K Number
K142873

Validate with FDA (Live)

Manufacturer
Date Cleared
2015-03-12

(162 days)

Product Code
Regulation Number
878.4750
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The OptiFix™ Absorbable Fixation System is indicated for the approximation of soft tissue and fixation of surgical mesh to tissues during open or laparoscopic surgical procedures, such as hernia repair.

Device Description

The OptiFix™ Absorbable Fixation System is a sterile (via gamma) single use product that is comprised of a deployment component and an absorbable fastener component. Two product ordering codes are to be packaged for distribution; each contains the same ergonomically designed deployment device. The variation will be the preloaded fasteners; either 15 or 30. The shaft of the OptiFix™ Absorbable Fixation System is 39 cm in length and is designed for use with 5mm trocars. The fasteners are designed with retention features and are manufactured from Poly (D, L)-lactide.

AI/ML Overview

The provided document is a 510(k) summary for the OptiFix™ Absorbable Fixation System. It details the device's characteristics and compares it to a predicate device to demonstrate substantial equivalence, rather than providing the results of a primary clinical study establishing acceptance criteria and device performance against those criteria. Therefore, most of the questions cannot be answered directly from the provided text.

However, I can extract information related to the performance data provided to support the substantial equivalence determination, specifically focusing on the non-clinical tests and their "acceptance criteria" which is implicitly "met the established acceptance criteria" in the context of the document.

Here's what can be extracted and what cannot:

1. A table of acceptance criteria and the reported device performance

The document states: "All samples tested met the established acceptance criteria." However, the specific quantitative acceptance criteria for each test are not provided. The "reported device performance" is summarized as the device passing these tests.

Acceptance Criteria (Not explicitly quantified in document)Reported Device Performance (as stated in document)
Established acceptance criteria for Actuation (trigger) torqueMet established acceptance criteria
Established acceptance criteria for Fastener deploymentMet established acceptance criteria
Established acceptance criteria for Fastener gap heightMet established acceptance criteria
Established acceptance criteria for Ball burst testingMet established acceptance criteria
Established acceptance criteria for Mesh compatibility testingMet established acceptance criteria
Established acceptance criteria for Resorption profile of the fastenerMet established acceptance criteria

2. Sample size used for the test set and the data provenance

The document does not specify the sample size used for the non-clinical tests (e.g., how many devices were tested for "Actuation torque"). It only states "All samples tested."

The data provenance is from non-clinical bench testing performed by Davol Inc. for the proposed OptiFix™ Absorbable Fixation System and the predicate device.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This question is not applicable as there were no clinical studies described, and thus no "ground truth" established by experts in this context. The "truth" for the non-clinical tests would be the measurement results against engineering specifications, not expert consensus on medical images or diagnoses.

4. Adjudication method for the test set

Not applicable. There was no clinical study with human readers requiring adjudication.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. No MRMC study was performed, and this device is a surgical fixation system, not an AI-assisted diagnostic tool.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This question is not applicable as the device is a physical surgical fixation system, not an algorithm.

7. The type of ground truth used

For the non-clinical tests, the "ground truth" would be the engineering specifications and established test protocols that define what constitutes a passing performance for each parameter (e.g., a specific torque range, a specific deployment force, a certain gap height). The document indicates "All samples tested met the established acceptance criteria," implying such specifications were used.

For the biocompatibility evaluation, the ground truth was based on recognized standards: "FDA Blue Book Memorandum #G95-1 "Use of International Standard ISO-10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing." May 1, 1995, and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA."

8. The sample size for the training set

Not applicable. This device is a physical surgical device, not a machine learning algorithm that requires a "training set."

9. How the ground truth for the training set was established

Not applicable, for the same reason as point 8.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 12, 2015

Davol Incorporated Ms. Christine Lloyd Regulatory Affairs Specialist 100 Crossings Boulevard Warwick, Rhode Island 02886

Re: K142873

Trade/Device Name: OptiFix™ Absorbable Fixation System Regulation Number: 21 CFR 878.4750 Regulation Name: Implantable staple Regulatory Class: Class II Product Code: GDW Dated: February 6, 2015 Received: February 9, 2015

Dear Ms. Lloyd:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{1}------------------------------------------------

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

for

Sincerely yours,

David Krause -S

Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K142873

Device Name OptiFix™ Absorbable Fixation System

Indications for Use (Describe)

The OptiFix™ Absorbable Fixation System is indicated for the approximation of soft tissue and fixation of surgical mesh to tissues during open or laparoscopic surgical procedures, such as hernia repair.

Type of Use (Select one or both, as applicable)

☒ Prescription Use (Part 21 CFR 801 Subpart D)
□ Over-The-Counter Use (21 CFR 801 Sub

__ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

I. SUBMITTER

Davol Inc. 100 Crossings Boulevard Warwick, RI 02886

Contact Person:Christine Lloyd
Regulatory Affairs Specialist
Phone:(401) 825-8447
Fax:(401) 825-8753
E-mail:Christine.Lloyd@crbard.com

Date Prepared: March 12 2015

II. DEVICE

Name of Device:OptiFix™ Absorbable Fixation System [product ordering codes 0113126 (deployment device contains 30 fasteners) and 0113127 (deployment device contains 15 fasteners)]
Common or Usual Name:Staple Implantable/Implantable Staple
Classification Name:Implantable staple (21 CFR §878.4750)
Regulatory Class:II
Product Code:GDW

III. PREDICATE DEVICE

OptiFixTM Absorbable Fastener System, K132134; marketed by Davol, Inc., This predicate has not been subject to a design-related recall.

The reference device Davol Absorbable Fastem - SorbaFix™ (K082396) is used in this submission.

IV. DEVICE DESCRIPTION

The OptiFix™ Absorbable Fixation System is a sterile (via gamma) single use product that is comprised of a deployment component and an absorbable fastener component. Two product ordering codes are to be packaged for distribution; each contains the same ergonomically designed deployment device. The variation will be the preloaded fasteners; either 15 or 30. The shaft of the OptiFix™ Absorbable Fixation System is 39 cm in length and is designed for use with 5mm trocars. The fasteners are designed with retention features and are manufactured from Poly (D, L)-lactide.

SECTION 7

{4}------------------------------------------------

V. INDICATIONS FOR USE

The OptiFix™ Absorbable Fixation System is indicated for the approximation of soft tissue and fixation of surgical mesh to tissues during open or laparoscopic surgical procedures, such as hernia repair.

The Indications for Use statement for the subject device is identical to the predicate device. Both the subject and predicate devices have the same intended use for the approximation of soft tissue and fixation of surgical mesh to tissues during open or laparoscopic surgical procedures, such as hernia repair.

VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

The approximation of soft tissue and fixation of surgical mesh to tissues during open or laparoscopic surgical procedures, such as hernia repair using the mechanical actuation (trigger) force to guide the fastener and launch the implant is the principle for both the proposed and predicate devices.

At a high level, the proposed and predicate devices are based on the following same elements:

  • Deployment component mechanical component used to guide the fastener into tissue ●
  • Use of a mechanical component for positioning and launching the implant ●
  • User-controlled mechanical trigger to guide the fastener (implant) ●
  • Fastener component - absorbable component based on the same material used to fixate surgical mesh to tissue

At a high level, the following differences exist between the proposed and predicate devices:

  • The proposed deployment device has a resistive trigger and the predicate has a rotating ● trigger.
  • The proposed fastener component has slight differences in geometry than the predicate.

VII. PERFORMANCE DATA

The following performance data were provided in support of the substantial equivalence determination.

Performance standards

No performance standards have been established for this device under Section 514 of the Federal Food, Drug, and Cosmetic Act.

{5}------------------------------------------------

Biocompatibility testing

The biocompatibility evaluation for the OptiFix™ Absorbable Fixation System was conducted in accordance with the FDA Blue Book Memorandum #G95-1 "Use of International Standard ISO-10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing." May 1, 1995, and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA.

Testing for the fastener included Cytotoxicity and the complete biocompatibility profile for the OptiFix fastener has been supported by full biocompatibility testing on the reference device, SorbaFix (K082396) as listed in the following. The fastener materials between the OptiFix and the reference device have been demonstrated to be substantially equivalent.

  • Sensitization ●
  • Intracutaneous Reactivity
  • Single-dose Systemic Toxicity ●
  • Rabbit Pyrogenicity
  • 14-Day Repeat Dose Intravenous Toxicity
  • Genotoxicity ●
  • Chronic Toxicity, 13 weeks
  • . Implantation. 52 weeks

Testing for the new materials within the deployment device included Cytotoxicity, Sensitization, Intracutaneous Reactivity, and Single-dose Systemic Toxicity.

The deployment component of the proposed OptiFix™ Absorbable Fixation System is determined to be tissue contacting for duration of less than 24 hours, while the absorbable fasteners are determined to be permanent implants.

Testing

The following non-clinical tests were completed for the proposed and predicate device. The OptiFix™ Absorbable Fixation System passed all the test requirements and demonstrated substantial equivalence to the test results of the predicate device.

  • Performance and Functional testing of the OptiFix™ Absorbable Fixation System-● deployment device and fastener
    • Actuation (trigger) torque .
    • . Fastener deployment
    • . Fastener gap height
    • . Ball burst testing
  • Mesh compatibility testing of the OptiFix™ Absorbable Fixation System
  • Resorption profile of the fastener

All samples tested met the established acceptance criteria.

SECTION 7

{6}------------------------------------------------

Animal Study

Animal studies were not performed on the proposed device for this submission. Based on identical materials and comparable mechanical strengths measured at the bench level, the in vivo safety and performance of the proposed device was evaluated via the animal and histological studies conducted on the reference device, the SorbaFix Fixation Device.

Clinical Studies

Clinical studies were not performed for the submission of this device nor were clinical studies performed for the predicate device.

Device FeaturesProposed deviceOptiFix™Predicate deviceOptiFix™ (K132134)
Intended UseSoft tissue fixationSoft tissue fixation
Indication For useIndicated for the approximationof soft tissue and fixation ofsurgical mesh to tissues duringopen or laparoscopic surgicalprocedures, such as hernia repair.Indicated for the approximation ofsoft tissue and fixation of surgicalmesh to tissues during open orlaparoscopic surgical procedures,such as hernia repair.
Fastener MaterialPoly (D,L) LactidePoly (D,L) Lactide
Fastener Violet DyeD & C Violet No. 2conforms to 21 CFR §74.3602D & C Violet No. 2conforms to 21 CFR §74.3602
Fastener BodyContactLong term implant (>30 days)contacting tissue and/or boneLong term implant (>30 days)contacting tissue and/or bone
FastenerShape/DesignPush Tack with retention featureson endPush Tack with retention featureson end
Fastener Dimensions6.7 mm overall fastener lengthFastener head: 3.7 mm length; 3.5mm thickness6.7 mm overall fastener lengthFastener head: 3.7 mm length; 3.5mm thickness
FastenerManufacturingMethodInjection-MoldedInjection-Molded
Fastener AbsorptionTime360 days360 days
Fastener Quantityper Device15 & 30 fasteners15 & 30 fasteners

Table 7-1: Device Substantial Equivalence – General Characteristics

Premarket Notification for OptiFix™ Absorbable Fixation System

{7}------------------------------------------------

Device FeaturesProposed deviceOptiFix™Predicate deviceOptiFix™ (K132134)
Fastener DeliverySystemPush - Impact tube pushesfasteners forward over a straightguidewirePush – Impact tube pushesfasteners forward over a straightguidewire
Deployment DeviceHandle designPistol/Gun shapePistol/Gun shape
Deployment DeviceShaft NominalLength39 cm in length39 cm in length
Deployment DeviceTriggerResistive triggerRotating trigger
Deployment DeviceTrigger StrokeSwift and fullSlow and progressive
Deployment DeviceGuidewireMoveableStationary
Device SterilizationGamma Irradiation(25 - 40 kGy)Gamma Irradiation(25 - 40 kGy)

VIII. CONCLUSIONS

The OptiFix™ Absorbable Fixation System is substantially equivalent to the legally marketed predicate device for the following reasons:

  • A) The same intended use and indications for use as the predicate device.
  • B) Both devices use a similar fixation technology to deliver the fasteners by compressing a trigger.
  • C) Both devices house similar synthetic absorbable fastener with retention head design.
  • D) Identical materials with long history of biocompatible use in medical instrumentation and implantation.
  • E) Identical technological characteristics to the predicate devices such as: Fastener dimensions trigger handle, penetration depth, and shaft length.
  • F) Same principle of operation.

The comparative analysis as well as the bench and preclinical testing results demonstrated that the OptiFix™ Absorbable Fixation System should perform as safe, and as effective, as the predicate device that is currently marketed for the same intended use.

SECTION 7

§ 878.4750 Implantable staple.

(a)
Identification. An implantable staple is a staple-like device intended to connect internal tissues to aid healing. It is not absorbable.(b)
Classification. Class II.