(102 days)
No
The description focuses on the optical and mechanical components and image acquisition process. There is no mention of AI, ML, or advanced image processing techniques that would typically indicate the use of such technologies for analysis or interpretation. The performance study compares image quality to a predicate device, not the performance of an AI/ML algorithm.
No
This device is indicated for imaging the retina, which is a diagnostic purpose, not a therapeutic one. Therapeutic devices are used for treating diseases or conditions.
Yes
The device is indicated for color imaging of a human retina, and the clinical performance data section describes comparing its images to a predicate device for normal subjects and subjects with retinal pathologies, indicating its use in assessing the retina.
No
The device description explicitly lists multiple hardware components (optical head, patient forehead rest, display, base, stand, illumination system with LEDs, imaging system with CMOS camera, anterior segment alignment system with cameras) and describes their physical interaction with the patient and the environment to acquire images. While it runs dedicated software, it is fundamentally a hardware device with integrated software.
Based on the provided information, the CenterVue DRSplus is not an In Vitro Diagnostic (IVD) device.
Here's why:
- IVD Definition: In Vitro Diagnostic devices are used to examine specimens (like blood, urine, or tissue) taken from the human body to provide information for diagnosis, monitoring, or screening.
- DRSplus Function: The DRSplus is a scanning ophthalmoscope that directly images the human retina in vivo (within the living body). It does not analyze specimens taken from the body.
- Intended Use: The intended use is for "color imaging of a human retina," which is a direct imaging procedure, not an in vitro analysis.
Therefore, the CenterVue DRSplus falls under the category of medical imaging devices, not IVDs.
N/A
Intended Use / Indications for Use
The CenterVue DRSplus is a confocal scanning ophthalmoscope indicated for color imaging of a human retina without the use of a mydriatic agent.
Product codes
MYC
Device Description
The DRSplus is a scanning ophthalmoscope which uses infrared and white light to obtain confocal images of the retina, without pharmacological dilation.
The DRSplus operates as a standalone unit, running a dedicated software application, is intended for prescription use only, and includes:
-
- an optical head;
- a patient forehead rest; 2.
- a displav: 3.
-
- a base;
-
- a stand.
The CenterVue DRSplus operates based on the following principles:
- An illumination system consisting of infrared (IR) LEDs, white LEDs and a green LED a) illuminates the patient eye with the following functions:
- · the IR LED allows the capture of IR photos, which are used for alignment and focusing purposes. The patient's retina is uniformly illuminated by a line in the horizontal direction. Along the optical path there is an oscillating mirror which scans the line in order to illuminate the retina.
- · Two IR LEDs are seen from the eye in a free viewing system. The two LEDs are equally shifted with respect to the machine's optical axis. The LEDS are switched on during all exams in order to enable pupil tracking.
- The white LEDs allow the capture of color photos. The retina is uniformly illuminated by a line in the horizontal direction. Along the optical path an oscillating mirror scans the line in order to illuminate the retina.
- · The green LED is used as fixation target.
- An imaging system collects back-reflected light from the retina and creates a highb) resolution image. A focusing lens is included in the imaging path to achieve optimal retinal focusing on a CMOS camera.
- An anterior seqment alignment system is included, using two cameras and the two IR c) LEDs. The LEDs illuminate the anterior segment by diffusion, whereas the cameras allow a stereoscopic reconstruction of the pupil's position to be obtained with respect to the instrument's frontlens.
The DRSplus interacts with the patient by directing infrared, white (for imaging) and green (for fixation) illumination into the patient's eye. The only part of the only part of the device that contacts the patient.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
confocal images of the retina
Anatomical Site
human retina
Indicated Patient Age Range
Not Found
Intended User / Care Setting
Prescription Use (Part 21 CFR 801 Subpart D)
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies
The imaging properties of the DRSplus were checked by comparing color images captured using the DRSplus and the predicate device EIDON. Images from randomly selected normal subjects and subjects with retinal pathologies were acquired at two different sites and included in the comparison. The two devices were alternated and patients were imaged without pharmacological pupil dilation. The comparison showed that the DRSplus provides images that are similar to those of the mentioned predicate device.
Key Metrics
Not Found
Predicate Device(s)
Reference Device(s)
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 886.1570 Ophthalmoscope.
(a)
Identification. An ophthalmoscope is an AC-powered or battery-powered device containing illumination and viewing optics intended to examine the media (cornea, aqueous, lens, and vitreous) and the retina of the eye.(b)
Classification. Class II (special controls). The device, when it is an AC-powered opthalmoscope, a battery-powered opthalmoscope, or a hand-held ophthalmoscope replacement battery, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.
0
November 15, 2019
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue. Underneath the FDA logo is the word "ADMINISTRATION".
CenterVue SpA Roberto Gabriotti Quality and Regulatory Manager Via S. Marco 9H Padova, 35129 It
Re: K192113
Trade/Device Name: DRSplus Regulation Number: 21 CFR 886.1570 Regulation Name: Ophthalmoscope Regulatory Class: Class II Product Code: MYC Dated: October 7, 2019 Received: October 10, 2019
Dear Roberto Gabriotti:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
1
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Tieuvi Nguyen, Ph.D., Acting Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure
2
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/202 See PRA Statement below.
K192113 510(k) Number (if known)
Device Name
DRSplus
Indications for Use (Describe)
The CenterVue DRSplus is a confocal scanning ophthalmoscope indicated for color imaging of a human retina without the use of a mydriatic agent.
Type of Use (Select one or both, as applicable)
▽ Prescription Use (Part 21 CFR 801 Subpart D)
7 Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff @fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
3
510(K) SUMMARY
Regulatory information
Device Name: | DRSplus |
---|---|
Type of 510(k) submission: | Traditional |
Date summary prepared: | November 8th, 2019 |
Manufacturer: | CenterVue S.p.A. |
via San Marco 9h | |
35129 Padova Italy | |
510(k) Submitter and Contact: | Mr. Roberto Gabriotti |
Centervue S.p.A. | |
Manager of Quality and Regulatory Affairs | |
Via San Marco 9H | |
35129 Padova Italy | |
Phone: +39 049 501 8399 | |
Fax: +39 049 501 8398 | |
E-mail: roberto.gabriotti@centervue.com | |
Product Code: | MYC |
Regulation Number: Classification | 886.1570 |
Name: | Ophthalmoscope, Laser, Scanning |
Panel: | Ophthalmic |
FDA Classification: | Class II |
Indications for use: | The CenterVue DRSplus is a confocal scanning |
ophthalmoscope indicated for color imaging of a human | |
retina without the use of a mydriatic agent. |
4
Device description
The DRSplus is a scanning ophthalmoscope which uses infrared and white light to obtain confocal images of the retina, without pharmacological dilation.
The DRSplus operates as a standalone unit, running a dedicated software application, is intended for prescription use only, and includes:
-
- an optical head;
- a patient forehead rest; 2.
- a displav: 3.
-
- a base;
-
- a stand.
The CenterVue DRSplus operates based on the following principles:
- An illumination system consisting of infrared (IR) LEDs, white LEDs and a green LED a) illuminates the patient eye with the following functions:
- · the IR LED allows the capture of IR photos, which are used for alignment and focusing purposes. The patient's retina is uniformly illuminated by a line in the horizontal direction. Along the optical path there is an oscillating mirror which scans the line in order to illuminate the retina.
- · Two IR LEDs are seen from the eye in a free viewing system. The two LEDs are equally shifted with respect to the machine's optical axis. The LEDS are switched on during all exams in order to enable pupil tracking.
- The white LEDs allow the capture of color photos. The retina is uniformly illuminated by a line in the horizontal direction. Along the optical path an oscillating mirror scans the line in order to illuminate the retina.
- · The green LED is used as fixation target.
- An imaging system collects back-reflected light from the retina and creates a highb) resolution image. A focusing lens is included in the imaging path to achieve optimal retinal focusing on a CMOS camera.
- An anterior seqment alignment system is included, using two cameras and the two IR c) LEDs. The LEDs illuminate the anterior segment by diffusion, whereas the cameras allow a stereoscopic reconstruction of the pupil's position to be obtained with respect to the instrument's frontlens.
The DRSplus interacts with the patient by directing infrared, white (for imaging) and green (for fixation) illumination into the patient's eye. The only part of the only part of the device that contacts the patient.
5
Technical Specifications
Class and type of applied part
Class I, Type B (according to IEC 60601-1).
IP classification:
IP20 (according to the degree of protection provided by the enclosure with respect to harmful penetration of particulate matter or water).
lmage acquisition:
|--|
- Field of individual image: 45° (H) x 40° (V) captured in a single exposure
- lmage size: 3600 x 2910 pixels (10 MP)
- Light sources: infrared LED (825-870 nm), white LED (420-675 nm) ●
- lmaging modalities: ●
- color, red-free Working distance: 25 mm ●
- Pixel pitch: 3.7 microns ●
Other features:
-
Automatic operation: auto-alignment, auto-focus, auto-exposure, auto-capture
-
Focus adjustment range:
-
Internal fixation target:
-
Display: • Hard disk:
-
10 positions 10.1" multi-touch, color SSD, 512 GB
-15 D to +15 D
Dimensions:
- · Weight: 11 Kg (24 lbs)
- Size (W x H x D): W 300mm x H 450mm x D 650mm
Power supply:
- . Voltage: 12 V DC; Power consumption: 60 W
6
Predicate devices
The predicate device selected for comparison with the CenterVue DRSplus is identified as follows:
Predicate Device: | |
---|---|
Device Name: | EIDON |
Manufacturer: | CenterVue S.p.A. |
510(k) Number: | K142047 |
Product Code: | MYC |
Classification Name: | Ophthalmoscope, Laser, Scanning |
Regulation No: | 886.1570 |
Reference Device: | |
Device Name: | DRS |
Manufacturer: | CenterVue S.p.A. |
510(k) Number: | K101935 |
Product Code: | HKI |
Classification Name: | Camera, ophthalmic |
Regulation No: | 886.1120 |
Comparison of technological characteristics with the predicate device(s)
The following technological differences exist between the subject device and the predicate device:
Difference | Equivalence discussion |
---|---|
Infrared reflectance imaging is available | |
with the DRSplus only for the purpose of | |
alignment and focusing | No additional concerns for safety and effectiveness |
as this is a | |
Field of view is smaller | No additional concerns for safety and effectiveness |
as the subject device is equivalent to the reference | |
device in terms of field | |
Minimum pupil diameter is slightly larger | No additional concerns for safety and effectiveness |
as the subject device is equivalent to the reference | |
device in terms of minimum pupil size | |
Significantly smaller dimensions and | |
lighter weight | No additional concerns for safety and effectiveness |
Different technology is used to scan the | |
illumination light onto the retina | No additional concerns for safety and effectiveness |
Such differences are believed to have no effect on the safety and effectiveness of the device.
1 Neither the EIDON nor the DRSplus contain lasers but rather use LEDs for confocal imaging
7
Performance data - Bench
The following performance data were provided in support of the substantial equivalence determination.
Electrical safety and electromagnetic compatibility (EMC) The device complies with the EN 60601-1:2006 and IEC 60601-1-2:2015 standards.
Software Verification and Validation Testing
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "moderate" level of concern, since a failure or latent flaw in the software could directly result in minor injury. The software also complies with the IEC 62304 standard for software life cycle processes.
Fundus cameras testing
The device complies with the ISO 10940:2009 standard for fundus cameras.
Light hazard testing
The device complies with the ISO 15004-1:2006 standard for ophthalmic instruments and with the ANSI Z80.36-2016 standard for light hazard protection.
Performance data - Clinical
The imaging properties of the DRSplus were checked by comparing color images captured using the DRSplus and the predicate device EIDON. Images from randomly selected normal subjects and subjects with retinal pathologies were acquired at two different sites and included in the comparison. The two devices were alternated and patients were imaged without pharmacological pupil dilation. The comparison showed that the DRSplus provides images that are similar to those of the mentioned predicate device.
Conclusions
Based on the information contained within this submission, it is concluded that the CenterVue DRSplus is substantially equivalent to the identified predicate devices already in interstate commerce within the USA, and that any differences that do exist have no effect on the safety and effectiveness of the device.