K Number
K191362
Manufacturer
Date Cleared
2020-05-20

(364 days)

Product Code
Regulation Number
880.5440
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The LifeFlow® Blood System is an intravenous administration set that is intended for the rapid* delivery of blood, blood components (red blood cells or plasma), and crystalloid and colloid resuscitative fluids from a container to a patient's vascular system. These devices may be used for any adult or pediatric patient, greater than 28 days old and greater than 3kg, with consideration given to adequacy of vascular anatomy, appropriateness for the solution being infused, and duration of therapy.

The delivery of blood products is intended for patients requiring volume replacement for hemorrhagic shock or life-threatening bleeding.

*Capable of rates greater than 100ml/min

Device Description

The LifeFlow® Blood System is a single use, disposable, intravenous administration set that is intended to deliver blood, blood components (red blood cells or plasma), and IV fluids from a container to a patient's vascular system rapidly using a hand pump. The set includes the handle, syringe, and tubing. The blood tubing features two IV spikes with roller clamps, blood filter chamber, force reduction tubing, and luer connector.

AI/ML Overview

This is a 510(k) premarket notification for the LifeFlow® Blood System, an intravenous administration set. The provided text describes the device, its intended use, and a summary of performance testing to establish substantial equivalence to a predicate device.

Here's an analysis of the acceptance criteria and supporting studies based on the provided document:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state "acceptance criteria" in a tabular format with corresponding "reported device performance." However, it describes several performance aspects the device was tested against, primarily to demonstrate substantial equivalence to the predicate and compliance with relevant standards. These can be inferred as the "acceptance criteria" for the device's functionality.

Acceptance Criteria (Inferred from testing)Reported Device PerformanceStudy that Proves the Device Meets Acceptance Criteria
Rapid Delivery CapabilityRates greater than 100ml/min (for blood, blood components, crystalloid, and colloid fluids)Functional testing: "Flow rate with LifeFlow® Handle activation"
Mechanical HemolysisCompared to reference device"Mechanical Hemolysis in comparison to the reference device"
Component Reliability-Functional testing: "reliability"
Human Packed Red Blood Cell Infusion Capability-Functional testing: "human packed red blood cell infusion capability"
BiocompatibilityComplies with ISO 10993-1, passed tests for cytotoxicity, sensitization, irritation, material-mediated pyrogenicity, acute systemic toxicity, hemocompatibility, particulates, and risk assessmentBiocompatibility evaluation per ISO 10993-1
Sterility Assurance Level (SAL)10⁻⁶Sterilization testing according to ANSI/AAMI/ISO 11137-2:2013
Pyrogenicity-Sterilization testing (including pyrogenicity testing)
Packaging and Shelf Life-Sterilization testing (including packaging and shelf life testing)
Luer Taper ConformityCompliant with ISO 594-1:1986 and ISO 594-2:1998Design verification in compliance to ISO 594-1:1986 and ISO 594-2:1998
Transfusion Set StandardsCompliant with ISO 1135-4:2015, specifically section 5.6 for blood filtersDesign verification in compliance to ISO 1135-4:2015
Air Embolism Prevention (AirCheck™)Ball float descends and forms a seal if ~20mL of air enters, stopping flow and indicating to user.Device description and implicitly, functional testing.
Force ReductionAllows for consistent fluid flow, decreases peak force and peak pressure.Device description and implicitly, functional testing.

2. Sample Size Used for the Test Set and Data Provenance

The document does not provide details on the sample sizes used for any of the specific tests (e.g., mechanical hemolysis, flow rate determination, reliability testing). It also does not specify the provenance of the data (country of origin, retrospective or prospective). All testing appears to be bench testing, which typically involves laboratory conditions rather than patient data.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This type of information is generally not applicable to bench testing of an administration set for fluid delivery. The "ground truth" for performance is established by objective measurements against engineering specifications and international standards, not by expert interpretation of patient data or images.

4. Adjudication Method for the Test Set

Not applicable, as the acceptance criteria are based on objective engineering measurements and compliance with standards, not on subjective assessments requiring adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

No, an MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic devices where human readers interpret medical images or data, and the goal is to assess improvement with AI assistance. The LifeFlow® Blood System is an intravenous administration set, not a diagnostic AI device.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

Not applicable. The LifeFlow® Blood System is a physical medical device, not an algorithm or AI system. It is designed for human operation (hand pump activation).

7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

For the performance testing mentioned (e.g., flow rate, hemolysis, biocompatibility), the ground truth is established by:

  • Objective engineering measurements: Data collected from bench tests related to fluid dynamics, material strength, etc.
  • Compliance with international standards: Verification against recognized standards like ISO 594-1, ISO 594-2, ISO 1135-4, and ISO 10993-1.

8. The Sample Size for the Training Set

Not applicable. The LifeFlow® Blood System is a physical medical device and does not involve machine learning algorithms that require a training set.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as no training set is used for this device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

May 20, 2020

410 Medical, Inc. % Danielle Besal Principal Consultant MRC Global 6075 Poplar Ave. Memphis, Tennessee 38119

Re: K191362

Trade/Device Name: LifeFlow® Blood System Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: Class II Product Code: FPA Dated: May 14, 2020 Received: May 15, 2020

Dear Danielle Besal:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

for Tina Kiang Ph.D. Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K191362

Device Name LifeFlow® Blood System

Indications for Use (Describe)

The LifeFlow® Blood System is an intravenous administration set that is intended for the rapid* delivery of blood, blood components (red blood cells or plasma), and crystalloid and colloid resuscitative fluids from a containt's vascular system. These devices may be used for any adult or pediatric patient, greater than 28 days old and greater than 3kg, with consideration given to adequacy of vascular anatomy, appropriateness for the solution being infused, and duration of therapy.

The delivery of blood products is intended for patients requiring volume replacement for hemorrhagic shock or lifethreatening bleeding.

*Capable of rates greater than 100ml/min

Type of Use (Select one or both, as applicable)
-------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

LifeFlow® Blood System K191362 May 19, 2020

Company:410 Medical, Inc.201 West Main St, Suite 207, Durham, NC 27701Phone: (844) 435-5450
EstablishmentRegistration:3012969104
Primary Contact:Danielle BesalPrincipal Consultant MRC GlobalPhone: (901)827-8670
Trade Name:LifeFlow® Blood System
Common Name:IV Administration Set
Classification:Class II
Regulation Number:21 CFR 880.5440
Panel:General Hospital
Product Code:FPA
Predicate Device:LifeFlow™ Rapid Infusion System (K153731)
Reference Device:IV Administration Sets with 200µm Blood Filter (K143082)

Device Description:

The LifeFlow® Blood System is a single use, disposable, intravenous administration set that is intended to deliver blood, blood components (red blood cells or plasma), and IV fluids from a container to a patient's vascular system rapidly using a hand pump. The set includes the handle, syringe, and tubing. The blood tubing features two IV spikes with roller clamps, blood filter chamber, force reduction tubing, and luer connector.

Indications for Use:

The LifeFlow® Blood System is an intravenous administration set that is intended for the rapid* delivery of blood, blood components (red blood cells or plasma), and crystalloid and colloid resuscitative fluids from a container to a patient's vascular system. These devices may be used for any adult or pediatric patient greater than 28 days old and greater than 3kg with consideration given to adequacy of vascular anatomy, appropriateness for the solution being infused, and duration of therapy.

The delivery of blood products is intended for patients reguiring volume replacement for hemorrhagic shock or life-threatening bleeding.

*Capable of rates greater than 100ml/min

Technological Characteristics Summary:

The subject LifeFlow® Blood System is substantially equivalent to the predicate device: LifeFlow™ Rapid Infusion System. The subject device components are similar to the predicate in terms of intended use,

{4}------------------------------------------------

geometry, and materials. The handle and syringe components are identical to the predicate, while the tubing differs to accommodate delivery of blood and blood components (red blood cells or plasma). The table below provides a comparison of technological characteristics. The minor differences in technological characteristics do not raise new questions of safety or effectiveness.

SUBJECTPREDICATEREFERENCE
LifeFlow® BloodSystemK191362LifeFlow™ RapidInfusion Device(K153731)IV AdministrationSets with 200 μmBlood Filter(K143082)Discussion
Product CodeIntended UseFPAAdministration ofblood, bloodcomponents (redblood cells orplasma), &crystalloid andcolloid resuscitativefluidsFPAAdministration ofcrystalloid andcolloid resuscitativefluidsFPAAdministration ofblood, bloodcomponents, & IVfluidsIdenticalDifference incompatible fluids;however, use of theproduct remainsidentical to thepredicate. Referencedevice is intended foruse with blood andblood components.
Indications for UseThe LifeFlow®Blood System is anintravenousadministration setthat is intended forthe rapid* deliveryof blood, bloodcomponents (redblood cells orplasma), andcrystalloid andcolloid resuscitativefluids from acontainer to apatient's vascularsystem. Thesedevices may beused for any adultor pediatric patient,greater than 28days old andgreater than 3kg,with considerationgiven to adequacyof vascularanatomy,appropriateness forthe solution beinginfused, andduration oftherapy.The LifeFlow™Rapid InfusionDevice is anintravenousadministration setwith Handleintended for rapid*delivery of fluidsfrom a containerinto a patient'svascular system.The device isintended to deliveronly crystalloid andcolloid resuscitativefluids. Thesedevices may beused for anypediatric or adultpatient populationwith considerationgiven to adequacyof vascularanatomy,appropriateness forthe solution beinginfused andduration oftherapy. *Capableof rates greaterthan 150 mL/minThe IVAdministrationSets with 200μmBlood Filter areused to deliverblood, bloodcomponents, andIV fluids from acontainer to apatient's vascularsystem. When thehand pumpcomponent isactivated, thedevice is intendedto deliver blood,blood productsand crystalloidand colloidresuscitativefluids. Thesedevices may beused for anypatientpopulation withconsiderationgiven to adequacyof vascularanatomy andappropriatenessfor the solutionbeing infused andDifference to theaddition of blood andblood components ascompatible fluids;however, the use ofthe product remainsidentical to thepredicate. Identicalto the subject device,the reference deviceis indicated fordelivery of blood andblood components.Difference inquantified flow rateversus predicate;however, theindicated subjectdevice flow rate issupported byperformance testing.Both subject andpredicate devices areindicated for similarpatient populations.The subject deviceprovides a morespecific patientpopulation than thepredicate.
The delivery ofblood products isintended forpatients requiringvolumereplacement forhemorrhagic shockor life-threateningbleeding.*Capable of ratesgreater than100mL/min.through a 20Gneedle.the duration oftherapy.
Patient PopulationAdults & pediatrics(greater than 28days old andgreater than 3kg)Adults & pediatricsSame as thepredicate withadditionalclarification ofpediatrics definitionfor subject device,which aligns with thepatient populationsevaluated in thesubject deviceperformance testing.
UsesSingleIdenticalIdentical to predicate
AdministrationMethodHand pumpactivationIdenticalIdentical to predicate
Hand Pump DesignHandle actuates asyringe to deliverfluid, whichautomatically refillswhen handle isreleasedIdenticalIdentical to predicate
BiocompatibilityISO 10993-1ISO 10993-1Identical to predicate
Sterilization MethodGamma (tubing setonly, handle is non-sterile)IdenticalIdentical to predicate
Sterility AssuranceLevel (SAL)10-6IdenticalIdentical to predicate
Components
BloodFilterMaterialPolyamideN/APolyamideDifference frompredicate due tosubject compatibilitywith blood; however,similar to referencedevice. The minordifference in filtersurface area versusthe reference devicedoes not raisedifferent questionsof safety or
Pore Size200μm200μm
SurfaceArea40cm260cm2
effectiveness as the
subject filter design
complies with ISO
1135-4 and was
verified through
bench testing.
TubingLength84"60"85"Identical diameters
ID0.094-0.170"0.094-0.170"0.120" & 0.190"to predicate,
MaterialNon-DEHP PVC &Non-DEHP PVCPVCincrease in length for
TPEcompatibility with
blood (similar to
reference device).
Air Chamber (AirYesNoDifference due to
Check)addition of air
chamber versus the
predicate. Reference
device has a drip
chamber; whereas,
the subject device air
chamber will stop
the flow of air if it
enters the chamber;
however, does not
raise new questions
of safety and
effectiveness.
Force ReductionYesNoDifference from
Tubing (Forcepredicate; however,
Reducer)
does not raise new
questions of safety or
effectiveness and
subject device has
been verified to yield
substantially
equivalent
performance to
predicate and
reference devices.
Spikes212Identical to predicate
spikes, only
difference is in
quantity to allow use
with two bags, which
is the same as the
reference device.
Clamps2 roller clamps & 11 pinch clamp3 roller clampsDifference due to
pinch clampsubject device ability
to connect to two IV
bags, similar to the
reference device.
The pinch clamp
component is

{5}------------------------------------------------

{6}------------------------------------------------

{7}------------------------------------------------

identical to that ofthe predicate deviceTwo roller clampswere added, whichmatches the 3 totalclamps of thereference device.This minor changedoes not raise newquestions of safety oreffectiveness.
Dual check valveYesYesIdentical to predicate
Luer connectorYesYesIdentical to predicate
Needleless access YconnectorYesYesIdentical to predicate

Difference In Technology:

  • . Air chamber - The subject device contains an air chamber ('AirCheck™') to prevent air from being delivered to the patient. AirCheck™ is located on the inlet tubing and assists the user in stopping the infusion if air trapped in the fluid bag enters the tubing. If a significant amount of air (approximately 20mL) enters the tubing, the ball float within AirCheck™ descends and forms a seal at the bottom of the chamber ('activates'). This activation prevents the LifeFlow® syringe from refilling and the handle from moving further (which provides a tactile indication to the user), while causing the flow of fluid to cease. The AirCheck™ chamber may be squeezed to purge any small volume of air. Since the subject device design requires constant user engagement, it has the advantage of the user immediately observing any activation of AirCheck™.
  • Force Reduction mechanism - The subject device's force reduction tubing, also known as the Force Reducer, allows for a more consistent flow of fluid through the IV catheter while the LifeFlow® Handle is cycled. The Force Reducer is a section of flexible tubing located in the outlet tubing that is designed to decrease the peak force applied by the user and reduce the peak pressure applied to fluid being infused. The Force Reducer is a passive mechanism that does not require conscious effort from the user to operate.
  • . Blood filter - The subject device blood chamber on the inlet tubing whereby blood enters and passes through a filter before flowing through the remaining tubing for administration to the patient. The filter is made from polyamide with a 200μm pore size and 40cm² surface area. The filter is compliant to section 5.6 of ISO 1135-4:2015 regarding filters for blood and blood components.

Performance Testing Summary:

Bench testing has verified the performance of the subject device is substantially equivalent to the predicate device. This testing included:

  • . Design verification in compliance to the following FDA recognized standards:
    • ISO 594-1:1986, Conical fittings with a 6% (Luer) taper for syringes, needles and certain o other medical equipment - Part 1: General requirements
    • ISO 594-2:1998, Conical fittings with a 6% (Luer) taper for syringes, needles and certain o other medical equipment - Part 2: Lock fittings

{8}------------------------------------------------

  • ISO 1135-4: 2015 Transfusion equipment for medical use, Part 4: Transfusion sets for o single use, gravity feed
  • Mechanical Hemolysis in comparison to the reference device
  • Functional testing included reliability, human packed red blood cell infusion capability, flow rate with LifeFlow® Handle activation
  • Human Factors usability testing
  • Guidance for Industry and FDA Staff – Intravascular Administration Sets Premarket Notification Submissions [510(k)], July 11, 2008

Biocompatibility

A biocompatibility evaluation was completed per ISO 10993-1, which included testing for cytotoxicity, sensitization, irritation, material-mediated pyrogenicity, acute systemic toxicity, hemocompatibility, particulates, and risk assessment.

Sterilization

The LifeFlow® Handle is not sterile, but the tubing is gamma sterilized. Sterilization testing were completed according to the FDA recognized ANSI/AAMI/ISO 11137-2:2013, Sterilization of healthcare products - Radiation - Part 2: Establishing the Sterilization Dose. Pyrogenicity (bacterial endotoxins), packaging, and shelf life testing were performed.

Substantial Equivalence Conclusions:

In conclusion, there are no changes in the use of the subject device compared to the predicate and the minor design changes do not raise new questions of safety and effectiveness; thus, the is substantially equivalent to the predicate.

§ 880.5440 Intravascular administration set.

(a)
Identification. An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.(b)
Classification. Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.