K Number
K191266
Device Name
VitalScan ANS
Manufacturer
Date Cleared
2020-01-15

(250 days)

Product Code
Regulation Number
870.2780
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

VitalScan ANS is intended for non-invasive measurements of pulse waveforms by photoelectric plethysmography (PPG), heart rate electrocardiograph (ECG), Heart Variability measurements (HRV) and blood pressure (NIBP) in response to paced respiration and controlled testing procedures for physician assessment of the Cardiovascular Systems.

VitalScan ANS also measures/calculates a patient's Brachial and Ankle Blood Pressure, Ankle Brachial Pressure Index (ABPI) and Ankle Brachial Index (ABI) and provides Pulse Volume Recording (PVR) / volume plethysmography for assessing development of peripheral arterial disease (PAD).

The device is intended for use on transitional adolescents and adults, including those with unilateral lower limb amputation, in medical clinics, healthcare practices and out-patient departments of hospitals.

The physician has the responsibility of interpreting the significance of the resulting data.

Device Description

VitalScan ANS system collects multi-parameter patient data including: Electrocardiography (ECG), Plethysmogram (PPG), Pulse Volume Recording (PVR), Blood Pressure, Heart Rate, and Peripheral capillary Oxygen saturation (SpO2).

The system comprises: USB plug and play device hardware and Software installed on a computer.

VitalScan ANS is intended to measure a patient's variations in the heart rate (R-R beat-to-beat intervals from ECG) and perform Heart Rate Variability (HRV) analysis to assess Autonomic Nervous System (ANS) Function;

VitalScan ANS also measure a patient's Brachial and Ankle Blood Pressure, Ankle Brachial Pressure Index (ABPI) and Ankle Brachial Index (ABI) and provides Pulse Volume Recording (PVR) / volume plethysmography for assess the risk of developing peripheral arterial disease (PAD).

The results are saved in a backup and can also be printed.

The device does not provide any direct diagnosis rather the device provides information to the physician for inclusion in their decision making process.

AI/ML Overview

The provided document is a 510(k) summary for the VitalScan ANS device. It primarily focuses on demonstrating substantial equivalence to predicate devices through technological characteristic comparison and non-clinical testing.

Here's an analysis of the provided information regarding acceptance criteria and device performance:

1. Table of Acceptance Criteria and Reported Device Performance

The document provides a comparison table (pages 7-8) that highlights the VitalScan ANS device's specifications against predicate devices. While this table shows performance characteristics, it does not explicitly state "acceptance criteria" in a quantifiable sense for the VitalScan ANS itself. Instead, it demonstrates that the VitalScan ANS either matches or is similar to the established performance of legally marketed predicate devices.

Here's an extracted and summarized table based on the direct comparison for the VitalScan ANS device's performance, using the predicate device values as a proxy for what would be considered acceptable based on substantial equivalence:

Feature/ParameterAcceptance Criteria (from Predicate/Reference)Reported Device Performance (VitalScan ANS)
ECG
MethodECG lead wires attached to disposable electrodes to the skinECG lead wires attached to disposable electrodes to the skin
Resolution16 bit (from ANSHA-QHRV1)24 bit
Input impedance> 20 Mohm (from ANSHA-QHRV1)> 20 Mohm
Common mode rejection-80dB - 100dB (from ANSHA-QHRV1)-80dB - 100dB
Sampling frequency800, 400 and 200 Hz (from ANSHA-QHRV1)1000, 500, 200 Hz
Channels3 channels (from ANSHA-QHRV1)3 or 12 channels
QRS detection99.8% (from ANSHA-QHRV1)99.8%
Permanent DisplayNot provided (from ANSHA-QHRV1)Not provided
Applied parts in contact with patientECG Electrodes (from ANSHA-QHRV1)ECG Electrodes
HEART RATE
MethodQRS detection (from ANSHA-QHRV1)QRS detection
Range40 - 200 bpm (from ANSHA-QHRV1)38 - 250 bpm
Accuracy± 2 bpm (from ANSHA-QHRV1)± 2 bpm
BLOOD PRESSURE
Measurement typesOscillometric measuring method during deflation of the cuff (from ABPI MD / Tonoport V)Oscillometric measuring method during deflation of the cuff
Measurement ranges (Systolic)60 to 260 mmHg (from Tonoport V)60 to 260 mmHg
Measurement ranges (Diastolic)40 to 220 mmHg (from Tonoport V)40 to 220 mmHg
Measurement ranges (HR)35 to 240 min-1 (from Tonoport V)35 to 240 min-1
Limit values of measurement errors (ABPI)± 0.1 (from ABPI MD)± 0.1
Limit values of measurement errors (Systolic)4.6 mmHg (Empirical Standard Deviation - from Tonoport V)4.6 mmHg (Empirical Standard Deviation)
Limit values of measurement errors (Diastolic)4.4 mmHg (Empirical Standard Deviation - from Tonoport V)4.4 mmHg (Empirical Standard Deviation)
Cuffs inflation and deflationAutomatic inflation using an air pump and deflation using an electromagnetic valve. Max 300 mmHg (from Tonoport V)Automatic inflation using an air pump and deflation using an electromagnetic valve. Max 300 mmHg
Pulse Volume / PlethysmographyPneumo-plethysmography method using the cuffs measuring the blood pressure values: Plethysmography displayed at the inflation and deflation pressure (from ABPI MD)Pneumo-plethysmography method using the cuffs measuring the blood pressure values: Plethysmography displayed at the inflation and deflation pressure
Applied parts in contact with patient3 cuffs, tubes and bladders (from ABPI MD) / 1 cuff, tube and bladder (from Tonoport V)2 or 4 cuffs, tubes and bladders
OXYGEN SATURATION
MethodPhoto plethysmogram on finger. Infrared: 910 nm @ 1.2 mW maximum average, Red: 660 nm @ 0.8 mW maximum average (from ANSHA-QHRV1)Photo plethysmogram on finger. Infrared: 910 nm @ 1.2 mW maximum average, Red: 660 nm @ 0.8 mW maximum average
Range70-100% (from ANSHA-QHRV1)70-100%
Accuracy± 2 digits (from 70-100%) (from ANSHA-QHRV1)± 2 digits (from 70-100%)

2. Sample size used for the test set and the data provenance

The document explicitly states: "No Clinical testing was necessary to determine substantial equivalence." (Page 6, section b2). This means there was no test set of patient data used to evaluate the device's clinical performance. The evaluation was based on non-clinical (bench) testing and comparison to predicate devices, not on direct human data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Since no clinical testing was performed, there were no experts used to establish ground truth for a clinical test set. The substantial equivalence was determined by comparing the device's technical specifications and safety/performance data (from bench testing) to those of already approved predicate devices.

4. Adjudication method for the test set

Not applicable, as no clinical test set was used.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is not an AI-assisted diagnostic tool that would typically involve human readers.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Yes, if we interpret "standalone" as the device operating according to its technical specifications and algorithms during non-clinical (bench) testing. The document states:

  • "Bench testing was carried out on the following characteristics: Electrocardiograph (ECG), Heart rate variability (R-R interval), Heart rate, SpO2 and Plethysmogram, Blood-Pressure Measurement accuracy, Communication, data transmission and storage integrity, Electromagnetic compatibility (EMC), Electrical safety testing, Software verification and validation testing, Biocompatibility verification." (Page 5, section b1)
  • And, "The VitalScan ANS device was tested and meets the applicable requirements of following performance Standards and is in accordance with FDA Class II Special Controls Guidance Document" followed by a list of relevant IEC and ISO standards (pages 5-6).

These non-clinical tests demonstrate the device's performance in a standalone capacity against established technical and safety standards.

7. The type of ground truth used

For the non-clinical testing, the "ground truth" would be established by the specifications and measurement accuracies defined in the referenced standards (e.g., IEC 60601-1, AAMI / ANSI 80601-2-30 for blood pressure accuracy). The device's measurements were compared against calibrated instruments or reference signals as part of these bench tests. The substantial equivalence argument also uses the established performance of the legally marketed predicate devices as a form of "ground truth" for acceptable performance.

8. The sample size for the training set

Not applicable. This device is not described as utilizing machine learning or artificial intelligence that would require a "training set" of data. Its functionalities are based on established physiological measurement principles and algorithms.

9. How the ground truth for the training set was established

Not applicable, as there is no training set described for this device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

January 15, 2020

Medeia, Inc. % Daniel Lehtonen Regulatory Consultant Compliance and Regulatory Services LLC 3771 Southbrook Dr Dayton, Ohio 45430

Re: K191266

Trade/Device Name: VitalScan ANS Regulation Number: 21 CFR 870.2780 Regulation Name: Hydraulic, Pneumatic, or Photoelectric Plethysmographs Regulatory Class: Class II Product Code: JOM, DPS, DXN Dated: December 11, 2019 Received: December 16, 2019

Dear Daniel Lehtonen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jennifer Shih Assistant Director (Acting) Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K191266

Device Name

VitalScan ANS

Indications for Use (Describe)

VitalScan ANS is intended for non-invasive measurements of pulse waveforms by photoelectric plethysmography (PPG), heart rate electrocardiograph (ECG), Heart Variability measurements (HRV) and blood pressure (NIBP) in response to paced respiration and controlled testing procedures for physician assessment of the Cardiovascular Systems.

VitalScan ANS also measures/calculates a patient's Brachial and Ankle Blood Pressure, Ankle Brachial Pressure Index (ABPI) and Ankle Brachial Index (ABI) and provides Pulse Volume Recording (PVR) / volume plethysmography for assessing development of peripheral arterial disease (PAD).

The device is intended for use on transitional adolescents and adults, including those with unilateral lower limb amputation, in medical clinics, healthcare practices and out-patient departments of hospitals.

The physician has the responsibility of interpreting the significance of the resulting data.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Medeia. The logo has the word "Medeia" in a sans-serif font, with the "M" in green and blue, and the rest of the letters in black. Below the word "Medeia" is the phrase "Designed for Clinicians" in a smaller, sans-serif font.

510(k) SUMMARY

This summary of 510(k) information is submitted in accordance with the Requirements of Safe Medical Device systems Act 1990 and 21 CFR Sec. 807.92

510(k) Number:

K191266

a1 APPLICANT INFORMATION:

Date Prepared:03 May 2019
Name:Medeia, Inc.
Address:7 W. Figueroa StreetSuite 300Santa Barbara, CA, 93101
Contact Person:Slav Danev
Phone Number:+1 800 433 4609
Fax Number:+1 800 433 4609
Email:danev@medeia.com

a2 NAME OF DEVICE:

Trade Name:VitalScan ANS
Common Name:Plethysmograph, photoelectric, pneumatic or hydraulic
Classification Name:Plethysmograph, photoelectric, pneumatic or hydraulic21 CFR 870.2780 (JOM)Electrocardiograph
21 CFR 870.2340 (DPS)Noninvasive blood pressure measurement system
21 CFR 870.1130 (DXN)

аЗ PREDICATE DEVICES:

Primary Predicate Device:K083735; ANSHA-QHRV1
Secondary Predicate Device:K101983; Critical Care Assessment
Secondary Predicate Device:K172655; ABPI MD
Reference Device:K012647; Tonoport V

The FDA database for recalls was searched on 20 March 2019 during the preparation of the 510(k) submission and no recalls for the devices noted above were found.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Medeia. The logo has the word "Medeia" in a sans-serif font, with the "M" in green and blue and the rest of the letters in black. Below the word "Medeia" is the phrase "Designed for Clinicians" in a smaller, blue sans-serif font.

INDICATIONS FOR USE: a4

VitalScan ANS is intended for non-invasive measurements of pulse waveforms by photoelectric plethysmography (PPG), heart rate electrocardiograph (ECG), Heart Variability measurements (HRV) and blood pressure (NIBP) in response to paced respiration and controlled testing procedures for physician assessment of the Cardiovascular Systems.

VitalScan ANS also measures/calculates a patient's Brachial and Ankle Blood Pressure, Ankle Brachial Pressure Index (ABPI) and Ankle Brachial Index (ABI) and provides Pulse Volume Recording (PVR) / volume plethysmography for assessing development of peripheral arterial disease (PAD).

The device is intended for use on transitional adolescents and adults, including those with unilateral lower limb amputation, in medical clinics, healthcare practices and out-patient departments of hospitals.

The physician has the responsibility of interpreting the significance of the resulting data.

a5 DESCRIPTION OF THE DEVICE:

VitalScan ANS system collects multi-parameter patient data including: Electrocardiography (ECG), Plethysmogram (PPG), Pulse Volume Recording (PVR), Blood Pressure, Heart Rate, and Peripheral capillary Oxygen saturation (SpO2).

The system comprises: USB plug and play device hardware and Software installed on a computer.

VitalScan ANS is intended to measure a patient's variations in the heart rate (R-R beat-to-beat intervals from ECG) and perform Heart Rate Variability (HRV) analysis to assess Autonomic Nervous System (ANS) Function;

VitalScan ANS also measure a patient's Brachial and Ankle Blood Pressure, Ankle Brachial Pressure Index (ABPI) and Ankle Brachial Index (ABI) and provides Pulse Volume Recording (PVR) / volume plethysmography for assess the risk of developing peripheral arterial disease (PAD).

The results are saved in a backup and can also be printed.

The device does not provide any direct diagnosis rather the device provides information to the physician for inclusion in their decision making process.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Medeia. The logo has a large letter "M" in green and blue, followed by the rest of the word "edeia" in black. Below the word "Medeia" is the phrase "Designed for Clinicians" in blue.

a6 TECHNOLOGICAL CHARACTERISTIC COMPARISON:

VitalScan ANS product share similar device characteristics, intended use, performance, specifications, sensors and is the same in design, function, and application to the predicate devices.

The comparison table, beginning on page 5, demonstrates that the VitalScan ANS device is substantially equivalent to the predicate devices. The nonclinical data support the safety of the device and the hardware and software verification and validation demonstrate that the VitalScan ANS device should perform as intended in the specified use conditions.

Based on comparisons of device technological characteristics, features, materials, intended use and performance the VitalScan ANS has been shown to be substantially equivalent to the commercially available predicate devices.

NON-CLINICAL TESTING: b1

The VitalScan ANS device was subjected to the following non-clinical performance testing:

Bench testing was carried out on the following characteristics:

  • Electrocardiograph (ECG)
  • Heart rate variability (R-R interval)
  • Heart rate
  • SpO2 and Plethysmogram ●
  • Blood-Pressure Measurement accuracy
  • Communication, data transmission and storage integrity
  • Electromagnetic compatibility (EMC)
  • Electrical safety testing
  • Software verification and validation testing
  • Biocompatibility verification

In addition to the above, usability testing was also conducted.

Referenced Standards and Performance Testing:

The VitalScan ANS device was tested and meets the applicable requirements of following performance Standards and is in accordance with FDA Class II Special Controls Guidance Document:

  • IEC 60601-1 Medical Electrical Equipment - Part 1: Basic safety and essential performance Ed 3.1 2005+A1:2012
  • . IEC 60601-1-2 Medical Electrical Equipment - Basic safety and essential performance-EMC-Edition 3: 2007-03
  • IEC 60601-1-6 Medical Electrical Equipment - Part 1-6: General Requirements for Basic Safety and Essential Performance - Collateral Standard: Usability; Edition 3.1 2013-10
  • AAMI / ANSI 60601-2-27 Particular requirements for the basic safety and essential performance of electrocardiographic monitoring equipment 2011
  • . ISO 80601-2-61 Particular requirements for basic safety and essential performance of pulse oximeter equipment 2011

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Medeia. The logo has the word "Medeia" in a sans-serif font, with the "M" in green and blue and the rest of the letters in black. Below the word "Medeia" is the phrase "Designed for Clinicians" in a smaller, blue sans-serif font.

  • . AAMI / ANSI / IEC 80601-2-30 Particular requirements for the basic safety and essential performance of automated noninvasive sphygmomanometers 2009 + A1:2013
  • . AAMI / ANSI EC57 - Testing and reporting performance results of cardiac rhythm and stsegment measurement algorithms 2012 [Select clauses to support device performance]

Software Verification and Validation Testing

Software verification and validation testing were conducted following the FDA guidance document for software contained in medical devices. The software was considered to be a "moderate" level of concern since a failure or latent flaw could indirectly result in a minor injury to the patient through incorrect or delayed information or through action of the operator.

b2 CLINICAL TESTING:

No Clinical testing was necessary to determine substantial equivalence.

b3 CONCLUSIONS DRAWN FROM TESTING:

Based on information obtained on the predicate device with reference to the design specification, electrical safety / EMC testing and intended use, the VitalScan ANS device was subjected to the same type of testing. The results support the conclusion that the VitalScan ANS device is substantially equivalent to the Predicate devices.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for Medeia. The logo has the word "Medeia" in black font, with the first letter "M" in green and blue. Below the word "Medeia" is the phrase "Designed for Clinicians" in a smaller blue font.

Comparison Table

In section a 3 Predicate Devices section of this Summent, The Critical Care Assessment (CCA) device (K101983) is included as a predicate. The Critical Care Assessment device is a re-labeled ANSHA-QHRV1 device with the only difference being the name of the following comparison table includes only the ANSHA-QHRV1 device for direct comparison as all rows for the CCA device would be identical.

FeatureVitalScan ANSPrimary PredicateSecondary PredicateReference DeviceSubstantial EquivalenceComments
ANSHA-QHRV1ABPI MDTonoport V
510(k) NumberSubject DeviceK083735K172655K012647--
Device nameVitalScan ANSANSHA-QHRV1ABPI MDTonoport V--
ManufacturerMedeia Ltd.Medeia Ltd.MESIPAR MedizintechnikGmbH--
RegulationNumbers21 CFR 870.278021 CFR 870.234021 CFR 870.113021 CFR 870.278021 CFR 870.234021 CFR 870.278021 CFR 870.1130CFR numbers of thesubject device includethose of the predicate andreference devices
Product CodesJOM, DPS, DXNJOM, DPSJOMDXNProduct Codes of thesubject device includethose of the predicate andreference devices
VitalScan ANSPrimary PredicateSecondary PredicateReference DeviceSubstantial EquivalenceComments
FeatureANSHA-QHRV1ABPI MDTonoport V
Indications for UseVitalScan ANS is intended for non-invasive measurements of pulsewaveforms by photoelectricplethysmography (PPG), heart rateelectrocardiograph (ECG), HeartVariability measurements (HRV)and blood pressure (NIBP) inresponse to paced respiration andcontrolled testing procedures forphysician assessment of theCardiovascular Systems.VitalScan ANS alsomeasures/calculates a patient'sBrachial and Ankle Blood Pressure,Ankle Brachial Pressure Index(ABPI) and Ankle Brachial Index(ABI) and provides Pulse VolumeRecording (PVR) / volumeplethysmography for assessingdevelopment of peripheral arterialdisease (PAD).The device is intended for use ontransitional adolescents andadults, including those withunilateral lower limb amputation,in medical clinics, healthcarepractices and out-patientdepartments of hospitals.The physician has theresponsibility of interpreting thesignificance of the resulting data.ANSHA-QHRV1 is intended fornon-invasive measurements ofpulse waveforms byphotoelectricplethysmography and heartrate electrocardiograph.The system is intended for useof patients in medical clinics,healthcare practices and inout-patient department ofhospitals.The ABPI MD system isindicated for use on adultsubjects at risk of having ordeveloping peripheral arterialdisease (PAD).The ABPI MD system isintended for the rapidmeasurement of ankle-brachial pressure index (ABPI),or ankle-brachial index (ABI)and pulse volume recording(PVR) / volumeplethysmography in adults.It is suitable for wound careassessment, for assessingsymptomatic PAD, and as ascreening device for PAD. Itmay also be used on patientswith venous or arterial ulcersprior to application ofcompression therapy.The ABPI MD system can beused on patients withunilateral lower limbamputation. The ABPI MDsystem is intended to be usedto spot-check patients. TheABPI MD system providedinformation regarding patientrisk. The physician has theresponsibility of makingproper judgements based onthis information.[Excerpt of IFU] The TonoportV is a compact, lightweight,patient-borne, non-invasiveblood pressure (NIBP) bolterusing the oscillometricmethod. The cuff is borne onthe upper arm and anelectrical pump inside thedevice generates the pressurein the cuff. Tonoport V isintended to be used formeasuring the systolic,diastolic, mean blood pressureand the heart rate of humanbeings for periods up to 30hours. The intended patientpopulations are adults andchildren (but not neonates)with a circumference of theupper arm in the range of 17cm to 42 cm. A measurementwith Tonoport V is combinedwith other measurements andmedical examinations at thepatient, so that a diagnosisabout the patient's healthcondition depends not alonefrom the measurement of theTonoport V.There are no differences in theintended use of the devices.The indications for use areclarified, but not changed.The measurements made bythe ANSHA-QHRV1 are usedfor the same purposes as theVitalScan ANS device - analysisof HRV, PWV, and PVR. TheVitalScan ANS uses the samealgorithms as the ANSHA-QHRV1 for PPG and ECGacquisition and analysis.The measurements made bythe ABPI-MD device are usedfor assessing symptomatic PADor for screening for PAD. TheVitalScan ANS makesmeasurements for the samepurposes.The VitalScan ANS uses thesame algorithm as theTonoport V for NIBPmeasurements.The VitalScan ANS andpredicate/reference devicesare used for the adultpopulation in professionalsettings and a medicalprofessional is responsible forinterpreting the data /determining its significancerelated to the patient's health.
Primary PredicateSecondary PredicateReference DeviceSubstantial Equivalence
FeatureVitalScan ANSANSHA-QHRV1ABPI MDTonoport VComments
MeasurementsHRV – by ECG usingelectrodes placed on thechest or on hands. Themonitoring program hasbeen developed according tothe standards andmathematical procedures forshort-term HRV Autonomicnervous system analysis aswell as for performing andevaluating blood pressure,blood flow and autonomicchallenge tests.Test Administration:Cardio respiratory Couplingand Autonomic Balance(Sympathetic andParasympathetic Tone),Autonomic dysregulationStress Analysis. Time: 4.30minutesParasympathetic BreathingReflex.Time: 1 minuteSympatheticVasoconstriction Reflex;Time: 1.30 minutesOrthostatic Test,Cardiovascular Reflex,Cardiovascular FitnessAnalysisTime: 5.00 minutesSystolic, Diastolic and AnkleBrachial pressures using theoscillometric methodHRV – by ECG usingelectrodes placed on thechest or on hands. Themonitoring program hasbeen developed according tothe standards andmathematical procedures forshort-term HRV Autonomicnervous system analysis aswell as for performing andevaluating blood flow andautonomic challenge tests.Test Administration:Cardio respiratory Couplingand Autonomic Balance(Sympathetic andParasympathetic Tone),Stress Analysis. Time: 4.30minutesParasympathetic BreathingReflex.Time: 1 minuteSympatheticVasoconstriction ReflexTime: 1.30 minutesOrthostatic Test,Cardiovascular Reflex,Fitness AnalysisTime: 5.00 minutesAnkle brachial pressure indexusing the oscillometricmethodSystolic blood pressure usingthe oscillometric methodDiastolic blood pressureusing the oscillometricmethodHeart rate using theoscillometric methodMeasuring the systolic,diastolic and mean bloodpressure and the heart rateusing the oscillometricmethodFor HRV / PWV / PVR theVitalScan ANS device isidentical to the ANSHA-QHRV1. The measurementsand test protocol are thesame between the twodevices.For ABPI / ABI / PADassessment the VitalScanANS is the same as theABPI-MD device. Bothdevices use the oscillometricmethod to makedeterminations.
VitalScan ANSPrimary PredicateSecondary PredicateReference DeviceSubstantial Equivalence
FeatureANSHA-QHRV1ABPI MDTonoport VComments
AlarmsNo alarms as part of thedeviceNo alarms as part of thedeviceNo alarms are part of thedeviceUnknownSame - no alarms areprovided for the predicatedevices.
GeneralSpecifications
Dimensions24 x 8.4 x 20 cm110 x 70 x 11 mm25 x 7.3 x 20 cm80 x 27 x 100 mmSimilar – dimensions / massdoesn't impact devicefunction or raise differentquestions of safety oreffectiveness.
Weight1000 grams100 grams600 grams199 grams
DisplayPCPC4.3″ color LCD screen30.0 mm x 10.5 mmSame as ANSHA-QHRV1
Energy source5VDC/1.6 A +USB from laptopUSB from laptopRechargeable lithiumpolymer batteryPowered by two AA sizebatteriesVitalScan ANS uses the USBfor powering the device andan external power supply forpowering the NIBP function
Target populationAdultAdultAdultAdults and children, but notneonatesSame
Where usedClinical environmentClinical environmentClinical environmentIn hospitalSame
PC DatatransmissionUSBUSBUSBUSB or RS-232Same
Temperature andhumidity rangeWorking environment:10 to 40°C, 30 to 75%Relative air humidity, IPX1protection.Transport and storage:-20 to 70°C, up to 90%Relative air humidity.Working environment:10 to 40°C, 30 to 75%Relative air humidity, IPX0protection.Transport and storage:-20 to 70°C, up to 90%Relative air humidity.Working environment:10 to 40°C, 30 to 85%Relative air humidity, IPX0protection.Transport and storage:0 to 60°C, up to 85% Relativeair humidity.Working environment:10 to 40°C, 30 to 75%Relative air humidity, IPX0protection.Transport and storage:-20 to 70°C, up to 90%Relative air humidity.Similar - Change from IPX0to IPX1 does not affectmeasurements. Change wasnecessitated by testing tothe device to ISO 80601-2-61which mandates IPX1 fordevices not used fortransport
Standards60601-160601-1-260601-1-660601-2-2780601-2-3080601-2-6160601-160601-2-27 (limited clauses)60601-160601-1-260601-1-680601-2-3080601-2-30Similar - the VitalScan ANSwas tested to the full suite ofstandards based on thefunctions included. Thistesting does not raisedifferent questions of safetyor effectiveness
Primary PredicateSecondary PredicateReference DeviceSubstantial Equivalence
FeatureVitalScan ANSANSHA-QHRV1ABPI MDTonoport VComments
PHYSIOLOGICALPARAMETERS
ECG
MethodECG lead wires attached todisposable electrodes to theskinECG lead wires attached todisposable electrodes to theskinSame as ANSHA-QHRV1
Resolution24 bit16 bitSimilar. Increased resolutionimproves noise reduction.There is no impact on devicefunction. No differentquestions of safety oreffectiveness are raised.
Input impedance> 20 Mohm> 20 MohmSame as ANSHA-QHRV1
Common moderejection-80dB - 100dB-80dB - 100dBSame as ANSHA-QHRV1
Sampling frequency1000, 500, 200 Hz800, 400 and 200 HzSimilar. For HRV analysis thedefault sampling frequency isset at 200 Hz via software inboth devices. The samplingfrequency options do notimpact device function. Nodifferent questions of safetyor effectiveness are raised.
Channels3 or 12 channels3 channelsSimilar to ANSHA-QHRV1.Added ECG leads does notimpact measurementfunctionality. Both 3 and 12lead cables were evaluatedto 60601-2-27
QRS detectionYes - 99.8%Yes - 99.8%Same as ANSHA-QHRV1
Permanent DisplayNot ProvidedNot ProvidedSame as ANSHA-QHRV1
Applied parts incontact with thepatientECG ElectrodesECG ElectrodesSame as ANSHA-QHRV1
FeatureVitalScan ANSPrimary PredicateANSHA-QHRV1Secondary PredicateABPI MDReference DeviceTonoport VSubstantial EquivalenceComments
HEART RATE
MethodQRS detectionQRS detectionOscillometric methodOscillometric methodSame as ANSHA-QHRV1
Range38 - 250 bpm40 - 200 bpm30 - 199 bpm35 - 240 bpmSimilar - minor changedoesn't impact devicefunction or raise differentquestions of safety oreffectiveness
Accuracy± 2 bpm± 2 bpm± 5% of valueNot SpecifiedSame as ANSHA-QHRV1
BLOOD PRESURE
Measurement typesOscillometric measuringmethod during deflation ofthe cuffOscillometric measuringmethod during deflation ofthe cuffOscillometric measuringmethod during deflation ofthe cuffSame as ABPI-MD
MeasurementrangesSystolic: 60 to 260 mmHgDiastolic: 40 to 220 mmHgBP Pulse rate (HR): 35 to 240min-139 – 242 mmHg Systolic40 – 180 mmHg Diastolic30 to 199 bpm (Pulse)Systolic: 60 to 260 mmHgDiastolic: 40 to 220 mmHgBP Pulse rate (HR): 35 to 240min-1Same as Tonoport V
Limit values ofmeasurementerrorsABPI: ± 0.1Empirical Standard DeviationSystolic 4.6 mmHgDiastolic 4.4 mmHg± 0.1 ABPI± 3 mmHg (BP)± 5% of value (Pulse)Empirical Standard DeviationSystolic 4.6 mmHgDiastolic 4.4 mmHgSame as Tonoport V
Cuffs inflation anddeflationAutomatic inflation using anair pump and deflation usingan electromagnetic valve.Max 300 mmHgAutomatic inflation using anair pump and deflation usingand electromagnetic valveAutomatic inflation using anair pump and deflation usingan electromagnetic valve.Max 300 mmHgSame as Tonoport V
Pulse Volume /PlethysmographyPneumo-plethysmographymethod using the cuffsmeasuring the bloodpressure values:Plethysmography displayedat the inflation and deflationpressurePneumo-plethysmographymethod using the cuffsmeasuring the bloodpressurevalues: Plethysmographydisplayed at the inflationand deflation pressureN/ASame as ABPI-MD
FeatureVitalScan ANSPrimary PredicateANSHA-QHRV1Secondary PredicateABPI MDReference DeviceTonoport VSubstantial Equivalence Comments
Applied parts in contact with the patient2 or 4 cuffs, tubes and bladders3 cuffs, tubes and bladders1 cuff, tube and bladderSimilar – The VitalScan ANS allows the user to place a cuff on each arm and each leg simultaneously rather than having to move a cuff from one limb to another.
OXYGEN SATURATION
MethodPhoto plethysmogram on finger.Infrared: 910 nm @ 1.2 mW maximum averageRed: 660 nm @ 0.8 mW maximum averagePhoto plethysmogram on finger.Infrared: 910 nm @ 1.2 mW maximum averageRed: 660 nm @ 0.8 mW maximum averageSame as ANSHA-QHRV1. Only difference between subject and predicate device is that the VitalScan ANS device was fully tested to 80601-2-61 for safety and performance
Range70-100%70-100%
Accuracy± 2 digits (from 70-100%)± 2 digits (from 70-100%)

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image contains the logo for Medeia. The logo has the word "Medeia" in bold black font, with the "M" in green and blue. Below the word "Medeia" is the phrase "Designed for Clinicians" in a smaller blue font.

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for Medeia. The logo has the word "Medeia" in a sans-serif font, with the first letter "M" in green and blue, and the rest of the letters in black. Below the word "Medeia" is the phrase "Designed for Clinicians" in a smaller, sans-serif font.

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image contains the logo for Medeia. The logo has a stylized "M" in green and blue, followed by the word "Medeia" in black. Below the word "Medeia" is the phrase "Designed for Clinicians" in blue.

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image contains the logo for Medeia. The logo has a stylized "M" in green and blue, followed by the word "Medeia" in black. Below the word "Medeia" is the phrase "Designed for Clinicians" in blue.

{12}------------------------------------------------

Image /page/12/Picture/0 description: The image contains the logo for Medeia. The logo has a stylized "M" in green and blue, followed by the word "Medeia" in black. Below the word "Medeia" is the phrase "Designed for Clinicians" in blue.

{13}------------------------------------------------

Image /page/13/Picture/0 description: The image contains the logo for Medeia. The logo has a stylized "M" in green and blue, followed by the word "Medeia" in black. Below the word "Medeia" is the phrase "Designed for Clinicians" in blue.

§ 870.2780 Hydraulic, pneumatic, or photoelectric plethysmographs.

(a)
Identification. A hydraulic, pneumatic, or photoelectric plethysmograph is a device used to estimate blood flow in a region of the body using hydraulic, pneumatic, or photoelectric measurement techniques.(b)
Classification. Class II (performance standards).