(371 days)
MINT™ is indicated for use in mid-face suspension surgery to temporarily fixate the cheek subcutaneous fat layer and SMAS layer in an elevated position for the treatment of moderate to severe nasolabial folds.
The MINT™ device is a sterile synthetic absorbable surgical suture comprised of polydioxanone, (GHoOJ)n. Polydioxanone has been found to be nonantigenic and to elicit only a slight tissue reaction during absorption. The pigment of the violet dye is D&C Violet No.2 (21CFR §74.3602). MINT™ is available in a range of gauge sizes and lengths. Each suture has bi-directional barbs along the long axis of the suture monofilament. The MINT™ Synthetic Absorbable PDO suture approximates tissues, without the need to tie surgical knots, by using the opposing barbs on the suture surface to embed in the tissues after the surgeon precisely places the suture within the tissues. Barbed suture lifting is a minimally invasive surgical technique for facial rejuvenation. The MINT™ Product Family includes models which are supplied with needles. For those models supplied with a needle, designated as "Lifting Thread Combined with needle" in the MINT™ Product Family instructions for use, the needle is used to make the insertion point and for threading of the suture in the patient's dermis.
Here's an analysis of the provided text to extract the requested information about device acceptance criteria and the supporting study:
Device: MINT™ Product Family (Absorbable Polydioxanone Surgical Suture) for mid-face suspension surgery.
Indication: Temporarily fixate the cheek subcutaneous fat layer and SMAS layer in an elevated position for the treatment of moderate to severe nasolabial folds.
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly present a "table of acceptance criteria" with quantifiable targets for the clinical study. However, the primary effectiveness endpoint and several secondary effectiveness evaluations serve as the de-facto acceptance criteria for clinical efficacy. The reported performance for these criteria is extracted below.
| Acceptance Criteria (Effectiveness Endpoint) | Reported Device Performance |
|---|---|
| Primary Effectiveness Endpoint: Ratio of subjects showing wrinkle improvement (WSRS: below -1 point) by independent evaluation at 12 weeks post-surgery. | 59 subjects (96.72%) showed improvement (WSRS: below -1 point). The lowest confidence level of 20.95% was greater than 0, indicating superiority. |
| Secondary Effectiveness Evaluation (Independent Evaluators): Ratio of subjects showing improvement (WSRS: below -1 point) at 4, 8, and 24 weeks. | Week 4: 100% (61 subjects) Week 8: 96.72% (59 subjects) Week 24: 90.16% (55 subjects) |
| Secondary Effectiveness Evaluation (Testers): Ratio of subjects showing improvement (WSRS: below -1 point) at 4, 8, 12, and 24 weeks. | Week 4: 100% (61 subjects) Week 8: 100% (61 subjects) Week 12: 100% (61 subjects) Week 24: 100% (61 subjects) |
| Mean WSRS Change from Baseline (Independent Evaluators): Statistical significance of decrease in WSRS at 4, 8, 12, and 24 weeks compared to baseline. | Baseline: 3.30 ± 0.45 points. Week 4: -1.56 ± 0.43 points Week 8: -1.41 ± 0.48 points Week 12: -1.23 ± 0.41 points Week 24: -1.20 ± 0.45 points. Statistically significant difference at all points (p<0.001). |
| Mean WSRS Change from Baseline (Testers): Statistical significance of decrease in WSRS at 4, 8, 12, and 24 weeks compared to baseline. | Baseline: 3.21 ± 0.41 points. Week 4: -1.23 ± 0.42 points Week 8: -1.25 ± 0.43 points Week 12: -1.18 ± 0.39 points Week 24: -1.16 ± 0.37 points. Statistically significant difference at all points (p<0.0001). |
| Mean GAIS Assessed by Evaluators: Satisfaction level (e.g., above a certain threshold) at 12 and 24 weeks. | Week 12: 1.93 ± 0.36 points Week 24: 1.85 ± 0.44 points. Distribution at Week 12: 86.89% (53 subjects) with 2 points ("very much improved"). Distribution at Week 24: 78.69% (48 subjects) with 2 points ("very much improved"). |
| Mean GAIS Assessed by Subjects: Satisfaction level (e.g., above a certain threshold) at 12 and 24 weeks. | Week 12 and 24: 1.92 ± 0.63 points. Distribution at Week 12: 57.38% (35 subjects) with 2 points. Distribution at Week 24: 52.46% (32 subjects) with 2 points. |
| Safety: Device is safe and effective as used according to the instructions for use. | Clinical investigation supports that the device is safe and effective. (No specific quantifiable safety endpoints are detailed in this summary, but the conclusion states it.) |
| Non-Clinical Performance: Compliance with USP standards for Suture diameter, Suture-Needle attachment, Suture tensile strength, and barb holding strength. | Complies with USP <861> Suture diameter, USP <871> Suture-Needle attachment, and USP <881> Suture tensile strength. Displays superior barb holding strength (20.94 ± 3.93 N) compared to predicate (9.90 ± 0.93 N). |
| Animal Performance (Absorption): Absorption timeframe. | In vivo absorption occurs between 180-220 days post-implantation (on Monosorb, which is identical to the sutures). |
| Animal Performance (Mechanical Strength): Tensile strength retention. | Tensile strength retention of MINT™ (USP 1) was 44.5% at 6 weeks. Considered approximately 50% at 6 weeks. Unable to be measured at 10 and 12 weeks due to severe degradation. |
2. Sample Size for the Test Set and Data Provenance
- Sample Size (Clinical Study): 62 male and female subjects (referred to as "FA set" in results, with 61 subjects for some analyses).
- Data Provenance: Prospective clinical study. The country of origin of the data is not explicitly stated, but the sponsor, Hans Biomed Corporation, is located in the Republic of Korea, suggesting the study may have been conducted there.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
- Number of Experts: Unspecified, but referred to as "independent, blinded assessors" and "blinded evaluators." Also, "testers" (who may or may not be the same as independent evaluators) were involved.
- Qualifications of Experts: Not specified. They are generally referred to as "evaluators" or "assessors."
4. Adjudication Method for the Test Set
The document mentions "Independent, blinded assessors" and that the analysis was based on "blinded evaluator Global Aesthetic Improvement Scale (GAIS) ratings" and "comparing the photos... by blinded evaluators." It also refers to "evaluation of independent evaluators" and "evaluation of testers."
However, no specific adjudication method (e.g., 2+1, 3+1, majority vote, etc.) for resolving disagreements between multiple evaluators is described. It implies individual independent ratings were collected and analyzed (e.g., averaging or using individual scores).
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done
- No, a MRMC comparative effectiveness study was not done in the sense of comparing human readers' performance with and without AI assistance, as this is a medical device (suture) and not an AI/CADe system for image analysis.
- The clinical study evaluates the device's effectiveness through expert assessment of clinical outcomes (WSRS, GAIS) performed directly on subjects or their photos, not through a reader-AI interaction for diagnosis or interpretation.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
- Yes, a standalone performance was done for the device in the context of its mechanical and biological properties.
- The "non-clinical performance data" (USP compliance, barb holding strength) and "animal performance data" (absorption, mechanical strength in vivo) represent standalone performance assessments of the physical characteristics of the suture itself, without human intervention in the device's inherent function.
- The clinical study then assessed its performance in vivo in human subjects, which is the ultimate "standalone" performance for a medical implant like a suture.
7. The Type of Ground Truth Used (Clinical Study)
- Expert Consensus and Subjective Scales: The primary ground truth for the clinical study was established using validated subjective rating scales like the 5-grade Wrinkle Severity Rating Scale (WSRS) and the Global Aesthetic Improvement Scale (GAIS). These were applied by:
- Independent, blinded assessors/evaluators.
- Testers (likely clinicians involved in the study).
- The subjects themselves (for GAIS).
- This represents a form of expert assessment/consensus based on observable clinical outcomes, rather than objective pathology or hard outcomes data like mortality.
8. The Sample Size for the Training Set
- Not Applicable. This document describes a medical device (surgical suture), not an AI/Machine Learning model. Therefore, there is no "training set" in the context of machine learning. The clinical study served as the primary evidence for the expanded indication.
9. How the Ground Truth for the Training Set Was Established
- Not Applicable, as there is no training set for an AI/ML model.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
September 9, 2020
Hans Biomed Corporation % Michael Dun Senior Consultant, Quality and Regulatory Affair Emergo Global Consulting, LLC 2500 Bee Cave Road Building 1, Suite 300 Austin, Texas 78746
Re: K192423
Trade/Device Name: MINT Product Family Regulation Number: 21 CFR 878.4840 Regulation Name: Absorbable Polydioxanone Surgical Suture Regulatory Class: Class II Product Code: NEW Dated: June 18, 2020 Received: August 10, 2020
Dear Michael Dun:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Cindy Chowdhury, Ph.D., M.B.A. Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K192423
Device Name MINTTM
Indications for Use (Describe)
MINT™ is indicated for use in mid-face suspension surgery to temporarily fixate the cheek subcutaneous fat layer and SMAS layer in an elevated position for the treatment of moderate to severe nasolabial folds.
| Type of Use (Select one or both, as applicable) |
|---|
| ------------------------------------------------- |
X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) Summary
MINT™ Product Family
1. Submission Sponsor
Hans Biomed Corporation
7, Jeongui-ro 8-gil, Songpa-gu
Seoul, 05836
Republic of Korea
Contact: Lucy Choi
Title: General Manager
2. Submission Correspondent
Emergo Global Consulting, LLC
2500 Bee Cave Road
Building 1, Suite 300
Austin, TX 78746
Office Phone: (512) 327.9997
Contact: Michael Dun
Title: Senior Consultant, Quality and Regulatory Affairs
3. Date Prepared
9 September2020
4. Device Identification
| Trade/Proprietary Name: | MINT™ Product Family |
|---|---|
| Common/Usual Name: | Absorbable surgical suture |
| Classification Name: | Absorbable polydioxanone surgical suture |
| Regulation Number: | 21 CFR §878.4840 |
{4}------------------------------------------------
| Product Code: | NEW |
|---|---|
| Device Class: | Class II |
| Classification Panel: | General & Plastic Surgery |
5. Legally Marketed Predicate Device(s)
The primary predicate is the Silhouette InstaLift™ device cleared under K163676 by Silhouette Lift, Inc. The reference device is the original MINT™ device cleared under K130191 by Hans Biomed Corporation. Neither the predicate nor reference devices have been subject to a design-related recall.
6. Indication for Use Statement
MINT™ is indicated for use in mid-face suspension surgery to temporarily fixate the cheek subcutaneous fat layer and SMAS layer in an elevated position for the treatment of moderate to severe nasolabial folds.
7. Device Description
The MINT™ device is a sterile synthetic absorbable surgical suture comprised of polydioxanone, (GHoOJ)n. Polydioxanone has been found to be nonantigenic and to elicit only a slight tissue reaction during absorption. The pigment of the violet dye is D&C Violet No.2 (21CFR §74.3602). MINT™ is available in a range of gauge sizes and lengths.
Each suture has bi-directional barbs along the long axis of the suture monofilament. The MINT™ Synthetic Absorbable PDO suture approximates tissues, without the need to tie surgical knots, by using the opposing barbs on the suture surface to embed in the tissues after the surgeon precisely places the suture within the tissues. Barbed suture lifting is a minimally invasive surgical technique for facial rejuvenation
While the formation of barbs in the MINT™ reduces the tensile strength relative to non-barbed suture material of the same size, tying knots in non-barbed suture materials also reduce their effective strength. For this reason, the strength of the MINT™ can be compared with USP knot strength of nonbarbed sutures and the USP size of MINT™ is 1 while its tensile strength is equivalent to that of USP 2-0, as demonstrated under the original 510(k) for the MINT™ device (K130191).
The MINT™ Product Family includes models which are supplied with needles. For those models supplied with a needle, designated as "Lifting Thread Combined with needle" in the MINT™ Product Family instructions for use, the needle is used to make the insertion point and for threading of the suture in the patient's dermis.
8. Substantial Equivalence Discussion
{5}------------------------------------------------
The MINT™ Product Family is currently indicated for use in soft tissue approximation where the use of absorbable sutures is appropriate (K130191). The purpose of this submission is to obtain clearance of a new indication for the MINT™ Product Family for use in "lifting" procedures which includes variants supplied with needles. Due to the difference in intended use, the cleared MINT™ device family (K130191) is presented as a reference device in this submission while the cleared Silhouette InstaLift™ device family (K163676) is presented as the predicate as it has an identical intended use to that for the subject MINT™ device and includes variants supplied with identical composition to the needles supplied with certain variants of the subject device. Additionally, while the predicate device is comprised of a different material (poly glycolide/L-lactide) and has a different number of barbs to that used in the subject device (polydioxanone (PDO)) questions of biocompatibility and clinical effectiveness are identical. Therefore, different questions of safety and effectiveness are not raised with the subject device in relation to these difference in technological characteristics.
While there are differences in the indications for use between the subject MINT™ Product Family and the predicate Silhouette InstaLift™ device family (K163676), these differences do not constitute differences in intended use or raise different questions of safety and effectiveness. Both devices are absorbable surgical sutures for use in soft tissue indicated for mid-face suspensions surgery to temporarily fixate the cheek in an elevated position. Both devices achieve nasolabial fold depth reduction as it is a clinical outcome of mid-face suspension surgery, as reported in the following published literature, copies of which accompany this submission:
-
Noone, R.B. (2006) Suture Suspension Malarplasty with SMAS Plication and Modified SMASectomy: A Simplified Approach to Midface Lifting. Plast. Reconstr. Surg. 117: 792.
-
Benito, J., et al. (2011) Facial Rejuvination and Improvement of Malar Projection Using Sutures with Absorbable Cones: Surgical Technique and Case Series. Aesth. Plast. Surg. 35: 248-253.
-
Paul, M. D. et al. (2006) The Evolution of the Midface Lift in Aesthetic Plastic Surgery. Plast. Reconstr. Surg. 117: 1809-1827.
The following table compares the MINT™ Product Family to the predicate device with respect to indications for use, principles of operation, technological characteristics, materials, and performance testing. The subject device does not raise any new issues of safety or effectiveness based on the similarities to the predicate device and shown by clinical and other scientific data presented in this submission.
| Manufacturer | Submission Device | Predicate Device | Significant Differences |
|---|---|---|---|
| Hans Biomed Corporation | SILHOUETTE LIFT, INC | ||
| Trade Name | MINT™ | Silhouette InstaLift™ | |
| Manufacturer | Submission Device | Predicate Device | Significant Differences |
| Hans Biomed Corporation | SILHOUETTE LIFT, INC | ||
| Trade Name | MINT™ | Silhouette InstaLift™ | |
| 510(k)Number | K192423 | K163676 | Not applicable |
| Product Code | NEW | GAM | Different;The difference in ProductCode does not raise newquestions of safety andeffectiveness. Both productcodes cover absorbablesurgical sutures. |
| RegulationNumber | 21 CFR 878.4840 | 21 CFR 878.4493 | Different;The difference inRegulation Number doesnot raise new questions ofsafety and effectiveness.Both regulations coverabsorbable, sterile flexiblesurgical sutures that areintended for use in softtissue approximation,coated or uncoated, andwith or without a standardneedle attached. |
| RegulationName | Suture, Surgical,Absorbable, Polydioxanone | Suture, Absorbable,Synthetic, Polyglycolic Acid | Different;The difference inRegulation Name does notraise new questions ofsafety and effectiveness.Both regulations coverabsorbable, sterile flexiblesurgical sutures that areintended for use in softtissue approximation,coated or uncoated, andwith or without a standardneedle attached. |
| Manufacturer | Submission Device | Predicate Device | Significant Differences |
| Hans Biomed Corporation | SILHOUETTE LIFT, INC | ||
| Trade Name | MINT™ | Silhouette InstaLift™ | |
| needle attached. | |||
| Intended Use | Absorbable surgical suturesfor use in soft tissue | Absorbable surgical suturesfor use in soft tissue | Identical |
| Indication foruse | MINT™ is indicated for usein mid-face suspensionsurgery to temporarilyfixate the cheeksubcutaneous fat layer andSMAS layer in an elevatedposition for the treatmentof moderate to severenasolabial folds. | The Silhouette InstaLift™device is indicated for use inmid-face suspension surgeryto temporarily fixate thecheek subdermis in anelevated position. | Similar. While there aredifferences, these do notconstitute differences inintended use or raisedifferent questions ofsafety and effectiveness.Both devices areabsorbable surgical suturesfor use in soft tissueindicated for mid-facesuspension surgery totemporarily fixate thecheek in an elevatedposition. Both devicesachieve nasolabial folddepth reduction as it is aclinical outcome of mid-face suspension surgery, asreported in publishedliterature referenced in this510(k). |
| Raw Material | Polydioxanone sutureStainless steel (SUS 304)needle | Poly glycolide/L-lactidesutureStainless steel (SUS 304)needle | Different; While there aredifferences in suturematerial to the predicatedevice K163676, these donot raise new questions ofsafety and effectiveness.Biocompatibility data on filefor the reference MINT™device family K130191demonstrate the device's |
| Manufacturer | Submission Device | Predicate Device | Significant Differences |
| Hans Biomed Corporation | SILHOUETTE LIFT, INC | ||
| Trade Name | MINT™ | Silhouette InstaLift™ | biocompatibility whileclinical data presented inthis submission alsodemonstrate equivalentsafety and performance. |
| SutureCharacteristic | Synthetic AbsorbableMonofilament | Synthetic AbsorbableMonofilament | Identical |
| Technique ofDeployment | Needle-based deployment | Needle-based deployment | Identical;Both include variantssupplied with needles. |
| TechnologicalCharacteristic | Bi-directional barbs alongthe long axis of the suturemonofilament | Bi-directional barbs along thelong axis of the suturemonofilament | Identical |
| Sterilization | Ethylene Oxide | Ethylene Oxide | Identical |
| Size (USP) | 1-0 | 1-0 | Identical |
| Absorbable | Absorbable | Absorbable | Identical |
| Patientcontact | Implant | Implant | Identical |
| Duration ofcontact | Over 30 Days | Over 30 Days | Identical |
| Number ofbarbs perlinear lengthof suture | 20 barbs per 10cm in allsutures. | 8 barbs in 30cm sutures.12 barbs in 27.5cm sutures.16 barbs in 26.8cm sutures. | Different; While there aredifferences in the numberof barbs to the predicatedevice K163676, thesedevices are substantiallyequivalent as no newquestions of safety andeffectiveness are raised, asdemonstrated by barbholding strength testingand clinical data presentedin this submission. |
| Manufacturer | Submission Device | Predicate Device | Significant Differences |
| Hans Biomed Corporation | SILHOUETTE LIFT, INC | ||
| Trade Name | MINTTM | Silhouette InstaLiftTM | |
| Barb Holding Strength(Average ± Standard Deviation) | $20.94 ± 3.93 N$ | $9.90 ± 0.93 N$ | Different; While there are differences in the average barb holding strength to predicate device K163676, the differences do not raise new questions of safety and effectiveness as the average barb holding strength for the subject MINTTM device is superior to that of the predicate. Clinical data presented in this submission also demonstrate equivalent safety and performance. |
| Barb shape | Cog shape | Cog shape | Identical |
| Barb size | 0.601mm | 0.601mm | Identical |
| Barb direction | A section and B section is opposite direction | A section and B section is opposite direction | Identical |
| Pattern of the barbs | Bi-directional barbs along the long axis of the suture | Bi-directional barbs along the long axis of the suture | Identical |
| Suture diameter | Compliant with USP <861> requirements | Compliant with USP <861> requirements | Identical |
| Suture tensile strength | Compliant with USP <881> requirements | Compliant with USP <881> requirements | Identical |
| Suture-Needle attachment | Compliant with USP <871> requirements | Compliant with USP <871> requirements | Identical |
Table 5A - Comparison of Characteristics
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
9. Non-Clinical Performance Data
Additional testing was conducted and relevant scientific data collated for this 510(k). Previous tests performed on the MINT™ reference device (K130191) included those for overall design, sterilization, biocompatibility, and the physical and performance testing as described in the guidance entitled Class II Special Controls Guidance Document: Surgical Sutures, confirming that the design output meets the
{10}------------------------------------------------
design inputs and specifications for the sutures. These data also apply to the subject device, in all aspects except the expanded indications for use and provision of variants with needles.
The non-clinical performance testing performed on the subject device MINT™ Product Family variants including needles included physical and performance and real-time stability testing was focused on generating scientific data to support the inclusion of the needles in addition to generating scientific data to support the expansion of suture variants in order to demonstrate compliance with USP <861> Suture diameter, USP <871> Suture-Needle attachment, and USP <881> Suture tensile strength requirements.
Additional comparative barb holding strength testing was performed to demonstrate that the MINT™ Product Family displays superior barb holding strength to that for the Silhouette InstaLift™ predicate device (K163676).
The results of non-clinical performance testing demonstrate substantial equivalency between the MINT™ Product Family (K192423) and the predicate Silhouette InstaLift™ device (K163676) as they both comply with all the requirements of USP <861> Suture diameter, USP <871> Suture-Needle attachment, and USP <881> Suture tensile strength, and that the MINT™ Product Family displays superior barb holding strength to that for the Silhouette InstaLift™.
10. Animal Performance Data
In vivo studies in animal models to demonstrate absorption and mechanical strength of the subject device were performed.
The in vivo absorption study was completed on 30 rats prepared and anesthetized in accordance with the procedure and was performed on Monosorb, identical to the sutures used in the MINT™ Product Family. The results of this study indicated in vivo absorption of the sutures occurs between 180°220 days post-implantation.
The in vivo mechanical strength study was completed on rats prepared and anesthetized in accordance with the procedure and performed on MINT™ barbed suture. The results of this study indicated that tensile strength retention of MINT™ (USP 1) was 44.5% at 6 weeks and was unable to be measured at 10 and 12 weeks post-implantation due to severe degradation. Based on the results of this study, the breaking strength retention of MINT™ is considered to be approximately 50% at 6 weeks.
11. Clinical Performance Data
A prospective clinical study was conducted to evaluate safety and effectiveness of the MINT™ Product Family to support mid-face suspension surgery to fixate the cheek sub dermis in an elevated position. The aim of this study was to evaluate the surgical outcomes associated with using MINT™ for improving nasolabial folds.
{11}------------------------------------------------
In this study, 62 male and female subjects were assessed pretreatment and compared to 4, 8, 12, and 24 weeks following treatment using the previously validated and published 5-grade Wrinkle Severity Rating Scale (WSRS) score and rating by independent, blinded assessors. The correction of nasolabial folds using MINT™ was evaluated using blinded evaluator Global Aesthetic Improvement Scale (GAIS) ratings.
The primary effectiveness evaluation was carried out at 12 weeks after surgery by comparing the photos of nasolabial folds of the patients in the study group by blinded evaluators. Twenty-four weeks were set for final safety and effectiveness evaluation period. The primary effectiveness evaluation was carried out at 12 weeks after surgery by comparing the photos of the patients in the study group.
The ratio of subjects in the study group who showed wrinkle improvement (WSRS: below -1 point) by independent evaluation at 12 weeks after the application of test device, the primary effectiveness endpoint, was analyzed. The results showed 59 subjects (96.72%) in FA set showed improvement (WSRS: below -1 point) and the lowest confidence level of 20.95% was greater than 0, indicating superiority.
The results of secondary effectiveness evaluation showed the ratio of the subjects in FA who showed improvement (WSRS: below -1 point) by the evaluation of independent evaluators after the application of test device was 100% (61 subjects) at week 4, 96.72% (59 subjects) at week 8 and 90.16% (55 subjects) at week 24. The ratio of the subjects in FA who showed improvement (WSRS: below -1 point) by the evaluation of testers was 100% (61 subjects) at week 4, 8, 12 and 24. This tendency was represented in the results of an analysis of PP test.
When comparing the WSRS evaluated by independent evaluators at 4, 8, 12 and 24 weeks after the application of test device with baseline of FA set was 3.30 ± 0.45 points; the difference in the mean of WSRS in comparison with baseline at each point was -1.56 ± 0.43 points at week 4, 1.41 ± 0.48 points at week 8, -1.23 ± 0.41 points at week 12, and -1.20 ± 0.45 points at week 24, indicating a decrease over time and statistically significant difference in comparison with baseline at all points of time (p<0.001). When comparing WSRS evaluated by testers at 4, 8, 12 and 24 weeks after the application of test device with the baseline by point of time, the mean of WSRS in FA set before the application of medical device was 3.21 ± 0.41 points; the difference in the mean of WSRS in comparison with the baseline at each point of time was -1.23 ± 0.42 points at week 4 and -1.25 ± 0.43 points at 8 weeks, indicating an increase in wrinkle improvement. It was -1.18 ± 0.39 points at week 12 and -1.16 ± 0.37 points at week 24 after the application of test device, indicating that the difference in the mean of WSRS after 12 weeks became smaller than that at week 8. Also, there was the statistically significant difference in comparison with the baseline at all points of time (p<0.0001).
The mean of GAS in FA set assessed by evaluators at 12 weeks after the application of test device in comparison with baseline was 1.93 ± 0.36 points at week 12 and 1.85 ± 0.44 points at week 24, indicating the high satisfaction level of evaluators. When analyzing the distribution of GAIS point evaluated by evaluators at each point of time, 2 points (very much improved) took the greatest
{12}------------------------------------------------
proportion- 53 subjects (86.89%) at week 12, and 48 subjects (78.69%) at week 24. The mean of GAIS evaluated by the subject at 12 and 24 weeks after the application of test device was 1.92 ± 0.63 points in FA set, indicating a high satisfaction level of subjects. When analyzing the distribution of GAIS point evaluated by evaluators at each point of time, 2 points took the greatest proportion - 35 subjects (57.38%) at week 12 and 32 subjects (52.46%) at week 24.
Results of the clinical investigation support the indications for the MINT™ Product Family to support mid-face suspension surgery to fixate the cheek sub dermis in an elevated position and is effective for the improvement of facial wrinkles. Clinical study conclusion confirms that the device is safe and effective as used according to the instructions for use.
12. Statement of Substantial Equivalence
The subject device is identical to the predicate in relations for use. The indications for use for the subject device were evaluated in a clinical study which demonstrated acceptable performance and safety, thereby supporting a finding of substantial equivalence. This clinical data also demonstrates that no different questions of safety and effectiveness were raised for differences in the number of barbs between the subject and predicate devices. Differences in raw material are addressed by the data presented in the 510(k) for the reference device along with the presented non-clinical, animal, and clinical performance data. The needles supplied with the subject device were evaluated in pre-clinical (bench) performance testing which demonstrate acceptable performance and safety.
§ 878.4840 Absorbable polydioxanone surgical suture.
(a)
Identification. An absorbable polydioxanone surgical suture is an absorbable, flexible, sterile, monofilament thread prepared from polyester polymer poly (p-dioxanone) and is intended for use in soft tissue approximation, including pediatric cardiovascular tissue where growth is expected to occur, and ophthalmic surgery. It may be coated or uncoated, undyed or dyed, and with or without a standard needle attached.(b)
Classification. Class II (special controls). The special control for the device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.