K Number
K192785
Date Cleared
2020-02-05

(128 days)

Product Code
Regulation Number
880.5725
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The OptiScanner® 5000 Glucose Monitoring System is an automated, bedside glucose monitoring device indicated for detecting trends and tracking patterns in persons (age 18 and older) in the surgical intensive care unit. The system collects a venous whole blood sample via connection to a central venous catheter, centrifuges the sample, and measures the plasma glucose concentration. It is not intended for the screening or diabetes mellitus but is indicated for use in determining dysglycemia. The OptiScanner® 5000 Glucose Monitoring System is for in vitro diagnostic use.

Device Description

The OptiScanner® 5000 Glucose Monitoring System ("OptiScanner") is an automated, bedside glucose monitoring device that quantitatively measures the concentration of glucose in the blood of patients in the Surgical Intensive Care Unit (SICU). In contrast to Point of Care (POC) glucose measuring devices that measure glucose using enzymatic techniques, the OptiScanner uses a direct, reagent-free, spectrophotometer method to quantify glucose. The system is comprised of the following three primary components:

  • OptiScanner Instrument
  • OptiScanner Transport Cart
  • OptiScanner Disposable Cartridge

The Instrument is the primary hardware component that houses all electrical, mechanical, analytical, and power subsystems. This includes the integrated pump, spectrometer, and the user interface. For mobility and easy access, the Instrument is mounted onto the chassis of a transport Cart. The Cart, in addition to holding the Instrument, holds batteries, IV pole(s) and bar code scanner. The Cartridge is a disposable, single patient use, sterile component containing the fluid pathway through which the blood is sampled, stored, processed, and analyzed. The Cartridge is the only component of the OptiScanner system that contacts patient blood. Integrated into the cartridge are tubing sets that are used to connect to the patient and to a saline bag. The Cartridge also includes a syringe that is intended to be pre-filled with heparin by the user for processing the blood samples. The Cartridge is inserted into the Instruments interface port that provides connections integrating the fluidic components of the Cartridge with the electro-mechanical subsystems of the Instrument.

AI/ML Overview

This is a 510(k) summary for the OptiScanner® 5000 Glucose Monitoring System (K192785). The submission is for an update to the device labeling (User Manual) to include Peripherally Inserted Central Catheters (PICC) compatibility and clarifying language regarding central venous access catheter port placement. The physical device is unchanged from its predicate (K162042). Therefore, the performance criteria and study results presented here refer to the performance of the predicate device, K162042, which remains the same for K192785.

Here's a breakdown of the acceptance criteria and study information:

1. Table of Acceptance Criteria (from the predicate device K162042) and Reported Device Performance (Table 1: Comparison of Subject Device to Predicate):

Device FeatureAcceptance Criteria (Predicate OptiScanner 5000 (K162042))Reported Device Performance (Subject Device OptiScanner 5000)Comments/Proof of Meeting Criteria
Measuring range40 - 400 mg/dL40 - 400 mg/dLUnchanged. The device retains the same measuring range as its predicate.
Hematocrit range15-60 %15-60 %Unchanged. The device supports the same hematocrit range as its predicate.
System accuracy (MARD)Not explicitly stated as an "acceptance criterion" but reported for predicate.MARD 7.28%Based on pivotal trial results with 160 SICU patients (2,804 matched OptiScanner / YSI pairs) from the predicate device (K162042). This value is considered acceptable for the device's intended use in the SICU for detecting trends and tracking patterns of glucose. This implies that the study results met the internal accuracy requirements for FDA clearance.
Compatibility with PICCs (new for K192785)PICCs should perform adequately for blood draw within acceptance criteria limits and be mechanically compatible.PICCs evaluated performed adequately for their ability to draw blood and were mechanically compatible.In vitro design verification tests were conducted, evaluating mechanical compatibility and blood draw performance relative to Central Venous Catheters (CVCs). This testing demonstrated that the PICCs can be used with the OptiScanner system in accordance with its intended use and in an equivalent manner to CVCs.

2. Sample Size Used for the Test Set and Data Provenance:

  • Sample Size for System Accuracy (Pivotal Trial from K162042): 160 SICU patients (resulting in 2,804 matched OptiScanner / YSI pairs).
  • Data Provenance: The document does not explicitly state the country of origin but implies a clinical study environment, likely in the US, given the FDA submission. The study for K162042 was a prospective clinical trial as it involved patients in a SICU.
  • Sample Size for PICC Compatibility (K192785): Not explicitly stated, but described as "in vitro design verification tests." This suggests a benchtop study with a set number of PICCs rather than a patient-based test set for this specific update.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

  • For the system accuracy reported as MARD 7.28%, the ground truth was established using the YSI analyzer. A YSI analyzer is a laboratory reference method, implying that the ground truth was established by certified laboratory personnel following standardized laboratory procedures. The number and specific qualifications of the technicians performing the YSI measurements are not provided in this summary.
  • For the PICC compatibility testing, the ground truth was based on pre-defined technical specifications and performance criteria for blood draw and mechanical compatibility, not expert consensus on patient data.

4. Adjudication Method for the Test Set:

  • For System Accuracy (K162042): Not applicable in the context of expert adjudication for diagnoses. The YSI analyzer served as the objective reference standard. Differences were quantified against this standard.
  • For PICC Compatibility (K192785): Not applicable as it involved in vitro design verification tests against predefined engineering and performance specifications.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:

  • No, a multi-reader multi-case (MRMC) comparative effectiveness study was not conducted for this device. The OptiScanner 5000 is an automated glucose monitoring system; it does not involve human readers interpreting results in a diagnostic imaging or similar context where MRMC studies are typically performed.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

  • Yes, this device is inherently a standalone algorithm/device performance study. The OptiScanner® 5000 Glucose Monitoring System is an automated device that measures plasma glucose concentration without direct human interpretation of the measurement result itself. Physicians use the numerical output from the device. The reported MARD value is a measure of the device's standalone accuracy against a reference method.

7. The Type of Ground Truth Used:

  • The primary ground truth for the glucose measurements (system accuracy) was laboratory reference standard (YSI analyzer) for the predicate device's pivotal trial. This is considered a gold standard for glucose measurement.
  • For the K192785 submission update regarding PICC compatibility, the ground truth was established by in vitro design verification tests against pre-defined performance specifications for blood draw and mechanical compatibility.

8. The Sample Size for the Training Set:

  • The document does not explicitly state the sample size for a training set. However, for a Spectrophotometric detection method, the "training" would typically involve developing and refining the algorithms/models that convert the spectral data into glucose concentrations. This development and internal validation would likely occur using a substantial amount of diverse blood samples, but these are generally separate from the formal clinical validation study (like the 160 SICU patients). The summary focuses on the clinical performance, not the internal development data.

9. How the Ground Truth for the Training Set Was Established:

  • Since the document does not specify a training set sample size, it does not detail how the ground truth for such a set was established. However, given the measurement technology (spectrophotometric), the ground truth for any development or internal validation would also be established by comparison to a highly accurate laboratory reference method (like YSI), similar to the method used for the clinical test set. This would ensure the Spectrophotometric model's accuracy against recognized standards.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

February 5, 2020

Optiscan Biomedical Corporation Charles Zimliki Vice President of Regulatory, Quality, and Clinical Affairs 24590 Clawiter Road Hayward, California 94545

Re: K192785

Trade/Device Name: OptiScanner® 5000 Glucose Monitoring System Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion Pump Regulatory Class: Class II Product Code: LZF, PYV Dated: January 2, 2020 Received: January 2, 2020

Dear Charles Zimliki:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Marianela Perez-Torres, Ph.D. Acting Deputy Director Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K192785

Device Name

OptiScanner® 5000 Glucose Monitoring System

Indications for Use (Describe)

The OptiScanner® 5000 Glucose Monitoring System is an automated, bedside glucose monitoring device indicated for detecting trends and tracking patterns in persons (age 18 and older) in the surgical intensive care unit. The system collects a venous whole blood sample via connection to a central venous catheter, centrifuges the sample, and measures the plasma glucose concentration. It is not intended for the screening or diabetes mellitus but is indicated for use in determining dysglycemia. The OptiScanner® 5000 Glucose Monitoring System is for in vitro diagnostic use.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

This K192785 510(k) Summary was prepared in accordance with 21 CFR 807.92.

Submitter 1.

Applicant:OptiScan Biomedical Corporation24590 Clawiter RoadHayward, CA 94545
OfficialCorrespondent:Charles ZimlikiVice President of Regulatory, Quality, and Clinical AffairsPhone: (240) 350-6690Email: czimliki@optiscancorp.com
Date Prepared:31 January 2020

2. Device Information

Trade Name:OptiScanner® 5000 Glucose Monitoring System
510(k) Number:K192785
Common Name:Automated, in-line, bedside, glucose monitoring system
Device Classification:Name: Infusion Pump
Regulation No.: 21 CFR 880.5725
Product Code: LZF – Pump, Infusion Analytic Sampling
Class: II
Name: Glucose Test System
Regulation No.: 21 CFR §862.1345
Product Code: PYV – Hospital Continuous Glucose Monitoring
System
Class: II

3. Predicate Device Information

Trade Name:OptiScanner® 5000 Glucose Monitoring System (K162042)
510(k) Number:K162042
Common Name:Automated, in-line, bedside, glucose monitoring system
Device Classification:Name: Infusion PumpRegulation No.: 21 CFR 880.5725Product Code: LZF – Pump, Infusion Analytic Sampling

{4}------------------------------------------------

Class: II
Name: Glucose Test SystemRegulation No.: 21 CFR §862.1345Product Code: PYV – Hospital Continuous Glucose MonitoringSystemClass: II

Device Description 4.

The main features of the OptiScanner® 5000 Glucose Monitoring System are described below. There have been no changes to the OptiScanner Instrument, Cart, or Cartridge. They are identical to those cleared in K162042.

The OptiScanner® 5000 Glucose Monitoring System ("OptiScanner") is an automated, bedside glucose monitoring device that quantitatively measures the concentration of glucose in the blood of patients in the Surgical Intensive Care Unit (SICU). In contrast to Point of Care (POC) glucose measuring devices that measure glucose using enzymatic techniques, the OptiScanner uses a direct, reagent-free, spectrophotometer method to quantify glucose. The system is comprised of the following three primary components:

  • OptiScanner Instrument
  • OptiScanner Transport Cart
  • OptiScanner Disposable Cartridge

The Instrument is the primary hardware component that houses all electrical, mechanical, analytical, and power subsystems. This includes the integrated pump, spectrometer, and the user interface. For mobility and easy access, the Instrument is mounted onto the chassis of a transport Cart. The Cart, in addition to holding the Instrument, holds batteries, IV pole(s) and bar code scanner. The Cartridge is a disposable, single patient use, sterile component containing the fluid pathway through which the blood is sampled, stored, processed, and analyzed. The Cartridge is the only component of the OptiScanner system that contacts patient blood. Integrated into the cartridge are tubing sets that are used to connect to the patient and to a saline bag. The Cartridge also includes a syringe that is intended to be pre-filled with heparin by the user for processing the blood samples. The Cartridge is inserted into the Instruments interface port that provides connections integrating the fluidic components of the Cartridge with the electro-mechanical subsystems of the Instrument.

5. Indications for Use

The OptiScanner® 5000 Glucose Monitoring System is an automated, bedside glucose monitoring device indicated for detecting trends and tracking patterns in persons (age 18 and older) in the surgical intensive care unit. The system collects a venous whole blood sample via connection to a central venous catheter, centrifuges the sample, and measures the plasma glucose

{5}------------------------------------------------

concentration. It is not intended for the screening or diagnosis of diabetes mellitus but is indicated for use in determining dysglycemia. The OptiScanner® 5000 Glucose Monitoring System is for in vitro diagnostic use.

Comparison of Technical Characteristics with the Predicate Device 6.

There are no differences in the technological characteristics between the proposed device and the predicate device (K162042). The purpose of this Traditional 510(k) is to seek an update to the device labeling (User Manual) to:

    1. Update the listing of central venous catheters in "IV Catheters" section that can be used with the OptiScanner 5000 when performing the patient blood draw
    1. Add clarification to "Catheter Placement" section regarding the description of the "proximal port" of the central venous catheter

The physical device remains unchanged from the predicate (K162042), with the only change being a revision to the device labeling. Table 1 provides a technological comparison between the proposed device and the predicate device.

{6}------------------------------------------------

Device FeatureSubject DeviceOptiScanner 5000Predicate DeviceOptiScanner 5000 (K162042)Comments
User Manual(OptiScanner Instrument)The OptiScanner® 5000 User's ManualThe OptiScanner® 5000 User's ManualMODIFIEDSubject of this 510(k) filing:Chapter 2:"IV Catheters" (including Figure 41)updated to include additional IVcatheters which have been successfullyevaluated for compatibility with theOptiScanner system."Catheter Placement" updated to addclarifying text.No other revisions have been made tothe User Manual
Instructions for Use(OptiScanner DisposableCartridge)Instructions for Use, OptiScanner® 5000Instructions for Use, OptiScanner® 5000Unchanged
Product Labeling(OptiScanner DisposableCartridge)● Disposable Cartridge, Tyvek/Box● Disposable Cartridge, Shipper● Disposable Cartridge, Tyvek/Box● Disposable Cartridge, ShipperUnchanged
Device FeatureSubject DeviceOptiScanner 5000Predicate DeviceOptiScanner 5000 (K162042)Comments
Indications for UseThe OptiScanner® 5000 GlucoseMonitoring System is an automated,bedside glucose monitoring deviceindicated for detecting trends andtracking patterns in persons (age 18 andolder) in the surgical intensive care unit.The system collects a venous wholeblood sample via connection to a centralvenous catheter, centrifuges the sample,and measures the plasma glucoseconcentration. It is not intended for thescreening or diagnosis of diabetesmellitus but is indicated for use indetermining dysglycemia. TheOptiScanner® 5000 Glucose MonitoringSystem is for in vitro diagnostic use.The OptiScanner® 5000 GlucoseMonitoring System is an automated,bedside glucose monitoring deviceindicated for detecting trends andtracking patterns in persons (age 18 andolder) in the surgical intensive care unit.The system collects a venous wholeblood sample via connection to a centralvenous catheter, centrifuges the sample,and measures the plasma glucoseconcentration. It is not intended for thescreening or diagnosis of diabetesmellitus but is indicated for use indetermining dysglycemia. TheOptiScanner® 5000 Glucose MonitoringSystem is for in vitro diagnostic use.Unchanged
Intended UseThe OptiScanner® 5000 GlucoseMonitoring System is intended for in-vitro diagnostic useThe OptiScanner® 5000 GlucoseMonitoring System is intended for in-vitro diagnostic useUnchanged
Target patient populationPatients in the surgical intensive careunitPatients in the surgical intensive careunitUnchanged
Key indication-modifyingstatement"…not intended for the screening ordiagnosis of diabetes mellitus but isindicated for use in determiningdysglycemia""…not intended for the screening ordiagnosis of diabetes mellitus but isindicated for use in determiningdysglycemia"Unchanged
Point of careIn-hospitalIn-hospitalUnchanged
Glucose quantificationmethodSpectrophotometric detection of glucoseSpectrophotometric detection of glucoseUnchanged
Single use disposables?Yes (sampling cartridge)Yes (sampling cartridge)Unchanged
Duration of single usedisposables72 hours72 hoursUnchanged
Device FeatureSubject DeviceOptiScanner 5000Predicate DeviceOptiScanner 5000 (K162042)Comments
Source of sampleVenous bloodVenous bloodUnchanged
Point of access for sampleCentral venous catheter (CVC),Peripherally Inserted Central Catheters(PICC) or multi-lumen access catheter(MAC)Central venous catheter (CVC) or multi-lumen access catheter (MAC)Added access by Peripherally InsertedCentral Catheters (PICC)
Invasive / indwellingcomponents?No. Connects to central venous catheter(CVC), peripherally inserted centralcatheters (PICC), or multi-lumen accesscatheter (MAC) line.No. Connects to central venous catheter(CVC) or multi-lumen access catheter(MAC) line.Added Peripherally Inserted CentralCatheters (PICC)
Sample acquisition methodBlood draw (3 mL) through connectionto venous access. 0.17 mL retained,heparinized and analyzed. Rest of bloodreturned to patient with 2.5 mL salineflush.Blood draw (3 mL) through connectionto venous access. 0.17 mL retained,heparinized and analyzed. Rest of bloodreturned to patient with 2.5 mL salineflush.Unchanged
Sample size required foranalysis0.17 mL0.17 mLUnchanged
Average fluid infused intopatient during monitoring(excluding blood return)10 mL saline / hour(assumes measurement every 15 min)10 mL saline / hour(assumes measurement every 15 min)Unchanged
Sampling frequency15 min15 minUnchanged
Measuring range40 - 400 mg/dL40 - 400 mg/dLUnchanged
Hematocrit range15-60 %15-60 %Unchanged
Requires use of heparinYes, but discarded with portion of bloodsample that is analyzed (0.17 mL).Heparin is not infused into patient.Yes, but discarded with portion of bloodsample that is analyzed (0.17 mL).Heparin is not infused into patient.Unchanged
Physician-determined highand low glucose values?YesYesUnchanged
Alarms
Device FeatureSubject DeviceOptiScanner 5000Predicate DeviceOptiScanner 5000 (K162042)Comments
High / low glucoseYesYesUnchanged
Air in line / occlusionYesYesUnchanged
CalibrationAt factory(None required under normal use)At factory(None required under normal use)Unchanged
System accuracyMARD 7.28% Based on pivotal trialresults with 160 SICU patients (2,804matched OptiScanner / YSI pairs).MARD 7.28% Based on pivotal trialresults with 160 SICU patients (2,804matched OptiScanner / YSI pairs).Unchanged
Temperature sensitiveNoNoUnchanged
Data storageYes(72 hours of data)Yes(72 hours of data)Unchanged
Back-up battery powerYes(3 hours of operation w/o AC power)Yes(3 hours of operation w/o AC power)Unchanged
Single patient usedisposablesYesYesUnchanged

Table 1: Comparison of Subject Device to Predicate

{7}------------------------------------------------

{8}------------------------------------------------

{9}------------------------------------------------

{10}------------------------------------------------

7. Performance Standards

No performance standards have been established by the Agency to date that apply to this device.

Summary of Non-Clinical Testing 8.

A comprehensive bench testing program was originally executed as reported in K162042 for the OptiScanner 5000 System to verify the function, performance, and safety of the device. This Traditional 510(k) seeks FDA clearance of revisions the product labeling (User Manual) to include Peripherally Inserted Central Catheters (PICC) for which compatibility has been demonstrated, and to add clarifying language to the User Manual regarding optimal placement of the central venous access catheter port. All other elements of the OptiScanner system are unchanged and as a result only a subset of the testing performed in the Original 510(k) filing was repeated to support the proposed change. The compatibility requirements of PICCs were assessed through in vitro design verification tests, evaluating the mechanical compatibility of the PICCs to the OptiScanner patient connector as well as blood draw performance relative to Central Venous Catheters (CVCs). Test results demonstrated that the PICCs evaluated performed adequately for their ability to draw blood within the acceptance criteria limits, and that the evaluated PICCs were mechanically compatible with the patient connector on the disposable cartridge. Overall, the test results demonstrate that PICCs can be used with the OptiScanner system in accordance with its intended use and in an equivalent manor to its use with CVCs.

Summary of Clinical Testing 9.

No clinical data were required for this submission.

10. Conclusion

The revised labeling that OptiScan is requesting for the proposed OptiScanner® 5000 Glucose Monitoring System is to update the listing of central catheters to which compatibility with the OptiScanner system has been established. Additionally, the labeling updates provide clarification regarding the description of the "proximal port" of the central catheter. The submitted information demonstrate substantial equivalence to the predicate.

§ 880.5725 Infusion pump.

(a)
Identification. An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.(b)
Classification. Class II (performance standards).