K Number
K192183
Manufacturer
Date Cleared
2020-04-23

(255 days)

Product Code
Regulation Number
876.5130
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The RocaTub ureteral catheter is intended for use during ureteral catheterization for drainage, opacification of the upper urinary tract (retrograde ureteropyelography) and flushing procedures for diagnostics or interventional endourology procedures (stone management, stricture management) on adults.

Device Description

The Rocamed RocaTub Catheters consist of a flexible tube, tapered, perforated and with position marks. It is radio-opaque and hydrophilic coated. The device includes a stylet for easing insertion into the patient if needed and a connector for syringe for flushing procedures. The device is offered in various sizes, from 4 to 7 Fr in 80cm length, and with various tip shapes.

AI/ML Overview

The provided text is a 510(k) summary for the RocaTub Ureteral Catheter, a medical device. It focuses on demonstrating substantial equivalence to a predicate device rather than presenting a study of the device's performance against specific acceptance criteria in the manner typically seen for AI/ML-driven diagnostic devices.

Therefore, for this specific submission, many of the requested fields regarding AI/ML study design are not applicable.

Here's an analysis based on the provided document:

1. Table of Acceptance Criteria and Reported Device Performance:

The document does not explicitly state quantitative acceptance criteria for device performance in the form of a table that would be common for diagnostic algorithms (e.g., accuracy, sensitivity, specificity). Instead, it focuses on demonstrating that the device meets "necessary specifications" through bench testing and compliance with recognized standards.

Acceptance Criteria (Implicit)Reported Device Performance
Mechanical Strength for Intended UseAll bench testing results confirmed that the products met the necessary specifications. (Specific details of "necessary specifications" and exact performance values are not provided in this summary, but are implied to be met).
Radiopacity (per ASTM F640-12)Performance testing done according to ASTM F640-12. (Implies compliance, specific results not provided).
Biocompatibility (per ISO 10993)Confirmed in accordance with ISO 10993. An analysis of material sheets/safety data sheets has been performed. (Implies compliance, specific results not provided).
Sterilization Adoption (per recognized industry standards)Sterilization adoption performed in accordance with recognized industry standards. (Implies compliance, specific results not provided).
Shelf LifeValidated shelf life of 2 years.
Safety and Effectiveness"The results of these evaluations demonstrate that the RocaTub Ureteral Catheter are safe and effective when used in accordance with their intended use and labeling." "The RocaTub Ureteral Catheter are safe and effective as the corresponding predicate device when used in accordance with their intended use and labeling." (This is a summary conclusion based on all testing and comparison to predicate, not a specific performance metric).

2. Sample Size Used for the Test Set and Data Provenance:

This information is not applicable in the context of this 510(k) summary, as it describes a non-AI/ML medical device (a ureteral catheter). The "test set" for this type of device refers to units of the manufactured product subjected to various bench tests, not a dataset of patient cases or images. The provenance would be the manufacturing facility.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

This is not applicable. Ground truth, in the context of an AI/ML diagnostic device, refers to definitive diagnoses often established by expert clinicians or pathological findings. For a physical medical device like a catheter, "ground truth" relates to material properties, design specifications, and functional performance measured through standardized tests, not expert interpretation of cases.

4. Adjudication Method for the Test Set:

This is not applicable. Adjudication methods like 2+1 or 3+1 are used in studies involving expert interpretation discrepancies, typically for diagnostic imaging or clinical assessment. This device's evaluation relies on objective measurements from bench testing.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

This is not applicable. The RocaTub Ureteral Catheter is a physical medical device, not an AI-assisted diagnostic tool.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

This is not applicable. This is not an algorithm.

7. The Type of Ground Truth Used:

For this device, the "ground truth" is established through:

  • Standardized Test Methods: Adherence to recognized industry standards (e.g., EN 1618:1997 for catheter properties, ASTM F640-12 for radiopacity, ASTM D412A:2015 for polymeric geomembranes).
  • Biocompatibility Testing: Conducted in accordance with ISO 10993.
  • Validated Manufacturing Processes: Sterilization validation and shelf-life validation.
  • Design Specifications: The inherent design parameters of the catheter (e.g., diameter, length, material, tip shape).

8. The Sample Size for the Training Set:

This is not applicable. This device does not involve a "training set" in the context of AI/ML.

9. How the Ground Truth for the Training Set Was Established:

This is not applicable. This device does not involve a "training set" in the context of AI/ML.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

April 23, 2020

Promepal Sam Mohamed Rekik QRA Manager 9 Avenue Albert II Monaco, 98000 MONACO

Re: K192183

Trade/Device Name: RocaTub Ureteral Catheter Regulation Number: 21 CFR 876.5130 Regulation Name: Urological Catheter and accessories Regulatory Class: Class II Product Code: EYB Dated: March 13, 2020 Received: March 23, 2020

Dear Mohamed Rekik:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

for Purva Pandya Acting Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES

Indications for Use

510(k) Number (if known) K192183

Device Name RocaTub Ureteral catheter

Indications for Use (Describe)

The RocaTub ureteral catheter is intended for use during ureteral catheterization for drainage, opacification of the upper urinary tract (retrograde ureteropyelography) and flushing procedures or interventional endourology procedures (stone management, stricture management) on adults.

Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (7/17) PSC Publishing Services (301) 443-6740 EF

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

{3}------------------------------------------------

K192183 Page 1 of 4

Submitter: Promepla SAM RocaTub Ureteral catheter Traditional 510(k)

510 (k) Summary

A. Submitter Information

Submitter's Name:PROMEPLA SAM
Address9 Avenue Albert II98000 MonacoMONACO (Principality of)
Contact PersonMohamed REKIKQRA Manager
Contact Person's email:mr@promepla.com
Contact Person's Number(377) 979-842-44
Contact Person's Fax(377) 920-561-50
Submission dateApril 17th, 2020

B. Device Name

Trade Name of the Device:RocaTub Ureteral Catheter
Common Name:Catheter, Ureteral, Gastro-Urology
Classification Name:Urological catheter and accessories
Device Class:2
Panel:Gastroenterology/Urology
Product CodeEYB
Classification Regulation21 CFR 876.5130
Official Contact personMohamed REKIK

C. Predicate Device

Product nameManufacturer510(k) number
1Ureteric cathetersColoplast corpK182122

D. Device Description:

The Rocamed RocaTub Catheters consist of a flexible tube, tapered, perforated and with position marks. It is radio-opaque and hydrophilic coated. The device includes a stylet for easing insertion into the patient if needed and a connector for syringe for flushing procedures. The device is offered in various sizes, from 4 to 7 Fr in 80cm length, and with various tip shapes.

ReferenceDiameter(Fr)Length(cm)Designation of the product
ROTA4200ST480ROCATUB CATHETER - 2 EYES - Straight/Open distal tip - 80 cm - 4Fr

{4}------------------------------------------------

ROTA5200ST580ROCATUB CATHETER - 2 EYES - Straight/Open distal tip - 80 cm - 5Fr
ROTA6200ST680ROCATUB CATHETER - 2 EYES - Straight/Open distal tip - 80 cm - 6Fr
ROTA7200ST780ROCATUB CATHETER - 2 EYES - Straight/Open distal tip - 80 cm - 7Fr
ROTG4000ST480ROCATUB CATHETER - NO EYES - Straight/Open distal tip - 80 cm - 4Fr
ROTG5000ST580ROCATUB CATHETER - NO EYES - Straight/Open distal tip - 80 cm - 5Fr
ROTG6000ST680ROCATUB CATHETER - NO EYES - Straight/Open distal tip - 80 cm - 6Fr
ROTG7000ST780ROCATUB CATHETER - NO EYES - Straight/Open distal tip - 80 cm - 7Fr
ROTH4000ST480ROCATUB CATHETER - NO EYES - Curved/Open distal tip - 80 cm - 4Fr
ROTH5000ST580ROCATUB CATHETER - NO EYES - Curved/Open distal tip - 80 cm - 5Fr
ROTH6000ST680ROCATUB CATHETER - NO EYES - Curved/Open distal tip - 80 cm - 6Fr
ROTH7000ST780ROCATUB CATHETER - NO EYES - Curved/Open distal tip - 80 cm - 7Fr

E. Indications for Use:

The RocaTub ureteral catheter is intended for use during ureteral catheterization for drainage, opacification of the upper urinary tract (retrograde ureteropyelography) and flushing procedures for diagnostics or interventional endourology procedures (stone management, stricture management) on adults.

F. Summary of Non-Clinical Performance Testing:

In support of this 510(k) premarket notification, Promepla SAM has conducted bench testing to demonstrate that the RocaTub Ureteral Catheter provide adequate mechanical strength for their intended use.

All bench testing results confirmed that the products described in this submission met the necessary specification. Performance testing has been done according to FDA-recognized standards or guidance document: EN 1618:1997 Catheters Other than Intravascular Catheters - Test Methods For Common Properties, ASTM F640-12 Standard Test Methods for Determining Radiopacity for Medical Use, ASTM D412A:2015 A quide to Polymeric Geomembranes: A Practical Approach.

In addition, the biocompatibility of the devices has been confirmed in accordance with ISO 10993, and the company has conducted sterilization adoption in accordance with recognized industry standards. The RocaTub ureteral catheter has a validated shelf life of 2 years.

A list of the tests performed to support substantial equivalence is provided below:

  • -Sterilization Validation;
  • -Biocompatibility;
  • -Device verification and validation
  • Transportation adoption Validation; -
  • Shelf life adoption Validation. -

The results of these evaluations demonstrate that the RocaTub Ureteral Catheter are safe and effective when used in accordance with their intended use and labeling.

{5}------------------------------------------------

G. Comparison of Technological Characteristics:

The RocaTub Ureteral Catheters are substantially equivalent to Coloplast currently marketed the Ureteric Catheters, regarding to the intended use, design and sterilization process.

Product NameRocaTub UreteralCatheterUreteric Catheters
510(k) NumberK192183K182122
Product CodeEYBEYB
Regulation NameUreteral CatheterUreteral Catheter
ManufacturerPROMEPLA SAMColoplast Corp
Intended UseRocaTub catheters areintended for use duringureteral catherization,drainage, opacification ofthe upper urinary tract(retrogradeureteropyelography) andflushing proceduresThe Ureteric Catheters forretrogradeureteropyelography areintended for injection ofcontrast medium or salineduring endourologicalprocedures
DiameterFrom 4 Fr to 7 FrFrom 5 Fr to 7Fr
DisposableYESYES
SterileYESYES
Stylet for ease of insertionYESYES
Suitable for use with aguidewireYESYES
Connection with a syringeYESYES
MaterialThermoplastic PolyurethanePolyether block amide
Tip shapeStraight or open distal tipStraight or open distal tip
Side eyesWith or withoutWith or without
Ink MarksYESYES
Hydrophilic coatingYESNO
Type of sterilizationEthylene oxideEthylene oxide

{6}------------------------------------------------

Submitter: Promepla SAM RocaTub Ureteral catheter Traditional 510(k)

*Green Boxes means that the characteristics are the same between the subject device and the predicate.

H. Conclusion

Promepla SAM has demonstrated that the proposed RocaTub Ureteral Catheter are substantially equivalent to Coloplast currently marketed the Ureteric Catheters, cleared under premarket notification number K182122. The differences do not affect the performances and the safety of the patient as biocompatibility tests have been performed to ensure that the raw material used are biocompatible. An analysis of material sheets/safety data sheets has been performed as well as a risk analysis. Plus, the RocaTub Ureteral Catheter have a hydrophilic coating which when activated allow to insert more easily the device into the patient than the predicate device. So, the RocaTub Ureteral Catheter are safe and effective as the corresponding predicate device when used in accordance with their intended use and labeling.

§ 876.5130 Urological catheter and accessories.

(a)
Identification. A urological catheter and accessories is a flexible tubular device that is inserted through the urethra and used to pass fluids to or from the urinary tract. This generic type of device includes radiopaque urological catheters, ureteral catheters, urethral catheters, coudé catheters, balloon retention type catheters, straight catheters, upper urinary tract catheters, double lumen female urethrographic catheters, disposable ureteral catheters, male urethrographic catheters, and urological catheter accessories including ureteral catheter stylets, ureteral catheter adapters, ureteral catheter holders, ureteral catheter stylets, ureteral catheterization trays, and the gastro-urological irrigation tray (for urological use).(b)
Classification. (1) Class II (performance standards).(2) Class I for the ureteral stylet (guidewire), stylet for gastrourological catheter, ureteral catheter adapter, ureteral catheter connector, and ureteral catheter holder. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.