(64 days)
The Motion Correction System is an accessory to a Magnetic Resonance Imaging (MRI) scanner. The system is a sensorbased technology for tracking of patient movement during an MRI session. It provides the current position of the patient in real-time to the MRI scanner for further data processing.
User: Professional Use
Environment: Hospital, doctor's office, or any facility that uses an MRI device
KinetiCor defines the Motion Correction System as an accessory to a medical device. It is intended to supplement and/or augment the performance of the MAGNETOM Skyra 3T MRI scanner.
During an MRI scan, if the patient's head does not remain still during the scan, the resulting images may not be sharp. This could prevent a physician from analyzing the images, therefore requiring additional scanning.
Software has been developed to somewhat reduce these effects.
KinetiCor has developed a Motion Correction technology using a track the true position of the patient's head. These coordinates are provided to the MRI scanner, which adjusts the MR images.
The Motion Correction System delivers head pose tracking and correction in 6 degrees of freedom for MR applications.
- . Optimized for high field MR environments, the technology has been FDA Cleared for use in the MAGNETOM Vida, 510(k) K183254, cleared January 18, 2019. Customized optics, electronics and enclosures minimize Electromagnetic Interference (EMI) and Radiofrequency (RF) interference and isolate the camera system for stable tracking accuracy.
- . Proprietary optical machine vision tracking technology provides tracking data to the MRI scanner with a static accuracy of approximately 0.1 mm and 0.1 deg. This means that the MR pulse sequences are updated in real-time.
- The impact of the Motion Correction technology can be illustrated by taking a closer look at the corrected and uncorrected images in the figure below.
The Motion Correction System Hardware consists of the following:
- Camera Enclosure (incorporating four Cameras) located inside the distal end of the MRI scanner; ●
- Wiring Conduit - provides a protected location for wiring to exit the distal end of the MRI scanner;
- Processing Unit - electronics with software, located outside the MRI room;
- External Wiring - used to connect processing unit to Camera Enclosure;
- . Tracking Markers - single use markers located on patient nose bridge (Referred to as "Nose Marker for Inline Motion Correction" in predicate device);
- External Monitor - if requested, located inside MRI Control Room;
- Network System located outside the MRI room.
Here's an analysis of the acceptance criteria and study information for the KinetiCor Motion Correction System (K193324) based on the provided text, structured as requested:
1. Table of Acceptance Criteria and Reported Device Performance
The provided text does not explicitly state specific quantitative acceptance criteria or a reported device performance table in the typical format of a clinical study outcome. Instead, it describes the technology's performance capabilities and its substantial equivalence to a predicate device.
However, based on the description, we can infer some performance claims:
| Acceptance Criteria (Inferred from device description) | Reported Device Performance (From device description) |
|---|---|
| Static Tracking Accuracy (position) | Approximately 0.1 mm |
| Static Tracking Accuracy (angle/degrees of freedom) | Approximately 0.1 deg (for 6 degrees of freedom) |
| Real-time update of MR pulse sequences | MR pulse sequences are updated in real-time based on tracking data |
| Ability to correct for patient head motion | Motion correction using KinetiCor technology results in sharper MR images |
| Equivalence to predicate device for motion correction | Provides the same motion correction hardware and technology as the predicate device |
| Safety and Effectiveness | Device is safe, as effective, and performs as well as or better than the predicate device |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size for Test Set: The document does not specify a numerical sample size for a test set in terms of patient numbers or a dataset size for evaluating device performance. The non-clinical tests mention "Software Verification and Validation testing" and "Design Verification and Validation testing" but don't provide details on the data used.
- Data Provenance: The document does not specify the country of origin of data or whether it was retrospective or prospective for any performance evaluation directly linked to the KinetiCor Motion Correction System. The comparative images (corrected vs. uncorrected MRI scans) are presented as illustrative examples, not as part of a formal study with specified provenance.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
The document does not provide information on the number of experts, their qualifications, or how ground truth was established by experts for a specific test set. The visual comparison of "corrected" and "uncorrected" images implies a qualitative assessment of image quality, but this isn't attributed to a formal expert review process with specified criteria or adjudicators.
4. Adjudication Method for the Test Set
The document does not describe any adjudication method (e.g., 2+1, 3+1) for a test set.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
The document does not mention a Multi-Reader Multi-Case (MRMC) comparative effectiveness study comparing human readers with and without AI assistance. The device is a motion correction system directly influencing image quality, rather than an AI diagnostic aid that assists human readers in interpretation.
6. Standalone (Algorithm Only Without Human-in-the-Loop) Performance
Yes, a standalone performance assessment was effectively done for the core function of the motion correction system, though not presented as a formal study in the typical sense of a diagnostic algorithm. The description states:
- "Proprietary optical machine vision tracking technology provides tracking data to the MRI scanner with a static accuracy of approximately 0.1 mm and 0.1 deg."
- "The MR pulse sequences are updated in real-time."
- The comparison of "Motion correction using KinetiCor technology" versus "MR images obtained when the motion was not corrected" (Figure on page 5) directly illustrates the system's performance in isolation from human interpretation. The effectiveness is demonstrated by the improved image quality produced by the system itself.
7. Type of Ground Truth Used
The ground truth for the motion correction capabilities appears to be based on:
- Quantitative measurements: The stated static accuracy of 0.1 mm and 0.1 deg implies a technical measurement against a known, precise reference or standard for validating the tracking technology.
- Visual comparison/Qualitative assessment: The "corrected" versus "uncorrected" image examples serve as a qualitative ground truth demonstrating the visual improvement in image quality due to the system's function. The "similar motion performed during both scans" indicates a control for comparing the effect of the system.
- The device is framed as an accessory to improve image quality, and its "effectiveness" is linked to this improvement, rather than a diagnostic outcome for which pathology or outcomes data would be the ground truth.
8. Sample Size for the Training Set
The document does not specify a sample size for any training set. It mentions "Proprietary optical machine vision tracking technology," which implies algorithms that might have been trained, but no details are provided.
9. How the Ground Truth for the Training Set Was Established
The document does not provide information on how the ground truth for any training set was established.
{0}------------------------------------------------
February 4, 2020
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
KinetiCor, Inc. % Mr. Scott R. Herr Sr. Director - RA/QA 3465 Waialae Avenue, Suite 300A HONOLULU HI 96816
Re: K193324
Trade/Device Name: Motion Correction System Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: Class II Product Code: LNH Dated: November 27, 2019 Received: December 6, 2019
Dear Mr. Herr:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
{1}------------------------------------------------
https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K193324
Device Name Motion Correction System
Indications for Use (Describe)
The Motion Correction System is an accessory to a Magnetic Resonance Imaging (MRI) scanner. The system is a sensorbased technology for tracking of patient movement during an MRI session. It provides the current position of the patient in real-time to the MRI scanner for further data processing.
User: Professional Use
Environment: Hospital, doctor's office, or any facility that uses an MRI device
KinetiCor defines the Motion Correction System as an accessory to a medical device. It is intended to supplement and/or augment the performance of the MAGNETOM Skyra 3T MRI scanner.
| Type of Use (Select one or both, as applicable) |
|---|
| ------------------------------------------------- |
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Kineticor. The logo is in blue and features the word "KINETICOR" in a stylized font. Below the main word is the phrase "BIOMETRIC INTELLIGENCE" in a smaller, sans-serif font. The logo appears to be for a company that specializes in biometric intelligence.
DATE: 2019-11-27
TITLE: Section 5: 510(k) Summary
510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of the Safe Medical Devices Act 1990 and 21 CFR Section 807.92.
General Information 1.
Establishment: KinetiCor Inc. 9965 Businesspark Ave., Suite B San Diego, CA 92131 Phone: (858) 800-1025 E-mail: www.kineticor.com Registration Number: TBD
- Date Prepared: November 27, 2019; Minor update January 30, 2020
Manufacturer: KinetiCor, Inc. 3454 Waialae Avenue, Suite 300A Honolulu, HI 96816 Phone: (808) 800-1025
2. Contact Information
Scott R. Herr Sr. Director - RA/QA 9965 Businesspark Ave., Suite B San Diego, CA 92131 Phone: (858) 800-1025 E-mail: scott.herr@kineticor.com
3. Device Name and Classification
| Device Name: | Motion Correction System |
|---|---|
| Trade Name: | Motion Correction System |
| Classification Name: | Magnetic Resonance Diagnostic Device (MRDD) |
| Classification Panel: | Radiology |
| CFR Code: | 21 CFR Sec. 892.1000 |
| Classification: | Class II |
| Product Code: | Primary: LNH |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the logo for "Kineticor Biometric Intelligence". The word "Kineticor" is in a large, blue, sans-serif font. Below "Kineticor" is the phrase "Biometric Intelligence" in a smaller, sans-serif font, also in blue. The overall design is clean and professional.
DATE: 2019-11-27
TITLE: Section 5: 510(k) Summary
4. Legally Marketed Predicate Device (21 CFR §807.92(a)(3))
| Trade Name: | MAGNETOM Vida |
|---|---|
| 510(k) Number: | K183254 cleared January 18, 2019 |
| Classification Name: | Magnetic Resonance Diagnostic Device (MRDD) |
| Classification Panel: | Radiology |
| CFR Code: | 21 CFR Sec. 892.1000 |
| Classification: | Class II |
| Product Code: | Primary: LNH, Secondary: LNI, MOS |
5. INDICATIONS FOR USE (21 CFR §807.92(a)(5))
The Motion Correction System is an accessory to a Magnetic Resonance Imaging (MRI) scanner. The system is a sensor-based technology for tracking of patient movement during an MRI session. It provides the current position of the patient in real-time to the MRI scanner for further data processing.
User: Professional Use
Environment: Hospital, doctor's office, or any facility that uses an MRI device
KinetiCor defines the Motion Correction System as an accessory to a medical device. It is intended to support supplement and/or augment the performance of the MAGNETOM Skyra 3T MRI scanner.
6. Device Description (21 CFR §807.92(a)(4))
During an MRI scan, if the patient's head does not remain still during the scan, the resulting images may not be sharp. This could prevent a physician from analyzing the images, therefore requiring additional scanning.
Software has been developed to somewhat reduce these effects.
KinetiCor has developed a Motion Correction technology using a track the true position of the patient's head. These coordinates are provided to the MRI scanner, which adjusts the MR images.
The Motion Correction System delivers head pose tracking and correction in 6 degrees of freedom for MR applications.
- . Optimized for high field MR environments, the technology has been FDA Cleared for use in the MAGNETOM Vida, 510(k) K183254, cleared January 18, 2019. Customized optics, electronics and
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the logo for Kineticor Biometric Intelligence. The word "KINETICOR" is written in a blue sans-serif font, with the "COR" portion of the word being a lighter shade of blue. Below the word "KINETICOR" is the phrase "BIOMETRIC INTELLIGENCE" in a smaller, sans-serif font, also in blue.
TITLE: Section 5: 510(k) Summary
enclosures minimize Electromagnetic Interference (EMI) and Radiofrequency (RF) interference and isolate the camera system for stable tracking accuracy.
- . Proprietary optical machine vision tracking technology provides tracking data to the MRI scanner with a static accuracy of approximately 0.1 mm and 0.1 deg. This means that the MR pulse sequences are updated in real-time.
- The impact of the Motion Correction technology can be illustrated by taking a closer look at the corrected and uncorrected images in the figure below.
Image /page/5/Picture/7 description: The image contains two sagittal MRI scans of a human brain. The left scan appears to be a T1-weighted image, showing good anatomical detail with clear differentiation between gray and white matter. The right scan seems to be a T2-weighted or FLAIR image, where fluid appears bright, and there is less contrast between gray and white matter. Both scans include a scale indicating a 10cm reference.
(a)
(b)
-
(a) Motion correction using KinetiCor technology and (b) MR images obtained when the motion was not corrected and acquired using the conventional way. Similar motion performed during both scans.
KinetiCor is offering this accessory to the following MRI scanners that are installed in the field: -
. Siemens MAGNETOM Skyra 3T (510(k) K153343 cleared April 15, 2016)
Hardware
The Motion Correction System Hardware consists of the following:
- Camera Enclosure (incorporating four Cameras) located inside the distal end of the MRI scanner; ●
- Wiring Conduit - provides a protected location for wiring to exit the distal end of the MRI scanner;
- Processing Unit - electronics with software, located outside the MRI room;
- External Wiring - used to connect processing unit to Camera Enclosure;
- . Tracking Markers - single use markers located on patient nose bridge (Referred to as "Nose Marker for Inline Motion Correction" in predicate device);
- External Monitor - if requested, located inside MRI Control Room;
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image contains the logo for Kineticor. The logo is in blue and consists of the word "KINETICOR" in large, bold letters. Below the word "KINETICOR" is the phrase "BIOMETRIC INTELLIGENCE" in smaller letters. The logo is simple and modern.
TITLE: Section 5: 510(k) Summary
- Network System located outside the MRI room.
7. RISK MANAGEMENT, GENERAL SAFETY AND EFFECTIVENESS
The device labeling contains instructions for use and any necessary cautions and warnings to provide for safe and effective use of the device. The Motion Correction System has an own Instruction for Use.
Risk management is ensured via a risk analysis, which is used to identify potential hazards and mitigations. These potential hazards are controlled by design means, protection measures and user instructions. To confirm that the measures are effective and that the product meets its intended uses, verification of requirements and standards, and validation of the clinical workflow was performed. KinetiCor adheres to recognized and established industry practice and relevant international standards where indicated.
8. Technological Characteristics and substantial equivalence (21 CFR §807.92(a)) (6))
The KinetiCor Motion Correction System is substantially equivalent to the predicate device as the Camera Enclosure, Processing Unit electronics, Motion Correction Software and Tracking Markers are the same as in the predicate device. The camera enclosure, electronics and software for the MAGNETOM Vida are installed at the Siemens Factory on a new device prior to first customer use, whereas the KinetiCor Motion Correction System accessory is installed on the MAGNETOM Skyra 3T System, which is already in the field and in use.
While there are some differences in characteristics between the subject device and predicate device, including off-the shelf software applications (XPACE), wiring conduit configuration, external wiring, and external monitor; these differences have been tested and the conclusions from the non-clinical data suggests that the features bear an equivalent safety and performance profile to that of the predicate device.
9. Nonclinical Tests
The following performance testing was conducted on the subject accessory to a medical device:
- -Software Verification and Validation testing was performed in accordance with FDA guidance document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" dated May 11, 2005 and KinetiCor Quality System Procedure QSP-106 Quality Lifecycle Procedure.
- -Design Verification and Validation testing was performed in accordance with 21 CFR 820 and KinetiCor Quality System Procedure QSP-102 Design Control.
10. Substantial Equivalence
There are very few differences between the KinetiCor Motion Correction System installed on a Skyra MRI and the predicate – almost all are due to the installing the KinetiCor accessory as a retrofit to existing product in the field versus assembly of the KinetiCor accessory at the factory in a brand new MAGNETOM Vida scanner.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Kineticor Biometric Intelligence. The word "KINETICOR" is written in a blue sans-serif font, with the "COR" portion of the word in a lighter shade of blue. Below the word "KINETICOR" is the phrase "BIOMETRIC INTELLIGENCE" in a smaller, sans-serif font.
DATE: 2019-11-27
TITLE: Section 5: 510(k) Summary
The KinetiCor Motion Correction System as an accessory to a medical device is substantially equivalent to the MAGNETOM VIDA with software syngo MR XA11B and KinetiCor's Motion Correction accessories, which was 510(k) cleared on January 18, 2019 (K183254). As stated in that submission, "TFL with inline Motion Correction: Tracking of motion of the head during 3D MPRAGE head scans with a nose marker and a camera system. The MR system uses the tracking information to compensate for the detected motion." In fact, the designs of the nose marker, electronics and the camera enclosure system are the same as in the MAGNETOM Vida and the KinetiCor Motion Correction System.
11. Conclusions (21 CFR §807.92(b)(3))
The KinetiCor Motion Correction System as an accessory to a medical device provides the same motion correction hardware and technology as the predicate device system. The KinetiCor Motion Correction System provides motion correction features to potentially improve the magnetic resonance image if patient movement occurs during one specific sequence. It is a currently a feature in the predicate new device and a proposed accessory to an existing device.
KinetiCor believes that the Motion Correction System as an accessory to a medical device is substantially equivalent to the currently marketed device MAGNETOM VIDA with software syngo MR XA11B and KinetiCor's Motion Correction accessories.
The subject device does not raise any new questions with respect to safety or effectiveness. The conclusions drawn from the nonclinical and clinical tests demonstrate that the device is safe, as effective, and performs as well as or better than the legally marketed device identified in paragraph (a)(3) of this section.
§ 892.1000 Magnetic resonance diagnostic device.
(a)
Identification. A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).(b)
Classification. Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.