17 years where YSI was used.
- Age Groups in Study 2 (for SMBG comparison): 8 pediatric subjects (4-5 years old) only had SMBG as comparator.
- Data Provenance: Prospective, multi-center studies conducted at five (adult) and four (pediatric) centers in the United States.
- Ascorbic Acid Interference Study: 60 subjects (adults, age 18 and older) with diabetes, across 4 sites.
- Precision Study (from Clinical Studies):
- Adults (18+): 146 subjects, 26,791 paired readings.
- Pediatric 4-5 years: 7 subjects, 248 paired readings.
- Pediatric 6-17 years: 130 subjects, 10,623 paired readings.
- Clinical Accuracy Studies (Study 1 & Study 2):
-
Number of Experts and Qualifications for Ground Truth:
- The document implies that the ground truth for glucose values was established using a "laboratory glucose reference" (YSI) analyzer. While YSI analyzers are highly accurate, the document does not explicitly state the number of experts or their qualifications for establishing the ground truth. It refers to the YSI as the comparator method, which is a laboratory instrument, not human experts. For pediatric subjects aged 4-5 years, Self-Monitoring Blood Glucose (SMBG) tests were used as the comparator method.
-
Adjudication Method for the Test Set:
- Not applicable in the traditional sense of medical image adjudication. The "ground truth" was established by a laboratory reference method (YSI for most cases, SMBG for some pediatric cases). There's no indication of an adjudication process involving multiple human readers to resolve discrepancies in ground truth measurements. The comparison is between the device and the reference instrument.
-
Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
- No, an MRMC comparative effectiveness study was not done. This type of study typically involves human readers interpreting cases with and without AI assistance. The FreeStyle Libre 2 is a standalone diagnostic device that measures glucose directly, so its performance is compared against a reference standard (YSI, SMBG), not human interpretation with or without AI.
-
Standalone (Algorithm Only) Performance:
- Yes, the studies primarily assessed standalone performance. The device automatically provides continuous glucose measurements. The "iCGM glucose values" are the output of the algorithm. The intent is for the device to replace blood glucose testing for treatment decisions, which is a standalone use case. Human interpretation of trends and patterns is part of the overall management, but the device's accuracy itself is assessed in a standalone manner against the reference.
-
Type of Ground Truth Used:
- Laboratory Reference and Self-Monitoring Blood Glucose (SMBG):
- For adult subjects and pediatric subjects aged 6-17 years, the primary ground truth was established by YSI glucose measurements. YSI (Yellow Springs Instrument) is a common laboratory method considered a highly accurate reference for glucose.
- For pediatric subjects aged 4-5 years, the ground truth was established by SMBG (Self-Monitoring Blood Glucose) tests.
- Laboratory Reference and Self-Monitoring Blood Glucose (SMBG):
-
Sample Size for the Training Set:
- The document mentions "clinical evaluation" and "pivotal clinical studies" but does not explicitly state the sample size of a training set specifically for algorithm development or machine learning. The provided clinical studies (Study 1 and Study 2) describe the test sets used for performance evaluation, not necessarily for training the algorithm. CGM devices typically use complex calibration algorithms developed with extensive internal data that is not always detailed in 510(k) summaries.
-
How the Ground Truth for the Training Set Was Established:
- As the document does not provide details on a specific training set or its size, it also does not elaborate on how the ground truth for such a set was established. We can infer that similar high-accuracy laboratory reference methods (like YSI) would have been used during the development and calibration phases of the device, prior to the validation studies described in the submission.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
June 12, 2020
Abbott Diabetes Care Inc. Naveen Thuramalla DVP, Quality Assurance and Regulatory Affairs 1360 South Loop Road Alameda, CA 94502
Re: K193371
Trade/Device Name: FreeStyle Libre 2 Flash Glucose Monitoring System Regulation Number: 21 CFR 862.1355 Regulation Name: Integrated Continuous Glucose Monitoring System Regulatory Class: Class II Product Code: QLG, NBW Dated: May 8, 2020 Received: May 11, 2020
Dear Naveen Thuramalla:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Kellie B. Kelm, Ph.D. Acting Director Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K193371
Device Name
FreeStyle Libre 2 Flash Glucose Monitoring System
Indications for Use (Describe)
The FreeStyle Libre 2 Flash Glucose Monitoring System is a continuous glucose monitoring (CGM) device with real time alarms capability indicated for the management of diabetes in persons age 4 and older. It is intended to replace blood glucose testing for diabetes treatment decisions, unless otherwise indicated.
The System also detects trends and tracks patterns and aids in the detection of episodes of hyperglycemia and hypoglycemia, facilitating both acute and long-term therapy adjustments. Interpretation of the System readings should be based on the glucose trends and several sequential readings over time.
The System is also intended to autonomously communicate with digitally connected devices. The System can be used alone or in conjunction with these digitally connected devices where the user manually controls actions for therapy decisions.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| ☑ Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Abbott Diabetes Care. The logo consists of a blue abstract shape on the left, followed by the text "Abbott" in bold black font. Below "Abbott" is the text "Diabetes Care" in a smaller, non-bold font. The logo is clean and professional, representing Abbott's presence in the diabetes care industry.
1. 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.
The assigned 510(k) number is: K193371
1.1 Submitter:
Abbott Diabetes Care, Inc. 1360 South Loop Road Alameda, CA 94502
Contact: Naveen Thuramalla Title: DVP Quality Assurance and Regulatory Affairs Phone: 510-239-2618 Fax: (510) 864-4791 Email: naveen.thuramalla@abbott.com
1.2 Device Names and Classification:
| Name of Device: | FreeStyle Libre 2 Flash Glucose Monitoring System |
|---|---|
| Common Name: | Integrated Continuous Glucose Monitoring System, FactoryCalibrated, Not for use with automated insulin delivery systems |
| Regulatory Section: | 21 CFR 862.1355, 21 CFR 862.1345 |
| Classification: | Class II |
| Product Code(s): | QLG, NBW |
| Review Panel: | Clinical Chemistry |
1.3 Predicate Device
Dexcom G6 Continuous Glucose Monitoring System, DEN170088 Predicate Device:
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the logo for Abbott Diabetes Care. The logo consists of a stylized blue "a" on the left, followed by the text "Abbott" in bold, black letters. Below "Abbott" is the text "Diabetes Care" in a smaller, non-bold font, also in black.
1.4 Indications for Use:
Indications for Use:
The FreeStyle Libre 2 Flash Glucose Monitoring System is a continuous glucose monitoring (CGM) device with real time alarms capability indicated for the management of diabetes in persons age 4 and older. It is intended to replace blood glucose testing for diabetes treatment decisions, unless otherwise indicated.
The System also detects trends and tracks patterns and aids in the detection of episodes of hyperglycemia and hypoglycemia, facilitating both acute and long-term therapy adjustments. Interpretation of the System readings should be based on the glucose trends and several sequential readings over time.
The System is also intended to autonomously communicate with digitally connected devices. The System can be used alone or in conjunction with these digitally connected devices where the user manually controls actions for therapy decisions.
Contraindications
- Automated Insulin Delivery: The System must not be used with automated insulin ● dosing (AID) systems, including closed loop and insulin suspend systems.
- MRI/CT/Diathermy: The System must be removed prior to Magnetic Resonance ● Imaging (MRI). Computed Tomography (CT) scan. or high-frequency electrical heat (diathermy) treatment. The effect of MRI, CT scans, or diathermy on the performance of the System has not been evaluated. The exposure may damage the Sensor and may impact proper function of the device which could cause incorrect readings.
1.5 Device Description
The FreeStyle Libre 2 Flash Glucose Monitoring System (herein referred to as the 'FreeStyle Libre 2 System or 'System') is an integrated continuous glucose monitoring system (iCGM) that provides continuous glucose measurements every minute to provide glucose levels, trends and alerts. The FreeStyle Libre 2 System consists of two primary components: a sensor that transmits via Bluetooth Low Energy (BLE), and a BLE enabled display device (Reader). User-initiated RFID scanning provides the user with real-time glucose measurements (glucose values) accompanied by trend information (glucose arrows) and historical glucose information (glucose graph). The user may use the sensor glucose results and information provided by the System in making treatment decisions. The System also provides configurable alerts and alarms designed to warn the user of Low Glucose, High Glucose or Signal Loss.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Abbott Diabetes Care. The logo consists of a blue stylized letter "a" on the left, followed by the text "Abbott" in bold, black font on the top line, and "Diabetes Care" in a smaller, regular black font on the bottom line. The logo is clean and professional, conveying a sense of trust and reliability.
FreeStyle Libre 2 Sensor
- The Sensor is single use and disposable. The Sensor is provided as two secondary components, Sensor Applicator and Sensor Pack (sterile device), which are used to assemble and apply the Sensor to the back of the user's arm. During Sensor application, the sensor tail is inserted about 5.5 millimeters below the surface of the skin through the guidance of a needle. The needle is retracted back into the applicator after insertion. The Sensor continuously measures glucose concentration in interstitial fluid every minute and has an 8-hour memory capacity. The Sensor is factory calibrated, does not require fingerstick calibration, and can be worn for up to 14 days.
FreeStyle Libre 2 Reader
- The Reader is a small handheld device that uses RFID communication to start new ● sensors, to scan sensors to display and record data and uses BLE communication to issue alarms that notify the user when glucose values pass high or low thresholds, when enabled. The Reader also includes a built-in strip port with blood glucose functionality as a convenience for users (which is intended to work with the FreeStyle Precision Neo Blood Glucose test strips, cleared under K171941), and a user interface that includes event logging features.
The FreeStyle Libre 2 System is an interoperable connected device that can communicate glucose readings and other information wirelessly and from interoperable electronic interfaces. This system is not compatible with AID systems, including insulin suspend systems.
Test Principle 1.5.1
The FreeStyle Libre 2 Flash Glucose Monitoring System uses an electrochemical sensor to monitor glucose levels in the interstitial fluid (ISF). The sensor is held in place with an adhesive pad and incorporates both the subcutaneously implanted sensor and associated electronics. The sensor uses a glucose oxidase enzyme to oxidize glucose and transfer electrons to a metal electrode, producing a current. The strength of the current is proportional to the amount of glucose present in the subcutaneous space. The FreeStyle Libre 2 System converts the electrical current signal to a glucose value (in mg/dL) for display to the user on the handheld Reader.
Substantial Equivalence 1.6
A. Predicate Device Name:
Dexcom G6 Continuous Glucose Monitoring System
B. Predicate 510(k) Number(s):
C. Comparison with Predicate:
{6}------------------------------------------------
The similarities and differences between the FreeStyle Libre 2 Flash Glucose Monitoring System and the predicate device, Dexcom G6 (DEN170088) are highlighted in the tables below:
| Similarities | ||
|---|---|---|
| Item | FreeStyle Libre 2 FlashGlucose Monitoring System | Dexcom G6 ContinuousGlucose Monitoring System(DEN170088) |
| Indications for Use | The FreeStyle Libre 2 FlashGlucose Monitoring System is acontinuous glucose monitoring(CGM) device with real timealarms capability indicated forthe management of diabetes inpersons age 4 and older. It isintended to replace bloodglucose testing for diabetestreatment decisions, unlessotherwise indicated.The System also detects trendsand tracks patterns and aids inthe detection of episodes ofhyperglycemia andhypoglycemia, facilitating bothacute and long-term therapyadjustments. Interpretation ofthe System readings should bebased on the glucose trends andseveral sequential readings overtime.The System is also intended toautonomously communicate with digitally connecteddevices. The System can beused alone or in conjunctionwith these digitally connecteddevices where the user manuallycontrols actions for therapydecisions. | The Dexcom G6 ContinuousGlucose Monitoring System(Dexcom G6 System) is a realtime, continuous glucosemonitoring device indicated forthe management of diabetes inpersons age 2 years and older.The Dexcom G6 System isintended to replace fingerstickblood glucose testing fordiabetes treatment decisions.Interpretation of the DexcomG6 System results should bebased on the glucose trends andseveral sequential readings overtime. The Dexcom G6 Systemalso aids in the detection ofepisodes of hyperglycemia andhypoglycemia, facilitating bothacute and long-term therapyadjustments.The Dexcom G6 System is alsointended to autonomouslycommunicate with digitally connected devices, includingautomated insulin dosing (AID)systems. The Dexcom G6System can be used alone or inconjunction with these digitallyconnected medical devices forthe purpose of managingdiabetes. |
| Device Type | Integrated CGM | Same |
| Principle of Operation | Amperometric measurement ofcurrent proportional to glucoseconcentration in interstitial fluidvia glucose oxidase chemicalreaction | Same |
| Sample Type | Interstitial fluid | Same |
| Similarities | ||
| Item | FreeStyle Libre 2 FlashGlucose Monitoring System | Dexcom G6 ContinuousGlucose Monitoring System(DEN170088) |
| Enzyme | Glucose oxidase | Same |
| Clinical Application | Management of diabetesmellitus | Same |
| Clinical Setting/Sites of Use | Home use | Same |
| Data Displayed | Current glucose value, currentglucose trend, graph with recentglucose history, user enteredevents | Same |
| Differences | ||
| Item | FreeStyle Libre 2 FlashGlucose Monitoring System | Dexcom G6 Continuous GlucoseMonitoring System(DEN170088) |
| Situations where fingersticktest is required to confirmsensor reading (adjunctiveuse) | The user's symptoms donot match the glucosevalues displayed by thedevice.The device does not show | The user's symptoms do notmatch the glucose valuesdisplayed by the device.The device does not show aglucose value or a trendarrow. |
| Compatibility with connecteddevices | Compatible with digitallyconnected devices where theuser manually controlsactions for therapy decisions | Compatible with digitallyconnected devices, includingautomated insulin dosing (AID)systems |
| Sensor Calibration | Factory calibrated | Factory calibrated, with optionaluser calibration |
| Glucose reading updateinterval | Every 1 minute | Every 5 minutes |
| Wireless communicationsprotocol | Near Field Communication(NFC): (13.56 MHz RFID)Bluetooth Low Energy (BLE)4.0 | Bluetooth Core Specification v4.0 |
| Communications range | 20 feet unobstructed | 20 feet unobstructed |
| Glucose Trend Arrow | ↑, > +2 mg/dL/min↗, +1 and +2 mg/dL/min→, -1 to +1 mg/dL/min↘, -2 to -1 mg/dL/min↓, < -2 mg/dL/min | ↑↑, > +3mg/dL/min↑, +2 to +3 mg/dL/min↗, +1 to +2mg/dL/min→, -1 to +1 mg/dL/min↘, -2 to -1 mg/dL/min↓, -3 to -2 mg/dL/min↓↓, < -3mg/dL/min |
| Anatomical sensor wearlocations | Back of the upper arm | Abdomen (age 2+ years) or upperbuttocks (age 2-17 years) |
| Sensor warm up time | 1 hour | 2 hours |
| Sensor life | Up to 14 days (automaticsensor shut off) | Up to 10 days (automatic sensorshut off) |
| Storage Conditions(Sensor) | Temperature: 36°F – 82°FHumidity: 10-90% RH | Temperature: 36°F – 86°FHumidity: 10%-90% RH |
| Intended Use Population | Persons with diabetes age 4and above | Persons with diabetes age 2+years |
| Alerts and Alarms | Low Glucose Alarm, HighGlucose Alarm, Signal LossAlarm, Scan Error, SensorError | Urgent low glucose (55mg/dL), predictable lowglucose, threshold lowglucose, threshold high |
| Differences | ||
| Item | FreeStyle Libre 2 FlashGlucose Monitoring System | Dexcom G6 Continuous GlucoseMonitoring System(DEN170088) |
| For Low and High Glucosealarms, a user-initiated actionis required to see glucosereading. | glucose, rising rate ofglucose, falling rate ofglucose, signal loss, sensorfailure, transmitter failure. | |
| Primary Display Device | Reader | Hardware receiver or mobile appinstalled on compatible smartdevice |
| Trend Graph Glucose History | 8 hours, 24-hour graph andother reports can be used toview logged data | 1,3, 6, 12, and 24 hours |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image contains the logo for Abbott Diabetes Care. The logo consists of a blue, stylized "A" on the left, followed by the text "Abbott" in bold, black letters. Below "Abbott" is the text "Diabetes Care" in a smaller, non-bold font, also in black. The overall design is clean and corporate, representing the brand identity of Abbott Diabetes Care.
{8}------------------------------------------------
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for Abbott Diabetes Care. The logo consists of a blue square with a stylized "a" inside, followed by the words "Abbott" in bold black letters and "Diabetes Care" in smaller black letters below. The logo is simple and modern, and the colors are clean and professional.
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image contains the logo for Abbott Diabetes Care. The logo consists of a stylized blue "a" on the left, followed by the text "Abbott" in bold black font on the right. Below "Abbott" is the text "Diabetes Care" in a smaller, non-bold black font. The logo is simple and modern, with a focus on the company's name and area of expertise.
Summary of Performance Testing 1.7
Analytical Performance 1.7.1
a. Reproducibility/Precision
iCGM performance was evaluated in clinical studies described below in Section 1.7.3. In Study 1, all subjects wore two sensors concurrently, one on the back of each upper arm, to evaluate device precision. A total of 26,791 paired readings from 146 subjects were evaluated. Mean coefficient of variation (Mean %CV) was 5.7%.
In Study 2, all subjects wore two sensors concurrently, one on the back of each upper arm, to evaluate device precision. For pediatric age 4-5, a total of 248 paired readings from 7 subjects were evaluated; mean coefficient of variation (Mean %CV) was 4.8%. For pediatric age 6-17, a total of 10,623 paired readings from 130 subjects were evaluated; mean %CV was 5.8%.
Precision by subject age group:
| Subject agegroup | Coefficientof Variation(%) | PairedAbsoluteDifference(mg/dL) | Paired AbsoluteRelativeDifference (%) | Number ofSubjects | Number ofPairedReadings |
|---|---|---|---|---|---|
| Adults (18+) | 5.7 | 12.4 | 8.1 | 146 | 26791 |
| Pediatric 4-5years | 4.8 | 10.7 | 6.7 | 7 | 248 |
| Pediatric 6-17years | 5.8 | 13.0 | 8.2 | 130 | 10623 |
b. Linearity/Assay Reportable Range
The reportable range for the FreeStyle Libre 2 System is 40 to 400 mg/dL.
c. Traceability, Stability, Expected Values (Control, Calibrators, or Methods)
The FreeStyle Libre 2 sensor has a storage life of 9 months. Shelf-life was evaluated at 36° -82° Fahrenheit within the humidity range of 10% - 90%.
d. Detection Limit
If a glucose measurement is less than 40 mg/dL, the result will be displayed by the system as 'LO'. If the glucose measurement exceeds 400 mg/dL, the result will be displayed as 'HI'. Data supporting this claimed measurement range was generated in the clinical study described in Section 1.7.3 below.
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image contains the logo for Abbott Diabetes Care. On the left is a blue stylized letter 'a'. To the right of the logo is the text 'Abbott' in bold black font, with the text 'Diabetes Care' in a smaller, non-bold black font below it.
e. Analytical Specificity:
Certain endogenous and exogenous substances in the interstitial fluid may interfere with iCGM measurements. The types of potential interference and the extent of bias are dependent on the test principle of the sensor technology. For this technology, ascorbic acid (vitamin C) has been shown to interfere with the system performance when present in the body at concentrations higher than the recommended daily allowance. During the pivotal clinical studies, all medications and vitamins for which the subject was using or had used over the course of the study along with frequency of use were recorded. Both studies did not include exclusion criteria based on medications or vitamin use.
A clinical study was conducted to evaluate ascorbic acid interference for the FreeStyle Libre 2 System. It is a prospective, multi-center, single-arm study that enrolled 60 subjects at 4 sites. The study was designed to evaluate performance of the System in people with diabetes (age 18 and older) taking ascorbic acid. Each subject wore two sensors, one on the back of each upper arm, for a period of up to 10 days.
Subjects participated in one clinic session between sensor wear days 4 and 10. During the inclinic visit, data was collected over a 13-hour period. Each subject had a one-hour baseline phase where venous blood was collected every 10 minutes. After this first hour, a dose of 1000 mg ascorbic acid was given with a meal and venous samples were collected every 20 minutes for the next four hours. Subjects then received a second dose of 1000 mg ascorbic acid with a meal and measurements were taken every 20 minutes for another 4 hours. Subsequently, a third dose of 1000 mg ascorbic acid was given with a meal and measurements were taken every 20 minutes for 4 more hours.
Accuracy performance of the System was assessed by the mean difference (mg/dL) between GM and YSI readings pre and post- ascorbic acid doses. For each matched pair of CGM and Y SI, the difference, absolute difference, relative difference and absolute relative difference is calculated. The biases for all three doses are calculated by subtracting the baseline bias at pre-dose from the post-dose bias as shown in the 'Difference from Baseline' column below.
{12}------------------------------------------------
| Time Interval | Difference (Bias, mg/dL) | Relative Difference(%) | ||||||
|---|---|---|---|---|---|---|---|---|
| (Hour)Relative toDose | DifferencefromBaseline | Mean | Median | SD | DifferencefromBaseline | Mean | Median | SD |
| Pre-dose (0) | 0.0 | -2.4 | -2.3 | 13.5 | 0.0 | -1.7 | -1.6 | 9.05 |
| Post 1st dose:0-1 hour (1) | -2.9 | -5.3 | -3.3 | 14.3 | -1.8 | -3.5 | -2.7 | 10.2 |
| Post 1st dose:1-2 hour (2) | 5.4 | 2.9 | 3.0 | 17.0 | 3.8 | 2.1 | 1.5 | 12.4 |
| Post 1st dose:2-3 hour (3) | 9.3 | 6.9 | 5.6 | 15.4 | 6.2 | 4.5 | 3.8 | 11.1 |
| Post 1st dose:3-4 hour (4) | 8.7 | 6.2 | 5.8 | 13.5 | 5.9 | 4.2 | 4.1 | 9.96 |
| Post 2nd dose:0-1 hour (5) | 12.4 | 10.0 | 9.0 | 14.2 | 9.2 | 7.5 | 6.7 | 10.9 |
| Post 2nd dose:1-2 hour (6) | 16.6 | 14.1 | 13.0 | 15.2 | 11.7 | 9.9 | 9.7 | 11.5 |
| Post 2nd dose:2-3 hour (7) | 18.4 | 15.9 | 15.5 | 16.2 | 12.8 | 11.1 | 9.9 | 12.0 |
| Post 2nd dose:3-4 hour (8) | 14.9 | 12.4 | 12.8 | 13.7 | 9.9 | 8.2 | 7.9 | 11.1 |
| Post 3rd dose:0-1 hour (9) | 14.5 | 12.1 | 12.3 | 15.9 | 9.2 | 7.5 | 8.0 | 11.3 |
| Post 3rd dose:1-2 hour (10) | 18.9 | 16.5 | 17.8 | 17.7 | 11.6 | 9.9 | 10.5 | 12.1 |
| Post 3rd dose:2-3 hour (11) | 19.7 | 17.3 | 16.3 | 19.0 | 12.8 | 11.1 | 12.0 | 13.1 |
| Post 3rd dose:3-4 hour (12) | 19.0 | 16.5 | 16.0 | 17.9 | 12.8 | 11.1 | 10.5 | 13.4 |
The figure below shows the mean bias from the baseline at different time points. The results show that the maximum bias was observed at around 3 hours after the intake of each dose of vitamin C, with a maximum average bias of 9.3 mg/dL after the first 1000 mg dose. The maximum average biases were 18.4 mg/dL and 19.7 mg/dL after the second and third doses respectively.
{13}------------------------------------------------
Image /page/13/Picture/0 description: The image contains the logo for Abbott Diabetes Care. The logo consists of a blue, stylized letter "a" on the left, followed by the text "Abbott" in bold, black letters. Below "Abbott" is the text "Diabetes Care" in a smaller, non-bold font, also in black.
Image /page/13/Figure/2 description: The image shows a line graph with "Hour" on the x-axis and "Bias, mg/dL" on the y-axis. The y-axis ranges from -10 to 50, while the x-axis ranges from 0 to 12. The line starts at approximately 0 at hour 0, dips to around -3 at hour 1, and then gradually increases to approximately 19 at hour 12.
Based on the results of the clinical evaluation, the following statements have been placed in the device labeling:
- Taking ascorbic acid (vitamin C) supplements while wearing the Sensor may falsely raise Sensor glucose readings. Taking more than 500 mg of ascorbic acid per day may affect the Sensor readings which could cause you to miss a severe low glucose event. Ascorbic acid can be found in supplements including multivitamins. Some supplements, including cold remedies such as Airborne® and Emergen-C®, may contain high doses of 1000 mg of ascorbic acid and should not be taken while using the Sensor. See your healthcare professional to understand how long ascorbic acid is active in your body.
- A clinical study was conducted to evaluate the effect of ascorbic acid on Sensor ● performance. Data from 57 adult subjects with diabetes was collected over a 13-hour period. Each subject had a one-hour baseline phase where venous blood was collected every 10 minutes. After this first hour, a dose of 1000 mg ascorbic acid was given with a meal and venous samples were collected every 20 minutes for the next four hours. A maximum average sensor bias of 9.3 mg/dL was observed around 3 hours after the 1000 mg ascorbic acid dose. Subjects then received a second dose of 1000 mg ascorbic acid with a meal and the same process was continued for another 4 hours. A third dose of 1000 mg ascorbic acid was then given and study subjects were followed for 4 more hours. After the second dose of ascorbic acid the maximum average sensor bias increased, with minimal change in sensor bias after the third dose, suggesting that saturation had occurred by the second 1000 mg dose of ascorbic acid. The maximum average sensor bias after the three 1000 mg doses of ascorbic acid was less than 20 mg/dL.
{14}------------------------------------------------
Image /page/14/Picture/0 description: The image contains the logo for Abbott Diabetes Care. The logo consists of a stylized blue "a" on the left, followed by the text "Abbott" in bold, black letters. Below "Abbott" is the text "Diabetes Care" in a smaller, non-bold font, also in black.
1.7.2 Comparison Studies
-
I. Method comparison with predicate device.
Not applicable -
II. Matrix comparison:
Not applicable. Interstitial fluid is the only indicated matrix
1.7.3 Clinical Studies
| Study Name | Patient Population | Study Objective |
|---|---|---|
| FreeStyle Libre Flash GlucoseMonitoring System AccuracyStudy(Study 1) | 18 + years of ageType 1 or Type 2 Diabetes | Pivotal study to evaluate safetyand effectiveness of theFreeStyle Libre Flash GlucoseMonitoring System incomparison to laboratoryglucose reference. |
| Effectiveness and Safety Studyof the FreeStyle Libre FlashGlucose Monitoring System inPediatric Populations(Study 2) | 4 - 17 years of ageType 1 or Type 2 Diabetes | Pivotal study to evaluate safetyand effectiveness of theFreeStyle Libre Flash GlucoseMonitoring System in pediatricpopulations in comparison tolaboratory glucose reference.SMBG was the comparatormethod used for pediatricsubjects aged 4-5 years. |
To demonstrate the accuracy performance of the FreeStyle Libre 2 Flash Glucose Monitoring System, two prospective clinical studies (Study 1, adult: Study 2, pediatric) were conducted at five and four centers, respectively. All study centers were located in the United States. Results from these studies are presented concurrently.
Study 1 enrolled 146 adult subjects with diabetes. Of these, 91.1% were diagnosed with Type 1 diabetes, while 8.9% were diagnosed with Type 2 diabetes. All subjects required insulin to manage their diabetes. Subjects wore 2 sensors for up to 14 days. Subjects took part in up to three ten-hour clinical sessions. Clinic sessions for each subject took place during four distinct periods: days 1, 2 and 3; days 9 and 12, and days 13 and 14. During each clinic session, each subject's glucose was manipulated to achieve sustained periods of either hypoglycemia or hyperglycemia, or both.
{15}------------------------------------------------
Image /page/15/Picture/0 description: The image contains the logo for Abbott Diabetes Care. The logo consists of a blue abstract symbol on the left, followed by the word "Abbott" in bold, black font. Below "Abbott" is the phrase "Diabetes Care" in a smaller, non-bold, black font. The logo is clean and professional, suggesting a focus on healthcare and technology.
Study 2 enrolled 139 pediatric subjects with diabetes. Of these, 98.6% were diagnosed with Type 1 diabetes, while 1.4% with Type 2 diabetes. All subjects required insulin to manage their diabetes. Subjects wore 2 sensors for up to 14 days. Subjects age six and older took part in up to two eight-hour clinical sessions. No YSI comparator measurements were obtained for subjects 4-5 years of age, and performance for these subjects was compared against self-monitoring blood glucose (SMBG) tests. Clinic sessions for each subject took place during four distinct periods: days 1, 2 and 3; days 7 and 8; days 9 and 12, and days 13 and 14. During each clinic session, each subject's glucose was manipulated to achieve sustained periods of either hypoglycemia or hyperglycemia, or both.
In Study 1 and Study 2, accuracy of the FreeStyle Libre 2 Flash Glucose Monitoring System was evaluated by comparing iCGM glucose values to reference glucose values. Glucose values were obtained from the system and from the comparator at the same or similar time. In both studies, absolute differences in mg/dL of values compared to the comparator method were calculated for all values below 70 mg/dL. For values of 70 mg/dL and above, percentage differences compared to the comparator method were calculated.
| iCGMGlucose Range(mg/dL) | No. Pair | No.Subject | Percentwithin15 mg/dL (95%LCL) | Percentwithin40 mg/dL(95% LCL) | Percentwithin15%(95% LCL) | Percentwithin40%(95% LCL) |
|---|---|---|---|---|---|---|
| <70 | 3530 | 130 | 89.0 (86.7) | 99.4 (99.1) | ||
| 70-180 | 7785 | 144 | 75.9 (73.1) | 99.6 (99.4) | ||
| >180 | 7420 | 141 | 91.5 (89.4) | 100.0 (99.9) |
Percent and Point Accuracy by iCGM Glucose Range (Study 1, Adult)
| Percent and Point Accuracy by iCGM Glucose Range (Study 2, Pediatric*) | ||||
|---|---|---|---|---|
| iCGMGlucose Range(mg/dL) | No.Pairs | No.Subject | Percentwithin15 mg/dL (95%LCL) | Percentwithin40 mg/dL(95% LCL) | Percentwithin15%(95% LCL) | Percentwithin40%(95% LCL) |
|---|---|---|---|---|---|---|
| <70 | 1002 | 77 | 82.7 (77.2) | 98.6 (97.5) | ||
| 70-180 | 2690 | 124 | 78.0 (74.8) | 99.3 (98.9) | ||
| >180 | 2854 | 124 | 87.3 (84.4) | 99.7 (99.5) |
- Results only include children 6-17 years of age with YSI measurement as comparator. SMBG measurements were obtained as comparator for children 4-5 years of age.
Percent and Point Accuracy by YSI Glucose Range (Study 1. Adult)
| YSIGlucose Range(mg/dL) | No. Pair | No. Subject | Percentwithin15 mg/dL(95% LCL) | Percentwithin40 mg/dL(95% LCL) | Percentwithin15%(95% LCL) | Percentwithin40%(95% LCL) |
|---|---|---|---|---|---|---|
| <70 | 3468 | 130 | 95.3 (93.9) | 100.0 (100.0) | ||
| 70-180 | 7504 | 144 | 76.5 (73.8) | 99.6 (99.4) | ||
| >180 | 7763 | 141 | 88.9 (86.3) | 99.9 (99.8) |
{16}------------------------------------------------
Image /page/16/Picture/0 description: The image shows the logo for Abbott Diabetes Care. The logo consists of a blue square with a stylized "a" inside, followed by the text "Abbott" in bold black letters. Below "Abbott" is the text "Diabetes Care" in a smaller, non-bold font.
| Percent and Point Accuracy by YSI Glucose Range (Study 2, Pediatric*) | |
|---|---|
| ----------------------------------------------------------------------- | -- |
| YSIGlucose Range(mg/dL) | No.Pairs | No.Subject | Percentwithin15 mg/dL(95% LCL) | Percentwithin40 mg/dL(95% LCL) | Percentwithin15%(95% LCL) | Percentwithin40%(95% LCL) |
|---|---|---|---|---|---|---|
| <70 | 882 | 67 | 94.9 (92.1) | 100.0 (100.0) | ||
| 70-180 | 2743 | 124 | 75.6 (72.2) | 99.0 (98.5) | ||
| >180 | 2921 | 122 | 87.7 (84.9) | 99.1 (97.9) |
- Results only include children 6-17 years of age with YSI measurement as comparator. SMBG measurements were obtained as comparator for children 4-5 years of age.
Percent values within 20% of the comparator method were calculated across the measuring range overall.
| iCGM GlucoseRange (mg/dL) | No. Pairs | No. Subject | Percentwithin 20%(95% LCL) |
|---|---|---|---|
| Adult (18 yearsand up) | 18735 | 144 | 90.2 (88.7) |
| Pediatric (6-17years) | 6546 | 129 | 90.3 (88.1) |
| Pediatric (4-5 yearsold)* | 341 | 8 | 85.4 (80.3) |
Percent of iCGM values within 20% of YSI reference
*Only SMBG measurements were obtained as comparator for children 4-5 years of age
Percent values within 15%/15 mg/dL, 20%/20 mg/dL stratified by glucose ranges of <54, 54-69, 70-180. 181-250 and >250 mg/dL for iCGM and laboratory comparator were also provided.
| CGMGlucoseLevel(mg/dL) | Numberof CGM-ReferencePairs | PercentWithin±15mg/dL | PercentWithin±20mg/dL | PercentWithin±40mg/dL | PercentWithin±15% | PercentWithin±20% | PercentWithin±40% | Meanbias(mg/dL) | MARI(%) |
|---|---|---|---|---|---|---|---|---|---|
| <54 | 518 | 85.9 | 93.8 | 99.4 | -6.4 | 13.8 | |||
| 54-69 | 3012 | 89.5 | 94.2 | 99.1 | -3.3 | 10.8 | |||
| 70-180 | 7785 | 76.5 | 86.6 | 99.2 | -4.8 | 10.6 | |||
| 181-250 | 3037 | 89.1 | 95.0 | 99.9 | -10.1 | 7.8 | |||
| >250 | 4383 | 94.0 | 97.9 | 100.0 | -6.3 | 6.1 |
Accuracy to YSI within CGM Glucose Ranges (Study 1, Adult)
{17}------------------------------------------------
Image /page/17/Picture/0 description: The image contains the logo for Abbott Diabetes Care. The logo consists of a blue abstract symbol on the left, followed by the text "Abbott" in bold, and "Diabetes Care" in a smaller font size below "Abbott". The text is black, contrasting with the blue symbol.
| Accuracy to YSI within CGM Glucose Ranges (Study 2, Pediatric*) | |||||
|---|---|---|---|---|---|
| CGMGlucoseLevel(mg/dL) | Numberof CGM-ReferencePairs | PercentWithin±15mg/dL | PercentWithin±20mg/dL | PercentWithin±40mg/dL | PercentWithin±15% | PercentWithin±20% | PercentWithin±40% | Meanbias(mg/dL) | MARD(%) |
|---|---|---|---|---|---|---|---|---|---|
| <54 | 139 | 71.9 | 79.1 | 97.1 | -9.9 | 17.1 | |||
| 54-69 | 863 | 86.4 | 90.5 | 97.1 | -4.9 | 12.0 | |||
| 70-180 | 2690 | 77.4 | 87.6 | 98.7 | -3.4 | 10.6 | |||
| 181-250 | 1236 | 86.0 | 94.7 | 99.7 | -8.9 | 8.3 | |||
| >250 | 1618 | 92.2 | 97.7 | 99.8 | -2.2 | 7.2 |
- Results only include children 6-17 years of age with YSI measurement as comparator. SMBG measurements were obtained as comparator for children 4-5 years of age.
| YSIGlucoseLevel(mg/dL) | Numberof CGM-ReferencePairs | PercentWithin±15mg/dL | PercentWithin±20mg/dL | PercentWithin±40mg/dL | PercentWithin±15% | PercentWithin±20% | PercentWithin±40% | Meanbias(mg/dL) | MARD(%) |
|---|---|---|---|---|---|---|---|---|---|
| <54 | 440 | 91.1 | 97.5 | 100.0 | 7.4 | 15.5 | |||
| 54-69 | 3028 | 94.7 | 98.6 | 100.0 | 1.5 | 10.2 | |||
| 70-180 | 7504 | 77.5 | 86.9 | 99.4 | -4.8 | 10.4 | |||
| 181-250 | 2937 | 87.9 | 93.7 | 99.7 | -8.0 | 8.0 | |||
| >250 | 4826 | 90.9 | 95.9 | 99.7 | -11.8 | 6.9 |
Accuracy to YSI within YSI Glucose Ranges (Study 1, Adult)
Accuracy to YSI within YSI Glucose Ranges (Study 2, Pediatric*)
| YSIGlucoseLevel(mg/dL) | Numberof CGM-ReferencePairs | PercentWithin±15mg/dL | PercentWithin±20mg/dL | PercentWithin±40mg/dL | PercentWithin±15% | PercentWithin±20% | PercentWithin±40% | Meanbias(mg/dL) | MARD(%) |
|---|---|---|---|---|---|---|---|---|---|
| <54 | 131 | 93.9 | 98.5 | 100.0 | 6.6 | 14.2 | |||
| 54-69 | 751 | 96.5 | 98.8 | 100.0 | 1.0 | 9.3 | |||
| 70-180 | 2743 | 74.3 | 84.8 | 98.0 | -3.0 | 11.4 | |||
| 181-250 | 1104 | 86.6 | 92.9 | 99.0 | -3.9 | 8.4 | |||
| >250 | 1817 | 90.2 | 97.5 | 99.9 | -10.2 | 7.6 |
- Results only include children 6-17 years of age with YSI measurement as comparator. SMBG measurements were obtained as comparator for children 4-5 years of age.
{18}------------------------------------------------
Image /page/18/Picture/0 description: The image contains the logo for Abbott Diabetes Care. The logo consists of a stylized blue "a" on the left, followed by the text "Abbott" in bold black font on the top line and "Diabetes Care" in a smaller, regular black font on the second line. The logo is clean and professional, conveying a sense of trust and reliability.
Concurrence of iCGM values compared to the comparator method across the entire measuring range was also evaluated. iCGM glucose ranges of <40, 40-60, 61-80, 81-120, 121-160, 161-200, 201-250, 251-300, 351-400 and >400 mg/dL were evaluated against the comparator glucose ranges and percent of iCGM values within those ranges were reported.
| YSI (mg/dL) | |||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| CGM(mg/dL) | <40 | 40-60 | 61 -80 | 81 -120 | 121-160 | 161 -200 | 201-250 | 251 -300 | 301-350 | 351-400 | >400 | N | |
| <40 | 20.0 | 20.0 | 40.0 | 20.0 | . | . | . | . | . | . | . | 5 | |
| 40-60 | 0.4 | 52.9 | 43.3 | 3.3 | . | 0.1 | . | . | . | . | . | 1889 | |
| 61-80 | . | 18.9 | 62.7 | 18.1 | 0.4 | 0.0 | . | . | . | . | . | 3090 | |
| 81-120 | . | 0.2 | 11.0 | 70.1 | 17.8 | 0.8 | 0.1 | . | . | . | . | 3040 | |
| 121-160 | . | . | 0.1 | 9.1 | 69.9 | 18.9 | 1.6 | 0.3 | 0.2 | . | . | 2407 | |
| 161-200 | . | . | . | . | 10.6 | 60.6 | 26.9 | 1.6 | 0.3 | . | . | 1745 | |
| 201-250 | . | . | . | . | . | 7.0 | 65.5 | 25.6 | 1.9 | 0.1 | . | 2181 | |
| 251-300 | . | . | . | . | . | 0.1 | 8.4 | 66.9 | 22.7 | 1.8 | 0.1 | 2327 | |
| 301-350 | . | . | . | . | . | . | 0.4 | 13.6 | 68.8 | 16.0 | 1.2 | 1522 | |
| 351-400 | . | . | . | . | . | . | . | 0.6 | 27.5 | 63.3 | 8.6 | 534 | |
| >400 | . | . | . | . | . | . | . | . | 2.5 | 62.8 | 34.7 | 121 |
Concurrence Analysis by CGM Glucose Level (Study 1, Adult)
Concurrence Analysis by CGM Glucose Level (Study 2, Pediatric*)
| YSI (mg/dL) | ||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| CGM(mg/dL) | <40 | 40-60 | 61-80 | 81-120 | 121-160 | 161-200 | 201-250 | 251-300 | 301-350 | 351-400 | >400 | N |
| <40 | . | 50.0 | 50.0 | . | . | . | . | . | . | . | . | 2 |
| 40-60 | 0.6 | 48.6 | 42.5 | 7.8 | 0.6 | . | . | . | . | . | . | 527 |
| 61-80 | . | 12.1 | 61.9 | 24.3 | 1.7 | . | . | . | . | . | . | 915 |
| 81-120 | . | 0.2 | 11.2 | 69.0 | 18.2 | 1.3 | 0.1 | . | . | . | . | 1006 |
| 121-160 | . | . | . | 11.4 | 71.0 | 15.8 | 1.8 | . | . | . | . | 868 |
| 161-200 | . | . | . | 0.1 | 18.2 | 61.3 | 20.1 | 0.3 | . | . | . | 703 |
| 201-250 | . | . | . | . | 0.2 | 9.6 | 55.3 | 33.6 | 1.2 | 0.1 | . | 909 |
| 251-300 | . | . | . | . | . | 0.1 | 14.1 | 60.8 | 23.7 | 1.3 | . | 818 |
| 301-350 | . | . | . | . | . | . | 0.3 | 24.8 | 58.2 | 16.5 | 0.2 | 593 |
| 351-400 | . | . | . | . | . | 1.0 | . | 0.5 | 33.8 | 59.4 | 5.3 | 207 |
| >400 | . | . | . | . | . | . | . | 4.4 | 6.7 | 33.3 | 55.6 | 45 |
- Results only include children 6-17 years of age with YSI measurement as comparator. SMBG measurements were obtained as comparator for children 4-5 years of age.
{19}------------------------------------------------
Image /page/19/Picture/0 description: The image contains the logo for Abbott Diabetes Care. The logo consists of a stylized blue "a" on the left, followed by the text "Abbott" in bold, black letters. Below "Abbott" is the text "Diabetes Care" in a smaller, non-bold font, also in black.
| CGM (mg/dL) | |||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| YSI(mg/dL) | <40 | 40-60 | 61 -80 | 81 -120 | 121-160 | 161-200 | 201-250 | 251-300 | 301-350 | 351-400 | >400 | N | |
| <40 | 12.5 | 87.5 | . | . | . | . | . | . | . | . | . | 8 | |
| 40-60 | 0.1 | 62.9 | 36.6 | 0.4 | . | . | . | . | . | . | . | 1591 | |
| 61-80 | 0.1 | 26.4 | 62.6 | 10.8 | 0.1 | . | . | . | . | . | . | 3093 | |
| 81-120 | 0.0 | 2.1 | 18.8 | 71.7 | 7.3 | . | . | . | . | . | . | 2971 | |
| 121-160 | . | . | 0.5 | 22.3 | 69.6 | 7.7 | . | . | . | . | . | 2418 | |
| 161-200 | . | 0.1 | 0.1 | 1.5 | 26.9 | 62.5 | 9.0 | 0.1 | . | . | . | 1694 | |
| 201-250 | . | . | 0.1 | 1.8 | 21.9 | 66.8 | 9.1 | 0.3 | . | . | . | 2139 | |
| 251-300 | . | . | . | . | 0.3 | 1.2 | 23.7 | 66.0 | 8.8 | 0.1 | . | 2359 | |
| 301-350 | . | . | . | . | 0.3 | 0.3 | 2.3 | 29.8 | 58.9 | 8.3 | 0.2 | 1777 | |
| 351-400 | . | . | . | . | . | . | 0.3 | 6.1 | 34.7 | 48.1 | 10.8 | 703 | |
| >400 | . | . | . | . | . | . | . | 1.9 | 16.7 | 42.6 | 38.9 | 108 |
Concurrence Analysis by YSI Glucose Level (Study 1, Adult)
Concurrence Analysis by YSI Glucose Level (Study 2, Pediatric*)
| CGM (mg/dL) | ||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| YSI(mg/dL) | <40 | 40-60 | 61 -80 | 81 -120 | 121-160 | 161-200 | 201-250 | 251-300 | 301-350 | 351-400 | >400 | N |
| <40 | . | 100.0 | . | . | . | . | . | . | . | . | . | 3 |
| 40-60 | 0.3 | 69.2 | 30.0 | 0.5 | . | . | . | . | . | . | . | 370 |
| 61-80 | 0.1 | 24.8 | 62.6 | 12.5 | . | . | . | . | . | . | . | 904 |
| 81-120 | . | 3.9 | 21.0 | 65.7 | 9.4 | 0.1 | . | . | . | . | . | 1057 |
| 121-160 | . | 0.3 | 1.7 | 19.3 | 65.0 | 13.5 | 0.2 | . | . | . | . | 948 |
| 161-200 | . | . | . | 1.9 | 20.4 | 64.2 | 13.0 | 0.1 | . | 0.3 | . | 671 |
| 201-250 | . | . | 0.1 | 2.1 | 18.1 | 64.7 | 14.8 | 0.3 | . | . | . | 778 |
| 251-300 | . | . | . | . | . | 0.2 | 32.0 | 52.1 | 15.4 | 0.1 | 0.2 | 954 |
| 301-350 | . | . | . | . | . | . | 1.8 | 31.1 | 55.4 | 11.2 | 0.5 | 623 |
| 351-400 | . | . | . | . | . | . | 0.4 | 4.4 | 39.5 | 49.6 | 6.0 | 248 |
| >400 | . | . | . | . | . | . | . | . | 2.7 | 29.7 | 67.6 | 37 |
- Results only include children 6-17 years of age with YSI measurement as comparator. SMBG measurements were obtained as comparator for children 4-5 years of age.
{20}------------------------------------------------
Image /page/20/Picture/0 description: The image shows the logo for Abbott Diabetes Care. The logo consists of a blue abstract shape on the left, followed by the words "Abbott" in bold black letters on the top line and "Diabetes Care" in smaller, regular black letters on the second line. The logo is simple and professional, conveying a sense of trust and reliability.
Trend Accuracy
Trend accuracy describes the accuracy of the sensor during times of rapidly changing glucose and is characterized by slopes, such as from > 2mg/dL to <-2 mg/dL. Trend accuracy was assessed by the concurrence rate of the glucose rate of changes in mg/dL of glucose per minutes) determined by the iCGM trend arrow and the corresponding comparator values for each iCGM-comparator measurement pair.
| CGM(mg/dL/min) | YSI (mg/dL/min) | N | |||||
|---|---|---|---|---|---|---|---|
| <-2 | [-2, -1) | [-1, 0) | [0, 1] | (1, 2] | >2 | ||
| <-2 | 34.4 | 44.9 | 18.3 | 2.2 | 0.3 | . | 323 |
| [-2, -1) | 6.8 | 46.5 | 41.2 | 4.0 | 0.9 | 0.6 | 1090 |
| [-1, 0) | 1.2 | 8.3 | 67.1 | 19.7 | 2.6 | 1.2 | 9389 |
| [0, 1] | 0.9 | 3.4 | 26.0 | 46.9 | 15.5 | 7.3 | 5420 |
| (1, 2] | 0.1 | 1.7 | 7.7 | 31.6 | 38.4 | 20.5 | 1151 |
| >2 | 0.1 | 0.2 | 3.1 | 14.6 | 32.9 | 49.0 | 881 |
Concurrence Analysis by Glucose Rate of Change (Study 1, Adult)
| CGM(mg/dL/min) | <-2 | [-2, -1) | [-1, 0) | [0, 1] | (1, 2] | >2 | N |
|---|---|---|---|---|---|---|---|
| <-2 | 44.1 | 44.7 | 8.8 | 2.4 | . | . | 170 |
| [-2, -1) | 11.4 | 49.5 | 32.8 | 5.2 | 0.4 | 0.6 | 463 |
| [-1, 0) | 2.1 | 11.2 | 60.0 | 20.8 | 3.9 | 1.9 | 2587 |
| [0, 1] | 1.4 | 5.6 | 25.2 | 43.2 | 14.8 | 9.7 | 2095 |
| (1, 2] | 0.2 | 2.6 | 10.4 | 29.7 | 35.5 | 21.5 | 498 |
| >2 | . | 0.9 | 4.2 | 15.0 | 29.7 | 50.2 | 448 |
- Results only include children 6-17 years of age with YSI measurement as comparator. SMBG measurements were obtained as comparator for children 4-5 years of age.
Agreement When iCGM Reads 'LO' or 'HI'
The FreeStyle Libre 2 Glucose Monitoring System reports glucose readings between 40 and 400 mg/dL. When the system determines that the glucose reading is below 40 mg/dL, it will display 'LO' whenever the sensor is scanned. When the system determines that the glucose reading is above 400 mg/dL, it will display 'HI' whenever the sensor is scanned. Because the system does not display glucose values below 40 mg/dL or above 400 mg/dL, the comparisons to the actual blood glucose levels (as determined by the comparator) when the iCGM value is classified as 'LO' or 'HI' are evaluated separately. The cumulative percentages of when the comparator values were less than certain glucose values (for 'LO') and when comparator values were more that certain glucose values (for 'HI') are presented in the tables below.
{21}------------------------------------------------
Image /page/21/Picture/0 description: The image shows the logo for Abbott Diabetes Care. The logo consists of a blue abstract shape on the left, followed by the text "Abbott" in bold black font on the top line and "Diabetes Care" in a smaller, regular black font on the second line. The logo is simple and professional, conveying a sense of trust and reliability.
| Concurrence Analysis with 'LO' CGM Reading (Study 1: Adult) | ||||
|---|---|---|---|---|
| CGM-ReferencePairs | <50 | <60 | <70 | <80 | ≥ 80 | N |
|---|---|---|---|---|---|---|
| n | 1 | 2 | 2 | 4 | 1 | 5 |
| Cumulative% | 20.0 | 40.0 | 40.0 | 80.0 | 20.0 |
Concurrence Analysis with 'LO' CGM Reading (Study 2: Pediatric*)
| CGM-ReferencePairs | YSI Comparator (mg/dL) | N | ||||
|---|---|---|---|---|---|---|
| <50 | <60 | <70 | <80 | ≥ 80 | ||
| n | 0 | 1 | 2 | 2 | 0 | 2 |
| Cumulative% | 0.0 | 50.0 | 100.0 | 100.0 | 0.0 |
- Results only include children 6-17 years of age with YSI measurement as comparator. SMBG measurements were obtained as comparator for children 4-5 years of age.
Concurrence Analysis with 'HI' CGM Reading (Study 1: Adult)
| CGM-Reference Pairs | YSI Comparator (mg/dL) | N | |||
|---|---|---|---|---|---|
| >350 | >300 | >250 | ≤250 | ||
| n | 118 | 121 | 121 | 0 | 121 |
| Cumulative% | 97.5 | 100.0 | 100.0 | 0.0 |
Concurrence Analysis with 'HI' CGM Reading (Study 2: Pediatric*)
| CGM-ReferencePairs | YSI Comparator (mg/dL) | N | |||
|---|---|---|---|---|---|
| >350 | >300 | >250 | ≤250 | ||
| n | 40 | 43 | 45 | 0 | 45 |
| Cumulative% | 88.9 | 95.6 | 100.0 | 0.0 |
- Results only include children 6-17 years of age with YSI measurement as comparator. SMBG measurements were obtained as comparator for children 4-5 years of age.
{22}------------------------------------------------
Image /page/22/Picture/0 description: The image shows the logo for Abbott Diabetes Care. The logo consists of a blue square with a white "A" inside, followed by the words "Abbott" in bold black letters and "Diabetes Care" in smaller black letters below it. The logo is simple and modern, and the colors are clean and professional.
Alarm Performance
The tables in this section show the accuracy of the System's Low and High Glucose Alarms. The Alarm Rate tells you how often the alarm is right or wrong. The Detection Rate tells you how often the System recognizes and notifies you about a low or high glucose event.
Low Glucose Alarm Performance
True Alarm Rate
Tells you: When you got a low glucose alarm, were you actually low? Definition: Percentage of time the alarm issued and blood glucose was below the alarm level within 15 minutes before or after the alarm.
False Alarm Rate
Tells you: Did you get a low glucose alarm that you shouldn't have? Definition: Percentage of time the alarm issued and blood glucose was not below the alarm level within 15 minutes before or after the alarm.
Detection Rate
Tells you: When you were low, did you get a low glucose alarm? Definition: Percentage of time blood glucose was below the alarm level and the alarm issued within 15 minutes before or after the glucose event.
Missed Detection Rate
Tells you: When you were low, did you miss a low glucose alarm? Definition: Percentage of time blood glucose was below the alarm level and the alarm didn't issue within 15 minutes before or after the glucose event.
For example, the Adult study found that for a Low Glucose alarm level set to 70 mg/dL: 86.0% of the time a low glucose alarm was received when blood glucose was below the alarm level but 14.0% of the time a low glucose alarm was received when blood glucose was not below the alarm level.
89.3% of the time blood glucose was below the alarm level and a low glucose alarm was appropriately issued but 10.7% of the time the glucose event was missed and no alarm was issued.
{23}------------------------------------------------
Image /page/23/Picture/0 description: The image contains the logo for Abbott Diabetes Care. The logo consists of a blue abstract shape on the left, followed by the text "Abbott" in bold, and "Diabetes Care" in a smaller font size below "Abbott". The text is black.
| Alarm Rate | Detection Rate | |||||
|---|---|---|---|---|---|---|
| Low GlucoseAlarm level(mg/dL) | Number ofEvents (n) | TrueAlarmRate (%) | FalseAlarmRate (%) | Number ofEvents (n) | CorrectDetection Rate(%) | MissedDetection Rate(%) |
| 60 | 9861 | 72.6 | 27.4 | 1527 | 75.7 | 24.3 |
| 70 | 21504 | 86.0 | 14.0 | 3652 | 89.3 | 10.7 |
| 80 | 32784 | 91.3 | 8.7 | 4753 | 97.3 | 2.7 |
| 90 | 41299 | 93.6 | 6.4 | 5591 | 98.5 | 1.5 |
Low Glucose Alarm Performance (Adult)
Low Glucose Alarm Performance (Pediatric*)
| Low GlucoseAlarm level(mg/dL) | Number ofEvents (n) | TrueAlarmRate (%) | FalseAlarmRate (%) | Number ofEvents (n) | CorrectDetection Rate(%) | MissedDetection Rate(%) |
|---|---|---|---|---|---|---|
| 60 | 2780 | 62.9 | 37.1 | 373 | 87.4 | 12.6 |
| 70 | 6363 | 80.3 | 19.7 | 963 | 93.5 | 6.5 |
| 80 | 9747 | 85.6 | 14.4 | 1318 | 96.4 | 3.6 |
| 90 | 12550 | 92.2 | 7.8 | 1656 | 97.3 | 2.7 |
- Results only include children 6-17 years of age with YSI measurement as comparator. SMBG measurements were obtained as comparator for children 4-5 years of age.
High Glucose Alarm Performance
True Alarm Rate
Tells you: When you got a high glucose alarm, were you actually high? Definition: Percentage of time the alarm issued and blood glucose was above the alarm level within 15 minutes before or after the alarm.
False Alarm Rate
Tells you: Did you get a high glucose alarm that you shouldn't have? Definition: Percentage of time the alarm issued and blood glucose was not above the alarm level within 15 minutes before or after the alarm.
Detection Rate
Tells you: When you were high, did you get a high glucose alarm? Definition: Percentage of time blood glucose was above the alarm level and the alarm issued within 15 minutes before or after the glucose event.
Missed Detection Rate
Tells you: When you were high, did you miss a high glucose alarm?
Definition: Amount of time blood glucose was above the alarm level and the alarm didn't issue within 15 minutes before or after the glucose event.
{24}------------------------------------------------
Image /page/24/Picture/0 description: The image contains the logo for Abbott Diabetes Care. The logo consists of a blue stylized letter "a" on the left, followed by the text "Abbott" in bold black font on the right. Below "Abbott" is the text "Diabetes Care" in a smaller, non-bold black font.
For example, the Adult study found that for a High Glucose alarm level set to 200 mg/dL: 99.2% of the time a high glucose alarm was received when blood glucose was above the alarm level but 0.8% of the time a high glucose alarm was received when blood glucose was not above the alarm level.
97.1% of the time blood glucose was above the alarm level and a high glucose alarm was appropriately issued but 2.9% of the time the glucose event was missed and no alarm was issued.
| High GlucoseAlarm level(mg/dL) | Alarm Rate | Detection Rate | ||||
|---|---|---|---|---|---|---|
| Number ofEvents (n) | TrueAlarmRate (%) | FalseAlarmRate (%) | Number ofEvents (n) | CorrectDetection Rate(%) | MissedDetection Rate(%) | |
| 120 | 105544 | 99.1 | 0.9 | 11417 | 98.2 | 1.8 |
| 140 | 93574 | 99.1 | 0.9 | 10152 | 98.1 | 1.9 |
| 180 | 74290 | 99.2 | 0.8 | 8080 | 97.8 | 2.2 |
| 200 | 66039 | 99.2 | 0.8 | 7269 | 97.1 | 2.9 |
| 220 | 57549 | 99.0 | 1.0 | 6390 | 96.9 | 3.1 |
| 240 | 48733 | 98.4 | 1.6 | 5550 | 95.6 | 4.4 |
| 300 | 21512 | 96.3 | 3.7 | 2672 | 90.0 | 10.0 |
High Glucose Alarm Performance (Adult)
| Alarm Rate | Detection Rate | |||||
|---|---|---|---|---|---|---|
| High GlucoseAlarm level(mg/dL) | Number ofEvents (n) | TrueAlarmRate (%) | FalseAlarmRate (%) | Number ofEvents (n) | CorrectDetection Rate(%) | MissedDetection Rate(%) |
| 120 | 34176 | 98.8 | 1.2 | 4441 | 98.2 | 1.8 |
| 140 | 30107 | 98.0 | 2.0 | 3945 | 98.4 | 1.6 |
| 180 | 22430 | 98.4 | 1.6 | 3125 | 98.0 | 2.0 |
| 200 | 19425 | 98.0 | 2.0 | 2791 | 98.0 | 2.0 |
| 220 | 16371 | 98.2 | 1.8 | 2492 | 96.9 | 3.1 |
| 240 | 13559 | 98.0 | 2.0 | 2172 | 95.7 | 4.3 |
| 300 | 6064 | 90.8 | 9.2 | 962 | 91.0 | 9.0 |
- Results only include children 6-17 years of age with YSI measurement as comparator. SMBG measurements were obtained as comparator for children 4-5 years of age.
{25}------------------------------------------------
Image /page/25/Picture/0 description: The image shows the logo for Abbott Diabetes Care. The logo consists of a blue stylized "a" on the left, followed by the text "Abbott" in bold black font on the top line and "Diabetes Care" in a smaller, regular black font on the bottom line. The logo is simple and professional, conveying a sense of trust and reliability.
Sensor Stability
Sensor stability describes the performance of the sensor lifetime. Sensors can be worn for up to 14 days. Performance was estimated by calculating the percentage of FreeStyle Libre 2 readings within 15 mg/dL or 15% (15/15%), 20 mg/dL or 20% (20/20%) and 40 mg/dL or 405 (40/40%) of the comparator values at the beginning Adult: Days 1-3, Pediatric: Day 1-2) Early Middle (Adult: Days 7-8, Pediatric: Day 7-8), Late Middle (Adult: Days 9-12, Pediatric: Day 9-12), and End (Adult: Days 13-14, Pediatric: Day 13-14). The mean of the absolute relative differences was evaluated over 14 day life of the sensor within measuring range.
| Wear Period | Number ofCGM-referencepairs | MARD (%) | Within ±15%/ ±15mg/dL | Within ±20%/ ±20mg/dL | Within ±40%/ ±40mg/dL |
|---|---|---|---|---|---|
| Beginning | 6955 | 9.9 | 83.4 | 90.4 | 99.3 |
| Early Middle | 4522 | 8.5 | 87.7 | 94.5 | 99.8 |
| Late Middle | 3503 | 8.8 | 86.8 | 93.4 | 99.7 |
| End | 3755 | 9.1 | 86.4 | 92.9 | 100.0 |
Sensor Accuracy Relative to YSI Comparator over the wear duration (Study 1: Adult)
| Sensor Accuracy Relative to YSI Comparator over the wear duration (Study 2: Pediatric*) | |||
|---|---|---|---|
| ----------------------------------------------------------------------------------------- | -- | -- | -- |
| Wear Period | Number ofCGM-referencepairs | MARD (%) | Within ±15%/ ±15mg/dL | Within ±20%/ ±20mg/dL | Within ±40%/ ±40mg/dL |
|---|---|---|---|---|---|
| Beginning | 1828 | 10.7 | 79.6 | 88.5 | 98.6 |
| Early Middle | 1642 | 8.0 | 89.5 | 94.2 | 98.5 |
| Late Middle | 1534 | 9.7 | 83.6 | 92.9 | 99.5 |
| End | 1542 | 10.2 | 82.6 | 91.1 | 99.3 |
- Results only include children 6-17 years of age with YSI measurement as comparator. SMBG measurements were obtained as comparator for children 4-5 years of age.
{26}------------------------------------------------
Image /page/26/Picture/0 description: The image contains the logo for Abbott Diabetes Care. The logo consists of a stylized blue "a" on the left, followed by the text "Abbott" in bold, black letters. Below "Abbott" is the text "Diabetes Care" in a smaller, non-bold font, also in black.
Sensor Life
The Sensor can be worn for up to 14 days. 146 Sensors were evaluated in the Adult Study and 139 Sensors were evaluated in the Pediatric Study to determine how many days of readings each Sensor provided. Of the 146 Sensors in the Adult study, 71.1% lasted until the final day of use. 6 Sensors (4.1%) had "early sensor shut-off" where the Sensor algorithm detected that the Sensors did not function as intended and presented the user with a "Replace Sensor'' message. In the Pediatric study, 78.1% of the Sensors lasted until the final day of use. 3 Sensors (2.2%) had "early sensor shut-off" where the Sensor algorithm detected that the Sensors did not function as intended and presented the user with a "Replace Sensor" message.
| Day of Wear | Number of Sensors | Survival Rate (%) |
|---|---|---|
| 1 | 145 | 99.3 |
| 2 | 142 | 97.3 |
| 3 | 140 | 95.9 |
| 4 | 137 | 93.8 |
| 5 | 134 | 93.8 |
| 6 | 133 | 91.1 |
| 7 | 132 | 90.4 |
| 8 | 127 | 87.0 |
| 9 | 123 | 84.9 |
| 10 | 119 | 82.2 |
| 11 | 112 | 77.3 |
| 12 | 111 | 76.6 |
| 13 | 104 | 71.8 |
| 14 | 100 | 71.1 |
Sensor Survival Rate Over Wear Duration (Study 1: Adult)
{27}------------------------------------------------
Image /page/27/Picture/0 description: The image shows the logo for Abbott Diabetes Care. The logo consists of a blue stylized "a" on the left, followed by the words "Abbott" in bold black font on the top line and "Diabetes Care" in a smaller, regular black font on the bottom line. The logo is simple and clean, with a focus on the company name and its area of expertise.
| Day of Wear | Number of Sensors | Survival Rate (%) |
|---|---|---|
| 1 | 137 | 98.6 |
| 2 | 136 | 97.8 |
| 3 | 134 | 97.1 |
| 4 | 133 | 96.4 |
| 5 | 133 | 96.4 |
| 6 | 133 | 96.4 |
| 7 | 133 | 96.4 |
| 8 | 131 | 94.9 |
| 9 | 126 | 91.3 |
| 10 | 124 | 89.9 |
| 11 | 122 | 88.4 |
| 12 | 120 | 87.0 |
| 13 | 114 | 83.4 |
| 14 | 104 | 78.1 |
Sensor Survival Rate Over Wear Duration (Study 2: Pediatric)
{28}------------------------------------------------
Image /page/28/Picture/0 description: The image contains the logo for Abbott Diabetes Care. The logo consists of a blue abstract symbol on the left, followed by the text "Abbott" in bold, and "Diabetes Care" in a smaller font size below it. The text is black, contrasting with the blue symbol.
Glucose Reading Availability
The System is designed to show a Sensor glucose reading after each scan that is performed throughout the wear period after the start-up time. As such, the capture rate characterizes the reliability of the communication between components of the system.
| Glucose Reading Capture Rate Over Wear Duration (Study 1: Adult) | ||
|---|---|---|
| Day of Wear | Number of Sensors | Capture Rate (%) |
| 1 | 146 | 98.3 |
| 2 | 145 | 98.1 |
| 3 | 143 | 98.3 |
| 4 | 140 | 98.3 |
| 5 | 138 | 98.4 |
| 6 | 135 | 98.3 |
| 7 | 134 | 98.4 |
| 8 | 131 | 98.4 |
| 9 | 128 | 98.4 |
| 10 | 123 | 98.4 |
| 11 | 120 | 98.4 |
| 12 | 113 | 98.5 |
| 13 | 112 | 98.5 |
| 14 | 104 | 98.6 |
Glucose Reading Capture Rate Over Wear Duration (Study 1: Adult)
{29}------------------------------------------------
Image /page/29/Picture/0 description: The image contains the logo for Abbott Diabetes Care. The logo consists of a stylized blue "a" on the left, followed by the text "Abbott" in bold black font on the top line and "Diabetes Care" in a smaller, regular black font on the second line. The logo is simple and professional, conveying a sense of trust and reliability.
| Day of Wear | Number of Sensors | Survival Rate (%) |
|---|---|---|
| 1 | 139 | 94.6 |
| 2 | 137 | 94.9 |
| 3 | 136 | 95.2 |
| 4 | 133 | 95.3 |
| 5 | 134 | 95.5 |
| 6 | 133 | 95.6 |
| 7 | 133 | 96.0 |
| 8 | 133 | 95.9 |
| 9 | 130 | 95.7 |
| 10 | 125 | 95.6 |
| 11 | 125 | 95.6 |
| 12 | 122 | 95.8 |
| 13 | 119 | 95.9 |
| 14 | 116 | 95.8 |
Glucose Reading Capture Rate Over Wear Duration (Study 2: Pediatric)
{30}------------------------------------------------
Image /page/30/Picture/0 description: The image contains the logo for Abbott Diabetes Care. The logo consists of a blue square with a stylized "a" inside, followed by the text "Abbott" in bold, black letters. Below "Abbott" is the text "Diabetes Care" in a smaller, non-bold font, also in black.
Human Factors 1.7.4
Human factors and usability testing of the FreeStyle Libre 2 System was conducted to determine whether the user interface design and labeling would impact the performance of the device. Human factors testing was conducted in accordance with:
- FDA Guidance titled Applying Human Factors and Usability Engineering to Medical Devices, dated February 3, 2016
- ANSI/AAMI/IEC 62366: Medical devices Application of Usability Engineering to ● Medical Devices
- IEC 60601-1-6: Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability
1.7.5 Software Verification and Validation
Software verification and validation testing was conducted to confirm that the software used in the FreeStyle Libre 2 System performed in accordance with established specifications, EN 62304 and FDA Guidance document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices," May 11, 2005. Evaluation activities included unit, system integration (SIT), and system level testing which verified functionality of the device against established software requirements. Results of the software executed protocols for FreeStyle Libre 2 met the acceptance criteria and therefore supports that the System's embedded software is acceptable for its intended use.
1.7.6 Biocompatibility
Biocompatibility testing in accordance with ISO10993-1and FDA Guidance "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices Part 1: Evaluation and testing within a risk management process," June 16, 2016 was performed on Sensor materials including the outer Sensor, casing, adhesive pad, and sensor tail. Biological evaluation included cytotoxicity, genotoxicity, irritation, sensitization, and system toxicity testing. All biocompatibility testing met the acceptance criteria.
Sterilization 1.7.7
Electron beam sterilization vas performed per ISO11137-1 and ISO 11137-2. Sterilization validation confirmed that the Sterility Assurance Level (SAL) of 10-6 is achieved with the selected target dose of 25kGy. The sterilization dose was established by the VDmax25 method described in ISO 11137-2. All sterilization testing met the acceptance criteria and supports sterility of the patient contact components of the device.
{31}------------------------------------------------
Image /page/31/Picture/0 description: The image contains the logo for Abbott Diabetes Care. On the left is a blue stylized letter 'a'. To the right of the logo is the text 'Abbott' in bold black font, with the text 'Diabetes Care' in a smaller, non-bold black font below it.
Shelf-Life and Stability 1.7.8
Sensor Kit shelf life stability testing provided objective evidence to support a shelf life of 9 months. The testing conducted satisfied specifications. Additional testing was conducted to support sensor kit storage temperature range of 36°F and 82°F. Results demonstrated no impact to sensor stability when stored at these storage temperatures.
1.7.9 Packaging Integrity/Shipping Integrity
Packaging and shipping integrity testing was performed for Sensor Kits and Reader Kits in accordance with EN ISO 11607-1, EN ISO 11607-2, ASTM D4169-14, and ASTM D4332-13 guidelines. All tested units passed the testing requirements of all distribution tests.
1.7.10 Electromagnetic Compatibility
Electromagnetic compatibility (EMC) testing was performed for the FreeStyle Libre 2 System to verify that the system is able to withstand the electromagnetic interference in compliance with IEC 60601-1-2, IEC CISPR 11 and Federal Communication Commission Regulations Part 15.209 and 15.225. EMC coexistence testing was also performed to confirm that the Reader and Patch remain functional and perform within acceptable limits while in the presence of common radiating electronic devices in accordance with FDA guidance Radio Frequency Wireless Technology in Medical Devices. Testing was also performed to demonstrate compliance of the FreeStyle Libre 2 System with Category M of RTCA DO-160:
- The FreeStyle Libre 2 Reader and Patch underwent electromagnetic compatibility (EMC) and electromagnetic immunity (EMI) testing and both demonstrated compliance with IEC 60601-1-2:2014.
- The FreeStyle Libre 2 Reader and Patch were exposed to electrostatic events (i.e., static electricity discharges from operators directly and from personnel to adjacent objects) in accordance with IEC 61000-4-2 Edition. Electrostatic discharge testing was performed at ± 8 kV for contact discharge and at ± 15 kV for air discharge. The FreeStyle Libre 2 System met the performance requirements for the ESD Immunity Test.
- Testing was performed to determine compliance with Federal Communications ● Commission (FCC) standards. The FreeStyle Libre 2 System successfully demonstrated compliance with FCC Part 15 Subpart C §15.247 (2016) and FCC Part 15 Subpart B (2016).
- The FreeStyle Libre 2 System demonstrated compliance with airworthiness . requirements per the Federal Aviation Administration (FAA) Advisory Circular RTCA/DO-160 Edition G section 21, Category M (RF Emission specification).
- The FreeStyle Libre 2 System underwent coexistence testing in the presence of . common RF interfering devices that are likely to be encountered by users in a home environment. A representative set of devices known to operate in the same frequency band (2.4 GHz) was selected. The test results showed that the FreeStyle Libre 2 System could tolerate interference generated by these RF interfering devices and still meet the target performance criteria.
{32}------------------------------------------------
Image /page/32/Picture/0 description: The image shows the logo for Abbott Diabetes Care. The logo consists of a stylized blue letter "a" on the left, followed by the text "Abbott" in bold black font on the top line and "Diabetes Care" in a smaller, regular black font on the second line. The logo is clean and professional, conveying a sense of trust and reliability.
- Conducted emissions and radiated emissions testing for the FreeStyle Libre 2 System ● was performed in accordance with IEC/EN 60601-1-2:2014 and CISPR 11. The FreeStyle Libre 2 System demonstrated that maximum emissions did not exceed the limits established for residential or home use (Class B).
1.7.11 Electrical Safety
The basic safety and essential performance of the FreeStyle Libre 2 System was evaluated to IEC 60601-1:2006/A1:2013. Tested units included the FreeStyle Libre 2 Reader and Patch, and both demonstrated compliance to the requirements of IEC 60601-1:2006/A1:2013.
1.7.12 Environmental Testing
Environmental testing on the FreeStyle Libre 2 System was performed in accordance with IEC 60601-1 to ensure the device specifications for operating temperature, operating humidity, operating pressure, impact resistance, vibration resistance, shock resistance, drop resistance, and storage conditions were met.
1.7.13 Test Strip Claim Support
The FreeStyle Precision Neo Blood Glucose Test Strips cleared under K171941 are compatible for use with the FreeStyle Libre 2 built-in meter for quantitative measurement of glucose from fresh capillary whole blood drawn from fingertips. Studies to support the use of the FreeStyle Libre 2 Reader's built-in meter with the FreeStyle Precision Neo Blood Glucose Test Strips (K171941) demonstrated that all acceptance criteria were met as defined by specification.
1.7.14 Interoperability
The FreeStyle Libre 2 System incorporates an approach for interoperability developed in alignment with FDA guidance, "Design Considerations and Pre-market Submission Recommendations for Interoperable Medical Devices," September 6, 2017, which includes expectations, requirements, and interface specifications to potential interoperable devices. In addition, the ADC approach to interoperability includes working with connected device companies regarding contractual approaches, interfaces for data communication and exchange, and post-market reporting procedures and responsibilities with respect to managing complaints (e.g., who is responsible for investigating and reporting complaints, malfunctions, and adverse events).
1.7.15 Cybersecurity
ADC has provided cybersecurity risk management documentation for the FreeStyle Libre 2 System that includes analysis of confidentiality, integrity, and availability for data, information and software related to the System. For each identified threat and vulnerability risk event scenario, risk assessment of impact to confidentiality, integrity, and availability was performed and documented within the cybersecurity risk management documentation. Appropriate risk mitigation controls have been implemented and tested.
{33}------------------------------------------------
Image /page/33/Picture/0 description: The image contains the logo for Abbott Diabetes Care. The logo consists of a blue stylized letter "a" on the left, followed by the text "Abbott" in bold black font on the right. Below "Abbott" is the text "Diabetes Care" in a smaller, non-bold black font.
In addition, ADC has controls and processes in place to ensure continued support for keeping the device secure and to ensure that the device firmware, software and components are malware free. Additional controls are also in place in manufacturing through distribution to ensure that the medical device firmware and software are malware free from point of origin to the hands of the end user.
1.8 Proposed Labeling
The FreeStyle Libre 2 Flash Glucose Monitoring System labeling is sufficient and satisfies the requirements of 21 CFR Parts 801 and 809, and the special controls for this type of device.
1.9 Conclusion
The FreeStyle Libre 2 Flash Glucose Monitoring System has the same intended use and clinical application as the predicate device. There are no differences in technological characteristics that raise questions of safety or effectiveness. The FreeStyle Libre 2 System provides significant benefits to users (including glucose readings, trends, real time alerts, a short sensor warm-up period, factory calibration, and the ability to autonomously communicate with digitally connected devices where the user manually controls actions for therapy decisions) that outweigh any potential risks associated with iCGM technology. Clinical studies in adult and pediatric patients have demonstrated that the FreeStyle Libre 2 System provides accurate and reliable glucose readings in accordance with the iCGM special controls relating to accuracy and does not post any clinically significant risks not seen with the predicate device. Accordingly, based on this and the data provided in this pre-market notification, the subject device and predicate device have been shown to be substantially equivalent.
§ 862.1355 Integrated continuous glucose monitoring system.
(a)
Identification. An integrated continuous glucose monitoring system (iCGM) is intended to automatically measure glucose in bodily fluids continuously or frequently for a specified period of time. iCGM systems are designed to reliably and securely transmit glucose measurement data to digitally connected devices, including automated insulin dosing systems, and are intended to be used alone or in conjunction with these digitally connected medical devices for the purpose of managing a disease or condition related to glycemic control.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Design verification and validation must include the following:
(i) Robust clinical data demonstrating the accuracy of the device in the intended use population.
(ii) The clinical data must include a comparison between iCGM values and blood glucose values in specimens collected in parallel that are measured on an FDA-accepted laboratory-based glucose measurement method that is precise and accurate, and that is traceable to a higher order (
e.g., an internationally recognized reference material and/or method).(iii) The clinical data must be obtained from a clinical study designed to fully represent the performance of the device throughout the intended use population and throughout the measuring range of the device.
(iv) Clinical study results must demonstrate consistent analytical and clinical performance throughout the sensor wear period.
(v) Clinical study results in the adult population must meet the following performance requirements:
(A) For all iCGM measurements less than 70 milligrams/deciliter (mg/dL), the percentage of iCGM measurements within ±15 mg/dL of the corresponding blood glucose value must be calculated, and the lower one-sided 95 percent confidence bound must exceed 85 percent.
(B) For all iCGM measurements from 70 mg/dL to 180 mg/dL, the percentage of iCGM measurements within ±15 percent of the corresponding blood glucose value must be calculated, and the lower one-sided 95 percent confidence bound must exceed 70 percent.
(C) For all iCGM measurements greater than 180 mg/dL, the percentage of iCGM measurements within ±15 percent of the corresponding blood glucose value must be calculated, and the lower one-sided 95 percent confidence bound must exceed 80 percent.
(D) For all iCGM measurements less than 70 mg/dL, the percentage of iCGM measurements within ±40 mg/dL of the corresponding blood glucose value must be calculated, and the lower one-sided 95 percent confidence bound must exceed 98 percent.
(E) For all iCGM measurements from 70 mg/dL to 180 mg/dL, the percentage of iCGM measurements within ±40 percent of the corresponding blood glucose value must be calculated, and the lower one-sided 95 percent confidence bound must exceed 99 percent.
(F) For all iCGM measurements greater than180 mg/dL, the percentage of iCGM measurements within ±40 percent of the corresponding blood glucose value must be calculated, and the lower one-sided 95 percent confidence bound must exceed 99 percent.
(G) Throughout the device measuring range, the percentage of iCGM measurements within ±20 percent of the corresponding blood glucose value must be calculated, and the lower one-sided 95 percent confidence bound must exceed 87 percent.
(H) When iCGM values are less than 70 mg/dL, no corresponding blood glucose value shall read above 180 mg/dL.
(I) When iCGM values are greater than 180 mg/dL, no corresponding blood glucose value shall read less than 70 mg/dL.
(J) There shall be no more than 1 percent of iCGM measurements that indicate a positive glucose rate of change greater than 1 mg/dL per minute (/min) when the corresponding true negative glucose rate of change is less than −2 mg/dL/min as determined by the corresponding blood glucose measurements.
(K) There shall be no more than 1 percent of iCGM measurements that indicate a negative glucose rate of change less than −1 mg/dL/min when the corresponding true positive glucose rate of change is greater than 2 mg/dL/min as determined by the corresponding blood glucose measurements.
(vi) Data demonstrating similar accuracy and rate of change performance of the iCGM in the pediatric population as compared to that in the adult population, or alternatively a clinical and/or technical justification for why pediatric data are not needed, must be provided and determined by FDA to be acceptable and appropriate.
(vii) Data must demonstrate that throughout the claimed sensor life, the device does not allow clinically significant gaps in sensor data availability that would prevent any digitally connected devices from achieving their intended use.
(2) Design verification and validation must include a detailed strategy to ensure secure and reliable means of iCGM data transmission to provide real-time glucose readings at clinically meaningful time intervals to devices intended to receive the iCGM glucose data.
(3) Design verification and validation must include adequate controls established during manufacturing and at product release to ensure the released product meets the performance specifications as defined in paragraphs (b)(1) and (b)(2) of this section.
(4) The device must demonstrate clinically acceptable performance in the presence of clinically relevant levels of potential interfering substances that are reasonably present in the intended use population, including but not limited to endogenous substances and metabolites, foods, dietary supplements, and medications.
(5) The device must include appropriate measures to ensure that disposable sensors cannot be used beyond its claimed sensor wear period.
(6) Design verification and validation must include results obtained through a usability study that demonstrates that the intended user can use the device safely and obtain the expected glucose measurement accuracy.
(7) The labeling required under § 809.10(b) of this chapter must include a separate description of the following sensor performance data observed in the clinical study performed in conformance with paragraph (b)(1) of this section for each intended use population, in addition to separate sensor performance data for each different iCGM insertion or use sites (
e.g., abdomen, arm, buttock):(i) A description of the accuracy in the following blood glucose concentration ranges: less than 54 mg/dL, 54 mg/dL to less than 70 mg/dL, 70 to 180 mg/dL, greater than 180 to 250 mg/dL, and greater than 250 mg/dL.
(ii) A description of the accuracy of positive and negative rate of change data.
(iii) A description of the frequency and duration of gaps in sensor data.
(iv) A description of the true, false, missed, and correct alert rates and a description of the available glucose concentration alert settings, if applicable.
(v) A description of the observed duration of iCGM life for the device.