K Number
K190719
Device Name
Artemis Eye System
Manufacturer
Date Cleared
2019-08-30

(163 days)

Product Code
Regulation Number
882.1480
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The Artemis™ Eye System is indicated to provide visualization and illumination of intracranial tissue and fluids during diagnostic and therapeutic procedures.
Device Description
The Artemis Eye System is indicated to provide visualization and illumination of intracranial tissue and fluids during diagnostic and therapeutic procedures. The System consists of two components: - Artemis Eye a single-use neuro-endoscope consisting of a camera, two pathways for . irrigation and/or drain, and a working channel for a surgical tool. - Artemis Eye Tablet a reusable component that connects to the Artemis Eye and . displays live imaging captured by the Artemis Eye camera.
More Information

No
The description focuses on visualization and illumination, and the performance studies are standard for medical devices, not indicative of AI/ML. There are no mentions of AI, ML, or related concepts.

No.
The Artemis™ Eye System is indicated for visualization and illumination during diagnostic and therapeutic procedures, but it does not directly provide therapy itself. It is a tool for a healthcare professional to perform diagnostic or therapeutic procedures.

No
The device provides visualization and illumination during diagnostic and therapeutic procedures, but it does not perform the diagnosis itself. It is a tool for a clinician to use in those procedures.

No

The device description clearly states the system consists of two components: a single-use neuro-endoscope (Artemis Eye) and a reusable tablet (Artemis Eye Tablet). The neuro-endoscope is a hardware component containing a camera and other physical elements. The performance studies also include hardware-related testing such as sterilization, biocompatibility, bench-top performance, shelf-life, and electrical safety/EMC testing, in addition to software verification and validation.

Based on the provided information, the Artemis™ Eye System is not an IVD (In Vitro Diagnostic) device.

Here's why:

  • Intended Use: The intended use is "to provide visualization and illumination of intracranial tissue and fluids during diagnostic and therapeutic procedures." This describes a device used during a procedure to aid the clinician, not a device that analyzes samples in vitro (outside the body) to diagnose a condition.
  • Device Description: The description details a neuro-endoscope and a tablet for displaying live imaging. This aligns with a visualization tool used during surgery or diagnostic procedures, not a device for analyzing biological samples.
  • Lack of IVD Characteristics: There is no mention of analyzing biological samples (blood, urine, tissue samples, etc.), performing tests on these samples, or providing diagnostic information based on such analysis.

IVD devices are specifically designed to examine specimens derived from the human body to provide information for the diagnosis, prevention, monitoring, treatment, or alleviation of disease. The Artemis Eye System's function is to provide visual access to an anatomical site in vivo.

N/A

Intended Use / Indications for Use

The Artemis™ Eye System is indicated to provide visualization and illumination of intracranial tissue and fluids during diagnostic and therapeutic procedures.

Product codes (comma separated list FDA assigned to the subject device)

GWG

Device Description

The Artemis Eye System is indicated to provide visualization and illumination of intracranial tissue and fluids during diagnostic and therapeutic procedures. The System consists of two components:

  • Artemis Eye a single-use neuro-endoscope consisting of a camera, two pathways for . irrigation and/or drain, and a working channel for a surgical tool.
  • Artemis Eye Tablet a reusable component that connects to the Artemis Eye and . displays live imaging captured by the Artemis Eye camera.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

intracranial tissue and fluids

Indicated Patient Age Range

Not Found

Intended User / Care Setting

Not Found

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Performance testing was conducted to evaluate the physical and mechanical properties of the Artemis Eye System and to demonstrate substantial equivalence to the predicate device. The following tests were performed and all tests passed successfully:

  • Dimensional / visual inspection ●
  • . Design Feature Testing
  • . Simulated Use
  • Destructive Testing .

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.

K002572

Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.

K162475

Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).

Not Found

§ 882.1480 Neurological endoscope.

(a)
Identification. A neurological endoscope is an instrument with a light source used to view the inside of the ventricles of the brain.(b)
Classification. Class II (performance standards).

0

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

August 30, 2019

Penumbra, Inc. Aditi Kolla Regulatory Affairs Specialist One Penumbra Place Alameda, California 94502

Re: K190719

Trade/Device Name: Artemis Eye System Regulation Number: 21 CFR 882.1480 Regulation Name: Neurological Endoscope Regulatory Class: Class II Product Code: GWG Dated: July 30, 2019 Received: August 1, 2019

Dear Aditi Kolla:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

1

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Matthew Krueger Assistant Director DHT5A: Division of Neurosurgical. Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

2

Indications for Use

510(k) Number (if known) K190719

Device Name Artemis™ Eye System

Indications for Use (Describe)

The Artemis™ Eye System is indicated to provide visualization and illumination of intracranial tissue and fluids during diagnostic and therapeutic procedures.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

3

Image /page/3/Picture/1 description: The image shows the word "Penumbra" in a bold, red font. To the right of the word is a red circle with a white "P" inside. The logo is simple and modern, with a focus on the company name and a stylized initial.

1 510(k) Summary

(as required by 21 CFR 807.92)

Pursuant to Section 12, Part (a)(i)(3A) of the Safe Medical Devices Act of 1990,

Penumbra, Inc. is providing the summary of Substantial Equivalence for the Artemis™ Eye System.

1.1 Sponsor/Applicant Name and Address

Penumbra, Inc. One Penumbra Place Alameda, CA 94502 USA

1.2 Sponsor Contact Information

Aditi Kolla Regulatory Affairs Specialist Phone: (510) 995-2010 FAX: (510) 217-6414 Email: akolla@penumbrainc.com

Date of Preparation of 510(k) Summary 1.3

August 29, 2019

Device Trade or Proprietary Name 1.4

Artemis™ Eye System

Primary Device Classification 1.5

Regulatory Class: II Classification Panel: Neurology Common Name: Neurological Endoscope Classification Name: Endoscope, neurological Regulation Number: 21 CFR 882.1480 Product Code: GWG

1.6 Predicate & Reference Devices

| 510(k)
Number | Clearance Date | Device | Name of
Manufacturer |
|------------------|----------------------|--------------------------------------------------------|-------------------------|
| Predicate Device | | | |
| K002572 | November 16,
2000 | Medtronic High
Resolution Channel
Neuroendoscope | Medtronic PS Medical |

4

Image /page/4/Picture/1 description: The image shows the logo for Penumbra, a healthcare company. The word "Penumbra" is written in a red, sans-serif font. To the right of the word is a red circle with a white "P" inside. The logo is simple and modern.

| 510(k)
Number | Clearance Date | Device | Name of
Manufacturer |
|------------------|-----------------------|-------------------------------------|-------------------------|
| K162475 | September 16,
2016 | Trice Mi-eye 2, Mi-eye 2
Monitor | Trice Medical Inc. |

1.7 Predicate Comparison

Table 1: Predicate Device Comparison
--------------------------------------------
AttributePredicate DeviceSubject Device
Trade NameMedtronic High Resolution
Channel NeuroendoscopePenumbra Artemis Eye System
510(k) No.K002572To be determined
ClassificationClass II, GWG (882.1480)Class II, GWG (882.1480)
Indications For UseThe High Resolution Channel
Neuro-endoscope is indicated for
diagnostic and intraoperative
procedures where the physician
desires direct vision of intracranial
tissue where cerebrospinal fluid
(CSF) may be contacted.The Artemis Eye System is
indicated to provide visualization
and illumination of intracranial
tissue and fluids during diagnostic
and therapeutic procedures.
Intended UseIntended for visualization of
intracranial tissue during diagnostic
and intraoperative procedures.SAME
PortabilityHand-HeldSAME
WeightNot published in the 510(k)
Summary≤ 300 g
Video OutputLCD monitorSAME
Illumination Light
Source1 mm CAMLite Light CableSurface Mount LED
Neuroendoscope Probe
MaterialStainless SteelSAME
Working Length13 cm or 21.6 cm15 cm
Total Length17.2 cm or 25.8 cm25 cm
Working Tube/Probe
OD4.2 mm6 mm
Working Channel
Diameter2.15 mm3.1 mm
Camera ResolutionHigh (30,000 optical fibres) or
Standard (10,000 optical fibres)200 pixels x 200 pixels

5

Image /page/5/Picture/1 description: The image shows the word "Penumbra" in a red font, followed by a red circle with a white "P" inside. The word "Penumbra" is written in a sans-serif font. The "P" in the circle is also in a sans-serif font. The logo is likely for a company or organization called Penumbra.

Table 1: Predicate Device Comparison
AttributePredicate DeviceSubject Device
Trade NameMedtronic High Resolution
Channel NeuroendoscopePenumbra Artemis Eye System
Video Output/ Tablet
DimensionsNot Applicable31.0 x 20.0 x 3.8 cm
(12.2 x 7.9 x 1.5 in.)
Display Screen
ResolutionNot Applicable≥ 1920 pixels x 1280 pixels
Video Output/ Tablet
Battery LifeNot Applicable3 hours (may vary based upon
usage)
AccessoriesPeelaway Introducer SheathStopcock
Neuro-Endoscope UseSingle-UseSAME
SterilizationEOSAME
Shelf-Life36-MonthsArtemis Eye: 12-Month data
currently available

Device Description 1.8

The Artemis Eye System is indicated to provide visualization and illumination of intracranial tissue and fluids during diagnostic and therapeutic procedures. The System consists of two components:

  • Artemis Eye a single-use neuro-endoscope consisting of a camera, two pathways for . irrigation and/or drain, and a working channel for a surgical tool.
  • Artemis Eye Tablet a reusable component that connects to the Artemis Eye and . displays live imaging captured by the Artemis Eye camera.

1.9 Indications for Use

The Artemis™ Eye System is indicated to provide visualization and illumination of intracranial tissue and fluids during diagnostic and therapeutic procedures.

1.10 Summary of Non-Clinical Data

Pursuant to Section 12, Part (a)(i)(3A) of the Safe Medical Devices Act of 1990, a summary of any information regarding substantial equivalence of the devices follows.

Included in this section is a summary description of the testing, which substantiates the performance of the subject Artemis Eye System.

6

Image /page/6/Picture/1 description: The image shows the word "Penumbra" in a red font. To the right of the word is a red circle with a white "P" inside. The logo is simple and modern.

1.10.1 Sterilization

The subject Artemis Eye has proved to be sterile in accordance with EN ISO 11135:2014.

1.10.2 Biocompatibility Testing

Biocompatibility test studies were selected in accordance Guidance for Industry and FDA Staff: Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" (issued June 16, 2016). All studies were conducted pursuant to 21 CFR, Part 58, Good Laboratory Practices. The following tests were successfully conducted on the patient-contacting Artemis Eye component of the subject device:

TestMethodResults
In Vitro CytotoxicityISO Elution Test (MEM Extract)Non-Toxic
SensitizationMagnusson-Kligman MethodNon-Sensitizing
Irritation (Intracutaneous
Reactivity)ISO Intracutaneous (Intradermal)
Injection TestNon-Irritant
Systemic Toxicity (Acute)
Acute Systemic ToxicityISO Acute Systemic Injection TestNon-Toxic
Material Mediated
PyrogenUSP Material-Mediated Rabbit
Pyrogen TestNon-pyrogenic
Hemo-compatibility
In-Vitro HemolysisASTM Method (Indirect & Direct
Contact)Non-Hemolytic
CoagulationPT and PTT TestNon-Thrombogenic

1.10.3 Bench-Top Performance (Design Verification)

Performance testing was conducted to evaluate the physical and mechanical properties of the Artemis Eye System and to demonstrate substantial equivalence to the predicate device. The following tests were performed and all tests passed successfully:

  • Dimensional / visual inspection ●
  • . Design Feature Testing
  • . Simulated Use
  • Destructive Testing .

7

Image /page/7/Picture/1 description: The image shows the word "Penumbra" in red font. To the right of the word is a red circle with a white "P" inside. The "P" has a horizontal line through it. The logo is simple and modern.

1.10.4 Shelf-life

The physical and mechanical performance of the Artemis Eye was evaluated for 12 months through accelerated aging studies to demonstrate device stability for the duration of the labeled shelf-life. Test results confirm the subject device will maintain device performance for the entirety of the proposed shelf-life.

1.10.5 Software Verification & Validation

Software verification and validation testing and documentation for the Artemis Eye System was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" (issued May 11, 2005). The software for this device was considered as a "major" level of concern.

1.10.6 Electrical Safety/EMC Testing

Electrical safety and EMC testing were conducted on the Artemis Eye System. The system complies with the requirements of IEC/EN 60601-1, IEC/EN 60601-1-2, IEC 60601-1-6, and IEC/EN 62366.

1.11 Summary of Substantial Equivalence

The subject Artemis Eye System is substantially equivalent to the predicate device. The subject device has an identical intended use as the predicate device. The subject and the predicate devices differ slightly in regards to minor technological variations, while maintaining the same fundamental scientific technology. However, these differences do not raise different questions of safety and effectiveness.

The device testing described in the 510(k) Summary demonstrate the subject devices are substantially equivalent to the predicate device in regards to operating principle, fundamental technology and device performance. It also demonstrates the Artemis Eye System should perform as intended in the specified use conditions.