(28 days)
Not Found
No
The 510(k) summary describes a standard guiding catheter and its physical properties and performance tests. There is no mention of AI, ML, image processing, or any data-driven algorithms.
No
The device is described as providing a pathway for other therapeutic devices, rather than being therapeutic itself.
No
The device is described as providing a "pathway through which therapeutic devices are introduced," indicating its function as an access tool for treatment, not for diagnosis.
No
The device description clearly states it is an "intravascular catheter" and describes physical characteristics like outer diameter, curves, and lengths. The performance studies also detail bench tests related to the physical properties of the catheter. This indicates it is a physical medical device, not software-only.
Based on the provided information, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- Intended Use: The intended use is to "provide a pathway through which therapeutic devices are introduced" within the coronary or peripheral vascular system. This describes a device used in vivo (within the body) to facilitate a medical procedure, not a device used in vitro (outside the body) to examine specimens from the body for diagnostic purposes.
- Device Description: The description reinforces its use as an intravascular catheter, designed to be inserted into blood vessels.
- Lack of IVD Characteristics: There is no mention of analyzing biological samples, detecting analytes, or providing diagnostic information based on testing specimens.
Therefore, the Medtronic Guiding Catheter is a medical device used for interventional procedures, not an in vitro diagnostic device.
N/A
Intended Use / Indications for Use
The Medtronic Guiding Catheter is designed to provide a pathway through which therapeutic devices are introduced. The guiding catheter is intended to be used in the coronary or peripheral vascular system.
Product codes (comma separated list FDA assigned to the subject device)
DQY
Device Description
Medtronic 6F Taiga Guiding Catheter is an intravascular catheter that is supplied sterile and intended for single use only. The primary function of the 6F Taiga Guiding Catheter is to provide a pathway through which therapeutic devices are introduced. Medtronic 6F Taiga Guiding Catheter has a 6F outer diameter (0.082") and is be available in variety of curves and lengths, identical to the predicate Medtronic Taiga Guide Catheter. The various curves and lengths available, will allow the catheter to accommodate the varying patient vascular system anatomies
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
coronary or peripheral vascular system
Indicated Patient Age Range
Not Found
Intended User / Care Setting
Not Found
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Performance/ Bench Testing: The following bench tests were performed on the predicate Taiga Guide Catheters:
- Shaft Inner Diameter
- Shaft Outer Diameter
- Segments Outer Diameter
- Sleeve Outer Diameter
- Overlap Segment/Shaft OD
- Usable Length
- Soft Tip Exposed Length
- Shaft Stiffness @ Room Temp
- Shaft Kink @ Room Temp
- Shaft Stiffness @ Body Temp
- Shaft Kink @ Body Temp
- Shaft Torque
- Shaft Torque to separation
- Shaft Crush
- Shaft Tensile
- Primary Stiffness
- Secondary Stiffness
- Arch Stiffness @ Room Temp
- Arch Kink @ Room Temp
- Arch Stiffness @ Body Temp
- Arch Kink @ Body Temp
- Segment Crush
- Ostial Engagement
- Secondary Curve Retention
- Primary Curve Retention
- Segment Tensile
- Soft Tip Tensile*
- Soft Tip Compression*
- Hub Tensile
- Pressure and Leak Resistance
- Inner Lubricity
*These tests, Soft Tip Tensile and Soft Tip Compression, were performed on the subject device to demonstrate substantial equivalence to the predicate Taiga device for the modification in this premarket notification.
Biocompatibility Testing: Pursuant to the ISO 10993-1 Biological evaluation of medical devices- Part 1: Evaluation and testing within a risk management process; the biocompatibility testing for the predicate Taiga Guide Catheter represents the modified Taiga Guide Catheter has met all requirements.
The performance testing along with biocompatibility testing demonstrated that the modified Taiga Guide Catheters are substantially equivalent to the predicate device.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.
Medtronic 6F Taiga™ Guiding Catheter - K083422
Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).
Not Found
§ 870.1250 Percutaneous catheter.
(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).
0
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. Underneath the square are the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
September 20, 2019
Medtronic Vascular Colleen Mullins, MSRA, RAC Principal Regulatory Affairs Specialist 37a Cherry Hill Drive Danvers, Massachusetts 01923
Re: K192296
Trade/Device Name: Medtronic 6F Taiga Guiding Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY Dated: August 22, 2019 Received: August 23, 2019
Dear Colleen Mullins:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
1
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/ training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory-assistance/contact-usdivision-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Lydia Glaw Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
2
Indications for Use
510(k) Number (if known) K192296
Device Name Medtronic 6F Taiga Guiding Catheter
Indications for Use (Describe)
The Medtronic Guiding Catheter is designed to provide a pathway through which therapeutic devices are introduced. The guiding catheter is intended to be used in the coronary or peripheral vascular system.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) |
---|
Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
3
K192296 Special 510(k) Summary
| Submitter: | Medtronic Vascular
37A Cherry Hill Drive
Danvers, MA 01923,
USA |
|-------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Colleen Mullins
Principal Regulatory Affairs Specialist
37A Cherry Hill Drive
Danvers, MA 01923
USA
Phone: (978)-739-3267
Email: colleen.mullins@medtronic.com |
| Date Prepared: | September 20, 2019 |
| Trade Name: | Medtronic 6F Taiga™ Guiding Catheter |
| Common Name: | Guiding Catheter |
| Classification
Name: | Percutaneous Catheter
Product Code DQY. |
| Predicate Device: | Medtronic 6F Taiga™ Guiding Catheter - K083422 |
| Device
Description: | Medtronic 6F Taiga Guiding Catheter is an intravascular
catheter that is supplied sterile and intended for single use
only. The primary function of the 6F Taiga Guiding Catheter
is to provide a pathway through which therapeutic devices are
introduced. Medtronic 6F Taiga Guiding Catheter has a 6F
outer diameter (0.082") and is be available in variety of
curves and lengths, identical to the predicate Medtronic Taiga
Guide Catheter. The various curves and lengths available, will
allow the catheter to accommodate the varying patient
vascular system anatomies |
| Statement of
Intended Use: | The Medtronic Guiding Catheter is designed to provide a
pathway through which therapeutic devices are introduced.
The guiding catheter is intended to be used in the coronary or
peripheral vascular system. |
4
Summary of Technological Characteristics: | The 6F Taiga Guiding Catheter consists of the following technical characteristics: Catheter length: 30-130 cmCatheter size: 6 FrenchDesign Components/ Construction: Luer HubStrain ReliefShaft comprising of: Inner linerBraid wireOuter jacketIntermediate layerSoft tipDistal SegmentsSleeve |
---|---|
Compared to the predicate device, the subject device was modified to include a specification related to the soft tip component. | |
Summary of Non-clinical Data: | The device performance/ bench tests and biocompatibility tests were conducted in accordance to the relevant FDA guidance to demonstrate substantial equivalence to the legally marketed predicate devices. |
Performance/ Bench Testing: The following bench tests were performed on the predicate Taiga Guide Catheters: Shaft Inner DiameterShaft Outer DiameterSegments Outer DiameterSleeve Outer DiameterOverlap Segment/Shaft ODUsable LengthSoft Tip Exposed LengthShaft Stiffness @ Room TempShaft Kink @ Room TempShaft Stiffness @ Body TempShaft Kink @ Body TempShaft TorqueShaft KinkShaft Torque to separationShaft CrushShaft TensilePrimary Stiffness |
5
20) Arch Kink @ Room Temp | |
---|---|
21) Arch Stiffness @ Body Temp | |
22) Arch Kink @ Body Temp | |
23) Segment Crush | |
24) Ostial Engagement | |
25) Secondary Curve Retention | |
26) Primary Curve Retention | |
27) Segment Tensile | |
28) Soft Tip Tensile* | |
29) Soft Tip Compression* | |
30) Hub Tensile | |
31) Pressure and Leak Resistance | |
32) Inner Lubricity | |
*These tests, Soft Tip Tensile and Soft Tip Compression, | |
were performed on the subject device to demonstrate | |
substantial equivalence to the predicate Taiga device for the | |
modification in this premarket notification. | |
Biocompatibility Testing: Pursuant to the ISO 10993-1 | |
Biological evaluation of medical devices- Part 1: Evaluation | |
and testing within a risk management process; the | |
biocompatibility testing for the predicate Taiga Guide | |
Catheter represents the modified Taiga Guide Catheter has | |
met all requirements. | |
The performance testing along with biocompatibility testing | |
demonstrated that the modified Taiga Guide Catheters are | |
substantially equivalent to the predicate device. | |
Summary of | |
Clinical Data: | No clinical investigation has been performed on the modified |
device. | |
Conclusion from | |
Data: | Medtronic Vascular has demonstrated that the modified Taiga |
Guide Catheters are substantially equivalent to the legally | |
marketed predicate devices based on its intended use and | |
technological characteristics. |
-
Secondary Stiffness
-
Arch Stiffness @ Room Temp