(416 days)
uAl EasyTriage-Rib is a radiological computer-assisted triage and notification software device for analysis of CT chest images. The device is intended to assist hospital networks and trained radiologists in workflow triage by flagging and prioritizing trauma studies with suspected positive findings of multiple (3 or more) acute rib fractures.
uAl EasyTriage-Rib is a radiological computer-assisted triage and notification software device indicated for analysis of CT chest images. The device is intended to assist hospital networks and trained radiologists in workflow triage by flagging and prioritizing studies with suspected positive findings of multiple (3 or more) acute rib fractures. The device consists of the following two modules: (1) uAl EasyTriage-Rib Server; and (2) uAl EasyTriage-Rib Studylist Application that provides the user interface in which notifications from the application are received.
The information provided describes the uAI EasyTriage-Rib device and its performance study to meet acceptance criteria for identifying multiple (3 or more) acute rib fractures in CT chest images.
Here's a breakdown of the requested information:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are implied by the reported performance metrics. The specific "acceptance criteria" are not explicitly stated as numerical targets in the provided text. However, a general statement is made: "The results show that it can detect rib fractures and reach the preset standard." Given the context of a 510(k) summary, the reported sensitivity, specificity, and AUC, along with a comparable time-to-notification to the predicate device, are the performance benchmarks that demonstrate achievement of that standard.
| Performance Metric | Acceptance Criteria (Implied) | Reported Device Performance | Comments |
|---|---|---|---|
| Sensitivity | High | 92.7% (95% CI: 84.8%-97.3%) | Achieved high sensitivity as a crucial consideration for a time-critical condition. |
| Specificity | Adequate | 84.7% (95% CI: 77.0%-90.7%) | Specificity was noted to be affected by the difficulty in distinguishing acute from chronic fractures, but considered acceptable given clinical relevance of reviewing chronic fractures. |
| AUC | High | 0.939 (95% CI: 0.906, 0.972) | Indicates high discriminative power. |
| Time-to-notification (Average) | Comparable to predicate device | 69.56 seconds | Comparable to predicate device (HealthVCF: 61.36 seconds), suggesting timely notifications. |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size for Test Set: 200 cases
- Data Provenance:
- Country of Origin: Multiple US clinical sites (explicitly stated).
- Retrospective or Prospective: Retrospective (explicitly stated).
3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts
This information is not explicitly stated in the provided text. The document mentions "trained radiologists" being involved in clinical decision-making but does not specify the number or qualifications of experts used to establish the ground truth for the test set.
4. Adjudication Method for the Test Set
This information is not explicitly stated in the provided text.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance
A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done with human readers. The study focused on the standalone performance of the AI algorithm and a comparison of its "time-to-notification" with a predicate device, not on how human readers improve with AI assistance.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
Yes, a standalone study was done. The reported sensitivity, specificity, and AUC are measures of the algorithm's performance in identifying the target condition without human intervention in the analysis. The device "uses an artificial intelligence algorithm to analyze images and highlight studies with suspected multiple (3 or more) acute rib fractures in a standalone application for study list prioritization or triage in parallel to ongoing standard of care."
7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)
The document does not explicitly state the method used to establish the ground truth for the 200 test cases. It is often implied to be expert consensus by radiologists in such studies, but this is not confirmed in the text.
8. The Sample Size for the Training Set
The sample size for the training set is not provided in the text. The document only mentions that the deep learning algorithm was "trained on medical images."
9. How the Ground Truth for the Training Set Was Established
This information is not provided in the text.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. Food & Drug Administration".
January 15, 2021
Shanghai United Imaging Intelligence Co., Ltd. % Zhao Xiaojing Quality & Regulatory Manager No. 199, Huanke Road Shanghai, Shanghai 201210 CHINA
Re: K193271
Trade/Device Name: uAI EasyTriage-Rib Regulation Number: 21 CFR 892.2080 Regulation Name: Radiological computer aided triage and notification software Regulatory Class: Class II Product Code: QFM Dated: December 8, 2020 Received: December 8, 2020
Dear Zhao Xiaojing:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR
{1}------------------------------------------------
- for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
510(k) Number (if known) K193271 Device Name
uAl EasyTriage-Rib
Indications for Use (Describe)
uAl EasyTriage-Rib is a radiological computer-assisted triage and notification software device for analysis of CT chest images. The device is intended to assist hospital networks and trained radiologists in workflow triage by flagging and prioritizing trauma studies with suspected positive findings of multiple (3 or more) acute rib fracture(s).
uAl EasyTriage-Rib uses an artificial intelligence algorithm to analyze images and highlight studies with suspected multiple (3 or more) acute rib fracture(s) in a standalone application for study list prioritization or triage in parallel to ongoing standard of care. The user is presented with notifications of cases with suspected findings. Notifications include compressed preview images that are meant for informational purposes only and not intended for diagnostic use beyond notification. The device does not alter the original medical image and is not intended to be used as a diaqnostic device.
The results of uAl EasyTriage-Rib, in conjunction with other clinical information and professional judgment, are to be used to assist with triage/prioritization of medical images. Notified radiologists who read the original medical images are responsible for the diagnostic decision.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number"
{3}------------------------------------------------
{4}------------------------------------------------
510(k) SUMMARY
K193271
Shanghai United Imaging Intelligence Co., Ltd.'s uAl EasyTriage-Rib
Submitter:
Shanghai United Imaging Intelligence Co., Ltd. Floor 23-26, No.701 Yunjin Road, Xuhui District, Shanghai Phone: +86 13917486296 Contact Person: ZHAO Xiaojing
Date Prepared: January 14, 2021
Name of Device: uAI EasyTriage-Rib
Common or Usual Name/ Classification Name: Radiological Computer-Assisted Prioritization Software For Lesions
Regulatory Class: Class II
Product Code: QFM (21 C.F.R. 892.2080)
Predicate Device: Zebra Medical Vision Ltd.'s HealthVCF (K192901)
Device Description
uAl EasyTriage-Rib is a radiological computer-assisted triage and notification software device indicated for analysis of CT chest images. The device is intended to assist hospital networks and trained radiologists in workflow triage by flagging and prioritizing studies with suspected positive findings of multiple (3 or more) acute rib fractures. The device consists of the following two modules: (1) uAl EasyTriage-Rib Server; and (2) uAl EasyTriage-Rib Studylist Application that provides the user interface in which notifications from the application are received.
Intended Use / Indications for Use
uAl EasyTriage-Rib is a radiological computer-assisted triage and notification software device for analysis of CT chest images. The device is intended to assist hospital networks and trained radiologists in workflow triage by flagging and prioritizing trauma studies with suspected positive findings of multiple (3 or more) acute rib fractures.
uAl EasyTriage-Rib uses an artificial intelligence algorithm to analyze images and highlight studies with suspected multiple (3 or more) acute rib fractures in a standalone application for study list prioritization or triage in parallel to ongoing standard of care. The user is presented with notifications of cases with suspected findings. Notifications include compressed preview images that are meant for informational purposes only and not intended for diagnostic use beyond notification. The device does not alter the original medical image and is not intended to be used as a diagnostic device.
{5}------------------------------------------------
The results of uAl EasyTriage-Rib, in conjunction with other clinical information and professional judgment, are to be used to assist with triage/prioritization of medical images. Notified radiologists who read the original medical images are responsible for the diagnostic decision.
Justification for Time Criticality of Indication
The device aims to triage multiple (3 or more) rib fractures since the condition of multiple rib fractures is time sensitive in clinical practice. Specifically,
- The presence of 3 or more rib fractures is highly predictive of poor clinical outcomes including respiratory failure and overall mortality [1-10].
- . The presence of 3 or more rib fractures is incorporated into US clinical quidelines for trauma patient management [1, 11, 15-24].
- Flail chest, which occurs in a subset of patients with 3 or more rib fractures, is a ● potentially life threatening condition that requires prompt management [1, 14,15, 25-281.
Accordingly, rib fracture is a time-critical condition that is appropriate to prioritize for review. In this setting, high sensitivity is a crucial consideration so that all appropriate cases may be identified and promptly interpreted.
Comparison of Technological Characteristics with the Predicate Device
HealthVCF (K192901) is the predicate device. The subject and predicate device are both radiological computer-assisted triage and notification software. Both devices are artificial intelligence algorithms incorporated software packages that analyze CT images for fracture(s). Both devices process images intended to aid in prioritization and triage of radiological medical images. The subject device is intended to provide notifications for cases with suspected positive findings of multiple (3 or more) acute rib fractures by analysis of CT chest images and the predicate device is intended to analyze chest and abdominal CT scans and flags those that are suggestive of the presence of at least one vertebral compression at the exam level. This difference does not affect the intended use of both devices, which is to prioritize timesensitive fractures for trained clinician review.
Both software devices provide passive notifications to a clinician of the availability of time sensitive radiological medical images for review based on computer aided image analysis performed by the device's AI algorithm. The subject device flags cases with the suspected positive findings on the Studylist Application on the workstations of the radiologist. Those notifications work in parallel to the standard of care. They prompt the radiologist to start preemptive triage of a flagged case, upon which he may turn to the local PACS to perform the review. In addition, both devices show preview images for positive findings.
{6}------------------------------------------------
The predicate and subject devices process CT images using similar techniques and a similar artificial intelligence algorithm. Specifically, the subject and predicate software utilize a deep learning algorithm trained on medical images. The deep-learning process allows for high accuracy in the detection of initial suspect positive findings. As a system, the uAl EasyTriage-Rib raises the same types of safety and effectiveness questions as the predicate; namely, accurate detection of findings within the reviewed and processed study on which a physician can base a clinically useful triage/prioritization assessment considering all available clinical information.
lt is important to note that, like the predicate, the device does not remove cases from a reading queue. Again, both devices operate in parallel with the standard of care, which remains the default option for all cases.
| TechnologicalCharacteristics | Subject DeviceuAl EasyTriage-Rib(K193271) | Predicate DeviceHealthVCF(K192901) | Summary |
|---|---|---|---|
| Indication forUse/IntendedUse | uAl EasyTriage-Rib is aradiological computer-assisted triage andnotification softwaredevice for analysis of CTchest images. The deviceis intended to assisthospital networks andtrained radiologists inworkflow triage byflagging and prioritizingtrauma studies withsuspected positivefindings of multiple (3 ormore) acute rib fractures.uAl EasyTriage-Rib usesan artificial intelligencealgorithm to analyzeimages and highlightstudies with suspectedmultiple (3 or more) acuterib fractures in astandalone application forstudy list prioritization ortriage in parallel toongoing standard of care.The user is presentedwith notifications of caseswith suspected findings | HealthVCF is a passivenotification forprioritization-only,parallel-workflowsoftware tool used byclinicians to prioritizespecific patients withinthe standard-of-carebone health setting forsuspected vertebralcompression fractures.HealthVCF uses anartificial intelligencealgorithm to analyzechest and abdominalCT scans and flagsthose that aresuggestive of thepresence of at leastone vertebralcompression at theexam level. Theseflags are viewed by theclinician in Bone Healthand Fracture LiaisonService programs inthe medical setting viaa worklist applicationon their Picture | Similar exceptfor lesion type.Both findingsareappropriatelytime sensitive.Performancedata willsupport uAlEasyTriage-Ribindications. |
| TechnologicalCharacteristics | Subject DeviceuAl EasyTriage-Rib(K193271) | Predicate DeviceHealthVCF(K192901) | Summary |
| Notifications includecompressed previewimages that are meant forinformational purposesonly and not intended fordiagnostic use beyondnotification. The devicedoes not alter the originalmedical image and is notintended to be used as adiagnostic device.The results of uAlEasyTriage-Rib, inconjunction with otherclinical information andprofessional judgment,are to be used to assistwith triage/prioritization ofmedical images. Notifiedradiologists who read theoriginal medical imagesare responsible for thediagnostic decision. | Archiving andCommunicationSystem (PACS).HealthVCF does notsend a proactive alertdirectly to the user.HealthVCF does notprovide diagnosticinformation beyondtriage and prioritization,it does not removecases from theradiology worklist, andshould not be used inplace of full patientevaluation, or reliedupon to make orconfirm diagnosis. | ||
| Notification-only,parallel workflowtool | Yes | Yes | Same, bothdevicesproducepassivenotifications |
| User | Radiologist | Bone Health Clinician | Radiologists arecommon usersfor productsunder productcode QFM forRadiologicalComputer-AssistedPrioritizationSoftware ForLesions |
| Identify patientswith pre-specified clinicalcondition | Yes | Yes | Same |
| Clinical condition | Multiple (3 or more) acuterib fractures | Vertebral compressionfracture | Different butboth findings |
| TechnologicalCharacteristics | Subject DeviceuAl EasyTriage-Rib(K193271) | Predicate DeviceHealthVCF(K192901) | Summary |
| indicate pre-specifiedclinicalconditions fortriage | |||
| Alert to finding | Yes; notification flaggedfor review | Yes; notificationflagged for review | Same |
| Independent ofstandard of careworkflow | Yes; No cases areremoved from worklist | Yes; No cases areremoved from worklist | Same |
| Modality | CT | CT | Same |
| Body part | Chest | Chest and abdomen | Similar, bothinclude "chest". |
| ArtificialIntelligencealgorithm | Yes | Yes | Same |
| Limited toanalysis ofimaging data | Yes | Yes | Same |
| Aids promptidentification ofcases withindicatedfindings | Yes | Yes | Same |
| Where resultsare received | Workstation | PACS / Workstation | Different butboth provide apassivenotification tothe workstationof the presenceof suspectedfinding in thescan |
A table comparing the key features of the subject and predicate devices is provided below.
{7}------------------------------------------------
{8}------------------------------------------------
Performance Data
UII conducted a retrospective, blinded, multicenter study with the uAl EasyTriage-Rib software with the primary endpoint to evaluate the software's performance in identifying CT chest images containing multiple (3 or more) acute rib fractures in 200 cases from multiple US clinical sites. The 200 cases had >1mm slice thickness and were from GE and Siemens scanners. The sensitivity was 92.7% (95% Cl: 84.8%-97.3%) and specificity was 84.7% (95% Cl: 77.0%-90.7%). The AUC was 0.939 (95% Cl: 0.906, 0.972).
An important consideration with these data is the presence of chronic rib fractures in the
{9}------------------------------------------------
dataset and the difficulty in distinguishing these findings from acute rib fractures, resulting in a decrease of specificity when acute fractures are the target condition. Specifically, certain chronic fractures can present as a pseudoarthrosis and/or malunion, findings that are difficult to distinguish from acute fractures. Accordingly, such findings are clinically relevant to review so as to exclude acute fracture.
Overall, the benefit-risk profile is favorable, and reflects the benefit of detecting 3 or more acute rib fractures with the high degree of sensitivity, and alerting the radiologist to the presence of this low incidence condition so that the study can be promptly interpreted.
In addition, a secondary endpoint measure was uAl EasyTriage-Rib's potential clinical benefit of worklist prioritization. For that purpose, we tested all the 76 true positive studies from clinical data set to compare the time-to-notification metric with Zebra Medical Vision Ltd.'s HealthVCF (K192901). The uAl EasyTriage-Rib time-to-notification is defined from the beginning of downloading the DICOM data from the PACS to the time of notification shown in the Studylist.
As shown in the table below, the average time-to-notification of uAl EasyTriage-Rib among 76 true positive studies 69.56 seconds is comparable to the time-to-notification of the HealthVCF software documented for an average of 61.36 seconds, suggesting that the radiologist can receive a notification timely on the status of studies with potential rib fracture findings with the help of uAI EasyTriage-Rib.
| Time-to-notification | Average performance time(seconds) |
|---|---|
| uAl EasyTriage-Rib | 69.56 |
| HealthVCF | 61.36 |
In summary, the performance on 200 cases establishes the achievement of effective triage by the uAl EasyTriage-Rib as well as effective notification functionality of the application, as compared to the time-to-notification of HealthVCF. The results show that it can detect rib fractures and reach the preset standard.
Conclusions
The uAI EasyTriage-Rib is as safe and effective as the HealthVCF. The uAI EasyTriage-Rib has the same intended uses and similar indications, technological characteristics, and principles of operation as its predicate device. The minor differences in indications do not alter the intended use of the device and do not affect its safety and effectiveness when used as labeled. In addition, the minor technological differences between the uAl EasyTriage-Rib and its predicate device raise no new issues of safety or effectiveness. Performance data demonstrate that the uAl EasyTriage-Rib is as safe and effective as the HealthVCF. Thus, the uAl EasyTriage-Rib is substantially equivalent.
{10}------------------------------------------------
REFERENCES:
[1] Karlson KA et al. Initial Evaluation and Management of Rib Fractures. In: UpToDate, Post TW (Ed), UpToDate, Waltham, MA.
[2] Brasel KJ, Guse CE, Layde P, Weigelt JA. Rib fractures: relationship with pneumonia and mortality. Crit Care Med. 2006;34(6):1642-1646. doi: 10.1097/ 01.CCM.0000217926. 40975.4B
[3] Bulger EM, Arneson MA, Mock CN, Jurkovich GJ. Rib fractures in the elderly. J Trauma. 2000;48(6):1040-1047. doi:10.1097/00005373-200006000-00007
[4] Flagel BT, Luchette FA, Reed RL, et al. Half-a-dozen ribs: the breakpoint for mortality. Surgery. 2005;138(4):717-725. doi:10.1016/j.surg.2005.07.022
[5] Jones KM, Reed RL 2nd, Luchette FA. The ribs or not the ribs: which influences mortality. Am J Surg. 2011;202(5):598-604. doi:10.1016/j.amjsurg.2010.09.029
[6] Battle CE, Hutchings H, James K, Evans PA. The risk factors for the development of complications during the recovery phase following blunt chest wall trauma: a retrospective study. Injury. 2013;44(9):1171-1176. doi:10.1016/j.injury.2012.05.019
[7] Holcomb JB, McMullin NR, Kozar RA, Lygas MH, Moore FA. Morbidity from rib fractures increases after age 45. J Am Coll Surg. 2003;196(4):549-555. doi:10.1016/S1072-7515(02)01894-X
[8] Stawicki SP, Grossman MD, Hoey BA, Miller DL, Reed JF 3rd. Rib fractures in the elderly: a marker of injury severity. J Am Geriatr Soc. 2004;52(5):805-808. doi:10.111/j.1532-5415.2004.52223.x
[9] Battle CE, Hutchings H, Evans PA. Risk factors that predict mortality in patients with blunt chest wall trauma: a systematic review and meta-analysis. Injury. 2012;43(1):8-17. doi:10.1016/j.injury.2011.01.004
(10) Shulzhenko NO, Zens TJ, Beems MV, et al. Number of rib fractures thresholds independently predict outcomes in older patients with blunt trauma. Surgery. 2017;161(4):1083 - 1089. worse doi:10.1016/j.surq.2016.10.018
[11] Sirmali M, Türüt H, Topcu S, et al. A comprehensive analysis of traumatic rib fractures: morbidity, mortality and management. Eur J Cardiothorac Surg. 2003;24(1):133-138. doi:10.1016/s1010-7940(03)00256-2
[12] Ziegler DW, Agarwal NN. The mortality of rib fractures. J Trauma. 1994;37(6):975-979. doi:10.1097/00005373-199412000-00018
[13] Oikonomou A, Prassopoulos P. CT imaging of blunt chest trauma. Insights into imaging. 2011 Jun 1;2(3):281-95.
[14] Talbot BS, Gange Jr CP, Chaturvedi A, Klionsky N, Hobbs SK, Chaturvedi A. Traumatic rib injury: patterns, imaging pitfalls, complications, and treatment. Radiographics. 2017 Mar;37(2):628-51
[15] Brasel KJ, Moore EE, Albrecht RA, et al. Western Trauma Association Critical Decisions in Trauma: Management of rib fractures. J J Trauma Acute Care Surg. 2017;82(1):200 - 203. doi:10.1097/TA.000000000000001301
[16] Lee RB, Bass SM, Morris JA Jr, MacKenzie EJ. Three or more rib fractures as an indicator for transfer to a Level I trauma center: a population-based study. J Trauma. 1990:30(6):689-694. doi:10.1097/00005373-199006000-00006
{11}------------------------------------------------
[17] Barnea Y, Kashtan H, Skornick Y, Werbin N. Isolated rib fractures in elderly patients: mortality and morbidity. Can J Surg. 2002:45(1):43-46.
[18] Sarani B et al. Inpatient Management of Traumatic Rib Fractures. In: UpToDate, Post TW (Ed), UpToDate, Waltham. MA.
[19] Pieracci FM, Rodil M, Stovall RT, Johnson JL, Biffl WL, Mauffrey C, Moore EE, Jurkovich GJ, Surgical stabilization of severe rib fractures. Journal of Trauma and Acute Care Surgery. 2015 Apr 1:78(4):883-7.
[20] Majercik S, Cannon Q, Granger SR, VanBoerum DH, White TW. Long-term patient outcomes after surgical stabilization of rib fractures. Am J Surq. 2014:208(1):88-92. doi:10.1016/i.amisurg.2013.08.051
[21] Bhatnagar A, Mayberry J, Nirula R. Rib fracture fixation for flail chest: what is the benefit?. J Am Coll Surg. 2012;215(2):201-205. doi:10.1016/j.jamcollsurg.2012.02.023
[22] Pieracci FM, Lin Y, Rodil M, et al. A prospective, controlled clinical evaluation of surgical stabilization of severe rib fractures. J Trauma Acute Care Surq. 2016:80(2):187-194. doi:10.1097/TA.00000000000025
[23] Kasotakis G. Hasenboehler EA. Streib EW. Patel N. Patel MB. Alarcon L. Bosarge PL. Love J. Love J. Haut ER. Como JJ. Operative fixation of rib fractures after blunt trauma: a practice management guideline from the Eastern Association for the Surgery of Trauma. Journal of Trauma and Acute Care Surgery. 2017 Mar 1;82(3):618-26.
[24] Nirula R, Diaz JJ Jr, Trunkey DD, Mayberry JC. Rib fracture repair: indications, technical issues, and future directions. World J Surq. 2009:33(1):14-22. doi:10.1007/s00268-008-9770-y
[25] Collins J. Chest wall trauma. J Thorac Imaging 2000;15:112-119.
[26] Davignon K, Kwo J, Bigatello LM. Pathophysiology and management of the flail chest. Minerva Anestesiol. 2004;70(4):193-199
[27] Dehghan N, de Mestral C, McKee MD, Schemitsch EH, Nathens A. Flail chest injuries: a review of outcomes and treatment practices from the National Trauma Data Bank. J Trauma Acute Care Surg. 2014;76(2):462-468. doi:10.1097/TA.000000000000000000000000
[28] Velmahos GC, Vassiliu P, Chan LS, Murray JA, Berne TV, Demetriades D. Influence of flail chest on outcome among patients with severe thoracic cage trauma. Int Surg. 2002;87(4):240-244.
§ 892.2080 Radiological computer aided triage and notification software.
(a)
Identification. Radiological computer aided triage and notification software is an image processing prescription device intended to aid in prioritization and triage of radiological medical images. The device notifies a designated list of clinicians of the availability of time sensitive radiological medical images for review based on computer aided image analysis of those images performed by the device. The device does not mark, highlight, or direct users' attention to a specific location in the original image. The device does not remove cases from a reading queue. The device operates in parallel with the standard of care, which remains the default option for all cases.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Design verification and validation must include:
(i) A detailed description of the notification and triage algorithms and all underlying image analysis algorithms including, but not limited to, a detailed description of the algorithm inputs and outputs, each major component or block, how the algorithm affects or relates to clinical practice or patient care, and any algorithm limitations.
(ii) A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide effective triage (
e.g., improved time to review of prioritized images for pre-specified clinicians).(iii) Results from performance testing that demonstrate that the device will provide effective triage. The performance assessment must be based on an appropriate measure to estimate the clinical effectiveness. The test dataset must contain sufficient numbers of cases from important cohorts (
e.g., subsets defined by clinically relevant confounders, effect modifiers, associated diseases, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals for these individual subsets can be characterized with the device for the intended use population and imaging equipment.(iv) Stand-alone performance testing protocols and results of the device.
(v) Appropriate software documentation (
e.g., device hazard analysis; software requirements specification document; software design specification document; traceability analysis; description of verification and validation activities including system level test protocol, pass/fail criteria, and results).(2) Labeling must include the following:
(i) A detailed description of the patient population for which the device is indicated for use;
(ii) A detailed description of the intended user and user training that addresses appropriate use protocols for the device;
(iii) Discussion of warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (
e.g., poor image quality for certain subpopulations), as applicable;(iv) A detailed description of compatible imaging hardware, imaging protocols, and requirements for input images;
(v) Device operating instructions; and
(vi) A detailed summary of the performance testing, including: test methods, dataset characteristics, triage effectiveness (
e.g., improved time to review of prioritized images for pre-specified clinicians), diagnostic accuracy of algorithms informing triage decision, and results with associated statistical uncertainty (e.g., confidence intervals), including a summary of subanalyses on case distributions stratified by relevant confounders, such as lesion and organ characteristics, disease stages, and imaging equipment.