(56 days)
For in vitro diagnostic use in the quantitative determination of cortisol in serum, plasma (EDTA and lithium heparin), and urine using the ADVIA Centaur® XP system.
Measurements of cortisol are used in the diagnosis and treatment of disorders of the adrenal gland.
The ADVIA Centaur Cortisol (COR) assay is a competitive immunoassay using direct chemiluminescent technology. Results are determined using a calibration curve that is generated specifically on each instrument by a 2-point calibration and a master curve with the reagent bar code. The ADVIA Centaur Cortisol (COR) assay is intended for use on the ADVIA Centaur family of analyzers. The ADVIA Centaur Calibrator E is a set of 2 level calibrators for the assay. Siemens recommends the use of commercially available quality control materials with at least two levels (low and high).
This document is focused on the ADVIA Centaur Cortisol (COR) assay, specifically the addition of plasma (EDTA and lithium heparin) as a sample claim. It seeks to demonstrate substantial equivalence to an existing device (K142723) which already had claims for serum and urine.
Acceptance Criteria and Reported Device Performance
The core of the study is to prove that the performance of the assay with the new plasma sample types is equivalent to its performance with serum (the established sample type). The primary acceptance criteria for this type of submission involve showing a strong correlation and minimal bias between the new sample type and the established one.
Table of Acceptance Criteria and Reported Device Performance:
| Performance Metric | Acceptance Criteria (Implicit from Industry Standards like CLSI EP09-A3) | Reported Device Performance (ADVIA Centaur Cortisol (COR) with new plasma claims vs. Serum) |
|---|---|---|
| Correlation Coefficient (r) | Typically, a correlation coefficient (r) close to 1.00 (e.g., >0.975 or >0.98 is often considered good for method comparison studies) indicating a strong linear relationship between the new and established methods. | Dipotassium EDTA Plasma vs. Serum: 1.00 Lithium-Heparin Plasma vs. Serum: 1.00 |
| Slope (from Deming Regression) | A slope close to 1.00 (e.g., 0.95 to 1.05) indicating proportional agreement between the new and established methods. | Dipotassium EDTA Plasma vs. Serum: 0.95 Lithium-Heparin Plasma vs. Serum: 0.96 |
| Intercept (from Deming Regression) | An intercept close to 0 (e.g., within a predefined range that is considered clinically insignificant) indicating fixed bias between the new and established methods. | Dipotassium EDTA Plasma vs. Serum: 0.24 µg/dL Lithium-Heparin Plasma vs. Serum: 0.56 µg/dL |
| Bias from Interferents | Bias should be within acceptable limits for clinical significance (e.g., < ±10% or clinically insignificant change). While not explicitly stated as an acceptance criterion in terms of a percentage, the results are presented to demonstrate minimal impact, aligning with good laboratory practice and CLSI EP07-ed3. | Dipotassium EDTA (9.0 mg/mL): At 12.94 µg/dL: 0.5% bias At 50.39 µg/dL: -1.1% bias Heparin (75 U/mL): At 7.85 µg/dL: 2.9% bias At 46.50 µg/dL: -0.1% bias |
Study Details:
-
Sample sizes used for the test set and data provenance:
- Dipotassium EDTA Plasma vs. Serum: N = 83 samples
- Lithium-Heparin Plasma vs. Serum: N = 99 samples
- Data Provenance: The document does not explicitly state the country of origin or whether the data was retrospective or prospective. However, given that it's a 510(k) submission for commercial use, it's highly probable these were prospective studies conducted in a controlled environment, likely in the US or a country with similar regulatory standards for medical device development.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- This submission is for an in vitro diagnostic (IVD) quantitative assay, not an AI/imaging device requiring expert interpretation for ground truth.
- The "ground truth" for this type of device is established by the measurement itself using a reference method or the established method (serum measurement in this case). Therefore, there were no "experts" in the sense of radiologists or pathologists establishing subjective ground truth on the test set. The comparison is between different sample types on the same device.
-
Adjudication method for the test set:
- Not applicable. This is a quantitative assay comparison, not subject to human adjudication methods like consensus reading for imaging. The comparison is statistical (Deming regression, bias calculation) between objective numerical results.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not applicable. This is not an AI/imaging device. It's a chemistry assay for quantitative determination of cortisol.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Not applicable. This is a fully automated in vitro diagnostic assay. While the instrument performs the measurement "stand-alone" in terms of algorithm/chemistry, it's not an AI diagnostic algorithm in the context of imaging or clinical decision support where such a distinction is typically made. The performance presented is the "standalone" or instrument performance.
-
The type of ground truth used:
- The "ground truth" here is the measurement obtained from the established sample type (serum) using the same ADVIA Centaur Cortisol (COR) assay. The study aims to show equivalence of the new sample types (plasma) to this established serum measurement. This is a form of comparative measurement validation.
-
The sample size for the training set:
- Not applicable in the AI/machine learning sense. This is a traditional IVD assay based on competitive immunoassay and chemiluminescent technology. There is no "training set" in the context of deep learning models. The assay's chemical and optical parameters are inherently "tuned" during its development, but not using a machine learning training dataset.
-
How the ground truth for the training set was established:
- Not applicable due to the nature of the device (traditional IVD assay). The development and calibration of the assay involve rigorous internal validation using well-characterized samples and reference methods, but it's not a "training set" with "ground truth" established by human experts in the ML sense. The calibration curve is generated on each instrument via a 2-point calibration and a master curve. The master curve materials have 7 levels of cortisol, likely used for establishing the assay's quantitative response profile.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square, the words "U.S. FOOD & DRUG" are written in blue, with the word "ADMINISTRATION" written in a smaller font size below.
November 25, 2019
Siemens Healthcare Diagnostics, Inc. Ian Thompson Regulatory Affairs Specialist 511 Benedict Avenue Tarrytown, NY 10591
Re: K192788
Trade/Device Name: ADVIA Centaur Cortisol (COR) Regulation Number: 21 CFR 862.1205 Regulation Name: Cortisol (Hydrocortisone and Hydroxycorticosterone) Test System Regulatory Class: Class II Product Code: JFT Dated: September 26, 2019 Received: September 30, 2019
Dear Ian Thompson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Marianela Perez-Torres, Ph.D. Acting Deputy Director Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K192788
Device Name ADVIA Centaur Cortisol (COR)
Indications for Use (Describe)
For in vitro diagnostic use in the quantitative determination of cortisol in serum, plasma (EDTA and lithium heparin), and urine using the ADVIA Centaur® XP system.
Measurements of cortisol are used in the diagnosis and treatment of disorders of the adrenal gland.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
|---|---|
| ------------------------------------------------------------ | ----------------------------------------------------------- |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
This 510(k) Summary of Safety and Effectiveness is being submitted in accordance with the requirements of 21 CFR 807.92 and the Safe Medical Device Act of 1990.
The assigned 510(k) Number is: K192788
Date Prepared 1.
September 26, 2019
Applicant Information 2.
- Contact: lan Thompson Regulatory Affairs Specialist
| Address: | Siemens Healthcare Diagnostics Inc.511 Benedict AvenueTarrytown, NY 10591 |
|---|---|
| Phone: | 914-524-2296 |
Fax: 914-524-2101
Email ian thompson@siemens-healthineers.com
Regulatory Information 3.
Table 1. Regulatory Information for ADVIA Centaur® Cortisol (COR)
| Trade Name | ADVIA Centaur® Cortisol (COR) |
|---|---|
| Device | Fluorometric, Cortisol |
| Regulation Description | Cortisol (hydrocortisone and hydroxycorticosterone) test system |
| FDA Classification | Class II |
| Review Panel | Clinical Chemistry (75) |
| Product Code | JFT |
| Regulation Number | 21 CFR 862.1205 |
4. Predicate Device Information
Predicate Device Name: ADVIA Centaur® Cortisol (COR)
510(k) Number: K142723
The ADVIA Centaur Cortisol (COR) assay with the plasma (EDTA and lithium heparin) sample claim is substantially equivalent to the ADVIA Centaur Cortisol (COR) assay that was cleared under 510(k) K142723, as shown below in the Substantial Equivalence Information section.
5. Intended Use / Indications for Use
For in vitro diagnostic use in the quantitative determination of cortisol in serum, plasma (EDTA and lithium heparin), and urine using the ADVIA Centaur® XP system.
Measurements of cortisol are used in the diagnosis and treatment of disorders of the adrenal gland.
Special Conditions for Use Statement(s): For prescription use only
{4}------------------------------------------------
Device Description 6.
The ADVIA Centaur Cortisol (COR) assay is a competitive immunoassay using direct chemiluminescent technology. Results are determined using a calibration curve that is generated specifically on each instrument by a 2-point calibration and a master curve with the reagent bar code. The ADVIA Centaur Cortisol (COR) assay is intended for use on the ADVIA Centaur family of analyzers. The ADVIA Centaur Calibrator E is a set of 2 level calibrators for the assay. Siemens recommends the use of commercially available quality control materials with at least two levels (low and high).
The ADVIA Centaur COR reagent kit contains the following:
-
ADVIA Centaur ReadyPack® primary reagent pack contains Lite Reagent and Solid Phase ● Reagent
Materials Required but Not provided -
ADVIA Centaur Calibrator E: consists of 2 levels (low and high) of multi-analyte calibrators: ● lyophilized human plasma spiked with analytes (cortisol, progesterone, and testosterone), sodium azide (0.1%) and preservatives.
Optional Reagents -
ADVIA Centaur Multi-Diluent 3 is a human plasma solution with sodium azide (0.1%) ●
-
ADVIA Centaur COR Master Curve Material is a set of 7 levels of cortisol (MCM1-7) spiked . in lyophilized human plasma and sodium azide (0.1%).
-
Cortisol Urine Reconstitution Buffer is a protein buffer solution with sodium azide (0.1%). ●
7. Purpose of the Submission
The purpose of this submission is for the addition of plasma (EDTA and lithium heparin) sample claim for the ADVIA Centaur Cortisol (COR) assay.
Substantial Equivalence Information - Comparison of Candidate 8. Device and Predicate Device
The following table demonstrates substantial equivalence between the ADVIA Centaur Cortisol (COR) assay (Candidate Device) that has modified Instructions for Use (Package Inserts) with the addition of the plasma (EDTA and lithium) sample claim and the currently marketed ADVIA Centaur Cortisol (COR) assay (Predicate Device) that was cleared under 510(k) K142723.
| Trade Name | Candidate Device | Predicate Device |
|---|---|---|
| ADVIA Centaur Cortisol (COR)(Addition of Plasma Claim) | ADVIA Centaur Cortisol (COR)(Unmodified Labeling) | |
| Intended Use | For in vitro diagnostic use in the quantitative determination of cortisol in serum, plasma (EDTA and lithium heparin), and urine using the ADVIA Centaur® XP system. | For in vitro diagnostic use in the quantitative determination of cortisol in serum or urine using the ADVIA Centaur® XP system. |
| Indications for Use | Measurements of cortisol are used in the diagnosis and treatment of disorders of the adrenal gland. | Same |
| Measurement | Quantitative | Same |
| Assay Range | Serum and Plasma: 0.50–75 µg/dLUrine: 0.50–53 µg/dL | Serum: 0.50-75 µg/dLUrine: 0.50-53 µg/dL |
| Operating Principle | Competitive immunoassay | Same |
| Technology | Direct chemiluminescent | Same |
{5}------------------------------------------------
510(k) Summary of Safety and Effectiveness
| Sample Type | Serum, Plasma (EDTA and lithiumheparin), Urine | Serum, Urine |
|---|---|---|
| Sample Volume | 20 µL (serum and plasma) | 20 µL (serum) |
| Traceability/Standardization | Internal Standards traceable toGCMS | Same |
| Calibration | 2-point | Same |
| Calibrator/Levels | Calibrator E/2 levels | Same |
| Controls/Levels | Commercial Controls/3 levels | Same |
| Master CurveMaterials | Seven levels (MCM1–7) | Same |
9. Standard/Guidance Document References
The following recognized standards from Clinical Laboratory Standards Institute (CLSI) were used as a basis of the study procedures described in this submission:
- Method Comparison and Bias Estimation Using Patient Samples; Approved Guideline— ● Third Edition (CLSI EP09-A3).
- Interference Testing in Clinical Chemistry; Approved Guideline-Third Edition (CLSI EP07ed3).
Test Principle 10.
The ADVIA Centaur Cortisol (COR) assav is a competitive immunoassay using direct chemiluminescent technology. Cortisol in the patient sample competes with acridinium ester labeled cortisol in the Lite Reagent for binding to polyclonal rabbit anti-cortisol antibody in the Solid Phase. The polyclonal rabbit antibody is bound to monoclonal mouse anti-rabbit antibody, which is covalently coupled to paramagnetic particles in the Solid Phase.
11. Performance Characteristics
The inclusion of the plasma (EDTA plasma and lithium heparin) sample claim in the Instructions for Use (Package Inserts) for the ADVIA Centaur Cortisol (COR) assay was demonstrated by testing the performance characteristics with the following studies:
- Specimen Equivalence by Method Comparison
- . Interferences: EDTA and Heparin
The plasma (EDTA plasma and lithium heparin) sample claim for the ADVIA Centaur Cortisol (COR) assay does not require the collection of additional analytical performance data. Therefore, all analytical performance data previously reviewed for the ADVIA Centaur Cortisol (COR) assay continues to apply to this assay, because the assay was not modified. These performance data are cross-referenced to the 510(k) submission for the plasma claim for the ADVIA Centaur Cortisol assay (K142723).
Specifically, the following studies are not needed for the purpose of this submission:
- Detection Capability (LoB, LoD, LoQ) ●
- Linearity/Assay Range ●
- Precision ●
- Dilution Recovery ●
- Method Comparison with Predicate Device
- Expected Values (Reference Intervals) ●
- Calibrator/Assay Traceability
{6}------------------------------------------------
- Reagent On-Board Stability and Calibration Interval ●
- Cross reactivity ●
- Interfering substances ●
- Shelf Life Stability .
11.1 Specimen Equivalence by Method Comparison
Specimen equivalency was determined with the Deming linear regression model in accordance with CLSI Document EP09-A3 for testing completed using the ADVIA Centaur Cortisol (COR) assay.
| Comparison | N* | Sample Interval | Slope | Intercept | CorrelationCoefficient (r) |
|---|---|---|---|---|---|
| Dipotassium EDTA Plasma vs. Serum | 83 | 0.29-67.06 μg/dL | 0.95 | 0.24 μg/dL | 1.00 |
| Lithium-Heparin Plasma vs. Serum | 99 | 0.29-67.06 μg/dL | 0.96 | 0.56 μg/dL | 1.00 |
- N = Number of samples tested.
11.2 Interferences: EDTA and Heparin
Interference testing for EDTA and heparin was performed in accordance with CLSI guideline EP07-ed3 using the ADVIA Centaur Cortisol (COR) assay. The following results were obtained:
| Interferent | InterferentConcentration | Analyte Concentration(ug/dL) | Bias (%) |
|---|---|---|---|
| Dipotassium EDTA | 9.0 mg/mL | 12.94 | 0.5 |
| 50.39 | -1.1 | ||
| Heparin | 75 U/mL | 7.85 | 2.9 |
| 46.50 | -0.1 |
11.3 Clinical Studies
Not applicable.
11.4 Clinical Cut-off
Not applicable.
Conclusions 12.
The ADVIA Centaur Cortisol (COR) assay with the addition of the plasma (EDTA plasma and lithium heparin) sample claim in the Instructions for Use (package insert) is substantially equivalent to the currently marketed ADVIA Centaur Cortisol (COR) assay (K142723).
§ 862.1205 Cortisol (hydrocortisone and hydroxycorticosterone) test system.
(a)
Identification. A cortisol (hydrocortisone and hydroxycorticosterone) test system is a device intended to measure the cortisol hormones secreted by the adrenal gland in plasma and urine. Measurements of cortisol are used in the diagnosis and treatment of disorders of the adrenal gland.(b)
Classification. Class II.