K Number
K191912
Device Name
BiWaze Cough
Manufacturer
Date Cleared
2020-03-27

(254 days)

Product Code
Regulation Number
868.5905
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

This device is designed for use on patients unable to cough or clear secretions effectively due to reduced peak cough exspiratory flow, resulting from high spinal cord injuries, neuromuscular deficits or severe fatigue associated with intrinsic lung disease. It may be used either with a facemask, mouthpiece, or an adapter to a patient's endotrached tube or tracheostomy tube. For use in hospital, institutional setting, or at home. For use on adult patients 3 years old and up.

Device Description

The BiWaze Cough is a device intended for clearing bronchopulmonary secretions. The therapy provided by BiWaze Cough mimics a cough and consists of three phases which mimic a cough; inhale, exhale, and pause phase. The inhale phase is positive airway pressure to expand the lungs. Then the exhale phase is a sudden shift to negative pressure to pull the air out of lungs. Finally, the pause phase provides positive pressure which keeps the airways open in between the therapy cycle. The Peak Inspiratory Flow can be selected on three different levels: High, Medium, Low. The device can be operated via a power supply and is also battery operated. Performance is controlled from a touch screen panel in manual or automatic modes. The device is controlled by software algorithms, and error messages are displayed in cases where the normal functioning doesn't occur. Advanced features include inspiratory trigger and oscillations.

AI/ML Overview

The provided text does not contain information about acceptance criteria and the study proving a device meets these criteria in the context of an AI/Machine Learning device. Instead, it's a 510(k) summary for a medical device called "BiWaze Cough," which is a non-continuous ventilator.

The document discusses the device's substantial equivalence to predicate devices based on its technical characteristics, safety protocols, and performance data from bench testing and software code reviews, not from a clinical study involving human patients or ground truth established by experts.

Therefore, I cannot provide the requested information about acceptance criteria, test set sample size, expert ground truth, adjudication methods, MRMC studies, standalone performance, training set details, or how ground truth was established for a machine learning model, as this information is not present in the provided text for this specific device.

The "Performance Data" section explicitly states: "Performance testing was conducted on BiWaze Cough and the device was found substantially equivalent to the predicate device. Verification activities have been performed to verify that the device modifications did not affect the safety and effectiveness of the subject device. This included bench testing, software unit testing, and code reviews." This confirms that the evaluation was engineering-based for a physical device, not an AI/ML model.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of the square.

March 27, 2020

ABMRC LLC Sneha P OA/RA Lead 239. Fair Child Street. Daniel Island Charleston, South Carolina 29492

Re: K191912

Trade/Device Name: BiWaze Cough Regulation Number: 21 CFR 868.5905 Regulation Name: Noncontinuous Ventilator (IPPB) Regulatory Class: Class II Product Code: NHJ Dated: February 21, 2020 Received: February 26, 2020

Dear Sneha P:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Michael Ryan Division Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K191912

Device Name BiWaze Cough

Indications for Use (Describe)

This device is designed for use on patients unable to cough or clear secretions effectively due to reduced peak cough exspiratory flow, resulting from high spinal cord injuries, neuromuscular deficits or severe fatigue associated with intrinsic lung disease. It may be used either with a facemask, mouthpiece, or an adapter to a patient's endotrached tube or tracheostomy tube. For use in hospital, institutional setting, or at home. For use on adult patients 3 years old and up.

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D)☒ Prescription Use (Part 21 CFR 801 Subpart D)
☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)☐ Over-The-Counter Use (21 CFR 801 Subpart C)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

Submitter:ABMRC LLC239, Fair Child St, Daniel Island,Charleston, South Carolina 29492www.abmrespiratorycare.com
Contact:Chad M BoerstDirectorABMRC LLCEmail: chadboerst@abmrespiratorycare.com
Prepared By:Sneha S PQA/RA LeadABMRC LLCEmail: sneha.sp@abmrespiratorycare.com
Date Prepared:March 27, 2020
Device Proprietary Name:BiWaze Cough
Device Common Name:Secretion Clearance Device
Classification Regulation Number:21 CFR 868.5905
Classification Panel:Anesthesiology
Device Classification Code/Name:NHJ - Non-continuous ventilator (IPPB)
Classification:Class II
Predicate Device:Philips Respironics Cough Assist T70K12195-
Reference Devices:Pegaso CoughK140598
Lung Assist Vital CoughK132988

Device Description:

The BiWaze Cough is a device intended for clearing bronchopulmonary secretions. The therapy provided by BiWaze Cough mimics a cough and consists of three phases which mimic a cough; inhale, exhale, and pause phase. The inhale phase is positive airway pressure to expand the lungs. Then the exhale phase is a sudden shift to negative pressure to pull the air out of lungs. Finally, the pause phase provides positive pressure which keeps the airways open in between the therapy cycle.

{4}------------------------------------------------

The Peak Inspiratory Flow can be selected on three different levels: High, Medium, Low.

The device can be operated via a power supply and is also battery operated. Performance is controlled from a touch screen panel in manual or automatic modes. The device is controlled by software algorithms, and error messages are displayed in cases where the normal functioning doesn't occur. Advanced features include inspiratory trigger and oscillations.

Indications for Use:

This device is designed for use on patients unable to cough or clear secretions effectively due to reduced peak cough expiratory flow, resulting from high spinal cord injuries, neuromuscular deficits or severe fatigue associated with intrinsic lung disease. It may be used either with a facemask, mouthpiece, or an adapter to a patient's endotracheal tube or tracheostomy tube. For use in hospital, institutional setting, or at home. For use on adult patients and pediatric patients 3 years old and up.

Substantial Equivalence Determination:

The table below summarizes the key technical characteristics of BiWaze Cough and the predicate and reference devices listed in the submission:

{5}------------------------------------------------

TechnologicalCharacteristicBiWaze CoughK191912Philips RespironicsCoughAssist T70K121955Pegaso CoughK140598Lung Assist Vital CoughK132988
CFR Classification868.5905 NHJ868.5905 NHJ868.5905 NHJ868.5905 NHJ
Classification Panel andClassAnesthesiologyClass IIAnesthesiologyClass IIAnesthesiologyClass IIAnesthesiologyClass II
Classification NameDevice, positive pressurebreathing, intermittent(IPPB)Device, positive pressurebreathing, intermittent(IPPB)Device, positivepressure breathing,intermittent (IPPB)Device, positive pressurebreathing, intermittent(IPPB)
Environments of UseHospital, InstitutionalSetting, homecareHospital, InstitutionalSetting, homecareHospital, InstitutionalSetting, homecareHospital, InstitutionalSetting, homecare
Patient PopulationAdult or pediatric patientunable to cough or clearsecretions effectivelyAdult or pediatric patientunable to cough or clearsecretions effectivelyAdult or pediatricpatient unable to coughor clear secretionseffectivelyFor use on adult orpediatric patients
Delivery TypeNon-Invasive or InvasiveNon-Invasive or InvasiveNon-Invasive orInvasive-
Modes of OperationAuto and ManualAuto and ManualAuto and ManualAuto and Manual
Inhalation Pressure0 to 70 cmH2O0 to 70 cmH2O0 to 70 cmH2O0 to 50 cmH2O
Exhalation Pressure0 to -70 cmH2O0 to -70 cmH2O0 to -70 cmH2O0 to -50 cmH2O
Inhale FlowLow, Medium, HighLow, Medium, HighLow, Medium, HighLow, Medium, High
Pause Time0 to 5 seconds0 to 5 seconds0 to 9.9 seconds0 to 5 seconds
Phases of Therapy CycleInsufflation, Exsufflation,PauseInsufflation, Exsufflation,PauseInsufflation,Exsufflation, PauseInsufflation, Exsufflation,Pause
Safety protocolsDynamic Flow and Pressurecontrol. Sensor malfunctiondetection. Altitudecorrection. Pressure andDynamic stability Analysis.Flow and Pressure basedOscillation DetectionExtreme Flow Rate Controland Response SensorDynamic Flow andPressure control.Manufacturer SoftwareCalibration eliminates-
Flow rate control forsmooth pressure outputwithout fluctuations. Valvedesign ensures spontaneousbreathing in event of powerfailure/standby whenpatient is connectedMalfunction Stabilityall undesiredoscillations. Sensormalfunction detection
Oscillations/VibrationsFrequencyUp to 20 HzN/AUp to 10 HzUp to 20 Hz
Therapy Features –Inspiratory TriggerYesYes (CoughTrak)Yes (Easy Start/AutoSync)-
Remote Data AccessAn internal memory storesall data. Encrypted andsecure data transfer throughUSB (2.0) and Wi-Fi(WiLink8 802.11 a/b/g/n +MIMO)Secure Digital (SD) cardAn internal memorystores therapies data.RS232/USB adaptertransmits to a PCtherapies and technicaldata.-
Remote ControlA remote-control accessory(foot pedal) and/oriOS/Android app is providedto initiate manual therapy bymeans of a remote-controlinterfaceA remote-control accessory(foot pedal) is provided toinitiate manual therapy bymeans of a wired remote-control interface-
Energy Source100-240 V ac50/60 Hz100-240 V ac50/60 Hz100/240 V ac50/60 Hz100-230 V ac50-60 Hz
Patient InterfacesFacemask, Mouthpiece, anadapter to endotracheal tubeor tracheostomy tubeFacemask, Mouthpiece,endotracheal tube ortracheostomy tubeFacemask, Mouthpiece,an adapter toendotracheal tube ortracheostomy tubeFacemask, Mouthpiece, anadapter to endotrachealtube or tracheostomy tube

{6}------------------------------------------------

{7}------------------------------------------------

Comparison of Characteristics with respect to Predicate Device:

The BiWaze Cough device has similar features and indications for use when compared to the Predicate Device, Phillips Respironics Cough Assist T70 and the reference devices, Pegaso Cough and Lung Assist Vital Cough. The core capabilities of the BiWaze Cough and its fundamental scientific technology remain unaltered compared to the Phillips Respironics Cough Assist T70. The device modifications discussed do not alter the BiWaze Cough's safety or effectiveness and neither do they change its intended use compared to the predicate.

The table below provides a description of the modifications to the BiWaze Cough device that are subject of this 510(k) submission:

Device FeaturesDescription
User InterfaceA new graphic touch display and a new user interface with hierarchical menu system.
SIMILARITIESThe predicate device has similar parameters displayed in the main screen.
InspiratoryTriggerA feature for patients who can provide spontaneous breathing effort to trigger the Cough sequence, instead of using manual or auto modes therapy. This software feature monitors the device outlet flow and pressure and initiates the insufflation phase of therapy delivery when the flow increases over a set threshold indicative of patient effort.
SIMILARITIESThe Inspiratory Trigger feature is based on the detection of the patient's inspiratory efforts. For this detection a digital flow and pressure sensor is used. The principle is similar to the Respironics CoughAssist T70's CoughTrak and Pegaso Cough's Auto Sync/Easy Start.
Custom CycleSettingsA feature to customize each breath cycle in terms of pressure, time and oscillations.
REMARKSThis feature is added to provide more flexibility in terms of setting therapy parameters for every individual cycle. This feature is similar to Respironics' T70 Advanced Automatic Mode.The device can save up to 10 custom therapy profiles.
Custom ProfilesSIMILARITIESThe Custom Profiles feature is provided so that User can select a pre-defined set of therapy parameters. This feature is similar to Respironics CoughAssist T70's Preset feature.
A feature that delivers a pressure oscillation based on frequency and amplitude set points.
OscillationsSIMILARITIESThe feature is similar to the Percussor feature of the Pegaso Cough device.
Data ManagementAll therapy data is stored in an internal memory. On request, data can be securely transferred over authenticated link through USB (2.0), Wi-Fi (WiLink8 802.11 b/g/n)
REMARKSEncrypted data is securely transferred through either USB 2.0, Wi-Fi (WiLink8 802.11 b/g/n) along with additional authentication information.
Wired Remote ControlA remote-control accessory (foot pedal) is provided to initiate manual therapy by means of a wired remote-control interface.
SIMILARITIESThis feature/accessory is similar to that of Respironics CoughAssist T70 device.
Wireless Remote ControlA remote-control accessory (mobile app) is provided to initiate manual therapy by means of a wireless (BLE) remote-control interface.
Power ManagementAn internal rechargeable battery is included.

{8}------------------------------------------------

Performance Data:

Performance testing was conducted on BiWaze Cough and the device was found substantially equivalent to the predicate device.

Verification activities have been performed to verify that the device modifications did not affect the safety and effectiveness of the subject device. This included bench testing, software unit testing, and code reviews.

The BiWaze Cough device was designed and tested according to the following standards:

  • AAMI ANSI ES 60601-1 Medical Electrical Equipment Part 1: General Requirements . for Basic Safety and Essential Performance
  • IEC 60601-1-2 Medical Electrical Equipment Part 1-2: General Requirements for ● Basic Safety and Essential Performance - Collateral Standard: Electromagnetic Disturbances - Requirements and Tests
  • . IEC 60601-1-6 Medical Electrical Equipment - Part 1-6: General Requirements for Basic Safety and Essential Performance - Collateral Standard: Usability
  • IEC 60601-1-11 Medical Electrical Equipment - Part 1-11: General Requirements for Basic Safety and Essential Performance - Collateral Standard: Requirements for Medical Electrical Equipment and Medical Electrical Systems Used in the Home Healthcare Environment
  • IEC 62366-1 Medical Devices Part 1: Application of Usability Engineering to Medical ● Devices
  • . ISO 10993-1 Biological evaluation of medical devices - Part 1: Evaluation and Testing within a Risk Management Process
  • ISO 18562-1 Biocompatibility Evaluation of Breathing Gas Pathways in Healthcare ● Applications - Part 1: Evaluation and Testing within a Risk Management Process

{9}------------------------------------------------

  • IEC 62304 Medical Device Software Software Life Cycle Processes ●
  • ISO 14971 Medical Devices Application of Risk Management to Medical Devices ●
  • IEEE C63.27:2017 American National Standard for Evaluation of Wireless Coexistence ●

Conclusion:

In summary, bench testing and software code reviews have confirmed that the BiWaze Cough device performs substantially equivalent to the cited predicate device. The indications for use, technological characteristics, and principles of operation are similar to the predicate device. The BiWaze Cough device is substantially equivalent to the predicate device.

§ 868.5905 Noncontinuous ventilator (IPPB).

(a)
Identification. A noncontinuous ventilator (intermittent positive pressure breathing-IPPB) is a device intended to deliver intermittently an aerosol to a patient's lungs or to assist a patient's breathing.(b)
Classification. Class II (performance standards).