K Number
K191115
Date Cleared
2019-06-12

(47 days)

Product Code
Regulation Number
892.1550
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ultrasound diagnostic system and probes are designed to obtain ultrasound images and analyze body fluid. The clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Intraoperative, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-raginal, Muscular-Skeletal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric, Trans-esophageal (Cardiac) and Peripheral vessel.

Device Description

The RS85 is a general purpose, mobile, software controlled, diagnostic ultrasound system. Its function is to acquire ultrasound data and to display the data as 2D mode, Color Doppler imaging, Power Doppler imaging (including Directional Power Doppler mode; S-Flow), PW Spectral Doppler mode, CW Spectral Doppler mode, Harmonic imaging, Tissue Doppler imaging, Tissue Doppler Wave, 3D imaging mode (real-time 4D imaging mode), Elastoscan* Mode or as a combination of these modes. The RS85 also gives the operator the ability to measure anatomical structures and offers analysis packages that provide information that may aid in making a diagnosis by competent health care professionals. The RS85 has real time acoustic output display with two basic indices, a mechanical index and a thermal index, which are both automatically displayed.

AI/ML Overview

The provided text is a 510(k) summary for the SAMSUNG MEDISON RS85 Diagnostic Ultrasound System. It mostly focuses on claiming substantial equivalence to a predicate device (RS85 Diagnostic Ultrasound System K182932) by demonstrating that the proposed device has the same intended use, imaging capabilities, technological characteristics, and safety and effectiveness.

Here's an analysis of the acceptance criteria and study information based only on the provided document:

1. A table of acceptance criteria and the reported device performance

The document does not explicitly state acceptance criteria in terms of specific performance metrics or thresholds. Instead, it claims substantial equivalence to its predicate device (RS85 Diagnostic Ultrasound System K182932). The "device performance" reported is that the "proposed RS85 is same to the predicate device with regards to intended use, imaging capabilities, technological characteristics and safety and effectiveness."

The table provided lists comparisons between the proposed RS85 (version V1.05) and its predicate (K182932). This table indicates:

  • Clinical Application: No addition (meaning identical to the predicate).
  • Operation of modes: No addition (meaning identical to the predicate).
  • Applied transducers: No addition (meaning identical to the predicate).
  • SW Features: Improvement of cleared software (Setup&Preference, Patient Management, Measurement, ScanMode(CEUS+), S-Fusion).
  • HW Features: No addition (meaning identical to the predicate).

Since no specific quantitative acceptance criteria or performance numbers are given for the device, a table of acceptance criteria vs. reported performance cannot be fully constructed in the traditional sense of numerical targets. The "acceptance criterion" appears to be equivalence to the predicate device's existing performance, which is implicitly considered acceptable.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document states, "The subject of this premarket submission, RS85, did not require clinical studies to support substantial equivalence." This indicates that no dedicated clinical test set with a specific sample size was used for this 510(k) submission. Instead, the submission relies on non-clinical tests and the established safety and effectiveness of its predicate device through technical comparison.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Since no clinical studies were performed for this submission (as stated in section 11 "Summary of Clinical Tests"), there is no information provided on experts establishing ground truth for a test set. The substantial equivalence claim is based on non-clinical performance and a comparison to a previously cleared device.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

As per item 2 and 3, no clinical test set was used, and therefore, no adjudication method is mentioned.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

The document explicitly states, "The subject of this premarket submission, RS85, did not require clinical studies to support substantial equivalence." Furthermore, the described improvements in "SW Features" (e.g., Setup&Preference, Patient Management, Measurement, ScanMode(CEUS+), S-Fusion) are functional enhancements to the ultrasound system itself, not features implying AI assistance for human image interpretation that would warrant an MRMC study. Therefore, no MRMC comparative effectiveness study was done, and no effect size for human reader improvement with/without AI assistance is provided.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

The device is an ultrasound system with software features, not a standalone algorithm. The "SW Features" listed are improvements to the existing cleared software. There is no indication of a standalone algorithm-only performance study.

7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

As no clinical study was conducted for this submission, no information on the type of ground truth used is provided. The submission relies on non-clinical testing and comparison to the predicate device, K182932, whose ground truth establishment would have been part of its original clearance.

8. The sample size for the training set

The document does not describe any specific "training set" for an algorithm. The software features are described as "improvement of the cleared Software" for a diagnostic ultrasound system. This implies software updates or enhancements to an existing system, rather than the development of a new AI/ML algorithm that would undergo a distinct training phase. Therefore, no sample size for a training set is provided.

9. How the ground truth for the training set was established

As no training set is mentioned in the context of an algorithm, no information is provided on how ground truth for a training set was established.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

SAMSUNG MEDISON CO., LTD. % Ji Yea Lee Regulatory Affairs Specialist 3366, Hanseo-ro, Nam-myeon, Hongcheon-gun, Gangwon-do 25108 REPUBLIC OF KOREA

Re: K191115

Trade/Device Name: RS85 Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: Class II Product Code: IYN, IYO, ITX Dated: May 17, 2019 Received: May 21, 2019

Dear Ji Yea Lee:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

June 12, 2019

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K191115

Device Name RS85 Diagnostic Ultrasound System

Indications for Use (Describe)

The ultrasound diagnostic system and probes are designed to obtain ultrasound images and analyze body fluid. The clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Intraoperative, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-raginal, Muscular-Skeletal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric, Trans-esophageal (Cardiac) and Peripheral vessel.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (7/17)

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

{3}------------------------------------------------

510(k) No.:

Device Name: RS85 Diagnostic Ultrasound System Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)PPPPNote 1Notes 2, 3, 4, 7, 8, 9, 11, 14
Abdominal (See Note 10)PPPPPNote 1Notes 2, 4, 5, 6, 7, 8, 9, 11, 12, 14, 15
Intra-operative (See Note 6)PPPPNote 1Note 7, 9, 11
Intra-operative (Neuro.)PPPPNote 1Note 7, 9, 11
Fetal Imaging& OtherLaparoscopic
PediatricPPPPPNote 1Notes 2, 7, 8, 9, 11
Small Organ (See Note 5)PPPPNote 1Note 2, 5, 6, 7, 8, 9, 11, 12, 14
Neonatal CephalicPPPPNote 1Notes 8, 9, 11
Adult CephalicPPPPPNote 1Notes 7
Trans-rectalPPPPNote 1Notes 2, 7, 8, 9, 11, 12
Trans-vaginalPPPPNote 1Notes 2, 7, 8, 9, 11, 12
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)PPPPNote 1Note 2, 5, 6, 7, 8, 9, 11, 14
Musculo-skel. (Superfic.)PPPPNote 1Note 2, 5, 6, 7, 8, 9, 11, 14
Intra-luminal
Other (See Note 13)PPPPNote 1Notes 2, 7, 8, 9, 12
Cardiac AdultPPPPPNote 1Notes 4, 7
CardiacCardiac PediatricPPPPPNote 1Notes 4, 7
Trans-esophageal (Cardiac)PPPPPNote 1Notes 4, 7
Other (spec.)
PeripheralVesselPeripheral vesselPPPPPNote 1Note 2, 5, 6, 7, 8, 9, 11, 14
Other (spec.)

N= new indication; P= previously cleared by FDA K182932; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+PPI, B+CW, B+C+PW, B+PD+PW, B+PD+PW, B+PPI+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

  • Note 2: Includes imaging for guidance of biopsy
    Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

  • Note 8: 3D imaging
    Note 9: Spatial Compound Imaging

  • Note 10: Includes Renal, Gynecology/Pelvis
    Note 11: Panoramic imaging

Note 12: ElastoScan

Note 13: Includes Urology/Prostate

Note 14: S-Fusion

{4}------------------------------------------------

510(k) No.:

Device Name: CA1-7A for use with RS85

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)PPPPNote 1Notes 2,4,7,8,9,11,14
Abdominal (See Note 10)PPPPNote 1Notes 2,6,7,8,9,11,14,15
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
PediatricPPPPNote 1Notes 2,6,7,8,9,11,14
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)PPPPNote 1Notes 2,6,7,8,9,11,14
Musculo-skel. (Superfic.)
Intra-luminal
Other (See Note 13)PPPPNote 1Notes 2,6,7,8,9,11,14
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vesselPPPPNote 1Notes 2,6,7,8,9,11,14
Other (spec.)

N= new indication; P= previously cleared by FDA K182932; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+PPI, B+CW, B+CW, B+C+PW, B+PD+PW, B+PPH+PW, B+PPH+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

  • Note 2: Includes imaging for guidance of biopsy
    Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: Spatial Compound Imaging

Note 10: Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging

Note 12: ElastoScan

Note 13: Includes Urology/Prostate

Note 14: S-Fusion

{5}------------------------------------------------

510(k) No.:

Device Name: CF4-9 for use with RS85

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
GeneralSpecificBMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
(Track I only)(Tracks I & III)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)
Abdominal (See Note 10)PPPPNote 1Notes 7, 8, 9, 11
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
PediatricPPPPNote 1Notes 7, 8, 9, 11
Small Organ (See Note 5)
Neonatal CephalicPPPPNote 1Notes 7, 8, 9, 11
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (See Note 13)
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralPeripheral vesselPPPPNote 1Notes 8, 9, 11
VesselOther (spec.)

N= new indication; P= previously cleared by FDA K182932; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+PPI, B+CW, B+C+PW, B+PD+PW, B+PD+PW, B+PPI+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

  • Note 2: Includes imaging for guidance of biopsy
    Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: Spatial Compound Imaging

Note 10: Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging

Note 12: ElastoScan

Note 13: Includes Urology/Prostate

Note 14: S-Fusion Note 15: S-Shearwave

{6}------------------------------------------------

510(k) No.:

Device Name: CA2-8A for use with RS85

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)PPPPPNote 1Notes 2, 4, 7, 8, 9, 11
Abdominal (See Note 10)PPPPPNote 1Notes 2, 4, 7, 8, 9, 11
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (See Note 13)
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vessel
Other (spec.)

N= new indication; P= previously cleared by FDA K182932; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+PPI, B+CW, B+C+PW, B+PD+PW, B+PPI+PW, B+PY+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

  • Note 2: Includes imaging for guidance of biopsy
    Note 3: Includes infertility monitoring of follicle development

  • Note 4: Color M-mode
    Note 5: For example: thyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

  • Note 6: Abdominal organs and peripheral vessel

  • Note 7: Tissue Harmonic Imaging (THI)

  • Note 8: 3D imaging

  • Note 9: MultiVision (Spatial Compound Imaging)

  • Note 10: Includes Renal, Gynecology/Pelvis

  • Note 11: Panoramic imaging

  • Note 12: ElastoScan

  • Note 13: Includes Urology/Prostate

  • Note 14: S-Fusion Note 15: S-Shearwave

{7}------------------------------------------------

510(k) No.:

Device Name: CA3-10A for use with RS85

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)PPPPPNote 1Notes 2, 3, 7, 8, 9, 11
Abdominal (See Note 10)PPPPPNote 1Notes 2, 7, 8, 9, 11
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal ImagingLaparoscopic
&OtherPediatricPPPPPNote 1Notes 2, 7, 8, 9, 11
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)PPPPPNote 1Notes 2, 7, 8, 9, 11
Musculo-skel. (Superfic.)PPPPPNote 1Notes 2, 7, 8, 9, 11
Intra-luminal
Other (See Note 13)
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralPeripheral vesselPPPPPNote 1Notes 2, 7, 8, 9, 11
VesselOther (spec.)

N= new indication; P= previously cleared by FDA K182932; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+PPI, B+CV, B+C+PW, B+C+PW, B+DPD+PW, B+PPH+PW, B+TD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

  • Note 2: Includes imaging for guidance of biopsy
    Note 3: Includes infertility monitoring of follicle development

  • Note 4: Color M-mode
    Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

  • Note 6: Abdominal organs and peripheral vessel

  • Note 7: Tissue Harmonic Imaging (THI)

  • Note 8: 3D imaging

  • Note 9: MultiVision (Spatial Compound Imaging)

  • Note 10: Includes Renal, Gynecology/Pelvis

  • Note 11: Panoramic imaging

  • Note 12: ElastoScan

  • Note 13: Includes Urology/Prostate

  • Note 14: S-Fusion Note 15: S-Shearwave

Concurrence of Center for Devices and Radiological Health (CDRH) Prescription Use (Per 21 CFR 801.109)

{8}------------------------------------------------

510(k) No.:

Device Name: E3-12A for use with RS85

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)PPPPPNote 1Notes 2, 7, 8, 9
Abdominal (See Note 10)PPPPPNote 1Notes 2, 7, 8, 9, 12
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectalPPPPPNote 1Notes 2, 7, 8, 9, 12
Trans-vaginalPPPPPNote 1Notes 2, 7, 8, 9, 12
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (See Note 13)PPPPPNote 1Notes 2, 7, 8, 9, 12
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vessel
Other (spec.)

N= new indication; P= previously cleared by FDA K182932; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+PPL, B+CW, B+CW, B+C+PW, B+PD+PW, B+PPH+PW, B+PPH+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: Spatial Compound Imaging

Note 10: Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging

Note 12: ElastoScan

Note 13: Includes Urology/Prostate

Note 14: S-Fusion Note 15: S-Shearwave

Note TSTB Shear Wave

Note 2: Includes imaging for guidance of biopsy

{9}------------------------------------------------

510(k) No.:

Device Name: L3-12A for use with RS85

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal Imaging& OtherFetal/Obstetrics (See Note 3)
Abdominal (See Note 10)PPPPNote 1Note 2, 5, 6, 7, 8, 9, 11, 14
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Laparoscopic
Pediatric
Small Organ (See Note 5)PPPPNote 1Note 2, 5, 6, 7, 8, 9, 11, 12, 14
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)PPPPNote 1Note 2, 5, 6, 7, 8, 9, 11, 14
Musculo-skel. (Superfic.)PPPPNote 1Note 2, 5, 6, 7, 8, 9, 11, 14
Intra-luminal
Other (See Note 13)
CardiacCardiac Adult
Cardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vesselPPPPNote 1Note 2, 5, 6, 7, 8, 9, 11, 14
Other (spec.)

N= new indication; P= previously cleared by FDA K182932; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+PPI, B+CW, B+CW, B+C+PW, B+PD+PW, B+PPH+PW, B+PPH+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

  • Note 2: Includes imaging for guidance of biopsy
    Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: Spatial Compound Imaging

Note 10: Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging

Note 12: ElastoScan

Note 13: Includes Urology/Prostate

Note 14: S-Fusion

{10}------------------------------------------------

510(k) No.:

Device Name: EA2-11B for use with RS85

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)PPPPPNote 1Note 2, 7, 8, 9
Abdominal (See Note 12)PPPPPNote 1Note 2, 6, 7, 8, 9, 12
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectalPPPPPNote 1Note 2, 7, 8, 9, 12
Trans-vaginalPPPPPNote 1Note 2, 7, 8, 9, 12
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (See Note 13)PPPPPNote 1Note 2, 7, 8, 9, 12
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vessel
Other (spec.)

N= new indication; P= previously cleared by FDA K182932; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+TD, B+CW, B+C+PW, B+C+PW, B+PD+PW, B+PPH+PW, B+TD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

  • Note 2: Includes imaging for guidance of biopsy
    Note 3: Includes infertility monitoring of follicle development

  • Note 4: Color M-mode
    Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

  • Note 7: Tissue Harmonic Imaging (THI)
  • Note 8: 3D imaging
  • Note 9: MultiVision (Spatial Compound Imaging)
  • Note 10: Includes Renal, Gynecology/Pelvis
  • Note 11: Panoramic imaging
  • Note 12: ElastoScan
  • Note 13: Includes Urology/Prostate
  • Note 14: S-Fusion
  • Note 15: S-Shearwave

{11}------------------------------------------------

510(k) No.:

Device Name: LA2-9A for use with RS85

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)
Abdominal (See Note 10)PPPPNote 1Note 2, 5, 6, 7, 8, 9, 11
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (See Note 5)PPPPNote 1Note 2, 5, 6, 7, 8, 9, 11,12,15
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)PPPPNote 1Note 2, 5, 6, 7, 8, 9, 11,12,15
Musculo-skel. (Superfic.)PPPPNote 1Note 2, 5, 6, 7, 8, 9, 11,12,15
Intra-luminal
Other (See Note 13)
CardiacCardiac Adult
Cardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vesselPPPPNote 1Note 2, 5, 6, 7, 8, 9, 11
Other (spec.)

N= new indication; P= previously cleared by FDA K182932; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+PPI, B+CW, B+C+PW, B+PD+PW, B+PD+PW, B+PPI+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

  • Note 2: Includes imaging for guidance of biopsy
    Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: Spatial Compound Imaging

Note 10: Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging

Note 12: ElastoScan

Note 13: Includes Urology/Prostate

Note 14: S-Fusion

{12}------------------------------------------------

510(k) No.:

Device Name: LA3-16A for use with RS85

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)
Abdominal (See Note 10)
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (See Note 5)PPPPPNote 1Note 2, 5, 6, 7, 9, 11, 12
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)PPPPPNote 1Note 2, 5, 6, 7, 9, 11
Musculo-skel. (Superfic.)PPPPPNote 1Note 2, 5, 6, 7, 9, 11
Intra-luminal
Other (See Note 13)
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vesselPPPPPNote 1Note 2, 5, 6, 7, 9, 11
Other (spec.)

N= new indication; P= previously cleared by FDA K182932; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+PPI, B+CV, B+C+PW, B+C+PW, B+DPD+PW, B+PPH+PW, B+TD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

  • Note 2: Includes imaging for guidance of biopsy
    Note 3: Includes infertility monitoring of follicle development

  • Note 4: Color M-mode
    Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

  • Note 6: Abdominal organs and peripheral vessel
    Note 7: Tissue Harmonic Imaging (THI)

  • Note 8: 3D imaging

  • Note 9: MultiVision (Spatial Compound Imaging)

  • Note 10: Includes Renal, Gynecology/Pelvis

  • Note 11: Panoramic imaging

  • Note 12: ElastoScan

  • Note 13: Includes Urology/Prostate

  • Note 14: S-Fusion Note 15: S-Shearwave

Concurrence of Center for Devices and Radiological Health (CDRH) Prescription Use (Per 21 CFR 801.109)

{13}------------------------------------------------

510(k) No.:

Device Name: LA3-16AI for use with RS85

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)
Abdominal (See Note 10)
Intra-operative (See Note 6)PPPPNote 1Note 7, 9, 11
Intra-operative (Neuro.)PPPPNote 1Note 7, 9, 11
Fetal ImagingLaparoscopic
&OtherPediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)PPPPNote 1Note 7, 9, 11
Musculo-skel. (Superfic.)PPPPNote 1Note 7, 9, 11
Intra-luminal
Other (See Note 13)
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralPeripheral vesselPPPPNote 1Note 7, 9, 11
VesselOther (spec.)

N= new indication; P= previously cleared by FDA K182932; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+PPI, B+CV, B+C+PW, B+C+PW, B+DPD+PW, B+PPH+PW, B+TD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

  • Note 2: Includes imaging for guidance of biopsy
    Note 3: Includes infertility monitoring of follicle development

  • Note 4: Color M-mode
    Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

  • Note 6: Abdominal organs and peripheral vessel

  • Note 7: Tissue Harmonic Imaging (THI)

  • Note 8: 3D imaging

  • Note 9: MultiVision (Spatial Compound Imaging)

  • Note 10: Includes Renal, Gynecology/Pelvis

  • Note 11: Panoramic imaging

  • Note 12: ElastoScan

  • Note 13: Includes Urology/Prostate

  • Note 14: S-Fusion Note 15: S-Shearwave

{14}------------------------------------------------

510(k) No.:

Device Name: LA4-18B for use with RS85

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)
Abdominal (See Note 10)
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (See Note 5)PPPPPNote 1Note 2, 5, 6, 7, 8, 9, 11, 12, 14
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)PPPPPNote 1Note 2, 5, 6, 7, 8, 9, 11, 14
Musculo-skel. (Superfic.)PPPPPNote 1Note 2, 5, 6, 7, 8, 9, 11, 14
Intra-luminal
Other (See Note 13)
CardiacCardiac Adult
Cardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vesselPPPPPNote 1Note 2, 5, 6, 7, 8, 9, 11, 14
Other (spec.)

N= new indication; P= previously cleared by FDA K182932; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+PPI, B+CV, B+C+PW, B+C+PW, B+DPD+PW, B+PPH+PW, B+TD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

  • Note 2: Includes imaging for guidance of biopsy
    Note 3: Includes infertility monitoring of follicle development

  • Note 4: Color M-mode
    Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

  • Note 6: Abdominal organs and peripheral vessel

  • Note 7: Tissue Harmonic Imaging (THI)

  • Note 8: 3D imaging

  • Note 9: MultiVision (Spatial Compound Imaging)

  • Note 10: Includes Renal, Gynecology/Pelvis

  • Note 11: Panoramic imaging

  • Note 12: ElastoScan

  • Note 13: Includes Urology/Prostate

  • Note 14: S-Fusion Note 15: S-Shearwave

{15}------------------------------------------------

510(k) No.:

Device Name: LM4-15B for use with RS85

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)
Abdominal (See Note 10)PPPPNote 1Note 2, 5, 6, 7 , 8, 9, 11
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal ImagingLaparoscopic
&OtherPediatric
Small Organ (See Note 5)PPPPNote 1Note 2, 5, 6, 7, 8, 9, 11, 12
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)PPPPNote 1Note 2, 5, 6, 7, 8, 9, 11
Musculo-skel. (Superfic.)PPPPNote 1Note 2, 5, 6, 7, 8, 9, 11
Intra-luminal
Other (See Note 13)
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralPeripheral vesselPPPPNote 1Note 2, 5, 6, 7, 8, 9, 11
VesselOther (spec.)

N= new indication; P= previously cleared by FDA K182932; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+PPI, B+CV, B+C+PW, B+C+PW, B+DPD+PW, B+PPH+PW, B+TD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

  • Note 2: Includes imaging for guidance of biopsy
    Note 3: Includes infertility monitoring of follicle development

  • Note 4: Color M-mode
    Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

  • Note 6: Abdominal organs and peripheral vessel

  • Note 7: Tissue Harmonic Imaging (THI)

  • Note 8: 3D imaging

  • Note 9: MultiVision (Spatial Compound Imaging)

  • Note 10: Includes Renal, Gynecology/Pelvis

  • Note 11: Panoramic imaging

  • Note 12: ElastoScan

  • Note 13: Includes Urology/Prostate

  • Note 14: S-Fusion Note 15: S-Shearwave

Concurrence of Center for Devices and Radiological Health (CDRH) Prescription Use (Per 21 CFR 801.109)

{16}------------------------------------------------

510(k) No.:

Device Name: PM1-6A for use with RS85

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)
Abdominal (See Note 10)PPPPPNote 1Note 7
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult CephalicPPPPPNote 1Note 7
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (See Note 13)
Cardiac AdultPPPPPNote 1Note 4, 7
CardiacCardiac PediatricPPPPPNote 1Note 4, 7
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vessel
Other (spec.)

N= new indication; P= previously cleared by FDA K182932; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+PPI, B+CW, B+C+PW, B+PD+PW, B+PD+PW, B+PPI+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

  • Note 2: Includes imaging for guidance of biopsy
    Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: Spatial Compound Imaging

Note 10: Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging

Note 12: ElastoScan

Note 13: Includes Urology/Prostate

Note 14: S-Fusion Note 15: S-Shearwave

{17}------------------------------------------------

510(k) No.:

Device Name: PA3-8B for use with RS85

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)
Abdominal (See Note 10)PPPPPNote 1Note 7
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
PediatricPPPPPNote 1Note 7
Small Organ (See Note 5)
Neonatal CephalicPPPPPNote 1Note 7
Adult CephalicPPPPPNote 1Note 7
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (See Note 13)
Cardiac AdultPPPPPNote 1Note 4, 7
CardiacCardiac PediatricPPPPPNote 1Note 4, 7
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vessel
Other (spec.)

N= new indication: P= previously cleared by FDA K182932; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+TD, B+CW, B+C+PW, B+C+PW, B+PD+PW, B+PPH+PW, B+TD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

  • Note 2: Includes imaging for guidance of biopsy
    Note 3: Includes infertility monitoring of follicle development

  • Note 4: Color M-mode
    Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

  • Note 7: Tissue Harmonic Imaging (THI)
  • Note 8: 3D imaging
  • Note 9: MultiVision (Spatial Compound Imaging)
  • Note 10: Includes Renal, Gynecology/Pelvis
  • Note 11: Panoramic imaging
  • Note 12: ElastoScan
  • Note 13: Includes Urology/Prostate
  • Note 14: S-Fusion
  • Note 15: S-Shearwave

{18}------------------------------------------------

510(k) No.:

Device Name: PA4-12B for use with RS85

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)
Abdominal (See Note 10)PPPPPNote 1Note 7
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
PediatricPPPPPNote 1Note 7
Small Organ (See Note 5)
Neonatal CephalicPPPPPNote 1Note 7
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (See Note 13)
Cardiac AdultPPPPPNote 1Note 4, 7
CardiacCardiac PediatricPPPPPNote 1Note 4, 7
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vessel
Other (spec.)

N= new indication; P= previously cleared by FDA K182932; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+PPI, B+CV, B+C+PW, B+C+PW, B+PD+PW, B+PPH+PW, B+C+M, Dual/Ouad. B+C+CW. B+PD+CW. B+E. B+B/C

  • Note 2: Includes imaging for guidance of biopsy
    Note 3: Includes infertility monitoring of follicle development

  • Note 4: Color M-mode
    Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

  • Note 6: Abdominal organs and peripheral vessel

  • Note 8: 3D imaging

  • Note 9: MultiVision (Spatial Compound Imaging)

  • Note 10: Includes Renal, Gynecology/Pelvis

  • Note 11: Panoramic imaging

  • Note 12: ElastoScan

  • Note 13: Includes Urology/Prostate

  • Note 14: S-Fusion

  • Note 15: S-Shearwave

Note 7: Tissue Harmonic Imaging (THI)

{19}------------------------------------------------

510(k) No.:

Device Name: MMPT3-7 for use with RS85

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal Imaging& OtherFetal/Obstetrics (See Note 3)
Abdominal (See Note 12)
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Laparoscopic
Pediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (spec.) (See Note 13)
CardiacCardiac Adult
Cardiac Pediatric
Trans-esophageal (Cardiac)PPPPPNote 1Note 4, 7
Other (spec.)
PeripheralVesselPeripheral vessel
Other (spec.)

N= new indication; P= previously cleared by FDA K182932; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+PPI, B+TD, B+C+PW, B+C+PW, B+DPD+PW, B+PPH+PW, B+TD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

  • Note 2: Includes imaging for guidance of biopsy
    Note 3: Includes infertility monitoring of follicle development

  • Note 4: Color M-mode
    Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

  • Note 6: Abdominal organs and peripheral vessel

  • Note 7: Tissue Harmonic Imaging (THI)

  • Note 8: 3D imaging

  • Note 9: MultiVision (Spatial Compound Imaging)

  • Note 10: Includes Renal, Gynecology/Pelvis

  • Note 11: Panoramic imaging

  • Note 12: ElastoScan

  • Note 13: Includes Urology/Prostate

  • Note 14: S-Fusion Note 15: S-Shearwave

{20}------------------------------------------------

510(k) No.:

Device Name: CV1-8A for use with RS85

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal Imaging& OtherFetal/Obstetrics (See Note 3)PPPPNote 1Note 2, 4, 7, 8, 9, 11
Abdominal (See Note 10)PPPPNote 1Note 2, 4, 7, 8, 9
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Laparoscopic
Pediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (See Note 13)
CardiacCardiac Adult
Cardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vessel
Other (spec.)

N= new indication; P= previously cleared by FDA K182932; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+PPI, B+TD, B+C+PW, B+C+PW, B+DPD+PW, B+PPH+PW, B+TD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

  • Note 2: Includes imaging for guidance of biopsy
    Note 3: Includes infertility monitoring of follicle development

  • Note 4: Color M-mode
    Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

  • Note 6: Abdominal organs and peripheral vessel

  • Note 7: Tissue Harmonic Imaging (THI)

  • Note 8: 3D imaging

  • Note 9: MultiVision (Spatial Compound Imaging)

  • Note 10: Includes Renal, Gynecology/Pelvis

  • Note 11: Panoramic imaging

  • Note 12: ElastoScan

  • Note 13: Includes Urology/Prostate

  • Note 14: S-Fusion Note 15: S-Shearwave

{21}------------------------------------------------

510(k) No.:

Device Name: V5-9 for use with RS85

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal Imaging& OtherFetal/Obstetrics (See Note 3)PPPPNote 1Note 2, 7, 8, 9
Abdominal (See Note 10)PPPPNote 1Note 2, 7, 8, 9, 12
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Laparoscopic
Pediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectalPPPPNote 1Note 2, 7, 8, 9, 12
Trans-vaginalPPPPNote 1Note 2, 7, 8, 9, 12
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (See Note 13)PPPPNote 1Note 2, 7, 8, 9, 12
CardiacCardiac Adult
Cardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vessel
Other (spec.)

N= new indication; P= previously cleared by FDA K182932; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+PPI, B+CW, B+C+PW, B+PD+PW, B+PD+PW, B+PPI+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

  • Note 2: Includes imaging for guidance of biopsy
    Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: Spatial Compound Imaging

Note 10: Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging

Note 12: ElastoScan

Note 13: Includes Urology/Prostate

Note 14: S-Fusion

{22}------------------------------------------------

510(k) No.:

Device Name: LV3-14A for use with RS85

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics ( See Note 3 )
Abdominal ( See Note 10 )
Intra-operative ( See Note 6 )
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ ( See Note 5 )PPPPNote 1Note 2, 5, 6, 7, 8, 9, 11
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)PPPPNote 1Note 2, 5, 6, 7, 8, 9, 11
Musculo-skel. (Superfic.)PPPPNote 1Note 2, 5, 6, 7, 8, 9, 11
Intra-luminal
Other ( See Note 13 )
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vesselPPPPNote 1Note 2, 5, 6, 7, 8, 9, 11
Other (spec.)

N= new indication; P= previously cleared by FDA K182932; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+PPI, B+CW, B+C+PW, B+PD+PW, B+PD+PW, B+PPI+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

  • Note 2: Includes imaging for guidance of biopsy
    Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

  • Note 8: 3D imaging
    Note 9: Spatial Compound Imaging

Note 10: Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging

Note 12: ElastoScan

Note 13: Includes Urology/Prostate

Note 14: S-Fusion

{23}------------------------------------------------

510(k) No.:

Device Name: CW6.0 for use with RS85

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)
Abdominal (See Note 10)
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging&OtherLaparoscopic
Pediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (See Note 13)
Cardiac AdultP
CardiacCardiac PediatricP
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralPeripheral vesselP
VesselOther (spec.)

N= new indication; P= previously cleared by FDA K182932; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+PPI, B+CW, B+C+PW, B+PD+PW, B+PD+PW, B+PPI+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

  • Note 2: Includes imaging for guidance of biopsy
    Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: Spatial Compound Imaging

Note 10: Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging

Note 12: ElastoScan

Note 13: Includes Urology/Prostate

Note 14: S-Fusion

{24}------------------------------------------------

510(k) No.:

Device Name: DP2B for use with RS85

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal Imaging& OtherFetal/Obstetrics (See Note 3)
Abdominal (See Note 10)
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Laparoscopic
Pediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (See Note 13)
CardiacCardiac AdultP
Cardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vessel
Other (spec.)

N= new indication; P= previously cleared by FDA K182932; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+PPI, B+TD, B+C+PW, B+C+PW, B+DPD+PW, B+PPH+PW, B+TD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

  • Note 2: Includes imaging for guidance of biopsy
    Note 3: Includes infertility monitoring of follicle development

  • Note 4: Color M-mode
    Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

  • Note 6: Abdominal organs and peripheral vessel

  • Note 7: Tissue Harmonic Imaging (THI)

  • Note 8: 3D imaging

  • Note 9: MultiVision (Spatial Compound Imaging)

  • Note 10: Includes Renal, Gynecology/Pelvis

  • Note 11: Panoramic imaging

  • Note 12: ElastoScan

  • Note 13: Includes Urology/Prostate

  • Note 14: S-Fusion Note 15: S-Shearwave

{25}------------------------------------------------

510(k) No.:

Device Name: DP8B for use with RS85

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal Imaging&OtherFetal/Obstetrics (See Note 3)
Abdominal (See Note 12)
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Laparoscopic
Pediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (See Note 13)
Cardiac AdultP
CardiacCardiac PediatricP
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vesselP
Other (spec.)

N= new indication: P= previously cleared by FDA K182932; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+TD, B+CW, B+C+PW, B+C+PW, B+PD+PW, B+PPH+PW, B+TD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

  • Note 2: Includes imaging for guidance of biopsy
    Note 3: Includes infertility monitoring of follicle development

  • Note 4: Color M-mode
    Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

  • Note 6: Abdominal organs and peripheral vessel
    Note 7: Tissue Harmonic Imaging (THI)

  • Note 8: 3D imaging

  • Note 9: MultiVision (Spatial Compound Imaging)

  • Note 10: Includes Renal, Gynecology/Pelvis

  • Note 11: Panoramic imaging

  • Note 12: ElastoScan

  • Note 13: Includes Urology/Prostate

  • Note 14: S-Fusion

  • Note 15: S-Shearwave

{26}------------------------------------------------

510(k) No.:

Device Name: EV3-10B for use with RS85

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)PPPPPNote 1Note 2, 7, 8, 9, 11
Abdominal (See Note 10)PPPPPNote 1Note 2, 7, 8, 9, 11, 12
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal ImagingLaparoscopic
&OtherPediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectalPPPPPNote 1Note 2, 7, 8, 9, 11, 12
Trans-vaginalPPPPPNote 1Note 2, 7, 8, 9, 11, 12
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (See Note 13)PPPPPNote 1Note 2, 7, 8, 9, 11, 12
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralPeripheral vessel
VesselOther (spec.)

N= new indication; P= previously cleared by FDA K182932; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+PPI, B+TD, B+C+PW, B+C+PW, B+DPD+PW, B+PPH+PW, B+TD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

  • Note 2: Includes imaging for guidance of biopsy
    Note 3: Includes infertility monitoring of follicle development

  • Note 4: Color M-mode
    Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

  • Note 6: Abdominal organs and peripheral vessel
    Note 7: Tissue Harmonic Imaging (THI)

  • Note 8: 3D imaging
    Note 9: MultiVision (Spatial Compound Imaging)

  • Note 10: Includes Renal, Gynecology/Pelvis

  • Note 11: Panoramic imaging

  • Note 12: ElastoScan

  • Note 13: Includes Urology/Prostate

Note 14: S-Fusion Note 15: S-Shearwave

{27}------------------------------------------------

510(K) Summary: K191115

In accordance with 21 CFR 807.92 the following summary of information is provided:

  • Date Prepared April 25, 2019 1.
    1. Manufacturer SAMSUNG MEDISON CO., LTD. 3366, Hanseo-ro, Nam-myeon, Hongcheon-gun, Gangwon-do 25108, REPUBLIC OF KOREA
  • Primary Contact Person 3. Ji Yea Lee Regulatory Affairs Specialist Phone: +82.2.2.2194.1594 Fax: +82. 31.8017.9576 Email: jiyea722.lee@samsungmedison.com

Secondary Contact Person Scully Kim Regulatory Affairs Specialist Phone: +82.2.2194.1312 Fax: +82. 31.8017.9576 Email: scully.kim@samsungmedison.com

    1. Proposed Device
    • Common/Usual Name: Diagnostic Ultrasound System and Accessories
    • Proprietary Name: RS85 Diagnostic Ultrasound System
    • Common Name: Diagnostic Ultrasound System
    • Classification Names: system, imaging, pulsed doppler, ultrasonic
    • Product Code: IYN, IYO, ITX
    • Regulation: 892.1550, 892.1560, 892.1570
    1. Predicate Device RS85 Diagnostic Ultrasound System (K182932)
    1. Device Description

The RS85 is a general purpose, mobile, software controlled, diagnostic ultrasound system. Its function is to acquire ultrasound data and to display the data as 2D mode, Color Doppler imaging, Power Doppler imaging (including Directional Power Doppler mode; S-Flow), PW Spectral Doppler mode, CW Spectral Doppler mode, Harmonic imaging, Tissue Doppler imaging, Tissue Doppler Wave, 3D imaging mode (real-time 4D imaging mode), Elastoscan* Mode or as a combination of these modes. The RS85 also gives the operator the ability to measure anatomical structures and offers analysis packages that provide information that may aid in making a diagnosis by competent health care professionals. The RS85 has real time acoustic output display with two basic indices, a mechanical index and a thermal index, which are both automatically displayed.

    1. Indications for Use
      The ultrasound diagnostic system and probes are designed to obtain ultrasound images and analyze body

{28}------------------------------------------------

fluids.

The clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Intraoperative, Pediatric, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-rectal, Trans-vaginal, Muscular-Skeletal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric, Trans-esophageal (Cardiac) and Peripheral vessel.

The Indications for use of the proposed device is the same as predicates.

    1. Technology
      The RS85 employs the same fundamental scientific technology as its predicate device(s).
    1. Determination of Substantial Equivalence
      The proposed RS85 is same to the predicate device with regards to intended use, imaging capabilities, technological characteristics and safety and effectiveness.

A comparison of the proposed device to the currently marketed predicates is provided in the table below.

RS85Addition of V1.05Remarks
Clinicalapplication• No addition
Operation ofmodes• No addition
Appliedtransducers• No addition
SW Features• Improvement of the cleared Software:Setup&Preference, PatientManagement, Measurement,ScanMode(CEUS+), S-Fusion
HW Features• No addition

{29}------------------------------------------------

Device name(510(k) clearance)RS85(Under Review)RS85(Κ182932)Predicate
ManufacturerSamsung Medison, Co.,LTDSamsung Medison, Co.,LTD
ClinicalApplication- Fetal/Obstetrics- Abdominal- Gynecology- Intra-operative- Pediatric- Small Organ, Neonatal Cephalic- Adult Cephalic- Trans-rectal- Trans-vaginal- Trans-esoph. (non-Cardiac)- Muscular-Skeletal (Conventional,Superficial)- Urology- Cardiac Adult- Cardiac Pediatric- Trans-esophageal (Cardiac)- Peripheral vessel- Fetal/Obstetrics- Abdominal- Gynecology- Intra-operative- Pediatric- Small Organ, Neonatal Cephalic- Adult Cephalic- Trans-rectal- Trans-vaginal- Trans-esoph. (non-Cardiac)- Muscular-Skeletal (Conventional,Superficial)- Urology- Cardiac Adult- Cardiac Pediatric- Trans-esophageal (Cardiac)- Peripheral vessel
Scanhead Types- Linear Array- Curved Linear Array- Endocavity- Phased Array- Static Probes- Linear Array- Curved Linear Array- Endocavity- Phased Array- Static Probes
ScanheadFrequency1.0 ~ 20.0 MHz1.0 ~ 20.0 MHz
Acoustic OutputDisplay & FDALimits- Display Feature for Higher Output–Track3- MI Output DisplayTI Output Display- Display Feature for Higher Output–Track3- MI Output DisplayTI Output Display
Modes ofOperation- B-mode- M-mode- Pulsed wave (PW) Doppler- Continuous wave (CW) Doppler- Color Doppler- Power Amplitude Doppler- Tissue Harmonic Imaging- 3D/4D imaging mode- Combined modes- B-mode- M-mode- Pulsed wave (PW) Doppler- Continuous wave (CW) Doppler- Color Doppler- Power Amplitude Doppler- Tissue Harmonic Imaging- 3D/4D imaging mode- Combined modes
#TransmitChannels192192
#Receive Channels192192
- LCD Monitor (LED Backlight unit)- 256 gray shades on monitor- 100-240VAC, 1100VA, 50/60Hz- LCD Monitor (LED Backlight unit)- 256 gray shades on monitor- 100-240VAC, 1100VA, 50/60Hz
Product SafetyCertification- IEC 60601-1- UL 60601-1- CSA C22.2 No.601.1- IEC 60601-2-37- IEC 60601-1- UL 60601-1- CSA C22.2 No.601.1- IEC 60601-2-37
EMC ComplianceIEC 60601-1-2IEC 60601-1-2
Acoustic OutputDisplay StandardTrack 3Track 3
BiocompatibilityComplianceISO10993-1ISO10993-1
SoftwareFunctionality- Q Scan- ClearVision- MultiVision- Panoramic- Needle Mate- Auto IMT+- Strain+- Stress Echo- Elastoscan- E-Thyroid- E-Breast- E-Strain- S-Shearwave- S-Detect for Breast- S-Detect for Thyroid- HQ Vision- S-Fusion- Clear Track- Virtual Track- S-3D ArterialAnalysis- ADVR- 3D Imaging- (Volume Data Acquisition)- 3D Imaging presentation- 3D Cine/4D Cine- 3D Rendering MPR- 3D XI MSV/Oblique View- Volume CT- 3D MagiCut- Volume Calculation- (VOCAL, XI VOCAL)- XI STIC- HDVI- Realistic Vue- Natural Vue- Q Scan- ClearVision- MultiVision- Panoramic- Needle Mate- Auto IMT+- Strain+- Stress Echo- Elastoscan- E-Thyroid- E-Breast- E-Strain- S-Shearwave- S-Detect for Breast- S-Detect for Thyroid- HQ Vision- S-Fusion- Clear Track- Virtual Track- S-3D ArterialAnalysis- ADVR- 3D Imaging- (Volume Data Acquisition)- 3D Imaging presentation- 3D Cine/4D Cine- 3D Rendering MPR- 3D XI MSV/Oblique View- Volume CT- 3D MagiCut- Volume Calculation- (VOCAL, XI VOCAL)- XI STIC- HDVI- Realistic Vue- Natural Vue
- CEUS+- CEUS+
- MV-Flow- MV-Flow
- S-Shearwave Imaging- S-Shearwave Imaging
Transducer- CA1-7A- CA1-7A
- CF4-9- CF4-9
- CA2-8A- CA2-8A
- CA3-10A- CA3-10A
- E3-12A- E3-12A
- L3-12A- L3-12A
- EA2-11B- EA2-11B
- LA2-9A- LA2-9A
- LA3-16A- LA3-16A
- LA3-16AI- LA3-16AI
- LA4-18B- LA4-18B
- LM4-15B- LM4-15B
- PM1-6A- PM1-6A
- PA3-8B- PA3-8B
- PA4-12B- PA4-12B
- MMPT3-7- MMPT3-7
- CV1-8A- CV1-8A
- V5-9- V5-9
- LV3-14A- LV3-14A
- CW6.0- CW6.0
- DP2B- DP2B
- DP8B- DP8B
- EV3-10B- EV3-10B
Biopsy Guide- BP-KIT-029- BP-KIT-029
- BP-KIT-030- BP-KIT-030
- BP-KIT-043- BP-KIT-043
- BP-KIT-053- BP-KIT-053
- BP-KIT-054- BP-KIT-054
- BP-KIT-055- BP-KIT-055
- BP-KIT-057- BP-KIT-057
- BP-KIT-058- BP-KIT-058
- BP-KIT-059- BP-KIT-059
- BP-KIT-060- BP-KIT-060
- BP-KIT-067- BP-KIT-067
- BP-KIT-068- BP-KIT-068
- BP-KIT-069- BP-KIT-069
- BP-KIT-071- BP-KIT-071
- BP-KIT-075- BP-KIT-075
- BP-KIT-083- BP-KIT-083
DigitalStorage/TransferStation- Foot Switch- Foot Switch
- ECG (defibrillation proof)- ECG
- Gel Warmer- Gel Warmer
- Clear Track/ Virtual Track- Clear Track/ Virtual Track
System StandSystem Stand
- WiFi Card- WiFi Card

{30}------------------------------------------------

{31}------------------------------------------------

{32}------------------------------------------------

10. Summary of Non-Clinical Test

The proposed device has been evaluated for acoustic output, biocompatibility, cleaning and disinfection effectiveness as well as thermal, electromagnetic and mechanical safety, and has been found to conform with applicable medical device safety standards. The RS85 and its applications comply with voluntary standards.

TestStandards and FDA Guidance
Risk ManagementISO 14971 Second edition 2007 Medical devices - Application of risk management tomedical devices
Electrical SafetyThe RS85 Ultrasound System with defibrillation-proof ECG electrode was evaluated perthe following standards.ANSI AAMI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012Medical Electrical Equipment - Part 1: General Requirements for basic safety andessential performance.
ElectromagneticCompatibilityIEC60601-1-2: 2014(4th Edition) Medical Electrical Equipment -- Part 1-2: GeneralRequirements For Basic Safety And Essential Performance -- Collateral Standard:Electromagnetic Disturbances -- Requirements And Tests
BiocompatibilityISO 10993-1 Fourth edition 2009-10-15 Biological evaluation of medical devices - Part1: Evaluation and testing within a risk management process.
ReprocessingMedical DevicesFDA Guidance: Reprocessing Medical Devices in Health Care Settings: ValidationMethods and Labeling issued on March 17, 2015, revised June 9, 2017.
Software/Firmware-driven FunctionalityAll migrated software functionality was evaluated using the same test criteria as thepredicate for all applicable imaging modes to ensure that migration into a new systemdesign did not compromise image quality with respect to the intended use of eachfeature.Guidance for the Content of Premarket Submissions for Software Contained in MedicalDevices issued on May 11, 2005
Ultrasound SafetyInformation for Manufacturers Seeking Marketing Clearance ofDiagnostic Ultrasound Systems and Transducers issued September 9, 2008IEC60601-2-37:2007 + A1:2015, Particular requirements for the safety of ultrasonicmedical diagnostic and monitoring equipmentNEMA UD 2-2004 (R2009)Acoustic Output Measurement Standard for Diagnostic Ultrasound Equipment Revision3NEMA UD 3-2004 (R2009)Standard for Real-Time Display of Thermal and Mechanical Acoustic Output Indices onDiagnostic Ultrasound Equipment, Revision 2

{33}------------------------------------------------

11. Summary of Clinical Tests

The subject of this premarket submission, RS85, did not require clinical studies to support substantial equivalence.

12. Conclusion

SAMSUNG MEDISON considers the RS85 to be as safe, as effective, and performance is substantially equivalent to the predicate device.

END of 510(K) Summary

§ 892.1550 Ultrasonic pulsed doppler imaging system.

(a)
Identification. An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.