K Number
K191945
Device Name
KOWA nonmyd 8
Manufacturer
Date Cleared
2019-09-10

(50 days)

Product Code
Regulation Number
886.1120
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

KOWA nonmyd 8 is indicated for true color, infrared and fundus autofluorescent (FAF) imaging of a human retina without the use of a mydriatic agent.

Device Description

The KOWA nonmyd 8 is an instrument that enables retinal examination and retinal image capturing using infrared rays without requiring the patients to take any mydriatics. The KOWA nonmyd 8 is used with KOWA VK-2s software (510(k) number is K190056) which is designated filing software installed in a computer and enables to save, register or print the captured images. The KOWA nonmyd 8 has Color mode and FAF mode for photographing. Those are the functions to capture a normal color image and the Fundus Autofluorescence (FAF) image.

AI/ML Overview

The KOWA nonmyd 8 is an ophthalmic camera used for true color, infrared, and fundus autofluorescent (FAF) imaging of the human retina without pupil dilation.

Here's an analysis of the provided information regarding its acceptance criteria and supporting study:


  1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance Criterion (Implicit)Reported Device Performance
    Electrical Safety and Electromagnetic Compatibility (EMC): Compliance with relevant standards.Complies with IEC60601-1:2005 and IEC60601-1-2:2014.
    Biocompatibility: No new biocompatibility concerns for patient contact parts (forehead rest, chin rest).Forehead rest is identical to KOWA nonmyd 7 (K053026); chin rest is identical to KOWA DR-1α (K190573). Nature and duration of body contact are identical to reference devices. Determined no new biocompatibility concern.
    Software Verification and Validation: Compliance with FDA guidance for software in medical devices.Software has been validated according to FDA guidance "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices."
    Optical Radiation Safety: Classification regarding light hazard and compliance with optical safety standards.Classified in Group 1 of continuous wave instrument according to ANSI Z80.36-2016. Complies with ISO 10940:2009. Risk of radiation is low level, even with higher output power for photographing compared to the predicate device.
    FAF Image Quality: FAF image quality similar to or better than a legally marketed reference device (Canon CR-2 Plus AF).75% of images taken by KOWA nonmyd 8 were superior or equivalent to the Canon CR-2 Plus AF (K123208). The remaining 25% were inferior but diagnosable. The clinical testing demonstrates that FAF image quality of the proposed device is similar to that of the Canon CR-2 Plus AF (K123208) and no new concern was raised.
  2. Sample size used for the test set and data provenance:

    • Sample Size: 8 subjects.
    • Data Provenance: Not explicitly stated, but the study was conducted to compare the KOWA nonmyd 8 with the Canon CR-2 Plus AF, which implies prospective data collection for this purpose. The country of origin is not mentioned.
  3. Number of experts used to establish the ground truth for the test set and their qualifications:

    • Number of Experts: One doctor (referred to as "a doctor").
    • Qualifications: "a doctor" is mentioned, indicating a medical professional capable of evaluating diagnostic image quality. Specific qualifications (e.g., years of experience, subspecialty) are not provided.
  4. Adjudication method for the test set:

    • The method described is that "a doctor determined that the FAF image is not good quality for diagnosis" to prompt re-takes, and then the FAF images from the KOWA nonmyd 8 were "evaluated by comparing image quality with the FAF image from the Canon CR-2 Plus AF (K123208)." This suggests a single reader evaluation rather than a multi-reader adjudication method (like 2+1 or 3+1).
  5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No MRMC comparative effectiveness study was done in the context of human readers improving with or without AI assistance. The study described is a comparison of image quality between two devices (KOWA nonmyd 8 and an existing Canon device). There is no mention of AI.
  6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • This question is not applicable as the KOWA nonmyd 8 is an ophthalmic camera, not an AI algorithm. The performance evaluation focuses on the image quality produced by the camera.
  7. The type of ground truth used:

    • Expert Consensus: The "ground truth" for FAF image quality comparison was established by a single doctor's determination of diagnostic quality and subsequent comparison of images between the two devices. This leans towards expert opinion/evaluation. There is no mention of pathology or outcomes data being used as ground truth for image quality.
  8. The sample size for the training set:

    • Not applicable. The document describes a comparison study of a new photographic device (KOWA nonmyd 8) against an existing one, not the training of an algorithm via a training set.
  9. How the ground truth for the training set was established:

    • Not applicable, as there was no training set for an algorithm.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left, there is a seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern. To the right of the seal, there is a blue square with the letters "FDA" in white. Next to the blue square, the words "U.S. FOOD & DRUG ADMINISTRATION" are written in blue.

Kowa Company, Ltd. Nariaki Morita Manager of Development Management Dept 3-1. Chofugaoka 3-Chome Chofu, 1820021 JAPAN

Re: K191945

Trade/Device Name: KOWA nonmyd 8 Regulation Number: 21 CFR 886.1120 Regulation Name: Ophthalmic Camera Regulatory Class: Class II Product Code: HKI Dated: July 17, 2019 Received: July 22, 2019

Dear Nariaki Morita:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-devicereporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Elvin Ng Acting Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K191945

Device Name KOWA nonmyd 8

Indications for Use (Describe)

KOWA nonmyd 8 is indicated for true color, infrared and fundus autofluorescent (FAF) imaging of a human retina without the use of a mydriatic agent.

Type of Use (Select one or both, as applicable)

Exportation Use (Part 21 CFR 601.1, Subpart D)
Own Use, Contract Use (21 CFR 601.1, Subpart E)

X Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for Kowa Company, Ltd. The word "Kowa" is written in red and enclosed in a red oval. The words "Kowa Company, Ltd." are written in black, with a cursive font.

Chofu Factory

3-1 Chofugaoka 3-chome, Chofu City Tokyo 182-0021, Japan ……………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………

Phone:81-42(440)7612 FAX : 81-42(440)7618

510k Summary

a. Owner/Company name, address

Kowa Company, Ltd. 3-1, Chofugaoka 3-Chome Chofu Tokyo, JAPAN 182-0021 Phone: +81-42-440-7612 +81-42-440-7618 Fax:

b. Contact

Nariaki Morita

Kowa Company, Ltd. Medical Device Division 3-1, Chofugaoka 3-Chome Chofu Tokyo, JAPAN 182-0021 Phone: +81-42-440-7627 Fax: +81-42-440-7618 Email: n-morita@kowa.co.jp

c. Date prepared

June 17, 2019

d. Name of device

Trade Name: KOWA nonmyd 8 Regulation description: Ophthalmic Camera Regulation number: 21 CFR 886.1120 Product code: HKI

June 17, 2019

510(k) Summary Page 1 OF 9

{4}------------------------------------------------

e. Predicate and Reference Devices

Predicate Device
Trade name:KOWA nonmyd WX
510(k) number:K101628
Regulation description:Ophthalmic camera
Regulation number:21 CFR 886.1120
Product code:HKI

Reference Devices

The proposed device is equipped with FAF imaging modality. However, the predicate device does not include such imaging modality. Following device is referred for the feature of FAF imaging.

Trade name:Canon Digital Camera CR-2 Plus AF
510(k) number:K123208
Regulation description:Ophthalmic camera
Regulation number:21 CFR 886.1120
Product code:HKI

The forehead rest of the proposed device is identical to the forehead rest of following device;

Trade name:KOWA nonmyd 7
510(k) number:K053026
Regulation description:Ophthalmic camera
Regulation number:21 CFR 886.1120
Product code:HKI

The chin rest of the proposed device is identical to the chin rest of the following device;

Trade name:KOWA DR-1α
510(k) number:K190573
Regulation description:Ophthalmic camera
Regulation number:21 CFR 886.1120

June 17, 2019

{5}------------------------------------------------

HKI Product code:

f. Description of the device

The KOWA nonmyd 8 is an instrument that enables retinal examination and retinal image capturing using infrared rays without requiring the patients to take any mydriatics. The KOWA nonmyd 8 is used with KOWA VK-2s software (510(k) number is K190056) which is designated filing software installed in a computer and enables to save, register or print the captured images.

The KOWA nonmyd 8 has Color mode and FAF mode for photographing. Those are the functions to capture a normal color image and the Fundus Autofluorescence (FAF) image.

g. Indications for Use Statement

KOWA nonmyd 8 is indicated for true color, infrared and fundus autofluorescent (FAF) imaging of a human retina without the use of a mydriatic agent.

h. Discussion of substantial equivalence

Comparative Information

The KOWA nonmyd 8 is an instrument that enables retinal examination and retinal image capturing using infrared rays without requiring the patients to take any mydriatics. The predicate device can also capture retinal imaqes without mydratic.

Indications for Use Statement

The indications for use statement of the KOWA nonmyd 8 is not identical to that of the predicate device because the KOWA nonmyd 8 is not equipped with stereo imaging modality. The plane retinal images of the predicate device are identical to images using the color mode of the proposed device. The stereo imaging is not essential modality of an ophthalmic camera, nor does it affect safety and effectiveness of the KOWA nonmyd 8.

In addition, the indications for use statement of the KOWA nonmyd 8 does not include storage of the images because the KOWA nonmyd 8 is intended to be used with VK-2s software (K190056) which provides storage function.

{6}------------------------------------------------

Based on above, the differences of the indications for use statement do not alter intended use. Both the proposed and predicate devices are intended for use with retinal image capturing without mydriatic.

Technological characteristics

The proposed and predicate devices have the following same fundamental technologies;

  • Capturing non-mydriatic color retinal images ●
  • Small pupil mode
  • Alignment and focusing methods
  • Observation system

The predicate device has no FAF imaging mode. FAF mode is included in the Canon CR-2 Plus AF (K123208). The proposed device has no stereo mode.

Following table shows comprehensive comparison among the proposed, predicate and Canon CR-2 Plus AF(K123208).

{7}------------------------------------------------

Table 1 Comparison Table
Device NameKOWAnonmyd 8KOWAnonmyd WXCanon DigitalCameraCR-2 Plus AF
510(k) numberK101628K123208
Indications for usestatementKOWA nonmyd 8is indicated fortrue color,infrared andfundusautofluorescent(FAF) imaging ofa human retinawithout the useof a mydriaticagent.KOWA nonmydWX is intendedfor use with planeand stereo retinalimage capturingwithout mydriatic.The retinal imagecan be stored toan image filingdevice throughserial interface.The device isintended to beused for takingdigital images ofthe retina of thehuman eye withouta mydriatic. CR-2Plus AF has thefollowingphotographymodes: color, redfree, cobalt digitaland fundusautofluorescence(FAF).
Function
Photography modeColorNormal*Color
FAFStereoRed Free
Small pupilSmall pupilCobalt
FAF
Small pupil
Storing andDisplaying Images(Software name)YES(KOWA VK-2ssoftware(K190056))YES(Portable VK-2software)YES(unknown)
Transfer Imagesto external deviceYESYESunknown
Retinal camera specification
Saved ImagesFormatJPEGJPEGunknown
Field of view45 degree45 degree forplane image45 degree

{8}------------------------------------------------

34 degree forstereo image
Working distanceSame asKOWA nonmydWX30mm35mm
Working distancedetection methodSame asKOWA nonmydWXAnterior(Observation)Retinal (Focusingon bright spots)Anterior(Observation)Fundus (Workingdistance dots)
Minimum diameterof pupilNormal mode:$φ$ 4.0mmSmall pupilmode: $φ$ 3.3mmNormal mode:$φ$ 4.0mmSmall pupil mode:$φ$ 3.5mmNormal mode:$φ$ 4.0mmSmall pupil mode:$φ$ 3.5mm
DioptercompensationSame asKOWA nonmydWX-32 ~ +35D-31 ~ +33D
Internal eyefixation NavigationSame asKOWA nonmydWXFixation targetselecting by 11pointsLED Dot Matrix
Camera forobservation1/3 inch CCDcamera1/3 inch CCDcameraUnknown
ObservationsystemSame asKOWA nonmydWXLCDCamera unitmonitor
FocusingSame asKOWA nonmydWXBy alignment ofthe split linesBy alignment ofthe split lines
Filter for FAFPresentNot applicablePresent
Observation lightSourceSame asKOWA nonmydWXInfrared LEDlampLED
PhotographingLight SourceXenon flash lamp(Max 150W)Xenon flash lamp(Max 50W)Xenon tube
PowerconsumptionSame asKOWA nonmydWX150VAunknown
Dimension
DimensionSame asKOWA nonmydWX310(W) x 504(D)x548(H) mm305(W) x 513(D) x500(H) mm
WeightSame asKOWA nonmydWX21kg (excludingthe attached SLRcamera)19.9kg
Other
Observation mediaSame asKOWA nonmydWX5.7 inch LCDMonitorunknown
WX
Record mediaSame asKOWA nonmydWXFlash memorycardunknown
ConformedStandardsIEC60601-1:2005 +A1:2012IEC60601-1-2:2014ANSI Z80.36ISO10940:2009IEC60601-1:2005+A1:2012IEC60601-1-2:2007ISO15004-2:2007ISO10940:2009unknown

{9}------------------------------------------------

*Normal mode of the predicate device is equivalent to color mode of the proposed device.

The differences do not alter the intended use of the proposed device nor do they affect the safety and effectiveness of the proposed device relative to the predicate. The performance data to prove the safety and performance of KOWA nonmyd 8 are provided below.

Performance Data

The following performance data were provided in support of the substantial equivalence determination.

Electrical safety and electromagnetic compatibility (EMC)

Electrical safety and EMC testing were conducted on the KOWA nonmyd 8. The proposed device complies with the IEC60601-1:2005 and IEC60601-1-2:2014.

Biocompatibility

The forehead rest and the chin rest of the proposed device contact intact patient skin for a very short time. The forehead rest of the proposed device is identical to the forehead rest of the KOWA nonmyd 7(K053026) and the chin rest of the proposed device is identical to the chin rest of the KOWA DR-1α(Κ190573). Nature of body contact and contact duration of the proposed device are identical to those of the reference devices in regard to the forehead rest and the chin rest. We determined that the forehead rest and the chin rest of the proposed device have no new biocompatibility concern.

Software verification and validation testing

The software of the proposed device has been validated according to FDA guidance

{10}------------------------------------------------

entitled "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices."

Optical radiation safety

KOWA performed estimation of the light hazard and evaluation as to whether the KOWA nonmyd 8 satisfied the requirements of ANSI Z80.36-2016. As a result, the KOWA nonmyd 8 is classified in Group 1 of the continuous wave instrument. Although the output power of photographing of KOWA nonmyd 8 is higher than that of the predicate device, the risk of radiation is low level.

Furthermore, the KOWA nonmyd 8 complies with ISO 10940:2009 as with the predicate device.

Clinical Study

A comparative study regarding FAF images quality was conducted using the proposed and the Canon CR-2 Plus AF (K123208) devices. 8 subjects were accepted for this study. For those subjects, one color (with appropriate strobe intensity) image and one FAF image (with ± 0 strobe intensity) by the Canon CR-2 Plus AF (K123208) were taken. If a doctor determined that the FAF image is not good quality for diagnosis, one more FAF image with appropriate strobe intensity was taken by the Canon CR-2 Plus AF (K123208). And then, for the same subjects, one color (with appropriate strobe intensity) and three FAF images (with ± 0 and with -2 & -1, -1 & +1 or +1 & +2) by the KOWA nonmyd 8 were taken. The best FAF image for diagnosis taken with the KOWA nonmyd 8 was selected. The FAF image from the KOWA nonmyd 8 was evaluated by comparing image quality with the FAF image from the Canon CR-2 Plus AF (K123208).

The KOWA nonmyd 8 were superior or equivalent to the Canon CR-2 Plus AF (K123208) for 6 subjects and were inferior but diagnosable for 2 subjects. Therefore, 75 % images by the KOWA nonmyd 8 were superior or equivalent to the Canon CR-2 Plus AF (K123208).

i. Conclusion

The non-clinical testing demonstrates that performance except FAF imaging of the KOWA nonmyd 8 was comparable to the predicate device and any new concern reqarding safety and performance was not raised for the proposed device. The

{11}------------------------------------------------

clinical testing demonstrates that FAF image quality of the proposed device is similar to that of the Canon CR-2 Plus AF (K123208) and no new concern was raised for the proposed device. Based on the information described above, we conclude that the KOWA nonmyd 8 is substantially equivalent to the predicate device.

§ 886.1120 Ophthalmic camera.

(a)
Identification. An ophthalmic camera is an AC-powered device intended to take photographs of the eye and the surrounding area.(b)
Classification. Class II (special controls). The device, when it is a photorefractor or a general-use ophthalmic camera, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.