(22 days)
The C2 CryoBalloon™ EndoGrip is intended to provide access for endoscopic device passage and exchange throughout the endoscopic procedure.
The C2 CryoBalloon™ EndoGrip is an accessory product that is used in conjunction with therapeutic endoscopes and therapeutic catheters (the C2 CryoBalloon™ Catheter of the C2 CryoBalloon™ Ablation System). The C2 CryoBalloon™ EndoGrip consists of a Clip Base, Rubber Grip, and Compression Spring. This is supplied non-sterile. The EndoGrip is used to anchor catheters in place relative to the scope. In a clinical application, once the endoscope is delivered into the gastric anatomy and positioned at its desired location the physician may insert the catheter through the endoscope. When the catheter is located properly in its axial position, the EndoGrip is positioned against the biopsy valve by depressing the clip mechanism and sitting the catheter on the rubber grips. The revised design allows for placing the EndoGrip after the CryoBalloon catheter is in place in the scope and for repositioning without having to completely withdraw and reinsert the catheter. The C2 CryoBalloon™ EndoGrip is designed for one-time, continuous application use (single patient) in conjunction with a therapeutic endoscope (3.7mm accessory channel ID) or a C2 Sidecar (FG -1011), which serves as an external channel for therapeutic devices when a therapeutic endoscope is unavailable for use.
The PENTAX Medical C2 CryoBalloon™ EndoGrip (K193036) is an accessory for endoscopic procedures. The information provided in the 510(k) summary focuses on demonstrating substantial equivalence to a predicate device through performance testing rather than an AI/ML-driven study. Therefore, sections typically relevant to AI/ML device studies (such as MRMC studies, effect size of AI assistance, sample size for training sets, etc.) are not applicable here.
Here's a breakdown of the available information regarding acceptance criteria and performance testing:
1. A table of acceptance criteria and the reported device performance
| Test Category | Acceptance Criteria | Reported Device Performance |
|---|---|---|
| Physical Characteristics | Verification of dimensions of the device. | All tests passed and met predetermined acceptance criteria. |
| Simulated Use, Performance, & Reliability | - Device can be attached to and detached from an endoscope. - Withstands 10 endoscopic delivery cycles. - Compatible with endoscopic devices. - Resists dislodgement forces. - Resists kinking. | All tests passed and met predetermined acceptance criteria. The device performs as intended in the specified use conditions and performs comparably to the legally marketed predicate device. |
| Usability Testing | Validation of user needs. | All tests passed and met predetermined acceptance criteria. |
2. Sample size used for the test set and the data provenance
The document does not specify the exact sample size for the test set for dimensional verification, simulated use, and usability testing. It refers to these as "evaluations" and "testing" without providing numerical details for the quantity of devices or cycles tested.
3. Number of experts used to establish the ground truth for the test set and their qualifications
This information is not provided. The performance testing appears to be primarily engineering and functional verification, rather than clinical efficacy studies requiring expert ground truth for interpretation of medical images or data.
4. Adjudication method for the test set
Not applicable. The described tests are functional and physical performance evaluations where outcomes are objective (e.g., successful attachment, resistance to dislodgement) rather than subjective assessments requiring adjudication.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done
No, an MRMC comparative effectiveness study was not done. The document describes performance testing for an endoscopic accessory, not an AI/ML-driven diagnostic or therapeutic device that would typically involve human-in-the-loop performance evaluation.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
No, this is not applicable. The device is a physical accessory (EndoGrip), not an algorithm. Therefore, "standalone" performance in the context of an algorithm is not relevant.
7. The type of ground truth used
The "ground truth" for the performance testing cited would be direct physical measurements and engineering specifications, or the successful completion of specified functional tasks. For example:
- Physical characteristic evaluations: Ground truth would be the design specifications and manufacturing tolerances.
- Simulated use, performance, and reliability testing: Ground truth would be the defined functional requirements (e.g., "device can be attached," "withstand 10 cycles," "resist dislodgement").
- Usability testing: Ground truth would be the user needs and successful completion of tasks by users.
8. The sample size for the training set
Not applicable. This is a physical medical device, not an AI/ML system that requires a training set.
9. How the ground truth for the training set was established
Not applicable, as there is no training set for this device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is a symbol representing the Department of Health & Human Services-USA, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The FDA logo is a symbol of the agency's mission to protect and promote public health through the regulation and supervision of food safety, tobacco, pharmaceuticals, medical devices, and other related products.
November 22, 2019
PENTAX Medical, A Division of PENTAX of America, Inc. Gurvinder Nanda Senior Director of Regulatory Affairs 303 Convention Way, Suite 1 Redwood City, California 94063
Re: K193036
Trade/Device Name: C2 CryoBalloon EndoGrip Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: Class II Product Code: OCX Dated: October 30, 2019 Received: October 31, 2019
Dear Gurvinder Nanda:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K193036
Device Name C2 CryoBalloon™ EndoGrip
Indications for Use (Describe)
The C2 CryoBalloon™ EndoGrip is intended to provide access for endoscopic device passage and exchange throughout the endoscopic procedure.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
PENTAX Medical C2 CryoBalloon™ EndoGrip (K193036)
Image /page/3/Picture/1 description: The image shows the logo for Pentax Medical. The word "PENTAX" is in red, bold letters, and it is positioned above the word "MEDICAL", which is in gray, bold letters. There is a gray line separating the two words.
510(k) Summary
This 510(k) Summary is being submitted in accordance with the requirements of 21 CFR 807.92.
I. SUBMITTER
PENTAX Medical 303 Convention Way, Suite 1 Redwood City, CA 94063
Phone: 650-316-8601 Fax: 650-316-8601
Gurvinder Singh Nanda, Ph.D. Contact Person: Senior Director of Regulatory Affairs
Date Prepared: October 30, 2019
II. DEVICE
Name of Device: C2 CryoBalloon™ EndoGrip Common Name: EndoGrip Classification Name: Endoscopic and Accessories Classification Code: 21 CFR §876.1500 Regulatory Class: Class II Product Code: OCX
III. PREDICATE DEVICE
BOA Endoscopic Valve (K152140)
This predicate device has not been subject to a design-related recall.
IV. DEVICE DESCRIPTION
The C2 CryoBalloon™ EndoGrip is an accessory product that is used in conjunction with therapeutic endoscopes and therapeutic catheters (the C2 CryoBalloon™ Catheter of the C2 CryoBalloon™ Ablation System). The C2 CryoBalloon™ EndoGrip consists of a Clip Base, Rubber Grip, and Compression Spring. This is supplied non-sterile. The EndoGrip is used to anchor catheters in place relative to the scope. In a clinical application, once the endoscope is delivered into the gastric anatomy and positioned at its
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for Pentax Medical. The word "PENTAX" is in red, bold, sans-serif font. Below it is a gray line, and below that is the word "MEDICAL" in gray, bold, sans-serif font.
desired location the physician may insert the catheter through the endoscope. When the catheter is located properly in its axial position, the EndoGrip is positioned against the biopsy valve by depressing the clip mechanism and sitting the catheter on the rubber grips. The revised design allows for placing the EndoGrip after the CryoBalloon catheter is in place in the scope and for repositioning without having to completely withdraw and reinsert the catheter. The C2 CryoBalloon™ EndoGrip is designed for one-time, continuous application use (single patient) in conjunction with a therapeutic endoscope (3.7mm accessory channel ID) or a C2 Sidecar (FG -1011), which serves as an external channel for therapeutic devices when a therapeutic endoscope is unavailable for use.
V. INDICATIONS FOR USE
The C2 CryoBalloon™ EndoGrip is intended to provide access for endoscopic device passage and exchange throughout the endoscopic procedure.
VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The subject C2 CrvoBalloon™ EndoGrip has similar technological characteristics to the legally marketed predicate C2 BOA Endoscopic Valve (K152140). The subject and predicate devices are based on the following same technological elements:
- Attached against the proximal biopsy channel port of an endoscopic device to ● provide endoscopic device passage and exchange
- . Used in hospitals, sub-acute care institutions, surgery centers, or doctor's office where endoscopic procedures may be performed
- . Used in single patient (disposable) undergoing endoscopic procedures
The following technological differences exist between the modified device and the predicate device:
The modified device is secured through the use of a spring-clip mechanism, and the predicate device is secured through the use of a twist-Tuohy tightening mechanism.
PERFORMANCE DATA VII.
The performance testing for the C2 CryoBalloon™ EndoGrip included:
-
- Physical characteristic evaluations for dimensional verification
-
- Simulated use, performance, and reliability testing were performed to verify the device can be attached to and detached from endoscope, withstand 10
{5}------------------------------------------------
PENTAX Medical C2 CryoBalloon™ EndoGrip (K193036)
Image /page/5/Picture/1 description: The image contains the logo for Pentax Medical. The word "PENTAX" is in red, bold, sans-serif font. A gray line is underneath the word "PENTAX". Below the gray line, the word "MEDICAL" is in gray, bold, sans-serif font.
endoscopic delivery cycles, compatible with endoscopic devices, resist dislodgement forces, and resist kinking.
-
- Usability testing for validation of user needs
All the tests passed and met the predetermined acceptance criteria.
- Usability testing for validation of user needs
VIII. CONCLUSION
The subject C2 CryoBalloon™ EndoGrip has the same clinical attributes, technological characteristics and expected performance as the legally marketed predicate, C2 BOA Endoscopic valve (K152140). The performance data demonstrate that the subject C2 CryoBalloon™ EndoGrip performs as intended in the specified use conditions and performs comparably to the legally marketed predicate device that is currently marketed for the same intended use.
§ 876.1500 Endoscope and accessories.
(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.