(165 days)
The Clear-Aligner system is indicated for the treatment of malocclusion in patients with permanent dentition. The Clear-Aligner system positions teeth by way of continuous gentle force.
A dental health care professional (e.q. orthodontist or dentist), using a standard personal computer prescribes the Clear-Aligner system based on an assessment of thee patient's teeth, determines a course of treatment with the system, takes molds of the patient's teeth, and completes a prescription form. Utilizing standard dental software used for tooth alignment, CDB then designs a series of plastic trays intended to gradually realign the patient's teeth in accordance with the physician's prescription. The prescribing physician reviews and approves the treatment plan before the molds are produced. Once approved, CDB produces the trays, which are formed of clear, thin, thermoformed, copolyester plastic. The trays sent back to the dental health care professional, who then provides them to the patient, confirming fit and design. Over a period of months, additional trays are provided sequentially to the patient by the dental health professional to gradually move the target teeth to the desired position. The dental care professional monitors treatment from the moment the first aligner is delivered to when the final aligner is delivered. The trays are held in place by pressure and can be removed by the patient at any time. This technology is essentially identical to that used by a number of sequential alignment systems, including the predicates referenced in Table 1.
The provided text describes the acceptance criteria and study for the "Clear-Aligners" device. However, it does not detail a multi-reader multi-case (MRMC) comparative effectiveness study or a standalone algorithm performance study, as the device is a physical product (orthodontic aligners) rather than a diagnostic AI. The "performance testing" section refers to bench testing of the manufacturing process, not clinical performance for diagnostic accuracy.
Here's a summary of the available information:
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria | Reported Device Performance |
|---|---|
| Functional Characteristics: | |
| Fit of the aligner against the treatment plan as determined by a dental health professional | The acceptance results observed during verification and validation testing have shown that the devices are capable of performing to their stated intended use and specification requirements. |
| Process consistency from the treatment plan file to 3D printed mold to final aligner | Measured against internal acceptance criteria derived by considering all sources of variation. The acceptance results observed during verification and validation testing have shown that the devices are capable of performing to their stated intended use and specification requirements. |
| Biocompatibility Testing (for mucosal membrane contacting for >30 days): | |
| Cytotoxicity | Met the requirements of the study protocols; material is considered non-cytotoxic. |
| Sensitization | Met the requirements of the study protocols; material is considered non-sensitizing. |
| Irritation | Met the requirements of the study protocols; material is considered not an intracutaneous irritant. |
| Software Functionality (for ordering process and treatment planning): | Software usage is considered of moderate concern. Software testing has been performed to verify the function, supporting a substantial equivalence decision. (Specific performance metrics for software are not provided in this document). |
2. Sample Size Used for the Test Set and Data Provenance
The document does not explicitly mention a "test set" in the context of clinical data, as this is a device for physical tooth alignment, not a diagnostic algorithm. The performance testing refers to bench testing for manufacturing and fit. Therefore, information on data provenance (country of origin, retrospective/prospective) for a clinical test set is not applicable or provided.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
Not applicable. The performance testing involves "dental health professional" assessment of aligner fit, and manufacturing process consistency, not diagnostic expert consensus for a clinical ground truth.
4. Adjudication Method for the Test Set
Not applicable. There is no mention of a diagnostic test set requiring adjudication.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No such study was conducted or is applicable to this device, as it is a physical orthodontic aligner, not an AI-powered diagnostic tool. The document focuses on the substantial equivalence of the aligners themselves to predicate devices in terms of function, materials, and manufacturing process.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. The Clear-Aligners are a physical medical device. The software mentioned is for ordering and treatment planning, not for standalone diagnostic performance.
7. The Type of Ground Truth Used
For the functional characteristics (fit, process consistency), the "ground truth" seems to be the pre-established internal specification and acceptance criteria, along with assessment by a dental health professional. For biocompatibility, the ground truth is established by the results of standardized biological tests (Cytotoxicity, Sensitization, Irritation).
8. The Sample Size for the Training Set
Not applicable. There is no "training set" in the context of an AI/ML algorithm for this physical device.
9. How the Ground Truth for the Training Set Was Established
Not applicable.
Overall Study Information:
The study described is primarily focused on bench testing to ensure the manufacturing accuracy and functional characteristics of the Clear-Aligners, and biocompatibility testing of the materials used. The approach taken is to demonstrate substantial equivalence to already legally marketed predicate devices (Clear Correct and Invisalign) based on similar indications for use, technological characteristics, mechanism of action, materials, and performance. The document explicitly states: "Further testing was not performed given there is sufficient information available from the scientific literature and from the predicate devices to demonstrate that sequential aligners provide reasonable assurance of substantial equivalence."
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 20, 2019
CDB Corporation Leah Lehman Quality & Regulatory Manager 9201 Industrial Blvd Leland, North Carolina 28451
Re: K191823
Trade/Device Name: Clear-Aligners Regulation Number: 21 CFR 872.5470 Regulation Name: Orthodontic Plastic Bracket Regulatory Class: Class II Product Code: NXC Dated: November 18, 2019 Received: November 19, 2019
Dear Leah Lehman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Srinivas "Nandu" Nandkumar, Ph.D. Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K191823
Device Name Clear-Aligners
Indications for Use (Describe)
The Clear-Aligner system is indicated for the treatment of malocclusion in patients with permanent dentition. The Clear-Aligner system positions teeth by way of continuous gentle force.
Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/2 description: The image shows the logo for CDB Ceramic Dental Brackets. The logo consists of a stylized blue triangle on the left and the text "CDB CERAMIC DENTAL BRACKETS" on the right. Below the logo is the text "510(k) Summary" in a large, bold font.
This 510(k) Summary is submitted in accordance with the requirements of 21CFR 807.92.
| Submitter: | CDB Corporation9201 Industrial Blvd, NELeland, NC 28451 USAOwner/Operator No.: 9002860Establishment Registration No.: 1054415 |
|---|---|
| Primary Contact: | Leah M LehmanQuality & Regulatory ManagerPhone: 910.383.6464llehman@cdbcorp.net |
| Secondary Contact: | Jens RumsfeldManaging DirectorPhone: 910.383.6464jrumsfeld@cdbcorp.net |
| Date Prepared: | December 19, 2019 |
| 510k Submission: | K191823 |
| Device Name: | Clear-Aligners |
| Common Name: | Sequential Aligner |
| Classification Name: | Orthodontic Plastic Bracket |
| Regulation Number: | 21 CFR 872.5470 |
| Product Code: | NXC |
| Classification Panel: | Dental Products Panel 76 |
| Classification: | Medical Device, Class II |
Indications for Use:
The Clear-Aligner system is indicated for the treatment of malocclusion in patients with permanent dentition. The Clear-Aligner system positions teeth by way of continuous gentle force.
{4}------------------------------------------------
Predicate Devices:
Clear Correct, Clear Correct Inc., K082556 (primary predicate) Invisalign, Align Technology, K981095 (reference device)
Device Description:
A dental health care professional (e.q. orthodontist or dentist), using a standard personal computer prescribes the Clear-Aligner system based on an assessment of thee patient's teeth, determines a course of treatment with the system, takes molds of the patient's teeth, and completes a prescription form. Utilizing standard dental software used for tooth alignment, CDB then designs a series of plastic trays intended to gradually realign the patient's teeth in accordance with the physician's prescription. The prescribing physician reviews and approves the treatment plan before the molds are produced. Once approved, CDB produces the trays, which are formed of clear, thin, thermoformed, copolyester plastic. The trays sent back to the dental health care professional, who then provides them to the patient, confirming fit and design. Over a period of months, additional trays are provided sequentially to the patient by the dental health professional to gradually move the target teeth to the desired position. The dental care professional monitors treatment from the moment the first aligner is delivered to when the final aligner is delivered. The trays are held in place by pressure and can be removed by the patient at any time. This technology is essentially identical to that used by a number of sequential alignment systems, including the predicates referenced in Table 1.
Technoloqical Characteristics:
Treatment of tooth malocclusions via a series of plastic appliances that incrementally moves teeth to a desired end-state is the technological principle for both the subject and predicate devices. A comparison between technological characteristics of the Clear-Aligners and that of legally marketed predicate devices has been performed. The results of this comparison demonstrate that the design, technology, materials, and composition of the Clear-Aligners are substantially equivalent to the predicate devices.
{5}------------------------------------------------
Mechanism of Action:
The mechanism of action is similar to the predicate devices and supports a determination of substantial equivalence. Orthodontic tooth movement occurs through forces applied by the appliance to the dentition as each tooth follows the programmed displacement based on a dental health professional's prescription.
Performance Testing:
Bench testing was performed to ensure the accuracy of the final product carried through the entire process from the initial scan through treatment planning and manufacturing of the model to the final thermoformed aligner. Functional characteristics tested and the standard by which they were tested include the following, all pre-established specification and acceptance criteria:
- . Fit of the aligner against the treatment plan as determined by dental health professional
- Process consistency from the treatment plan file to 3D printed mold to final . aligner was measured against internal acceptance criteria derived by considering all sources of variation
The acceptance results observed during the verification and validation testing that has been conducted on these devices has shown that they are capable of performing to their stated intended use and specification requirements.
Further testing was not performed given there is sufficient information available from the scientific literature and from the predicate devices to demonstrate that sequential aligners provide reasonable assurance of substantial equivalence.
{6}------------------------------------------------
Biocompatibility Testing:
The biological evaluation for the device was conducted in accordance with the US FDA Docket Number FDA-2013-D-0350, "Use of International Standard ISO 10993-1, Biological evaluation and testing within a risk management process - Guidance for Industry and Food and Drug Administration Staff" as recognized by FDA. The aligner is considered mucosal membrane contacting for a duration of greater than 30 days. The battery of testing included the following tests:
- Cytotoxicity ●
- Sensitization
- Irritation ●
The results of the testing met the requirements of the study protocols and the material is considered non-cytotoxic, non-sensitizing, and is not an intracutaneous irritant. The results of the studies further support the determination of substantial equivalence.
Software Consideration
There are two areas in the Clear-Aligner process where software is used; they are the ordering process and treatment planning. The software usage in these areas is considered of moderate concern. Software testing has been performed to verify the function of the Clear-Aligners, which supports a substantial equivalence decision.
Substantial Equivalence Comparison:
The following table compares the Clear-Aligner system to the predicate devices, Clear Correct (primary predicate) and the Align System (reference device), with respect to intended use, technological characteristics, and principles of operation.
{7}------------------------------------------------
| Feature | Clear-AlignersSubmission Device | Clear CorrectPrimary Predicate | Align SystemReference Device |
|---|---|---|---|
| 510(k) Number | K191823 | K082556 | K981095 |
| Manufacturer | CDB Corporation | Clear Correct Inc. | Align Technology |
| Regulation Number | 872.5470 | 872.5470 | 872.5470 |
| Device Classification | Aligner, Sequential | Aligner, Sequential | Aligner, Sequential |
| Product Code | NXC | NXC | NXC |
| Device Class | II | II | II |
| Indications for Use | The Clear-Alignersystem is indicatedfor the treatment ofmalocclusion inpatients withpermanent dentition.The Clear-Alignersystem positionsteeth by way ofcontinuous gentleforce. | The Clear CorrectSystem is indicatedfor the treatment oftooth malocclusion inpatients withpermanent dentition(i.e. all secondmolars). The ClearCorrect Systempositions teeth byway of continuousgentle force. | The Align System isindicated for thetreatment of toothmalocclusion in patientswith permanent dentition(i.e. all second molars).The Align Systempositions teeth by wayof continuous gentleforce. |
| Mode of Action | Alignment of teeth bysequential use ofpreformed plastictrays. | Alignment of teeth bysequential use ofpreformed plastictrays. | Alignment of teeth bysequential use ofpreformed plastic trays. |
| Material | ThermoformedCopolyester | ThermoformedPolycarbonate | ThermoformedPolycarbonate |
| Material Properties | Demonstratessufficient tensilestrength, elasticity,ductility, chemicalresistance, andclarity for use as aclear tray aligner | Demonstratessufficient tensilestrength, elasticity,ductility, chemicalresistance, and clarityfor use as a clear trayaligner | Demonstrates sufficienttensile strength,elasticity, ductility,chemical resistance,and clarity for use as aclear tray aligner |
| Software Used forOrdering Workflow | Yes | Yes | Yes |
| OTC or Rx | Rx | Rx | Rx |
Table 1. Predicate Device Information Comparison
Substantial Equivalence Conclusion:
The conclusion drawn from the data included in this submission demonstrates that the Clear-Aligner system is substantially equivalent to the predicate devices in indications for use, design, technological characteristics, mechanism of action, performance, materials, and biocompatibility.
§ 872.5470 Orthodontic plastic bracket.
(a)
Identification. An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.(b)
Classification. Class II.