K Number
K193083

Validate with FDA (Live)

Date Cleared
2019-12-03

(28 days)

Product Code
Regulation Number
888.3040
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The M.U.S.T. Sacral Iliac Screw and Pelvic Trauma System is intended for use in skeletally mature patients for fracture fixation of small and long bones of the pelvis, and for sacroiliac joint fusion for patients suffering from sacrolliac joint disruptions and degenerative sacroiliitis.

Device Description

The M.U.S.T. Sacral Iliac Screw and Pelvic Trauma System is designed for sacroiliac joint fusion in degenerative sacroiliitis, as well as for the fixation of small and long bone fractures in trauma cases. The system consists of various sized screws, manufactured from Ti6Al4V ELI and coated with rough Hydroxyapatite (HA). The HA coating allows for biological fixation and potentially leads to arthrodesis. The sacroiliac (SI) joint screws are hollow-body threaded fusion devices with a multiplefenestrated pattern shaft to promote arthrodesis, a self-tapping design to facilitate screw insertion, and a tapered tip to aid in guidance through pilot hole. Radial windowed slots along the shaft facilitate bone ingrowth after implantation and "bone filling" after insertion. The SI joint screws are provided in standard and headless designs.

AI/ML Overview

The provided document is a 510(k) premarket notification for a medical device, the M.U.S.T. Sacral Iliac Screw and Pelvic Trauma System. It outlines the device's indications for use, its description, and a comparison with a predicate device.

However, this document does not describe a study involving an AI or algorithm-based device. It pertains to a physical medical implant (screws) and its performance is evaluated through material and mechanical testing, as well as a cadaver model, not through metrics associated with algorithmic performance (e.g., sensitivity, specificity, AUC).

Therefore, I cannot provide the requested information regarding acceptance criteria and study details for an AI-based device, as this document does not refer to such a device.

To directly address the numbered points based on the provided document's content, I can say:

  1. A table of acceptance criteria and the reported device performance: Not applicable. The document lists performance testing related to material properties and mechanical strength (e.g., ASTM F543-13, ASTM F2193-14, ISO 13779-3), but it does not present specific acceptance criteria or reported numerical performance metrics in a table format. It states that "design verification testing supports there are no new risks associated with the subject devices" and that the device is "as safe and effective as the predicate device."
  2. Sample size used for the test set and the data provenance: A "cadaver model" was used for design validation on the headless screws. The sample size for this cadaver model is not specified, nor is the provenance of the cadavers (e.g., country of origin, retrospective or prospective).
  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. The study involved a cadaver model for design validation, not interpretation of data requiring expert consensus or ground truth establishment in the context of an AI device.
  4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
  5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This document does not describe an AI or algorithm-based device, nor a study involving human readers.
  6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This document does not describe an AI or algorithm-based device.
  7. The type of ground truth used (expert consensus, pathology, outcomes data, etc): Not applicable in the context of AI performance. The design validation was performed using a "cadaver model."
  8. The sample size for the training set: Not applicable. This document does not describe an AI or algorithm-based device with a training set.
  9. How the ground truth for the training set was established: Not applicable. This document does not describe an AI or algorithm-based device.

{0}------------------------------------------------

December 3, 2019

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Medacta International SA % Mr. Chris Lussier Director, Quality and Regulatory Medacta USA 3973 Delp Street Memphis, Tennessee 38118

Re: K193083

Trade/Device Name: M.U.S.T. Sacral Iliac Screw and Pelvic Trauma System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: OUR, HWC, HTN Dated: November 5, 2019 Received: November 5, 2019

Dear Mr. Lussier:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for

(Ronald Jean) Vacant Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K193083

Device Name

M.U.S.T Sacral Iliac Screw and Pelvic Trauma System

Indications for Use (Describe)

The M.U.S.T. Sacral Iliac Screw and Pelvic Trauma System is intended for use in skeletally mature patients for fracture fixation of small and long bones of the pelvis, and for sacroiliac joint fusion for patients suffering from sacrolliac joint disruptions and degenerative sacroiliitis.

Type of Use (Select one or both, as applicable)

[X] Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

{3}------------------------------------------------

3.0 510(k) Summary

I. Submitter

Medacta International SA Strada Regina 6874 Castel San Pietro (CH) Switzerland Phone (+41) 91 696 60 60 Fax (+41) 91 696 60 66

Contact Person: Stefano Baj, Regulatory and Compliance Director, Medacta International SA Applicant Correspondent: Chris Lussier, Director of Quality and Regulatory, Medacta USA Date Prepared: November 5, 2019

Date Revised: December 3, 2019

II. Device

Device Proprietary Name:M.U.S.T. Sacral Iliac Screw and Pelvic Trauma System
Common or Usual Name:Pelvic Joint Fixation
Classification Name:Smooth or threaded metallic bone fixation fastener
Primary Product Code:OUR
Secondary Product Codes:HWC, HTN
Regulation Number:21 CFR 888.3040, 21 CFR 888.3030
Device ClassificationII

III. Predicate Device

Substantial equivalence is claimed to the following device:

  • M.U.S.T. Sacral Iliac Screw and Pelvic Trauma System, K171595, Medacta International SA (Primary Predicate)

IV. Device Description

The M.U.S.T. Sacral Iliac Screw and Pelvic Trauma System is designed for sacroiliac joint fusion in degenerative sacroiliitis, as well as for the fixation of small and long bone fractures in trauma cases.

The M.U.S.T. Sacral Iliac Screw and Pelvic Trauma System consists of various sized screws, manufactured from Ti6Al4V ELI and coated with rough Hydroxyapatite (HA). The HA coating allows for biological fixation and potentially leads to arthrodesis.

{4}------------------------------------------------

The sacroiliac (SI) joint screws are hollow-body threaded fusion devices with a multiplefenestrated pattern shaft to promote arthrodesis, a self-tapping design to facilitate screw insertion, and a tapered tip to aid in guidance through pilot hole. Radial windowed slots along the shaft facilitate bone ingrowth after implantation and "bone filling" after insertion.

The SI joint screws are provided in standard and headless designs. The standard screw can be coupled to modular washers of different diameters to optimize and stabilize the contact between the screw head and the cortical bone and to improve the compression activity of the screw. The standard screws are provided sterile in three diameters (8, 9, and 10 mm) and multiple lengths (25 - 80 mm).

The headless screw has an anatomical shape which allows for full insertion into the bone. The headless screws are provided sterile in three diameters (7.5, 9, and 11 mm) and multiple lengths (30 – 75 mm).

V. Indications for Use

The M.U.S.T. Sacral Iliac Screw and Pelvic Trauma System is intended for use in skeletally mature patients for fracture fixation of small and long bones of the pelvis, and for sacroiliac joint fusion for patients suffering from sacroiliac joint disruptions and degenerative sacroiliitis.

VI. Comparison of Technological Characteristics

The M.U.S.T. Sacral Iliac Screw and Pelvic Trauma System SI joint screws and the predicate devices share the following characteristics:

  • length;
  • material of construction;
  • coating and coating composition;
  • biocompatibility;
  • sterility:
  • shelf life; and
  • packaging.

The M.U.S.T. Sacral Iliac Screw and Pelvic Trauma System SI joint screws are technologically different from the predicate device with respect to provision of headless joint screws and location of the radial windowed slots on the screw tip and screw shaft.

These differences do not raise different questions of safety or effectiveness and design verification testing supports there are no new risks associated with the subject devices.

{5}------------------------------------------------

VII. Performance Data

The introduction of headless screws and the change in location of the fenestrations does not create a new worst case; therefore, the following performance testing from the predicate device was leveraged to support this submission:

  • ASTM F543-13 Standard Specification and Test Methods for Metallic Medical Bone Screws:
  • ASTM F2193-14 Standard Specifications and Test Methods for Components Used in the Surgical Fixation of the Spinal Skeletal System;
  • ISO 13779-3: 2008 Implants for Surgery Hydroxyapatite Part 3: Chemical Analysis ● and Characterization of Crystallinity and Phase Purity;
  • ASTM F1185-03 (Reapproved 2014) Standard Specification for Composition of Hydroxyapatite for Surgical Implants;
  • Bacterial Endotoxin Testing (LAL test and USP <151>); ●
  • sterilization validation; and
  • shelf-life testing.

Design validation on the headless screws was undertaken using a cadaver model.

VIII. Conclusion

The information provided within this submission supports that the M.U.S.T. Sacral Iliac Screw and Pelvic Trauma System standard and headless SI joint screws are as safe and effective as the predicate device; therefore, the M.U.S.T. Sacral Iliac Screw and Pelvic Trauma System standard and headless SI joint screws are substantially equivalent to the predicate device.

§ 888.3040 Smooth or threaded metallic bone fixation fastener.

(a)
Identification. A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.