K Number
K192604

Validate with FDA (Live)

Date Cleared
2020-06-18

(272 days)

Product Code
Regulation Number
868.5895
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Servo-air ventilator system is intended for respiratory support, monitoring and treatment of pediatric and adult patients to be used only by healthcare providers to be used only in professional healthcare facilities and for transport within these facilities

Device Description

The Servo-air 4.0 Ventilator System consists of a Patient Unit where gases are mixed and administered, and a User Interface where the settings are made and ventilation is monitored. The Servo-air 4.0 Ventilator System is based on the cleared predicate device Servo-U/n 2.1 Ventilator System (K180098), with some modifications. The ventilation modes in the Servo-air 4.0 are the same as the predicate device. Standard configurations of available modes and optional modes do differ between the Servo-air, and the cleared predicate device Servo-U/n 2.1. The ventilator delivers controlled or supported breaths to the patient, with constant flow, constant pressure, using a set oxygen concentration. Servo-air contains a dedicated controller circuit for the Aerogen Solo nebulizer (included as standard). The Servo-air 4.0 Ventilator System will produce visual and audible alarms if any parameter varies beyond pre-set or default limits and log alarm recordings. The system contains provisions for battery modules to supply the system in the case of mains power failure or during intra-hospital transport.

AI/ML Overview

This document does not contain an AI/ML device or its associated performance study. It is a 510(k) premarket notification for a Servo-air 4.0 Ventilator System. The information provided describes the ventilator, its comparison to a predicate device (Servo-U/n 2.1 Ventilator System), and non-clinical testing performed to demonstrate substantial equivalence, focusing on engineering changes and performance specifications of a physical device.

Therefore, I cannot provide the requested information regarding acceptance criteria and the study that proves an AI/ML device meets acceptance criteria.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo. On the right is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

June 18, 2020

Maquet Critica Care AB % Mark Smith Sr. Regulatory Affairs Specialist Getinge 45 Barbour Pond Drive Wayne, New Jersey 07470

Re: K192604

Trade/Device Name: Servo-air 4.0 Ventilator System Regulation Number: 21 CFR 868.5895 Regulation Name: Continuous Ventilator Regulatory Class: Class II Product Code: CBK Dated: May 15, 2020 Received: May 18, 2020

Dear Mark Smith:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

James J. Lee Ph.D. Assistant Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K192604

Device Name Servo-air 4.0 Ventilator System

Indications for Use (Describe)

  • The Servo-air ventilator system is
  • · intended for respiratory support, monitoring and treatment of pediatric and adult patients
  • to be used only by healthcare providers
  • · to be used only in professional healthcare facilities and for transport within these facilities

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CER 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "GETINGE" in a sans-serif font. To the right of the word is a stylized four-pointed star or asterisk symbol. The text and symbol are in a dark blue color.

510(k) SUMMARY

as required by section 21 CFR 807.92

Device ownerMaquet Critical Care ABRöntgenvägen 2SE-171 54 Solna, SwedenTel: (011) 46 10 335 7300
Contact Persons for this submission:Mr. David ArdaniusRegulatory Affairs ManagerPhone: direct: (011) 46 10 335 7300Email: david.ardanius@getinge.com
Application Correspondent:Mr. Mark N. SmithManager, Regulatory AffairsGetinge45 Barbour Pond DriveWayne, NJ 07470Email: mark.n.smith@getinge.comPhone: 585-272-5274

Date prepared: September 13, 2019 Updated: May 7, 2020

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Getinge. The logo consists of the word "GETINGE" in a bold, sans-serif font. To the right of the wordmark is an abstract symbol that resembles a stylized asterisk or a four-way directional arrow, composed of curved shapes in varying shades of blue.

Trade Name: Servo-air 4.0 Ventilator System Model: Servo-air Model no: 68 82 000

Device Classification

Common Name Classification Number Class Regulation Number 510(k) Number

Ventilator, continuous, facility use CBK II 21 CFR 868.5895 K192604

Predicate Device Identification

Maquet Servo-U/n Ventilator System K180098

Reference Device Identification

NA

Indications for Use

The Servo-air ventilator system is:

  • intended for respiratory support, monitoring and treatment of pediatric and adult . patients
  • to be used only by healthcare providers ●
  • to be used only in professional healthcare facilities and for transport within these ● facilities

{5}------------------------------------------------

GETING

Device Description

The Servo-air 4.0 Ventilator System consists of a Patient Unit where gases are mixed and administered, and a User Interface where the settings are made and ventilation is monitored.

The Servo-air 4.0 Ventilator System is based on the cleared predicate device Servo-U/n 2.1 Ventilator System (K180098), with some modifications. The ventilation modes in the Servo-air 4.0 are the same as the predicate device. Standard configurations of available modes and optional modes do differ between the Servo-air, and the cleared predicate device Servo-U/n 2.1.

The ventilator delivers controlled or supported breaths to the patient, with constant flow, constant pressure, using a set oxygen concentration.

Servo-air contains a dedicated controller circuit for the Aerogen Solo nebulizer (included as standard).

The Servo-air 4.0 Ventilator System will produce visual and audible alarms if any parameter varies beyond pre-set or default limits and log alarm recordings.

The system contains provisions for battery modules to supply the system in the case of mains power failure or during intra-hospital transport.

PREDICATE DEVICESUBJECT DEVICEComment
510(k) NumberK180098K192604-
Device nameSERVO-U 2.1Servo-air 4.0-
ManufacturerMaquet Critical Care ABMaquet Critical Care ABIdentical
Device ClassificationCBKCBKIdentical
Indications for UseThe SERVO-U 2.1 ventilator system is:The SERVO-air 4.0 ventilator system is:Identical forincluded parts
intended for respiratory support, monitoring and treatment of neonatal, pediatric and adult patientsintended for respiratory support, monitoring and treatment of pediatric and adult patientsNeonatal patient category excluded for subject device
to be used only by healthcare providersto be used only by healthcare providers
to be used only in professional healthcare facilities and for transport within these facilitiesto be used only in professional healthcare facilities and for transport within these facilities
For NAVA and Edi monitoring, it is in addition intended:*to provide monitoring of the patient's breathing drive*to improve synchrony between theNo NAVA and Edi monitoring for subject device)

COMPARISSON TO PREDICATE

{6}------------------------------------------------

GETINGE

PREDICATE DEVICESUBJECT DEVICEComment
510(k) NumberK180098K192604-
Device nameSERVO-U 2.1Servo-air 4.0-
ventilator system and patient whenthe electrical signal from the brain tothe diaphragm is active*for use on all patients with nocontraindication forinsertion/exchange ofa nasogastric tube
Patient rangeNeonatal to adult patient populationsPediatric to adult patientpopulationsIdentical forincluded parts
Reduced;Neonatal patientcategory excludedfor subject device
Ventilation Modes• Volume Control (VC)• Pressure Control (PC)• Volume Support (VS)• Pressure Support (PS)• Pressure Support (CPAP)• Pressure Regulated VolumeControl (PRVC)• SIMV (SIMV(PC)+PS,SIMV(PRVC)+PS,SIMV(VC)+PS)• Automode (PC-PS, PRVC-VS,VC-VS)• Bi-Vent/APRV• NIV (NIV PC, NIV PS)• NAVA• NIV NAVA• Nasal CPAP• Volume Control (VC)• Pressure Control (PC)• Volume Support (VS)• Pressure Support (PS)• Pressure Support (CPAP)• Pressure Regulated VolumeControl (PRVC)• SIMV (SIMV(PC)+PS,SIMV(PRVC)+PS,SIMV(VC)+PS)• Automode (PC-PS, PRVC-VS,VC-VS)• Bi-Vent/APRV• NIV (NIV PC, NIV PS)Identical forincluded partsNAVA, NIVNAVA and NasalCPAP are notincluded in subjectdevice.
Breathing Therapies• High Flow therapy• High Flow therapyIdentical

Comparison of Intended Use/Indications Use

There are no changes to the Intended Use or Indications for Use between the subject device and the predicate device, with exception to the removal of NAVA which is not part of the indications for use for the subject device.

Comparison of Technology Characteristics

The Servo-air 4.0 is based on the cleared predicate device Servo-U/n 2.1 (K180098), with adaptions to be wall-gas independent, and modifications noted in the description below. Software algorithms for ventilation and alarms are the same. Furthermore, the ventilation modes in the Servo-air 4.0 are the same as the cleared predicate device.

The following changes have been made to Servo-air 4.0 compared to the cleared predicate device Servo-U/n 2.1 (K180098):

{7}------------------------------------------------

GETING

. Mechanics and attachment of the panel

Compared to the predicate device the panel has been changed from a 15" capacitive touch screen to a 12" resistive touch screen. The general layout, functionality and screen resolution (1024x768) remains the same compared to the subject device.

The panel is fixed to the device via hinges instead of being mounted to either the mobile cart, a table or railing, as for the predicate device. As part of the change of panel changed has also been done to the PCBs.

The alarm indicator on the panel of the subject device flashes with the same colors and frequency as the alarm indicator on the screen, compared to the predicate device where the lamp does not flash together with the alarm.

The subject device has different physical dimensions. The casing has been changed from cast aluminum to steel and plastic.

The carrier is based on the same system as the predicate device, but the shelf and the pillar attached to the base of the carrier has been modified to fit the dimensions of the subject device.

Software and user interface ●

Software updates have been made to accommodate the modifications to the system. The Servo-air software is considered a Major Level of Concern. Software verification was performed according to FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" and according to the standard IEC 62304 "Medical device software - Software life-cycle processes".

Components ●

Compared to the predicate device the Handle, Mounting kit for Compressor Mini on Mobile Cart, and Pendant/bed holder have been removed. The handle is now integrated in the design of the subject device which is also not intended to be bedside mounted and thus the Pendant/bed holder is removed. The Compressor Mini is not an accessory for the subject device so the mounting kit is excluded.

The following components have been changed; Support Arm, Mobile cart, Humidifier Holder and shelf base to accommodate the physical changes to the cart.

New components specific for the subject device include the battery module, battery charger, cabinet, air inlet filter and dust filter.

{8}------------------------------------------------

GETING

. Power system

The subject device maximum power consumption has increased from 80W to 100W and the power supply electronics have been modified to accommodate this.

The battery has been changed from NiMH to Li-ion batteries. However, the battery system backup time range for the subject device is within the range specified for the predicate device.

● Blower module

The Servo-air adds the use of a turbine/blower module with an ambient humidity and temperature sensor. The turbine generates a controlled airflow from the ambient air and is controlled by the blower module. An air flow meter measurement of the mixed (air+O2) gas flow is used.

As a result of these changes some performance specifications are also changed for the subject device.

1.The noise level has increased from <40dB (LpA) to <49dB (LpA) andfrom <51db (LWA) to <57dB (LWA), compared to the predicate device.These levels are still below the recommendations described in ISO 60601-1-8.
2.The air supply performance has been improved so that the ventilator cansupply 240 l/min.
3.The Maximum pressure setting has been reduced from 120 to 100 cmH2O.The maximum airway pressure is consequently also reduced from 125 to100 cmH2O. These changes do not have any significant clinical impact,since in clinical use 100 cmH2O is already considered high.
4.The upper limit for tidal volume Air/O2 has been changed from 4000 ml to2000 ml for the adult patient category. For the pediatric patient categorythe lower limit has been raised from 10 ml to 20 ml.
5.The upper limit for monitored inspiratory tidal volume Air/O2 has beenlowered from 4000 ml to 2000 ml for the adult patient category. For thepediatric patient category the lower limit has been raised from 10 ml to 20ml.
6.The HFT flow range for the pediatric patient category has been changedfrom 0.5 – 30 l/min to 2 – 30 l/min.

{9}------------------------------------------------

GETINGE

Pre-use check ●

The pre-use check has been adapted to the design changes for the subject device.

Non-clinical Testing and Performance

Maquet Critical Care has conducted risk analysis and performed necessary verification and validation activities to demonstrate that the design output of the modified devices meet the design input requirements:

Software

  • . Code review
  • Static code analysis ●
  • Unit tests ●
  • . Integration tests

Performance

  • System testing ●
  • Regression ●
  • Free User testing ●
  • . Waveform testing

Biocompatibility

  • . Volatile Organic Compounds
  • Particulate Testing ●

Human Factors Validation Testing

The following product standards are included in the verification:

  • ANSI/AAMI ES 60601-1:2005 + A1:2012, Medical electrical equipment . Part 1: General requirements for basic safety and essential performance, Recognition Number 19-4
  • IEC 60601-1-2:2014, Medical electrical equipment Part 1-2: General . requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and Test, Recognition Number 19-8

{10}------------------------------------------------

GETINGE

  • IEC 60601-1-8:2006 + A1:2012, Medical electrical equipment Part 1-8: ● General requirements for basic safety and essential performance - Collateral Standard: General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems, Recognition Number 5-76
  • ISO 80601-2-12:2011. Medical electrical equipment -- Part 2-1: Particular ● Requirements For The Safety Of Lung Ventilators - Critical Care Ventilators, Recognition Number 1-98
  • ISO 80601-2-55:2018. Medical electrical equipment -- Part 2-55: Particular . requirements for the basic safety and essential performance of respiratory gas monitors, Recognition Number 1-140
  • EN13544-1:2007, this standard is not recognized by FDA .

Biocompatibility evaluation of the Servo-air 4.0 is in accordance with AAMI / ANSI / ISO 10993-1:2009, recognition number 2-156 included the extent of recognition. The connector to High-Pressure gas is in accordance with CGA V-5:2008, Recognition Number 1-81.

Conclusion for Substantial Equivalence

The subject device and the predicate device have the same intended use. There are no new type questions of safety and effectiveness for the subject device compared to the cleared predicate device. MAQUET has conducted risk analysis and performed the necessary verification and validation activities to demonstrate that the design outputs meet the design input requirements and the appropriate product standards. MAQUET concludes that the performance data for the subject device shows that it is substantially equivalent to the cleared predicate device.

§ 868.5895 Continuous ventilator.

(a)
Identification. A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.(b)
Classification. Class II (performance standards).