K Number
K191068
Date Cleared
2019-07-17

(86 days)

Product Code
Regulation Number
878.5400
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Powersculp laser lipolysis system is intended for non-invasive lipolysis of the flank and abdomen to achieve disruption of adipocyte cells intended for non-invasive aesthetic use to achieve a desired aesthetic affect. This treatment is intended for individuals with a Body Mass Index (BMI) of 30 or less.

Device Description

The Powersculp laser lipolysis system is a diode laser system. Electrically efficient semiconductors generate optical radiation (1060 mm) which is used to deliver laser energy to subcutaneous tissue layers. The Powersculp laser lipolysis system is capable of peak powers of 50W. The main components of Powersculp laser lipolysis system are a console and four applicators.

AI/ML Overview

The provided text describes a 510(k) premarket notification for the "Powersculp laser lipolysis system" and primarily focuses on demonstrating substantial equivalence to predicate devices rather than providing detailed acceptance criteria and study results for de novo performance validation as one might see for an AI/ML device.

Therefore, the document does not contain the specific information required to answer questions 1-9 comprehensively for a device proving it meets acceptance criteria through clinical studies. The submission explicitly states that clinical tests were not performed for the Powersculp given its substantial equivalence to already marketed devices.

However, I can extract information related to the device's technical specifications and how the manufacturer states the device meets safety and effectiveness requirements by relying on existing predicate devices and non-clinical testing.

Here's a breakdown of what can be inferred or directly stated from the document, and where information is explicitly lacking for the prompt's requirements:


Acceptance Criteria and Device Performance (Based on "Substantial Equivalence" to Predicates)

Since this is a 510(k) submission, the "acceptance criteria" are implicitly met by demonstrating substantial equivalence (SE) to legally marketed predicate devices. The performance is not measured against explicit clinical thresholds in this document, but rather by showing comparable technology and safety.

1. Table of Acceptance Criteria and Reported Device Performance

As this is a 510(k) based on substantial equivalence, there are no specific performance acceptance criteria like sensitivity, specificity, or accuracy for a diagnostic AI. Instead, the acceptance is based on demonstrating the proposed device is as safe and effective as the predicates. The "performance" is implicitly tied to the general performance of the predicate devices.

The table below summarizes the comparative attributes that, when deemed "similar" or "same" to the predicate, imply acceptance for a 510(k) submission.

Attribute/Criterion (Implied for SE)Powersculp Laser Lipolysis System (Proposed Device)Predicate Devices (K160470 & K150230 SculpSure)Comparison/Outcome
Indications for UseNon-invasive lipolysis of the flank and abdomen to achieve disruption of adipocyte cells for aesthetic affect (BMI <= 30).Identical wording.Same - Crucial for SE
Operating TheoryDiode laser system; electrically efficient semiconductors generate optical radiation (1060 nm) which is used to deliver laser energy to subcutaneous tissue layers. Peak powers of 50W.Diode laser system; electrically efficient semiconductors generate optical radiation (1060 nm or 1064 nm) used to deliver laser energy to subcutaneous tissue layers.Same - "Although there are differences in expression among the three devices, the technical principles are essentially the same."
Structure & Main ComponentsConsole and four applicators.Console and four applicators (K160470) or console and applicator (K150230).Same (as primary predicate K160470)
Laser TypeDiode laserDiode laserSame
Lipolysis MethodHeat-assistedHeat-assistedSame
Wavelength1060nm ± 20 nm (infrared)1060 ± 20 nm (infrared) or 1064nm.Same
Pulse Width (laser ON time)CW (Continuous Wave)CWSame
CoolingContact coolingContact coolingSame
Attachment to PatientBeltBeltSame
Software ControlYesYesSame
EMC and Electrical Safety ComplianceIEC 60601-1-2, ANSI AAMI ES60601-1, IEC 60825-1, IEC 60601-2-22Same standards tested.Same (compliance with same standards)
BiocompatibilityAll patient contacting materials assessed as per ISO 10993-1 and found to be biocompatible.Implied for predicate devices.Meets standard (for proposed device)
Spot Size4 × 8 cm² (A single applicator of four applicators)4 × 6 cm² (K160470 - 4X applicators; K150230 - 3X applicators).Similar - "The spot size of applicator does not affect the safety and effectiveness." (More precisely, the number of applicators that can be used simultaneously is similar to the primary predicate, but the individual spot size differs, yet is deemed safe).
Power DensityUp to 0.7-1.7 W/cm²Up to 1.4 W/cm² (K160470); 1.7 W/cm² (K150230).Similar - "The power density of the proposed device is customizable, and the maximum power density is same as the secondary predicate device K150230, so this difference is not affect safety and effectiveness."
Peak Power50W (per applicator)30W (per applicator, K160470); 40W (per applicator, K150230).Similar - "The peak power is different, which related to power density and spot size. The difference does not affect safety and effectiveness."
Power SupplyAC100-240V, 50/60Hz, 15A200-240V~, Single Phase, 20A (K160470); 120V, 20A (K150230).Similar - "The power supply is different, not affect safety and effectiveness."

2. Sample size used for the test set and the data provenance

  • Sample Size: N/A for clinical testing. The submission states: "Lotuxs belives that the proposed device Powersculp laser lipolysis system PSP050 does not need to carry out clinical tests, and the clinical study data that have been the legally marketed device can be used for reference."
  • Data Provenance: N/A for clinical testing of the proposed device itself. Any referenced clinical data would be from the predicate devices (likely retrospective studies conducted by their manufacturers).

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • N/A. No de novo clinical test set was used for the proposed device, as no clinical studies were performed.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • N/A. No de novo clinical test set was used.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • N/A. This device is a laser system for lipolysis, not an AI-assisted diagnostic or imaging device for human readers. No MRMC study was mentioned or required.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • N/A. This applies to an algorithm/AI, not a physical laser system. The "performance" of the device is its ability to perform lipolysis, not to interpret data.

7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

  • N/A. For this 510(k), the "ground truth" for demonstrating safety and effectiveness relies on the established safety and effectiveness of the predicate devices and robust non-clinical (bench) testing of the proposed device itself. The non-clinical tests included:
    • Software verification and validation (demonstrated performance as intended per FDA guidance).
    • Power output testing (confirmed specification).
    • Electromagnetic compatibility (EMC) and electrical safety testing (confirmed compliance with standards ANSI AAMI ES60601-1, IEC 60601-1-2, IEC 60601-2-22, IEC 60825-1).
    • Biocompatibility assessment of patient-contacting materials (per ISO 10993-1).

8. The sample size for the training set

  • N/A. This device has no "training set" in the context of an AI/ML model.

9. How the ground truth for the training set was established

  • N/A. Does not apply to this type of device.

Summary of how the device meets acceptance criteria (implicit in a 510(k)):

The Powersculp laser lipolysis system meets the implicit acceptance criteria for a 510(k) clearance by demonstrating substantial equivalence (SE) to legally marketed predicate devices (K160470 and K150230 Sculpsure). The manufacturer asserted that the proposed device has the same indications for use, fundamental operating theory, structure, components, laser type, lipolysis method, wavelength, pulse width, cooling method, and patient attachment method as the predicate devices.

Minor differences in spot size, power density, peak power, and power supply were argued not to affect safety and effectiveness. Non-clinical tests, including software verification and validation, power output testing, electromagnetic compatibility, electrical safety testing, and biocompatibility assessment, were performed to demonstrate that the device itself is safe and performs as intended, in line with recognized standards. This approach, outlined in Section 9 and 10 of the K191068 submission, negates the need for new clinical trials for this specific device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the symbol of the Department of Health & Human Services on the left and the text "FDA U.S. FOOD & DRUG ADMINISTRATION" on the right. The text is in a sans-serif font, with "FDA" in a blue square and the rest of the text in blue.

July 17, 2019

Wuhan Lotuxs Technology Co., Ltd. % Jinghua Zhou Regulation Control Manager Guangzhou Junyi Information Technology Co., Ltd. Room 215, Huaming Building, Chebei Road Guangzhou, China 511660

Re: K191068

Trade/Device Name: Powersculp laser lipolysis system Regulation Number: 21 CFR 878.5400 Regulation Name: Low Level Laser System for Aesthetic Use Regulatory Class: Class II Product Code: PKT Dated: April 22, 2019 Received: April 22, 2019

Dear Jinghua Zhou:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

for

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Neil Ogden Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K191068

Device Name

Powersculp laser lipolysis system

Indications for Use (Describe)

The Powersculp laser lipolysis system is intended for non-invasive lipolysis of the flank and abdomen to achieve disruption of adipocyte cells intended for non-invasive a desired a desired aesthetic affect. This treatment is intended for individuals with a Body Mass Index (BMI) of 30 or less.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

|× Prescription Use (Part 21 CFR 801 Subpart D)

__ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Section 5 - 510(k) Summary

K191068

Date of Summary Preparation: April 15, 2019

1. Submitter's Identifications

Submitter's Name: Wuhan Lotuxs Technology Co., Ltd. Address: 5F, E2 Building, NO 999 Gaoxin Avenue, Future City, East lake High-Tech Development Zone, Wuhan 430206, China Contact Person: Na Wu Contact Title: QA Manager Contact Email Address: na.wu@lotuxs.com Telephone: +86-27-87619668 Fax: +86-27-87515058

2. Correspondent's Identifications

Correspondent's Name: Guangzhou Junyi Information Technology Co., Ltd. Address: Room 215, Huaming Building, Chebei Road, Guangzhou, P.R. China ZIP Code: 511660 Contact Person: Jinghua Zhou Contact Title: Regulation Control Manager Contact E-mail Address: admanzhou@126.com Telephone: +86-20-82329549 Fax: +86-20-82329549

3. Name of the Device

Device Classification Name: Laser for disruption of adipocyte cells for aesthetic use Product Name: Low level laser system for aesthetic use Trade Name: Powersculp laser lipolysis system Model: PSP050 Classification Panel: General & Plastic Surgery Product Code: PKT Regulation Number: 21 CFR 878.5400 Device Classification: Class II

4. The Predicate Devices

Primary Predicate device: K160470 SculpSure Secondary predicate device: K150230 Sculpsure

5. Device Description

The Powersculp laser lipolysis system is a diode laser system. Electrically efficient semiconductors generate optical radiation (1060 mm) which is used to deliver laser

{4}------------------------------------------------

Wuhan Lotuxs Technology Co., Ltd.

energy to subcutaneous tissue layers. The Powersculp laser lipolysis system is capable of peak powers of 50W. The main components of Powersculp laser lipolysis system are a console and four applicators.

6. Intended Use of Device

The Powersculp laser lipolysis system is intended for non-invasive lipolysis of the flank and abdomen to achieve disruption of adipocyte cells intended for non-invasive aesthetic use to achieve a desired aesthetic affect. This treatment is intended for individuals with a Body Mass Index (BMI) of 30 or less.

{5}------------------------------------------------

Wuhan Lotuxs Technology Co., Ltd.

7. Summary of Substantial Equivalence

Table 1

Proposed devicePrimary predicate deviceSecondary predicate deviceComparison
510k Number------K160470K150230------
Product CodePKTPKTPKTSame
Proprietary NamePowersculp laser lipolysis systemSculpsureSculpsure------
ModelPSP050//------
ManufacturerWuhan Lotuxs Technology Co., Ltd.Cynosure, Inc.Cynosure, Inc.------
Indications for useThe Powersculp laser lipolysis system is intended for non-invasive lipolysis of the flank and abdomen to achieve disruption of adipocyte cells intended for non-invasive aesthetic use to achieve a desired aesthetic affect. This treatment is intended for individuals with a Body Mass Index (BMI) of 30 or less.The Cynosure SculpSure is a diode laser system intended for noninvasive lipolysis of the abdomen and flanks in individuals with a Body Mass Index (BMI) of 30 or less. The device is intended to affect the appearance of visible fat bulges in the abdomen and flanks.The Cynosure SculpSure is intended for non-invasive lipolysis of the flanks to achieve disruption of adipocyte cells intended for non-invasive aesthetic use to achieve a desired aesthetic affect. Intended for individuals with a Body Mass Index (BMI) of 30 or less.SameThe indications for use of the proposed device are same as those of predicate device K160470 and K150230.
Operating theoryThe Powersculp laser lipolysis system is a diode laser system. Electrically efficientThe Cynosure SculpSure is a diode laser system. The main components ofThe Cynosure SculpSure is a diode laser system. Electrically efficientSameAlthough there are differences in expression
semiconductors generate optical radiation (1060 nm) which is used to deliver laser energy to subcutaneous tissue layers. The Powersculp laser lipolysis system is capable of peak powers of 50W. The main components of Powersculp laser lipolysis system are a console and four applicators.SculpSure are a console and four applicators that deliver the laser energy to the patient. Electrically efficient semiconductors generate opticalradiation (1060 nm) which is used to deliver laser energy tosubcutaneous tissue layers.semiconductors generate optical radiation (1064 nm) which is used to directly irradiate the skin's surface. The Sculpsure is intended for noninvasive lipolysis of the flanks to achieve disruption of adipocyte cells intended for non-invasive aesthetic use to achieve a desired aesthetic affect.among the three devices, the technical principles are essentially the same.
Structure and main componentsThe main components of Powersculp laser lipolysis system are a console and four applicators.The main components of SculpSure are a console and four applicators.The main components of SculpSure are a console and applicator.SameThe main components of the proposed device are same as those of the primary predicate device K160470.
Laser typediode laserdiode laserdiode laserSame
Lipolysis methodHeat-assistedHeat-assistedHeat-assistedSame
Wavelength1060nm±20 nm (infrared)1060 ±20 nm (infrared)1064nmSame
Spot size4 × 8 cm² (A single applicator of four applicators)4 × 6 cm² on each of the applicator heads (4X)4 × 6 cm² (3X)SimilarThe amount of applicator is same between the primary predicate device K160470 and the proposed device.
Pulse Width (laserON time)CWCWCWThe spot size of applicatordoes not affect the safetyand effectiveness.
Power densityUp to 0.7-1.7W/ cm²Up to 1.4 W/cm²1.7W/ cm²SimilarThe power density of theproposed device iscustomizable, and themaximum power density issame as the secondarypredicate device K150230,so this defference is notaffect safety andeffectiveness.
Power supplyAC100-240V, 50/60Hz, 15A200-240V~, Single Phase,20A120V, 20ASimilarThe power supply isdifferent, not affect safetyand effectiveness.
Peak power50W (per applicator)30W (per applicator)40W (per applicator)SimilarThe peak power is different,which related to powerdensity and spot size. Thedifference does not affectsafety and effectiveness.
CoolingContact coolingContact coolingContact coolingSame
Attachment topatientbeltbeltbeltSame
Software controlYesYesYesSame
ElectromagneticIEC 60601-1-2IEC 60601-1-2IEC 60601-1-2Same
compatibility andANSI AAMI ES60601-1ANSI AAMI ES60601-1ANSI AAMI ES60601-1
electrical safefyIEC 60825-1IEC 60825-1IEC 60825-1
complianceIEC 60601-2-22IEC 60601-2-22IEC 60601-2-22
Discussion forSubstantiallyEquivalent (SE)The proposed device PSP050 has the same as the predicate device: indications for use, operating theory, structure andmain components, laser type, lipolysis method, wavelength, power width, cooling method, attachment to patient. Thedifferences only exist in such contents: spot size, power density, power supply, peak power that both can be controlled inrange of application. These minor differences between proposed device and predicate device raise no new issue of safetyand effectiveness. According to the non-clinical test results, the proposed devices are as safe, as effective and perform aswell as the predicate device.So the proposed device is Substantially Equivalent (SE) to the predicate device which is US legally market device.

{6}------------------------------------------------

Wuhan Lotuxs Technology Co., Ltd.

{7}------------------------------------------------

Wuhan Lotuxs Technology Co., Ltd.

{8}------------------------------------------------

Wuhan Lotuxs Technology Co., Ltd.

{9}------------------------------------------------

Wuhan Lotuxs Technology Co., Ltd.

8. Non-Clinical Tests Submitted:

Software verification and validation was performed, and it was demonstrated that the software performs as intended according to the FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices. Testing confirmed that the power output meets specification. Electromagnetic compatibility and electrical safety testing was performed per standards ANSI AAMI ES60601-1, IEC 60601-1-2, IEC 60601-2-22 and IEC 60825-1. Results confirmed the device meets the standards. All patient contacting materials were assessed as per ISO 10993-1 and found to be biocompatible.

9. Clinical Tests Submitted:

According to "8. Clinical Testing" of "Class II Special Controls Guidance Document: Low Level Laser System for Aesthetic Use", and operating instruction and "7. Summary of Substantial Equivalence" submitted of the proposed device, the proposed device PSP050 is Substantially Equivalent (SE) to the predicate device, includes: indication for use, design (such as operating theory, structure and main component, software control), technology (such as laser type, lipolysis method, wavelength, power width, cooling method, attachment to patient), the non-clinical tests complied with the requirements of relevant recognized standards, and passed the bench tests (such as performance tests, storage condition tests), so Lotuxs belives that the proposed device Powersculp laser lipolysis system PSP050 does not need to carry out clinical tests, and the clinical study data that have been the legally marketed device can be used for reference.

10. Conclusions drawn from clinical and non-clinical tests submitted:

Lotuxs believes that Powersculp laser lipolysis system PSP050 is substantially equivalent to its predicate devices with same indications for use, similar technological characteristics. The non-clinical data for Powersculp laser lipolysis system PSP050 supports the safety of the device and the biocompatibility, hardware and software verification and validation demonstrate that the Powersculp laser lipolysis system PSP050 should perform as intended in the specified use conditions.

--- End of this section ---

§ 878.5400 Low level laser system for aesthetic use

(a)
Identification. A Low Level Laser System for Aesthetic Use is a device using low level laser energy for the disruption of adipocyte cells within the fat layer for the release of fat and lipids from these cells for noninvasive aesthetic use.(b)
Classification. Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance for Industry and Food and Drug Administration Staff; Class II Special Controls Guidance Document: Low Level Laser System for Aesthetic Use.” See § 878.1(e) for the availability of this guidance document.