K Number
K191533
Date Cleared
2019-08-28

(79 days)

Product Code
Regulation Number
862.1680
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

ADVIA Centaur® Testosterone II (TSTII)

The ADVIA Centaur® Testosterone II (TSTI) assav is for in the quantitative determination of total testosterone (bound and unbound) in human serum and plasma using the ADVIA Centaur XP system.

Measurements of testosterone are used in the diagnosis and treatment of disorders involving the male sex hormones (androgens), including primary and secondary hypogonadism, delayed or precocious puberty, impotence in males and, in females, hirsutism (excessive hair) and virilization) due to tumors, polycystic ovaries, and adrenogenital syndromes.

ADVIA Centaur® SHBG

The ADVIA Centaur® SHBG assay is an in vitro diagnostic immunoassay for the quantitative determination of sex hormone-binding globulin (SHBG) in human serum and plasma using the ADVIA Centaur XP system.

The ADVIA Centaur SHBG assay is intended for use as an aid in the diagnosis of androgen disorders.

Device Description

The re-standardized ADVIA Centaur® Testosterone II (TSTII) and the ADVIA Centaur® SHBG are in vitro diagnostic immunoassays for use on the ADVIA Centaur® XP system. The ADVIA Centaur® Testosterone II (TSTII) assay quantitatively determines total testosterone in human serum and plasma. The ADVIA Centaur® SHBG assay quantitatively determines sex hormone-binding globulin (SHBG) in human serum and plasma. The submission includes details on the ingredients of the reagents for both assays.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the ADVIA Centaur® Testosterone II (TSTII) and ADVIA Centaur® SHBG based on the provided text, formatted as requested:

1. Table of Acceptance Criteria and Reported Device Performance:

Performance CharacteristicAcceptance Criteria (ADVIA Centaur® Testosterone II (TSTII))Reported Device Performance (ADVIA Centaur® Testosterone II (TSTII))Acceptance Criteria (ADVIA Centaur® SHBG)Reported Device Performance (ADVIA Centaur® SHBG)
Detection Limits
Limit of Blank (LoB)(Implicitly good performance relative to predicate)2.50 ng/dL(Not explicitly provided in this document)(Previously established, see K151986 & K091867)
Limit of Detection (LoD)(Implicitly good performance relative to predicate)5.00 ng/dL (95% probability)(Not explicitly provided in this document)(Previously established, see K151986 & K091867)
Limit of Quantitation (LoQ)(Total CV of 20%)7.00 ng/dL (at 20% CV)(Not explicitly provided in this document)(Previously established, see K151986 & K091867)
Precision(Implicitly good performance relative to predicate)See table below for detailed CVs across various samples and concentrations. Generally, Within-Run CVs are lower than Within-Lab CVs.(Not explicitly provided in this document)(Previously established, see K151986 & K091867)
LinearityBias from linear fit estimate < 10%Bias from the linear fit estimate was <10% for all 9 samples.(Not explicitly provided in this document)(Previously established, see K151986 & K091867)
Method ComparisonAcceptable correlation with predicate device and reference methodVs ID-LC-MS/MS: y = 0.97(x) - 0.22 ng/dL (r = 0.98) Vs Predicate ADVIA Centaur TSTII: y = 1.04(x) - 4.14 ng/dL (r = 1.00) Vs Dimension Vista LOCI Total Testosterone (Adult & Pediatric): y = 1.01(x) - 3.32 ng/dL (r = 0.99) Vs Dimension Vista LOCI Total Testosterone (Pediatric only): y = 1.01(x) - 4.74 ng/dL (r = 1.00)(Implicitly good agreement with predicate with new reference intervals not affecting analytical performance)(Previously established, see K151986 & K091867; this submission focused solely on new reference intervals)
Reference IntervalsEstablished in accordance with CLSI guideline EP28-A3cNew reference intervals established and provided in detail for adult and pediatric populations, stratified by age and Tanner stage.Established in accordance with CLSI guideline EP28-A3cNew reference intervals established for adult male and female populations.
Biotin InterferenceNo significant effect on the assay at 30 ng/mL% Bias ranged from 2% to 6% for 30 ng/mL biotin at various analyte concentrations. Higher biotin levels (e.g., 250, 500 ng/mL) showed significant bias.Less than or equal to 10% change in results at 300 ng/mL% Bias ranged from -3% to -9% at 300 ng/mL biotin for various analyte concentrations. Higher biotin levels (e.g., 600, 1200 ng/mL) showed greater bias.

ADVIA Centaur® Testosterone II (TSTII) Precision Results:

SamplenMean (ng/dL)SD (Repeatability)CV (Repeatability)SD (Within-Lab)CV (Within-Lab)
Control 180257.389.063.514.385.6
Control 280636.5745.147.153.128.3
Control 3801021.9362.286.181.027.9
MDP18020.921.36.21.868.9
MDP28073.573.544.85.247.1
MDP380312.8113.44.323.997.7
MDP480776.6442.595.558.267.5
MDP5801123.8359.225.3109.579.7
Female Patient Serum Pool8015.861.8411.62.3915.1

2. Sample sizes used for the test set and the data provenance:

  • ADVIA Centaur® Testosterone II (TSTII):

    • Precision: 80 replicates per sample, totaling 9 samples (3 controls, 5 medical decision pools, 1 female patient serum pool). Data provenance not explicitly stated, but typically internal lab data.
    • Linearity: 9 serially diluted samples. Data provenance not explicitly stated.
    • Method Comparison:
      • Vs ID-LC-MS/MS: 108 adult male and female serum samples (7.27-1394.00 ng/dL). Data provenance not explicitly stated, but implied to be patient samples.
      • Vs Predicate ADVIA Centaur TSTII: 108 individual male and female adult serum samples (10.93-1439.77 ng/dL). Data provenance not explicitly stated.
      • Vs Dimension Vista LOCI Total Testosterone: 124 serum samples (79 adult, 45 pediatric). Data provenance not explicitly stated.
        • Pediatric: 45 samples (31 females, ages 7-18; 14 males, ages 22 months-18 years).
    • Reference Intervals: Samples from apparently healthy, normal adult females and males, and pediatric populations. Collected prospectively.
      • Adult Male (Under 50): 250 samples
      • Adult Male (50 and Over): 135 samples
      • Adult Female (Under 50): 224 samples
      • Adult Female (50 and Over): 151 samples
      • Pediatric Male (by Tanner Stage): 101-129 samples per stage
      • Pediatric Male (by Age): 27-149 samples per age group
      • Pediatric Female (by Tanner Stage): 49-138 samples per stage
      • Pediatric Female (by Age): 145-174 samples per age group
    • Biotin Interference: Not explicitly stated, but typically in-house spiked samples.
  • ADVIA Centaur® SHBG:

    • Reference Intervals: Samples from apparently healthy, normal adult females and males. (Presumably collected prospectively, similar to TSTII).
      • Adult Male (Under 50): 250 samples
      • Adult Male (50 and Over): 135 samples
      • Adult Female (Under 50): 224 samples
      • Adult Female (50 and Over): 151 samples
    • Biotin Interference: Not explicitly stated, but typically in-house spiked samples.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • For both devices: The ground truth for the ADVIA Centaur® Testosterone II (TSTII) method comparison was established using ID-LC-MS/MS (Isotope Dilution Liquid Chromatography-Mass Spectrometry/Mass Spectrometry), with samples assigned concentrations by the CDC HoSt RMP ID-LC/MS/MS. This is an analytical reference method, not expert human readers.
  • For the ADVIA Centaur® SHBG, the ground truth for traceability/standardization is stated as WHO 2nd International Standard (08/226), an internationally recognized analytical standard.

No human experts were used to establish the ground truth for the test set in the traditional sense of image or clinical interpretation. The methods employed are laboratory analytical techniques. When discussing reference intervals for pediatric subjects, it mentions that "Samples were also clinically characterized according to Tanner Stage," which would implicitly involve clinical experts (pediatricians or endocrinologists) for classification, but the number and qualifications are not specified in this document.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

  • None. The ground truth was established by objective analytical reference methods (ID-LC-MS/MS) and international standards (WHO).

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • Not applicable. This document describes laboratory diagnostic devices (immunoassays) for quantitative determination of biomarkers, not AI-assisted image analysis or clinical decision support tools that involve human readers/interpreters in a comparative effectiveness study.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

  • Yes, this describes standalone performance. The ADVIA Centaur® Testosterone II (TSTII) and ADVIA Centaur® SHBG are automated in-vitro diagnostic assays. The performance metrics (precision, linearity, method comparison, detection limits, etc.) represent the performance of the algorithm/instrument itself without human intervention in the measurement process, beyond sample loading and general operation according to the Instructions for Use.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

  • Analytical Reference Methods and International Standards:
    • For the ADVIA Centaur® Testosterone II (TSTII), the primary ground truth for method comparison was ID-LC-MS/MS (specifically, CDC HoSt RMP ID-LC/MS/MS).
    • For the ADVIA Centaur® SHBG, the primary ground truth for traceability/standardization is the WHO 2nd International Standard (08/226).
    • For reference intervals, apparently healthy individuals were identified, and for pediatric subjects, Tanner Stage classification was used, which is a clinical assessment.

8. The sample size for the training set:

  • Not explicitly stated in this document. The document discusses the performance of a re-standardized assay (ADVIA Centaur® Testosterone II (TSTII)) and new reference intervals for another (ADVIA Centaur® SHBG). It does not detail the development or "training" process of the original assays or any underlying algorithms, which would typically involve a separate, often larger, dataset. The studies described are primarily for validation of the modified device and new reference intervals.

9. How the ground truth for the training set was established:

  • Not explicitly stated in this document. As mentioned above, this document focuses on validation studies of a modified/re-standardized device and new reference intervals for an existing device, rather than the initial development and "training" of the assays. The original development would have involved similar analytical ground truth methods, but the specifics are not provided here.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains two logos. The logo on the left is the Department of Health & Human Services - USA logo. The logo on the right is the FDA U.S. Food & Drug Administration logo. The FDA logo is in blue.

September 20, 2019

Siemens Healthcare Diagnostics Inc. Paul DaSilva Regulatory Affairs Specialist 511 Benedict Avenue Tarrytown, NY 10591

Re: K191533

Trade/Device Name: ADVIA Centaur® Testosterone II (TSTII), ADVIA Centaur® SHBG Regulation Number: 21 CFR 862.1680 Regulation Name: Testosterone test system Regulatory Class: Class I, reserved Product Code: CDZ Dated: June 7, 2019 Received: June 10, 2019

Dear Paul DaSilva:

This letter corrects our substantially equivalent letter of August 28, 2019.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Kellie B. Kelm -S

Kellie B. Kelm, Ph.D. Acting Director Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K191533

Device Name ADVIA Centaur® Testosterone II (TSTII) ADVIA Centaur® SHBG

Indications for Use (Describe)

ADVIA Centaur® Testosterone II (TSTII)

The ADVIA Centaur® Testosterone II (TSTI) assav is for in the quantitative determination of total testosterone (bound and unbound) in human serum and plasma using the ADVIA Centaur XP system.

Measurements of testosterone are used in the diagnosis and treatment of disorders involving the male sex hormones (androgens), including primary and secondary hypogonadism, delayed or precocious puberty, impotence in males and, in females, hirsutism (excessive hair) and virilization) due to tumors, polycystic ovaries, and adrenogenital syndromes.

ADVIA Centaur® SHBG

The ADVIA Centaur® SHBG assay is an in vitro diagnostic immunoassay for the quantitative determination of sex hormone-binding globulin (SHBG) in human serum and plasma using the ADVIA Centaur XP system.

The ADVIA Centaur SHBG assay is intended for use as an aid in the diagnosis of androgen disorders.

Type of Use (Select one or both, as applicable)
[X] Prescription Use (Part 21 CFR 801 Subpart D) [ ] Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

This 510(k) Summary of Safety and Effectiveness is being submitted in accordance with the requirements of 21 CFR 807.92 and the Safe Medical Device Act of 1990.

The assigned 510(k) Number is: K191533

Date Prepared 1.

June 7, 2019

Applicant Information 2.

Contact:Paul DaSilvaRegulatory Affairs Specialist
Address:Siemens Healthcare Diagnostics Inc.511 Benedict AvenueTarrytown, NY 10591-5097
Phone:914-524-3141
Fax:914-524-2101
Email:paul.dasilva@siemens-healthineers.com

Regulatory Information 3.

Table 1. Regulatory Information for ADVIA Centaur® Testosterone II (TSTII)

Trade NameADVIA Centaur® Testosterone II (TSTII)
Common NameRadioimmunoassay, testosterones and dihydrotestosterone
Classification NameTestosterone test system
FDA ClassificationClass I (Reserved)
Review PanelClinical Chemistry
Product CodeCDZ
Regulation Number862.1680

Table 2. Regulatory Information for ADVIA Centaur® SHBG

Trade NameADVIA Centaur® SHBG
Common NameRadioimmunoassay, testosterones and dihydrotestosterone
Classification NameTestosterone test system
FDA ClassificationClass I (Reserved)
Review PanelClinical Chemistry (75)
Product CodeCDZ
Regulation Number862.1680

{4}------------------------------------------------

Predicate Device Information 4.

ADVIA Centaur® Testosterone II (TSTII)

Predicate Device Name: ADVIA Centaur® Testosterone II (TSTII)

510(k) Number: K151986

The re-standardized ADVIA Centaur® Testosterone II (TSTII) is substantially equivalent to the ADVIA Centaur® Testosterone II (TSTII) that was cleared under 510(k) (K151986), as shown below in the Substantial Equivalence Information section.

ADVIA Centaur® SHBG

Predicate Device Name: ADVIA Centaur® SHBG

510(k) Number: K151986

The ADVIA Centaur® SHBG with new reference intervals in the Instructions for Use (Package Insert) does not require any other device modifications (i.e. no change to design or manufacturing process). Therefore, as shown below in the Substantial Equivalence Information section, the predicate and subject devices are the same.

5. Intended Use / Indications for Use

ADVIA Centaur® Testosterone II (TSTII)

The ADVIA Centaur® Testosterone II (TSTII) assay is for in vitro diagnostic use in the quantitative determination of total testosterone (bound and unbound) in human serum and plasma using the ADVIA Centaur® XP system.

Measurements of testosterone are used in the diagnosis and treatment of disorders involving the male sex hormones (androgens), including primary and secondary hypogonadism, delayed or precocious puberty, impotence in males and, in females, hirsutism (excessive hair) and virilization (masculinization) due to tumors, polycystic ovaries. and adrenogenital syndromes.

ADVIA Centaur® SHBG

The ADVIA Centaur® SHBG assay is an in vitro diagnostic immunoassay for the quantitative determination of sex hormone-binding globulin (SHBG) in human serum and plasma using the ADVIA Centaur® XP system.

The ADVIA Centaur® SHBG assay is intended for use as an aid in the diagnosis of androgen disorders.

Device Description 6.

Table 3. Summary of Ingredients of the Re-standardized ADVIA Centaur® Testosterone II (TSTII) Components

ComponentVolumeIngredients
ADVIA Centaur ® Testosterone II (TSTII) Primary Reagent ReadyPack (included in assay kit)
ADVIA Centaur ®Testosterone II (TSTII)Lite Reagent10.0 mL/packAcridinium ester-labeled hapten (36 µg/mL) inbuffered saline with preservatives

{5}------------------------------------------------

Table 3. Summary of Ingredients of the Re-standardized ADVIA Centaur® Testosterone
II (TSTII) Components
ComponentVolumeIngredients
ADVIA Centaur®Testosterone II (TSTII)Solid Phase Reagent17.0 mL/packStreptavidin-coated latex particles (0.33 g/L) inbuffered saline with preservatives
ADVIA Centaur® Testosterone II (TSTII) Ancillary Reagent ReadyPack (included in assaykit)
ADVIA Centaur®Testosterone II (TSTII)Releasing Agent10.0 mL/packSteroid releasing agent (0.4 $ \mu $ g/mL) and biotinylatedsheep monoclonal anti-testosterone antibody (27 $ \mu $ g/L) in buffered saline and preservatives
Table 4. Summary of Ingredients of the ADVIA Centaur® SHBG Components
-------------------------------------------------------------------------------------
ComponentVolumeIngredients
ADVIA Centaur® SHBG Primary Reagent ReadyPack (included in assay kit)
ADVIA Centaur® SHBGLite Reagent3.0 mL/packMouse monoclonal anti-SHBG antibody (~130 ng/mL) labeled with acridinium ester in HEPES-buffered saline with bovine serum albumin, mouse serum, sodium azide (< 0.1%), surfactant, protein stabilizers and preservatives
ADVIA Centaur® SHBGSolid Phase Reagent11.0 mL/packStreptavidin coupled to latex paramagnetic particles (~150 µg/mL) in HEPES buffered saline with bovine serum albumin, mouse serum, sodium azide (< 0.1%), surfactant, protein stabilizers and preservatives
ADVIA Centaur® SHBGAncillary Well Reagent3.0 mL/packBiotinylated monoclonal mouse anti-SHBG antibody (~6 µg/mL) in HEPES buffered saline with bovine serum albumin, mouse serum, sodium azide (< 0.1%), surfactant, protein stabilizers and preservatives

Purpose of the Submission 7.

The purpose of this submission is to submit a re-standardized ADVIA Centaur® Testosterone II (TSTII) and to submit the new reference intervals data for the ADVIA Centaur® SHBG that was cleared under K151986. For the ADVIA Centaur® SHBG the only change is the new reference intervals.

Substantial Equivalence Information - Comparison of 8. Predicate Device and Modified Device

Both the re-standardized ADVIA Centaur® Testosterone II (TSTII) (Modified Device) and the current ADVIA Centaur® Testosterone II (TSTII) (Predicate Device cleared under K151986) employ the same prepackaged reagents for use on automated test systems. The Intended Use / Indications for Use, assay principle, and reagent formulations are the same. The major differences between the Modified and Predicate devices are the re-standardization and change in reference interval data. However, despite these differences, the performance and accuracy of the Modified Device are substantially equivalent to that of the Predicate

{6}------------------------------------------------

Device. The method comparison study between the Modified Device and the Predicate Device demonstrate acceptable correlation between these assays. A comparison of these assays is shown in the following table:

ltemRe-standardized ADVIA Centaur®Testosterone II (TSTII) (CandidateDevice)ADVIA Centaur® Testosterone II(TSTII) (Predicate Device -K151986)
Intended UseFor in vitro diagnostic use in thequantitative determination of totaltestosterone (bound and unbound)in serum using the ADVIA Centaurand ADVIA Centaur XP Systems.Same
Indications for UseMeasurements of testosterone areused in the diagnosis and treatmentof disorders involving the male sexhormones (androgens), includingprimary and secondaryhypogonadism, delayed orprecocious puberty, impotence inmales and, in females hirsutism(excessive hair) and virilization(masculinization) due to tumors,polycystic ovaries, andadrenogenital syndromes.Same
MethodologyChemiluminescenceSame
Assay ProtocolCompetitive immunoassaySame
Traceability/StandardizationID-LC-MS/MSSame
Specimen TypeHuman serum and plasmaSame
Lower Limit ofMeasuring RangeLoQSame
Sample Volume20 µLSame
Measuring Range7.0-1500 ng/dLSame
Calibration2-point calibrationSame
Table 5. Comparison of ADVIA Centaur® Testosterone II (TSTII) to Predicate
------------------------------------------------------------------------------

The following table demonstrates substantial equivalence between the ADVIA Centaur® SHBG (Candidate Device) with new reference intervals in the Instructions for Use (Package Insert) and the currently marketed ADVIA Centaur® SHBG (Predicate Device) that was cleared under 510(k) K151986.

{7}------------------------------------------------

ltemADVIA Centaur® SHBG: WHO08/226 2nd IS(Candidate Device - NewReference Intervals)ADVIA Centaur® SHBG: WHO08/226 2nd IS(Predicate Device - UnmodifiedLabeling)
Intended UseThe ADVIA Centaur® SHBG assayis an in vitro diagnosticimmunoassay for the quantitativedetermination of sex hormone-binding globulin (SHBG) in humanserum and plasma using the ADVIACentaur® XP system.Same
Indications for UseMeasurement of SHBG are used inthe diagnosis and treatment ofdisorders involving the male sexhormones (androgens), includingprimary and secondaryhypogonadism, delayed orprecocious puberty, impotence inmales and, in females, hirsutism(excessive hair) and virilization(masculinization) due to tumors,polycystic ovaries, andadrenogenital syndromes.Same
MethodologyChemiluminescenceSame
Assay ProtocolSandwich immunoassaySame
Traceability/StandardizationWHO 2nd International Standard(08/226)Same
Specimen TypeHuman serum and plasmaSame
Sample Volume20 µLSame
Measuring Range1.60-180 nmol/LSame
Calibration2-point calibrationSame

Table 6. Comparison of ADVIA Centaur® SHBG to Predicate

Standard/Guidance Document References 9.

The following recognized standards from Clinical Laboratory Standards Institute (CLSI) were used as a basis of the study procedures described in this submission:

  • Evaluation of Precision Performance of Quantitative Measurement Procedures; ■ Approved Guideline - Third Edition (CLSI EP05-A3, 2004; Recognition Number 7-251)
  • . Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach; Approved Guideline - 1st Edition (CLSI EP06-A, 2003; Recognition Number 7-193)
  • . Measurement Procedure Comparison And Bias Estimation Using Patient Samples -Third Edition (CLSI EP9-A3, 2013; Recognition Number 7-245)
  • Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures: ■ Approved Guideline -- Second Edition (CLSI EP17-A2, 2013; Recognition Number 7-233)

{8}------------------------------------------------

  • I Defining, Establishing and Verifying Reference Intervals in the Clinical Laboratory; Approved Guideline - Third Edition (CLSI EP28-A3c - formerly C28-A3c, 2010; Recognition Number 7-224)

10. Performance Characteristics: Re-standardized ADVIA Centaur® Testosterone II (TSTII)

Substantial equivalence of the re-standardized ADVIA Centaur® Testosterone II (TSTII) demonstrated by testing several performance characteristics with the following studies:

  • . Detection Capability (LoB. LoD. and LoQ)
  • . Linearity
  • Precision
  • Method Comparison
  • Reference Interval ●

In addition, the following study was performed:

  • Biotin Interference ●
    The performance studies listed below were determined not to be impacted by the restandardization of the ADVIA Centaur® Testosterone II (TSTII). These studies are from the original 510(k) K151986 and are used in the labeling for the re-standardized ADVIA Centaur® Testosterone II (TSTII).

  • Dilution Recovery

  • . Cross reactivity

  • . Interfering substances

  • . Matrix Comparison

  • Reagent shelf life stability ●

  • Reagent On-Board Stability and Calibration Interval

  • . Open Vial Stability

Detection Limit 10.1

The limit of blank (LoB), limit of detection (LoD), and the limit of quantitation (LoQ) were determined as described in CLSI protocol EP17-A2. The Re-standardized ADVIA Centaur® Testosterone II (TSTII) has an LoB of 2.50 ng/dL, an LoD of 5.00 ng/dL, and an LoQ of 7.00 ng/dL. The LoB is defined as the highest measurement result that is likely to be observed for a blank sample. The LoD is defined as the lowest concentration of testosterone that can be detected with 95% probability. The LoQ is defined as the lowest concentration of testosterone that can be detected at a total CV of 20%.

10.2 Precision

A 20-day precision study was performed according to CLSI EP5-A3. Samples included three levels of controls, 5 medical decision pools and 1 female patient serum pool. Samples were assayed twice a day in replicates of 2, for 20 days (n = 80 replicates per sample). The following results were obtained:

{9}------------------------------------------------

Repeatability(Within-Run)Within-Lab(Total Precision)
SamplenMean(ng/dL)SDCVSDCV
Control 180257.389.063.514.385.6
Control 280636.5745.147.153.128.3
Control 3801021.9362.286.181.027.9
MDP18020.921.36.21.868.9
MDP28073.573.544.85.247.1
MDP380312.8113.44.323.997.7
MDP480776.6442.595.558.267.5
MDP5801123.8359.225.3109.579.7
Female PatientSerum Pool8015.861.8411.62.3915.1

10.3 Linearity

A linearity study was performed using the modified device according to CLSI EP06-A using 9 serially diluted samples spanning the assay range. The mean was taken from each sample tested in duplicate. As presented below, the bias from the linear fit estimate was <10%.

SampleExpectedDose (ng/dL)MeanObservedDose (ng/dL)WeightedLinear FitEstimateDeviationfromLinear Fit(Percentage)DeviationFromLinear Fit
16.946.946.95-0.15-0.01
211.6511.9511.682.330.27
316.3518.8416.4014.882.44
4110.00102.03110.39-7.57-8.35
5213.06228.35213.826.7914.53
6419.18430.72420.692.3810.03
7831.42819.33834.43-1.81-15.10
81243.661317.801248.165.5869.64
91655.901655.901661.90-0.36-6.00

The weighted linear regression equation is presented below.

Observed = 1.004(Expected) - 0.015 ng/dL

10.4 Method Comparison

A method comparison study was performed by comparing the Re-standardized ADVIA Centaur® Testosterone II (TSTII) to the ID-LC-MS/MS with 108 adult male and female serum samples distributed over the assay range (7.27-1394.00 ng/dL). The samples were assigned testosterone concentrations by the CDC HoSt RMP ID-LC/MS/MS. The study duration was 3 days using day 1 calibration and all samples run in singlicate. The analysis was performed

{10}------------------------------------------------

using Weighted Deming regression. The regression equation from the analysis is presented below.

Re-standardized ADVIA Centaur® Testosterone II (TSTII) (y) = 0.97 (x) (ID-LC-MS/MS)-0.22 ng/dL (r = 0.98)

For the re-standardized ADVIA Centaur® Testosterone II (TSTII) and the current ADVIA Centaur® Testosterone II (TSTII) method comparison study, an analysis was performed on a total of 108 individual male and female adult serum samples spanning the assay range (10.93-14.39.77 ng/dL) The analysis was performed using Weighted Deming regression. The regression equation from the analysis is presented below.

Re-standardized ADVIA Centaur® Testosterone II (TSTII) (y) = 1.04 (x) (ADVIA Centaur TSTII) - 4.14 ng/dL (r = 1.00)

A further method comparison study was performed by comparing the Re-standardized ADVIA Centaur® Testosterone II (TSTII) to the Dimension Vista LOCI Total Testosterone with 124 serum samples (79 samples from adults and 45 samples from pediatric subjects) ranging from 9.00–984.00 ng/dL (0.31–34.12 nmol/L). An analysis was also performed with the 45 samples from pediatric subjects (31 females, ages 7 years to 18 years; and 14 males, ages 22 months to 18 years) ranging from 11.00–762.00 ng/dL. The analyses were performed using Weighted Deming regression. The regression equations from the analyses are:

Adult and pediatric specimens

Re-standardized ADVIA Centaur® Testosterone II (TSTII) (y) = 1.01(x) (Dimension Vista TST) - 3.32 ng/dL (r = 0.99)

Pediatric specimens

Re-standardized ADVIA Centaur® Testosterone II (TSTII) (y) = 1.01(x) (Dimension Vista TST) - 4.74 ng/dL (r = 1.00)

10.5 Reference Intervals

Reference intervals for the ADVIA Centaur® Testosterone II (TSTII), ADVIA Centaur® SHBG and the Free Androgen Index (FAI) were obtained using samples from apparently healthy, normal adult females and males. The testosterone, SHBG, and FAI results provided in the tables below were obtained using results from the ADVIA Centaur® TSTII and ADVIA Centaur® SHBG. Central 90% reference intervals were established in accordance with the CLSI guideline EP28-A3c using a non-parametric approach for sample sizes of at least 120. For populations with a sample size of 40–119, the reference interval was calculated using an approach to accommodate the smaller sample size.

Central 90% reference intervals characterized by age and Tanner stage were established for the ADVIA Centaur® Testosterone II (TSTII) for a pediatric population (children and adolescents) in accordance with the CLSI guideline EP28-A3c using a non-parametric approach for sample sizes of at least 120. For populations with a sample size of < 120, the central 90% reference interval was calculated using an approach to accommodate the smaller sample size. Samples were collected prospectively from apparently healthy pediatric subjects (good

endocrinological health) using predefined inclusion criteria.

Samples were also clinically characterized according to Tanner Stage.

{11}------------------------------------------------

Testosterone

ng/dLnmol/L
Central 90th Ref IntervalCentral 90th Ref Interval
TSTIINMedian5th95thMedian5th95th
AdultUnder 50250409.72197.44669.5814.226.8523.23
Male50 and Over135377.46187.72684.1913.106.5123.74
AdultUnder 5022418.018.3835.010.620.291.21
Female50 and Over15114.18< 7.0035.920.49< 0.241.25
Pediatric MaleTanner Stage 1101< 7.00< 7.0013.06< 0.24< 0.240.45
Tanner Stage 278< 7.00< 7.0079.13< 0.24< 0.242.75
Tanner Stage 36459.67< 7.00499.182.07< 0.2417.32
Tanner Stage 488376.8479.10747.1713.082.7425.93
Tanner Stage 5129451.17224.83669.6515.667.8023.24
Age 2-10147< 7.00< 7.0010.50< 0.24< 0.240.36
Age 113410.73< 7.00478.500.37< 0.2416.60
Age 1235132.47< 7.00487.974.60< 0.2416.93
Age 1334199.028.28549.796.910.2919.08
Age 1434228.398.91535.347.930.3118.58
Age 1527327.8965.96756.5011.382.2926.25
Age 16-21149453.86228.16710.7415.757.9224.66
Pediatric FemaleTanner Stage 1138< 7.00< 7.0010.06< 0.24< 0.240.35
Tanner Stage 2608.17< 7.0030.110.28< 0.241.04
Tanner Stage 34912.98< 7.0030.490.45< 0.241.06
Tanner Stage 49817.37< 7.0035.190.60< 0.241.22
Tanner Stage 513319.7611.8039.300.690.411.36
Age 2-10159< 7.00< 7.0011.86< 0.24< 0.240.41
Age 11-1517412.95< 7.0027.570.45< 0.240.96
Age 16-2114519.8111.7843.340.690.411.50

SHBG

SHBGNµg/mLnmol/L
Central 90th Ref IntervalCentral 90th Ref Interval
Median5th95thMedian5th95th
AdultUnder 502502.261.105.1823.8011.5454.49
Male50 and Over1353.511.656.7936.9117.3371.50
AdultUnder 502244.441.6813.1346.7217.69138.26
Female50 and Over1514.552.2510.5147.8623.65110.61

FAI

FAINPercent (%)
Central 90th Ref Interval
Median5th95th
AdultUnder 5025055.86%26.18%107.07%
Male50 and Over13536.01%17.38%60.86%
AdultUnder 502241.37%0.33%4.37%
Female50 and Over1511.05%0.31%2.53%

Siemens Healthcare Diagnostics Inc., Tarrytown, NY Re-standardized ADVIA Centaur TSTII and new Reference Intervals for ADVIA Centaur SHBG Traditional 510(k) K191533

{12}------------------------------------------------

10.6 Biotin Interference

Specimens that contain biotin at a concentration of 30 ng/mL demonstrate no significant effect on the assay. A summary of the results is presented below.

AnalyteConcentration(ng/dL)Biotin Test Level (ng/mL)
305010025050012001500
% Bias
83.2746238437039816837
289.242719107441>AR*>AR*
  • AR = Assay Range

Performance Characteristics: ADVIA Centaur® SHBG 11.

The change in reference intervals in the Instructions for Use (Package Inserts) for the ADVIA Centaur® SHBG that was cleared in 2016 under 510(k) K151986 does not require the collection of additional analytical performance data. Therefore, all analytical performance data previously reviewed for the ADVIA Centaur® SHBG continues to apply to this assay. because the assay was not modified. These performance data are cross-referenced to the 510(k) submission for the ADVIA Centaur® SHBG (K151986 and K091867).

Specifically, the following studies are not needed for the purpose of this submission:

  • Detection Capability (LoB, LoD, LoQ)
  • Linearity ●
  • Precision
  • . Dilution Recovery
  • Method Comparison
  • Calibrator/Assay Traceability ●
  • Reagent On-Board Stability and Calibration Interval
  • Cross reactivity ●
  • Interfering substances
  • Matrix Comparison
  • Shelf Life Stabilitv ●
  • Onboard Stability ●
  • Open Vial Stability ●

Clinical Studies 11.1

Not applicable.

11.2 Clinical Cut-off

Not applicable.

11.3 Biotin Interference

Specimens that contain biotin at a concentration of 300 ng/mL demonstrate a less than or equal to 10% change in results. A summary of the results is presented below.

{13}------------------------------------------------

AnalyteConcentration(nmol/L)Biotin Test Level (ng/mL)
38751503006001200
% Bias
19.6-3-3-9-7-7-25
44.1-21-3-7-9-26

Conclusions 12.

Based on the results of comparative testing, the re-standardized ADVIA Centaur® Testosterone II (TSTII) is substantially equivalent in principle and performance to the currently-marketed predicate device, the ADVIA Centaur® Testosterone II (TSTII) cleared under 510(k) K151986.

The ADVIA Centaur® SHBG with the new reference intervals in the Instructions for Use (package insert) is substantially equivalent to the currently marketed ADVIA Centaur® SHBG (K151986). The new reference intervals that will replace the current reference intervals in the Instructions for Use (Package Inserts) for the currently marketed assay do not require collection of additional analytical performance data.

§ 862.1680 Testosterone test system.

(a)
Identification. A testosterone test system is a device intended to measure testosterone (a male sex hormone) in serum, plasma, and urine. Measurement of testosterone are used in the diagnosis and treatment of disorders involving the male sex hormones (androgens), including primary and secondary hypogonadism, delayed or precocious puberty, impotence in males and, in females hirsutism (excessive hair) and virilization (masculinization) due to tumors, polycystic ovaries, and adrenogenital syndromes.(b)
Classification. Class I.