Intended Use

The ultrasound diagnostic system and probes are designed to obtain ultrasound images and analyze body fluid. The clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Intraoperative, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-vaginal, Muscular-Skeletal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric, Trans-esophageal (Cardiac) and Peripheral vessel.

Device Description

The RS85 is a general purpose, mobile, software controlled, diagnostic ultrasound system. Its function is to acquire ultrasound data and to display the data as 2D mode. Color Doppler imaging. Power Doppler imaging (including Directional Power Doppler mode: S-Flow), PW Spectral Doppler mode, CW Spectral Doppler mode, Harmonic imaging, Tissue Doppler imaging, Tissue Doppler Wave, 3D imaging mode (real-time 4D imaging mode), Elastoscan* Mode or as a combination of these modes. The RS85 also gives the operator the ability to measure anatomical structures and offers analysis packages that provide information that may aid in making a diagnosis by competent health care professionals. The RS85 has real time acoustic output display with two basic indices, a mechanical index and a thermal index, which are both automatically displayed.

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and study information:

Response:

Based on the provided 510(k) summary for the SAMSUNG MEDISON RS85 Diagnostic Ultrasound System (K182932), the device's acceptance criteria are primarily related to its substantial equivalence to a predicate device, K173204 (also an RS85 Diagnostic Ultrasound System). The submission did not involve clinical studies to demonstrate substantial equivalence, but rather relied on non-clinical tests and a comparison of technological characteristics.

1. Table of Acceptance Criteria and Reported Device Performance

Given that no clinical studies were performed, the "reported device performance" in the typical sense of metrics like accuracy, sensitivity, or specificity is not available for this submission. Instead, the acceptance criteria are based on equivalence to the predicate device in terms of features and conformity to relevant standards.

Acceptance Criterion (Implicit)Reported Device Performance (as demonstrated)
Intended Use Equivalence: Same clinical applications and modes of operation as predicate.Met: Device has the same stated Indications for Use and Modes of Operation (pages 3-26).
Technological Characteristics Equivalence: Similar scanhead types, frequency range, acoustic output display, modes of operation, transmit/receive channels, system characteristics, software functionality, transducers, and biopsy guides as predicate.Met: Detailed comparison table on page 28-29 shows identical specifications and listed features between the proposed and predicate RS85 device (K173204). New features like "S-Fusion" and "MV Index" are stated to be substantially equivalent to features on other reference devices (page 31).
Safety and Effectiveness Equivalence: Conformity to applicable medical device safety standards.Met: Non-clinical tests evaluated acoustic output, biocompatibility, cleaning/disinfection, thermal, electromagnetic, and mechanical safety, confirming compliance with standards like ISO 14971, ANSI AAMI ES60601-1, IEC60601-1-2, ISO 10993-1, and NEMA UD 2/3 (pages 31-32).

2. Sample Size Used for the Test Set and Data Provenance

No test set of patient data (e.g., images) was used or described for performance evaluation in this 510(k) summary. The submission focuses on hardware and software equivalence to a previously cleared device. Therefore, there is no information on sample size or data provenance (country of origin, retrospective/prospective data) relating to clinical performance.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

Not applicable, as no test set requiring ground truth establishment by experts was used.

4. Adjudication Method for the Test Set

Not applicable, as no test set requiring adjudication was used.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No MRMC comparative effectiveness study was done. The submission explicitly states: "The subject of this premarket submission, RS85, did not require clinical studies to support substantial equivalence." (page 32). There is no mention of AI integration or assessment of human reader improvement.

6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

No standalone performance study of an algorithm was conducted or described, as this submission is for a diagnostic ultrasound system and not an AI algorithm.

7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

Not applicable, as no clinical studies with ground truth were conducted. The "ground truth" for this submission focuses on demonstrating that the device's technical specifications and intended uses are equivalent to a predicate device and that it complies with relevant safety and performance standards.

8. The Sample Size for the Training Set

Not applicable. This submission does not describe the development or training of any machine learning or AI models. It pertains to a diagnostic ultrasound system.

9. How the Ground Truth for the Training Set Was Established

Not applicable. This submission does not describe the development or training of any machine learning or AI models.

{0}------------------------------------------------

February 15, 2019

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services seal on the left and the FDA acronym with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of it. The full name reads "U.S. Food & Drug Administration" with the word "Administration" on the second line.

SAMSUNG MEDISON CO., LTD. % Ji Yea Lee Regulatory Affairs Specialist 3366, Hanseo-ro, Nam-myeon, Hongcheon-gun, Gangwon-do 25108 REPUBLIC OF KOREA

Re: K182932

Trade/Device Name: RS85 Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: Class II Product Code: IYN, IYO, ITX Dated: January 11, 2019 Received: January 15, 2019

Dear Ji Yea Lee:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Michael D'Hara For

Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

510(k) Number (if known) K182932

Device Name RS85 Diagnostic Ultrasound System

Indications for Use (Describe)

The ultrasound diagnostic system and probes are designed to obtain ultrasound images and analyze body fluid. The clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Intraoperative, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-vaginal, Muscular-Skeletal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric, Trans-esophageal (Cardiac) and Peripheral vessel.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) No.:

Device Name: RS85 Diagnostic Ultrasound System

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)PPPPPNote 1Notes 2, 3, 4, 7, 8, 9, 11, 14
Abdominal (See Note 10)PPPPPNote 1Notes 2, 4, 5, 6, 7, 8, 9, 11, 12, 14, 15
Intra-operative (See Note 6)PPPPNote 1Note 7, 9, 11
Intra-operative (Neuro.)PPPPNote 1Note 7, 9, 11
Fetal Imaging& OtherLaparoscopic
PediatricPPPPPNote 1Notes 2, 7, 8, 9, 11
Small Organ (See Note 5)PPPPNote 1Note 2, 5, 6, 7, 8, 9, 11, 12, 14
Neonatal CephalicPPPPNote 1Notes 8, 9, 11
Adult CephalicPPPPPNote 1Notes 7
Trans-rectalPPPPNote 1Notes 2, 7, 8, 9, 11, 12
Trans-vaginalPPPPNote 1Notes 2, 7, 8, 9, 11, 12
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)PPPPNote 1Note 2, 5, 6, 7, 8, 9, 11, 14
Musculo-skel. (Superfic.)PPPPNote 1Note 2, 5, 6, 7, 8, 9, 11, 14
Intra-luminal
Other (See Note 13)PPPPNote 1Notes 2, 7, 8, 9, 12
Cardiac AdultPPPPPNote 1Notes 4, 7
CardiacCardiac PediatricPPPPPNote 1Notes 4, 7
Trans-esophageal (Cardiac)PPPPPNote 1Notes 4, 7
Other (spec.)
PeripheralVesselPeripheral vesselPPPPPNote 1Note 2, 5, 6, 7, 8, 9, 11, 14
Other (spec.)

N= new indication; P= previously cleared by FDA; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+PPI, B+CW, B+C+PW, B+PD+PW, B+DPD+PW, B+PPI+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

  • Note 2: Includes imaging for guidance of biopsy
    Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

  • Note 8: 3D imaging
    Note 9: Spatial Compound Imaging

Note 10: Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging

Note 12: ElastoScan

Note 13: Includes Urology/Prostate

Note 14: S-Fusion

Note 15: S-Shearwave

{4}------------------------------------------------

510(k) No.:

Device Name: CA1-7A for use with RS85

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)PPPPNote 1Notes 2,4,7,8,9,11,14
Abdominal (See Note 10)PPPPNote 1Notes 2,6,7,8,9,11,14,15
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
PediatricPPPPNote 1Notes 2,6,7,8,9,11,14
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)PPPPNote 1Notes 2,6,7,8,9,11,14
Musculo-skel. (Superfic.)
Intra-luminal
Other (See Note 13)PPPPNote 1Notes 2,6,7,8,9,11,14
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vesselPPPPNote 1Notes 2,6,7,8,9,11,14
Other (spec.)

N= new indication; P= previously cleared by FDA K173204; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPL, B+CW, B+CW, B+C+PW, B+PD+PW, B+DPD+PW, B+PPI+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

  • Note 2: Includes imaging for guidance of biopsy
    Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: Spatial Compound Imaging

Note 10: Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging

Note 12: ElastoScan

Note 13: Includes Urology/Prostate

Note 14: S-Fusion Note 15: S-Shearwave

{5}------------------------------------------------

510(k) No.:

Device Name: CF4-9 for use with RS85

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)
Abdominal (See Note 10)PPPPNote 1Notes 7, 8, 9, 11
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
PediatricPPPPNote 1Notes 7, 8, 9, 11
Small Organ (See Note 5)
Neonatal CephalicPPPPNote 1Notes 7, 8, 9, 11
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (See Note 13)
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vesselPPPPNote 1Notes 8, 9, 11
Other (spec.)

N= new indication; P= previously cleared by FDA K173204; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+PPI, B+CW, B+C+PW, B+PD+PW, B+DPD+PW, B+PPI+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

  • Note 2: Includes imaging for guidance of biopsy
    Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: Spatial Compound Imaging

Note 10: Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging

Note 12: ElastoScan

Note 13: Includes Urology/Prostate

Note 14: S-Fusion Note 15: S-Shearwave

{6}------------------------------------------------

510(k) No.:

Device Name: CA2-8A for use with RS85

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)PPPPNote 1Notes 2, 4, 7, 8, 9, 11
Abdominal (See Note 10)PPPPNote 1Notes 2, 4, 7, 8, 9, 11
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (See Note 13)
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vessel
Other (spec.)

N= new indication; P= previously cleared by FDA K173204; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+DPI, B+CW, B+C+PW, B+PD+PW, B+PPF+PW, B+PPI+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

  • Note 2: Includes imaging for guidance of biopsy
    Note 3: Includes infertility monitoring of follicle development

  • Note 4: Color M-mode
    Note 5: For example: thyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

  • Note 6: Abdominal organs and peripheral vessel

  • Note 7: Tissue Harmonic Imaging (THI)

  • Note 8: 3D imaging

  • Note 9: MultiVision (Spatial Compound Imaging)

  • Note 10: Includes Renal, Gynecology/Pelvis

  • Note 11: Panoramic imaging

  • Note 12: ElastoScan

  • Note 13: Includes Urology/Prostate

  • Note 14: S-Fusion Note 15: S-Shearwave

{7}------------------------------------------------

510(k) No.:

Device Name: CA3-10A for use with RS85

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)PPPPPNote 1Notes 2, 3, 7, 8, 9, 11
Abdominal (See Note 10)PPPPPNote 1Notes 2, 7, 8, 9, 11
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
PediatricPPPPPNote 1Notes 2, 7, 8, 9, 11
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)PPPPPNote 1Notes 2, 7, 8, 9, 11
Musculo-skel. (Superfic.)PPPPPNote 1Notes 2, 7, 8, 9, 11
Intra-luminal
Other (See Note 13)
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vesselPPPPPNote 1Notes 2, 7, 8, 9, 11
Other (spec.)

N= new indication; P= previously cleared by FDA K173204; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+PPI, B+CV, B+C+PW, B+PD+PW, B+DPD+PW, B+PPH+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

  • Note 2: Includes imaging for guidance of biopsy
    Note 3: Includes infertility monitoring of follicle development

  • Note 4: Color M-mode
    Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

  • Note 6: Abdominal organs and peripheral vessel
    Note 7: Tissue Harmonic Imaging (THI)

  • Note 8: 3D imaging
    Note 9: MultiVision (Spatial Compound Imaging)

  • Note 10: Includes Renal, Gynecology/Pelvis

  • Note 11: Panoramic imaging

  • Note 12: ElastoScan

  • Note 13: Includes Urology/Prostate Note 14: S-Fusion

Note 15: S-Shearwave

{8}------------------------------------------------

510(k) No.:

Device Name: E3-12A for use with RS85

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)PPPPPNote 1Notes 2, 7, 8, 9
Abdominal (See Note 10)PPPPPNote 1Notes 2, 7, 8, 9, 12
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectalPPPPPNote 1Notes 2, 7, 8, 9, 12
Trans-vaginalPPPPPNote 1Notes 2, 7, 8, 9, 12
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (See Note 13)PPPPPNote 1Notes 2, 7, 8, 9, 12
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vessel
Other (spec.)

N= new indication; P= previously cleared by FDA K173204; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+PPI, B+CW, B+C+PW, B+PD+PW, B+DPD+PW, B+PPI+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

  • Note 2: Includes imaging for guidance of biopsy
    Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: Spatial Compound Imaging

Note 10: Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging

Note 12: ElastoScan

Note 13: Includes Urology/Prostate

Note 14: S-Fusion Note 15: S-Shearwave

{9}------------------------------------------------

510(k) No.:

Device Name: L3-12A for use with RS85

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)
Abdominal (See Note 10)PPPPNote 1Note 2, 5, 6, 7, 8, 9, 11, 14
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (See Note 5)PPPPNote 1Note 2, 5, 6, 7, 8, 9, 11, 12, 14
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)PPPPNote 1Note 2, 5, 6, 7, 8, 9, 11, 14
Musculo-skel. (Superfic.)PPPPNote 1Note 2, 5, 6, 7, 8, 9, 11, 14
Intra-luminal
Other (See Note 13)
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vesselPPPPNote 1Note 2, 5, 6, 7, 8, 9, 11, 14
Other (spec.)

N= new indication; P= previously cleared by FDA K173204; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPL, B+CW, B+CW, B+C+PW, B+PD+PW, B+DPD+PW, B+PPI+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

  • Note 2: Includes imaging for guidance of biopsy
    Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: Spatial Compound Imaging

Note 10: Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging

Note 12: ElastoScan

Note 13: Includes Urology/Prostate

Note 14: S-Fusion Note 15: S-Shearwave

{10}------------------------------------------------

510(k) No.:

Device Name: EA2-11B for use with RS85

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)PPPPPNote 1Note 2, 7, 8, 9
Abdominal (See Note 12)PPPPPNote 1Note 2, 6, 7, 8, 9, 12
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectalPPPPPNote 1Note 2, 7, 8, 9, 12
Trans-vaginalPPPPPNote 1Note 2, 7, 8, 9, 12
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (See Note 13)PPPPPNote 1Note 2, 7, 8, 9, 12
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vessel
Other (spec.)

N= new indication; P= previously cleared by FDA K173204; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+ID, B+CV, B+C+PW, B+PD+PW, B+PPH+W, B+PPH+PW, B+ID+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

  • Note 2: Includes imaging for guidance of biopsy
    Note 3: Includes infertility monitoring of follicle development

  • Note 4: Color M-mode
    Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

  • Note 6: Abdominal organs and peripheral vessel
    Note 7: Tissue Harmonic Imaging (THI)

  • Note 8: 3D imaging

  • Note 9: MultiVision (Spatial Compound Imaging)

  • Note 10: Includes Renal, Gynecology/Pelvis

  • Note 11: Panoramic imaging

  • Note 12: ElastoScan

  • Note 13: Includes Urology/Prostate

  • Note 14: S-Fusion

  • Note 15: S-Shearwave

{11}------------------------------------------------

510(k) No.:

Device Name: LA2-9A for use with RS85

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)
Abdominal (See Note 10)PPPPNote 1Note 2, 5, 6, 7, 8, 9, 11
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (See Note 5)PPPPNote 1Note 2, 5, 6, 7, 8, 9, 11,12,15
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)PPPPNote 1Note 2, 5, 6, 7, 8, 9, 11,12,15
Musculo-skel. (Superfic.)PPPPNote 1Note 2, 5, 6, 7, 8, 9, 11,12,15
Intra-luminal
Other (See Note 13)
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vesselPPPPNote 1Note 2, 5, 6, 7, 8, 9, 11
Other (spec.)

N= new indication; P= previously cleared by FDA K173204; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+PPI, B+CW, B+C+PW, B+PD+PW, B+DPD+PW, B+PPI+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

  • Note 2: Includes imaging for guidance of biopsy
    Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: Spatial Compound Imaging

Note 10: Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging

Note 12: ElastoScan

Note 13: Includes Urology/Prostate

Note 14: S-Fusion Note 15: S-Shearwave

{12}------------------------------------------------

510(k) No.:

Device Name: LA3-16A for use with RS85

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
GeneralSpecificBMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
(Track I only)(Tracks I & III)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)
Abdominal (See Note 10)
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging&OtherLaparoscopic
Pediatric
Small Organ (See Note 5)PPPPNote 1Note 2, 5, 6, 7, 9, 11, 12
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)PPPPNote 1Note 2, 5, 6, 7, 9, 11
Musculo-skel. (Superfic.)PPPPNote 1Note 2, 5, 6, 7, 9, 11
Intra-luminal
Other (See Note 13)
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vesselPPPPNote 1Note 2, 5, 6, 7, 9, 11
Other (spec.)

N= new indication; P= previously cleared by FDA K173204; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+PPI, B+CV, B+C+PW, B+PD+PW, B+DPD+PW, B+PPH+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

  • Note 2: Includes imaging for guidance of biopsy
    Note 3: Includes infertility monitoring of follicle development

  • Note 4: Color M-mode
    Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

  • Note 6: Abdominal organs and peripheral vessel

  • Note 7: Tissue Harmonic Imaging (THI)

  • Note 8: 3D imaging

  • Note 9: MultiVision (Spatial Compound Imaging)

  • Note 10: Includes Renal, Gynecology/Pelvis

  • Note 11: Panoramic imaging

  • Note 12: ElastoScan

  • Note 13: Includes Urology/Prostate

  • Note 14: S-Fusion Note 15: S-Shearwave

{13}------------------------------------------------

510(k) No.:

Device Name: LA3-16AI for use with RS85

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal Imaging& OtherFetal/Obstetrics (See Note 3)
Abdominal (See Note 10)
Intra-operative (See Note 6)PPPPNote 1Note 7, 9, 11
Intra-operative (Neuro.)PPPPNote 1Note 7, 9, 11
Laparoscopic
Pediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)PPPPNote 1Note 7, 9, 11
Musculo-skel. (Superfic.)PPPPNote 1Note 7, 9, 11
Intra-luminal
Other (See Note 13)
CardiacCardiac Adult
Cardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vesselPPPPNote 1Note 7, 9, 11
Other (spec.)

N= new indication; P= previously cleared by FDA K173204; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+PPI, B+CV, B+C+PW, B+PD+PW, B+DPD+PW, B+PPH+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

  • Note 2: Includes imaging for guidance of biopsy
    Note 3: Includes infertility monitoring of follicle development

  • Note 4: Color M-mode
    Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

  • Note 6: Abdominal organs and peripheral vessel
    Note 7: Tissue Harmonic Imaging (THI)

  • Note 8: 3D imaging

  • Note 9: MultiVision (Spatial Compound Imaging)

  • Note 10: Includes Renal, Gynecology/Pelvis

  • Note 11: Panoramic imaging

  • Note 12: ElastoScan

  • Note 13: Includes Urology/Prostate Note 14: S-Fusion

  • Note 15: S-Shearwave

{14}------------------------------------------------

510(k) No.:

Device Name: LA4-18B for use with RS85

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)
Abdominal (See Note 10)
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (See Note 5)PPPPNote 1Note 2, 5, 6, 7, 8, 9, 11, 12, 14
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)PPPPNote 1Note 2, 5, 6, 7, 8, 9, 11, 14
Musculo-skel. (Superfic.)PPPPNote 1Note 2, 5, 6, 7, 8, 9, 11, 14
Intra-luminal
Other (See Note 13)
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vesselPPPPNote 1Note 2, 5, 6, 7, 8, 9, 11, 14
Other (spec.)

N= new indication; P= previously cleared by FDA K173204; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+PPI, B+CV, B+C+PW, B+PD+PW, B+DPD+PW, B+PPH+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

  • Note 2: Includes imaging for guidance of biopsy
    Note 3: Includes infertility monitoring of follicle development

  • Note 4: Color M-mode
    Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

  • Note 6: Abdominal organs and peripheral vessel
    Note 7: Tissue Harmonic Imaging (THI)

  • Note 8: 3D imaging
    Note 9: MultiVision (Spatial Compound Imaging)

  • Note 10: Includes Renal, Gynecology/Pelvis

  • Note 11: Panoramic imaging

  • Note 12: ElastoScan

  • Note 13: Includes Urology/Prostate

Note 14: S-Fusion Note 15: S-Shearwave

{15}------------------------------------------------

510(k) No.: K162094 Device Name: LM4-15B for use with RS85 Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)
Abdominal (See Note 10)PPPPNote 1Note 2, 5, 6, 7, 8, 9, 11
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (See Note 5)PPPPNote 1Note 2, 5, 6, 7, 8, 9, 11, 12
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)PPPPNote 1Note 2, 5, 6, 7, 8, 9, 11
Musculo-skel. (Superfic.)PPPPNote 1Note 2, 5, 6, 7, 8, 9, 11
Intra-luminal
Other (See Note 13)
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vesselPPPPNote 1Note 2, 5, 6, 7, 8, 9, 11
Other (spec.)

N= new indication; P= previously cleared by FDA K173204; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+PPI, B+CV, B+C+PW, B+PD+PW, B+DPD+PW, B+PPH+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

  • Note 2: Includes imaging for guidance of biopsy
    Note 3: Includes infertility monitoring of follicle development

  • Note 4: Color M-mode
    Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

  • Note 6: Abdominal organs and peripheral vessel
    Note 7: Tissue Harmonic Imaging (THI)

  • Note 8: 3D imaging
    Note 9: MultiVision (Spatial Compound Imaging)

  • Note 10: Includes Renal, Gynecology/Pelvis

  • Note 11: Panoramic imaging

  • Note 12: ElastoScan Note 13: Includes Urology/Prostate

  • Note 14: S-Fusion

  • Note 15: S-Shearwave

{16}------------------------------------------------

510(k) No.:

Device Name: PM1-6A for use with RS85

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal Imaging& OtherFetal/Obstetrics (See Note 3)
Abdominal (See Note 10)PPPPPNote 1Note 7
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Laparoscopic
Pediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult CephalicPPPPPNote 1Note 7
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (See Note 13)
CardiacCardiac AdultPPPPPNote 1Note 4, 7
Cardiac PediatricPPPPPNote 1Note 4, 7
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vessel
Other (spec.)

N= new indication; P= previously cleared by FDA K173204; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+PPI, B+CW, B+C+PW, B+PD+PW, B+DPD+PW, B+PPI+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

  • Note 2: Includes imaging for guidance of biopsy
    Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: Spatial Compound Imaging

Note 10: Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging

Note 12: ElastoScan

Note 13: Includes Urology/Prostate

Note 14: S-Fusion Note 15: S-Shearwave

{17}------------------------------------------------

510(k) No.:

Device Name: PA3-8B for use with RS85

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)
Abdominal (See Note 10)PPPPPNote 1Note 7
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
PediatricPPPPPNote 1Note 7
Small Organ (See Note 5)
Neonatal CephalicPPPPPNote 1Note 7
Adult CephalicPPPPPNote 1Note 7
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (See Note 13)
Cardiac AdultPPPPPNote 1Note 4, 7
CardiacCardiac PediatricPPPPPNote 1Note 4, 7
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vessel
Other (spec.)

N= new indication; P= previously cleared by FDA K173204; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+ID, B+CV, B+C+PW, B+PD+PW, B+PPH+W, B+PPH+PW, B+ID+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

  • Note 2: Includes imaging for guidance of biopsy
    Note 3: Includes infertility monitoring of follicle development

  • Note 4: Color M-mode
    Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

  • Note 7: Tissue Harmonic Imaging (THI)
  • Note 8: 3D imaging
  • Note 9: MultiVision (Spatial Compound Imaging)
  • Note 10: Includes Renal, Gynecology/Pelvis
  • Note 11: Panoramic imaging
  • Note 12: ElastoScan
  • Note 13: Includes Urology/Prostate
  • Note 14: S-Fusion
  • Note 15: S-Shearwave

{18}------------------------------------------------

510(k) No.:

Device Name: PA4-12B for use with RS85

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)
Abdominal (See Note 10)PPPPPNote 1Note 7
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
PediatricPPPPPNote 1Note 7
Small Organ (See Note 5)
Neonatal CephalicPPPPPNote 1Note 7
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (See Note 13)
Cardiac AdultPPPPPNote 1Note 4, 7
CardiacCardiac PediatricPPPPPNote 1Note 4, 7
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vessel
Other (spec.)

N= new indication; P= previously cleared by FDA K173204; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+TD, B+CV, B+C+PW, B+PD+PW, B+PPI+PW, B+PPH+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

  • Note 2: Includes imaging for guidance of biopsy
    Note 3: Includes infertility monitoring of follicle development

  • Note 4: Color M-mode
    Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

  • Note 6: Abdominal organs and peripheral vessel

  • Note 7: Tissue Harmonic Imaging (THI)

  • Note 8: 3D imaging

  • Note 9: MultiVision (Spatial Compound Imaging)

  • Note 10: Includes Renal, Gynecology/Pelvis

  • Note 11: Panoramic imaging

  • Note 12: ElastoScan

  • Note 13: Includes Urology/Prostate

  • Note 14: S-Fusion

  • Note 15: S-Shearwave

{19}------------------------------------------------

510(k) No.:

Device Name: MMPT3-7 for use with RS85

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal Imaging& OtherFetal/Obstetrics (See Note 3)
Abdominal (See Note 12)
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Laparoscopic
Pediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (spec.) (See Note 13)
CardiacCardiac Adult
Cardiac Pediatric
Trans-esophageal (Cardiac)PPPPPNote 1Note 4, 7
Other (spec.)
PeripheralVesselPeripheral vessel
Other (spec.)

N= new indication; P= previously cleared by FDA K173204; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+PPI, B+CV, B+C+PW, B+PD+PW, B+DPD+PW, B+PPH+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

  • Note 2: Includes imaging for guidance of biopsy
    Note 3: Includes infertility monitoring of follicle development

  • Note 4: Color M-mode
    Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

  • Note 6: Abdominal organs and peripheral vessel

  • Note 7: Tissue Harmonic Imaging (THI)

  • Note 8: 3D imaging

Note 9: MultiVision (Spatial Compound Imaging)

  • Note 10: Includes Renal, Gynecology/Pelvis
  • Note 11: Panoramic imaging
  • Note 12: ElastoScan
  • Note 13: Includes Urology/Prostate
  • Note 14: S-Fusion Note 15: S-Shearwave

{20}------------------------------------------------

510(k) No.:

Device Name: CV1-8A for use with RS85

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)PPPPPNote 1Note 2, 4, 7, 8, 9, 11
Abdominal (See Note 10)PPPPPNote 1Note 2, 4, 7, 8, 9
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (See Note 13)
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vessel
Other (spec.)

N= new indication; P= previously cleared by FDA K173204; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+PPI, B+CV, B+C+PW, B+PD+PW, B+DPD+PW, B+PPH+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

  • Note 2: Includes imaging for guidance of biopsy
    Note 3: Includes infertility monitoring of follicle development

  • Note 4: Color M-mode
    Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

  • Note 6: Abdominal organs and peripheral vessel
    Note 7: Tissue Harmonic Imaging (THI)

  • Note 8: 3D imaging
    Note 9: MultiVision (Spatial Compound Imaging)

  • Note 10: Includes Renal, Gynecology/Pelvis

  • Note 11: Panoramic imaging

  • Note 12: ElastoScan

  • Note 13: Includes Urology/Prostate Note 14: S-Fusion

Note 15: S-Shearwave

{21}------------------------------------------------

510(k) No.:

Device Name: V5-9 for use with RS85

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)PPPPPNote 1Note 2, 7, 8, 9
Abdominal (See Note 10)PPPPPNote 1Note 2, 7, 8, 9, 12
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectalPPPPPNote 1Note 2, 7, 8, 9, 12
Trans-vaginalPPPPPNote 1Note 2, 7, 8, 9, 12
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (See Note 13)PPPPPNote 1Note 2, 7, 8, 9, 12
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vessel
Other (spec.)

N= new indication; P= previously cleared by FDA K173204; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+PPI, B+CW, B+C+PW, B+PD+PW, B+DPD+PW, B+PPH+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

  • Note 2: Includes imaging for guidance of biopsy
    Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: Spatial Compound Imaging

Note 10: Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging

Note 12: ElastoScan

Note 13: Includes Urology/Prostate

Note 14: S-Fusion Note 15: S-Shearwave

{22}------------------------------------------------

510(k) No.:

Device Name: LV3-14A for use with RS85

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
GeneralSpecificBMPWDCWDColorCombined*Other
(Track I only)(Tracks I & III)Doppler*(Spec.)(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)
Abdominal (See Note 10)
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal ImagingLaparoscopic
&OtherPediatric
Small Organ (See Note 5)PPPPPNote 1Note 2, 5, 6, 7, 8, 9, 11
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)PPPPPNote 1Note 2, 5, 6, 7, 8, 9, 11
Musculo-skel. (Superfic.)PPPPPNote 1Note 2, 5, 6, 7, 8, 9, 11
Intra-luminal
Other (See Note 13)
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralPeripheral vesselPPPPPNote 1Note 2, 5, 6, 7, 8, 9, 11
VesselOther (spec.)

N= new indication; P= previously cleared by FDA K173204; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+PPI, B+CW, B+C+PW, B+PD+PW, B+DPD+PW, B+PPI+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

  • Note 2: Includes imaging for guidance of biopsy
    Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: Spatial Compound Imaging

Note 10: Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging

Note 12: ElastoScan

Note 13: Includes Urology/Prostate

Note 14: S-Fusion

Note 15: S-Shearwave

{23}------------------------------------------------

510(k) No.:

Device Name: CW6.0 for use with RS85

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal Imaging& OtherFetal/Obstetrics (See Note 3)
Abdominal (See Note 10)
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Laparoscopic
Pediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (See Note 13)
CardiacCardiac AdultP
Cardiac PediatricP
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vesselP
Other (spec.)

N= new indication; P= previously cleared by FDA K173204; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+PPI, B+CW, B+C+PW, B+PD+PW, B+DPD+PW, B+PPH+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

  • Note 2: Includes imaging for guidance of biopsy
    Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: Spatial Compound Imaging

Note 10: Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging

Note 12: ElastoScan

Note 13: Includes Urology/Prostate

Note 14: S-Fusion Note 15: S-Shearwave

{24}------------------------------------------------

510(k) No.:

Device Name: DP2B for use with RS85

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal Imaging& OtherFetal/Obstetrics (See Note 3)
Abdominal (See Note 10)
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Laparoscopic
Pediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (See Note 13)
CardiacCardiac AdultP
Cardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vessel
Other (spec.)

N= new indication; P= previously cleared by FDA K173204; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+PPI, B+CV, B+C+PW, B+PD+PW, B+DPD+PW, B+PPH+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

  • Note 2: Includes imaging for guidance of biopsy
    Note 3: Includes infertility monitoring of follicle development

  • Note 4: Color M-mode
    Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

  • Note 6: Abdominal organs and peripheral vessel

  • Note 7: Tissue Harmonic Imaging (THI)

  • Note 8: 3D imaging

  • Note 10: Includes Renal, Gynecology/Pelvis

  • Note 11: Panoramic imaging

  • Note 12: ElastoScan

  • Note 13: Includes Urology/Prostate Note 14: S-Fusion

  • Note 15: S-Shearwave

Note 9: MultiVision (Spatial Compound Imaging)

{25}------------------------------------------------

510(k) No.:

Device Name: DP8B for use with RS85

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal Imaging& OtherFetal/Obstetrics (See Note 3)
Abdominal (See Note 12)
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Laparoscopic
Pediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (See Note 13)
CardiacCardiac AdultP
Cardiac PediatricP
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vesselP
Other (spec.)

N= new indication; P= previously cleared by FDA K173204; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+ID, B+CV, B+C+PW, B+PD+PW, B+PPH+W, B+PPH+PW, B+ID+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

  • Note 2: Includes imaging for guidance of biopsy
    Note 3: Includes infertility monitoring of follicle development

  • Note 4: Color M-mode
    Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

  • Note 6: Abdominal organs and peripheral vessel
    Note 7: Tissue Harmonic Imaging (THI)

  • Note 8: 3D imaging

  • Note 9: MultiVision (Spatial Compound Imaging)

  • Note 10: Includes Renal, Gynecology/Pelvis

  • Note 11: Panoramic imaging

  • Note 12: ElastoScan

  • Note 13: Includes Urology/Prostate

  • Note 14: S-Fusion

  • Note 15: S-Shearwave

{26}------------------------------------------------

510(k) No.:

Device Name: EV3-10B for use with RS85

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)PPPPPNote 1Note 2, 7, 8, 9, 11
Abdominal (See Note 10)PPPPNote 1Note 2, 7, 8, 9, 11, 12
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectalPPPPNote 1Note 2, 7, 8, 9, 11, 12
Trans-vaginalPPPPNote 1Note 2, 7, 8, 9, 11, 12
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (See Note 13)PPPPNote 1Note 2, 7, 8, 9, 11, 12
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vessel
Other (spec.)

N= new indication; P= previously cleared by FDA K173204; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+PPI, B+CV, B+C+PW, B+PD+PW, B+DPD+PW, B+PPH+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

  • Note 2: Includes imaging for guidance of biopsy
    Note 3: Includes infertility monitoring of follicle development

  • Note 4: Color M-mode
    Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

  • Note 6: Abdominal organs and peripheral vessel

  • Note 7: Tissue Harmonic Imaging (THI)

  • Note 8: 3D imaging

  • Note 9: MultiVision (Spatial Compound Imaging)

  • Note 10: Includes Renal, Gynecology/Pelvis

  • Note 11: Panoramic imaging

  • Note 12: ElastoScan

  • Note 13: Includes Urology/Prostate Note 14: S-Fusion

  • Note 15: S-Shearwave

{27}------------------------------------------------

510(K) Summary: K182932

In accordance with 21 CFR 807.92 the following summary of information is provided:

  • Date Prepared January 13, 2019 1.
    1. Manufacturer SAMSUNG MEDISON CO., LTD. 3366, Hanseo-ro, Nam-myeon, Hongcheon-gun, Gangwon-do 25108, REPUBLIC OF KOREA
    1. Primary Contact Person Ji Yea Lee Regulatory Affairs Specialist Phone: +82.2.2.2194.1594 Fax: +82. 31.8017.9573 Email: jiyea722.lee@samsungmedison.com
    1. Proposed Device
    • Common/Usual Name: Diagnostic Ultrasound System and Accessories
    • Proprietary Name: RS85 Diagnostic Ultrasound System
    • Common Name: Diagnostic Ultrasound System
    • Classification Names: system, imaging, pulsed doppler, ultrasonic
    • Product Code: IYN, IYO, ITX
    • Regulation: 892.1550, 892.1560, 892.1570
    1. Predicate Device
    • Primary RS85 Diagnostic Ultrasound System (K173204) ।
    • i Reference - LOGIQ E9 Diagnostic Ultrasound System (K142160)

-Aplio i900/i800/i700/i600 Diagnostic Ultrasound System (K182679)

    1. Device Description
      The RS85 is a general purpose, mobile, software controlled, diagnostic ultrasound system. Its function is to acquire ultrasound data and to display the data as 2D mode. Color Doppler imaging. Power Doppler imaging (including Directional Power Doppler mode: S-Flow), PW Spectral Doppler mode, CW Spectral Doppler mode, Harmonic imaging, Tissue Doppler imaging, Tissue Doppler Wave, 3D imaging mode (real-time 4D imaging mode), Elastoscan* Mode or as a combination of these modes. The RS85 also gives the operator the ability to measure anatomical structures and offers analysis packages that provide information that may aid in making a diagnosis by competent health care professionals. The RS85 has real time acoustic output display with two basic indices, a mechanical index and a thermal index, which are both automatically displayed.
    1. Indications for Use
      The ultrasound diagnostic system and probes are designed to obtain ultrasound images and analyze body fluids.

The clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Intraoperative, Pediatric, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-rectal, Trans-vaginal, Muscular-Skeletal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric, Trans-esophageal (Cardiac) and Peripheral vessel.

The Indications for use of the proposed device is the same as predicates.

{28}------------------------------------------------

8. Technology

The RS85 employs the same fundamental scientific technology as its predicate device(s).

    1. Determination of Substantial Equivalence
      The proposed RS85 is substantially equivalent to the predicate devices with regards to intended use, imaging capabilities, technological characteristics and safety and effectiveness.

A comparison of the proposed device to the currently marketed predicates is provided in the table below.

FeatureSubject DevicePredicate Device
Device name(510(k) clearance)RS85(Proposed Device)RS85(K173204)Primary
ManufacturerSamsung Medison, Co.,LTDSamsung Medison, Co.,LTD
ClinicalApplication- Fetal/Obstetrics- Abdominal- Gynecology- Intra-operative- Pediatric- Small Organ,- Neonatal Cephalic- Adult Cephalic- Trans-rectal- Trans-vaginal- Trans-esoph. (non-Cardiac)- Muscular-Skeletal (Conventional,Superficial)- Urology- Cardiac Adult- Cardiac Pediatric- Trans-esophageal (Cardiac)- Peripheral vessel- Fetal/Obstetrics- Abdominal- Gynecology- Intra-operative- Pediatric- Small Organ,- Neonatal Cephalic- Adult Cephalic- Trans-rectal- Trans-vaginal- Trans-esoph. (non-Cardiac)- Muscular-Skeletal (Conventional,Superficial)- Urology- Cardiac Adult- Cardiac Pediatric- Trans-esophageal (Cardiac)- Peripheral vessel
Scanhead Types- Linear Array- Curved Linear Array- Endocavity- Phased Array- Static Probes- Linear Array- Curved Linear Array- Endocavity- Phased Array- Static Probes
ScanheadFrequency1.0 ~ 20.0 MHz1.0 ~ 20.0 MHz
Acoustic OutputDisplay & FDALimits- Display Feature for Higher Output-Track3- MI Output DisplayTI Output Display- Display Feature for Higher Output-Track3- MI Output DisplayTI Output Display
Modes of Operation- B-mode- M-mode- Pulsed wave (PW) Doppler- Continuous wave (CW) Doppler- Color Doppler- B-mode- M-mode- Pulsed wave (PW) Doppler- Continuous wave (CW) Doppler- Color Doppler
- Tissue Harmonic Imaging- Tissue Harmonic Imaging
- 3D/4D imaging mode- 3D/4D imaging mode
- Combined modes- Combined modes
#Transmit Channels192192
#Receive Channels192192
System- Beamformer 192- Beamformer 192
Characteristics:- Mobile cart- Mobile cart
- LCD Monitor (LED Backlight unit)- LCD Monitor (LED Backlight unit)
- 256 gray shades on monitor- 256 gray shades on monitor
- 100-240VAC, 1100VA, 50/60Hz- 100-240VAC, 1100VA, 50/60Hz
Acoustic OutputTrack 3Track 3
Display Standard
Software- Q Scan- Q Scan
Functionality- ClearVision- ClearVision
- MultiVision- MultiVision
- Panoramic- Panoramic
- Needle Mate- Needle Mate
- Auto IMT+- Auto IMT+
- Strain+- Strain+
- Stress Echo- Stress Echo
- Elastoscan- Elastoscan
- E-Thyroid- E-Thyroid
- E-Breast- E-Breast
- E-Strain- E-Strain
- S-Shearwave- S-Shearwave
- S-Detect for Breast- S-Detect for Breast
- S-Detect for Thyroid- S-Detect for Thyroid
- HQ Vision- HQ Vision
- S-Fusion- S-Fusion
Transducer- Clear Track- Virtual Track- S-3D ArterialAnalysis- ADVR- 3D Imaging- (Volume Data Acquisition)- 3D Imaging presentation- 3D Cine/4D Cine- 3D Rendering MPR- 3D XI MSV/Oblique View- Volume CT- 3D MagiCut- Volume Calculation- (VOCAL, XI VOCAL)- XI STIC- HDVI- Realistic Vue- Natural Vue- CEUS+- MV-Flow- S-Shearwave Imaging- CA1-7A- CF4-9- CA2-8A- CA3-10A- E3-12A- L3-12A- EA2-11B- LA2-9A- LA3-16A- LA3-16AI- LA4-18B- LM4-15B- PM1-6A- PA3-8B- PA4-12B- MMPT3-7- CV1-8A- V5-9- LV3-14A- CW6.0- DP2B- DP8B- EV3-10B- Clear Track- Virtual Track- S-3D ArterialAnalysis- ADVR- 3D Imaging- (Volume Data Acquisition)- 3D Imaging presentation- 3D Cine/4D Cine- 3D Rendering MPR- 3D XI MSV/Oblique View- Volume CT- 3D MagiCut- Volume Calculation- (VOCAL, XI VOCAL)- XI STIC- HDVI- Realistic Vue- Natural Vue- CEUS+- MV-Flow- S-Shearwave Imaging- CA1-7A- CF4-9- CA2-8A- CA3-10A- E3-12A- L3-12A- EA2-11B- LA2-9A- LA3-16A- LA3-16AI- LA4-18B- LM4-15B- PM1-6A- PA3-8B- PA4-12B- MMPT3-7- CV1-8A- V5-9- LV3-14A- CW6.0- DP2B- DP8B- EV3-10B
Biopsy Guide- BP-KIT-029- BP-KIT-030- BP-KIT-043- BP-KIT-029- BP-KIT-030- BP-KIT-043
- BP-KIT-053- BP-KIT-053
- BP-KIT-054- BP-KIT-054
- BP-KIT-055- BP-KIT-055
- BP-KIT-057- BP-KIT-057
- BP-KIT-058- BP-KIT-058
- BP-KIT-059- BP-KIT-059
- BP-KIT-060- BP-KIT-060
- BP-KIT-067- BP-KIT-067
- BP-KIT-068- BP-KIT-068
- BP-KIT-069- BP-KIT-069
- BP-KIT-071- BP-KIT-071
- BP-KIT-075- BP-KIT-075
- BP-KIT-083
Digital- Foot Switch- Foot Switch
Storage/Transfer- ECG (defibrillation proof)- ECG
Station- Gel Warmer- Gel Warmer
- Clear Track/ Virtual Track- Clear Track/ Virtual Track
- System StandSystem Stand
- WLAN Card- WLAN Card

{29}------------------------------------------------

{30}------------------------------------------------



{31}------------------------------------------------

  • The PET/CT modality of the S-Fusion feature is substantially equivalent to PET/CT modality of Volume Navigation feature of the reference device, LOGIQ E9 (K142160).

  • The MV Index is substantially equivalent to Vascularity Index of the reference device, Aplio i900/i800/i700/i600 (K182679).

10. Summary of Non-Clinical Test

The proposed device has been evaluated for acoustic output, biocompatibility, cleaning and disinfection effectiveness as well as thermal, electromagnetic and mechanical safety, and has been found to conform with applicable medical device safety standards. The RS85 and its applications comply with voluntary standards.

TestStandards and FDA Guidance
Risk ManagementISO 14971 Second edition 2007 Medical devices - Application of risk management tomedical devices
Electrical SafetyThe RS85 Ultrasound System with defibrillation-proof ECG electrode was evaluated perthe following standards.ANSI AAMI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012Medical Electrical Equipment - Part 1: General Requirements for basic safety andessential performance.
ElectromagneticCompatibilityIEC60601-1-2: 2014(4th Edition) Medical Electrical Equipment -- Part 1-2: GeneralRequirements For Basic Safety And Essential Performance -- Collateral Standard:Electromagnetic Disturbances -- Requirements And Tests
BiocompatibilityISO 10993-1 Fourth edition 2009-10-15 Biological evaluation of medical devices - Part1: Evaluation and testing within a risk management process.

{32}------------------------------------------------

ReprocessingMedical DevicesFDA Guidance: Reprocessing Medical Devices in Health Care Settings: ValidationMethods and Labeling issued on March 17, 2015, revised June 9, 2017.
Software/Firmware-driven FunctionalityAll migrated software functionality was evaluated using the same test criteria as thepredicate for all applicable imaging modes to ensure that migration into a new systemdesign did not compromise image quality with respect to the intended use of eachfeature.
Software/Firmware-driven FunctionalityGuidance for the Content of Premarket Submissions for Software Contained in MedicalDevices issued on May 11, 2005
Ultrasound SafetyInformation for Manufacturers Seeking Marketing Clearance ofDiagnostic Ultrasound Systems and Transducers issued September 9, 2008
IEC60601-2-37:2007 + A1:2015, Particular requirements for the safety of ultrasonicmedical diagnostic and monitoring equipment
NEMA UD 2-2004 (R2009)Acoustic Output Measurement Standard for Diagnostic Ultrasound Equipment Revision3
NEMA UD 3-2004 (R2009)Standard for Real-Time Display of Thermal and Mechanical Acoustic Output Indices onDiagnostic Ultrasound Equipment, Revision 2

11. Summary of Clinical Tests

The subject of this premarket submission, RS85, did not require clinical studies to support substantial equivalence.

12. Conclusion

SAMSUNG MEDISON considers the RS85 to be as safe, as effective, and performance is substantially equivalent to the predicate device.

END of 510(K) Summary

§ 892.1550 Ultrasonic pulsed doppler imaging system.

(a)
Identification. An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)