K Number
K192932
Device Name
FDR D-EVO III flat panel detector system
Date Cleared
2019-11-12

(26 days)

Product Code
Regulation Number
892.1680
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The Wireless/Wired FDR D-EVO III flat panel detector system is intended to capture for display radiographic images of human anatomy. It is intended for use in general projection radiographic applications including pediatric and neonatal exams wherever conventional film/screen or CR systems may be used. The FDR D-EVO III is not intended for mammography, fluoroscopy, tomography, and angiography applications.
Device Description
The subject device FDR D-EVO III Flat Panel Detector System (DR-ID1800) is a portable digital detector system that interfaces with, and acquires and digitizes x-ray exposures from, standard radiographic systems. DR-ID1800 is designed to be used in any environment that would typically use a radiographic cassette for examinations of adults, pediatrics and neonates. The detector models support both wireless and wired/tethered data communication between the detector and the system. Detectors can be placed in a wall bucky for upright exams, a table bucky for recumbent exams, or removed from the bucky for non-grid or free cassette exams. The software that supports the functions of the FDR D-EVO III Flat Panel Detector System is unchanged from the predicate device, cleared under K142003.
More Information

No
The summary explicitly states that the software is unchanged from the predicate device and there are no mentions of AI, DNN, or ML. The performance studies focus on non-clinical testing and equivalence to the predicate, without describing any AI/ML specific evaluations.

No
This device is intended to capture and display radiographic images, which is diagnostic, not therapeutic. It does not provide any treatment or therapy.

No

This device captures radiographic images for display, which are then interpreted by a human for diagnosis. The device itself does not perform any diagnostic analysis or interpretation of the images.

No

The device description explicitly states it is a "portable digital detector system" and a "Flat Panel Detector System," which are hardware components. While it includes software, it is not solely software.

No, this device is not an IVD (In Vitro Diagnostic).

Here's why:

  • IVD Definition: In Vitro Diagnostics are medical devices used to perform tests on samples taken from the human body, such as blood, urine, or tissue, to detect diseases, conditions, or infections.
  • Device Function: The description clearly states that the FDR D-EVO III flat panel detector system is intended to "capture for display radiographic images of human anatomy." This is an imaging device that uses X-rays to create images of the inside of the body.
  • No Sample Analysis: The device does not analyze any biological samples. It directly captures images from the patient's body.

Therefore, based on the provided information, the FDR D-EVO III is a medical imaging device, not an In Vitro Diagnostic device.

N/A

Intended Use / Indications for Use

The Wireless/Wired FDR D-EVO III flat panel detector system is intended to capture for display radiographic images of human anatomy. It is intended for use in general projection radiographic applications including pediatric and neonatal exams wherever conventional film/screen or CR systems may be used. The FDR D-EVO III is not intended for mammography, fluoroscopy, tomography, and angiography applications.

Product codes

MQB

Device Description

The subject device FDR D-EVO III Flat Panel Detector System (DR-ID1800) is a portable digital detector system that interfaces with, and acquires and digitizes x-ray exposures from, standard radiographic systems. DR-ID1800 is designed to be used in any environment that would typically use a radiographic cassette for examinations of adults, pediatrics and neonates. The detector models support both wireless and wired/tethered data communication between the detector and the system. Detectors can be placed in a wall bucky for upright exams, a table bucky for recumbent exams, or removed from the bucky for non-grid or free cassette exams. The software that supports the functions of the FDR D-EVO III Flat Panel Detector System is unchanged from the predicate device, cleared under K142003.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

X-ray

Anatomical Site

human anatomy

Indicated Patient Age Range

pediatric and neonatal exams

Intended User / Care Setting

Not Found

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Non-clinical Performance Data: FDR D-EVO III Flat Panel Detector System (DR-ID1800) conforms to the voluntary standards such as AAMI/ANSI ES60601-1, IEC 60601-1-2, IEC 62304, IEC 62366-1, IEC 62494-1 and DICOM. In addition, the FDA's Guidance for the Submission of 510(k)'s for Solid State X-ray Imaging Devices (September 1, 2016) was followed to describe the detector characteristics, and Radio Frequency Wireless Technology in Medical Devices (August 14, 2013) was followed to test wireless features. As required by the risk analysis, necessary verification and validation activities were performed including software testing, and the results were satisfactory. Furthermore, the image quality evaluation confirmed that the image quality of the FDR D-EVO III Flat Panel Detector System (DR-ID1800) is substantially equivalent to that of the predicate device.

Clinical Performance Data: No clinical study has been performed. The substantial equivalence has been demonstrated by non-clinical studies.

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s)

K142003

Reference Device(s)

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information

Not Found

§ 892.1680 Stationary x-ray system.

(a)
Identification. A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

0

November 12, 2019

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo includes the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG' in blue, and 'ADMINISTRATION' in a smaller font size below.

FUJIFILM Corporation % Ms. Kamila Sak Regulatory Affairs Specialist FUJIFILM Medical Systems U.S.A, Inc. 81 Hartwell Avenue, Suite 300 LEXINGTON MA 02421

Re: K192932

Trade/Device Name: FDR D-EVO III flat panel detector system Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: Class II Product Code: MQB Dated: October 16, 2019 Received: October 17, 2019

Dear Ms. Sak:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-

1

Page 2

combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

2

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known)

K192932

Device Name

FDR D-EVO III flat panel detector system

Indications for Use (Describe)

The Wireless/Wired FDR D-EVO III flat panel detector system is intended to capture for display radiographic images of human anatomy. It is intended for use in general projection radiographic applications including pediatric and neonatal exams wherever conventional film/screen or CR systems may be used. The FDR D-EVO III is not intended for mammography, fluoroscopy, tomography, and angiography applications.

Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Food and Druq Administration
Office of Chief Information Officer
PRAStaff(@fda.hhs.govDepartment of Health and Human Services
Paperwork Reduction Act (PRA) Staff
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."

==============================================================================================================================================================================

3

510(k) Summary K192932

Date Prepared: November 05, 2019

Submitter's Information:

FUJIFILM Corporation 798 Miyanodai Kaisei-Machi Ashigarakami-Gun, Kanagawa, 258-8538, Japan FDA Establishment Registration Number: 3001722928

Contact Person:

Kamila Sak Specialist, Regulatory Affairs Telephone: (347) 577-2309 Email: kamila.sak@fujifilm.com

Device Name and Classification:

Product Name:FUJIFILM FDR D-EVO III Flat Panel Detector System
Model Number:DR-ID 1800
Classification Name:Stationary x-ray system
Classification Panel:Radiology
CFR Section:21 CFR 892.1680
Device Class:Class II
Product Code:MQB

Predicate Device:

| Product Name | FUJIFILM FDR D-EVO II Flat Panel Detector System
(DR-ID 1200) |
|----------------------|------------------------------------------------------------------|
| 510(k) Number | K142003 |
| Classification Name | Stationary x-ray system |
| Classification Panel | Radiology |
| CFR Section | 21 CFR 892.1680 |
| Device Class | Class II |
| Product Code | MQB |

The subject device FDR D-EVO III Flat Panel Detector System (DR-ID1800) is essentially a modified version of legally marketed FDR D-EVO II Flat Panel Detector System (DR-

4

ID1200). The predicate device DR-ID 1200 had received 510(k) clearance via K142003 on October 21, 2014, and was documented internally several times after that.

Indications for Use:

The Wireless/Wired FDR D-EVO III flat panel detector system is intended to capture for display radiographic images of human anatomy. It is intended for use in general projection radiographic applications including pediatric and neonatal exams wherever conventional film/screen or CR systems may be used. The FDR D-EVO III is not intended for mammography, fluoroscopy, tomography, and angiography applications.

Description of the Device:

The subject device FDR D-EVO III Flat Panel Detector System (DR-ID1800) is a portable digital detector system that interfaces with, and acquires and digitizes x-ray exposures from, standard radiographic systems. DR-ID1800 is designed to be used in any environment that would typically use a radiographic cassette for examinations of adults, pediatrics and neonates. The detector models support both wireless and wired/tethered data communication between the detector and the system. Detectors can be placed in a wall bucky for upright exams, a table bucky for recumbent exams, or removed from the bucky for non-grid or free cassette exams. The software that supports the functions of the FDR D-EVO III Flat Panel Detector System is unchanged from the predicate device, cleared under K142003.

Comparison of Technological Characteristics:

A comparison of the technological characteristics between the subject device and predicate device is provided below:

5

| Comparison Item | Subject Device
FDR D-EVO III FPD system (DR-ID1800) | Predicate Device
FDR D-EVO II FPD system (DR-ID 1200) | Discussion
&
Conclusion |
|----------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for
Use | The Wireless/Wired FDR D-EVO III flat panel
detector system is intended to capture for display
radiographic images of human anatomy. It is
intended for use in general projection
radiographic applications including pediatric and
neonatal exams wherever conventional
film/screen or CR systems may be used. The
FDR D-EVO III is not intended for
mammography, fluoroscopy, tomography, and
angiography applications. | The Wireless/Wired FDR D-EVO II flat panel detector
system is intended to capture for display radiographic
images of human anatomy. It is intended for use in
general projection radiographic applications including
pediatric and neonatal exams wherever conventional
film /screen or CR systems may be used. The FDR D-
EVO II is not intended for mammography, fluoroscopy,
tomography, and angiography applications. | Same(Except device name) |
| Detector characteristics | | | |
| Panel variation | DR-ID 1801SE (14" x 17")
DR-ID 1802SE (17" x 17") | DR-ID 1201SE (14" x 17"),DR-ID 1202SE (17" x 17")
DR-ID 1211SE (14" x 17"),DR-ID 1212SE (17" x 17")
DR-ID 1213SE (9" x 11") | Described as below |
| Scintillator | GOS (Gadolinium oxysulfide) | GOS (Gadolinium oxysulfide);
DR-ID 1201SE (14" x 17")
DR-ID 1202SE (17" x 17")
CsI(Secium Iodide);
DR-ID 1211SE (14" x 17")
DR-ID 1212SE (17" x 17")
DR-ID 1213SE (9" x 11") | The subject device has only
GOS based detector.
No impact on safety or
effectiveness of the device. |
| Dimensions
(Detector exterior) | DR-ID1801SE
: 38.4 cm x 46 cm x 1.5 cm
DR-ID1802SE
: 46 cm x 46 cm x 1.5 cm | DR-ID1201SE, DR-ID1211SE
: 38.4 cm x 46 cm x 1.5 cm
DR-ID1202SE, DR-ID1212SE
: 46 cm x 46 cm x 1.5 cm
DR-ID1213SE
: 26.8 cm x 32.8 cm x 1.5 cm | Same* |
| Number of Pixels | DR-ID1801SE: 2336 x 2836
DR-ID1802SE: 2832 x 2836 | DR-ID1201SE, DR-ID1211SE
: 2336 x 2836
DR-ID1202SE, DR-ID1212SE
: 2832 x 2836
DR-ID1213SE
:1920 x 1536 | Same* |
| Pixel Size | 150 μm | 150 μm | Same |
| Comparison Item | Subject Device
FDR D-EVO III FPD system (DR-ID1800) | Predicate Device
FDR D-EVO II FPD system (DR-ID 1200) | Discussion
&
Conclusion |
| X-ray Conversion | Indirect conversion (a-Si) | Indirect conversion (a-Si) | Same |
| TFT sensor
substrate | Film-based TFT substrate | Glass-based TFT substrate | Only the material of
substrate is changed, there
is no change in the
mechanism. No impact on
safety or effectiveness of
the device. |
| DQE (RQA5, 1
lp/mm, 1mR) -
detector alone,
without tabletop | 33%
Measurement tolerance (±10%) | 31%
Measurement tolerance (±10%) | Slightly improved DQE by
using Film-based TFT
substrate
This difference of DQE
does not impact clinical
Image Quality. |
| MTF (RQA5, 1
lp/mm) | Two options are available by setting.
-Normal 60%
-High 75% | Two options are available by setting.
-Normal 60%
-High 75% | Same |
| Appearance
characteristics | - Adopt more rounded shape

  • Changed alighnment of LEDs/Operation button
    that is easier to use | - Rounded corners and curved edges are adopted.
  • LEDs/Operation button is placed. | This difference of detector
    appearance does not
    impact on safety or
    effectiveness of the device. |
    | Detector Weight | DR-ID1801SE:
    : Approx. 1.8kg (without battery)
    DR-ID1802SE
    : Approx. 2.1kg (without battery) | DR-ID1201SE
    : Approx. 2.3kg (without battery)
    DR-ID1202SE
    : Approx. 2.9kg (without battery)
    DR-ID1211SE
    : Approx. 2.4kg (without battery)
    DR-ID1212SE
    : Approx. 3.0kg (without battery)
    DR-ID1213SE
    : Approx. 1.3kg (without battery) | This difference of detector
    weight does not impact on
    safety or effectiveness of
    the device. |
    | Comparison Item | Subject Device
    FDR D-EVO III FPD system (DR-ID1800) | Predicate Device
    FDR D-EVO II FPD system (DR-ID 1200) | Discussion &
    Conclusion |
    | Software Charactisctic | | | |
    | Workstation | FDX Console Version 13.0 | FDX Console Version 8.0 and above (Latest version:
    12.1) | This difference of the
    workstation is only to add
    this subject detector as the
    connectable Flat panel
    detector options. No impact
    on safety or effectiveness of
    the device. |

6

7

*Note: Csl scintillator has not been adopted in DR-ID 1800 yet. Then, we discuss and conclude the substantial equivalence by excluding Csl scintillator.

8

Substantial Equivalence:

Both the subject device and predicate device have the same Indication for Use. There are some differences, but the differences are minor and do not affect fundamental scientific technology, principles of operation, safety and effectiveness, and image quality. Therefore, the subject device FDR D-EVO III Flat Panel Detector System (DR-ID1800) can be considered to be substantially equivalent to the predicate device FDR D-EVO II Flat Panel Detector System (DR-ID1200) limited to the flat panel detector with GOS scintillator.

Summary Of Studies:

Non-clinical Performance Data: FDR D-EVO III Flat Panel Detector System (DR-ID1800) conforms to the voluntary standards such as AAMI/ANSI ES60601-1, IEC 60601-1-2, IEC 62304, IEC 62366-1, IEC 62494-1 and DICOM. In addition, the FDA's Guidance for the Submission of 510(k)'s for Solid State X-ray Imaging Devices (September 1, 2016) was followed to describe the detector characteristics, and Radio Frequency Wireless Technology in Medical Devices (August 14, 2013) was followed to test wireless features. As required by the risk analysis, necessary verification and validation activities were performed including software testing, and the results were satisfactory. Furthermore, the image quality evaluation confirmed that the image quality of the FDR D-EVO III Flat Panel Detector System (DR-ID1800) is substantially equivalent to that of the predicate device.

Clinical Performance Data: No clinical study has been performed. The substantial equivalence has been demonstrated by non-clinical studies.

Conclusion:

This Special 510(k) premarket notification submission has demonstrated Substantial Equivalence as defined and understood in the Federal Food Drug and Cosmetic Act and various guidance documents issued by the Center for Devices and Radiological Health. We conclude the subject device to be as safe and effective as the predicate device.