K Number
K191352
Manufacturer
Date Cleared
2019-08-08

(80 days)

Product Code
Regulation Number
866.3393
Panel
MI
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The binx health io CT/NG Assay, when tested using the binx health io Instrument, is a fully automated, rapid, qualitative test intended for use in point-of-care or clinical laboratory settings for the detection of Chlamydia trachomatis and Neisseria gonorrhoeae DNA in female vaginal swab specimens collected either by a clinician or self-collected by a patient in a clinical setting, to aid in the diagnosis of symptomatic or asymptomatic infection in female patients with Chlamydia trachomatis and/or Neisseria gonorrhoeae.

Device Description

The binx health io CT/NG Assay System (the "binx io System", "binx io CT/NG Assay" or the "System") is a rapid qualitative in vitro diagnostic system consisting of the following:

  1. The binx io Instrument for running the Cartridge (the "Instrument")
  2. The binx io CT/NG Cartridge (the "CT/NG Cartridge", "Cartridge" or "Cartridges"), which contains all the necessary reagents to perform the binx io CT/NG Assay (the "Assay") on the binx io Instrument
  3. A single-use, fixed-volume transfer pipet (packaged with the Cartridge) for transferring the sample to the Cartridge
  4. A female Vaginal Swab Specimen Collection Kit consisting of a swab and a sample Collection tube containing preservation medium (the "Vaginal Swab Specimen Collection Kit")

The binx io CT/NG Cartridge is a single-use assay-specific cartridge for use on a single patient. All reagents are contained in the Cartridge as a combination of liquid reagents in blister packs and dried reagents. The Instrument is a small, bench top, fully integrated Instrument that uses air pressure to open and close valves on the CT/NG Cartridge which, in turn, controls the movement of solutions within the Cartridge; the Instrument takes full control of the Cartridges once they are inserted. The operation of the Instrument requires a minimal number of steps that a user follows via a graphical user interface (GUI) screen to load the Cartridge onto the Instrument. Once the Cartridge is loaded, no further interaction by the user is required as no sample preparation is needed. Turnaround time from adding a raw patient sample to a result on the Instrument takes about 30 minutes.

The Vaginal Swab Collection Kit consists of a sterile flocked swab and a tube of preservative medium. The Cartridge has a visual sample loading indicator window which turns from light to dark to confirm to the user that a sample has been added to the Cartridge.

The Cartridge has three fully automated assay steps, (i) sample preparation to isolate and purify target DNA, (ii) ultra-rapid polymerase chain reaction (PCR), which amplifies specific regions of DNA from the target organisms, and (iii) a proprietary electrochemical detection to identify the presence of amplified DNA.

When the specimen is added to the Cartridge, it is automatically mixed with a lysis solution to disrupt the cells present and release DNA which also rehydrates the Internal Process Control (IPC) sample. DNA extraction takes place and the eluted DNA is transferred to a homogenization chamber.

Ultra-rapid PCR is carried out using sequence-specific primers for CT, NG (two separate genomic targets) and the IPC.

Amplified target DNA is detected by hybridization to electrochemically labeled probes and cleavage of the label using a double-strand specific exonuclease. The free label diffuses to the electrode surface and generates an electrical current measured at a distinct voltage in nano Amps (nA) for each electrochemical label used.

The presence of a measurable peak to a fixed cut-off parameter for each target returns a qualitative result with no requirement for interpretation or calculations.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text.

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are not explicitly stated as numerical targets before the results are presented. However, the study aims to demonstrate substantial equivalence to predicate devices and acceptable clinical performance. We can infer the performance targets from the reported results and the fact that the device received clearance. The performance is reported as sensitivity, specificity, and predictive values against a Composite Infected Status (CIS).

CriterionTarget Performance (Implied for Clearance)Reported Device Performance (binx health io CT/NG Assay)
Chlamydia trachomatis (CT)
Overall SensitivityHigh (e.g., >90%)96.1% (95% CI: 91.2% - 98.3%)
Overall SpecificityHigh (e.g., >98%)99.1% (95% CI: 98.4% - 99.5%)
PPV (Asymptomatic - 9.5% prev)High (context-dependent)92.9% (84.1% - 97.6%)
NPV (Asymptomatic - 9.5% prev)Very High (context-dependent)99.7% (98.9% - 100.0%)
PPV (Symptomatic - 7.6% prev)High (context-dependent)88.1% (77.8% - 94.7%)
NPV (Symptomatic - 7.6% prev)Very High (context-dependent)99.6% (98.8% - 99.9%)
Neisseria gonorrhoeae (NG)
Overall SensitivityHigh (e.g., >95%)100.0% (95% CI: 92.1% - 100.0%)
Overall SpecificityHigh (e.g., >99%)99.9% (95% CI: 99.5% - 100.0%)
PPV (Asymptomatic - 2.3% prev)High (context-dependent)94.1% (71.3% - 99.9%)
NPV (Asymptomatic - 2.3% prev)Very High (context-dependent)100.0% (99.5% - 100.0%)
PPV (Symptomatic - 3.5% prev)High (context-dependent)96.7% (82.8% - 99.9%)
NPV (Symptomatic - 3.5% prev)Very High (context-dependent)100.0% (99.5% - 100.0%)
Invalid Result RateLow (e.g.,

§ 866.3393 Device to detect nucleic acids from non-viral microorganism(s) causing sexually transmitted infections and associated resistance marker(s).

(a)
Identification. A device to detect nucleic acids from non-viral microorganism(s) causing sexually transmitted infections and associated resistance marker(s) is an in vitro diagnostic device intended for the detection and identification of nucleic acids from non-viral microorganism(s) and their associated resistance markers in clinical specimens collected from patients suspected of sexually transmitted infections. The device is intended to aid in the diagnosis of non-viral sexually transmitted infections in conjunction with other clinical and laboratory data. These devices do not provide confirmation of antibiotic susceptibility since mechanisms of resistance may exist that are not detected by the device.(b)
Classification. Class II (special controls). The special controls for this device are:(1) The intended use for the labeling required under § 809.10 of this chapter must include a detailed description of targets the device detects, the results provided to the user, the clinical indications appropriate for test use, and the specific population(s) for which the device is intended.
(2) Any sample collection device used must be FDA-cleared, -approved, or -classified as 510(k) exempt (standalone or as part of a test system) for the collection of specimen types claimed by this device; alternatively, the sample collection device must be cleared in a premarket submission as a part of this device.
(3) The labeling required under § 809.10(b) of this chapter must include:
(i) A detailed device description, including reagents, instruments, ancillary materials, all control elements, and a detailed explanation of the methodology, including all pre-analytical methods for processing of specimens;
(ii) Detailed discussion of the performance characteristics of the device for all claimed specimen types based on analytical studies, including Limit of Detection, inclusivity, cross-reactivity, interfering substances, competitive inhibition, carryover/cross contamination, specimen stability, within lab precision, and reproducibility, as appropriate;
(iii) Detailed descriptions of the test procedure, the interpretation of test results for clinical specimens, and acceptance criteria for any quality control testing;
(iv) Limiting statements indicating that:
(A) A negative test result does not preclude the possibility of infection;
(B) The test results should be interpreted in conjunction with other clinical and laboratory data available to the clinician;
(C) Reliable results are dependent on adequate specimen collection, transport, storage, and processing. Failure to observe proper procedures in any one of these steps can lead to incorrect results; and
(D) If appropriate (
e.g., recommended by the Centers for Disease Control and Prevention, by current well-accepted clinical guidelines, or by published peer reviewed research), that the clinical performance is inferior in a specific clinical subpopulation or for a specific claimed specimen type; and(v) If the device is intended to detect antimicrobial resistance markers, limiting statements, as appropriate, indicating that:
(A) Negative results for claimed resistance markers do not indicate susceptibility of detected microorganisms, as resistance markers not measured by the assay or other potential mechanisms of antibiotic resistance may be present;
(B) Detection of resistance markers cannot be definitively linked to specific microorganisms and the source of a detected resistance marker may be an organism not detected by the assay, including colonizing flora;
(C) Detection of antibiotic resistance markers may not correlate with phenotypic gene expression; and
(D) Therapeutic failure or success cannot be determined based on the assay results, since nucleic acid may persist following appropriate antimicrobial therapy.
(4) Design verification and validation must include:
(i) Detailed device description documentation, including methodology from obtaining sample to result, design of primer/probe sequences, rationale for target sequence selection, and computational path from collected raw data to reported result (
e.g., how collected raw signals are converted into a reported result).(ii) Detailed documentation of analytical studies, including, Limit of Detection, inclusivity, cross-reactivity, microbial interference, interfering substances, competitive inhibition, carryover/cross contamination, specimen stability, within lab precision, and reproducibility, as appropriate.
(iii) Detailed documentation and performance results from a clinical study that includes prospective (sequential) samples for each claimed specimen type and, when determined to be appropriate by FDA, additional characterized clinical samples. The study must be performed on a study population consistent with the intended use population and compare the device performance to results obtained from FDA accepted comparator methods. Documentation from the clinical studies must include the clinical study protocol (including a predefined statistical analysis plan) study report, testing results, and results of all statistical analyses.
(iv) A detailed description of the impact of any software, including software applications and hardware-based devices that incorporate software, on the device's functions.