K Number
K191400

Validate with FDA (Live)

Date Cleared
2019-06-21

(28 days)

Product Code
Regulation Number
882.4560
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The V-TAG is intended to assist with stereotactic guidance, placement, and fixation for the operation of surgical instruments or devices during the planning and operation of neurological procedures performed in conjunction with preoperative and perioperative MR imaging. These procedures include laser coagulation, biopsies, catheter placement and electrode placement procedures.

Device Description

The V-TAG™ device is a single-use, skull-mounted, rigid trajectory array guide used in stereotactic surgical procedures. The V-TAG can be manipulated to provide a wide range of surgical trajectories into the head. An image-quided neuronavigational system loaded with a stereotactic surgical plan is used to align the V-TAG to the planned trajectory; once aligned, the V-TAG is securely locked to provide a stable instrument guide. After positioning of the V-TAG, magnetic resonance imaging (MRI) is used to confirm alignment with the planned trajectory. Adjustment of the V-TAG may be performed in the MRI scanner to refine or change the trajectory based on updated intraoperative imaging. Intracranial placement of a neurosurgical device or surgical instrument using the V-TAG device is only to be performed after MRI confirmation of the trajectory.

AI/ML Overview

The provided text describes a 510(k) premarket notification for the V-TAG (Voyager Trajectory Array Guide) device, a modification to a previously cleared predicate device (K180854). The core purpose of this submission is to modify the instructions for use to allow for trajectory adjustments in the MRI scanner based on intraoperative imaging.

Here's an analysis of the acceptance criteria and study information provided:

1. Table of Acceptance Criteria and Reported Device Performance:

Acceptance CriteriaReported Device Performance
X, Y, Z errors ≤ 2.0mmAll placements met the acceptance criteria.
Angular error of ≤ 2°All placements met the acceptance criteria.

2. Sample Size Used for the Test Set and Data Provenance:

  • Sample Size: The document states, "For each trial, the beginning alignment was intentionally set to an error greater than 10mm." It does not explicitly state the total number of "trials" or "placements" conducted. It implies multiple trials but lacks a specific numerical count for the test set.
  • Data Provenance: Not specified. The document does not mention the country of origin of the data or whether it was retrospective or prospective. Given the context of a 510(k) summary for a medical device, it is highly likely to be prospective testing specifically for this submission.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

  • This information is not provided in the document. The testing described appears to be a technical accuracy assessment (measuring alignment deviations) rather than an expert-adjudicated clinical outcome study.

4. Adjudication Method for the Test Set:

  • This information is not provided and is not applicable in the context of the described technical accuracy testing. There's no indication of human expert adjudication for the "ground truth" errors, as the errors were likely directly measured against the planned trajectory.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and the effect size of how much human readers improve with AI vs without AI assistance:

  • An MRMC comparative effectiveness study was not conducted. This device is a surgical guidance tool, not an AI-assisted diagnostic or interpretation system that would typically warrant such a study. The study focuses on the device's physical accuracy in guiding instruments.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

  • This question is not directly applicable as the V-TAG is a physical surgical guidance device, not primarily an algorithm or software that operates in a "standalone" fashion. While it works in conjunction with an image-guided neuronavigational system, the performance evaluation here is on the V-TAG's physical accuracy when manipulated by a human in a simulated or real surgical setting, confirmed by imaging. The "algorithm only" concept usually applies to diagnostic interpretations or automated decision-making.

7. The Type of Ground Truth Used:

  • The ground truth for the accuracy testing appears to be the pre-planned trajectory of the center lumen to the target. The "projected errors were then calculated," implying a measurement of deviation from this intended target.

8. The Sample Size for the Training Set:

  • This information is not applicable as the described study is a performance validation of a physical device's mechanical accuracy, not a machine learning model that requires a training set.

9. How the Ground Truth for the Training Set Was Established:

  • This information is not applicable as there is no training set mentioned or implied for this type of device validation.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

June 21, 2019

MRI Interventions, Inc. % John Smith Partner Hogan Lovells US LLP 555 Thirteenth St. NW Washington, District of Columbia 20004

Re: K191400

Trade/Device Name: Voyager Trajectory Array Guide (V-TAG) Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: HAW Dated: May 24, 2019 Received: May 24, 2019

Dear John Smith:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Matthew Krueger, M.S.E. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use Statement

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page

510(k) Number (if known)

K191400

Device Name V-TAG

Indications for Use (Describe)

The V-TAG is intended to assist with stereotactic guidance, placement, and fixation for the operation of surgical instruments or devices during the planning and operation of neurological procedures performed in conjunction with preoperative and perioperative MR imaging. These procedures include laser coagulation, biopsies, catheter placement and electrode placement procedures.

Type of Use (Select one or both, as applicable)

🇿 Prescription Use (Part 21 CFR 801 Subpart D)

□ Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.qov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (1/14)

{3}------------------------------------------------

K191400 - Page 1 of 2

510(k) SUMMARY

MRI Interventions V-TAG

Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared

Hogan Lovells, US LLP 555 Thirteenth Street, NW Washington, DC 20004 Phone: 202-637-5600 Fax: 202-637-5910

Contact Person: John J. Smith, M.D., J.D.

Date Prepared: May 24, 2019

Name of Device and Name/Address of Sponsor

ClearPoint System™ MRI Interventions, Inc. 5 Musick Irvine, CA 92618

Common or Usual Name: Neurological Stereotaxic Instrument

Classification: 21 C.F.R. §882.4560

Product Code: HAW

Predicate Device: Voyager Trajectory Array Guide (V-TAG) (K180854)

Purpose of the Special 510(k) notice.

The V-TAG is a modification to the predicate V-TAG (K180854).

Intended Use

The V-TAG is intended to assist with stereotactic guidance, placement, and fixation for the operation of surgical instruments or devices during the planning and operation of neurological procedures performed in conjunction with preoperative and perioperative MR imaging. These procedures include laser coagulation, biopsies, catheter placement and electrode placement procedures.

Device Description

The V-TAG™ device is a single-use, skull-mounted, rigid trajectory array guide used in stereotactic surgical procedures. The V-TAG can be manipulated to provide a wide range of

{4}------------------------------------------------

surgical trajectories into the head. An image-quided neuronavigational system loaded with a stereotactic surgical plan is used to align the V-TAG to the planned trajectory; once aligned, the V-TAG is securely locked to provide a stable instrument guide. After positioning of the V-TAG, magnetic resonance imaging (MRI) is used to confirm alignment with the planned trajectory. Adjustment of the V-TAG may be performed in the MRI scanner to refine or change the trajectory based on updated intraoperative imaging. Intracranial placement of a neurosurgical device or surgical instrument using the V-TAG device is only to be performed after MRI confirmation of the trajectory.

Technological Characteristics

The V-TAG device is a single use, MR compatible, sterile medical device consisting of a Base that is temporarily attached to the skull with four bone screws during the surgical procedure , a trajectory Guide containing seven channels to quide the insertion of 14-gauge neurosurgical devices or instruments, a Handle to assist with alignment using standard stereotactic surgical procedures, a Center Adaptor to identify the center trajectory of the Guide, an Imaging Cartridge to be filled with an FDA approved diluted gadolinium-based contrast agent (not included, no patient contact), and a Reducer for use with 16-gauge neurosurgical devices or instruments.

Performance Data

Accuracy testing was performed using the modified alignment technique for the V-TAG. For each trial, the beginning alignment was intentionally set to an error greater than 10mm. Alignment was then performed using rapid-refresh scanner sequences to manually align the center lumen of the V-TAG to the target. Projected errors were then calculated. If the projected error was more than 4mm, alignment was performed again. All placements met the acceptance criteria of X, Y, Z errors ≤2.0mm and angular error of ≤2°.

Substantial Equivalence

The modified V-TAG has the same intended use and similar indications, principles of operation, and technological characteristics as its predicate. The minor differences in the V-TAG's instructions for use do not raise any new questions of safety or effectiveness. Performance data demonstrates that the V-TAG is as safe and effective as its predicate. Thus, the V-TAG is substantially equivalent to its predicate device.

Conclusions

The V-TAG has the exact same intended use, indications for use, and substantially similar technological characteristics as the predicate. The purpose of this Special 510(k) notice was to modify the instructions for use of the device to allow for trajectory adjustments in the MRI scanner based on interoperative imaging. Accuracy testing provided in this submission demonstrate that the modified V-TAG accuracy is equivalent to that of the predicate device. Thus, the V-TAG is substantially equivalent.

§ 882.4560 Stereotaxic instrument.

(a)
Identification. A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.(b)
Classification. Class II (performance standards).