K Number
K190747

Validate with FDA (Live)

Device Name
Claritag
Manufacturer
Date Cleared
2019-10-31

(223 days)

Product Code
Regulation Number
878.4350
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Claritag is indicated for use in the treatment of achrochordons (skin tags).

Device Description

The Claritag device is indicated for the treatment of skin tags. Claritag® is a portable, hand operated device intended for use in the treatment of skin tag removal using a cryogen application system. The cryogen application methodology for skin tag removal is a widely accepted practice used by physicians for decades.

The Claritag device is provided in a kit containing the main device (canister containing aerosolized cryogen, fitted with two treatment head activation base, 20 disposal single use foam pads, tweezers (which are provided as an optional accessory) and Instructions for Use.

The mechanism of action for cryotherapy are the direct effects of freezing on the cells, and the vascular stasis which develops in the tissue after thawing.

AI/ML Overview

The provided text is a 510(k) summary for the Claritag device, a cryosurgical unit for treating skin tags. It primarily focuses on demonstrating substantial equivalence to a predicate device based on technological characteristics and non-clinical performance data.

However, the document does not contain information about acceptance criteria or a study proving the device meets those criteria for clinical performance (e.g., efficacy or safety in patients). The "Performance Data" section (page 4) exclusively details non-clinical performance testing, such as temperature measurements, depth of penetration in ballistic gel, shelf life, and package performance. These tests are primarily designed to show that the device functions as intended and safely, but not to measure its clinical effectiveness or human safety outcomes directly.

Therefore, I cannot fulfill your request for the following sections based on the provided text:

  1. A table of acceptance criteria and the reported device performance: No clinical acceptance criteria are mentioned, nor is clinical performance reported.
  2. Sample sizes used for the test set and the data provenance: No clinical test set data is presented.
  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: No ground truth establishment for a clinical test set is mentioned.
  4. Adjudication method for the test set: Not applicable as no clinical test set is discussed.
  5. If a multi reader multi case (MRMC) comparative effectiveness study was done: No MRMC study is mentioned.
  6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable, as this is a physical medical device, not an AI algorithm.
  7. The type of ground truth used (expert consensus, pathology, outcomes data, etc): No clinical ground truth is discussed.
  8. The sample size for the training set: Not applicable; this is a physical device, not a machine learning model with a training set.
  9. How the ground truth for the training set was established: Not applicable.

Summary of available information regarding performance:

The document reports on non-clinical performance data:

  • Claritag Predicate Comparison: Temperature of foam pads for both Claritag and the predicate device were measured ten times using a simulated skin model. Both met the minimum temperature requirement of -20°C.
  • Claritag Ballistic Gel Test: Depth of temperature penetration was measured at 1, 2, and 3 mm from the outer face of ballistic gelatin to confirm cryogenic gas freezing effect decreases with depth.
  • Shelf Life: Real-time data supports a one-year shelf life.
  • Package Performance Testing: Passed ISTA 6 -Fed Ex-A.
  • Claritag & Predicate Pad Temperature Comparison Test: Both achieved an average minimum temperature of -20°C.

Conclusion stated regarding performance:

The non-clinical performance data demonstrate that the Claritag device performs comparably to the predicate device. This data, along with comparison of technological characteristics, supports the substantial equivalence determination.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "U.S. Food & Drug Administration" in blue.

October 31, 2019

DGI Technologies % Russ Olsen VP Global OA/RA Health Policy Associates 690 Canton Street. Suite 302 Westwood, Massachusetts 02090

Re: K190747

Trade/Device Name: Claritag Regulation Number: 21 CFR 878.4350 Regulation Name: Cryosurgical Unit and Accessories Regulatory Class: Class II Product Code: GEH Dated: March 20, 2019 Received: March 22, 2019

Dear Russ Olsen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K190747

Device Name Claritag

Indications for Use (Describe)

Claritag is indicated for use in the treatment of achrochordons (skin tags).

Type of Use (Select one or both, as applicable)
-------------------------------------------------
☑ Registration Under Section 301, Sales, Distribution☐ Over-The-Counter Health/OTC Medical Device
----------------------------------------------------------------------------------------------------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

Prepared in accordance with 21 CFR 807.92

GENERAL INFORMATION

510(k) owner name address and telephone:

DGI Technologies 301 Madison Avenue Lakewood, NJ 08701 USA Telephone: 732-887-5087

Name of contact person info:

Russ Olsen Health Policy Associates, Inc. 690 Canton Street, Suite 302 Westwood, MA 02090 Telephone: 781.329.2993 Fax: 781 329.2968 rolsen@healthpolicyassociates.com

Date Prepared:

October 30, 2019

DEVICE INFORMATION:

Trade Name: Common Name: Classification Name: Regulation Number: Regulatory Class: Product Code:

Claritag® Portable aerosol cryosurgery device Cryosurgical Unit and Accessories 21 CFR 878.4350 Class II GEH

PREDICATE DEVICE INFORMATON:

Trade Name: Common Name: 510(k) Number: Decision Date:

Cool Renewal Cryosurgical Unit and Accessories K161296 February 9, 2011

DEVICE DESCRIPTION:

The Claritag device is indicated for the treatment of skin tags. Claritag® is a portable, hand operated device intended for use in the treatment of skin tag removal using a cryogen application system. The cryogen application methodology for skin tag removal is a widely accepted practice used by physicians for decades.

The Claritag device is provided in a kit containing the main device (canister containing aerosolized cryogen, fitted with two treatment head activation base, 20 disposal single use foam pads, tweezers (which are provided as an optional accessory) and Instructions for Use.

The mechanism of action for cryotherapy are the direct effects of freezing on the cells, and the vascular stasis which develops in the tissue after thawing.

{4}------------------------------------------------

INTENDED USE OF CLARITAG

The Claritaq device is indicated for use in the treatment of acrochordons (skin tags).

COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

The intended use and principles of operation of the Claritag device are substantially equivalent to the predicate device. Both devices incorporate the same basic cryosurgical design for skin tag removal, i.e., two foam pads treated with cryogen that enclose the skin tag and, after brief exposure, freezes the skin tag, The skin tag then thaws, and ultimately the skin tag tissue is destroyed. Both the Claritag device and the predicate achieve the minimum freeze temperature of -20°C, the widely accepted temperature required to destroy skin tags.

The differences between the Claritag device and the predicate are related to the cryogenic gas, and minor design elements. The Claritag device uses an alternative gas, which is more environmentally friendly versus the predicate device's cryogen gas. The Claritag device incorporates a dual head design with foam pads which wrap around, squeeze and freeze the skin tag. The predicate device uses "foam tipped skin tag tweezers" to wrap around, squeeze and freeze the skin tag. Any other differences are considered to be minor in nature and do not raise any new questions about safety and effectiveness.

PERFORMANCE DATA

The following non-clinical performance data are provided in support of the substantial equivalence determination:

Biocompatibility:

  • Cytotoxicity o
  • o Sensitization
  • O Irritation

Performance testing

  • Claritag Predicate Comparison: The temperature of the foam pads of the Claritag device and o the predicate device were measured ten times using a simulated skin model. Both Claritag device and predicate device measurements met the minimum temperature requirement of -20 °C.
  • Claritag Ballistic Gel Test: Depth of temperature penetration of the Claritan device was o measured at 1, 2 and 3 mm from the outer face of the ballistic gelatin. This test confirmed that the Claritag device's cryogenic gas freezing effect decreases with depth and demonstrates the safety and effectiveness of the Claritag in the treatment of acrochordons (skin tags).
  • Shelf Life: The Claritag device has real time data that represents a one-year shelf life o consistent with the predicate device.
  • O Package Performance Testing for Claritag device passed package performance testing pursuant to ISTA 6 -Fed Ex-A.
  • Claritag & Predicate Pad Temperature Comparison Test: Temperature testing of the Claritag O device and the predicate device both achieved an average minimum temperature of -20 °C.

Conclusion:

The non-clinical performance data demonstrate that the Claritag device performs comparably to the predicate device that is currently marketed for the same intended use. Based upon the information submitted in this Traditional 510(k) premarket notification, the Claritag device is substantially equivalent to the Cool Renewal (K161296).

§ 878.4350 Cryosurgical unit and accessories.

(a)
Identification —(1)Cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories. A cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold.(2)
Cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories. A cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures, including urological applications, by applying extreme cold.(3)
Cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories. A cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. The device is intended to treat disease conditions such as tumors, skin cancers, acne scars, or hemangiomas (benign tumors consisting of newly formed blood vessels) and various benign or malignant gynecological conditions affecting vulvar, vaginal, or cervical tissue. The device is not intended for urological applications.(b)
Classification. Class II.