K Number
K192248
Date Cleared
2019-11-25

(98 days)

Product Code
Regulation Number
888.3080
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Cortina™ [MAX] Lumbar Cage System is indicated for use in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1 whose condition requires the use of interbody fusion combined with supplemental fixation. The interior of the Cortina™ [MAX] Lumbar Cage System should be packed with autogenous bone graft (i.e. autograft). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment.

Device Description

The Cortina™ [MAX] Lumbar Cage System is an intervertebral fusion device made from medical grade PEEK with tantalum markers manufactured from EVONIK PEEK as described in MAF-1922 with the submissions of the predicate Cortina [MAX] (K171914 and K180431) or medical grade titanium per ASTM F136 with a TECOTEX® surface coating. The subject device is offered in a variety of styles and sizes to accommodate various patient anatomies. This submission is intended to add the titanium implants and expand the size offering to include larger footprints.

AI/ML Overview

1. Acceptance criteria and reported device performance

The document does not explicitly state numerical acceptance criteria for device performance. Instead, it argues that the new device, the Cortina™ [MAX] Lumbar Cage System made from titanium, requires no additional testing because it is identical in form, shape, and function to previously cleared PEEK devices (K171914/K180431). The rationale given is that titanium material produces implants with a higher yield load and stiffness when the shape is identical.

Similarly, for the PEEK devices, the document states that the addition of longer PLIF implants (36-42 mm) does not create a new mechanical worst-case compared to the previously tested 20x9x7mm PLIF (K171914/K180431), thus no additional testing is required for the additional length in the line extension.

The DLIF implants also do not need to be tested as a new worst-case because their design and footprint do not represent a new worst-case for either static or dynamic mechanical testing, and the smallest DLIF is significantly larger than what was previously tested.

Therefore, the reported "device performance" in this context is that the new iterations of the device (titanium material and expanded PEEK sizes) are expected to perform equivalently or better than the previously cleared predicate devices for which performance was established, without the need for new mechanical testing to demonstrate this equivalence.

2. Sample size used for the test set and data provenance

No new test set was used for this submission. The device's substantial equivalence is based on comparisons to previously cleared predicate devices. Therefore, there is no new data provenance to report beyond what would have been established for the predicate devices.

3. Number of experts used to establish the ground truth for the test set and their qualifications

Not applicable. As no new test set was directly used for this submission, no new ground truth was established by experts for this specific 510(k). The equivalence is based on technical characteristics and previous clearances.

4. Adjudication method for the test set

Not applicable. No new test set requiring adjudication was performed for this 510(k) submission.

5. Multi-reader multi-case (MRMC) comparative effectiveness study

No MRMC comparative effectiveness study was done. This submission focuses on substantial equivalence based on material and size modifications to an existing device, rather than a comparative effectiveness study involving human readers and AI.

6. Standalone (algorithm only without human-in-the-loop performance) study

Not applicable. This device is an intervertebral body fusion device (physical implant), not a software algorithm or AI device.

7. Type of ground truth used

For the purpose of this 510(k) submission, the "ground truth" relies on the established performance and safety of the predicate devices (Cortina™ [MAX] Lumbar Cage System - PEEK version, K171914/K180431, and other referenced devices). This would have been established through a combination of mechanical testing, material characterization, and potentially clinical data (if required for the predicates) to demonstrate their safety and effectiveness.

8. Sample size for the training set

Not applicable. This device is a physical medical implant, not an AI or machine learning algorithm. Therefore, there is no "training set" in the context of data for model development.

9. How the ground truth for the training set was established

Not applicable. As there is no training set, there is no ground truth to establish for it in this context.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 25, 2019

NeuroStructures, Inc. % Meredith May Vice President Empirical Testing Corp. 4628 Northpark Drive Colorado Springs, Colorado 80918

Re: K192248

Trade/Device Name: Cortina™ [MAX] Lumbar Cage System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: MAX Dated: October 28, 2019 Received: October 29, 2019

Dear Meredith May:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you. however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for

Brent Showalter, Ph.D. Assistant Director (Acting) DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

4. INDICATIONS FOR USE STATEMENT

DEPARTMENT OF HEALTH AND HUMAN SERVICES

Food and Drug Administration

Indications for Use

510(k) Number (if known) K192248

Device Name Cortina™ [MAX] Lumbar Cage System

Indications for Use (Describe)

The Cortina™ [MAX] Lumbar Cage System is indicated for use in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1 whose condition requires the use of interbody fusion combined with supplemental fixation. The interior of the Cortina™ [MAX] Lumbar Cage System should be packed with autogenous bone graft (i.e. autograft). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per
response, including the time to review instructions, search existing data sources, gather
and maintain the data needed and complete and review the collection of information.
Send comments regarding this burden estimate or any other aspect of this information
collection, including suggestions for reducing this burden, to:

Department of Health and Human Services

Food and Drug Administration

Office of Chief Information Officer

Paperwork Reduction Act (PRA) Staff

PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to
respond to, a collection of information unless it displays a currently valid OMB
number."

FORM FDA 3881 (7/17)Page 1 of 1PSC Publishing Services (301) 443-6740 EF
------------------------------------------------------------------------------

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020

{3}------------------------------------------------

5.510(K) SUMMARY

Submitter's NameNeurostructures, Inc.
Submitter's Address199 Technology Drive, Suite 1110Irvine, CA 92618
Submitter's Telephone800-352-6103
Company Contact PersonMoti Altarac
Contact PersonMeredith Lee May MS, RACEmpirical Consulting719-337-7579Mmay@EmpiricalConsulting.com
Date Summary was Prepared12-Aug-19
Trade or Proprietary NameCortina™ [MAX] Lumbar Cage System
Common or Usual NameIntervertebral Fusion Device With Bone Graft, Lumbar
ClassificationClass II per 21 CFR §888.3080_DeviceClassification
Product CodeMAX
Classification PanelDivision of Orthopedic Devices

DESCRIPTION OF THE DEVICE SUBJECT TO PREMARKET NOTIFICATION

The Cortina™ [MAX] Lumbar Cage System is an intervertebral fusion device made from medical grade PEEK with tantalum markers manufactured from EVONIK PEEK as described in MAF-1922 with the submissions of the predicate Cortina [MAX] (K171914 and K180431) or medical grade titanium per ASTM F136 with a TECOTEX® surface coating. The subject device is offered in a variety of styles and sizes to accommodate various patient anatomies. This submission is intended to add the titanium implants and expand the size offering to include larger footprints.

INDICATIONS FOR USE

The Cortina™ [MAX] Lumbar Cage System is indicated for use in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1 whose condition requires the use of interbody fusion combined with supplemental fixation. The interior of the Cortina™ [MAX] Lumbar Cage System should be packed with autogenous bone graft (i.e. autograft). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment.

TECHNOLOGICAL CHARACTERISTICS

The subject and predicate devices have nearly identical technological characteristics and the minor differences do not raise any new issues of safety and effectiveness. Specifically the following characteristics are identical between the subject and predicates:

{4}------------------------------------------------

  • Principles of Operation ●
  • Indications for Use ●
  • Implant Materials
  • Implant Sizes ●
  • Surgical Approach ●
510kNumberTrade or Proprietary or ModelNameManufacturerPredicateType
K171914/K180431Cortina™ [MAX] Lumbar CageSystemNeuroStructuresPrimary
K172064Ti-Diagon Oblique TLIFCamber Spine TechnologiesAdditional
K173080IVA CAGE Ti (ACIF, PLIF,TLIF, DLIF and ALIF)HUVEXELAdditional
K172320Cavetto® Cervical Cage SystemNeuroStructures, Inc.Reference
K142041PorOsteon Phusion MetalCervical CagePorOsteon, Inc.Reference

Table 5-1: Predicate Devices

PERFORMANCE DATA

The subject devices that are manufactured from titanium require no additional testing, for they are identical in form, shape, and function to the previously cleared devices manufactured from PEEK (K171914/K180431). Titanium material produces implants with a higher yield load and stiffness when the shape is identical, therefore no testing is required.

The subject devices that are manufactured from PEEK present no new worst case to what was to what was previously cleared (K171914/K180431). The addition of the longer (36-42 mm) PLIF implants does not create a new mechanical worst-case for the system as opposed to the 20x9x7mm PLIF tested for the previous clearance , therefore no additional testing is required for the additional length in the line extension.

The DLIF does not need to be tested as a new worst-case for the system, as the design and footprint do not represent a new worst-case for either static or dynamic mechanical testing. The smallest DLIF is significantly larger than what was previously tested for clearance. Additionally, the design of the DLIF is considerably simpler than the tested PLIF and has fewer areas of concern for mechanical integrity.

CONCLUSION

The overall technology characteristics and mechanical performance data lead to the conclusion that the Cavetto® [MAX] Cervical Cage System is substantially equivalent to the predicate device.

§ 888.3080 Intervertebral body fusion device.

(a)
Identification. An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.(b)
Classification. (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
Date premarket approval application (PMA) or notice of product development protocol (PDP) is required. Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.