(95 days)
The Alta™ Multipurpose Device is intended to immobilize, support, position, and transfer patients undergoing radiotherapy procedures using electrons, protons, including SBRT and SRS; imaging procedures such as x-ray, computed tomography, and magnetic resonance imaging; and other procedures involving transfer of a patient.
The Alta™ Multipurpose Device is designed to interface with the Symphony® Patient Transport System and other positioning devices, such as couch tops, thermoplastic masks, setup-specific treatment devices and adapters, and positioning supports and pads.
Qfix has developed a new device designed to streamline radiotherapy and imaging workflows, the Alta™ Multipurpose Device. The Alta Multipurpose Device is intended to immobilize, support, position, and transfer patients undergoing radiotherapy procedures using electrons, photons, protons, including SBRT and SRS; imaging procedures such as x-ray, computed tomography, and magnetic resonance imaging: and other procedures involving transfer of a patient. The Alta Multipurpose Device is designed to interface with the Symphony® Patient Transport System and other positioning devices, such as couch tops, thermoplastic masks, setup-specific treatment devices and adapters, and positioning supports and pads.
The Alta Multipurpose Device is a versatile device, providing access for a variety of clinical setups for transport, imaging, and treatment on a single device. The proprietary fiber-reinforced composite construction of the Alta Multipurpose Device provides strong and rigid support while remaining lightweight for ease of transport and use. The Alta Multipurpose Device boad-reaching compatibility with a variety of positioning and immobilization devices, including but not limited to:
- Headrests, adapters, and thermoplastics for head and neck applications, including intracranial, whole brain, and stereotactic radiosurgery (SRS),
- MR Coil Holders, ●
- Vacuum cushions for stereotactic body radiation therapy (SBRT), ●
- . Bridges and compression devices for SBRT.
- Supine positioning devices for breast and thorax applications,
- Thoracic and pelvis thermoplastics for thorax and pelvis applications, ●
- Upper and lower extremity positioning devices, like hand grips, arm positioners. ● and knee and foot blocks.
When combined with such other devices through its variety of indexing features, the Alta Multipurpose Device enables a diverse set of clinical applications within the fields of radiotherapy and diagnostic imaging all with the same device, providing significant flexibility, agility, and efficiency for clinicians. Additionally, the Alta Multipurpose Device can optionally be configured to utilize Symphony Air Drive™ technology to facilitate patient transfers with the Symphony Patient Transport System, further enabling more efficient clinical workflows.
1. Table of Acceptance Criteria and Reported Device Performance:
| Test | Acceptance Criteria | Device Performance |
|---|---|---|
| Verification of Hardware Specifications | Met all specified hardware requirements and functionalities. | Met: The Alta™ Multipurpose Device met all acceptance criteria for testing conducted and was appropriately validated to its intended use. (Implied from the general statement of conformity, specific performance details not provided). |
| Verification of MR Safety Characteristics | Demonstrated safe operation within Magnetic Resonance (MR) environments. | Met: The Alta™ Multipurpose Device is compatible with MR environments. (Implied from the general statement of conformity, specific performance details not provided). |
| Verification of Compatibility with Other Devices | Successfully interfaced with specified patient transport systems and positioning devices. | Met: The Alta™ Multipurpose Device is designed to interface with the Symphony® Patient Transport System and other positioning devices, and its broad-reaching compatibility was verified. (Implied from the general statement of conformity). |
| Verification of Attenuation Characteristics | Met acceptable levels of radiation attenuation for radiotherapy and imaging. | Met: The Alta™ Multipurpose Device is radiolucent and low attenuating. (Implied from the general statement of conformity, specific performance details not provided). |
| Load Rating | Withstood specified patient and accessory loads without structural failure. | Met: The proprietary fiber-reinforced composite construction provides strong and rigid support. (Implied from the general statement of conformity, specific performance details not provided). |
| Timed Workflow Studies | Achieved improvements in workflow efficiency during patient transfers and procedures. | Met: The subject device enables simplified and streamlined workflows and is designed to accommodate Symphony® Air Drive™ technology to facilitate patient transfers with the Symphony Patient Transport System, enabling more efficient clinical workflows. (Implied from the general statement of conformity). |
| Ease of Use/Ergonomic Assessments | Demonstrated improved ergonomics and usability features. | Met: The subject device has improved ergonomics and usability features. (Implied from the general statement of conformity). |
2. Sample Size Used for the Test Set and Data Provenance:
The document does not explicitly state the sample size used for the test set or the data provenance (e.g., country of origin, retrospective or prospective). It broadly mentions "Non-clinical bench testing was conducted," implying laboratory-based testing rather than studies involving human patients or specific data sets.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications:
This information is not provided in the document. The testing described is "non-clinical bench testing," which typically involves engineering and performance evaluations against predefined specifications, not expert-adjudicated ground truth as would be used for image interpretation or diagnostic accuracy studies.
4. Adjudication Method for the Test Set:
This information is not applicable and therefore not provided, as the testing described is non-clinical bench testing against predetermined technical specifications, not human interpretation requiring adjudication.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size:
No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The document focuses on the technical performance and substantial equivalence of the device, not its impact on human reader performance.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:
Yes, a standalone performance evaluation was done. The "Performance Standards and Testing" section describes "Non-clinical bench testing" conducted to verify hardware specifications, MR safety, compatibility, attenuation characteristics, load rating, workflow, and ergonomics. This testing evaluates the device itself, without human interaction as part of the primary performance metric.
7. Type of Ground Truth Used:
The ground truth for the non-clinical bench testing appears to be based on engineering specifications, industry standards, and predefined acceptance criteria for each test (e.g., specific load capacities, attenuation limits, compatibility requirements). It is not based on expert consensus, pathology, or outcomes data.
8. Sample Size for the Training Set:
This information is not applicable. The Alta™ Multipurpose Device is a physical medical device (immobilization, support, positioning, and transfer device), not an artificial intelligence (AI) algorithm that requires a training set.
9. How the Ground Truth for the Training Set Was Established:
This information is not applicable as the device is not an AI algorithm and therefore does not have a training set or associated ground truth establishment for a training set.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font size below that.
February 28, 2020
Qfix % Alexandra Low Smythe Senior Regulatory Affairs Specialist and Intellectual Property Specialist 440 Church Road AVONDALE PA 19311
Re: K193243
Trade/Device Name: Alta™ Multipurpose Device Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: Class II Product Code: IYE, LNH, JAI, JAK, LHN, KPS, OUO, FRZ Dated: February 10, 2020 Received: February 11, 2020
Dear Alexandra Low Smythe:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
{1}------------------------------------------------
https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known)
K193243
Device Name Alta™ Multipurpose Device
Indications for Use (Describe)
The Alta™ Multipurpose Device is intended to immobilize, support, position, and transfer patients undergoing radiotherapy procedures using electrons, protons, including SBRT and SRS; imaging procedures such as x-ray, computed tomography, and magnetic resonance imaging; and other procedures involving transfer of a patient.
The Alta™ Multipurpose Device is designed to interface with the Symphony® Patient Transport System and other positioning devices, such as couch tops, thermoplastic masks, setup-specific treatment devices and adapters, and positioning supports and pads.
| Type of Use (Select one or both, as applicable) |
|---|
| ------------------------------------------------- |
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Q-Fix. The logo is a large, red letter "Q" with the word "FIX" in black letters inside the "Q". The "I" in "FIX" is dotted with a black circle. There is a registered trademark symbol to the right of the "X".
510(K) SUMMARY
K193243
| I.GENERAL INFORMATIONEstablishment: | WFR/Aquaplast Corporation/Anholt Technologies Inc., Dba Qfix440 Church RoadAvondale, PA 19311 USA |
|---|---|
| Date Prepared: | November 22, 2019 |
| Manufacturer: | Qfix440 Church RoadAvondale, PA 19311 USARegistration Number: 2247992 |
| Contact Person: | Alexandra Low SmytheSenior Regulatory Affairs Specialist and Intellectual PropertySpecialistQfix440 Church RoadAvondale, PA 19311 USAPhone: 610 268-0585 Ext 736Fax: 610-268-0588 |
| Device Name: | Alta™ Multipurpose Device |
| Trade Name: | Alta™ Multipurpose Device |
| Common Name: | Multipurpose Positioning Solution and Accessories |
| Classification Name: | Accelerator, Linear, Medical |
| Classification Panel: | Radiology |
| Regulation Number: | 21 CFR § 892.5050 |
| Device Class: | II |
| Product Code: | Primary: IYESecondary: LNH, LHN, JAI, JAK, KPS, OUO, FRZ |
II.Safety and Effectiveness Information Supporting Substantial Equivalence
Indications for Use
The Alta™ Multipurpose Device is intended to immobilize, support, position, and transfer patients undergoing radiotherapy procedures using electrons, photons, protons, including SBRT and SRS; imaging procedures such as x-ray, computed tomography, and magnetic resonance imaging; and other procedures involving transfer of a patient.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image is a logo for a company called Q-Fix. The logo features a large, red letter "Q" with the word "Fix" in black letters superimposed over the bottom portion of the "Q". There is a registered trademark symbol to the right of the word "Fix". The logo is simple and eye-catching, and the use of red and black gives it a bold and professional look.
The Alta Multipurpose Device is designed to interface with the Symphony® Patient Transport System and other positioning devices, such as couch tops, thermoplastic masks, setup-specific treatment devices and adapters, and positioning supports and pads.
Device Description
Qfix has developed a new device designed to streamline radiotherapy and imaging workflows, the Alta™ Multipurpose Device. The Alta Multipurpose Device is intended to immobilize, support, position, and transfer patients undergoing radiotherapy procedures using electrons, photons, protons, including SBRT and SRS; imaging procedures such as x-ray, computed tomography, and magnetic resonance imaging: and other procedures involving transfer of a patient. The Alta Multipurpose Device is designed to interface with the Symphony® Patient Transport System and other positioning devices, such as couch tops, thermoplastic masks, setup-specific treatment devices and adapters, and positioning supports and pads.
The Alta Multipurpose Device is a versatile device, providing access for a variety of clinical setups for transport, imaging, and treatment on a single device. The proprietary fiber-reinforced composite construction of the Alta Multipurpose Device provides strong and rigid support while remaining lightweight for ease of transport and use. The Alta Multipurpose Device boad-reaching compatibility with a variety of positioning and immobilization devices, including but not limited to:
- Headrests, adapters, and thermoplastics for head and neck applications, including intracranial, whole brain, and stereotactic radiosurgery (SRS),
- MR Coil Holders, ●
- Vacuum cushions for stereotactic body radiation therapy (SBRT), ●
- . Bridges and compression devices for SBRT.
- Supine positioning devices for breast and thorax applications,
- Thoracic and pelvis thermoplastics for thorax and pelvis applications, ●
- Upper and lower extremity positioning devices, like hand grips, arm positioners. ● and knee and foot blocks.
When combined with such other devices through its variety of indexing features, the Alta Multipurpose Device enables a diverse set of clinical applications within the fields of radiotherapy and diagnostic imaging all with the same device, providing significant flexibility, agility, and efficiency for clinicians. Additionally, the Alta Multipurpose Device can optionally be configured to utilize Symphony Air Drive™ technology to facilitate patient transfers with the Symphony Patient Transport System, further enabling more efficient clinical workflows.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for Q-Fix. The logo consists of a large, red letter "Q" with the word "Fix" in black letters inside the "Q". The "Q" is stylized with a thick stroke, and the "Fix" is in a bold, sans-serif font. There is a small, black circle above the "Fix", and a registered trademark symbol to the right of the "Q".
Predicate information
The subject device, the Alta™ Multipurpose Device, includes all of the device properties belonging to the predicate device, Patient Positioning Devices from MacroMedics, BV. The predicate device information follows:
| Predicate Device Name | FDA ClearanceNumber and Date | Productcode | Manufacturer |
|---|---|---|---|
| Patient Positioning Devices | K142420, clearedMarch 27, 2015 | IYE | MacroMedics, BV |
To date, this predicate device has not been subject to a design-related recall per information that is publicly accessible in the FDA recall database.
Comparison to Predicate Device
The Alta™ Multipurpose Device bears many similarities to its predicate, the MacroMedics Patient Positioning Devices. The Alta Multipurpose Device has the same fundamental Intended Use and purpose as the predicate device. Both the subject device and the predicate device are compatible with radiation therapy environments and diagnostic imaging environments. Both devices are non-sterile, reusable devices. Both devices are primarily constructed from fiber-reinforced composites. Both devices are appropriate for use in magnetic resonance environments, where indicated.
Performance Standards and Testing
The FDA has not established performance standards for this product under Section 514 of the Food, Drug and Cosmetic Act. Testing and analysis has been conducted to show that the verification, validation, and safety requirements have been met and are inclusive of the following considerations:
- Verification of hardware specifications
- Verification of MR safety characteristics ●
- Verification of compatibility with other devices
- Verification of attenuation characteristics
- Load rating
- Timed workflow studies ●
- Ease of use/ergonomic assessments ●
Non-clinical bench testing was conducted to support the intended use and to confirm that the proposed device is safe and effective and that technological changes do not raise any new issues of safety or effectiveness over the predicate device.
The Alta™ Multipurpose Device met all acceptance criteria for testing conducted and was appropriately validated to its intended use.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for QFix. The logo consists of a large, red letter "Q" with the word "FIX" in black letters inside the "Q". The "X" in "FIX" is stylized to extend beyond the other letters. There is a registered trademark symbol to the right of the "X".
Safety and Effectiveness
Risk management is ensured via a risk analysis in compliance with ISO 14971:2007 to identify and provide reduction to potential hazards in a risk analysis beginning early in the design phase and continuing throughout the development of the product. These risks are controlled via measures realized in development, testing, and product labeling. To minimize risks, Qfix adheres to recognized and established industry practices and standards to minimize safety and performance risks. Furthermore, the operators and end users of the device are healthcare professionals familiar with and responsible for radiation therapy treatments, diagnostic imaging procedures, and other hospital procedures.
The device labeling contains instructions for use and any necessary cautions and warnings to provide for safe and effective use of the device.
Substantial Equivalence
The subject device, the Alta™ Multipurpose Device, offers the following improvements over the predicate device, the MacroMedics Patient Positioning Devices (cleared under K142420 on March 27, 2015).
- Subject device is radiolucent, low attenuating, and compatible with MR ● environments, all in one device
- Subject device accommodates MR imaging accessories such as coils and ● coil holders
- Subject device has improved ergonomics and usability features .
- Subject device enables simplified and streamlined workflows .
- Subiect device is designed to accommodate Symphony® Air Drive™ ● technology to facilitate patient transfers with the Symphony Patient Transport System, further enabling more efficient clinical workflows
However, the fundamental attributes of the subject device and the predicate device remain the same.
-
. Both devices have the same fundamental intended use.
The Alta Multipurpose Device has the same fundamental intended use and overall general functionality as it relates to patient positioning, and immobilization as the MacroMedics Patient Positioning Devices (cleared under K142420 on March 27, 2015). -
. Both the predicate and the subject device are intended to be used either independently or in conjunction with additional accessories, components, and standalone devices which are intended to immobilize, position and reposition patients undergoing radiation therapy and diagnostic imaging.
The predicate device is intended to be used in combination with other immobilization devices from MacroMedics. The Alta Multipurpose Device is intended to be used either independently or in conjunction with a number of existing positioning and immobilization devices via its indexing features,
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for Q-Fix. The logo is a red letter "Q" with the word "FIX" in black letters inside the "Q". There is a black dot above the "I" in "FIX", and a small "R" in a circle to the right of the "X".
creating a versatile, flexible solution for a diverse set of clinical applications in radiotherapy and diagnostic imaging.
● Both devices may be used in MR environments. Both the predicate device and subject device are intended for use in an MR environment where indicated. The predicate device provides for separate MRcompatible and non-MR compatible versions of its products, whereas the subject device is compatible with MR environments as well as x-ray, CT, and treatment environments on the same device.
- . Both devices are made of composite materials. Both the predicate device and subject device are made of composite materials.
The conclusions from the non-clinical data suggest that the subject device has the same fundamental technological characteristics with respect to the predicate device and exhibits an equivalent safety and performance profile as that of the predicate device. Therefore, Qfix is of the opinion that the Alta Multipurpose Device does not raise new questions of safety or effectiveness and, therefore, is substantially equivalent to the marketed predicate device.
N/A