(28 days)
The ACUSON P200 ultrasound imaging system is intended for the following applications: Fetal, Abdominal, Pediatric, Small Parts, OB/GYN (useful for visualization of the ovaries, follicles, uterus, and other pelvic structures), Adult, Pediatric and Neonatal Cardiac, Pelvic, Neonatal Cephalic, Musculoskeletal, Superficial Musculoskeletal, and Peripheral Vascular applications.
The system also provides the ability to measure anatomical structures; fetal, abdominal, pediatric, small organ, neonatal cephalic, cardiac (adult, pediatric and neonatal), trans-esophageal, transrectal, transvaginal, peripheral vessel, musculoskeletal (conventional), musculoskeletal (superficial) and calculation packages that provide information to the clinician that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes.
The ACUSON P200 Diagnostic Ultrasound System is a multi-purpose mobile, software controlled, diagnostic ultrasound system with an on-screen display of thermal and mechanical indices related to potential bio-effect mechanisms. Its function is to transmit and receive ultrasound echo data and display it in B-Mode. M-Mode. Pulsed (PW) Doppler Mode. Continuous (CW) Doppler Mode. Color Doppler Mode, Color M Mode, Doppler Tissue Mode, Amplitude Doppler Mode, a combination of modes and Harmonic Imaging on a Display.
The provided document does not contain information about specific acceptance criteria related to a device's performance that would be typically found in a clinical study for an AI/ML device (e.g., accuracy, sensitivity, specificity).
Instead, this 510(k) summary focuses on establishing substantial equivalence of the ACUSON P200 Diagnostic Ultrasound System to previously cleared predicate devices. This means that for general diagnostic ultrasound systems like this, clinical performance metrics that would require "acceptance criteria" based on a study, as understood for AI/ML devices, are typically not required. The substantial equivalence is demonstrated by showing that the new device has the same intended use and similar technological characteristics as the predicates, and that non-clinical tests (acoustic output, biocompatibility, safety, etc.) meet established standards.
Therefore, many of the specific questions you asked concerning acceptance criteria and study details (like sample size for test set, number of experts for ground truth, adjudication methods, MRMC studies, or standalone performance for an algorithm) are not applicable to the information provided in this 510(k) summary for the ACUSON P200 Diagnostic Ultrasound System.
The document explicitly states:
- "Since the ACUSON P200 Diagnostic Ultrasound System uses the same technology and principles as existing devices, clinical studies were not required to support substantial equivalence." (Page 21, Section 8)
However, I can extract the information that is present:
1. A table of acceptance criteria and the reported device performance
Since no specific performance-based acceptance criteria for clinical efficacy (like sensitivity or specificity) are provided for this ultrasound system, a table in the format you requested cannot be fully generated. The "performance" demonstrated for substantial equivalence is primarily through equivalence in intended use, technological characteristics, and compliance with general safety and performance standards.
The document states compliance with various standards as evidence of performance. These implicitly act as "acceptance criteria" for the non-clinical aspects.
| Acceptance Criterion (Type) | Reported Device Performance/Compliance |
|---|---|
| Non-Clinical Performance | |
| Acoustic Output | Conforms to AIUM/NEMA UD-3:2004, AIUM/NEMA UD-2:2004 (R2009), IEC 62359:2010 |
| Biocompatibility | Tested to ISO 10993-1 |
| Cleaning and Disinfection Effectiveness | Conforms to applicable medical device safety standards |
| Thermal Safety | Conforms to applicable medical device safety standards |
| Electromagnetic Safety | Conforms to AAMI ES60601-1:2005/(R)2012, IEC 60601-1:2005, IEC 60601-1-2 Edition 4.0 2014-02 |
| Mechanical Safety | Conforms to applicable medical device safety standards |
| Product Safety Standards | UL 60601-1 Certified, IEC 60601-2-18: Edition 3.0 2009-08, IEC 60601-2-37 Edition 2.1 2015 |
| Intended Use | |
| Fetal | Matches predicate (K180067) |
| Abdominal | Matches predicate (K180067) |
| Pediatric | Matches predicate (K180067, K183575) |
| Small Organ | Matches predicate (K180067) |
| Cardiac | Matches predicate (K180067, K183575) |
| Neonatal Cephalic | Matches predicate (K183575 via ACUSON S3000) |
| Trans-esophageal | Matches predicate (K183575 via ACUSON S3000) |
| Transrectal | Matches predicate (K180067) |
| Transvaginal | Matches predicate (K180067) |
| Peripheral vessel | Matches predicate (K180067) |
| Musculo-skeletal (conventional) | Matches predicate (K180067) |
| Musculo-skeletal (superficial) | Matches predicate (K180067) |
| Technological Characteristics | |
| Frequencies Supported | 1.0MHZ~18MHz (Matches predicate K180067) |
| Modes | B, M, PWD, CWD, PW DTI, Color Doppler, Power Doppler, Combined (BMDC) (Matches predicate K180067) |
| Features | Harmonics, Compounding, UltraArt Universal Image processing, Tissue Equalization (TEQ), Biopsy, Clarify, Cardiac Imaging, Speed of Sound, Protocols, DICOM, eSie Calcs, Wireless, Virtual Touch Strain Imaging, Virtual Touch Point Shear Wave (VTQ), Virtual Touch Shear Wave (VTIQ), Contrast Agent Imaging, eSie OB, eSie Left Heart, eSie Measure, eSie Follicle, Panoramic Imaging, Stress Echo, Syngo Velocity Vector Imaging (VVI), eSieLink, DICOM SR (Matches one or more predicates: K180067, K180039, K183575) |
| Touch Screen | 13.3" (Predicate K180067 lists 15") |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
Not applicable. Clinical studies were not required. The "test set" here would refer to the technical evaluations and comparisons to predicate devices.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
Not applicable. Clinical studies were not required.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable. Clinical studies were not required.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is a diagnostic ultrasound system, not an AI/ML diagnostic aid. Clinical studies were not required.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This is a diagnostic ultrasound system, not an AI/ML diagnostic aid. Clinical studies were not required.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
Not applicable. Clinical studies were not required. The "ground truth" for demonstrating substantial equivalence is based on established performance specifications, regulatory standards, and the technological characteristics of predicate devices.
8. The sample size for the training set
Not applicable. This is a diagnostic ultrasound system, not an AI/ML device that undergoes training with a dataset.
9. How the ground truth for the training set was established
Not applicable. This is a diagnostic ultrasound system, not an AI/ML device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Siemens Medical Solutions USA, Inc. % Prithul Bom Responsible Third Party Official Regulatory Technology Services, LLC 1000 Westgate Drive, Suite #510k SAINT PAUL MN 55114
August 15, 2019
Re: K191922
Trade/Device Name: ACUSON P200 Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: Class II Product Code: IYN, IYO, ITX Dated: July 17, 2019 Received: July 18, 2019
Dear Prithul Bom:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K191922
Device Name ACUSON P200 Diagnostic Ultrasound System
Indications for Use (Describe)
The ACUSON P200 ultrasound imaging system is intended for the following applications: Fetal, Abdominal, Pediatric, Small Parts, OB/GYN (useful for visualization of the ovaries, follicles, uterus, and other pelvic structures), Adult, Pediatric and Neonatal Cardiac, Pelvic, Neonatal Cephalic, Musculoskeletal, Superficial Musculoskeletal, and Peripheral Vascular applications.
The system also provides the ability to measure anatomical structures; fetal, abdominal, pediatric, small organ, neonatal cephalic, cardiac (adult, pediatric and neonatal), trans-esophageal, transvaginal, peripheral vessel, musculoskeletal (conventional), musculosketal (superficial) and calculation packages that provide information to the clinician that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| ------------------------------------------------- | -- |
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510 (k) Number (if known):
Device Name:
ACUSON P200 Diagnostic Ultrasound System
Intended Use:
Ultrasound imaging or fluid flow analysis of the human body as follows:
| Mode of Operation | ||||||||
|---|---|---|---|---|---|---|---|---|
| Clinical Application | B | M | PWD | CWD | ColorDoppler | AmplitudeDoppler | Combined(Specify) | Other(Specify) |
| Ophthalmic | ||||||||
| Fetal | P | P | P | P | P | BMDC | ||
| Abdominal | P | P | P | P | P | BMDC | ||
| Intraoperative | ||||||||
| IntraoperativeNeurological | ||||||||
| Pediatric | P | P | P | P | P | P | BMDC | |
| Small Organ (Note 1) | P | P | P | P | P | BMCD | ||
| Neonatal Cephalic | P | P | P | P | P | BMCD | ||
| Adult Cephalic | ||||||||
| Cardiac | P | P | P | P | P | P | BMCD | |
| Trans-esophageal | P | P | P | P | P | P | BMCD | |
| Transrectal | P | P | P | P | P | BMCD | ||
| Transvaginal | P | P | P | P | P | BMCD | ||
| Transurethral | ||||||||
| Intravascular | ||||||||
| Peripheral vessel(Note 2) | P | P | P | P | P | P | BMCD | |
| Laparoscopic | ||||||||
| Musculo-skeletalConventional | P | P | P | P | P | BMCD | ||
| Musculo-skeletalSuperficial | P | P | P | P | P | BMCD | ||
| Other (specify) |
N = new indication; P = previously cleared by FDA K180067; K180039; K187357
Note 1 Note 2 For example: Breast, Thyroid, Testis For example: Carotid, Arterial, Venous
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
{4}------------------------------------------------
510 (k) Number (if known):
Device Name: Intended Use:
4V1 Phased Array Transducer
Ultrasound imaging or fluid flow analysis of the human body as follows:
| Mode of Operation | ||||||||
|---|---|---|---|---|---|---|---|---|
| Clinical Application | B | M | PWD | CWD | ColorDoppler | AmplitudeDoppler | Combined(Specify) | Other(Specify) |
| Ophthalmic | ||||||||
| Fetal | P | P | P | P | P | BMDC | ||
| Abdominal | P | P | P | P | P | BMDC | ||
| Intraoperative | ||||||||
| IntraoperativeNeurological | ||||||||
| Pediatric | P | P | P | P | P | BMDC | ||
| Small Organ (Note 1) | ||||||||
| Neonatal Cephalic | ||||||||
| Adult Cephalic | ||||||||
| Cardiac | ||||||||
| Trans-esophageal | ||||||||
| Transrectal | ||||||||
| Transvaginal | ||||||||
| Transurethral | ||||||||
| Intravascular | ||||||||
| Peripheral vessel(Note 2) | ||||||||
| Laparoscopic | ||||||||
| Musculo-skeletalConventional | ||||||||
| Musculo-skeletalSuperficial | ||||||||
| Other (specify) |
N = new indication; P = previously cleared by FDA K180067
For example: Breast, Thyroid, Testis Note 1 Note 2 For example: Carotid, Arterial, Venous
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
{5}------------------------------------------------
510 (k) Number (if known):
Device Name:
5V1 Phased Array Transducer
Intended Use:
Ultrasound imaging or fluid flow analysis of the human body as follows:
| Mode of Operation | ||||||||
|---|---|---|---|---|---|---|---|---|
| Clinical Application | B | M | PWD | CWD | ColorDoppler | AmplitudeDoppler | Combined(Specify) | Other(Specify) |
| Ophthalmic | ||||||||
| Fetal | ||||||||
| Abdominal | ||||||||
| Intraoperative | ||||||||
| IntraoperativeNeurological | ||||||||
| Pediatric | P | P | P | P | P | P | BMCD | |
| Small Organ (Note 1) | ||||||||
| Neonatal Cephalic | ||||||||
| Adult Cephalic | ||||||||
| Cardiac | P | P | P | P | P | P | BMCD | |
| Trans-esophageal | ||||||||
| Transrectal | ||||||||
| Transvaginal | ||||||||
| Transurethral | ||||||||
| Intravascular | ||||||||
| Peripheral vessel(Note 2) | ||||||||
| Laparoscopic | ||||||||
| Musculo-skeletalConventional | ||||||||
| Musculo-skeletalSuperficial | ||||||||
| Other (specify) |
N = new indication; P = previously cleared by FDA K180067
Note 1
Note 2
For example: Breast, Thyroid, Testis
For example: Carotid, Arterial, Venous
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
{6}------------------------------------------------
510 (k) Number (if known):
Device Name:
8V3 Phased Array Transducer
Intended Use:
Ultrasound imaging or fluid flow analysis of the human body as follows:
| Mode of Operation | ||||||||
|---|---|---|---|---|---|---|---|---|
| Clinical Application | B | M | PWD | CWD | ColorDoppler | AmplitudeDoppler | Combined(Specify) | Other(Specify) |
| Ophthalmic | ||||||||
| Fetal | ||||||||
| Abdominal | P | P | P | P | P | BMCD | ||
| Intraoperative | ||||||||
| IntraoperativeNeurological | ||||||||
| Pediatric | P | P | P | P | P | P | BMCD | |
| Small Organ (Note 1) | ||||||||
| Neonatal Cephalic | ||||||||
| Adult Cephalic | ||||||||
| Cardiac | P | P | P | P | P | P | BMCD | |
| Trans-esophageal | ||||||||
| Transrectal | ||||||||
| Transvaginal | ||||||||
| Transurethral | ||||||||
| Intravascular | ||||||||
| Peripheral vessel(Note 2) | ||||||||
| Laparoscopic | ||||||||
| Musculo-skeletalConventional | ||||||||
| Musculo-skeletalSuperficial | ||||||||
| Other (specify) |
N = new indication; P = previously cleared by FDA K180067
Note 1 Note 2 For example: Breast, Thyroid, Testis For example: Carotid, Arterial, Venous
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
{7}------------------------------------------------
510 (k) Number (if known):
Device Name:
10V4 Curved Array Transducer
Intended Use:
Ultrasound imaging or fluid flow analysis of the human body as follows:
| Mode of Operation | ||||||||
|---|---|---|---|---|---|---|---|---|
| Clinical Application | B | M | PWD | CWD | ColorDoppler | AmplitudeDoppler | Combined(Specify) | Other(Specify) |
| Ophthalmic | ||||||||
| Fetal | ||||||||
| Abdominal | P | P | P | P | P | BMCD | ||
| Intraoperative | ||||||||
| IntraoperativeNeurological | ||||||||
| Pediatric | P | P | P | P | P | BMCD | ||
| Small Organ (Note 1) | ||||||||
| Neonatal Cephalic | P | P | P | P | P | BMCD | ||
| Adult Cephalic | ||||||||
| Cardiac | P | P | P | P | P | P | BMCD | |
| Trans-esophageal | ||||||||
| Transrectal | ||||||||
| Transvaginal | ||||||||
| Transurethral | ||||||||
| Intravascular | ||||||||
| Peripheral vessel(Note 2) | ||||||||
| Laparoscopic | ||||||||
| Musculo-skeletalConventional | ||||||||
| Musculo-skeletalSuperficial | ||||||||
| Other (specify) |
N = new indication; P = previously cleared by FDA K183575
Note 1 Note 2 For example: Breast, Thyroid, Testis For example: Carotid, Arterial, Venous
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
{8}------------------------------------------------
510 (k) Number (if known):
Device Name:
CW2 Continuous Wave Transducer
Intended Use:
Ultrasound imaging or fluid flow analysis of the human body as follows:
| Mode of Operation | ||||||||
|---|---|---|---|---|---|---|---|---|
| Clinical Application | B | M | PWD | CWD | ColorDoppler | AmplitudeDoppler | Combined(Specify) | Other(Specify) |
| Ophthalmic | ||||||||
| Fetal | ||||||||
| Abdominal | ||||||||
| Intraoperative | ||||||||
| IntraoperativeNeurological | ||||||||
| Pediatric | P | |||||||
| Small Organ (Note 1) | ||||||||
| Neonatal Cephalic | ||||||||
| Adult Cephalic | ||||||||
| Cardiac | P | |||||||
| Trans-esophageal | ||||||||
| Transrectal | ||||||||
| Transvaginal | ||||||||
| Transurethral | ||||||||
| Intravascular | ||||||||
| Peripheral vessel(Note 2) | ||||||||
| Laparoscopic | ||||||||
| Musculo-skeletalConventional | ||||||||
| Musculo-skeletalSuperficial | ||||||||
| Other (specify) |
N = new indication; P = previously cleared by FDA K183575
Note 1 Note 2 For example: Breast, Thyroid, Testis For example: Carotid, Arterial, Venous
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
{9}------------------------------------------------
510 (k) Number (if known):
CW5 Continuous Wave Transducer
Device Name: Intended Use:
Ultrasound imaging or fluid flow analysis of the human body as follows:
| Mode of Operation | ||||||||
|---|---|---|---|---|---|---|---|---|
| Clinical Application | B | M | PWD | CWD | ColorDoppler | AmplitudeDoppler | Combined(Specify) | Other(Specify) |
| Ophthalmic | ||||||||
| Fetal | ||||||||
| Abdominal | ||||||||
| Intraoperative | ||||||||
| IntraoperativeNeurological | ||||||||
| Pediatric | P | |||||||
| Small Organ (Note 1) | ||||||||
| Neonatal Cephalic | ||||||||
| Adult Cephalic | ||||||||
| Cardiac | ||||||||
| Trans-esophageal | ||||||||
| Transrectal | ||||||||
| Transvaginal | ||||||||
| Transurethral | ||||||||
| Intravascular | ||||||||
| Peripheral vessel(Note 2) | P | |||||||
| Laparoscopic | ||||||||
| Musculo-skeletalConventional | ||||||||
| Musculo-skeletalSuperficial | ||||||||
| Other (specify) |
N = new indication; P = previously cleared by FDA K183575
Note 1 For example: Breast, Thyroid, Testis
For example: Carotid, Arterial, Venous Note 2
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
{10}------------------------------------------------
510 (k) Number (if known):
10L4 Linear Array Transducer
Device Name: Intended Use:
Ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | |||||||
|---|---|---|---|---|---|---|---|---|
| B | M | PWD | CWD | ColorDoppler | AmplitudeDoppler | Combined(Specify) | Other(Specify) | |
| Ophthalmic | ||||||||
| Fetal | P | P | P | P | P | BMCD | ||
| Abdominal | P | P | P | P | P | BMCD | ||
| Intraoperative | ||||||||
| IntraoperativeNeurological | ||||||||
| Pediatric | P | P | P | P | P | BMCD | ||
| Small Organ (Note 1) | P | P | P | P | P | BMCD | ||
| Neonatal Cephalic | ||||||||
| Adult Cephalic | ||||||||
| Cardiac | ||||||||
| Trans-esophageal | ||||||||
| Transrectal | ||||||||
| Transvaginal | ||||||||
| Transurethral | ||||||||
| Intravascular | ||||||||
| Peripheral vessel(Note 2) | P | P | P | P | P | BMCD | ||
| Laparoscopic | ||||||||
| Musculo-skeletalConventional | P | P | P | P | P | BMCD | ||
| Musculo-skeletalSuperficial | P | P | P | P | P | BMCD | ||
| Other (specify) |
N = new indication; P = previously cleared by FDA K180067
Note 1 For example: Breast, Thyroid, Testis
For example: Carotid, Arterial, Venous Note 2
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
{11}------------------------------------------------
510 (k) Number (if known):
Device Name:
14L5 Linear Array Transducer
Intended Use:
Ultrasound imaging or fluid flow analysis of the human body as follows:
| Mode of Operation | ||||||||
|---|---|---|---|---|---|---|---|---|
| Clinical Application | B | M | PWD | CWD | ColorDoppler | AmplitudeDoppler | Combined(Specify) | Other(Specify) |
| Ophthalmic | ||||||||
| Fetal | ||||||||
| Abdominal | ||||||||
| Intraoperative | ||||||||
| IntraoperativeNeurological | ||||||||
| Pediatric | P | P | P | P | P | BMCD | ||
| Small Organ (Note 1) | P | P | P | P | P | BMCD | ||
| Neonatal Cephalic | ||||||||
| Adult Cephalic | ||||||||
| Cardiac | ||||||||
| Trans-esophageal | ||||||||
| Transrectal | ||||||||
| Transvaginal | ||||||||
| Transurethral | ||||||||
| Intravascular | ||||||||
| Peripheral vessel(Note 2) | P | P | P | P | P | BMCD | ||
| Laparoscopic | ||||||||
| Musculo-skeletalConventional | P | P | P | P | P | BMCD | ||
| Musculo-skeletalSuperficial | P | P | P | P | P | BMCD | ||
| Other (specify) |
N = new indication; P = previously cleared by FDA K180067
Note 1 Note 2 For example: Breast, Thyroid, Testis For example: Carotid, Arterial, Venous
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
{12}------------------------------------------------
510 (k) Number (if known):
Device Name:
18L6 Linear Array Transducer
Intended Use:
Ultrasound imaging or fluid flow analysis of the human body as follows:
| Mode of Operation | ||||||||
|---|---|---|---|---|---|---|---|---|
| Clinical Application | B | M | PWD | CWD | ColorDoppler | AmplitudeDoppler | Combined(Specify) | Other(Specify) |
| Ophthalmic | ||||||||
| Fetal | ||||||||
| Abdominal | ||||||||
| Intraoperative | ||||||||
| IntraoperativeNeurological | ||||||||
| Pediatric | P | P | P | P | P | BMCD | ||
| Small Organ (Note 1) | P | P | P | P | P | BMCD | ||
| Neonatal Cephalic | ||||||||
| Adult Cephalic | ||||||||
| Cardiac | ||||||||
| Trans-esophageal | ||||||||
| Transrectal | ||||||||
| Transvaginal | ||||||||
| Transurethral | ||||||||
| Intravascular | ||||||||
| Peripheral vessel(Note 2) | ||||||||
| Laparoscopic | ||||||||
| Musculo-skeletalConventional | P | P | P | P | P | BMCD | ||
| Musculo-skeletalSuperficial | P | P | P | P | P | BMCD | ||
| Other (specify) |
N = new indication; P = previously cleared by FDA K180067
Note 1 Note 2 For example: Breast, Thyroid, Testis For example: Carotid, Arterial, Venous
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
{13}------------------------------------------------
510 (k) Number (if known):
Device Name:
5C1 Curved Array Transducer
Intended Use:
Ultrasound imaging or fluid flow analysis of the human body as follows:
| Mode of Operation | ||||||||
|---|---|---|---|---|---|---|---|---|
| Clinical Application | B | M | PWD | CWD | ColorDoppler | AmplitudeDoppler | Combined(Specify) | Other(Specify) |
| Ophthalmic | ||||||||
| Fetal | P | P | P | P | P | BMCD | ||
| Abdominal | P | P | P | P | P | BMCD | ||
| Intraoperative | ||||||||
| IntraoperativeNeurological | ||||||||
| Pediatric | P | P | P | P | P | BMCD | ||
| Small Organ (Note 1) | ||||||||
| Neonatal Cephalic | ||||||||
| Adult Cephalic | ||||||||
| Cardiac | ||||||||
| Trans-esophageal | ||||||||
| Transrectal | ||||||||
| Transvaginal | ||||||||
| Transurethral | ||||||||
| Intravascular | ||||||||
| Peripheral vessel(Note 2) | ||||||||
| Laparoscopic | ||||||||
| Musculo-skeletalConventional | ||||||||
| Musculo-skeletalSuperficial | ||||||||
| Other (specify) |
N = new indication; P = previously cleared by FDA K180067
Note 1 Note 2 For example: Breast, Thyroid, Testis For example: Carotid, Arterial, Venous
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
{14}------------------------------------------------
510 (k) Number (if known):
Device Name:
9C3 Curved Array Transducer
Intended Use:
Ultrasound imaging or fluid flow analysis of the human body as follows:
| Mode of Operation | ||||||||
|---|---|---|---|---|---|---|---|---|
| Clinical Application | B | M | PWD | CWD | ColorDoppler | AmplitudeDoppler | Combined(Specify) | Other(Specify) |
| Ophthalmic | ||||||||
| Fetal | P | P | P | P | P | BMCD | ||
| Abdominal | P | P | P | P | P | BMCD | ||
| Intraoperative | ||||||||
| IntraoperativeNeurological | ||||||||
| Pediatric | P | P | P | P | P | BMCD | ||
| Small Organ (Note 1) | ||||||||
| Neonatal Cephalic | ||||||||
| Adult Cephalic | ||||||||
| Cardiac | ||||||||
| Trans-esophageal | ||||||||
| Transrectal | ||||||||
| Transvaginal | ||||||||
| Transurethral | ||||||||
| Intravascular | ||||||||
| Peripheral vessel(Note 2) | ||||||||
| Laparoscopic | ||||||||
| Musculo-skeletalConventional | P | P | P | P | P | BMCD | ||
| Musculo-skeletalSuperficial | ||||||||
| Other (specify) |
N = new indication; P = previously cleared by FDA K180067
Note 1 Note 2 For example: Breast, Thyroid, Testis For example: Carotid, Arterial, Venous
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
{15}------------------------------------------------
510 (k) Number (if known):
Device Name:
9EC4 Endocavity Transducer
Intended Use:
Ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | |||||||
|---|---|---|---|---|---|---|---|---|
| B | M | PWD | CWD | ColorDoppler | AmplitudeDoppler | Combined(Specify) | Other(Specify) | |
| Ophthalmic | ||||||||
| Fetal | ||||||||
| Abdominal | ||||||||
| Intraoperative | ||||||||
| IntraoperativeNeurological | ||||||||
| Pediatric | ||||||||
| Small Organ (Note 1) | ||||||||
| Neonatal Cephalic | ||||||||
| Adult Cephalic | ||||||||
| Cardiac | ||||||||
| Trans-esophageal | ||||||||
| Transrectal | P | P | P | P | P | BMCD | ||
| Transvaginal | P | P | P | P | P | BMCD | ||
| Transurethral | ||||||||
| Intravascular | ||||||||
| Peripheral vessel(Note 2) | ||||||||
| Laparoscopic | ||||||||
| Musculo-skeletalConventional | ||||||||
| Musculo-skeletalSuperficial | ||||||||
| Other (specify) |
N = new indication; P = previously cleared by FDA K180067
Note 1 Note 2 For example: Breast, Thyroid, Testis For example: Carotid, Arterial, Venous
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
{16}------------------------------------------------
510 (k) Number (if known):
Device Name:
V5Ms Multiplane TEE Transducer
Intended Use:
Ultrasound imaging or fluid flow analysis of the human body as follows:
| Mode of Operation | ||||||||
|---|---|---|---|---|---|---|---|---|
| Clinical Application | B | M | PWD | CWD | ColorDoppler | AmplitudeDoppler | Combined(Specify) | Other(Specify) |
| Ophthalmic | ||||||||
| Fetal | ||||||||
| Abdominal | ||||||||
| Intraoperative | ||||||||
| IntraoperativeNeurological | ||||||||
| Pediatric | ||||||||
| Small Organ (Note 1) | ||||||||
| Neonatal Cephalic | ||||||||
| Adult Cephalic | ||||||||
| Cardiac | ||||||||
| Trans-esophageal | P | P | P | P | P | P | BMCD | |
| Transrectal | ||||||||
| Transvaginal | ||||||||
| Transurethral | ||||||||
| Intravascular | ||||||||
| Peripheral vessel(Note 2) | ||||||||
| Laparoscopic | ||||||||
| Musculo-skeletalConventional | ||||||||
| Musculo-skeletalSuperficial | ||||||||
| Other (specify) |
N = new indication; P = previously cleared by FDA K183575
Note 1 Note 2 For example: Breast, Thyroid, Testis For example: Carotid, Arterial, Venous
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
{17}------------------------------------------------
510(K) SUMMARY
| Date: | August 1, 2019 | ||
|---|---|---|---|
| 1. Sponsor: | Siemens Medical Solutions USA, Inc.,Ultrasound Division685 East Middlefield RoadMountain View, California 94043 | ||
| Contact Person: | HyunJung LeeTel: (425) 281-5061 | ||
| 2. Device Name: | ACUSON P200 Diagnostic Ultrasound System | ||
| Common Name: | Diagnostic Ultrasound System with Accessories | ||
| Classification: | Regulatory Class:Review Category:Classification Panel: | IITier IIRadiology | |
| Ultrasonic Pulsed Doppler Imaging System | 892.1550 | 90-IYN | |
| Ultrasonic Pulsed Echo Imaging System | 892.1560 | 90-IYO | |
| Diagnostic Ultrasound Transducer | 892.1570 | 90-ITX | |
| Manufacturing Site: | Siemens Healthineers Ltd.2nd ~ 3rd Floor, 143, Sunhwan-ro, Jungwon-gu, Seongnam-si,Gyeonggi-do, Republic of Korea |
3. Legally Marketed Predicate Devices
The ACUSON P200 Diagnostic Ultrasound System is a multi-purpose diagnostic ultrasound system with accessories and proprietary software, and is substantially equivalent to the company's own ACUSON X800 (K180067) which is the primary predicate device.
The additional predicates are the indications for Neonatal Cephalic and Trans-esophageal uses which are cleared under ACUSON S3000 (K183575) and eSie Left Heart, eSie Measure, eSie Follicle and Stress Eco under ACUSON Juniper (K180039) as described in the table of section 6 Summary of Technological Characteristics.
{18}------------------------------------------------
4. Device Description
The ACUSON P200 Diagnostic Ultrasound System is a multi-purpose mobile, software controlled, diagnostic ultrasound system with an on-screen display of thermal and mechanical indices related to potential bio-effect mechanisms. Its function is to transmit and receive ultrasound echo data and display it in B-Mode. M-Mode. Pulsed (PW) Doppler Mode. Continuous (CW) Doppler Mode. Color Doppler Mode, Color M Mode, Doppler Tissue Mode, Amplitude Doppler Mode, a combination of modes and Harmonic Imaging on a Display.
5. Intended Use
The ACUSON P200 ultrasound imaging system is intended for the following applications: Fetal, Abdominal, Pediatric, Small Parts, OB/GYN (useful for visualization of the ovaries, follicles, uterus, and other pelvic structures). Adult. Pediatric and Neonatal Cardiac. Pelvic. Neonatal Cephalic. Vascular. Musculoskeletal, Superficial Musculoskeletal, and Peripheral Vascular applications.
The system also provides the ability to measure anatomical structures; fetal, abdominal, pediatric, small organ, neonatal cephalic, cardiac (adult, pediatric and neonatal), trans-esophageal, transrectal, transvaginal, peripheral vessel, musculoskeletal (conventional), musculoskeletal (superficial) and calculation packages that provide information to the clinician that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes.
6. Summary of Technological Characteristics
The ACUSON P200 Diagnostic Ultrasound System is a multi-purpose diagnostic ultrasound system with accessories and proprietary software, and is substantially equivalent to the company's own products, the ACUSON X800 (K180067). ACUSON Juniper (K180039) and ACUSON S3000 (K183575) with regards to both intended use and technological characteristics. Both the subject ultrasound system and the predicate ultrasound systems function in the same manner as all diagnostic ultrasound systems and transducers.
The foundation of the ACUSON P200 is the ACUSON X800 with features and transducers integrated with the ACUSON X800 hardware and the ACUSON P200 reuse software developed for X800 mainly as well as some transducers from Juniper and S3000 with a connector.
| Feature / Characteristic | ACUSON P200Subject device | ACUSON X800K180067 | ACUSONJuniperK1180039 | ACUSON S3000K183575 | |
|---|---|---|---|---|---|
| Indications for Use: | |||||
| ■ | Fetal | √ | √ | ||
| ■ | Abdominal | √ | √ | ||
| ■ | Pediatric | √ | √ | ||
| ■ | Small Organ | √ | √ | ||
| ■ | Cardiac | √ | √ | ||
| ■ | Neonatal Cephalic | √ | √ | ||
| ■ | Trans-esophageal | √ | √ | ||
| ■ | Transrectal | √ | √ | ||
| ■ | Transvaginal | √ | √ | ||
| ■ | Peripheral vessel | √ | √ | ||
| ■ | Musculo-skeletal(conventional) | √ | √ | ||
| Feature / Characteristic | ACUSON P200Subject device | ACUSON X800K180067 | ACUSONJuniperK1180039 | ACUSON S3000K183575 | |
| ■ | Musculo-skeletal (superficial) | √ | √ | ||
| Frequencies Supported: | √(1.0MHZ~18MHz) | √(1.0MHZ~18MHz) | |||
| Modes: | |||||
| ■ | B | √ | √ | ||
| ■ | M | √ | √ | ||
| ■ | PWD (Pulsed Wave Doppler) | √ | √ | ||
| ■ | CWD (Continuous WaveDoppler) | √ | √ | ||
| ■ | PW DTI (Doppler TissueImage) | √ | √ | ||
| ■ | Color Doppler | √ | √ | ||
| ■ | Power Doppler | √ | √ | ||
| ■ | Combined (BMDC) | √ | √ | ||
| Features: | |||||
| ■ | Harmonics | √ | √ | ||
| ■ | Compounding | √ | √ | ||
| ■ | UltraArt Universal Imageprocessing | √ | √ | ||
| ■ | Tissue Equalization (TEQ) | √ | √ | ||
| ■ | Biopsy | √ | √ | ||
| ■ | Clarify | √ | √ | ||
| ■ | Cardiac Imaging | √ | √ | ||
| ■ | Speed of Sound | √ | √ | ||
| ■ | Protocols | √ | √ | ||
| ■ | DICOM | √ | √ | ||
| ■ | eSie Calcs | √ | √ | ||
| ■ | Wireless | √ | √ | ||
| ■ | Virtual Touch Strain Imaging | √ | √ | ||
| ■ | Virtual Touch Point ShearWave (VTQ) | √ | √ | ||
| ■ | Virtual Touch Shear Wave(VTIQ) | √ | √ | ||
| ■ | Contrast Agent Imaging | √ | √ | ||
| ■ | eSie OB | √ | √ | ||
| ■ | eSie Left Heart | √ | √ | ||
| ■ | eSie Measure | √ | √ | ||
| ■ | eSie Follicle | √ | √ | ||
| ■ | Panoramic Imaging | √ | √ | ||
| ■ | Stress Echo | √ | √ | ||
| ■ | Syngo Velocity VectorImaging (VVI) | √ | √ | ||
| ■ | eSieLink | √ | √ | ||
| ■ | DICOM SR | √ | √ | ||
| Feature / Characteristic | ACUSON P200Subject device | ACUSON X800K180067 | ACUSONJuniperK1180039 | ACUSON S3000K183575 | |
| Touch Screen: adjustabledisplay | √ (13.3") | √ (15") | |||
| Output Display Standard(Track 3) | √ | √ | |||
| Patient Contact Materials | Tested to ISO10993-1 | Tested to ISO10993-1 | |||
| UL 60601-1 Certified | √ | √ | |||
| Indications for Use | √ | √ |
The submission device is substantially equivalent to the predicate with regards to both intended use and technological characteristics.
{19}------------------------------------------------
Siemens Medical Solutions USA, Inc. Ultrasound Division
{20}------------------------------------------------
7. A brief discussion of nonclinical tests submitted, referenced, or relied on in the 510(k) for a determination of substantial equivalence
The device has been evaluated for acoustic output, biocompatibility, cleaning and disinfection effectiveness as well as thermal, electromagnetic and mechanical safety and have been found to conform to applicable medical device safety standards. The systems comply with the following voluntary standards:
- 미 AIUM/NEMA UD-3:2004, Standard for Real-Time Display of Thermal and Mechanical Acoustic Output Indices on Diagnostic Ultrasound Equipment
- 미 AIUM/NEMA UD-2:2004 (R2009), Acoustic Output Measurement Standard for Diagnostic Ultrasound Equipment Revision 3
- 트 IEC 62359:2010, Ultrasonics - Field characterization - Test methods for the determination of thermal and mechanical indices related to medical diagnostic ultrasonic fields / This document and its separate amendments continue to be valid together with the consolidated version.
- 트 Safety and EMC Requirements for Medical Equipment
- AAMI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and o A2:2010/(R)2012 (Consolidated Text) Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1.2005)
- IEC 60601-1:2005, Medical electrical equipment Part 1: General requirements for o basic safety and essential performance / This document and its separate amendments continue to be valid together with the consolidated version
- IEC 60601-1-2 Edition 4.0 2014-02, Medical electrical equipment Part 1-2: General O requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
- IEC 60601-2-18: Edition 3.0 2009-08, Medical electrical equipment Part 2-18: O Particular requirements for the basic safety and essential performance of endoscopic equipment
- IEC 60601-2-37 Edition 2.1 2015, Medical electrical equipment Part 2-37: Particular O requirements for the basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment
- 트 ISO 10993-1 Fourth edition 2009-10-15, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process [Including: Technical Corrigendum 1 (2010)]
{21}------------------------------------------------
8. A summary discussion of the clinical tests submitted, referenced, or relied on for a determination of substantial equivalence.
Since the ACUSON P200 Diagnostic Ultrasound System uses the same technology and principles as existing devices, clinical studies were not required to support substantial equivalence.
9. Summary
Intended uses and other key features are consistent with traditional clinical practice and FDA guidelines. The design and development process of the manufacturer conforms to 21 CFR 820 Quality System Regulation and ISO 13485:2016 quality system standards. The product is designed to conform to applicable medical device safety standards and compliance is verified through independent evaluation with ongoing factory surveillance. Diagnostic ultrasound system has accumulated a long history of safe and effective performance. Therefore, it is the opinion of Siemens Medical Solutions USA, Inc. that the ACUSON P200 system is substantially equivalent with respect to safety and effectiveness to devices currently cleared for market.
§ 892.1550 Ultrasonic pulsed doppler imaging system.
(a)
Identification. An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.