K Number
K191287

Validate with FDA (Live)

Date Cleared
2019-07-16

(64 days)

Product Code
Regulation Number
878.4750
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The OptiFix AT Absorbable Fixation System with Articulating Technology is indicated for the approximation of soft tissue and fixation of surgical mesh to tissues during open or laparoscopic surgical procedures, such as hernia repair.

Device Description

The OptiFix™ AT Absorbable Fixation System with Articulating Technology, herein after referred to as "OptiFix™ AT", is a sterile (via gamma) single use device that is comprised of a deployment component and an absorbable fastener component. The device is available in either 15 or 30 preloaded fasteners. The tip of the shaft can be articulated, and the handle of the device can be rotated 360 degrees to facilitate access for fixation during surgery. The fasteners are designed with retention features and are manufactured from Poly (L-lactide and glycolide) and are dyed with D&C Violet No. 2.

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and study proving device performance:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state specific quantifiable acceptance criteria (e.g., minimum tensile strength, maximum deployment force) or numerical performance metrics for the OptiFix™ AT device. Instead, it uses qualitative statements about the device meeting established requirements.

Acceptance Criteria CategoryReported Device Performance
BiocompatibilitySuccessfully completed testing "in accordance with the FDA Guidance" and "International Standard ISO 10993-1."
Product Testing"Passed all the test requirements and demonstrated that the proposed device design meets product specifications and intended uses. All samples tested met the established acceptance criteria."
Package Integrity"Completed... in accordance with the following FDA consensus standards: ISO 11607-1+A1:2014" and "ISO 11607-2:2006+A1:2014."

2. Sample Sizes and Data Provenance

The document does not specify the sample sizes used for the product testing (design verification and validation).

  • Test Set Sample Size: Not specified.
  • Data Provenance: Not explicitly stated, but given the nature of pre-market submissions in the US, the testing would generally occur in the US or in adherence to US/international standards. It is retrospective in the sense that the testing was completed prior to the submission.

3. Number of Experts and Qualifications for Ground Truth Establishment

This information is not applicable as the device is a physical medical stapling system, not an AI/software device that requires expert-established ground truth for a test set. The performance is assessed through engineering and biological testing.

4. Adjudication Method for the Test Set

This information is not applicable as the device is a physical medical stapling system. Adjudication methods like 2+1 or 3+1 are typically used in clinical studies or for setting ground truth in imaging/AI studies, which were not conducted for this submission.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No, an MRMC comparative effectiveness study was not done. The document explicitly states: "Clinical studies were not performed for the submission of this device nor were clinical studies performed for the predicate device." MRMC studies are typically for evaluating diagnostic devices where human readers interpret results.

6. Standalone (Algorithm Only) Performance Study

No, a standalone performance study was not done. This device is a physical medical staple system and does not involve an algorithm.

7. Type of Ground Truth Used

The concept of "ground truth" as typically used in AI/diagnostic studies (e.g., expert consensus, pathology, outcomes data) is not directly applicable here. For this physical device, the "ground truth" for verifying performance relies on:

  • Engineering Specifications: The device's design, materials, and functional performance are compared against predetermined engineering specifications and safety standards.
  • Biocompatibility Standards: Performance against established international standards (ISO 10993).
  • Packaging Standards: Performance against established international standards (ISO 11607).

8. Sample Size for the Training Set

This information is not applicable. This is a physical medical device, not an AI algorithm, so there is no training set in the AI sense. The manufacturing process and quality control would involve internal testing of components and assemblies that could be analogous to "training" in a broader engineering sense, but not for an "AI training set."

9. How the Ground Truth for the Training Set Was Established

This information is not applicable for the same reason as point 8. The "ground truth" for the design and manufacturing of such a device is established through engineering design principles, material science, applicable regulatory standards, and risk assessments.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

July 16, 2019

Davol Inc., Subsidiary of C.R. Bard Angelica Hutchison Regulatory Affairs Specialist 100 Crossings Blvd Warwick, Rhode Island 02886

Re: K191287

Trade/Device Name: OptiFix AT Absorbable Fixation System with Articulating Technology - 30 Fasteners, OptiFix AT Absorbable Fixation System with Articulating Technology - 15 Fasteners Regulation Number: 21 CFR 878.4750 Regulation Name: Implantable Staple Regulatory Class: Class II Product Code: GDW Dated: June 14, 2019 Received: June 17, 2019

Dear Angelica Hutchison:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For Nina Mezu-Nwaba, PharmD., MPH., MSc, CAPT., United States Public Health Service Assistant Director (Acting), Plastic Surgery Implant Devices Team Division of Infection Control and Plastic Surgery Devices Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K191287

Device Name

OptiFix AT Absorbable Fixation System with Articulating Technology

Indications for Use (Describe)

The OptiFix AT Absorbable Fixation System with Articulating Technology is indicated for the approximation of soft tissue and fixation of surgical mesh to tissues during open or laparoscopic surgical procedures, such as hernia repair.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary Section 5

I. SUBMITTER

Davol Inc., a Subsidiary of C.R. Bard, Inc. 100 Crossings Boulevard Warwick, RI 02886

Angelica Hutchison Contact Person: Regulatory Affairs Specialist

Phone:(401) 825-8537
Fax:(401) 825-8765
E-mail:angelica.hutchison@bd.com

May 10, 2019 Date Prepared:

II. DEVICE

Name of Device:OptiFix™ AT Absorbable Fixation System
Common or Usual Name:Implantable Staple
Classification Name:Implantable Staple (21 CFR §878.4750)
Regulatory Class:II
Product Code:GDW

III. PREDICATE DEVICE

The predicate device for this submission is the OptiFix™ AT Absorbable Fixation System with Articulating Technology (K170278) cleared May 31, 2017. The predicate device has not been previously released to market and has not been subject to a design-related recall.

The reference device used in this submission is OptiFix™ Absorbable Fixation System (K142873), cheared on March 12, 2015; marketed by Davol Inc. This reference device has not been subject to a design related recall.

IV. DEVICE DESCRIPTION

The OptiFix™ AT Absorbable Fixation System with Articulating Technology, herein after referred to as "OptiFix™ AT", is a sterile (via gamma) single use device that is comprised of a deployment component and an absorbable fastener component. The device is available in either 15 or 30 preloaded fasteners. The tip of the shaft can be articulated, and the handle of the device can be rotated 360 degrees to facilitate access for fixation during surgery. The fasteners are designed with retention features and are manufactured from Poly (L-lactide and glycolide) and are dyed with D&C Violet No. 2.

The proposed device description is identical to what was previously cleared for the predicate device via K170278.

V. INDICATIONS FOR USE

The OptiFix™ AT Absorbable Fixation System with Articulating Technology is indicated for the The Open in - A - A - A soft tissue and fixation of surgical mesh to tissues during open and laparoscopic

{4}------------------------------------------------

surgical procedures, such as hernia repair. This indication is identical to the predicate device, OptiFix™ AT, K170278.

The intended use of the OptiFix™ AT device as covered by this Special 510(k) submission is identical to the predicate device, OptiFix™ AT, K170278. Both devices are intended for the approximation of to firsue and fixation of surgical mesh to tissues during open or laparoscopic surgical procedures, such as hernia repair.

VI. SUMMARY COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

The proposed OptiFix" AT device has the same indications for use, fundamental scientific technology and performance characteristics as the original OptiFix™ AT device, cleared via K170278. The and portormaned ogical characteristics of the OptiFix™ AT device are the same as the predicate device: device materials, delivery system design, principle of operation, biocompatibility, and sterilization. Modifications have been made to K170278 OptiFix™ AT device in order to optimize device performance. These modifications, which are the subject of this submission, are as follows:

  • Updated packaging configuration .
  • The Instructions for Use (IFU) was updated for clarification and to provide additional . information relating to changes to the device user interface
  • Minor dimensional and manufacturability changes to the fastener and delivery system to . optimize fastener deployment
  • Removal of the fastener level indicator .
Device FeaturesOptiFix™ AT(Predicate Device)K170278OptiFix™ AT(Proposed Device)
Intended UseSoft tissue fixationSame
Indication for UseIndicated for the approximation ofsoft tissue and fixation of surgicalmesh to tissues during open orlaparoscopic surgical procedures,such as hernia repair.Same
Fastener Materialand DyePoly(L-lactide-co-glycolide)(PLG)D&C Violet No. 2Conforms to 21 CFR §74.1602Same
Fastener BodyContactLong term implant (>30 days)contacting tissue and/or boneSame
FastenerShape/Design andDimensionsOverall fastener length: 7.1 mmFastener head: 3.5 mm diameterSame
FastenerAbsorption Time360 daysSame

Table 5-1: Summary Comparison of Predicate OptiFix™ AT (K170278) and Proposed OptiFix™ AT - General Characteristics

{5}------------------------------------------------

Device FeaturesOptiFix ™ AT (Predicate Device) K170278OptiFix ™ AT (Proposed Device)
Fastener Quantity per Device15 & 30 fastenersSame
Deployment Component Handle DesignPistol/Gun shape with 360 rotation optionSame
Deployment Component Shaft37 cm in length, 6 cm of articulationSame
Device SterilizationGamma IrradiationSame
Device Packaging ConfigurationPrimary Packaging: Device is placed into a cardboard insert which is inserted into a foil pouch. Secondary Packaging: Placed into a paperboard printed shelf carton.Primary Packaging: Device is placed into a PETG tray which is inserted into a foil pouch. Secondary Packaging: Same.

VII. PERFORMANCE DATA

Performance Standards

No performance standards have been established for this device under Section 514 of the Federal Food, Drug, and Cosmetic Act.

Biocompatibility Testing

The OptiFix™ AT device was evaluated and successfully completed biocompatibility testing in accordance with the FDA Guidance "Use of International Standard ISO 10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing." June 16, 2016; and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA.

Product Testing

The following non-clinical tests were completed for the proposed OptiFix™ AT device:

  • Design Verification Testing .
  • Design Validation Testing .

The proposed OptiFix™ AT device passed all the test requirements and demonstrated that the proposed device design meets product specifications and intended uses. All samples tested met the established acceptance criteria.

Package Integrity Testing

Package Qualification testing has been completed for the proposed OptiFix™ AT device in accordance with the following FDA consensus standards: ISO 11607-1+A1:2014 "Packaging for terminally sterilized medical devices Part 1: Requirements for materials, sterile barrier systems and packaging systems" and ISO 11607-2:2006+A1:2014 "Packaging for terminally sterilized medical devices Part 2: Validation requirements for forming, sealing, and assembly processes".

{6}------------------------------------------------

Animal Study

Allinial order. I he proposed Open in - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - in material and design. No additional animal studies were performed in support of this submission.

Clinical Studies

Clinical studies were not performed for the submission of this device nor were clinical studies performed for the predicate device.

VIII. CONCLUSIONS

The results of verification and validation testing demonstrate that the proposed OptiFix™ AT device is I he readly equivalent to the predicate device that has been previously cleared for the same intended uses.

§ 878.4750 Implantable staple.

(a)
Identification. An implantable staple is a staple-like device intended to connect internal tissues to aid healing. It is not absorbable.(b)
Classification. Class II.