K Number
K192547
Date Cleared
2020-01-17

(122 days)

Product Code
Regulation Number
862.1155
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

MAGLUMI 2000 HCG/S-HCG is an in vitro chemiluminescence immunoassay for the quantitative determination of total beta human chorionic gonadotropin (total ß-hCG) in human serum. The measurement of total ß-hCG is used as an aid in the early detection of pregnancy.

Device Description

MAGLUMI 2000 HCG/B-HCG kit consists of the following reagents: Magnetic Microbeads- coated with anti-HCG monoclonal antibody, containing BSA, NaN3 (<0.1%) Calibrator Low-Containing BSA and HCG antigen, NaN3(<0.1%) Calibrator High- Containing BSA and HCG antigen, NaN3(<0.1%) Buffer- containing BSA, NaN3 (<0.1%) ABEI Label- Anti-HCG monoclonal antibody labeled with ABEI, containing BSA, NaN3 (<0.1%) Control 1- Containing BSA and HCG antigen, NaN3 (<0.1%) Control 2- Containing BSA and HCG antigen, NaN3 (<0.1%)

AI/ML Overview

The provided text describes the performance characteristics of the MAGLUMI 2000 HCG/ß-HCG device, an in vitro chemiluminescence immunoassay for the quantitative determination of total beta human chorionic gonadotropin (total ß-hCG) in human serum. This information is presented as part of a 510(k) summary submitted to the FDA. While the document outlines various analytical performance studies, it does not explicitly define acceptance criteria in a quantitative table or refer to a multi-reader multi-case (MRMC) comparative effectiveness study.

However, we can infer acceptance based on the reported performance results aligning with standard clinical laboratory expectations and the successful substantial equivalence determination by the FDA. The study focuses on analytical performance characteristics rather than diagnostic accuracy involving human interpretation of results.

Here's an attempt to structure the information based on your request, identifying what is and isn't available in the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly present a table of acceptance criteria. Instead, it describes performance characteristics and indicates compliance with CLSI guidelines. We can infer the "acceptance" derived from the presented data and the overall conclusion of substantial equivalence.

Performance CharacteristicAcceptance Criteria (Inferred from CLSI/Clinical Practice)Reported Device Performance
PrecisionWithin-Run CV%: Generally < 10-15% for immunoassay, lower for higher concentrations.Control 1 (5.023 mIU/mL): SD=0.226, CV%=4.50%
Control 2 (37.135 mIU/mL): SD=1.529, CV%=4.12%
Control 3 (300.135 mIU/mL): SD=10.737, CV%=3.58%
Calibrator Low (2.876 mIU/mL): SD=0.199, CV%=6.92%
Calibrator High (1722.515 mIU/mL): SD=47.252, CV%=2.74%
Total (within instrument) CV%: Generally < 15-20%, lower for higher concentrations.Control 1: SD=0.321, CV%=6.39%
Control 2: SD=2.155, CV%=5.80%
Control 3: SD=16.757, CV%=5.58%
Calibrator Low: SD=0.257, CV%=8.94%
Calibrator High: SD=83.853, CV%=4.87%
Reproducibility (across instruments) CV%: Similar to Total CV%, reflecting inter-instrument consistency.Control 1: SD=0.326, CV%=6.49%
Control 2: SD=2.250, CV%=6.06%
Control 3: SD=17.665, CV%=5.89%
Calibrator Low: SD=0.277, CV%=9.65%
Calibrator High: SD=92.799, CV%=5.39%
LinearityR² > 0.99 for quantitative assays over measurement range.R² = 0.9932 (between 1.134 and 4680 mIU/mL)
Detection Limit (LOB)Determined by 95th percentile, should be clinically appropriate.0.302 mIU/mL (highest of 3 lots)
Detection Limit (LOD)Clinically appropriate for early detection of pregnancy.0.471 mIU/mL (highest of 3 lots)
Limit of Quantitation (LOQ)CV% no more than 20%, Bias no more than 15%.1.134 mIU/mL (highest of 3 lots)
InterferenceRecovery ±10% of initial value for various interferents.No significant interference observed at tested concentrations (details in text for TSH, LH, FSH, hGH, hCG α-subunit, bilirubin, hemoglobin, triglyceride, common drugs, HAMA, RF, total protein).
Hook EffectNo hook effect within clinically relevant high concentrations.No HOOK effect observed within 1,000,000 mIU/mL.
Dilution RecoveryPercent differences for diluted specimen versus expected concentration within 10%.Percent differences for diluted specimen versus expected concentration were within 10%.
Method ComparisonStrong correlation (high R²) and acceptable bias (slope near 1, intercept near 0) when compared to predicate device.Y = 0.988X + 1.995, R² = 0.993 (All three sites)
StabilityReagents/controls stable for specified period at specified conditions.Accelerated stability showed 12 months at 2-8°C. Real-time stability is ongoing.

2. Sample Size Used for the Test Set and Data Provenance

  • Precision Study: 80 samples analyzed per level (Control 1, Control 2, Control 3, Calibrator Low, Calibrator High, and 6 Native Serum Pools) on each of three instruments. Total N=240 samples per level across three instruments.
  • Linearity Study: Linearity samples prepared by mixing high and low level samples. Each sample was measured in quadruple on 3 lots of reagent. The exact number of distinct linearity samples (different concentrations) is not specified, but the range is 0.3 to 4680 mIU/mL.
  • Detection Limit Studies:
    • LOB: 80 measurements of HCG/ß-HCG negative serum samples using 3 different lots over 5 days.
    • LOD: Four levels of low samples, measured in 80 replicates over 5 days per sample using 3 lots of reagents.
    • LOQ: Six low serum samples, in six replicates per run, one run per day, over 5 days, using 3 lots of reagents.
  • Interference Study: Two base serum samples (6.0 mIU/mL and 100 mIU/mL HCG/ß-HCG) spiked with various cross-reactants. Also, human serum pools with 6.0 mIU/mL, 100 mIU/mL, and 2000 mIU/mL HCG/ß-HCG for endogenous substances and common drugs. Each tested using 3 lots of reagents. Exact number of distinct interferent samples not specified.
  • Hook Effect: Six samples with HCG/ß-HCG concentrations from 5000 to 1,000,000 mIU/mL prepared by spiking. Serial dilutions tested using 3 different lots.
  • On-board Dilution Recovery: Twelve serum samples with HCG/ß-HCG concentrations from 4475 to 223750 mIU/mL tested using three reagent lots and three instruments.
  • Method Comparison Study: 201 human serum samples with concentrations ranging from 1.1 to 4934 mIU/mL (as determined by the predicate device).
  • Expected Values/Reference Range: 431 serum samples from non-pregnant, apparently healthy females (20 years and older).

Data Provenance: The document does not explicitly state the country of origin for the data. Given the submitter's location (Shenzhen, China) and the FDA submission, it's likely the studies were conducted in China or involved samples from that region, but this is not explicitly confirmed. The studies are described in a manner typical of prospective performance evaluation studies for an in vitro diagnostic device, rather than retrospective analysis of pre-existing data.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

This type of information is not applicable to this submission. The device is an in vitro diagnostic immunoassay for quantitative measurement of a biomarker (hCG). Ground truth is established by the reference measurement procedure of the MAGLUMI 2000 instrument itself and validated through analytical performance studies (precision, linearity, detection limits, method comparison to a legally marketed predicate device, etc.) rather than through expert human interpretation of images or clinical findings.

4. Adjudication Method for the Test Set

Not applicable. As the device is an in vitro diagnostic assay providing a quantitative numerical result, there is no human interpretation or subjective assessment that would require an adjudication method. The "ground truth" for method comparison is the value obtained from the predicate device. For analytical performance studies, it's the objectively measured values and their statistical distributions.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

No. An MRMC study is typically conducted for image-based diagnostic aids (e.g., AI for radiology) where human readers interpret medical images. This device is a lab-based immunoassay that provides a quantitative numerical result. Therefore, an MRMC study is not relevant or applicable.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

Yes, the entire study is essentially a standalone performance evaluation. The MAGLUMI 2000 HCG/ß-HCG is an automated chemiluminescence immunoassay system. Its performance characteristics (precision, linearity, detection limits, interference, hook effect, dilution recovery, and method comparison) are evaluated as the direct output of the instrument and its reagents, without explicit human interpretation being part of the device's function or the primary subject of these studies. The quantitative output of the device itself constitutes its "performance."

7. The Type of Ground Truth Used

The ground truth for this device's performance evaluation is established through:

  • Reference Measurement Procedures/Known Concentrations: For studies like linearity, detection limits, and interference, known concentrations or "spiked" samples are used as a reference.
  • Measurement against a Predicate Device: For the method comparison study, the results from the Beckman Access Total B-HCG (5th IS) Assay (the predicate device) served as the comparative "ground truth" to establish substantial equivalence.
  • Statistical Analysis of Replicate Measurements: For precision, the statistical variation around the mean measured value for controls and patient samples provides the "ground truth" of the device's reproducibility.
  • Physiological/Clinical Samples: For determining expected values/reference ranges, samples from apparently healthy individuals are used.

8. The Sample Size for the Training Set

The document does not explicitly describe a "training set" in the context of an algorithm or machine learning model. This is an in vitro diagnostic assay, not an AI/ML-based device that typically undergoes a distinct training/validation/test split of data. The studies described are analytical verification and validation studies in a traditional medical device development sense.

9. How the Ground Truth for the Training Set was Established

Not applicable. As there is no defined "training set" for an AI/ML algorithm, this question is not relevant to the described device and its evaluation. The "training" of such a system would involve the manufacturer's internal assay development and optimization, which isn't part of a regulatory submission summary like this.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.

January 17, 2020

Shenzhen New Industries Biomedical Engineering Co., Ltd % Joe Shia Director LSI International Inc 504E Diamond Ave., Suite F Gaithersburg, MD 20877

Re: K192547

Trade/Device Name: MAGLUMI 2000 HCG/B-HCG Regulation Number: 21 CFR 862.1155 Regulation Name: Human Chorionic Gonadotropin (HCG) Test System Regulatory Class: Class II Product Code: DHA Dated: December 6, 2019 Received: December 10, 2019

Dear Joe Shia:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Marianela Perez-Torres, M.T., Ph.D. Acting Deputy Director Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K192547

Device Name MAGLUMI 2000 HCG/ß-HCG

Indications for Use (Describe)

MAGLUMI 2000 HCG/S-HCG is an in vitro chemiluminescence immunoassay for the quantitative determination of total beta human chorionic gonadotropin (total ß-hCG) in human serum. The measurement of total ß-hCG is used as an aid in the early detection of pregnancy.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K192547

510(k) SUMMARY

This summary of 510(k) safety and effectiveness information is submitted in accordance with the
requirements of 21 CFR 807.92

    1. Date: January 17, 2020 Shenzhen New Industries Biomedical Engineering Co., Ltd. 2. Submitter: No.16, Jinhui Road, Pingshan New District, Shenzhen China 518122 3. Contact person: Joe Shia LSI International Inc.
    • 504 East Diamond Ave., Suite F Gaithersburg, MD 20878 Telephone: 240-505-7880 Fax: 301-916-6213 Email:shiajl(@yahoo.com
    1. Device Name: MAGLUMI 2000 HCG/ß-HCG

Classification: Class II (assay)

Product CodeCFR #Product Name
DHA862.1155Human Chorionic Gonadotropin (HCG) Test System
    1. Predicate Devices:
      K130020, Beckman Access Total B-HCG (5th IS) Assay
    1. Device Description:
      MAGLUMI 2000 HCG/B-HCG kit consists of the following reagents: Magnetic Microbeads- coated with anti-HCG monoclonal antibody, containing BSA, NaN3 (<0.1%) Calibrator Low-Containing BSA and HCG antigen, NaN3(<0.1%) Calibrator High- Containing BSA and HCG antigen, NaN3(<0.1%) Buffer- containing BSA, NaN3 (<0.1%) ABEI Label- Anti-HCG monoclonal antibody labeled with ABEI, containing BSA, NaN3 (<0.1%) Control 1- Containing BSA and HCG antigen, NaN3 (<0.1%) Control 2- Containing BSA and HCG antigen, NaN3 (<0.1%)
    1. Intended Use:
      MAGLUMI 2000 HCG/ß-HCG is an in vitro chemiluminescence immunoassay for the quantitative

{4}------------------------------------------------

determination of total beta human chorionic gonadotropin (total ß-hCG) in human serum. The measurement of total ß-hCG is used as an aid in the early detection of pregnancy.

    1. Standard/Guidance Documents
      Clinical and Laboratory Standards Institute EP5-A2 - Evaluation of Precision Performance of Clinical Chemistry Devices-Approved Guideline-Second Edition.

Clinical and Laboratory Standards Institute EP6-A – Evaluation of the Linearity of Quantitative Analytical

Clinical and Laboratory Standards Institute EP17-A2: Evaluation of detection Capability for Clinical Laboratory Measurement Procedures

Clinical and Laboratory Standards Institute EP7-A2 - Interference Testing in Clinical Chemistry Clinical and Laboratory Standards Institute EP9-A2 – Method Comparison and Bias Estimation Using Patient Samples

    1. Substantial Equivalence Information

Assay Similarities

ItemPredicate DeviceCandidate Device
Intended Use/Indication forUseFor the quantitative determination ofhuman chorionic gonadotropin (β-subunit)(HCG/β-HCG) in human serum usingUniCel DxI chemiluminescenceimmunoassay analyzer. The measurementof HCG/β-HCG is used in the earlydetection of pregnancy.Same
Test principleSandwich ImmunoassaySame
MeasurementQuantitativeSame
TraceabilityWHO 5th International Standard 07/364Same
AutomatedYesSame

Assay Differences

ItemPredicate DeviceCandidate Device
SpecimenSerum or PlasmaSerum
Measuringrange0.6 – 1350 mIU/mLup to 270,000 mIU/mL with sample dilution1.132-4680 mIU/mL
CaptureantibodyParamagnetic particles coated with goatanti-mouse IgG: mouse monoclonal anti-ßhCG complexesMagnetic microbeads coated with anti-HCG monoclonal antibody

{5}------------------------------------------------

DetectionRabbit anti-βhCG alkaline phosphatase conjugateABEI-labeled Anti-HCG monoclonal antibody
Sample size25 ul15 ul
CalibrationUtilizes a stored calibration curve2 points
Calibrator6 levelsCalibrator Low and Calibrator High

10. Test Principle

The HCG/ß-HCG assay is an in vitro chemiluminescence immunoassay using FDA previously cleared MAGLUMI 2000 instrument (K162698). The assay is a sandwich chemiluminescence immunoassay.

The sample (or calibrator/control, if applicable), magnetic microbeads coated with anti-HCG monoclonal antibody are mixed thoroughly and incubated at 37°C to form immuno complexes, and then perform a wash cycle. After addition of ABEI labeled with anti-HCG monoclonal antibody, the complexes bound to the monoclonal antibody and incubate to form sandwich complexes. After precipitation in a magnetic field, decant the supernatant, and then perform another wash cycle. Subsequently, the Starter 1+2 are added to initiate a chemiluminescent reaction. The light signal is measured by a photomultiplier within 3 seconds as relative light units (RLUs), which is proportional to the concentration of HCG/β-HCG present in the sample (or calibrator/control, if applicable).

11. Performance Characteristics

    1. Analytical Performance
    • a. Precision

The precision was determined using the CLSI EP5-A3 protocol as a guide. The study was conducted on three different instruments with three controls, two calibrators, and 6 native patient serum pools. The data was collected over 20 days in duplicate with 2 runs per day with a total of 80 samples analyzed per level on each instrument. The results (in mIU/mL) obtained are summarized in the following tables:

SampleMean(mIU/mL)Within-RunBetween-RunBetween-DayTotal(within instrument)Reproducibility(across instruments)
(N=240)SDCV%SDCV%SDCV%SDCV%SDCV%
Control 15.0230.2264.500.0991.970.2054.080.3216.390.3266.49
Control 237.1351.5294.120.8142.191.2823.452.1555.802.2506.06
Control 3300.13510.7373.587.6662.5510.3323.4416.7575.5817.6655.89
Calibrator low2.8760.1996.920.0521.810.1545.350.2578.940.2779.65
Calibrator high1722.51547.2522.7443.6492.5353.7893.1283.8534.8792.7995.39
Native serum pool 11.8890.1377.250.0402.120.1015.350.1759.260.1859.79
Native serum pool 213.6250.5734.210.3702.720.5273.870.8626.330.8686.37
Native serum pool 3103.653.7963.662.6642.573.7103.585.9395.736.3776.15
Native serum pool 4770.93625.8013.3521.2932.7619.6382.5538.7915.0342.9315.57
Native serum pool 52443.44380.4593.2953.2572.1864.9442.66116.3094.76126.7355.19
Native serum pool 64555.709107.0242.3596.6002.12102.4522.25176.8683.88198.3244.35

{6}------------------------------------------------

b. Linearity

The linearity of the MAGLUMI HCG/ß-HCG method was determined following the CLSI EP6-A procedure. Linearity samples were prepared by mixing high level samples and low level samples through different volume ratios to span the whole measuring range with HCG/ß-HCG concentrations from 0.3 to 4680 mIU/mL. Each sample was measured in quadruple on 3 lots of reagent. Linearity was evaluated using regression analysis based on CLSI EP6-A.

The assays are linear between 1.134 and 4680 mIU/mL with the following relationship: Observed = 1.0243 (Expected) - 54.935, R2 = 0.9932

c. Stability

Accelerated stability study at 37°C showed that all controls are stable for 12 months at 2-8°C. Accelerated stability study at 37°C showed all that calibrators are stable for 12 months at 2-8°C. Accelerated stability study at 37°C showed that the reagent is stable for 12 months at 2-8°C. The real time stability at 2-8°C is on-going.

d. Detection Limit

Detection limit studies were performed following CLSI EP17-A guidelines. The limit of blank (LOB) is the 95th percentile value from 80 measurements of HCG/ß-HCG negative serum samples using 3 different lots of HCG/ß-HCG reagents over 5 days. The LOB corresponds to the concentration below which analyte-free samples are found with a probability of 95% and was determined to be 0.302 mIU/mL (highest of the 3 lots).

The limit of detection (LOD) is determined based on the LOB and the standard deviation of low concentration samples. The LOD corresponds to the lowest analyte concentration which can be detected. Four level of low samples were measured in 80 replicates over 5 days per sample using 3 lots of reagents. LOD was determined to be 0.471 mIU/mL (highest of the 3 lots).

The limit of quantitation (LOQ) was determined by measuring six low serum samples, in six replicates per run, one run per day, over 5 days, using 3 lots of reagents. LOQ is defined as the lowest analyte concentration that can be reproducibly measured with percent bias no more than 15% and CV% no more than 20% and was determined to be 1.134 mIU/mL (highest of the 3 lots).

Interference e.

A interference study was performed using two base serum samples containing HCG/ß-HCG of 6.0 mIU/mL and 100 mIU/mL respectively. These samples were spiked with various cross reactants and measured for these solutions using 3 lots of reagents. The

{7}------------------------------------------------

SubstancesHighest concentration tested without significant interference
TSH1000 mIU/L
LH500,000 mIU/L
FSH500,000 mIU/L
hGH25,000 mIU/L
hCG α-subunit500,000 mIU/L

following table shows interference of potential interference substances.

The effect of endogeous substances were evaluated using human serum pools. For each substance, three serum samples containing 6.0 mIU/mL and 2000 mIU/mL concentration of HCG/p-HCG were analyzed, the highest concentration of interferents were listed below at which no significant interference was observed.

Potential InterferentHighest concentration tested at which nosignificant interference is observed (mg/dL)
Conjugated bilirubin60
Unconjugated bilirubin42.5
Hemoglobin1000
triglyceride2000

The effect of common drugs and interference substances were evaluated using human serum pools. For each substance, three serum samples containing 6.0 mIU/mL, 100 mIU/mL and 2000 mIU/mL concentration of HCG/ß-HCG were analyzed. For all substances tested, no significant interference was defined as recovery ± 10% of initial value. The substances and the highest concentration tested which did not cause significant interference are listed below.

Potential InterferentHighest concentration tested at which nosignificant interference is observed (mg/dL)
Cefoxitin680
Ethanol790
Levodopa3.25
Metronidazole12.3
Ascorbic Acid (Vitamin C)16.95
Acetaminophen15.6
Biotin5
Cyclosporine0.6
Rifampicin6.5
Doxycycline3.2
Theophylline11.4
Ibuprofen50
EDTA-2Na4

{8}------------------------------------------------

The effect of human anti-mouse antibodies (HAMA), rheumatoid factor (RF) and human serum total protein was evaluated using human serum samples. Each potential interferent was added to HCG/B-HCG human serum samples and tested using 3 lots of reagents. For all substances tested, no significant interference was defined as recovery ± 10% of initial value. The potential interferents and the highest concentration tested which did not cause significant interference are listed below.

Potential InterferentHighest concentration tested at which nosignificant interference is observed
HAMA401 ng/mL
RF1745 IU/mL
Total protein15 g/dL
  • f. Hook Effect
    Six samples with HCG/ß-HCG concentrations from 5000 to 1000000 mIU/mL were prepared by spiking HCG/ß-HCG WHO 5th International Standard into HCG/ß-HCG free serum samples. Serial dilutions of these samples were made and tested using 3 different lots of the reagent. The obtained results indicated no HOOK effect was observed within HCG/ß-HCG concentration of 1000000 mIU/mL.

    1. On-board Dilution Recovery
      Verification studies were performed to determine the sample recovery after a 1:50 dilution is performed by both manually and automatically. Twelve serum samples with HCG/B-HCG concentrations from 4475 to 223750 mIU/mL as determined by the predicate device were tested using three reagent lots and three instruments with the 1:50 dilution. The percent differences for the diluted specimen versus the expected concentration were within 10%. The dilution study results support the claim that samples with HCG/0-HCG concentrations above 4680 mIU/mL may be diluted at 1:50 to obtain results up to 223750 mIU/mL.
    1. Comparison Studies
      A method comparison study was performed with 201 human serum samples with concentrations ranging from 1.1 to 4934 mIU/mL as determined by the predicate device. The comparison of the MAGLUMI HCG/B-HCG assay (y) with the predicate device, Beckman Access Total B-HCG assay (x), produced the following linear regression equations:
All three sites$Y = 0.988X + 1.995, R^2 =0.993$
Site 1$Y = 0.997X - 1.357, R^2 =0.995$
Site 2$Y = 0.997X + 7.614, R^2 =0.993$
Site 3$Y = 0.970X + 0.436, R^2 =0.992$
    1. Expected values/Reference range:

{9}------------------------------------------------

A total of 431 serum samples from non-pregnant, apparently healthy females (20 years and older) were tested according to the procedure in CLSI C28-A3. The expected normal range Maglumi HCG/ß-HCG is the following.

CategoryNumberMedian (mIU/mL)95th Percentile (mIU/mL)
20-39 years old138< 1.1< 1.1
≥40 years old, Pre-menopausal149< 1.11.6
≥51 years old, Post-menopausal1442.47.8

12. Conclusion

Based on the test principle and acceptable performance characteristics including precision, interference, specificity and method comparison of the device, it is concluded that the MAGLUMI 2000 HCG/ß-HCG is substantially equivalent to the predicate.

§ 862.1155 Human chorionic gonadotropin (HCG) test system.

(a)
Human chorionic gonadotropin (HCG) test system intended for the early detection of pregnancy —(1)Identification. A human chorionic gonadotropin (HCG) test system is a device intended for the early detection of pregnancy is intended to measure HCG, a placental hormone, in plasma or urine.(2)
Classification. Class II.(b)
Human chorionic gonadotropin (HCG) test system intended for any uses other than early detection of pregnancy —(1)Identification. A human chorionic goadotropin (HCG) test system is a device intended for any uses other than early detection of pregnancy (such as an aid in the diagnosis, prognosis, and management of treatment of persons with certain tumors or carcinomas) is intended to measure HCG, a placental hormone, in plasma or urine.(2)
Classification. Class III.(3)
Date PMA or notice of completion of a PDP is required. As of the enactment date of the amendments, May 28, 1976, an approval under section 515 of the act is required before the device described in paragraph (b)(1) may be commercially distributed. See § 862.3.