K Number
K191664
Manufacturer
Date Cleared
2019-08-30

(70 days)

Product Code
Regulation Number
870.1250
Panel
NE
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SelectFlex 072 Neurovascular Access System is indicated for the introduction of interventional devices into the peripheral and neurovasculature.

Device Description

The SelectFlex 072 Neurovascular Access System is a sterile, single-use intravascular catheter used to facilitate access to target vasculature during interventional procedures. The system is composed of a the SelectFlex 072 Catheter, a 3cc Inflation Syringe, a 7Fr Peel Away Introducer, and Hub Extension Line. The 072 SelectFlex Catheter has a usable length of 105cm. The SelectFlex 072 Catheter has variable stiffness along its length and has a dual mode stiffness mechanism on the distal portion of the catheter that is activated by the user allowing the device to transition between tracking and support modes. The distal end of the SelectFlex 072 Catheter has a hydrophilic coating.

AI/ML Overview

The provided document is a 510(k) summary for the SelectFlex 072 Neurovascular Access System, asserting its substantial equivalence to a predicate device (SelectFlex 072 Neurovascular Access System K181000). The summary focuses on comparing the new device to the predicate and presenting nonclinical bench test data. It does not detail a study proving the device meets acceptance criteria for an AI/ML-based medical device. The documentation pertains to a traditional medical device (a catheter system), not an AI-enabled device.

Therefore, the following information cannot be extracted from the provided text for an AI/ML-based device:

  • A table of acceptance criteria and the reported device performance: The document lists bench tests and states "PASS" for all, meaning all samples met "pre-determined acceptance criteria," but it does not specify the numerical or qualitative acceptance criteria for each test in a detailed table that would be typical for an AI/ML performance summary (e.g., sensitivity, specificity, AUC thresholds). The acceptance criteria for the new device are implicitly that they perform equivalently to the predicate device in the listed bench tests.
  • Sample size used for the test set and the data provenance: No information on test set sample size or data provenance in the context of an AI/ML study is provided. The bench tests would have their own sample sizes, but these are not for an AI model.
  • Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable as this is not an AI/ML study requiring expert-derived ground truth.
  • Adjudication method for the test set: Not applicable.
  • If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable.
  • If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
  • The type of ground truth used: Not applicable.
  • The sample size for the training set: Not applicable.
  • How the ground truth for the training set was established: Not applicable.

However, based on the document, I can provide information relevant to a traditional medical device's nonclinical performance evaluation:

1. Table of Acceptance Criteria and Reported Device Performance (Nonclinical Bench Testing):

The document states that all tests resulted in "PASS," meaning all samples met "pre-determined acceptance criteria." The specific numerical acceptance criteria themselves are not listed in this summary, only the outcome.

Test DescriptionTest MethodReported Device Performance (Result against unspecified acceptance criteria)
Packaging IntegrityTested per ISO 11607-1 and -2PASS (All samples met the pre-determined acceptance criteria)
Visual Surface RequirementsVisual inspection of catheter surfacesPASS (All samples met the pre-determined acceptance criteria)
Dimensional VerificationDevice dimensions were measured to confirm conformance to the product specificationPASS (All samples met the pre-determined acceptance criteria)
Liquid Leakage Under PressureTested per ISO 10555-1:2013 Annex CPASS (All samples met the pre-determined acceptance criteria)
Hub Aspiration Air LeakageTested per ISO 10555-1 2013 for Hub Aspiration Air LeakagePASS (All samples met the pre-determined acceptance criteria)
Simulated Use/UsabilityDevice preparation, delivery, access, was evaluated in a challenging neurovascular model. Simulated use testing includes a usability assessment with multiple physicians.PASS (All samples met the pre-determined acceptance criteria)
Flex FatigueTested per ISO 10555-1: 2013 for Flexural FatiguePASS (All samples met the pre-determined acceptance criteria)
Tip DeflectionThe amount of tip deflection under load was evaluated and compared to the predicate devicePASS (All samples met the pre-determined acceptance criteria)
Inflation FatigueTested per ISO 10555-1: 2013 for Inflation Fatigue - 20 inflation cyclesPASS (All samples met the pre-determined acceptance criteria)
Burst VolumeTested per ISO 10555-1: 2013 for Inflation Fatigue - tested to 2x the inflation volumePASS (All samples met the pre-determined acceptance criteria)
Torque TestTested per ISO 10555-1: 2013 for Torque TestingPASS (All samples met the pre-determined acceptance criteria)
Flow RateTested per ISO 10555-1: 2013 for Flow rate compared to the predicate devicePASS (All samples met the pre-determined acceptance criteria)
ISO 80369-7: Small Bore ConnectorsTested per ISO 80369-7 (dimensional, leakage, resistance to separation, etc.)PASS (All samples met the pre-determined acceptance criteria)
Corrosion ResistanceTested per ISO 10555-1 Annex A for corrosion resistancePASS (All samples met the pre-determined acceptance criteria)
Particulate CountEffluent tested per AAMI TIR42, USP 788 using multiple insertion and withdrawal cyclesPASS (All samples met the pre-determined acceptance criteria)
Coating Integrity, Lubricity, DurabilityTested in consideration of FDA CTQ for Hydrophilic Coated vascular cathetersPASS (All samples met the pre-determined acceptance criteria)
Peak Tensile TestingTested per ISO 10555-1 for tensile strength including all bonds/jointsPASS (All samples met the pre-determined acceptance criteria)
Device RemovalRemoval force in both modes was compared to a reference devicePASS (All samples met the pre-determined acceptance criteria)
RadiopacityRadiopacity was evaluated during simulated use testing confirming visualization under fluoroscopyPASS (All samples met the pre-determined acceptance criteria)
Shelf LifeAccelerated aging studies were completed to confirm the stability of the product and package. All testing noted above was performed on aged product to support the shelf life claimPASS (All samples met the pre-determined acceptance criteria)
CytotoxicityTested in accordance with ISO 10993-5, Neutral Red Uptake MethodPass (Noncytotoxic according to the predetermined acceptance criteria)
Intracutaneous IrritationTested in accordance with ISO 10993-10Pass (Test requirements for intracutaneous reactivity were met)
SensitizationTested in accordance with ISO 10993-10, Kligman Maximization TestPass (did not elicit a sensitization response)
Systemic ToxicityTested in accordance with ISO 10993-11Pass (Test requirements for systemic toxicity were met)
Material Mediated PyrogenicityTested in accordance with ISO 10993-11 and USP 40 Pyrogen TestPass (Nonpyrogenic, met the predetermined acceptance criteria)
HemolysisTested in accordance with ASTM F756-17 and ISO 10993-4 (direct and indirect methods)Pass (Non-hemolytic, met the predetermined acceptance criteria)
In Vitro HemocompatibilityTested in accordance with ISO 10993-4 (Direct Contact Method)Pass (Not expected to result in adverse effects in vivo)
Complement ActivationTested in accordance with ISO 10993-4 (SC5b-9 Complement Activation)Pass (Does not activate the complement system)
Un-activated Partial Thromboplastin TimeTested in accordance with ISO 10994-4 and ASTM F2382-04Pass (Does not have an effect on coagulation of human plasma)
ThrombogenicityTested in accordance with ISO 10994-4Pass (Demonstrates similar thromboresistance characteristics as control)
Bacterial EndotoxinTested in accordance with USP 40, NF 35, 2017. Bacterial Endotoxins TestPASS (All samples met the pre-determined acceptance criteria)

2. Sample sized used for the test set and the data provenance:

The document mentions "All samples met the pre-determined acceptance criteria" for various bench tests, but does not specify the sample size (n) for each test set used in the nonclinical bench testing. Data provenance is not applicable in the context of clinical data for an AI/ML device, as these are bench tests.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

For the "Simulated Use/Usability" test, the document states it "includes a usability assessment with multiple physicians." The specific number and qualifications of these physicians are not provided. This is not establishing "ground truth" in an AI/ML context, but rather assessing usability.

4. Adjudication method for the test set:

Not applicable in the context of bench testing for a traditional medical device.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done:

No, this type of study was not conducted as this is a non-AI/ML device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

Not applicable, as this is not an AI/ML device.

7. The type of ground truth used:

Not applicable, as this is not an AI/ML device. For the bench tests, the "ground truth" is defined by established international and national standards (e.g., ISO, ASTM, USP) and the device's own specifications.

8. The sample size for the training set:

Not applicable, as this is not an AI/ML device that requires a training set.

9. How the ground truth for the training set was established:

Not applicable, as this is not an AI/ML device.

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).