K Number
K193015
Date Cleared
2020-02-20

(114 days)

Product Code
Regulation Number
880.5970
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

HydroPICC is indicated for short- or long-term peripheral access to the central venous system for intravenous therapy, including but not limited to; the administration of fluids, medications, and nutrients; the sampling of blood; central venous pressure monitoring; and power injection of contrast media.

HydroPICC is rated for a Maximum Power Injection Flow Rate of 3.5mL/sec.

Device Description

The HydroPICC is a peripherally inserted central catheter (PICC) intended for short or long-term use. The HydroPICC is a 4Fr, 55cm long catheter suitable for periodic access to the superior vena cava (SVC) while maintaining a reduced level of thrombus accumulation. HydroPICC is supplied in a kit with the necessary accessories to insert and maintain the catheter.

AI/ML Overview

This document is a 510(k) K193015 Premarket Notification for the HydroPICC (Model 142) intravascular catheter, asserting its substantial equivalence to a previously cleared predicate device, HydroPICC (Model 141, K172885).

The submission focuses heavily on demonstrating equivalence and does NOT include a study involving AI assistance for human readers, nor a standalone AI algorithm performance study. Therefore, most of the requested information regarding AI study design, ground truth establishment for AI, expert involvement, and MRMC studies is not present in this document.

The document describes performance testing for a medical device (intravascular catheter), not an AI algorithm. The acceptance criteria and performance data described relate to physical and mechanical properties of the catheter, as distinct from the performance metrics (e.g., sensitivity, specificity, AUC) typically associated with AI/ML medical devices.

Here's the breakdown of the information that CAN be extracted, followed by a clear statement of what is NOT in the document:

1. Table of Acceptance Criteria and Reported Device Performance

The document states that "The subject devices met all predetermined acceptance criteria derived from the applicable guidance documents, standards and in-house protocols," but it does not provide a specific table with numerical acceptance criteria and reported performance values. Instead, it lists the types of tests performed and the standards followed.

Test CategorySpecific Tests PerformedAcceptance Criteria (Not explicitly quantified in text, stated as "met all predetermined criteria")Reported Device Performance (Not explicitly quantified in text, stated as "met all predetermined criteria")
Mechanical & Performance TestingPower Injection Flow RateDerived from standards/internal protocolsSuccessful/Compliant (implied)
Static Burst StrengthDerived from standards/internal protocolsSuccessful/Compliant (implied)
Multiple Power InjectionsDerived from standards/internal protocolsSuccessful/Compliant (implied)
Catheter LengthDerived from standards/internal protocolsSuccessful/Compliant (implied)
Dimensional Verification (including ID, OD, Length)Derived from standards/internal protocolsSuccessful/Compliant (implied)
Catheter Kink/Flex ResistanceDerived from standards/internal protocolsSuccessful/Compliant (implied)
Tensile Testing (of Catheter and Assembly)Derived from standards/internal protocolsSuccessful/Compliant (implied)
Particulate testingDerived from standards/internal protocolsSuccessful/Compliant (implied)
Shelf life testingDerived from standards/internal protocolsSuccessful/Compliant (implied)
Sterility testingDerived from standards/internal protocolsSuccessful/Compliant (implied)
Packaging distribution testingDerived from standards/internal protocolsSuccessful/Compliant (implied)
MR Compatibility testingDerived from standards/internal protocolsSuccessful/Compliant (implies HydroPICC-142 is labeled with MRI Safety Information, unlike predicate)
Evaluation of Humectant change on Thrombus accumulation compared to predicate (internal specifications)Derived from standards/internal protocolsSuccessful/Compliant (implied)
Biocompatibility TestingRisk assessment evaluation that the proposed modifications did not impact biocompatibilityDerived from standards/internal protocolsSuccessful/Compliant (implied)

2. Sample Size Used for the Test Set and Data Provenance

The document does not specify the sample size used for the test set for each individual test. It merely states that "The subject devices met all predetermined acceptance criteria." The provenance of the data (e.g., country of origin, retrospective/prospective) is not mentioned for this engineering/performance testing.

3. Number of Experts Used to Establish Ground Truth and Qualifications

Not applicable. This document describes the testing of a medical device (catheter), not the performance of an AI algorithm based on expert-labeled data.

4. Adjudication Method for the Test Set

Not applicable. This is for medical device performance (physical, mechanical, safety), not for AI algorithm evaluation requiring expert consensus/adjudication.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No MRMC comparative effectiveness study was done because this submission is not for an AI/ML medical device and does not involve human readers interpreting images with or without AI assistance.

6. Standalone Algorithm Performance

No standalone algorithm performance study was done relevant to AI/ML, as this is not an AI/ML device.

7. Type of Ground Truth Used

The "ground truth" for this device's performance testing is established by engineering standards and specifications (e.g., ISO, ASTM standards, FDA guidance documents, and internal protocols) for physical and mechanical properties, not expert consensus on medical images, pathology, or outcomes data in the context of an AI study.

8. Sample Size for the Training Set

Not applicable. This document describes testing for a manufactured medical device, not an AI/ML algorithm that requires a training set.

9. How the Ground Truth for the Training Set Was Established

Not applicable for the same reason as point 8.


Summary of what is NOT in the document regarding acceptance criteria and studies for an AI/ML device:

The provided document is a 510(k) premarket notification for a physical medical device (intravascular catheter). It details the substantial equivalence to a predicate device and describes the various engineering and performance tests conducted to ensure the device meets safety and performance standards. It does not contain any information about an AI/ML device, its acceptance criteria, expert consensus, ground truth establishment, or any studies related to its performance (standalone or human-in-the-loop). Therefore, the specific requirements of the prompt regarding AI are not addressed by this document.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters.

February 20, 2020

Access Vascular, Inc Brian Hanley Sr. Director, Operations 175 Middlesex Turnpike Suite 1A Bedford, Massachusetts 01730

Re: K193015

Trade/Device Name: HydroPICC Regulation Number: 21 CFR 880.5970 Regulation Name: Percutaneous, Implanted, Long-Term Intravascular Catheter Regulatory Class: Class II Product Code: LJS Dated: January 17, 2020 Received: January 21, 2020

Dear Brian Hanley:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Geeta Pamidimukkala Acting Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K193015

Device Name

HydroPICC

Indications for Use (Describe)

HydroPICC is indicated for short- or long-term peripheral access to the central venous system for intravenous therapy, including but not limited to; the administration of fluids, medications, and nutrients; the sampling of blood; central venous pressure monitoring; and power injection of contrast media.

HydroPICC is rated for a Maximum Power Injection Flow Rate of 3.5mL/sec.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Access Vascular. The word "ACCESS" is in a dark blue font, and the word "VASCULAR" is in a lighter blue font. The bottom of the "A" in "ACCESS" is connected to the "V" in "VASCULAR" by a blue line.

K193015 510(k) Summary for the HydroPICC PICC

Date prepared: 18 February, 2020

Submitter:

Access Vascular, Inc. 175 Middlesex Turnpike Suite 1A Bedford, MA 01730 Tel. 781-538-6594

Contact:

Brian Hanley Access Vascular, Inc. Tel. 781-538-6594

Subject Device

Trade Name:HydroPICC
Common Name:Intravascular Catheter
Regulation Number:21CFR§880.5970
Regulation Name:Percutaneous, Implanted, Long-Term Intravascular Catheter
Regulatory Class:Class II
Product Code:LJS
Classification Panel:General Hospital

Primary Predicate Device

Trade Name:HydroPICC
Manufacturer:Access Vascular
510(k) Reference:K172885
Common Name:Intravascular Catheter
Regulation Number:21 CFR§880.5970
Regulation Name:Percutaneous, Implanted, Long-Term Intravascular Catheter
Regulatory Class:Class II
Product Code:LJS
Classification Panel:General Hospital

{4}------------------------------------------------

Device Description

The device description of HydroPICC-142 is the same as the predicate device, HydroPICC-141 (K172885).

The HydroPICC is a peripherally inserted central catheter (PICC) intended for short or long-term use. The HydroPICC is a 4Fr, 55cm long catheter suitable for periodic access to the superior vena cava (SVC) while maintaining a reduced level of thrombus accumulation. HydroPICC is supplied in a kit with the necessary accessories to insert and maintain the catheter. The patient population for the proposed and predicate devices are identical, namely patients who require short- and long-term vascular access.

The following accessories are provided with the HydroPICC:

  • Tear-away Introducer
  • Introducer needle
  • Guidewire
  • Male Cap Plug
  • 60cm. Paper Tape Measure
  • Needle-Free Valve
  • Luer Lock Syringe
  • Adhesive Fixation Device - Bard StatLock
  • Scalpel
  • Transparent Film Dressing

Indications for Use

HydroPICC is indicated for short- or long-term peripheral access to the central venous system for intravenous therapy, including but not limited to; the administration of fluids, medications, and nutrients; the sampling of blood; central venous pressure monitoring; and power injection of contrast media. HydroPICC is rated for a Maximum Power Injection Flow Rate of 3.5ml/sec. The Indications for Use statement for the HydroPICC-142 is identical to the predicate device, HydroPICC-141 (K172885).

Comparison of Technological Characteristics with the Predicate Device

The technological characteristics of the HydroPICC-142 are substantially equivalent to the predicate, the HydroPICC-141 (K172885), in terms of intended use, application, user population, basic design, performance, and labeling. The hydration aid, nominal catheter dimensions, and product packaging were modified in order to reduce the time it takes for the catheter to hydrate during procedural setup. A radiopaque additive was removed from the suture wing overmold. The labeling of the HydroPICC-142 was revised to indicate MRI Safety Information following the evaluation of the device.

Briefly, both the subject and predicate device are,

{5}------------------------------------------------

  • intended for short- or long-term peripheral access to the central venous system for intravenous therapy, including but not limited to, the administration of fluids, medications, nutrients; the sampling of blood; for central venous pressure monitoring and for power injection of contrast media.
  • available in 4 French size
  • rated for maximum power injection flow rate up to 3.5mL/sec
  • available kitted with a range of procedural accessories for user convenient; and
  • demonstrative of enhanced resistance to blood component (platelet and thrombus) accumulation
HydroPICC-142Access VascularProposed DeviceHydroPICC-141Access VascularK172885
Specification
Intended Useintended for short- or long-termperipheral access to the centralvenous system for intravenoustherapy.Same
Indication for UseHydroPICC is indicated for short- orlong-term peripheral access to thecentral venous system forintravenous therapy, including butnot limited to; the administrationof fluids, medications, andnutrients; the sampling of blood;central venous pressuremonitoring; and power injection ofcontrast media.HydroPICC is rated for a MaximumPower Injection Flow Rate of3.5mL/sec.Same
Device ClassClass IISame
Product CodeLJSSame
Regulation Number880.5970Same
Prescription DeviceYesSame
Catheter TypePeripherally Inserted CentralCatheter (PICC)Same
Catheter Outer Diameter4 FrenchSame
Catheter Inner Diameter1mmSubstantially equivalent
Usable Catheter Length55cmSame
Guidewire compatibility0.018"Same
Catheter Shaft DesignNo TaperSame
Number of Catheter LumensSingleSame

Table 1 Summary of Technological Characteristics of Proposed and Predicate Devices

{6}------------------------------------------------

SpecificationHydroPICC-142Access VascularProposed DeviceHydroPICC-141Access VascularK172885
Key Device ComponentsCatheter Shaft, Suture Wing,Extension Tube, Luer HubSame
Short or Long Term AccessYesSame
Use with Power InjectionPower Settings Flow RateYesFlow rate: 3.5mL/secSame
Catheter MaterialsRadiopaque hydrophilic polyolcatheter with luer lock hub,polyurethane extension tub, andsuture wing without radiopaqueadditiveSame - suture wing containedradiopaque additive
X-Ray ConfirmationRequiredYesSame
Sterilization MethodEthylene OxideSame
Single UseYesSame
How suppliedConvenience kitSame
MRI SafetyEvaluated – labeled with MRISafety InformationNot evaluated – labeled withcontraindication to use in a MRenvironment

Performance Data

Verification and validation tests have been performed in accordance with Design Controls per 21 CFR §820.30. The subject devices met all predetermined acceptance criteria derived from the applicable guidance documents, standards and in-house protocols, which were used to determine appropriate methods for evaluating the performance of the device.

The following performance tests were performed in support of the substantial equivalence determination:

Mechanical and Performance Testing

The testing and test methods that were performed for HydroPICC-142 are substantially equivalent to testing and tests methods that were performed for the predicate, the HydroPICC-141 (K172885). Testing was conducted in accordance with the following FDA Guidance Document and international standards:

  • . FDA's "Guidance on Premarket Notification [510(k)] Submissions for Short-Term and Long-Term Intravenous Catheter"
  • . EN ISO 10555-1:2013 – "Intravascular Catheters – Sterile and Single-Use Catheters – Part 1: General Requirements"
  • . EN ISO 10555-3:2013 – "Intravascular Catheters – Sterile and Single-Use Catheters – Part 3 Central Venous Catheters:

{7}------------------------------------------------

  • . EN ISO 11135:2014 "Sterilization of Health-Care Products - Ethylene Oxide - Requirements for The Development, Validation and Routine Control Of A Sterilization Process For Medical Device"
  • . EN ISO 13868:2014 "Catheters – Test method for kinking of single lumen catheters and medical tubing"
  • . ASTM F2052:2015 "Standard Test Method for Measurement of Magnetically Induced Displacement Force on Medical Devices in the Magnetic Resonance Environment"
  • ASTM F2182: 2011 "Standard Test Method for Measurement of Radio Frequency Induced Heating on or Near Passive Implants During Magnetic Resonance Imaging"
  • . ASTM F2213: 2017 "Standard Test Method for Measurement of Magnetically Induced Torque on Medical Devices in the Magnetic Resonance Environment"
  • . ASTM F2119: 2013 "Standard Test Method for Evaluation of MR Image Artifacts from Passive Implants"
  • USP <788> Particulate Matter in Injections (Method 1)
  • Evaluation of Humectant change on Thrombus accumulation compared to predicate (internal specifications)

The following tests were conducted in accordance with the standards identified above as applicable:

  • . Power Injection Flow Rate
  • Static Burst Strength
  • Multiple Power Injections
  • Catheter Length
  • Dimensional Verification (including ID, OD, Length)
  • Catheter Kink/Flex Resistance
  • . Tensile Testing (of Catheter and Assembly)
  • Particulate testing
  • Shelf life testing
  • Sterility testing
  • Packaging distribution testing
  • MR Compatibility testing

Biocompatibility Testing

Risk assessment evaluation that the proposed modifications did not impact biocompatibility

Conclusion

Based on the performance testing, intended use, materials, design, and components the HydroPICC-142 is determined to be substantially equivalent to the predicate HydroPICC-141 (K172885).

§ 880.5970 Percutaneous, implanted, long-term intravascular catheter.

(a)
Identification. A percutaneous, implanted, long-term intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings, such as luer hubs, and accessories that facilitate the placement of the device. The device allows for repeated access to the vascular system for long-term use of 30 days or more, and it is intended for administration of fluids, medications, and nutrients; the sampling of blood; and monitoring blood pressure and temperature. The device may be constructed of metal, rubber, plastic, composite materials, or any combination of these materials and may be of single or multiple lumen design.(b)
Classification. Class II (special controls) Guidance Document: “Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters.”