(199 days)
The intended use of Blue Phantom PT is to move detectors in one dimension within a water tank. It transfers indicated ionization beams (proton beams) at a certain position into electrical signals.
The Blue Phantom PT system consists of a water tank with a one-dimensional servo including a common control unit (CCU PT) with integrated two-channel electrometer. An ionisation chamber (e.g. Stingray) is placed inside the Blue Phantom PT with a holder and is used as detector for measurements.
The Blue Phantom PT is intended to be used in the magnetic environment of a proton therapy treatment unit and its technology has been designed to meet those requirements.
On the horizontal Y-rail, detector holders for various detectors can be mounted on a sliding shoe. The detector can be positioned in the Y-direction for measuring horizontal beams.
The provided text is a 510(k) summary for the Blue Phantom PT device and discusses its substantial equivalence to a predicate device (WP 1D). It outlines performance testing but does not contain detailed acceptance criteria and performance data in the structured format requested. However, it does provide some device characteristics that can be interpreted as performance metrics.
Here's an attempt to extract and organize the information based on the provided text, recognizing that some requested details are not explicitly present:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly state "acceptance criteria" but rather lists "Device Characteristics" for both the predicate and the new device. We can infer that the Blue Phantom PT aiming to at least match or improve upon the predicate's characteristics constitutes a form of meeting "acceptance criteria."
| Characteristic | Acceptance Criteria (Inferred from Predicate WP 1D) | Reported Device Performance (Blue Phantom PT) |
|---|---|---|
| Position resolution | 0.1 mm | 0.1 mm |
| Position accuracy | ± 0.4 mm | ± 0.05 mm |
| Reproducibility | ± 0.1 mm | ± 0.03 mm |
| Maximum scan range | 25 cm | 38 cm max. |
| Positioning speed | 25 mm/s | Max. 25 mm/s |
| Scanning speed (continuous measurement) | Not applicable, not part of application scope (WP 1D) | max. 20 mm/s (Blue Phantom PT) |
| Electrometer | Separate device | Embedded in control unit CCU |
| Sensortype | Magneto-restrictive | Inductive |
| Radiation Type | Ionizing radiation (photon, electron), excluding proton | Proton beams |
2. Sample Size Used for the Test Set and Data Provenance
The document states: "BP PT was successfully tested to demonstrate safety and effectiveness and substantial equivalence to the predicate device. It was subject to the following tests: System test, Clinical environment test, Usability test, Unit test (Mechanics), Firmware test, Non-clinical test against the following standards."
- Test Set Sample Size: Not specified.
- Data Provenance: The document implies internal testing ("Internal testing") and testing by "external test house" for some standards. There is no information on country of origin or whether the data was retrospective or prospective in the context of device performance data, as this is a physical measurement device, not an AI/software device processing patient data.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
This section is not applicable (N/A) as the device is a physical measurement tool, not an AI or diagnostic device that relies on expert ground truth for interpretation of medical data. The ground truth would be physical standards or direct measurements.
4. Adjudication Method for the Test Set
N/A. This concept is typically relevant for studies involving human interpretation or subjective assessments, not for objective physical measurements of device characteristics. The "ground truth" for the performance metrics would be derived from calibrated measurement instruments, not human adjudication.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance
N/A. The Blue Phantom PT is a physical device for radiation dosimetry, not an AI-assisted diagnostic tool. Therefore, an MRMC study is not relevant.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
N/A. This refers to AI algorithm performance. The device is a hardware system with embedded firmware. Its performance is inherent to its design and operation.
7. The Type of Ground Truth Used
The ground truth for the performance metrics (e.g., position accuracy, reproducibility) would be established through physical standards, highly accurate calibration systems, and traceable measurement techniques. The document mentions testing against "standards" (e.g., IEC, ISO), which implies adherence to methods and limits defined by these standards.
8. The Sample Size for the Training Set
N/A. This is a physical device, not a machine learning model that requires a training set.
9. How the Ground Truth for the Training Set Was Established
N/A. (See point 8).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION".
January 23, 2020
IBA Dosimetry GmbH % Olaf Teichert Official Correspondent TUV SUD America, Inc. 1775 Old Highway 8 NW NEW BRIGHTON, MN 55112
Re: K191821
Trade/Device Name: Blue Phantom PT Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: Class II Product Code: LHN Dated: June 27, 2019 Received: July 8, 2019
Dear Olaf Teichert:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For
Thalia Mills, Ph.D. Chief Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K191821
Device Name Blue Phantom PT
Indications for Use (Describe)
The intended use of Blue Phantom PT is to move detectors in one dimension within a water tank. It transfers indicated ionization beams (proton beams) at a certain position into electrical signals.
Non-intended use of Blue Phantom PT
The Blue Phantom PT is not meant for measurement orthogonal to the proton beam (inline and crossline measurements).
| Type of Use (Select one or both, as applicable) | |
|---|---|
| ☒ Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image is a logo for an organization, possibly a company or non-profit. The logo features the letters "Iba" in a stylized, handwritten font. Above the letters, there is a curved line with a dot on top, resembling a person with their arms raised. The background is a solid, bright green color.
510(K) SUMMARY Blue Phantom PT
Image /page/3/Picture/2 description: The image shows the address of IBA Dosimetry GmbH. The address is Bahnhofstrasse 5, 90592 Schwarzenbruck, Germany. The company name is in a larger font than the address.
Image /page/3/Picture/3 description: The image contains three lines of text in a sans-serif font. The first line reads "PROTECT," the second line reads "ENHANCE," and the third line reads "SAVE LIVES." A plus sign is present to the right of the words "PROTECT" and "ENHANCE."
{4}------------------------------------------------
510(k) Summary
| Date of preparation | 1/22/2020 |
|---|---|
| Submitter | IBA Dosimetry GmbHBahnhofstr. 590592 SchwarzenbruckGermany |
| Contact Person | Andreas SuchiExecutive Director QRSAndreas.Suchi@iba-group.comPhone: +49 9128 607-943Fascimile: +49 9128 607-10 |
| Common Name | BP PT |
| Classification Name | Medical charged-particle radiation therapysystem |
| Trade Name | Blue Phantom PT |
| Class | II |
| Panel | Radiology |
| Classification regulation | 21CFR 892.5050 |
| Product Code | LHN |
| Predicate Device | WP 1D (K032594) |
1. Intended Use
1.1. Intended use of Blue Phantom PT
The intended use of Blue Phantom PT is to move detectors in one dimension within a water tank. It transfers indicated ionization beams (proton beams) at a certain position into electrical signals. Non-intended use of Blue Phantom PT:
The Blue Phantom PT is not meant for measurement orthogonal to the proton beam (inline and crossline measurements).
1.2. Intended use of WP 1D
The Clinical Reference Dosimetry Phantom WP1D is used to position various radiation detectors in water or air. It consists of a cubic tank and precision one-dimensional hand crank or motor driven servo. By design it is suitable to act as a phantom according to various dosimetric protocols (e.g. AAPM´s TG-51 or IAEA´s TRS-398). The device is intended to be used by experienced professionals entrusted with dosimetric functions only.
Image /page/4/Picture/11 description: The image shows a logo for an organization, possibly named "Iba". The logo is white and slightly blurred, set against a solid green background. The logo appears to be stylized text, with a small circle above the first letter, possibly representing a dot over the "i".
{5}------------------------------------------------
2. Product Description
2.1. Blue Phantom PT
The Blue Phantom PT system consists of a water tank with a one-dimensional servo including a common control unit (CCU PT) with integrated two-channel electrometer. An ionisation chamber (e.g. Stingray) is placed inside the Blue Phantom PT with a holder and is used as detector for measurements.
The Blue Phantom PT is intended to be used in the magnetic environment of a proton therapy treatment unit and its technology has been designed to meet those requirements.
Figure 1 shows the Blue Phantom PT and the CCU. Figure 1: BluePhantom PT with CCU
Image /page/5/Picture/6 description: The image shows two medical devices, the Blue Phantom PT water phantom and the CCU PT. The Blue Phantom PT water phantom is a clear, rectangular container filled with water. It has a thin window on one side and handles on top. A Y-axis servo with a motor is attached to the top of the container, and a Stingray chamber is visible inside. The CCU PT is a white, rectangular box with the letters CCU on the front.
On the horizontal Y-rail, detector holders for various detectors can be mounted on a sliding shoe. The detector can be positioned in the Y-direction for measuring horizontal beams.
2.2. WP 1D
The WP1D phantom consists of a cubic tank and a one-dimensional moving mechanics to move the detector up and down along the Z-axis. The tank has a water inlet/outlet that is equipped with a quick coupler for easy connection of the water-filling tube. Three adjustable feet support the tank and provide horizontal leveling adjustment.
3. Device Characteristics
| Characteristic | WP 1D | Blue Phantom PT |
|---|---|---|
| Operating Principle | WP 1D is a one-dimensionalwater phantom intended forreference dosimetry. A detector ispositioned in the phantom andcan be adjusted in differentdepths either electronically ormanually with a hand crank.In case of electronic positioning,the OmniPro-Accept 6 (K011763) | BP PT is a one-dimensionalwater phantom intended forreference dosimetry. A detectoris positioned in the phantom andis electronically adjusted indifferent depths.The myQA Accept (K011763)software is used for depth dose |
Table 1: Device Characteristics
(ba
{6}------------------------------------------------
| software is used for depth dosemeasurement and controlling thedetectors positioning. | measurement and controllingthe detectors positioning. | |
|---|---|---|
| Sensortype | Magneto-restrictive | Inductive |
| Electrometer | Separate device | Embedded in control unit CCU |
| Tank size | 36x42x36 cm | 51.4 x 27.7 x 26.8 cm |
| Tank wall material | Acrylic (PMMA) | Acrylic (PMMA) |
| Wall thickness | 10mm | 10mm |
| Thin window | N/A | 5mm / 3mm |
| Position resolution | 0.1mm | 0.1mm |
| Position accuracy | $\pm$ 0.4 mm | $\pm$ 0.05 mm |
| Reproducibility | $\pm$ 0.1 mm | $\pm$ 0.03 mm |
| Maximum scan range | 25 cm | 38 cm max. |
| Positioning speed | 25mm/s | Max. 25mm/s |
| Scanning speed(continuous measurement) | Not applicable, this is not part ofits application scope | max. 20mm/s |
4. Performance testing
BP PT was successfully tested to demonstrate safety and effectiveness and substantial equivalence to the predicate device. It was subject to the following tests:
- -System test
- Clinical environment test -
- -Usability test
- -Unit test (Mechanics)
- Firmware test -
- -Non-clinical test against the following standards:
| Standard | Test Method | Compliance |
|---|---|---|
| ISO 14971:2007 | Internal testing | yes |
| IEC 61010-1: 2010 | Tested by external test house | yes |
| IEC 62304:2006 + A1:2015 | Internal testing; FDA guidance documents have been applied | yes |
| IEC 62366-1:2015 | Internal testing | yes |
| IEC 61326-1:2012 | Tested by external test house | yes |
| AAMI RT2:2017 | Internal testing | yes |
Animal and clinical test were not required to demonstrate safety and effectiveness.
Requirements of the tests were met as specified in the test requirements and the applied standards.
Image /page/6/Picture/13 description: The image shows a logo with the letters 'iba' in a stylized, handwritten font. The letters are white and appear to be glowing slightly against a solid green background. The 'i' has a dot above it that is also stylized and slightly larger than a typical dot.
{7}------------------------------------------------
5. Conclusion
The operating principle of the BP PT and its predicate is identical and the performed tests proof that the differences between BP PT and WP 1D do not raise new questions of safety and effectiveness. The indication for use statements are slightly different, since BP PT is used for proton beams, while WP 1D is used for ionizing radiation like photon and electron radiation, but excluding proton radiation. This difference in radiation type is compensated with an equally safe and effective new inductive sensor.
Despite, the intended use of BP PT is more limited than of WP 1D. The evaluation and performed tests provide reasonable assurance that the differences are not critical to the intended use of the device, and that they do not affect the safety and effectiveness of the device when used as labeled.
Image /page/7/Picture/4 description: The image shows a logo with the word "Iba" in a stylized, handwritten font. The text is white and appears against a solid green background. The logo is simple and clean, with a focus on the text itself.
§ 892.5050 Medical charged-particle radiation therapy system.
(a)
Identification. A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.(b)
Classification. Class II. When intended for use as a quality control system, the film dosimetry system (film scanning system) included as an accessory to the device described in paragraph (a) of this section, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.