K Number
K191709
Manufacturer
Date Cleared
2020-03-05

(253 days)

Product Code
Regulation Number
888.3027
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

V-STRUT© Vertebral Implant is indicated for use in the treatment of vertebral fractures in the thoracic and lumbar spine from T9 to L5. It is intended to be used in combination with Teknimed F20® bone cement.

Device Description

V-STRUT© Vertebral Implant is part of V-STRUT© Transpedicular Vertebral System. V-STRUT© Transpedicular Vertebral System is composed of :

  • V-STRUT© Vertebral Implant,
  • -V-STRUT© Guide Wire,
  • -V-STRUT© Instrumentation Kit.
    V-STRUT© Vertebral Implant is a medical device to be placed in the vertebrae through a minimally invasive procedure. Two devices are implanted in each vertebra to be treated. Each implant is introduced posteriorly through the pedicle up to the anterior vertebral body wall.
    The implant is made of radio transparent polymer, PEEK (Polyetheretherketone as per ASTM F2026) and includes two visualizing markers made of tantalum (as per ASTM F560).
    V-STRUT© Vertebral Implant exists in 2 different diameters and 5 different lengths to accommodate individual patient's anatomy of thoracic and/or lumbar vertebrae from T9 to L5.
    V-STRUT© Vertebral Implant is provided sterile and is not reusable.
    V-STRUT© Vertebral Implanted using specific instrumentation provided with the implant and is combined in situ with a Polymethylmethacrylate (PMMA) bone cement, F20® manufactured by Teknimed SA (K103433).
AI/ML Overview

This document is a 510(k) summary for the V-STRUT© Vertebral Implant, which is a medical device. This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device rather than proving clinical effectiveness through a complex clinical trial with specific acceptance criteria for AI performance as would be seen for an AI/ML medical device.

Therefore, the requested information about acceptance criteria and the study proving the device meets them, particularly in the context of an AI/ML device (e.g., sample size, expert ground truth, MRMC studies, standalone performance), is not applicable to this 510(k) submission.

This document describes a traditional medical device (an implant) and its substantial equivalence is demonstrated through:

  • Biomechanical Performance Testing: Static and dynamic mechanical testing.
  • Biocompatibility Testing: According to ISO 10993-1, 10993-5, and USP 42-NF 37.
  • Sterility Testing: According to ISO 11137-1 and 11137-2.
  • Shelf-Life Testing: According to ASTM F1980.
  • Clinical Evaluation: Stating that "V-STRUT© Vertebral Implant has demonstrated to be substantially equivalent to its predicate device KIVA® VCF Treatment System. No additional safety or effectiveness issues as the ones expected and documented in the literature for predicates and references devices have been raised." This is not a prospective clinical trial with specific performance metrics designed to prove the device meets acceptance criteria related to a specific diagnosis or outcome as would be expected for an AI device.

To directly answer your numbered questions based on the provided document:

  1. A table of acceptance criteria and the reported device performance: This document does not contain such a table because it's for a physical implant, not an AI/ML diagnostic device with performance metrics like sensitivity, specificity, etc. The performance is assessed through the non-clinical tests listed above, demonstrating it functions as intended and is safe.
  2. Sample size used for the test set and the data provenance: Not applicable. There isn't a "test set" in the context of an AI/ML algorithm. The "testing" refers to lab-based biomechanical, biocompatibility, and sterility tests of the physical implant.
  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth as understood for AI/ML diagnostic interpretation (e.g., radiologist reads) is not relevant to this device.
  4. Adjudication method for the test set: Not applicable.
  5. If a multi reader multi case (MRMC) comparative effectiveness study was done: Not applicable. This type of study is for evaluating human performance with and without AI assistance for tasks like image interpretation, which is not the function of this implant.
  6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done: Not applicable. This is a physical implant, not an algorithm.
  7. The type of ground truth used: For the physical device, "ground truth" relates to standardized testing conditions and material properties (e.g., biomechanical strength, non-cytotoxicity), rather than a clinical diagnosis adjudication.
  8. The sample size for the training set: Not applicable. There is no training set for a physical implant.
  9. How the ground truth for the training set was established: Not applicable.

In summary, the provided document details a 510(k) submission for a vertebral implant, focusing on its substantial equivalence to a predicate device through non-clinical performance testing and a literature-based clinical evaluation, as opposed to the clinical trial and AI/ML specific performance metrics you've inquired about.

{0}------------------------------------------------

March 5, 2020

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services-USA. To the right of this symbol, there is a blue square with the letters "FDA" in white. Next to the blue square, the words "U.S. FOOD & DRUG ADMINISTRATION" are written in blue.

Hyprevention SAS Cécile Vienney President/CEO PTIB - Hopital Xavier Arnozan Avenue du Haut-Leveque PESSAC Cedex, 33604 Fr

Re: K191709

Trade/Device Name: V-STRUT© Vertebral Implant Regulation Number: 21 CFR 888.3027 Regulation Name: Polymethylmethacrylate (PMMA) Bone Cement Regulatory Class: Class II Product Code: NDN, LOD Dated: June 18, 2019 Received: June 26, 2019

Dear Ms. Vienney:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for

Laura Rose, Ph.D. Acting Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K191709

Device Name V-STRUT© Vertebral Implant

Indications for Use (Describe)

V-STRUT© Vertebral Implant is indicated for use in the treatment of vertebral fractures in the thoracic and lumbar spine from T9 to L5. It is intended to be used in combination with Teknimed F20® bone cement.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K191709 Page 1 of 4

hyprevention®

510(k) Summary

I. SUBMITTER

Hyprevention SAS

PTIB - Hôpital Xavier Arnozan Avenue du Haut Lévêque 33604 PESSAC CEDEX- France Phone : +33(0)5 57 10 28 52 Fax : +33 (0)5 67 07 10 26

Contact Person : Cécile VIENNEY Email: c.vienney@hyprevention.com

Date Prepared: March 05, 2020

II. DEVICE

Name of Device: V-STRUT© Vertebral Implant Common / Usual Name: V-STRUT© Requlation Number: 21 CFR 888.3027 Regulation Name: Polymethylmethacrylate (PMMA) bone cement Regulatory Class: II Product Code: NDN LOD

III. PREDICATE DEVICE

Primary predicate: KIVA® VCF Treatment System (K132817)

Reference Device/Clinical comparators: F20® (K103433), CORTOSS® Bone Augmentation Material (K080108).

The predicate device has not been subject to any design related recall.

IV. DEVICE DESCRIPTION

V-STRUT© Vertebral Implant is part of V-STRUT© Transpedicular Vertebral System. V-STRUT© Transpedicular Vertebral System is composed of :

  • V-STRUT© Vertebral Implant, ।
  • -V-STRUT© Guide Wire,
  • -V-STRUT© Instrumentation Kit.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the word "hyprevention" in a stylized font. The "y" in "hy" is orange, while the rest of the letters are black. There is a registered trademark symbol after the word.

V-STRUT© Vertebral Implant is a medical device to be placed in the vertebrae through a minimally invasive procedure. Two devices are implanted in each vertebra to be treated. Each implant is introduced posteriorly through the pedicle up to the anterior vertebral body wall.

The implant is made of radio transparent polymer, PEEK (Polyetheretherketone as per ASTM F2026) and includes two visualizing markers made of tantalum (as per ASTM F560).

V-STRUT© Vertebral Implant exists in 2 different diameters and 5 different lengths to accommodate individual patient's anatomy of thoracic and/or lumbar vertebrae from T9 to L5.

V-STRUT© Vertebral Implant is provided sterile and is not reusable.

V-STRUT© Vertebral Implanted using specific instrumentation provided with the implant and is combined in situ with a Polymethylmethacrylate (PMMA) bone cement, F20® manufactured by Teknimed SA (K103433).

V. INDICATIONS FOR USE

V-STRUT© Vertebral Implant is indicated for use in the treatment of vertebral fractures in the thoracic and lumbar spine from T9 to L5. It is intended to be used in combination with Teknimed F20® bone cement.

VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

The following table identifies technological characteristics shared between the primary predicate and the subject device.

Element of ComparisonV-STRUT©Transpedicular Vertebral SystemKIVA®VCF Treatment System
ClassificationClass II(V-STRUT© Vertebral Implant)Class II
Regulation21 CFR 888.302721 CFR 888.3027
Product CodeNDN, LODNDN, LOD
IndicationsFor use in the treatment of vertebral fractures in the thoracic and lumbar spine from T9 to L5. It is intended to be used in combination with Teknimed F20® bone cement.For use in the reduction and treatment of spinal fractures in the thoracic and/or lumbar spine from T6-L5. It is intended to be used in combination with Benvenue Vertebral Augmentation Cement Kit
1. System componentsV-STRUT© TranspedicularVertebral System:- V-STRUT© Vertebral Implant- V-STRUT© Guide Wire- V-STRUT© Instrumentation KitKIVA® VCF Treatment System:- KIVA® Implant- KIVA® Deployment System (coil)- Set of Access Instruments
Table 1 Comparison between V-STRUT© and its Predicate Device
----------------------------------------------------------------

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the word "hyprevention" in a stylized font. The "hy" portion of the word is connected by a curved line, and the "y" is colored in orange. A circled "R" symbol is located at the upper right of the word.

Element of ComparisonV-STRUT©Transpedicular Vertebral SystemKIVA®VCF Treatment System
2.Implant materialsPEEK-OPTIMA™ Polymer LT1(ASTM F2026).Provider INVIBIO.Tantalum markers.PEEK-OPTIMA™ Polymer LT1(ASTM F2026) with 15% bariumsulphate.Provider INVIBIO.Tantalum markers.
3.Implants sizesDiameter: 5.5 to 6.5 mmLength: 40 to 60 mmHeight: NAPedicle access diameter: 5 mmStacked spiral diameter: 20 mmHeight: up to 15 mm
4.Presentation of the deviceSterile, single use device(Implant only)Sterile, single use device
5.CombinationIt is intended to be used incombination with Teknimed F20®PMMA Bone cement (K103433).It is intended to be used incombination with BENVENUEMedical Vertebral AugmentationCement Kit.
6.Number of implants2 permanent devices per vertebrae1 permanent device per vertebrae
7.Fixation mechanismIn the Vertebral Body.Bone cement injection.Bilateral fracture support.Pedicle anchorage.In the Vertebral Body.Bone cement injection.Bilateral fracture support.

VII. PERFORMANCE DATA

The following performance testing was conducted to support substantial equivalence determination.

Biomechanical Performance Testing

The safety and performance of V-STRUT© Transpedicular Vertebral System have been substantiated through non-clinical mechanical (static and dynamic) for V-STRUT© Vertebral Implant and functional testing for the whole system. Results of the testing confirm that V-STRUT© Transpedicular Vertebral System can reliably perform as intended.

Biocompatibility

Biocompatibility testing was conducted and documented as recommended by FDA's Guidance for Industry and FDA Staff, "Use of International Standard ISO 10993-1, 'Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process."

V-STRUT© Vertebral Implant was tested to be non-pyrogenic according to USP 42 - NF 37 <151> Pyrogen (Test Rabbit) and BET testing.

V-STRUT© Vertebral Implant was tested to be non-cytotoxic according to ISO 10993-1 "Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity".

V-STRUT© Vertebral Implant is demonstrated to be biocompatible for use in human body and to be as safe as the predicate device.

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the word "hyprevention" in a stylized font. The "y" in "hy" is orange, while the rest of the letters are black. There is a registered trademark symbol to the upper right of the word.

Sterility

V-STRUT© Vertebral Implant is provided sterile to the end user. This device is for single use only. Each device is sterilized by gamma radiation in accordance with ISO 11137-1 "Sterilization of health care products - Radiation - Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices " and 11137-2 standards "Sterilization of health care products - Radiation - Part 2: Establishing the sterilization dose" to provide a Sterility Assurance Level (SAL) of 10-6.

Shelf-Life

V-STRUT© Vertebral Implant is released with a maximum shelf-life of 5 years from the date of sterilization. The shelf-life was determined in compliance with ASTM F1980 : "Standard Specification for accelerated aging of sterile barrier Systems for medical devices" and by performing a real-time aqing study.

Clinical Evaluation

Based on the clinical evaluation V-STRUT© Vertebral Implant has demonstrated to be substantially equivalent to its predicate device KIVA® VCF Treatment System. No additional safety or effectiveness issues as the ones expected and documented in the literature for predicates and references devices have been raised.

VIII. CONCLUSION

V-STRUT© Vertebral Implant compared to the predicate device KIVA® VCF Treatment System has demonstrated is safety, effectiveness and performance as intended. Thus, a conclusion of substantial equivalence to the predicate device is supported.

§ 888.3027 Polymethylmethacrylate (PMMA) bone cement.

(a)
Identification. Polymethylmethacrylate (PMMA) bone cement is a device intended to be implanted that is made from methylmethacrylate, polymethylmethacrylate, esters of methacrylic acid, or copolymers containing polymethylmethacrylate and polystyrene. The device is intended for use in arthroplastic procedures of the hip, knee, and other joints for the fixation of polymer or metallic prosthetic implants to living bone.(b)
Classification. Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Polymethylmethacrylate (PMMA) Bone Cement.”