K Number
K193321
Manufacturer
Date Cleared
2020-07-30

(241 days)

Product Code
Regulation Number
N/A
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The MY01 Continuous Compartmental Pressure Monitor is intended for real-time and continuous measurement of compartmental pressures. The measured compartmental pressures can be used as an aid in the diagnosis of compartment syndrome.

The trend arrows provide a general pressure direction and are for informational purposes only. Do not rely upon the trend arrows for the diagnosis of compartment syndrome.

Device Description

The MY01 Continuous Compartmental Pressure Monitor (MY01 device) is a device for measuring real-time and continuous compartmental pressured compartmental pressures should always be used along with the current standard of care as an aid in the diagnosis of compartment syndrome.

The device is supplied sterile for single patient use and can be used up to a maximum of 18 hours, at which point a warning message is displayed cautioning the user. It contains two major components that are referred to as the Introducer (plastic housing and 17gauge stainless-steel needle) and the Pressure Monitor. The Pressure Monitor consists of a capacitive Micro-Electro-Mechanical System (MEMS) pressure sensor, which allows for the measurement of intracompartmental pressure relative to a secondary atmospheric pressure sensor. It is embedded on a flexible PCB circuit, which extends via a lead-wire to a rigid PCB circuit within the Pressure Monitor. The Introducer allows for placement of the pressure sensor into muscle compartments. The Pressure Monitor continuously outputs pressure values on the LCD screen and can be attached on the patient's skin using the provided adhesive strip on the underside.

AI/ML Overview

The provided text describes the MY01 Continuous Compartmental Pressure Monitor. However, it does not contain detailed acceptance criteria or a study proving the device meets specific performance metrics in numerical terms. The text focuses on the device's substantial equivalence to a predicate device, regulatory compliance, and general performance testing (e.g., sterility, biocompatibility, electrical safety, software validation, human factors).

Therefore, I cannot populate the table with specific acceptance criteria and reported device performance, nor can I provide details on sample sizes, ground truth establishment for test/training sets, or MRMC studies, as this information is not present in the provided document.

Here's a summary of what can be extracted and inferred from the text, noting the limitations:

1. Table of Acceptance Criteria and Reported Device Performance:

Acceptance Criteria (Inferred from general performance testing)Reported Device Performance (Inferred from text)
Device sterility and packaging integrity validated to FDA recognized industry standards.MY01 device successfully met the criteria for maintenance of sterility and packaging integrity.
Biological safety evaluated in accordance with ISO 10993-1.Biological safety of final finished, sterile device was evaluated and found compliant.
Electrical safety and performance assessed in compliance with IEC 60601-1 (Ed. 3.1).Compliant with IEC 60601-1 (Ed. 3.1).
EMC assessed in compliance with IEC 60601-1-2 (Ed. 4).Compliant with IEC 60601-1-2 (Ed. 4) for a Group 1 Class B device.
Software verification and validation (moderate level of concern).Software verification and validation testing were conducted, and documentation provided.
Hardware and software device specifications verified for conformity to design input requirements and validated under simulated use conditions.MY01 device successfully met the criteria for safety of materials and ease of use in a simulated use.
Performance testing ensures equivalent level of safety and effectiveness as the predicate device.Extensive performance testing for the subject device assures that the MY01 device is at least as safe and effective as the previously cleared predicate device.

Missing Information: Numerical or specific performance metrics (e.g., accuracy, precision, bias of pressure measurements, sensitivity, specificity, or error rates for diagnosing compartment syndrome).

2. Sample size used for the test set and the data provenance:

  • Sample Size for Test Set: Not specified. The document mentions "simulated use conditions" for hardware/software validation and human factors, but no specific number of cases or patients for a test set validating the diagnostic aid indication.
  • Data Provenance: Not applicable, as no clinical study or specific dataset for diagnostic performance is detailed. The performance data focuses on device-specific engineering and safety tests.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not applicable. The document does not describe a clinical study with a test set requiring expert-established ground truth for diagnostic performance.

4. Adjudication method for the test set:

  • Not applicable. No clinical study with a test set requiring adjudication is described.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • No. The MY01 device is a Continuous Compartmental Pressure Monitor, a sensor-based medical device that aids in diagnosis, but it is not an AI/CAD (Computer-Aided Detection/Diagnosis) system that directly interprets medical images or data requiring human-in-the-loop performance measurement. Therefore, an MRMC study comparing human readers with and without AI assistance is not relevant to this device as described.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • Not applicable in the context of an AI/CAD system. The device itself performs continuous pressure measurements. Its "standalone" performance is assessed through its accuracy and reliability of pressure sensing, which are implied to be part of the "hardware and software device specifications... verified for conformity to design input requirements" and "performance testing." However, specific metrics for this standalone performance (e.g., accuracy against a gold standard pressure transducer) are not provided.

7. The type of ground truth used:

  • For engineering and safety tests (sterility, electrical safety, EMC, biocompatibility), the "ground truth" is defined by established international standards (e.g., ISO 14971, ISO 10993-1, IEC 60601-1, IEC 60601-1-2).
  • For hardware/software validation and human factors, the ground truth relates to "design input requirements" and "established performance criteria," implying internal specifications and usability guidelines.
  • The device's intended use is as an "aid in the diagnosis of compartment syndrome," but the document does not detail a clinical study where its diagnostic accuracy was directly measured against a clinical ground truth (e.g., surgical confirmation of compartment syndrome or expert clinical diagnosis).

8. The sample size for the training set:

  • Not applicable. The document describes a medical device, not a machine learning model that requires a training set in the conventional sense. Software verification and validation were performed, but this refers to traditional software engineering testing rather than AI model training.

9. How the ground truth for the training set was established:

  • Not applicable, as there is no "training set" for an AI model.

{0}------------------------------------------------

July 30, 2020

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo has the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" and "ADMINISTRATION" in blue text.

MY01, Inc. Charles Allan Chief Executive Officer 85 Rue Saint-Paul Ouest, Suite 200 Montreal, H2Y 3V4 Ca

Re: K193321

Trade/Device Name: MY01 Continuous Compartmental Pressure Monitor Regulatory Class: Unclassified Product Code: LXC Dated: July 29, 2020 Received: July 30, 2020

Dear Charles Allan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS)

{1}------------------------------------------------

regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Vesa Vuniqi Assistant Director DHT6A: Division of Joints Arthroplasty OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K193321

Device Name

MY01 Continuous Compartmental Pressure Monitor

Indications for Use (Describe)

The MY01 Continuous Compartmental Pressure Monitor is intended for real-time and continuous measurement of compartmental pressures. The measured compartmental pressures can be used as an aid in the diagnosis of compartment syndrome.

The trend arrows provide a general pressure direction and are for informational purposes only. Do not rely upon the trend arrows for the diagnosis of compartment syndrome.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CER 801 Subpart D)
Over-The-Counter Use (21 CER 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows a logo with the text "MY01" to the right of an abstract graphic. The graphic consists of an orange circle overlapping a light orange, elongated shape. The text is in a simple, sans-serif font and is black.

K193321

510(k) Summary for the MY01 Continuous Compartmental Pressure Monitor

In accordance with the requirements of 21 CFR 807.92 of the Federal Code of Regulations, the following information is a summary of safety and effectiveness for the MY01 Continuous Compartmental Pressure Monitor.

A. SUBMITTERS INFORMATION

Submitted by:MY01, Inc.85 Rue Saint-Paul Ouest, Suite 200,Montréal, Québec, H2Y 3V4, CANADA
Contact Person:Anthony SirgiAddress: 85 Rue Saint-Paul Ouest, Suite 200,Montréal, Québec, H2Y 3V4, CANADARegulatory Affairs and Compliance OfficerEmail: anthony.sirgi@my01.ioTel: (514) 963-6027

Date Prepared: November 27, 2019

B. DEVICE INFORMATION

Registration Number:TBD
Device Trade Name:MY01 Continuous Compartmental Pressure Monitor
Device Common Name:Monitor, Pressure, Intracompartmental
Classification Name:Unclassified
Classification Code:LXC
Classification Panel:Orthopedic
Regulation Number:Pre-Amendment, Unclassified

C. PREDICATE DEVICE

K131966Twin Star Extremity Compartment Monitor and Fluid Collection
System (ECM-III)

D. DEVICE DESCRIPTION

The MY01 Continuous Compartmental Pressure Monitor (MY01 device) is a device for measuring real-time and continuous compartmental pressured compartmental pressures should always be used along with the current standard of care as an aid in the diagnosis of compartment syndrome.

The device is supplied sterile for single patient use and can be used up to a maximum of 18 hours, at which point a warning message is displayed cautioning the user. It contains two major components that are referred to as the Introducer (plastic housing and 17gauge stainless-steel needle)

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows a logo with the text "MY01" to the right of a graphic. The graphic is an abstract shape in a light orange color, with a darker orange circle overlapping it. The text "MY01" is in a simple, sans-serif font and is black.

and the Pressure Monitor. The Pressure Monitor consists of a capacitive Micro-Electro-Mechanical System (MEMS) pressure sensor, which allows for the measurement of intracompartmental pressure relative to a secondary atmospheric pressure sensor. It is embedded on a flexible PCB circuit, which extends via a lead-wire to a rigid PCB circuit within the Pressure Monitor. The Introducer allows for placement of the pressure sensor into muscle compartments. The Pressure Monitor continuously outputs pressure values on the LCD screen and can be attached on the patient's skin using the provided adhesive strip on the underside.

E. INTENDED USE/INDICATIONS FOR USE

The MY01 Continuous Compartmental Pressure Monitor is intended for real-time and continuous measurement of compartmental pressures. The measured compartmental pressures can be used along with the current standard of care as an aid in the diagnosis of compartment syndrome. The trend arrows provide a general pressure direction and are for informational purposes only. Do not rely upon the trend arrows for the diagnosis of compartment syndrome.

This intended use for the MY01 subject device is identical to the predicate device, with the exception of the latter's ability for fluid withdrawal for analysis. Since, this is an additional indication, it does not impact the intended use for continuous measurement of compartment pressures and thus, does not affect the safety and effectiveness of the subject device.

F. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS

The MY01 device has the similar intended use and equivalent indications for use as the predicate Twin Star device.

The devices have the following technological similarities:

  • Use of continuously inserted measuring pressure sensor; ●
  • . Use of LCD display monitor to display pressure values in mmHg;
  • Sterile and single use introducer mechanism;
  • . Use of stainless-steel needle or trocar as introducers;
  • Similar insertion sites;
  • Use of software to provide pressure readings;
  • . Use of external or built-in battery for power source.

The following technological differences exist between the subject and predicate device:

  • . The predicate device uses fiber optics to sense a change in pressure while the MY01 device measures a change in sensor capacitance, however both devices use established effective methods incorporating membranes to measure relative pressure;

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows a logo with the text "MY01" to the right of an abstract graphic. The graphic consists of a light orange, elongated shape with a darker orange circle overlapping it. The text is in a simple, sans-serif font and is black in color.

  • . The predicate device uses a porous catheter to host the pressure sensor while MY01 device uses a flexible printed circuit to host the pressure sensor
  • . The predicate device includes a reusable monitor, a sterile catheter for single-patient and a vacuum pump that may be connected to a fluid collection catheter, while the MY01 device supplied sterile for single-patient use and does not provide a means for fluid collection

The MY01 device is substantially equivalent to the Twin Star predicate device based on intended use and indications for use, method of use for measuring intracompartmental pressure, and compliance to FDA recognized safety and performance consensus standars. The differences in technological characteristics are not significant and do not raise additional questions of safety and effectiveness as it can be demonstrated by a review of predicate device characteristics, design and development standards, and non-clinical performance testing. The differences in materials and manufacturing processes have been successfully assessed for safety and effectiveness using the same industry testing methods for the subject device.

G. PERFORMANCE DATA

Systematic risk analysis was conducted in accordance with ISO 14971 and risks were mitigated as far as possible considering device sterility at the point of care, the biological safety of device materials, electrical safety and electromagnetic compatibility (EMC), hardware and software component specifications, and human factors for the intended use:

  • . Device sterility, shelf life, and packaging integrity to the point of use were validated according to FDA recognized industry standards and guidance documents.
  • The biological safety of final finished, sterile device was evaluated in accordance with . FDA guidance on the Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" (2016), and the current FDA recognized standard ISO 10993-1. for externally communicating devices with limited contact duration (<24 h).
  • Electrical safety and performance requirements were assessed in compliance with the . current FDA recognized standard IEC 60601-1 (Ed. 3.1). Both the subject device and Twin Star predicate device fall under similar degree of protection and mode of operation.
  • EMC was assessed in compliance with the FDA recognized standard IEC 60601-1-2 . (Ed. 4). Both the predicate and subject device were found to comply with the limits for a Group 1 Class B device according to this standard.
  • Software verification and validation testing were conducted, and documentation was provided as recommended by FDA's Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices (2005) for a 'moderate' level of concern.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows a logo with an abstract design and the text "MY01". To the left of the text, there is an abstract shape in a light peach color, with an orange circle overlapping it. The text "MY01" is in a simple, sans-serif font and is black.

  • Hardware and software device specifications were verified for conformity to design . input requirements and were validated under simulated use conditions in accordance with FDA guidance on Applying Human Factors and Usability Engineering to Medical Devices (2016).
    The MY01 device successfully met the criteria for safety of materials, maintenance of sterility and packaging integrity and ease of use of the device in a simulated use. Additionally, the results of successful verification and validation testing in compliance with current FDA recognized consensus standards and guidance provide a high level of assurance that the MY01 device fulfills design input requirements and established performance criteria and demonstrate the effectiveness of risk mitigations for the intended use. Thus, the performance testing for the subject device, using acceptable methods, ensures equivalent level of safety and effectiveness as the predicate device.

H. CONCLUSION

Both the predicate device and the MY01 Continuous Compartmental Pressure Monitor are intended for continuous measurement of intracompartmental pressures, the values of which can be used as an aid in diagnosis of compartment syndrome. There are only minor technological differences between the two devices. However, the similarities in function, purpose, display method of pressure values and the extensive performance testing for the subject device assure that the MY01 device is at least as safe and effective as the previously cleared predicate device. Thus, we believe the subject device (MY01 device) is substantially equivalent to the predicate (Twin Star) device.

N/A