(88 days)
HS-890A 980mm Diode Laser Therapy Device is intended for use in surgical application requiring the vaporization , incision, excision, ablation, cutting and hemostasis or coagulation of soft tissue. Such as vascular lesions and dental procedure.
The Diode Laser Therapy Device is a desktop device which utilizes a semiconductor diode with invisible infrared radiation as a laser source (980nm). The laser power is delivered to the treatment via a delivery system. The proposed device includes power supply system, delivery system, control system, cooling system and laser system.
The provided document is a 510(k) summary for a medical device called the "980nm Diode Laser Therapy Device." This document details the device's characteristics, its intended use, and its comparison to a predicate device for the purpose of demonstrating substantial equivalence. It is important to note that the document does not describe acceptance criteria and a study proving a device meets those criteria in the context of an AI/machine learning-based device. Instead, it outlines the regulatory approval process for a traditional medical laser device.
Therefore, many of the requested items (e.g., sample size for test set, expert qualifications for ground truth, MRMC study, training set information) are not applicable to the content of this document.
However, I can extract the information that is available:
1. A table of acceptance criteria and the reported device performance:
The document does not present acceptance criteria in a table format with reported device performance in the way one would for an AI/ML device evaluating statistical metrics (e.g., sensitivity, specificity). Instead, the acceptance criteria are implicitly the adherence to established performance standards for medical laser devices, and the reported device performance is its successful compliance with these standards.
| Acceptance Criteria (Implied) | Reported Device Performance |
|---|---|
| Compliance with Electrical Safety standard IEC 60601-1 | Device complies with IEC 60601-1: 2005+corr.1:2006+Corr.2.2007+A1:2012 |
| Compliance with Electromagnetic Compatibility standard IEC 60601-1-2 | Device complies with IEC 60601-1-2:2014 |
| Compliance with Laser Safety standard IEC 60825-1 | Device complies with IEC 60825-1:2014 |
| Compliance with Particular Requirements for Laser Equipment standard IEC 60601-2-22 | Device complies with IEC 60601-2-22:2007(third edition) +A1:2012 |
| Device functions as intended within its specifications (e.g., wavelength, power output, pulse width, operation mode, delivery system, aiming beam, laser class, operation interface, power supply, dimensions, weight, safety classification, software) | Device specifications (e.g., 980nm±10nm wavelength, 1-30W output power, 5~400ms pulse width, etc.) are listed and compared to the predicate device, implying performance within these stated parameters. The "Non-Clinical Testing" section states that tests were performed to "verify that the proposed device met all design specification" and "The test result demonstrated that the proposed device complies with the following standards." |
| Substantial Equivalence to Predicate Device (K151890) | Concluded: "Base on the performance testing and validation studies that the subject device is substantially equivalent to the predicate device." |
2. Sample sized used for the test set and the data provenance:
This information is not applicable. The device is a physical laser therapy device, not an AI/ML algorithm that processes data. Testing involved engineering and safety compliance rather than data analysis.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
This information is not applicable. Ground truth, in the context of this device, would refer to the physical and electrical safety standards outlined, which are established by recognized regulatory and standardization bodies.
4. Adjudication method for the test set:
This information is not applicable. Adjudication methods like 2+1 or 3+1 are typically used in clinical review of imaging or complex diagnoses for AI/ML performance evaluation. The testing here is for compliance with technical standards.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
This information is not applicable. This is not an AI-assisted device, and no human reader studies are mentioned.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
This information is not applicable. The device is a standalone physical laser system, not an algorithm.
7. The type of ground truth used:
The "ground truth" for this device's acceptance is its compliance with established international and national standards for medical electrical equipment and laser safety. These include:
- IEC 60601-1 (General requirements for basic safety and essential performance)
- IEC 60601-1-2 (Electromagnetic disturbances)
- IEC 60825-1 (Safety of Laser products - Equipment classification and requirements)
- IEC 60601-2-22 (Particular requirements for the safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment)
The document explicitly states that "The test result demonstrated that the proposed device complies with the following standards."
8. The sample size for the training set:
Not applicable, as this is not an AI/ML device.
9. How the ground truth for the training set was established:
Not applicable, as this is not an AI/ML device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized human figure. To the right is the FDA logo, with the acronym "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in a larger, bolder font, and "ADMINISTRATION" in a smaller font below.
August 16, 2019
Shanghai Apolo Medical Technology Co., Ltd. % Claire Zhang Manager Shanghai Landlink Medical Information Technology Co., Ltd. Room 703, 705 , Building 1, West Guangzhong Road 555, Jingan District, Shanghai, 200071 China
Re: K191349
Trade/Device Name: 980nm Diode Laser Therapy Device Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: May 17, 2019 Received: May 20, 2019
Dear Claire Zhang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance)and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Neil R. P. Ogden, M.S. Acting Team Assistant Director Light Based Devices Team DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
510(k) Number (if known) K191349
Device Name Diode Laser Therapy Device
Indications for Use (Describe)
HS-890A 980mm Diode Laser Therapy Device is intended for use in surgical application requiring the vaporization , incision, excision, ablation, cutting and hemostasis or coagulation of soft tissue. Such as vascular lesions and dental procedure.
| Type of Use (Select one or both, as applicable) |
|---|
| Prescription Use (Part 21 CFR 801 Subpart D) |
| Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Section 3-510(k) summary
I Submitter
Shanghai Apolo Medical Technology Co., Ltd. 4F, Building A, No.388, Yindu Road, Xuhui District, Shanghai 200231,China
Establishment Registration Number: 3007120647
Contact person: Felix Li Position: Regulatory Affairs Phone: +86-138 4919 0618 Fax: +86-21-34622840 E-mail: liqiang@apolo.com.cn
II Proposed Device
| Trade Name of Device: | 980nm Diode Laser Therapy Device |
|---|---|
| Common name: | Powered Laser Surgical Instrument |
| Regulation Number: | 21 CFR 878.4810 |
| Regulatory Class: | Class II |
| Product code: | GEX |
| Review Panel: | General & Plastic Surgery |
III Predicate Devices
| 510(k) Number: | K151890 |
|---|---|
| Trade name: | Medical Diode Laser Systems |
| Common name: | Powered Laser Surgical Instrument |
| Classification: | Class II |
| Product Code: | GEX |
| Manufacturer | Wuhan Gigaa Optronics Technology Company., Ltd. |
IV Device description
The Diode Laser Therapy Device is a desktop device which utilizes a semiconductor diode with invisible infrared radiation as a laser source (980nm). The laser power is delivered to the treatment via a delivery system.
{4}------------------------------------------------
The proposed device includes power supply system, delivery system, control system, cooling system and laser system.
V Indication for use
The Diode Laser Therapy Device is intended for use in surgical application requiring the vaporization, incision, excision, ablation, cutting and hemostasis or coagulation of soft tissue. Such as: vascular lesions and dental procedure.
| Item | Proposed device | Predicate device(K151890) |
|---|---|---|
| Product name | Diode Laser Therapy Device | Medical Diode Laser System |
| Product Code | GEX | GEX |
| Regulation No. | 21 CFR 878.4810 | 21 CFR 878.4810 |
| Class | Class II | Class II |
| Model | HS-890A | VELAS II-30B |
| Indication for use | The Diode Laser Therapy Device is intended for use in surgical application requiring the vaporization, incision, excision, ablation, cutting and hemostasis or coagulation of soft tissue. Such as: vascular lesions and dental procedure. | The “VELAS II-30A/30B” are indicated for use in surgical applications requiring the vaporization, incision, excision, ablation, cutting and hemostasis, or coagulation of soft tissue in conjunction with endoscopic equipment for medical specialist including: Urology (BPH), Genitourinary (UJrology), Thoracic Surgery, Plastic Surgery and Dermatology, Aesthetics including Vascular lesions and hair removal, General Surgery, Ophthalmology, Orthopedics, Podiatry, Arthroscopy, Spinal Surgery, Gynecology, Pulmonary Surgery, Neurosurgery (PLDD), Gastroenterology, Head/neck/ENT and Radiology |
| Endovascular coagulation, OralSurgery and Dental Procedures,Lase assisted lipolysis. | ||
| Laser Type | GaA1As Diode Laser | GaA1As Diode Laser |
| Components | Laser system, Color touchscreen, Fiber, Footswitch | Laser system, Color touch screen,Fiber, Footswitch. |
| Wavelength | 980nm±10nm | VELAS II-30B: 980nm±10nm |
| Outputpower | 1-30W | 1-30W |
| Pulse width | 5~400ms | 10ms~2.5s |
| Operationmode | CW, single pulse, repeatpulse | CW, single pulse, repeat pulse |
| LightDeliverysystem | Fiber core diameter: 300µm,NA=0.22With SMA905 connector | Fiber core diameter: 400µm, 600µmNA≥0.22With SMA905 connector |
| Aimingbeam | 650nm, ≤2mW , adjustable | Diode laser of 635/532nm powermax.<5mW, adjustable. |
| Laser class | 4 | 4 |
| Operationinterface | Color LCD touch screen | Color LCD touch screen |
| Powersupply | 100-240VAC, 50/60Hz | 100-240VAC, 50/60Hz, |
| Dimension | 280mm[W]× 270mm[L]×370mm[H] | 400mm[W]×385mm[L]×200mm[H] |
| Weight | 8kg | 12.9kg |
| Safetyclassification | Class I | Class I |
| Software | Yes | Yes |
VI Comparison of technological characteristics with the predicate devices
{5}------------------------------------------------
VII Non-Clinical Testing
A battery of tests have been performed to verity that the proposed device met all design specification. The test result demonstrated that the proposed device complies with the following standards:
Electrical safety and electromagnetic compatibility
IEC 60601-1: 2005+corr.1:2006+Corr.2.2007+A1:2012 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
{6}------------------------------------------------
IEC 60601-1-2:2014 Medical electrical equipment -Part 1-2: General requirements for basic safety and essential performance-Collateral Standard: Electromagnetic disturbances - Requirements and tests
IEC 60825-1:2014 Safety of Laser products-Part 1: Equipment classification and requirements
IEC 60601-2-22:2007(third edition) +A1:2012 for use in conjunction with IEC 60601-1:2005 (third edition) +A1:2012 Medical electrical equipment - Part 2-22: Particular requirements for the safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment
VIII Clinical Testing
It is not applicable.
IX Conclusion
Base on the performance testing and validation studies that the subject device is substantially equivalent to the predicate device.
§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.
(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.