K Number
K190871

Validate with FDA (Live)

Date Cleared
2020-01-17

(289 days)

Product Code
Regulation Number
872.3640
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Bio-TackS Orthodontic Mini Implant is indicated for use as a fixed anchorage point for attachment of orthodontic appliances to facilitate the orthodontic movement of teeth. It is used temporarily and is removed after orthodontic treatment has been completed. Screws are intended for single use only.

Device Description

Fixed on jawbone the Bio-TackS Orthodontic Mini Implant is a screw-shaped product used as a fixed point for orthodontic treatment and it is applied in order to secure sufficient physical and mechanical fixing source for treatment.

AI/ML Overview

The provided text describes the Bio-TackS Orthodontic Mini Implant and its substantial equivalence to a predicate device, focusing on non-clinical performance and biocompatibility. It explicitly states that clinical testing was not required (Section 9, page 8). Therefore, the information requested about acceptance criteria and studies proving device performance with human data (e.g., sample size for test set, experts for ground truth, MRMC study, standalone performance, ground truth type for test/training sets) is not available within this document.

The document primarily focuses on demonstrating substantial equivalence through bench testing (non-clinical performance tests) and biocompatibility testing.

Here's the available information relevant to the acceptance criteria and the study that proves the device meets them, based on non-clinical data:

1. A table of acceptance criteria and the reported device performance:

The document doesn't explicitly state quantitative acceptance criteria for the "performance tests" in a table format. Instead, it describes performing tests according to a standard (ASTM F543) and comparing the subject device's results to the predicate device, concluding "similar performance."

Reported Device Performance (from Section 8, page 6-7):

TestAcceptance Criteria (Implicit: Similar to Predicate)Reported Device Performance
Sterilization and Shelf LifeSterilization validation per ISO 11137-1, 11137-2, 11137-3 (VDmax25 method). Shelf life demonstrated by accelerated aging with evaluations for seal peeling (ASTM F88), dye penetration (ASTM F1929), and sterility testing.Sterilization validation performed (ISO 11137-1, -2, -3, VDmax25 method). Shelf-life demonstrated by accelerated aging with seal peeling, dye penetration, and sterility testing.
Insertion-Removal Torque TestMechanical stability similar to predicate device (evaluated at 1440 degrees at 1 rpm).Both samples (BioTackS BSH1810S and Osstem OSSH1810) demonstrated "almost equivalent torque values" when inserted and removed.
Rotational Fracture Torque Strength TestMaterial yield strength similar to predicate device."The tapered implant fracture occurred in the same tread section." Both implants made from the same raw material (titanium alloy-ELI grade 23), indicating "material yield strengths of both products are the same." All screw-head types tested, results similar.
Axial Pull-Out Strength TestingEffectiveness of mechanical design and elimination similar to predicate device."The test data demonstrated similar results due to the similar pitch tread thickness of the samples" (BioTackS BSH1406S and Osstem OSSH1406).
Biocompatibility - CytotoxicityNon-cytotoxic (per ISO 10993-5)Non-cytotoxic
Biocompatibility - Acute Systemic ToxicityNon-acute systemic toxicity (per ISO 10993-11)Non-acute systemic toxicity
Biocompatibility - Intracutaneous (Intradermal) ReactivityNo irritation (per ISO 10993-10)None Irritation
Biocompatibility - Local Lymph Node Assay (LLNA)No hypersensitivity (per ISO 10993-10)Do not show any hypersensitivity

2. Sample size used for the test set and the data provenance:

  • Bench Tests (Performance Tests): For each specific test (Insertion-Removal torque, Rotational fracture torque, Axial pull-out), the document mentions testing specific models of the Bio-TackS and Osstem implants (e.g., BSH1810S, OSSH1810 for insertion-removal torque). It does not specify the exact number of samples tested for each condition but implies at least one sample of the "worst-case" or specific diameter/length combinations.
  • Biocompatibility Tests: Sample sizes are not explicitly stated for these tests (e.g., number of cells for cytotoxicity, number of animals for toxicity/reactivity tests).
  • Data Provenance: The tests are explicitly non-clinical bench and laboratory tests. No country of origin for the data is specified, but the applicant (BIO CETEC CO., LTD.) is from the Republic of Korea.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

This is not applicable as no clinical test set requiring human expert ground truth was conducted. The "ground truth" for the non-clinical tests is based on objective measurements against established standards (e.g., ASTM F543, ISO 11137, ISO 10993 series).

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

Not applicable, as no clinical test set requiring adjudication was conducted.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

Not applicable, as no clinical study (MRMC or otherwise) involving human readers or AI assistance was conducted.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

Not applicable, as this device (an orthodontic mini implant) is a physical medical device, not an algorithm or AI.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

For the non-clinical performance tests, the "ground truth" is measured physical and mechanical properties against established engineering and material science standards (ASTM F543), and for biocompatibility, it is biological response as assessed by ISO 10993 standards and laboratory assays.

8. The sample size for the training set:

Not applicable. This is a physical device, and the evaluation did not involve machine learning or AI models with training sets.

9. How the ground truth for the training set was established:

Not applicable (as above).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

January 17, 2020

BIO CETEC CO., LTD. % Dave Kim Medical Device Regulatory Affairs MTech Group Inc. 8310 Buffalo Speedway Houston, Texas 77025

Re: K190871

Trade/Device Name: Bio-TackS Orthodontic Mini Implant Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: OAT Dated: December 17, 2019 Received: December 17, 2019

Dear Dave Kim:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

for

Srinivas Nandkumar, Ph.D. Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K190871

Device Name Bio-TackS Orthodontic Mini Implant

Indications for Use (Describe)

The Bio-TackS Orthodontic Mini Implant is indicated for use as a fixed anchorage point for attachment of orthodontic appliances to facilitate the orthodontic movement of teeth. It is used temporarily and is removed after orthodontic treatment has been completed. Screws are intended for single use only.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K190871 510(k) Summary

This summary of 510(k) is being submitted in accordance with requirements of 21 CFR Part 807.92

Date: January 17, 2020

1. 510K Applicant / Submitter:

BIO CETEC CO., LTD. RM 1001-1002 New T Castle, 108, Gasan digital 2-ro, Geumcheon-gu, Seoul, REPUBLIC OF KOREA Tel: +82-2-704-2876 Fax: +82-2-704-2877

2. Submission Contact Person

MTech Group Inc. 7707 Fannin St. Ste 200, V111, Houston, TX 77054 Mr. Dave Kim Phone: 713-467-2607 Email: davekim@mtech-inc.net

3. Device

  • Trade / Device Name : Bio-TackS Orthodontic Mini Implant
  • Classification Name : Implant, Endosseous, Orthodontic
  • Regulation Number : 21 CFR 872.3640
  • Regulation Name : Endosseous dental implant
  • Regulatory Class : II
  • Product Code : OAT

4. Predicate Device

  • Trade / Device Name : Orthodontic Screw
  • 510(k) Number : K161197
  • Regulation Number : 21 CFR 872.3640
  • Regulation Name : Endosseous dental implant
  • Regulatory Class : II
  • Product Code : OAT

5. Description:

Fixed on jawbone the Bio-TackS Orthodontic Mini Implant is a screw-shaped product used as a fixed point for orthodontic treatment and it is applied in order to secure sufficient physical and mechanical fixing source for treatment.

6. Indications for Use

The Bio-TackS Orthodontic Mini Implant is indicated for use as a fixed anchorage point for attachment of orthodontic appliances to facilitate the orthodontic movement of teeth. It is used temporarily and is removed after orthodontic treatment has been completed. Screws are intended for single use only.

{4}------------------------------------------------

7. Substantial Equivalence Discussion:

The Bio-TackS Orthodontic mini implant is substantially equivalent to Osstem Orthodontic Screw (K161197). The following comparison table is presented to demonstrate substantial equivalence.

Subject DevicePredicate 1SEAnalysis
510(k) NumberK190871K161197-
Device NameBio-TackS Orthodontic Mini ImplantOrthodontic Screw-
Common NameOrthodontic mini implantOrthodontic mini implant-
ManufacturerBIOCETEC CO., LTD.OSSTEM Implant Co., Ltd.-
DesignDetails refer to below Table(*)--
Indication for UseThe Bio-TackS Orthodontic mini implant is indicated foruse as a fixed anchorage point for attachment oforthodontic appliances to facilitate the orthodonticmovement of teeth. It is used temporarily and is removedafter orthodontic treatment has been completed. Screws areintended for single use only.The Orthodontic Screw is indicated for useas a fixed anchorage point for attachment oforthodontic appliances to facilitate theorthodontic movement of teeth. It is usedtemporarily and is removed afterorthodontic treatment has been completed.Screws are intended for single use only.SE
Head StructureSimple Head, Through Hole, Small Head, Bracket Head,Hex Head, Hex Cross, Hex Open Cross, Simple Head HexStopper, Small Head Hex Stopper, Through Hole StopperSimple Head, Through Hole, Small Head,Bracket HeadDifferent
Body DiameterØ 1.4mm,Ø 1.6mm, Ø 1.8mmØ1.2mm, Ø 1.4mm,Ø 1.6mm, Ø 1.8mmSE
Length6mm, 8mm, 10mm6mm, 8mm, 10mmSE
Material of FixtureTi-6Al-4V ELI[ASTM F136]Ti-6Al-4V ELI[ASTM F136]SE
Surface TreatmentMachinedMachinedSE
SterilizationRadiationRadiationSE
Shelf-life5 years8 years-
Target PopulationProfessional use only - qualified dentists. Strictly reservedto specialized and trained users.Professional use only - qualified dentists.Strictly reserved to specialized and trainedusers.SE
Principles ofoperationOrthodontic screw is inserted into either jaw to help theorthodontist move the correct teeth and stop the wrongteeth from moving in the wrong direction.Orthodontic screw is inserted into either jawto help the orthodontist move the correctteeth and stop the wrong teeth from movingin the wrong direction.SE
Single useYesYesSE
Sterile PackagingYesYesSE
Target PopulationPatients in need of Dental orthodontics, maxillofacialsurgeryPatients in need of Dental orthodontics,maxillofacial surgerySE
Anatomical SiteTeethTeethSE
Location of UseUse only by professional orthodontistsUse only by professional orthodontistsSE
Bio-compatibilityAll user directly contacting materials are compliance withISO10993 requirements.All user directly contacting materials arecompliance with ISO10993 requirements.SE

(※) Images(Photos) of each type

{5}------------------------------------------------

NoTypeExternal Appearance
1Orthodontic Screw Simple HeadtypeImage: Orthodontic Screw Simple Head
2Orthodontic Screw Through HoletypeImage: Orthodontic Screw Through Hole
3Orthodontic Screw Small headtypeImage: Orthodontic Screw Small head
4Orthodontic Screw Bracket HeadtypeImage: Orthodontic Screw Bracket Head
5Orthodontic Screw Hex HeadtypeImage: Orthodontic Screw Hex Head
6Orthodontic Screw Hex CrosstypeImage: Orthodontic Screw Hex Cross

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows a close-up of an orthodontic screw. The screw has a hex open cross type head. The number 7 is also present in the image, likely indicating a product or item number.

Diameter-Length Combinations (in mm) for the Proposed Device:

Ø 1.6mm, Ø 1.8mm

    1. 1.4x6. 1.4x8
    1. 1.6x6, 1.6x8, 1.6x10
    1. 1.8x6. 1.8x8. 1.8x10

Bio-TackS Orthodontic Mini Implant and the predicate device (Osstem Orthodontic Screw (K161197)) have identical indication for use statements and the same intended use.

It shows equivalent specifications with the predicate devices in most of parameters.

However, there are differences in the device design characteristics (Head structure).
Bio-TackSOrthodontic Screw (K161197)
Body DiameterØ 1.4mm.011 1 1 1 0 1 0 1 0 1 1 0 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1Ø1.2mm. Ø 1.4mm.011 1 1 1 0 1 0 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1

The different head structure for the subject device include Hex Head, Hex Cross, Hex Open Cross, Simple Head Hex Stopper, Small Head Hex Stopper, Through Hole Stopper.

Ø 1.6mm, Ø 1.8mm

8. Performance Tests (Non-clinical)

Non-clinical performance tests were performed according to the test standard, ASTM F543 Standard Specification and Test Methods for Metallic Medical Bone Screws.

The following tests for performance of the subject device have been conducted;

Sterilization and Shelf Life:

Sterilization validation was performed according to ISO 11137-1. ISO 11137-2, and ISO 11137-3 using the VDmax25 method. Shelf life of the packaging was demonstrated by accelerated aging of the product with evaluations for seal peeling (ASTM F88), dye penetration (ASTM F1929), and sterility testing.

Insertion-Removal torque test

According to the test standard (ASTM F543-07 A1), Insertion torque / removal torque testing is performed to evaluate the mechanical stability of orthodontic mini implants. The largest diameter (1.8mm) and longest (10mm) screws are tested as the worst case; BSH1810S (BioTackS), OSSH1810 (Osstem Mini Implant). With the same diameter and length, both samples demonstrated almost equivalent torque values when inserted and removed at 1440 degrees at 1 rpm. A slight difference in the results is due to the different in drill-bit design and thread design.

Rotational fracture torque strength test

According to the test standard (ASTM F543-07 A1), Rotational fracture torque testing is performed to determine the material yield strength. The smallest diameter (1.4mm) with a 10mm length; BSH1408S (BioTacks), OSSH1408 (Osstem mini implant) were tested. The tapered implant fracture occurred in the same tread section. Both implants are made from the same raw material (titanium alloy-ELI grade 23), so the material yield strengths of both products are the same.

{7}------------------------------------------------

In addition, all screw-head types, with the same diameter and length, are tested.

Axial pull-out strength testing

According to the test standard (ASTM F543-07 A1), Pull-out strength testing is performed for the effectiveness of the mechanical design of the orthodontic mini implant and its elimination. The screw samples with the shortest (6mm) and smallest diameter (1.4mm) were tested: BSH1406S (BioTackS). OSSH1406 (Osstem mini implant). The test data demonstrated similar results due to the similar pitch tread thickness of the samples.

TestTypeDiameterLengthShape
Insertion-Removal torque testingBSH1810S(BioTackS)1.8 mm10 mmTaper
OSSH1810(Osstem)1.8 mm10 mmTaper
Rotational fracture torque testingBSH1408S(BioTackS)1.4 mm8 mmTaper
OSSH1408(Osstem)1.4 mm8 mmTaper
Axial Pull-out strength testingBSH1406S(BioTackS)1.4 mm6 mmTaper
OSSH1406(Osstem)1.4 mm6 mmTaper

Table 1. Description of orthodontic mini implant Used in This Study

Bench Test included rotational fracture torque strength testing for all screw-head types, with the same diameter and length. The test results were similar. Differences of Head structure do not affect effectiveness or safety of the Bio-TackS Orthodontic Mini Implant and Predicate Device.

The results of the bench testing demonstrate similar performance for Bio-TackS Orthodontic Mini Implant and the predicate device.

Biocompatibility testing accordance with ISO 10993-1 has been conducted for Bio-TackS Orthodontic Mini Implant.

  • Category : Implant Device for Tissue / Bone
  • Contact duration: C-permanent (>30d)
HumanContact PartTest ItemTest Report NumberTest StandardTest Result
Tissue / BoneCytotoxicity180600062 PPT-18-0028-01(E)ISO 10993-5Non-cytotoxic
Acute Systemic Toxicity180600062 PPT-18-0028-02(E)ISO 10993-11Non-acute systemic toxicity
Intracutaneous(Intradermal) Reactivity180600062 PPT-18-0028-03(E)ISO 10993-10None Irritation
Local Lymph NodeAssay (LLNA)180600062 PPT-18-0028-04(E)ISO 10993-10Do not show any hypersensitivity

Biocompatibility testing including Cytotoxicity, Sensitization or Intracutaneous Reactivity, Systemic Toxicity was completed according to the following standards:

ISO 10993-1:2009 Biological Evaluation of Medical Devices -Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process

ISO 10993-5:2009 Biological Evaluation of Medical Devices – Part 5 Cytotoxicity

ISO 10993-10:2010 Biological Evaluation of Medical Devices Part 10: Tests for irritation and skin sensitization

ISO 10993-11:2017 Biological Evaluation of Medical Devices Part 11: Tests for systemic toxicity test

{8}------------------------------------------------

The biocompatibility test results demonstrated that there is no new concern in the cytotoxicity, acute systemic toxicity, intracutaneous reactivity, local lymph node assay (LLNA).

Subchronic Toxicity Test, Genotoxicity Test, Implantation were not conducted due to the reasons as follows;

Products with titanium materials tested according to the contents of ASTM F 136, X2. BIOCOMPATIBILITY in Standard Specification for Wrought Titanium-6Aluminum-4Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant Applications of ASTM F136 do not have side effects on safety after a procedure.

X2. BIOCOMPATIBILITY

  • X2.1
    The alloy composition covered by this specification has been employed successfully in human implant annlications in contact with soft tissue and bone for over a decade. Due to the well-characterized level of biological response exhibited by this alloy, it has been used as a control material in Practice F 981.

SUMMARY OF CLINICAL TESTS 9.

Clinical testing was not required to demonstrate the substantial equivalence of the Bio-TackS Orthodontic Mini Implant to its predicate device.

10. Conclusions:

Based on the information above, Bio-TackS Orthodontic Mini Implant has same indications for use and intended use. Both the subject device and the predicate / reference devices have shown similar performance results in these bench tests

In conclusion, Bio-TackS Orthodontic Mini Implant is substantially equivalent to the predicate device as described herein.

§ 872.3640 Endosseous dental implant.

(a)
Identification. An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.(b)
Classification. (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
Classification. Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.