K Number
K180431

Validate with FDA (Live)

Date Cleared
2018-03-22

(34 days)

Product Code
Regulation Number
888.3080
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Cortina™ [MAX] Lumbar Cage System is indicated for use in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 lo S 1 whose condition requires the use of interbody fusion combined with supplemental fixation. The interior of the Cortina™ [MAX] Lumbar Cage System should be packed with autogenous bone graft (i.e. autograft). DDD is defined as back pain of discogenic origin with degeneration of the disc confim1ed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment.

Device Description

The Cortina™ [MAX] Lumbar Cage System is an intervertebral fusion device made from medical grade PEEK with titanium markers. The subject device is offered in a variety of styles and sizes to accommodate various patient anatomies.

AI/ML Overview

Here's a breakdown of the requested information based on the provided document:

1. Table of Acceptance Criteria and Reported Device Performance

Test ModeAcceptance Criteria (from ASTM standards)Reported Device Performance (as stated in document)
Static axial compression (per ASTM F2077-11)Not specified in the document; implied to meet the standard's requirements for devices of this type."the strength of the Cortina™ (MAX) Lumbar Cage System is sufficient for its intended use"
Dynamic axial compression (per ASTM F2077-11)Not specified in the document; implied to meet the standard's requirements for devices of this type."the strength of the Cortina™ (MAX) Lumbar Cage System is sufficient for its intended use"
Static subsidence (per ASTM F2267-04)Not specified in the document; implied to meet the standard's requirements for devices of this type."the strength of the Cortina™ (MAX) Lumbar Cage System is sufficient for its intended use"
Static expulsion (per ASTM DRAFT F-04.25.02.02)Not specified in the document; implied to meet the standard's requirements for devices of this type."the strength of the Cortina™ (MAX) Lumbar Cage System is sufficient for its intended use"

Note: The document only provides a general statement that the device's strength is sufficient and substantially equivalent to predicate devices, rather than specific numerical performance metrics against pre-defined acceptance criteria. The acceptance criteria are referenced implicitly by stating adherence to ASTM standards.

2. Sample Size Used for the Test Set and Data Provenance

The document describes non-clinical performance testing of mechanical properties. It does not refer to a "test set" in the context of patient data or clinical studies. Therefore, information regarding sample size for a test set, country of origin, or retrospective/prospective nature is not applicable to this submission, as it focuses on device characteristics, not clinical performance.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

This information is not applicable as the document describes non-clinical mechanical testing, not a study involving human interpretation or clinical ground truth.

4. Adjudication Method for the Test Set

This information is not applicable for the same reasons as above.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, an MRMC comparative effectiveness study was not done. The document focuses on non-clinical mechanical performance.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

This information is not applicable as the device is a physical medical implant (lumbar cage system), not an algorithm or an AI-powered system.

7. The Type of Ground Truth Used

The "ground truth" for the performance data in this document is established through physical testing against recognized mechanical engineering standards (ASTM standards). The document states that the testing validates the "strength" of the device.

8. The Sample Size for the Training Set

This information is not applicable as the document describes a physical medical device and its mechanical testing, not a machine learning model requiring a training set.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable for the same reasons as above.

Summary of the Study:

The provided document (K180431) describes a 510(k) premarket notification for the Cortina™ [MAX] Lumbar Cage System. The submission is primarily for a name change from the Cortina™ (K171914) to Cortina™ [MAX], with no changes to the device itself.

The "study" referenced in the document is a series of non-clinical performance tests designed to assess the mechanical strength and stability of the lumbar cage system. These tests were conducted according to established ASTM standards:

  • Static axial compression per ASTM F2077-11
  • Dynamic axial compression per ASTM F2077-11
  • Static subsidence per ASTM F2267-04
  • Static expulsion per ASTM DRAFT F-04.25.02.02

The acceptance criteria for these tests are implicitly the requirements of the respective ASTM standards for intervertebral body fusion devices. The reported device performance is a qualitative statement that "the strength of the Cortina™ (MAX) Lumbar Cage System is sufficient for its intended use and is substantially equivalent to legally marketed predicate devices."

This document does not describe a clinical study, an AI/algorithm-based device, or a study involving human readers or patient data for performance evaluation. It is solely focused on demonstrating the physical and mechanical equivalence of the device to its predicate through engineering testing.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

March 22. 2018

NeuroStructures, Inc. Kathleen Wong Director of RA & QA 16 Technology Drive. Suite 165 Irvine, California 92618

Re: K180431

Trade/Device Name: Cortina™ [MAX] Lumbar Cage System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: MAX Dated: February 15, 2018 Received: February 16, 2018

Dear Ms. Wong:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to Mav 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);

{1}------------------------------------------------

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K180431

Device Name Cortina™ [MAX] Lumbar Cage System

Indications for Use (Describe)

The Cortina™ [MAX] Lumbar Cage System is indicated for use in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 lo S 1 whose condition requires the use of interbody fusion combined with supplemental fixation. The interior of the Cortina™ [MAX] Lumbar Cage System should be packed with autogenous bone graft (i.e. autograft). DDD is defined as back pain of discogenic origin with degeneration of the disc confim1ed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment.

Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Submitter's NameNeurostructures, Inc.
Submitter's Address16 Technology Drive, Suite 165Irvine, CA 92618
Company Contact PersonKathleen Wongkw@neurostructures.com949.370.4497
Contact PersonKathleen Wongkw@neurostructures.com949.370.4497
Date Summary was Prepared15 February 2018
Trade or Proprietary NameCortina™ [MAX] Lumbar Cage System
Common or Usual NameIntervertebral Fusion Device With Bone Graft, Lumbar
ClassificationClass II per 21 CFR §888.3050 Device Classification
Product CodeMAX
Classification PanelDivision of Orthopedic Devices

K180431 - 510(K)SUMMARY

DESCRIPTION OF THE DEVICE SUBJECT TO PREMARKET NOTIFICATION

The Cortina™ [MAX] Lumbar Cage System is an intervertebral fusion device made from medical grade PEEK with titanium markers. The subject device is offered in a variety of styles and sizes to accommodate various patient anatomies.

This 510(k) is submitted only for the purposes of changing the name of the Cortina™ (K171914) to the Cortina™ MAX). The two devices are otherwise identical, and no changes whatsoever have been made to the Cortina™ (K171914 ).

INDICATIONS FOR USE

The Cortina™ [MAX] Lumbar Cage System is indicated for use in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1 whose condition requires the use of interbody fusion combined with supplemental fixation. The interior of the Cortina™ [MAX] Lumbar Cage System should be packed with autogenous bone graft (i.e. autograft). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment.

The indications for use for the Cortina™ [MAX] Lumbar Cage System is identical to that of the primary predicate and similar to that of the additional predicates noted in Table 5-1: Predicate Devices.

{4}------------------------------------------------

K180431 Page 2 of 2

TECHNOLOGICAL CHARACTERISTICS

The subject and primary predicate device are identical in technological characteristics. The subject and additional predicate devices have nearly identical technological characteristics and the minor differences do not raise any new issues of safety and effectiveness. Specifically, the following characteristics are identical between the subject and primary predicate:

  • Principles of Operation
  • Indications for Use
  • Implant Materials
  • Implant Sizes ●
  • Surgical Approach

Table 5-1: Predicate Devices

510k NumberTrade or Proprietary or Model NameManufacturerPredicate Type
K171914Cortina Lumbar Cage SystemNeuroStructuresPrimary

PERFORMANCE DATA

The Cortina™ [MAX] Lumbar Cage System has been tested in the following test modes:

  • . Static axial compression per ASTM F2077-11
  • Dynamic axial compression per ASTM F2077-11 .
  • Static subsidence per ASTM F2267-04 ●
  • . Static expulsion per ASTM DRAFT F-04.25.02.02

The results of this non-clinical testing show that the strength of the Cortina™ (MAX) Lumbar Cage System is sufficient for its intended use and is substantially equivalent to legally marketed predicate devices.

CONCLUSION

The overall technology characteristics and mechanical performance data lead to the conclusion that the Cortina™ [MAX] Lumbar Cage System is substantially equivalent to the predicate devices.

§ 888.3080 Intervertebral body fusion device.

(a)
Identification. An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.(b)
Classification. (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
Date premarket approval application (PMA) or notice of product development protocol (PDP) is required. Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.