K Number
K191955
Date Cleared
2020-08-05

(380 days)

Product Code
Regulation Number
870.5225
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The OM-A device is intended for the treatment of chronic stable angina that is refractory to optimal anti-anginal medical therapy and without options for revascularization. It is intended for use in healthy patients to provide improvement in vasodilation, and increased blood flow. It is intended for use under the oversight of a healthcare professional.

Device Description

Enhanced External Counterpulsation Device Plus OM-A is a computer-controlled system that applies external pressure via cuffs to the patient's lower extremities in synchronization with the patient's cardiac cycle. When the heart is in its relaxed state (diastole), pressure is applied sequentially; distal to proximal, from the lower legs (calves) to the lower thighs and then the upper thighs and buttocks, to propel blood back to the heart. The consequence is an increase in arterial blood pressure during diastole (diastolic augmentation) resulting in increased coronary perfusion pressure and coronary blood flow. Compression of the extremities also results in an increase in venous return to the heart. Just before the heart ejects blood (systole), air is released rapidly from all the cuffs simultaneously to release the externally applied pressure, allowing the compressed vessels to recover to their normal shape, thereby reducing vascular impedance. As a result, arterial pressure during systole is reduced (systolic unloading), as is cardiac workload. The patient's calves, lower thighs and buttocks are wrapped with cuffs containing air bladders. The patient's ECG is monitored via conventional, high-quality electrodes and detection of the R-wave is used to signal the System when to command sequential inflation of the cuffs. The start and duration of inflation, as well as the start of deflation, are adjustable by the Operator, within limits determined by the System. An algorithm is used to prevent the start of inflation during ejection of blood from the heart and to end inflation and begin deflation prior to the occurrence of the next heartbeat. The pressure applied by the cuffs is also adjustable by the Operator from 150mmHg to 300mmHg (for a patient with a heart rate of 60 bpm). The device also utilizes the following cleared components: ECG/EKG and SpO2: K123711

AI/ML Overview

The provided document is a 510(k) summary for the "Enhanced External Counter Pulsation Device Plus Omay-A." It focuses on demonstrating substantial equivalence to a predicate device rather than presenting a study proving a new device meets specific performance acceptance criteria for an AI or imaging diagnostic tool.

Therefore, many of the requested categories for describing an AI/imaging device study (e.g., acceptance criteria for diagnostic performance, sample size for test sets, expert ground truth establishment, MRMC studies, standalone performance, training set details) are not applicable to this document.

This document describes a medical device called an "Enhanced External Counter Pulsation Device," which is a therapy device, not a diagnostic AI or imaging device. The performance data provided are related to safety and efficacy standards (biocompatibility, electrical safety, EMC), and comparative functional features, not diagnostic performance metrics.

Specifically addressing your points based on the provided text:

1. A table of acceptance criteria and the reported device performance

  • Acceptance Criteria: The document does not define specific "acceptance criteria" in terms of diagnostic performance metrics (e.g., sensitivity, specificity, AUC) because it is a therapeutic device, not a diagnostic tool utilizing AI or advanced image analysis. The "acceptance" here is for market clearance based on substantial equivalence to an existing predicate device, primarily by meeting safety and performance standards (IEC 60601-1, IEC 60601-1-2) and demonstrating similar functional characteristics.
  • Reported Device Performance:
    • Biocompatibility: Passed Cytotoxicity, Sensitization, and Irritation tests according to ISO 10993-5 and ISO 10993-10.
    • Non-clinical (Safety & EMC): Passed tests according to IEC 60601-1-2005+CORR.1:2006+CORR.2:2007+A1:2012 (General requirements for basic safety and essential performance) and IEC 60601-1-2:2014 (Electromagnetic compatibility).
    • Functional Comparison to Predicate (K130439 P-ECP/TI):
      • Triggering Mechanism: Same (R-Wave trigger)
      • Microprocessor: Same (Windows Based)
      • Emergency System Power-down: Same (Red)
      • Pressure Setting: Subject device range (5mmHg to 350mmHg, Error: ± 10mmHg) is within predicate device range (1kPa~50kPa, Error: ± 2kPa)
      • Treatment Time: Subject device range (1min45min) is within predicate device range (1min60min)
      • Cuff System: Same (Three parts: calf, thigh, buttocks)
      • Major Components: Same (Base Unit, Air-Tubes, ECG/SPO2, three Cuffs)
      • Operating Environment: Same (10°C 30°C; Relative humidity: <85%; Atmospheric pressure: 70kPa106kPa)
      • Safety and EMC Standards: Same (IEC 60601-1, IEC 60601-1-2)
      • Biocompatibility Standards: Same (ISO 10993-5, ISO 10993-10)

2. Sample size used for the test set and the data provenance

  • Not applicable. This is a therapeutic device clearance, not a diagnostic AI study. The "test sets" here refer to samples for electrical safety, EMC, and biocompatibility testing, not patient data for diagnostic performance evaluation.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • Not applicable. Ground truth as understood in diagnostic AI is irrelevant here. The ground truth for electrical safety or biocompatibility is derived from established international standards and validated testing methodologies.

4. Adjudication method for the test set

  • Not applicable.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done

  • No. This type of study is for evaluating human reader performance with and without AI assistance, which is not relevant for this therapeutic device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • No. This concept applies to diagnostic algorithms. The device operates under the oversight of a healthcare professional as a therapeutic tool.

7. The type of ground truth used

  • Not applicable in the context of diagnostic performance. The "truth" for this device's clearance is its compliance with recognized safety and performance standards for medical devices and its substantial equivalence to an existing predicate device based on functional and safety characteristics.

8. The sample size for the training set

  • Not applicable. This device is not an AI model that requires a training set in the machine learning sense.

9. How the ground truth for the training set was established

  • Not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

August 5, 2020

Omay(Guangzhou)Med Technologies Co., Ltd. % Kevin Wang Consultant Chonconn Medical Device Consulting Co., Ltd. No. A415, Block A, NanShan Medical Devices Industrial Park Nanshan District Shenzhen. 518067 Cn

Re: K191955

Trade/Device Name: Enhanced External Counter Pulsation Device Plus Omay-A Regulation Number: 21 CFR 870.5225 Regulation Name: External counter-pulsating device Regulatory Class: Class II Product Code: DRN Dated: July 6, 2020 Received: July 6, 2020

Dear Kevin Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

for Fernando Aguel Assistant Director DHT2B: Division of Circulatory Support, Structural and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K191955

Device Name

Enhanced External Counter Pulsation Device Plus Omay-A

Indications for Use (Describe)

The OM-A device is intended for the treatment of chronic stable angina that is refractory to optimal anti-anginal medical therapy and without options for revascularization. It is intended for use in healthy patients to provide improvement in vasodilation, and increased blood flow. It is intended for use under the oversight of a healthcare professional.

Type of Use (Select one or both, as applicable)
-------------------------------------------------
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(K) Summary

Prepared in accordance with the requirements of 21 CFR Part 807.92

Prepared Date: 2019/11/05

1. Submission sponsor

Name: Omay (Guangzhou) Med Technologies Co., Ltd. Address: Room 302, 3F, No. 600-1, Guangshan 2nd Road, Tianhe Dist. Guangzhou, China. Contact person: Peixian Su Title: General Manager E-mail: omay@eecpcn.com Tel: +86-020-37092519

2. Submission correspondent

Name: Chonconn Medical Device Consulting Co., Ltd. Address: No. A415, Block A, NanShan Medical devices Industrial Park Nanshan District, Shenzhen, Guangdong, P.R. China 518067 Contact person: Kevin Wang E-mail: kevin@chonconn.com

Tel: +86-755 33941160

Trade/Device NameEnhanced External Counter Pulsation Device Plus Omay-A
ModelOM-A
Common NameExternal Counter-Pulsation Device
Regulatory ClassClass II
Classification870.5225 / Device, Counter-Pulsating, External / DRN
Submission typeTraditional 510(K)

3. Subject Device Information

Predicate Device 4.

  1. Primary Predicate Device Sponsor: Chongqing Psk Sci-tech Development Co., Ltd. Trade/Device Name: External Counterpulsation Device with SPO2 Monitoring Model: P-ECP/TI 510(k) #: K130439

  2. Reference Device Sponsor: Vamed Medical Instrument Co., Ltd. Device Name: External Counterpulsation System Model: ECP-MC3 510(k) #: K190683

5. Device Description

{4}------------------------------------------------

Enhanced External Counterpulsation Device Plus OM-A is a computer-controlled system that applies external pressure via cuffs to the patient's lower extremities in synchronization with the patient's cardiac cycle. When the heart is in its relaxed state (diastole), pressure is applied sequentially; distal to proximal, from the lower legs (calves) to the lower thighs and then the upper thighs and buttocks, to propel blood back to the heart. The consequence is an increase in arterial blood pressure during diastole (diastolic augmentation) resulting in increased coronary perfusion pressure and coronary blood flow. Compression of the extremities also results in an increase in venous return to the heart. Just before the heart ejects blood (systole), air is released rapidly from all the cuffs simultaneously to release the externally applied pressure, allowing the compressed vessels to recover to their normal shape, thereby reducing vascular impedance. As a result, arterial pressure during systole is reduced (systolic unloading), as is cardiac workload. The patient's calves, lower thighs and buttocks are wrapped with cuffs containing air bladders. The patient's ECG is monitored via conventional, high-quality electrodes and detection of the R-wave is used to signal the System when to command sequential inflation of the cuffs. The start and duration of inflation, as well as the start of deflation, are adjustable by the Operator, within limits determined by the System. An algorithm is used to prevent the start of inflation during ejection of blood from the heart and to end inflation and begin deflation prior to the occurrence of the next heartbeat. The pressure applied by the cuffs is also adjustable by the Operator from 150mmHg to 300mmHg (for a patient with a heart rate of 60 bpm). The device also utilizes the following cleared components:

ECG/EKG and SpO2: K123711

6. Intended use & Indication for use

The OM-A device is intended for the treatment of chronic stable angina that is refractory to optimal anti-anginal medical therapy and without options for revascularization. In addition, it is intended for use in healthy patients to provide improvement in vasodilation, and increased blood flow. It is intended for use under the oversight of a healthcare professional.

FeaturesSubject DevicePrimary Predicate DeviceConclusion
OM-AK130439 P-ECP/TI
ApplicantOmay (Guangzhou) MedTechnologies Co., Ltd.Chongqing PSK Sci-techDevelopment Co., Ltd./
Classificationn Regulation21CRF 870.522521CRF 870.5225Same
ProductCodeDRNDRNSame
CommonnameCounter-Pulsating, ExternalCounter-Pulsating, ExternalSame
TriggeringMechanismR-Wave triggerR-Wave triggerSame

7. Comparison to the Predicate Device

Comparison to Primary Predicate Device under K130436

{5}------------------------------------------------

FeaturesSubject DeviceOM-APrimary Predicate DeviceK130439 P-ECP/TIConclusion
MicroprocessorWindows BasedWindows BasedSame
EmergencySystempower-downRedRedSame
PressuresettingSetting range:5mmHg to 350mmHgSetting step: 1mmHgError: ± 10mmHgSetting range: 1kPa~50kPaSetting step: 1kPaError: ± 2kPaDifferent(1)
TreatmenttimeSetting range: 1min~45minSetting range: 1min~60minDifferent(2)
Cuff systemThree parts, calf, thigh andbuttocks.Three parts, calf, thigh andbuttocks.Same
Dimension/Weight2000mm × 800mm × 700mm,180kg2070mm × 1120mm × 1100mm,220kgDifferent(3)
MajorcomponentsBase Unit, Air-Tubes, ECGSPO2 and three CuffsBase Unit, Air-Tubes, ECG SPO2and three CuffsSame
OperatingEnvironment10°C 30°C;Relative humidity: less than85%;Atmospheric pressure:70kPa106kPa10°C 30°C;Relative humidity: less than 85%;Atmospheric pressure:70kPa106kPaSame
Safety andEMCIEC 60601-1IEC 60601-1-2IEC 60601-1IEC 60601-1-2Same
BiocompatibilityISO 10993-5ISO 10993-10ISO 10993-5ISO 10993-10Same

Justifications for differences between proposed device and the primary predicate device are shown as below:

Different (1) Pressure setting: The subject device's range of pressure is within predicate device. The error is less than predicate device. Thus, this difference will not cause any safety and effectiveness issues. Different (2) Treatment time: The subject device's range of treatment time is within predicate device. Thus, this difference will not cause any safety and effectiveness issues.

Different (3) Dimension/Weight: This feature is not related to clinical use and it is only a design consideration. Thus, this difference will not cause any safety and effectiveness issues.

Comparison to Reference Device under K190683

FeaturesSubject DeviceOM-AReference Device K190683 ECP-MC3Conclusion
ApplicantOmay (Guangzhou) MedVamed Medical Instrument Co.,/

{6}------------------------------------------------

FeaturesSubject DeviceOM-AReference Device K190683 ECP-MC3Conclusion
Classificationn RegulationTechnologies Co., Ltd.21CRF 870.5225Ltd.21CRF 870.5225Same
ProductCodeDRNDRNSame
CommonnameCounter-Pulsating, ExternalCounter-Pulsating, ExternalSame
Indicationfor useThe OM-A device is intended forthe treatment of chronic stableangina that is refractory tooptimal anti-anginal medicaltherapy and without options forrevascularization. In addition, itis intended for use in healthypatients to provide improvementin vasodilation, and increasedblood flow. It is intended for useunder the oversight of ahealthcare professional.The ECP-MC3 device is intendedfor the treatment of chronic stableangina that is refractory to optimalanti-anginal medical therapy andwithout options forrevascularization. In addition, it isintended for use in healthy patientsto provide improvement invasodilation, and increased bloodflow. It is intended for use under theoversight of a healthcareprofessional.Same

Performance Data 8.

The following performance data were provided in support of the substantial equivalence determination.

Biocompatibility testing

The biocompatibility evaluation for the OM-A was conducted in accordance with the FDA Guidance for Industry and Food and Drug Administration Staff: Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process". The battery of testing included the following tests:

  • . Cytotoxicity
  • o Sensitization
  • Irritation .

The subject devices are considered surface contacting for a duration of exceed 24 hours but not 30 days.

Non-clinical data

The OM-A Enhanced External Counter Pulsation Device Plus Omay-A have been tested according to the following standards:

  • . IEC 60601-1-2005+CORR.1:2006+CORR.2:2007+A1:2012, Medical Electrical Equipment-

{7}------------------------------------------------

Part 1: General requirements for basic safety and essential performance

  • IEC 60601-1-2:2014, Medical electrical equipment- Part 1-2: General requirements for basic . safety and essential performance- Collateral standard: Electromagnetic compatibility-Requirements and tests
    The test was selected to show substantial equivalence between the subject device and the predicate.

Clinical data

Clinical data were not required in this submission to support a finding of substantial equivalence.

9. Conclusion

We conclude that the information provided in this submission is sufficient to demonstrate that the OM-A Enhanced External Counter Pulsation Device is substantially equivalent to the predicate device which is currently marketed for the same intended use.

§ 870.5225 External counter-pulsating device.

(a)
Identification. An external counter-pulsating device is a noninvasive, prescription device used to assist the heart by applying positive or negative pressure to one or more of the body's limbs in synchrony with the heart cycle.(b)
Classification. (1) Class II (special controls) when the device is intended for the treatment of chronic stable angina that is refractory to optimal anti-anginal medical therapy and without options for revascularization. The special controls for this device are:(i) Nonclinical performance evaluation of the device must demonstrate a reasonable assurance of safety and effectiveness for applied pressure, synchronization of therapy with the appropriate phase of the cardiac cycle, and functionality of alarms during a device malfunction or an abnormal patient condition;
(ii) Reliabilities of the mechanical and electrical systems must be established through bench testing under simulated use conditions and matched by appropriate maintenance schedules;
(iii) Software design and verification and validation must be appropriately documented;
(iv) The skin-contacting components of the device must be demonstrated to be biocompatible;
(v) Appropriate analysis and testing must be conducted to verify electrical safety and electromagnetic compatibility of the device; and
(vi) Labeling must include a detailed summary of the device-related and procedure-related complications pertinent to use of the device.
(2) Class III (premarket approval) for the following intended uses: Unstable angina pectoris; acute myocardial infarction; cardiogenic shock; congestive heart failure; postoperative treatment of patients who have undergone coronary artery bypass surgery; peripheral arterial disease associated with ischemic ulcers rest pain or claudication, threatened gangrene, insufficient blood supply at an amputation site, persisting ischemia after embolectomy or bypass surgery, and/or pre- and post-arterial reconstruction to improve runoff; diabetes complicated by peripheral arterial disease or other conditions possibly related to arterial insufficiency including nocturnal leg cramps and/or necrobiosis diabeticorum; venous diseases, including prophylaxis of deep vein thrombophlebitis, edema (e.g., chronic lymphedema) and/or induration (e.g., stasis dermatitis) associated with chronic venous stasis, venous stasis ulcers, and/or thrombophlebitis; athletic injuries, including Charley horses, pulled muscles and/or edematous muscles; necrotizing cellulitis.
(c)
Date premarket approval application (PMA) or notice of completion of product development protocol (PDP) is required. A PMA or notice of completion of a PDP is required to be filed with FDA on or before March 31, 2014, for any external counter-pulsating device, with an intended use described in paragraph (b)(2) of this section, that was in commercial distribution before May 28, 1976, or that has, on or before March 31, 2014, been found to be substantially equivalent to any external counter-pulsating device, with an intended use described in paragraph (b)(2) of this section, that was in commercial distribution before May 28, 1976. Any other external counter-pulsating device with an intended use described in paragraph (b)(2) of this section shall have an approved PMA or declared completed PDP in effect before being placed in commercial distribution.