K Number
K190579

Validate with FDA (Live)

Date Cleared
2019-06-04

(90 days)

Product Code
Regulation Number
862.1345
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The TD-4183 Blood Glucose Monitoring System consists of the TD-4183 Blood Glucose meter and the TD-4183 Blood Glucose Test Strips.

The TD-4183 Blood Glucose Monitoring System is intended for use in the quantitative measurement of glucose in fresh capillary whole blood from the fingertip. This system is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes mellitus at home as an aid in monitoring the effectiveness of the diabetes control program. It is intended to be used by single person and should not be shared. It is not intended for the diagnosis of or screening for diabetes mellitus, and is not intended for use on neonates.

Device Description

The TD-4183 system kit includes the TD-4183 Blood Glucose meter with blood glucose measuring function and the TD-4183 Blood Glucose test strips. This system is a single-patient use blood glucose monitoring system intended to be used to quantitatively measure glucose in fresh capillary whole blood from the finger as an aid in monitoring the effectiveness of glucose control.

AI/ML Overview

The provided document is a 510(k) summary for the TaiDoc TD-4183 Blood Glucose Monitoring System. It describes the device, its intended use, and a comparison to a predicate device (TD-4277 Blood Glucose Monitoring System). It also briefly mentions non-clinical and clinical studies conducted to support the substantial equivalence claim.

However, the document does not contain the detailed acceptance criteria and the results of a study that proves the device meets those criteria in the format requested. Specifically:

  • No detailed table of acceptance criteria and reported device performance is provided. It only states that "all performance criteria were met."
  • No specific sample sizes for the test set or training set are given for any of the studies (non-clinical or clinical).
  • No information on data provenance (country of origin, retrospective/prospective) is provided.
  • No details are available regarding expert involvement (number of experts, qualifications) for ground truth establishment or adjudication methods.
  • No multi-reader multi-case (MRMC) comparative effectiveness study is mentioned, as this is a blood glucose monitoring system, not an imaging AI device where such studies are common.
  • No mention of a standalone algorithm-only performance study for the same reason.
  • No specific type of ground truth used is detailed, though for a blood glucose monitor, it would typically involve laboratory reference methods.
  • No information on how ground truth for the training set was established (as sensor calibration and performance is generally based on known glucose concentrations from laboratory standards).

The document is primarily a regulatory submission summary, focusing on demonstrating substantial equivalence to a predicate device rather than providing a detailed technical report of the studies performed.

Therefore, based on the provided text, I cannot complete the requested information in the specified format. The available information is too high-level for the detailed breakdown required by the prompt regarding acceptance criteria and study particulars.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.

June 4, 2019

TaiDoc Technology Corporation Sophia Wu Regulatory Affairs Vice President 6F. No.127. Wugong 2nd Rd., Wugu District New Taipei City, 24888 Taiwan

Re: K190579

Trade/Device Name: TD-4183 Blood Glucose Monitoring System Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose Test System Regulatory Class: Class II Product Code: NBW Dated: February 27, 2019 Received: March 6, 2019

Dear Sophia Wu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Kellie B. Kelm, Ph.D. Acting Director Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K190579

Device Name TD-4183 Blood Glucose Monitoring System

Indications for Use (Describe)

The TD-4183 Blood Glucose Monitoring System consists of the TD-4183 Blood Glucose meter and the TD-4183 Blood Glucose Test Strips.

The TD-4183 Blood Glucose Monitoring System is intended for use in the quantitative measurement of glucose in fresh capillary whole blood from the fingertip. This system is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes mellitus at home as an aid in monitoring the effectiveness of the diabetes control program. It is intended to be used by single person and should not be shared. It is not intended for the diagnosis of or screening for diabetes mellitus, and is not intended for use on neonates.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

| Prescription Use (Part 21 CFR 801 Subpart D)

|X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for TaiDoc, a company that specializes in medical devices. The logo consists of the word "TaiDoc" in gray letters, followed by three red hearts. Below the hearts is the company's website address, www.taidoc.com, also in gray.

510(k) Summary

Submission Number: K190579

In accordance with the requirements of 21 CFR 807.92, this summary is being provided to serve as the basis for the substantial equivalence determination.

Submitter information

ManufacturerTaiDoc Technology Corporation
Address6F, No. 127, Wugong 2nd Rd. Wugu Dist.New Taipei City, Taiwan 24888
Establishment Registration No.3004145393
Date PreparedMay 10, 2019
CorrespondentTaiDoc Technology Corporation
Correspondent ContactSophia Wu
TitleRegulatory Affairs Vice President
Phone+886-2-6625-8188

Proposed Device Information

Proprietary nameTD-4183 Blood Glucose Monitoring System
Common nameBlood Glucose Monitoring System
Product codeNBW, Blood Glucose Test System, Over-the-Counter
Classification panelClinical chemistry
Classification2
Regulation Number21 CFR §862.1345

Predicate Device Information

ManufacturerTaiDoc Technology Corporation
Proprietary NameTD-4277 Blood Glucose Monitoring System
Common NameBlood Glucose Monitoring System
510(k) NumberK100322

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for TaiDoc, a company that specializes in medical devices. The logo consists of the word "TaiDoc" in gray letters above three red hearts. Below the hearts is the company's website address, "www.taidoc.com," in white letters on a gray background.

Intended use

The TD-4183 Blood Glucose Monitoring System consists of the TD-4183 Blood Glucose meter and the TD-4183 Blood Glucose Test Strips.

The TD-4183 Blood Glucose Monitoring System is intended for use in the quantitative measurement of glucose in fresh capillary whole blood from the fingertip. This system is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes mellitus at home as an aid in monitoring the effectiveness of the diabetes control program. It is intended to be used by single person and should not be shared. It is not intended for the diagnosis of or screening for diabetes mellitus, and is not intended for use on neonates.

Device Description

The TD-4183 system kit includes the TD-4183 Blood Glucose meter with blood glucose measuring function and the TD-4183 Blood Glucose test strips. This system is a single-patient use blood glucose monitoring system intended to be used to quantitatively measure glucose in fresh capillary whole blood from the finger as an aid in monitoring the effectiveness of glucose control.

Test principle

The blood sample is applied to the absorbent hole at the end of the test strip to fill the window. When the blood sample is absorbed into the test strip, the blood glucose reacts with the reagent contained in the test strip and generates electrical current. When the meter receives the signal through the connector, the signal is then transferred into calculated blood glucose level and displayed on the LCD screen. The strength of the current produced by the reaction depends on the amount of glucose in the blood sample.

Summary of Technological Characteristics and Comparison to the Predicate

The TD-4183 Blood Glucose Monitoring System is substantially equivalent to the predicate device in terms of technological characteristics.

The similarities and differences between the predicate and proposed devices are summarized in Table 1 and 2 below.

CONFIDENTIAL

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for TaiDoc. The logo consists of the word "TaiDoc" in gray letters above three red hearts. Below the hearts is the website address "www.taidoc.com" in white letters on a gray background.

泰博科技股份有限公司 TaiDoc Technology Corp.

新北市24888五股區五工二路127號6樓 6F., No.127, Wugong 2nd Rd., Wugu Dist., New Taipei City 24888, Taiwan

Tel : +886-2-6625-8188 Fax : +886-2-6625-0288

Table 1: Similarities between the Predicate and Proposed Device
-------------------------------------------------------------------------
CharacteristicPredicate deviceTD-4277Proposed deviceTD-4183
Operation principleElectrochemical biosensortechnologythe same as the predicate device
Detection methodAmperometric glucose biosensorthe same as the predicate device
Mode of operationUser initiates test on meter, insertstest strip into meter, lances fingerwith lancing device, and places bloodsample on test strip.the same as the predicate device
Strip preparationStrips are stored in ahumidity-controlled vial (25-50 stripsper vial), user removes one strip fromvial and inserts in meter for testing.the same as the predicate device
Code calibrationNo coding requiredthe same as the predicate device
Strip enzymeFAD Glucose dehydrogenase ( E. coli )the same as the predicate device
Strip reaction time6 secondsthe same as the predicate device
Strip sample volume0.5 µLthe same as the predicate device
Measurement unitmg/dLthe same as the predicate device
Measurement modesAC (before meal)PC (after meal)Gen (not specified)QC (quality control)the same as the predicate device
User interfaceLCD and buttonsthe same as the predicate device
Display of test resultsLCD screenthe same as the predicate device
Software designMeter software controls themeasurement and calculation ofblood glucose and displays result onthe meter.the same as the predicate device

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for TaiDoc. The word "TaiDoc" is written in gray at the top of the image. Below the word are three red hearts. At the bottom of the image is the website address, www.taidoc.com, written in white on a gray background.

泰博科技股份有限公司 TaiDoc Technology Corp.

新北市24888五股區五工二路127號6樓 6F., No.127, Wugong 2nd Rd., Wugu Dist., New Taipei City 24888, Taiwan

Table 2: Differences between the Predicate and Proposed Device
CharacteristicPredicate deviceTD-4277Proposed deviceTD-4183
Physical AppearanceImage: Predicate device TD-4277Image: Proposed device TD-4183
Meter Size (mm)96 (L) x 61 (W) x 26 (H)90.3 (L) x 52.3 (W) x 18 (H)
Meter Weight (g)67.2 g46.8 g
Measurement range20-600 mg/dL20-650 mg/dL
Hematocrit range20-60%20-65%
Power source2 x 1.5V AAA batteries1 x 1.5V AAA batteries
Transmission functionUSBNA
Meter storage/transportationcondition-4°F to 140°F, < 95% R.H.-4°F to 140°F, 10% to 93% R.H.
Strip storage/transportationcondition35.6°F to 89.6°F, < 85% R.H.35.6°F to 86°F, 10% to 85% R.H.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the word "TaiDoc" in gray font with three red hearts underneath. The hearts are evenly spaced and aligned horizontally. The word "TaiDoc" is in a sans-serif font and is larger than the hearts. The overall design is simple and clean.

www.taidoc.com

泰博科技股份有限公司 TaiDoc Technology Corp.

新北市24888五股區五工二路127號6樓 6F., No.127, Wugong 2nd Rd., Wugu Dist., New Taipei City 24888, Taiwan

CharacteristicPredicate deviceTD-4277Proposed deviceTD-4183
Intended useTD-4277 Blood Glucose MonitoringSystem is intended for use in thequantitative measurement of glucosein fresh capillary whole blood fromthe finger.It is intended for use by healthcareprofessionals and people withdiabetes mellitus at home as an aid inmonitoring the effectiveness ofdiabetes control program.It is not intended for the diagnosis ofor screening for diabetes mellitus,and is not intended for use onneonates.Professionals may use the test stripsto test capillary and venous bloodsamples, but lay user may not testvenous blood samples.The TD-4183 Blood GlucoseMonitoring System consists of theTD-4183 Blood Glucose meter andthe TD-4183 Blood Glucose TestStrips.The TD-4183 Blood GlucoseMonitoring System is intended foruse in the quantitative measurementof glucose in fresh capillary wholeblood from the fingertip. This systemis intended for self-testing outside thebody (in vitro diagnostic use) bypeople with diabetes mellitus at homeas an aid in monitoring theeffectiveness of the diabetes controlprogram. It is intended to be used bysingle person and should not beshared. It is not intended for thediagnosis of or screening for diabetesmellitus, and is not intended for useon neonates.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for TaiDoc. The word "TaiDoc" is written in gray at the top of the image. Below the word are three red hearts. At the bottom of the image is the website address, www.taidoc.com.

Summary of Testing

Non-clinical and clinical studies were conducted to test, verify and validate the performance of the proposed device according to FDA Guidance issued on October 11, 2016: Self-Monitoring Blood Glucose Test Systems for Over-the-Counter Use. Results from these studies show that all performance criteria were met.

Non-Clinical Testing Summary: Design verification and validation testing was performed to ensure that the TD-4183 System met design specifications and requirements. Testing activities included electrical/mechanical safety tests, functional performance tests (precision, linearity, interference, flex studies) as well as disinfection, cleaning, and robustness studies. Software validation was performed for this moderate level of concern device per FDA Guidance Content of Premarket Submissions for Software Contained in Medical Devices.

Clinical Testing Summary: A user evaluation confirmed the system accuracy, operation according to design, and ease of use to support the intended use as described in the proposed labeling.

Conclusion

Based on the information provided in this submission, the TD-4183 Blood Glucose Monitoring System has been shown to be substantially equivalent to the TD-4277 Blood Glucose Monitoring System (K100322).

§ 862.1345 Glucose test system.

(a)
Identification. A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.(b)
Classification. Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.