K Number
K190942
Date Cleared
2020-01-07

(272 days)

Product Code
Regulation Number
880.6250
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Nitrile Powder Free patient examination glove is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hands or finger to prevent contamination between patient and examiner.

Device Description

"Disposable Powder Free Nitrile Examination Glove, Pink Color" and "Disposable Powder Free Nitrile Examination Glove, Black Color" are patient examination gloves made from nitrile compound, non-sterile (as per 21 CFR 880.6250, Class I). The principle operation of the medical device to provide single use barrier protection for the wearer and the device meets all the requirement specifications for Barrier Protection, tensile properties as defined in ASTM D6319-10, Standard specification for Nitrile Examination Gloves.

AI/ML Overview

This document is a 510(k) summary for a disposable powder-free nitrile examination glove, and as such, it focuses on demonstrating substantial equivalence to a predicate device based on bench testing. It does not describe an AI medical device, and therefore, many of the requested elements for AI/ML evaluation (such as sample sizes for test/training sets, data provenance, expert ground truth, adjudication methods, MRMC studies, or standalone performance of an algorithm) are not applicable.

However, I can extract the acceptance criteria and reported device performance from the provided text for the physical characteristics of the glove.

1. Table of acceptance criteria and the reported device performance

CharacteristicsAcceptance Criteria (Standard)Reported Device Performance
DimensionASTM D6319-10 (Reapproved 2015)
Length≥230mmComplies with ASTM D6319-10 (230mm min.)
Width (X-Small)70 ± 10 mmComplies with ASTM D6319-10 (70 ± 10 mm)
Width (Small)80 ± 10 mmComplies with ASTM D6319-10 (80 ± 10 mm)
Width (Medium)95 ± 10 mmComplies with ASTM D6319-10 (95 ± 10 mm)
Width (Large)110 ± 10 mmComplies with ASTM D6319-10 (110 ± 10 mm)
Width (X-Large)120 ± 10 mmComplies with ASTM D6319-10 (120 ± 10 mm)
Thickness (Finger tip)≥0.05mmComplies with ASTM D6319-10 (0.05mm min.)
Thickness (Palm)≥0.05mmComplies with ASTM D6319-10 (0.05mm min.)
Physical PropertiesASTM D6319-10 (Reapproved 2015)
Tensile strength (Before aging)≥14MPa14 MPa, min.
Tensile strength (After aging)≥14MPa14 MPa, min.
Elongated rate (Before aging)≥500%500% min.
Elongated rate (After aging)≥400%400% min.
Freedom from pinholesPassed Standard Acceptance Criteria (21 CFR 800.20, ASTM D6319-10, ASTM D5151-06)In accordance with ASTM D6319-10 and ASTM D5151-06, G-1, AQL 2.5
Powder Residual< 2 mg/glove (ASTM D6319-10, D6124-06)< 2mg per glove (Complies with ASTM D6319-10)
BiocompatibilityISO 10993-10, ISO 10993-5 criteria
Primary Skin IrritationNot a primary skin irritantPasses (Not a primary skin irritant)
Dermal sensitizationNot a primary skin sensitizerPasses (Not a primary skin sensitizer)
In vitro cytotoxicityNot cytotoxicPasses (Not cytotoxic)

2. Sample size used for the test set and the data provenance
The document refers to "bench testing" and compliance with ASTM standards and ISO standards. These standards typically involve specific sample sizes and testing methodologies defined within the standard itself. The document does not explicitly state the exact number of units tested for each specific criterion. Data provenance is not specified beyond "bench testing" implies laboratory testing of the physical product.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. This is a physical device (nitrile examination glove) and not an AI/ML medical device where expert ground truth for interpretation would be required. The "ground truth" here is the pass/fail result against established material and performance standards.

4. Adjudication method for the test set
Not applicable, as this is bench testing of physical properties against established standards, not an AI output requiring expert adjudication.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is not an AI medical device.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
Not applicable. This is not an AI medical device.

7. The type of ground truth used
The "ground truth" for the device's performance is based on established national and international standards for medical gloves, specifically:

  • ASTM D6319-10 (Reapproved 2015) - Standard specification for Nitrile Examination Gloves
  • 21 CFR 800.20 - Freedom from holes
  • ASTM D5151-06 (Reapproved 2015) - Test method for detection of holes in medical gloves
  • ASTM D6124-06 (Reaffirmation 2011) - Test method for residual powder on medical gloves
  • ISO 10993-10: Third Edition 2010-08-01 - Biological evaluation of medical devices – Part 10: Tests for irritation and skin sensitization
  • ISO 10993-5: Third Edition 2009-06 - Biological evaluation of medical devices – Part 5: Tests for in vitro cytotoxicity

8. The sample size for the training set
Not applicable. This is not an AI medical device.

9. How the ground truth for the training set was established
Not applicable. This is not an AI medical device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo. On the right is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

Ever Growth (Vietnam) Co. Ltd. % Elizabeth Deng U.S. Representative Elizabeth Deng 5748 Eaglewood Place, Rancho Cucamonga Rancho Cucamonga, California 91730

Re: K190942

Trade/Device Name: Disposable Powder Free Nitrile Examination Glove, Pink/Black Color Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I. reserved Product Code: LZA Dated: October 10, 2019 Received: October 10, 2019

Dear Elizabeth Deng:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products): good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For Elizabeth Claverie, M.S. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K190942

Device Name

Disposable Powder Free Nitrile Examination Glove, Pink Color

Indications for Use (Describe)

The Nitrile Powder Free patient examination glove is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hands or finger to prevent contamination between patient and examiner.

Type of Use (Select one or both, as applicable)

_ Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K190942

Device Name

Disposable Powder Free Nitrile Examination Glove, Black Color

Indications for Use (Describe)

The Nitrile Powder Free patient examination glove is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hands or finger to prevent contamination between patient and examiner.

Type of Use (Select one or both, as applicable)

_ Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary K190942

1.0 Submitter:

Submitter's name :Ever Growth (Vietnam) Co., Ltd.
Submitter's address :Long Khanh Industrial Zone, Binh Loc Ward, Long Khanh Township, Dong Nai,Vietnam
Phone number :84-61-3514025
Fax number :84 -61-3514023
Name of contact person:Ming Lee
Summary Preparation Date:Dec. 9th, 2019
2.0US Agent:
US representative nameElizabeth Deng
Company address5748 Eaglewood PlaceRancho Cucamonga, CaliforniaRancho Cucamonga, CA 91739
Telephone number909 4659188
Contact emailBaxianunited48@Yahoo.Com
3.0 Name of the Device
Proprietary/Trade name:Disposable Powder Free Nitrile Examination Glove, Pink ColorDisposable Powder Free Nitrile Examination Glove, Black Color
Common Name:Nitrile Examination Gloves
510(k) NumberK190942
Classification Name:Patient Examination Glove
Device Classification:Class I
Regulation Number:21 CFR 880.6250
Product Code:LZA
4.0 Predicate device
Device Name:Orange Non Sterile Powder Free Nitrile Examination Gloves
Company name:Central Medicare Sdn. Bhd.
510(K) Number:K172642

5.0 Device Description:

"Disposable Powder Free Nitrile Examination Glove, Pink Color" and "Disposable Powder Free Nitrile Examination Glove, Black Color" are patient examination gloves made from nitrile compound, non-sterile (as per 21 CFR 880.6250, Class I). The principle operation of the medical device to provide single use barrier protection for the wearer and the device meets all the requirement specifications for Barrier Protection, tensile properties as defined in ASTM D6319-10, Standard specification for Nitrile Examination Gloves.

6.0 Indications for Use:

The Nitrile Powder Free patient examination glove is a non-sterile disposable device intended for medical purposes

that is worn on the examiner's hands or finger to prevent contamination between patient and examiner.

7.0 Summary of the Technological Characteristics of the Device:

"Disposable Powder Free Nitrile Examination Glove, Pink Color" and "Disposable Powder Free Nitrile Examination Glove, Black Color" are summarized with the following technological characteristics compared to ASTM or equivalent standard.

Table 7.1 Summary of the Technological Characteristics
--------------------------------------------------------------
CharacteristicsStandard
DimensionASTM standard D 6319-10(Reapproved 2015)
Length ≥230mm
WidthX Small 70 $\pm$ 10 mmSmall 80 $\pm$ 10 mmMedium 95 $\pm$ 10 mmLarge 110 $\pm$ 10 mmX large 120 $\pm$ 10 mm

{5}------------------------------------------------

ThicknessFinger tip≥0.05mm
Palm≥0.05mm
Physical PropertiesASTM standard D 6319-10(Reapproved 2015)
Tensile strength (Before aging)≥14MPa
Tensile strength (After aging)≥14MPa
Elongated rate (Before aging)≥500%
Elongated rate (After aging)≥400%
Freedom from pinholes21 CFR 800.20Passed Standard Acceptance Criteria
ASTM standard D 6319-10(Reapproved 2015)
Test method in accordance with ASTM D5151-06(Reapproved 2015)
Powder ResidualASTM standard D6319-10(Reapproved 2015)Test method in accordance with D6124-06(Reaffirmation 2011)< 2 mg/glove
BiocompatibilityPrimary Skin Irritation in rabbitsISO 10993-10: Third Edition 2010-08-01PassesUnder the conditions of the study, the subjectdevice is not a primary skin irritant.
Dermal sensitization in the guinea pigISO 10993-10: Third Edition 2010-08-01PassesUnder the conditions of the study, the subjectdevice is not a primary skin sensitizer.
In vitro cytotoxicity accordance withISO 10993-5: Third Edition 2009-06PassesUnder the conditions of the study, the subjectdevice is not cytotoxic.

8.0 Based on Assessment of Non-Clinical Performance Data:

The following bench testing was conducted for design elements and performance characteristics deemed appropriate to demonstrate equivalence to the predicate device, Orange Non Sterile Powder Free Nitrile Examination Gloves made by Central Medicare Sdn Bhd. The subject device "Disposable Powder Free Nitrile Examination Glove, Pink Color" and "Disposable Powder Free Nitrile Examination Glove, Black Color" met with the acceptance criteria ensuring substantial equivalence to the predicate device. No new safety or performance issues were raised during testing:

  • Dimension per ASTM D6319-10 (Reapproved 2015) >
  • A Tensile strength(Before aging/After aging) and Elongation(Before aging/After aging) per ASTM D6319-10(Reapproved 2015)
  • A Water leak test on pinhole per ASTM D6319-10(Reapproved 2015) and per 21 CFR 800.20.
  • Powder Residual tests per ASTM D6319-10(Reapproved 2015)

  • Biocompatibility test per ISO 10993-10: Third Edition 2010-08-01 and ISO 10993-5: Third Edition 2009-06

9.0 Based on Assessment of Clinical Performance Data:

Clinical data was not needed to demonstrate that the subject glove is substantially equivalent to the predicate glove. So determination of substantial equivalence is not based on an assessment of clinical performance data.

10.0 Substantial Equivalence Comparison:

DeviceCharacteristicPredicate DeviceProposed DeviceComparison
Product nameOrange Non Sterile Powder Free NitrileExamination GlovesDisposable Powder Free Nitrile ExaminationGlove, Pink ColorDisposable Powder Free Nitrile ExaminationGlove, Black ColorN/A
510(K) No.K172642K190942N/A
Product OwnerCentral Medicare Sdn. Bhd.Ever Growth Enterprise CorporationDifferent
Product CodeLZALZAsame
Regulation21 CFR 880.625021 CFR 880.6250same
ClassIIsame
Intended UseOrange Non Sterile Powder Free NitrileExamination Gloves is a disposable deviceThe Nitrile Powder Free patient examinationglove is a non-sterile disposable devicesimilar
intended for medical purposes that is wornon the examiner's hands to preventcontamination between patient andexaminer.intended for medical purposes that is wornon the examiner's hands or finger toprevent contamination between patient andexaminer.
Power freeYesYessame
SizeX Small/ Small/ Medium/Large/X LargeX Small/ Small/ Medium/Large/X Largesimilar
Single UseYESYESsame
Non-SterileYESYESsame
Dimensions-LengthComplies with ASTM D6319-10230 mm min.Complies with ASTM D6319-10230 mm min.same
Dimensions -Palm WidthComplies with ASTM D6319-10X Small70±10Complies with ASTM D6319-10X Small70±10similar
Small80 ±10Small80 ±10
Medium95±10Medium95±10
Large110 ±10Large110 ±10
X large120 ±10X large120 ±10
Dimensions -ThicknessComplies with ASTM D6319-10Palm - 0.05mm min.Finger - 0.05 mm min.Complies with ASTM D6319-10Palm - 0.05mm min.Finger - 0.05 mm min.Same
PhysicalPropertiesTensile Strength:Before Aging 14 MPa, min.After Aging 14 MPa, min.Elongation:Before Aging 500% min.After Aging 400% min.Tensile Strength:Before Aging 14 MPa, min.After Aging 14 MPa, min.Elongation:Before Aging 500% min.After Aging 400% min.same
ResidualpowderComplies with ASTM D6319-10Max. 0.50 mg per gloveComplies with ASTM D6319-10< 2mg per glovesame
Freedomfrom HolesIn accordance with ASTM D6319-10 andASTM D5151-06(reapproved 2011), G-1,AQL 2.5In accordance with ASTM D6319-10 andASTM D5151-06(reapproved 2011), G-1,AQL 2.5Same
Bio-compatibilityISO 10993-10Under the conditions of the study, not anirritant and sensitizerISO 10993-10Skin sensitization test & Skin Irritation test:PassesISO 10993-5In vitro cytotoxicity test:Passessimilar

{6}------------------------------------------------

11.0 Conclusion:

The conclusions drawn from the nonclinical tests demonstrate that the proposed device is as safe, as effective, and performs as well as or better than the legally marketed predicated device.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.